Table II.
Image Acquisition Parameters | 2D Imaging | 3D Imaging |
---|---|---|
Tube Voltage (Nominal) | 100 kV + 0.1 mmCu + 3 mmAl | |
Additional Attenuation | 4.5 mm Cu | 1.7 mm Cu + 16 cm Acrylic Cylinder |
HVL | 6.1 mmAl | |
Fluence/Kerma (Φ/KA) | 2.89 ×105 x-rays/μGy/mm2 | |
X-ray Tube Output | 0.015 – 0.235 mAs | |
Frame Rate (Nominal) | 15 fps | |
SDD | 110 cm | |
Tube Output (@SDD) | 0.34 μGy/mAs | 0.25 μGy/mAs/view |
Projections/ Arc | - | 280-560 proj/ 360° |
CTDIw (16 cm acrylic) | - | 0.01-0.02 mGy/mAs |