Published in final edited form as: Med Phys. 2018 Sep 14;45(10):4652–4659. doi: 10.1002/mp.13138

Table 2.

Dose rate and imaging dose metrics for XFCT and transmission CT configurations

XFCT Transmission CT
Incident X-ray Spectrum 125 kVp, 1.8-mm Sn 62 kVp, 0.4-mm Al
Dose Rate at Isocenter (cGy/min) 24.9 @ 24 mA 40.2 @ 1.0 mA
Acquisition Time per Projection (s) 10 0.05
Dose per Projection (cGy) 1.87 (effective) 0.0335
Total Number of Projections 330 360
Total Imaging Dose (cGy) 617 (effective) 12.1