Abstract
Purpose:
To describe spectral-domain optical coherence tomography (SD-OCT) and adaptive optics (AO) imaging in hydroxychloroquine retinal toxicity.
Methods:
Two patients with long-term hydroxychloroquine use, subtle perifoveal ophthalmoscopic pigmentary changes, and bilateral perifoveal defects on automated Humphrey visual field (HVF) 10-2 perimetry were imaged using SD-OCT and AO.
Results:
SD-OCT images demonstrated loss of photoreceptor inner segment/outer segment (IS/OS) junction and a downward “sink-hole” displacement of inner retinal structures in areas of hydroxychloroquine toxicity corresponding to HVF 10-2 defects and ophthalmoscopic clinical examination findings. SD-OCT irregularities in the IS/OS junction were also seen in areas not detected on HVF 10-2. AO images showed disruption of the cone photoreceptor mosaic in areas corresponding to HVF 10-2 defects and SD-OCT IS/OS junction abnormalities. Additionally, irregularities in the cone photoreceptor density and mosaic were seen in areas with normal HVF 10-2 and SD-OCT findings.
Conclusions:
SD-OCT and AO detected abnormalities that correlate topographically with visual field loss from hydroxychloroquine toxicity as demonstrated by HVF 10-2 and may be useful in the detection of subclinical abnormalities that precede symptoms or objective visual field loss.
INTRODUCTION
Hydroxychloroquine sulfate (Plaquenil; Sanofi-Aventis, Bridgewater, New Jersey) is an antimalarial drug used extensively in treatment of autoimmune diseases such as systemic lupus erythematosus and rheumatoid arthritis. Retinal toxicity that may be irreversible1,2 can develop in some individuals who take hydroxychloroquine, especially when higher-risk criteria are present. Even after the drug is discontinued, retinal degeneration from hydroxychloroquine can continue to progress. For this reason, ophthalmic screening of patients is recommended to detect early retinopathy and discontinue the therapy.
Through the years, multiple modalities have been advocated for screening of hydroxychloroquine retinopathy.3–8 In response to a diversity of recommended screening regimens, the American Academy of Ophthalmology published preferred practice patterns (PPPs) for hydroxychloroquine retinopathy screening in 2002.9 These PPPs are based on stratifying risk of retinopathy in a patient based on multiple criteria. A baseline examination is recommended at the beginning of treatment. In low-risk patients in the first 5 years of treatment, screening should be done only as a component of normal ophthalmic examinations. High-risk patients should be screened annually. High-risk criteria consist of any of the following: (1) daily dose exceeding 6.5 mg/kg, (2) greater than 5 years duration of treatment, (3) high body mass index, (4) age 60 years or older, (5) renal or hepatic disease that may impair drug metabolism or excretion, and (6) concurrent retinal disease.10 Screening examinations for hydroxychloroquine retinopathy as recommended by the PPPs should include a comprehensive ophthalmic examination with either Amsler grid testing or Humphrey visual field (HVF) 10-2 perimetry (Carl Zeiss Meditec Inc, Dublin, California) testing. Other optional testing, including color vision testing, fundus photography, fluorescein angiography, electroretinography, and multifocal electroretinography (mERG), is left to the discretion of the screening clinician.10
Ideally, screening for retinopathy would be quick, noninvasive, and very sensitive and specific for detecting early hydroxychloroquine toxicity. New innovations in ocular imaging now allow for unprecedented views of retinal architecture to within microns. Recent advances in spectral-domain optical coherence tomography (SD-OCT) allow for cross-sectional imaging of the retina in vivo to resolutions of 5 to 10 μm.11–16 Adaptive optics (AO) imaging corrects for ocular aberrations and allows for direct visualization of the photoreceptor mosaic in vivo to resolutions of ≤2 μm.17–20 This report presents the results of SD-OCT and AO imaging on two patients with known hydroxychloroquine retinopathy.
METHODS
Two female patients with known hydroxychloroquine retinopathy were imaged using SD-OCT and AO imaging. Both patients were referred to the retina service for abnormal HVF 10-2 perimetry findings on hydroxychloroquine screening eye examinations. Comprehensive ophthalmic examinations, fundus photography, and mERG were performed, and both patients were diagnosed as having hydroxychloroquine retinopathy. Prior to SD-OCT and AO imaging, both patients provided informed consent. All research adhered to the tenets of the Declaration of Helsinki, and the study protocol was approved by the Children’s Hospital of Wisconsin Institutional Review Board.
Patient 1 was a 57-year-old woman who had been taking hydroxychloroquine, 400 mg daily, for 20 years at a dose of 6.15 mg/kg per day for systemic lupus erythematosus. She was asymptomatic and denied having red Amsler grid changes. Visual acuity was 20/20 in both eyes. On examination she had subtle perifoveal pigmentary changes in both eyes inferonasal to the fovea. HVF 10-2 showed a paracentral scotoma most apparent superotemporally in both eyes.
Patient 2 was a 61-year-old woman who had been taking hydroxychloroquine, 400 mg daily, for 16 years at a dose of approximately 8 mg/kg per day for Sjögren syndrome. She recently had lost about 10 pounds. Over the last year she had noted more difficulties with seeing. Visual acuity was 20/20 in both eyes. Clinical examination was remarkable for a distinct rim of depigmentation inferiorly in the perifoveal region of both eyes and other granular changes in the paracentral region. HVF 10-2 showed distinct paracentral scotomas most dense superiorly in both eyes.
Spectral domain imaging was performed using both the Spectralis SD-OCT (Heidelberg Engineering, Vista, California) and a Bioptigen SD-OCT (Bioptigen Inc, Durham, North Carolina). For Bioptigen SD-OCT, images were acquired and processed as follows. Each image was 6 or 7 mm in length, and B-scans were composed of 1,000 A-scans. One hundred B-scans were obtained in each scan set. Scan sets were then directly exported from the SD-OCT machine and read into ImageJ (http://rsb.info.nih.gov/ij/) for processing. Frames that were distorted on account of large saccades or eye blinks were removed. A rigid body registration using the ImageJ plug-in “StackReg” was applied to generate a stabilized frame sequence for subsequent averaging, and the scan length was calibrated by correcting for the patient’s axial length as measured by a Zeiss IOLMaster (Carl Zeiss Meditec Inc, Dublin, California).
Images of the photoreceptor mosaic were obtained using a newly developed AO ophthalmoscope housed at the Medical College of Wisconsin, Milwaukee. Each patient’s head was stabilized using a dental impression on a bite bar, and both eyes were dilated and cyclopleged using a combination of phenylephrine hydrochloride 2.5% and tropicamide 1%. In a continuous closed-loop fashion, the eyes’ monochromatic aberrations were measured over a 6.8-mm pupil with a Shack-Hartmann wavefront sensor and corrected for by using a 52-channel deformable mirror (Imagine Eyes, Orsay, France). A fiber-coupled near infrared source was used for imaging, which consisted of a 200-mW SLD (superluminescent diode; center wavelength of 837.8 nm and 14.1-nm spectral bandwidth FWHM [full width at half maximum]) and 110 m of multimode step index fiber (Fiberguide Industries, Stirling, New Jersey) to eliminate speckle noise of images.21 The patient was then instructed to fixate on a calibrated fixation light that could be translated to image different parts of the macula. Each imaging flash was 500 milliseconds, and the retinal image sequence that was available for processing consisted of 40 frames (6 milliseconds exposure per frame). The pattern of the retinal vasculature was used to confirm the image location, and the acquired images were 1.8° × 0.9°. Individual frames from each image sequence were registered and averaged using custom Matlab software (The MathWorks, Natick, Massachusetts).
RESULTS
SD-OCT images of all four eyes revealed abnormalities in the outer retina corresponding to areas of hydroxychloroquine retinopathy as identified by clinical funduscopic findings and HVF 10-2 defects. These changes included complete loss of the photoreceptor inner segment/outer segment (IS/OS) junction with relative preservation of the retinal pigment epithelium (RPE) and external limiting membrane and downward displacement, or “sink-hole effect,” of overlying inner retina layers in perifoveal areas corresponding to HVF 10-2 defects and ophthalmoscopic clinical examination findings (Figure 1). Additionally, outer retinal changes with a “moth-eaten” appearance to the photoreceptor IS/OS layer but with preservation of inner retinal layers were also identified in areas in which HVF 10-2 appeared normal (Figure 2). Central foveal architecture was preserved in both patients.
FIGURE 1.
Patient 1. Fluorescein angiogram (FA) and Bioptigen SD-OCT images of left eye. SD-OCT images are registered onto FA fundus photo. Upper SD-OCT image is the horizontal scan, and the lower SD-OCT image is the vertical scan. Numbers 1, 2, and 3 on the fundus photo represent locations where adaptive optics imaging was done. These same locations are also registered on the SD-OCT images. SD-OCT images show loss of photoreceptor inner segment/outer segment junctions and a “sink-hole” effect of the inner retinal structures in the perifoveal area.
FIGURE 2.
Patient 1. Spectralis SD-OCT of left eye. Area of “moth-eaten” photoreceptor inner segment/outer segment abnormalities can be seen between the arrows in an area that appeared normal on funduscopic examination and Humphrey visual field (HVF) 10-2.
AO images were taken at precise locations in the perifoveal area and registered to fundus images and SD-OCT images. Patient 1 had both eyes imaged. Patient 2 had the right eye imaged.
In patient 1, AO images were taken in the right eye 2° nasal to the fovea in an area that corresponded to the margin where outer segment photoreceptor IS/OS abnormalities were identified on SD-OCT. AO images confirmed there was a dramatic disruption of the cone mosaic in this location.
In the left eye of patient 1, three different areas were imaged by AO (Figures 1 and 3). Location 1 in Figure 1 is 1° nasal to the fovea. Photoreceptor spacing at this location is 0.881 ± 0.12 minutes of an arc (arc min) (mean ± 1 SD), which is normal18 (Figure 3A). Corresponding SD-OCT photoreceptor IS/OS findings also appear normal at this location (Figure 1).
FIGURE 3.
Adaptive optics (AO) images taken at locations 1, 2, and 3 as shown in Figure 1. A, Location 1 is 1° nasal to the fovea. Normal cone mosaic and normal cone spacing are seen. B and C, Location 3 is 3° temporal to the fovea. Photoreceptor spacing is 0.727 ± 0.13 arc min (mean ± 1 SD), which is decreased. Normal cone spacing is approximately 1.5 arc min at this location.18 D, Location 2 is 0.75° superior to the fovea. Cone mosaic shows patches on contiguousness. Cone spacing is 1.10 ± 0.15 arc min (mean ± 1 SD), which is increased. Normal cone spacing is approximately 0.8 arc min at this location.18
Location 2 in Figure 1 is 0.75° superior to the fovea. Outer retinal photoreceptor IS/OS findings appear normal on SD-OCT at this location. AO images show that the cone mosaic has patches of contiguousness (Figure 3D); however, the cone spacing in this area is 1.10 ± 0.15 arc min (mean ± 1 SD). The average spacing for normal retinas at this retinal eccentricity is approximately 0.8 arc min.18 The photoreceptor spacing at this location falls more than 2 SDs from the mean of normative cone data obtained from histological samples.22: Increased photoreceptor spacing indicates less photoreceptors are present.
Location 3 in Figure 1 is 3° temporal to the fovea in an area of photoreceptor IS/OS “moth-eaten” appearance as seen on SD-OCT and borderline HVF 10-2 abnormalities. The photoreceptor spacing was 0.727 ± 0.13 arc min (mean ± 1 SD). Normal cone spacing at this location is about 1.5 arc min.18 If completely packed, this cone spacing would correspond to a cone density of about 23,000 cones/mm2, whereas a spacing of 0.727 arc min would correspond to a cell spacing of 98,000 cells/mm2 (Figure 3B and C) This increased density suggests that structures imaged are actually rod photoreceptors.
Patient 2 had AO images at a location 2° temporal to the fovea. SD-OCT images at this location show that this is right at the margin of outer retinal photoreceptor IS/OS abnormalities. AO images at this area show an irregular, disrupted, patchy cone mosaic with areas of intermittent, near normal cone spacing in some areas. Cone spacing in this area was 1.13 ± 0.15 arc min (mean ± 1 SD) with normal retinal spacing in this area being 1.2 arc min,18 indicating that spacing in this individual at this location is within the normal range.
DISCUSSION
In the two patients described here, SD-OCT and AO imaging detected abnormalities from hydroxychloroquine retinopathy that correlate topographically with clinical funduscopic findings and visual field loss as demonstrated by HVF 10-2 perimetry. SD-OCT images showed complete loss of the photoreceptor IS/OS junction with relative preservation of the RPE and external limiting membrane as previously described by Rodriguez-Padilla and colleagues23; however, our SD-OCT images were acquired on two different commercially available SD-OCT machines instead of a research prototype. Additionally, our SD-OCT images reveal a distinct downward displacement, or “sink-hole effect,” of inner retina layers in areas overlying outer retinal abnormalities. Areas of outer retinal loss were very precise and localized to areas of toxicity in perifoveal areas corresponding to HVF 10-2 defects and ophthalmoscopic clinical examination findings. Adaptive optics images from an area with hydroxychloroquine retinopathy as seen both clinically and with HVF 10-2 defects showed disruption of the normal cone mosaic and quantitative alterations in cone packing.
SD-OCT and AO imaging also exhibited abnormalities in areas that appeared unaffected on HVF 10-2 and clinical funduscopic examination. These changes may represent a potential “preclinical” state of hydroxychloroquine toxicity. The exact mechanism of hydroxychloroquine retinal toxicity is not fully understood. Animal studies have shown that the first evidence of toxicity with chloroquine is seen in the retinal ganglion cells.24 However, histology has shown that perifoveal photoreceptor cells are most severely affected25 and that this may be a secondary effect due to disruption of RPE metabolism. Chloroquine and hydroxychloroquine are known to disrupt lysosomal function of the RPE, leading to increased accumulation of lipofuscin.26,27 Daily phagocystosis of photoreceptor outer segments occurs in the lysosomal apparatus of the RPE, and alterations in RPE metabolism may play a role in retinal toxicity.26,27 Mahon and associates27 also found an apparent accumulation of autophagic granules in cone photoreceptor cells exposed to chloroquine in an animal model. Rod photoreceptors did not show these granules, and the investigators hypothesize that this defective degradative capacity that appears more predominant in cone photoreceptors could also play a role in toxicity. Given this information, it makes sense that early toxicity would present with subtle changes in the photoreceptor mosaic, and that this would precede the dramatic loss of outer retinal structures seen with more severe hydroxychloroquine retinopathy. Evidence of a defective degradative capacity in cones may also suggest that they are preferentially affected more than rod photoreceptors, and this may be apparent even in early stages of toxicity.
In areas that appeared unaffected by funduscopic examination and with no visual field defects as seen on automated perimetry, our SD-OCT imaging showed areas of a characteristic “moth-eaten” appearance in the photoreceptor IS/OS layer with preservation of inner retina architecture (eg, Figure 1, location 3). AO imaging done at this same location (Figure 3B and C) showed a decrease in photoreceptor spacing, indicating an increased density of photoreceptor cells of approximately 98,000 cells/mm2. Normal cone density in this location is approximately 23,000 cones/mm2, making it more likely that most photoreceptors imaged at this location are rod photoreceptors instead of the cone photoreceptors. Densities calculated from AO imaging are similar to histological samples of normative rod photoreceptor data at this retinal eccentricity.22 This suggests that cone photoreceptors are missing or significantly diminished in number at this location and also that cone photoreceptors may be more susceptible to plaquenil toxicity than rod photoreceptors. This decrease of cone photoreceptors may be responsible for the “moth-eaten” photoreceptor IS/OS appearance seen on SD-OCT. These areas likely represent a “preclinical” stage of hydroxychloroquine retinopathy that with time will develop a severe loss of outer retinal structures and become apparent on clinical examination and visual field testing.
Adaptive optics imaging may also be more sensitive than SD-OCT in identifying these “preclinical” areas of retinal toxicity. In AO images taken at location 2 in the left eye of patient 1, the cone mosaic had patches of contiguousness, but cone spacing was increased to 1.10 ± 0.15 arc min (mean ± 1 SD). Average cone spacing at this retina eccentricity is approximately 0.8 arc min.18 Increased cone spacing indicates a reduced density of photoreceptors at this location. The photoreceptor spacing at this location also falls more than 2 SDs from the mean of normative cone data obtained from histological samples.22 This suggests that the structures imaged are morphologically compromised swollen cones or that the overall number of cone photoreceptors in this location is decreased. This may represent the first signs of hydroxychloroquine toxicity with cone photoreceptor dropout. SD-OCT images were normal at this location.
Current screening methods for hydroxychloroquine retinopathy are limited in that they detect toxicity only after retinal damage has occurred as seen on funduscopic abnormalities and visual field defects. Furthermore, funduscopic changes can be subtle and tend to be a late finding, Amsler grid testing is highly subjective, and automated perimetry can show a steep learning curve. Recently mERG has been shown to be abnormal in hydroxychloroquine retinopathy, and it may be able to detect subtle changes in earlier stages of toxicity.28–30 However, mERG is limited by clinical availability, patient cooperation, specialized training for administration and interpretation, and cost.
This study has shown that both SD-OCT and AO imaging are able to demonstrate abnormalities in the outer retina in hydroxychloroquine retinopathy. Furthermore, both SD-OCT and AO may be able to detect abnormalities at an early, “preclinical” stage of hydroxychloroquine toxicity. Obvious limitations of the study include its small sample size and its retrospective nature that allows one to look for abnormalities that may or may not be attributed to hydroxychloroquine. That being said, the noninvasive, amazing resolution and quantitative aspects of SD-OCT and AO imaging make them attractive as possible screening options for hydroxychloroquine toxicity. SD-OCT machines are becoming more common in clinical settings, and AO imaging development will likely result in a clinical model soon. These modalities may prove to be useful and effective tools for hydroxychloroquine retinal toxicity screening. Further research is needed.
ACKNOWLEDGMENTS
Funding/Support: Support was provided by grants EY001931 and EY017607 from the National Institutes of Health, Bethesda, Maryland; an unrestricted departmental grant from Research to Prevent Blindness Inc, New York, New York; the Marjorie Christiansen Foundation, Milwaukee, Wisconsin; Fight for Sight, New York, New York; the E. Matilda Ziegler Foundation for the Blind, Darien, Connecticut; the R. D. and Linda Peters Foundation, Brillion, Wisconsin; and the Gene and Ruth Posner Foundation, Milwaukee, Wisconsin. Dr Carroll is the recipient of a Career Development Award from Research to Prevent Blindness.
Financial Disclosures: Dr Stepien, Dr Han, and Dr Schell have none. Work of Dr Carroll, Dr Rha, and Dr Godara receives salary support from NIH grants EY001931 and EY017607.
Author Contributions: Design and conduct of the study (K.S., J.C.); Collection and management of data (K.S., D.H., J.C., J.R., P.G., J.S.); Analysis of data (K.S., J.C., P.G., J.S.); Preparation and review of manuscript (K.S., D.H., J.C., J.R., P.G., J.S.).
Conformity With Author Information: All research adhered to the tenets of the Declaration of Helsinki. The study protocol was approved by the Children’s Hospital of Wisconsin Institutional Review Board.
Other Acknowledgments: We would like to thank Dr Alfredo Dubra at the University of Rochester, Rochester, New York, for access to cone mosaic registration software.
REFERENCES
- 1.Mavrikakis I, Sfikakis PP, Mavrikakis E, et al. The incidence of irreversible retinal toxicity in patients treated with hydroxychloroquine: a reappraisal. Ophthalmology. 2003;110:1321–1326. doi: 10.1016/S0161-6420(03)00409-3. [DOI] [PubMed] [Google Scholar]
- 2.Bienfang D, Coblyn JS, Liang MH, Corzillius M. Hydroxychloroquine retinopathy despite regular ophthalmologic evaluation: a consecutive series. J Rheumatol. 2000;27:2703–2706. [PubMed] [Google Scholar]
- 3.Lai TY, Chan WM, Li H, Lai RY, Lam DS. Multifocal electroretinographic changes in patients receiving hydroxychloroquine therapy. Am J Ophthalmol. 2005;140:794–807. doi: 10.1016/j.ajo.2005.05.046. [DOI] [PubMed] [Google Scholar]
- 4.Lai TY, Ngai JW, Chan WM, Lam DS. Visual field and multifocal electroretinography and their correlations in patients on hydroxychloroquine therapy. Doc Ophthalmol. 2006;112:177–187. doi: 10.1007/s10633-006-9006-0. [DOI] [PubMed] [Google Scholar]
- 5.Levy GD, Munz SJ, Paschal J, Cohen HB, Pince KJ, Peterson T. Incidence of hydroxychloroquine retinopathy in 1,207 patients in a large multicenter outpatient practice. Arthritis Rheum. 1997;40:1482–1486. doi: 10.1002/art.1780400817. [DOI] [PubMed] [Google Scholar]
- 6.Marmor MF. The dilemma of hydroxychloroquine screening: new information from the multifocal ERG. Am J Ophthalmol. 2005;140:894–895. doi: 10.1016/j.ajo.2005.06.052. [DOI] [PubMed] [Google Scholar]
- 7.Ruiz RS, Saatci OA. Chloroquine and hydroxychloroquine retinopathy: how to follow affected patients. Ann Ophthalmol. 1991;23:290–291. [PubMed] [Google Scholar]
- 8.Bernstein HN. Ocular safety of hydroxychloroquine. Ann Ophthalmol. 1991;23:292–296. [PubMed] [Google Scholar]
- 9.Marmor MF, Carr RE, Easterbrook M, Farjo AA, Mieler WF. Recommendations on screening for chloroquine and hydroxychloroquine retinopathy: a report by the American Academy of Ophthalmology. Ophthalmology. 2002;109:1377–1382. doi: 10.1016/s0161-6420(02)01168-5. [DOI] [PubMed] [Google Scholar]
- 10.Marmor MF. New American Academy of Ophthalmology recommendations on screening for hydroxychloroquine retinopathy. Arthritis Rheum. 2003;48:1764. doi: 10.1002/art.10980. [DOI] [PubMed] [Google Scholar]
- 11.Potsaid B, Gorczynska I, Srinivasan VJ, et al. Ultrahigh speed spectral/Fourier domain OCT ophthalmic imaging at 70,000 to 312,500 axial scans per second. Opt Express. 2008;16:15149–15169. doi: 10.1364/oe.16.015149. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Srinivasan VJ, Wojtkowski M, Witkin AJ, et al. High-definition and 3-dimensional imaging of macular pathologies with high-speed ultrahigh-resolution optical coherence tomography. Ophthalmology. 2006;113:2054.e1–14. doi: 10.1016/j.ophtha.2006.05.046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Cense B, Chen TC, Nassif N, et al. Ultra-high speed and ultra-high resolution spectral-domain optical coherence tomography and optical Doppler tomography in ophthlamology. Bull Soc Belge Ophtalmol. 2006:123–132. [PubMed] [Google Scholar]
- 14.Nassif N, Cense B, Park BH, et al. In vivo human retinal imaging by ultrahigh-speed spectral domain optical coherence tomography. Opt Lett. 2004;29:480–482. doi: 10.1364/ol.29.000480. [DOI] [PubMed] [Google Scholar]
- 15.Schmidt-Erfurth U, Leitgeb RA, Michels S, et al. Three-dimensional ultrahigh-resolution optical coherence tomography of macular diseases. Invest Ophthalmol Vis Sci. 2005;46:3393–3402. doi: 10.1167/iovs.05-0370. [DOI] [PubMed] [Google Scholar]
- 16.Wojtkowski M, Srinivasan V, Fujimoto JG, et al. Three-dimensional retinal imaging with high-speed ultrahigh-resolution optical coherence tomography. Ophthalmology. 2005;112:1734–1746. doi: 10.1016/j.ophtha.2005.05.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Carroll J, Neitz M, Hofer H, Neitz J, Williams DR. Functional photoreceptor loss revealed with adaptive optics: an alternate cause of color blindness. Proc Natl Acad Sci U S A. 2004;101:8461–8466. doi: 10.1073/pnas.0401440101. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Duncan JL, Zhang Y, Gandhi J, et al. High-resolution imaging with adaptive optics in patients with inherited retinal degeneration. Invest Ophthalmol Vis Sci. 2007;48:3283–3291. doi: 10.1167/iovs.06-1422. [DOI] [PubMed] [Google Scholar]
- 19.Wolfing JI, Chung M, Carroll J, Roorda A, Williams DR. High-resolution retinal imaging of cone-rod dystrophy. Ophthalmology. 2006;113:1014–1019.e1. doi: 10.1016/j.ophtha.2006.01.056. [DOI] [PubMed] [Google Scholar]
- 20.Roorda A, Williams DR. The arrangement of the three cone classes in the living human eye. Nature. 1999;397:520–522. doi: 10.1038/17383. [DOI] [PubMed] [Google Scholar]
- 21.Rha J, Jonnal R, Thorn KE, Qu J, Zhang Y, Miller DT. Adaptive optics flood-illumination camera for high speed retinal imaging. Opt Express. 2006;14:4552–4569. doi: 10.1364/oe.14.004552. [DOI] [PubMed] [Google Scholar]
- 22.Curcio CA, Sloan KR, Kalina RE, Hendrickson AE. Human photoreceptor topography. J Comp Neurol. 1990;292:497–523. doi: 10.1002/cne.902920402. [DOI] [PubMed] [Google Scholar]
- 23.Rodriguez-Padilla JA, Hedges TR, 3rd, Monson B, et al. High-speed ultra-high-resolution optical coherence tomography findings in hydroxychloroquine retinopathy. Arch Ophthalmol. 2007;125:775–780. doi: 10.1001/archopht.125.6.775. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Rosenthal AR, Kolb H, Bergsma D, Huxsoll D, Hopkins JL. Chloroquine retinopathy in the rhesus monkey. Invest Ophthalmol Vis Sci. 1978;17:1158–1175. [PubMed] [Google Scholar]
- 25.Bernstein HN, Ginsberg J. The pathology of chloroquine retinopathy. Arch Ophthalmol. 1964;71:238–245. doi: 10.1001/archopht.1964.00970010254019. [DOI] [PubMed] [Google Scholar]
- 26.Sundelin SP, Terman A. Different effects of chloroquine and hydroxychloroquine on lysosomal function in cultured retinal pigment epithelial cells. APMIS. 2002;110:481–489. doi: 10.1034/j.1600-0463.2002.100606.x. [DOI] [PubMed] [Google Scholar]
- 27.Mahon GJ, Anderson HR, Gardiner TA, McFarlane S, Archer DB, Stitt AW. Chloroquine causes lysosomal dysfunction in neural retina and RPE: implications for retinopathy. Curr Eye Res. 2004;28:277–284. doi: 10.1076/ceyr.28.4.277.27835. [DOI] [PubMed] [Google Scholar]
- 28.Lyons JS, Severns ML. Detection of early hydroxychloroquine retinal toxicity enhanced by ring ratio analysis of multifocal electroretinography. Am J Ophthalmol. 2007;143:801–809. doi: 10.1016/j.ajo.2006.12.042. [DOI] [PubMed] [Google Scholar]
- 29.Lyons JS, Severns ML. Using multifocal ERG ring ratios to detect and follow Plaquenil retinal toxicity: a review. Review of mfERG ring ratios in Plaquenil toxicity. Doc Ophthalmol. 2009;118:29–36. doi: 10.1007/s10633-008-9130-0. [DOI] [PubMed] [Google Scholar]
- 30.Chang WH, Katz BJ, Warner JE, Vitale AT, Creel D, Digre KB. A novel method for screening the multifocal electroretonogram in patients using hydroxychloroquine. Retina. 2008;28:1478–1486. doi: 10.1097/IAE.0b013e318181445b. [DOI] [PubMed] [Google Scholar]