Skip to main content
. Author manuscript; available in PMC: 2018 Jul 30.
Published in final edited form as: Med Phys. 2018 May 28;45(7):3063–3075. doi: 10.1002/mp.12957

Table 7.

QT reflection and contrast to noise (CNR) measurements as a function of object size.

Object size (mm) Mean ROI intensity (RU) Reflection CNR mean ± s.d. Reflection CNR mean ± s.d. (dB)
1 1359800 2261.9 ± 350.4 33.5 ± 0.6
0.8 1111377 1825.1 ± 396.0 32.6 ± 0.9
0.55 1085997 1780.5 ± 218.1 32.5 ± 0.5
0.3 801039 1279.4 ± 149.6 31.1 ± 0.5
0.2 258373 325.1 ± 27.3 25.1 ± 0.4
0.1 189697 204.3 ± 11.6 23.1 ± 0.2