Skip to main content
. Author manuscript; available in PMC: 2010 Mar 11.
Published in final edited form as: Trends Amplif. 2008 Dec;12(4):300–315. doi: 10.1177/1084713808325880

Table 2.

Average and standard deviations for measured real ear insertion gain (REIG), target REIG calculated using the NAL-RP prescriptive method, and the deviation from target (measured REIG – target REIG) for right and left ears.

Frequency Right ear Left ear
Measured
REIG
Target
REIG
Difference Measured
REIG
Target
REIG
Difference
250 Hz 5.8 (6.2) 3.2 (3.9) 2.6 (5.7) 6.7 (9.2) 3.6 (4.2) 3.0 (8.2)
500 Hz 11.3 (9.3) 14.2 (5.0) −2.9 (8.8) 11.2 (10.8) 14.2 (5.7) −3.0 (8.6)
1000 Hz 21.6 (10.1) 26.1 (5.0) −4.5 (8.7) 20.6 (12.0) 26.1 (5.2) −5.5 (10.2)
2000 Hz 23.4 (9.7) 25.6 (4.8) −2.2 (8.5) 22.9 (10.4) 25.4 (4.3) −2.5 (9.4)
4000 Hz 16.1 (9.0) 25.9 (4.9) −9.8 (9.1) 15.6 (7.7) 25.6 (4.8) −10.0 (6.6)
6000 Hz 8.2 (11.8) 26.3 (5.7) −18.1 (14.0) 6.3 (14.6) 26.7 (5.9) −20.5 (13.5)