Abstract
Automated refraction with the Canon RK-1 Autoref keratometer was evaluated in 110 eyes (110 patients) six to eight weeks after they had undergone extracapsular cataract extraction with posterior chamber intraocular lens implantation and achieved a best corrected visual acuity of at least 6/12. Autorefraction readings were obtained in 100 (91%) of these eyes. The agreement between autorefraction and clinical refraction data was 98% for spherical equivalence less than 0.51 dioptres (D), 95% for sphere power less than 0.51D, 94% for cylinder power less than 0.51D, and 85% for cylinder axis less than 11 degrees. Autorefraction can provide acceptably accurate postoperative refraction values in pseudophakic eyes.
Full text
PDF


Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Dang M. S., Raj P. P. SRK II formula in the calculation of intraocular lens power. Br J Ophthalmol. 1989 Oct;73(10):823–826. doi: 10.1136/bjo.73.10.823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dufier J. L., Abitbol M., Pigamo F., Prete T., Paris J. P., Poitrenaud O. Evaluation des performances d'un réfractomètre objectif automatique sur 3,618 yeux. J Fr Ophtalmol. 1987;10(4):301–308. [PubMed] [Google Scholar]
- Ghose S., Nayak B. K., Singh J. P. Critical evaluation of the NR-1000F Auto Refractometer. Br J Ophthalmol. 1986 Mar;70(3):221–226. doi: 10.1136/bjo.70.3.221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guillon M. Automated refraction in aphakia--II. Its repeatability and accuracy compared to conventional techniques. Ophthalmic Physiol Opt. 1986;6(1):85–89. doi: 10.1016/0275-5408(86)90123-7. [DOI] [PubMed] [Google Scholar]
- Guyton D. L. Automated refractors-1983. Ophthalmology. 1983 Sep;Suppl:36–44. doi: 10.1016/s0161-6420(83)34441-9. [DOI] [PubMed] [Google Scholar]
- Hunold W., Auffarth G., Effert R. Klinische Anwendung des Canon Autorefraktometers R 10 zur Refraktionsbestimmung bei Schielkindern. Klin Monbl Augenheilkd. 1988 Jan;192(1):58–65. doi: 10.1055/s-2008-1050075. [DOI] [PubMed] [Google Scholar]
- Infra-red autorefractors--here to stay? Ophthalmic Physiol Opt. 1985;5(3):237–239. [PubMed] [Google Scholar]
- Klein B. E., Davis M. D., Segal P., Long J. A., Harris W. A., Haug G. A., Magli Y. L., Syrjala S. Diabetic retinopathy. Assessment of severity and progression. Ophthalmology. 1984 Jan;91(1):10–17. doi: 10.1016/s0161-6420(84)34374-3. [DOI] [PubMed] [Google Scholar]
- Reddy N. S., Agarwal S. Clinical evaluation of Topcon auto refractometer. Indian J Ophthalmol. 1987;35(5-6):407–410. [PubMed] [Google Scholar]
- Wesemann W., Rassow B. Automatic infrared refractors--a comparative study. Am J Optom Physiol Opt. 1987 Aug;64(8):627–638. doi: 10.1097/00006324-198708000-00011. [DOI] [PubMed] [Google Scholar]
- Wong E. K., Jr, Patella V. M., Pratt M. V., Myers S. W., Gaster R. N., Leopold I. H. Clinical evaluation of the Humphrey automatic refractor. Arch Ophthalmol. 1984 Jun;102(6):870–875. doi: 10.1001/archopht.1984.01040030690020. [DOI] [PubMed] [Google Scholar]
- Wood I. C. A review of autorefractors. Eye (Lond) 1987;1(Pt 4):529–535. doi: 10.1038/eye.1987.80. [DOI] [PubMed] [Google Scholar]
- Wood I. C., Papas E., Burghardt D., Hardwick G. A clinical evaluation of the Nidek Autorefractor. Ophthalmic Physiol Opt. 1984;4(2):169–178. [PubMed] [Google Scholar]

