Published in final edited form as: Med Phys. 2018 Nov 20;46(1):5–14. doi: 10.1002/mp.13246

Figure B1:

Figure B1:

Automated 2D-3D and manual 2D-2D Gamma failed pixel distribution and passing rate comparison for a prescription depth of 4.5cm. (a) Automated 3%, 3mm γ-index analysis with passing rate of 98.6%. (b) Manual 3%, 3mm γ-index analysis with passing rate of 98.3%. (c) Automated 2%, 2mm γ-index analysis with passing rate of 97.0%. (d) Manual 2%, 2mm γ-index analysis with passing rate of 95.8%.