Analysis

The Accuracy and Efficacy of the Dexcom G4 Platinum Continuous Glucose Monitoring System

Cornelis A J van Beers 1,✉, J H DeVries 2
PMCID: PMC4667343  PMID: 25917336

Abstract

In this issue of Journal of Diabetes Science and Technology, Nakamura and Balo report on accuracy and efficacy of the Dexcom G4 Platinum Continuous Glucose Monitoring System. The authors demonstrate good overall performance of this real-time continuous glucose monitoring (RT-CGM) system, although accuracy data of the next generation RT-CGM system, the G4AP, is already available. Also, now that MARDs seem to move to single-digit numbers, the question comes up how low we need to go with accuracy. Results of the study also showed a reduction in time spent in hypoglycemia, although the clinical relevance should be questioned. To date, few trials have demonstrated a reduction of severe hypoglycemia. Conventional RT-CGM, without threshold suspension or closing the loop, might be insufficient in preventing severe hypoglycemia.

Keywords: continuous glucose monitoring, continuous glucose monitoring accuracy, Dexcom G4 Platinum, diabetes mellitus


Real-time continuous glucose monitoring (RT-CGM) assists in optimizing glycemic control.1 RT-CGM systems assess interstitial glucose levels and provide glucose values every 5-10 minutes with trends and alarms. Randomized controlled trials investigating the effect of RT-CGM on glycemic control show that RT-CGM lowers HbA1c levels, especially in compliant, suboptimally controlled type 1 diabetes mellitus (T1DM) patients.2-6 There is also increasing evidence that RT-CGM reduces mild to moderate hypoglycemia in well-controlled patients7,8 and severe hypoglycemia in patients with impaired awareness of hypoglycemia.9,10 To date, the effect of RT-CGM on quality of life is not convincing,11 although RT-CGM might increase treatment satisfaction,6,12 decrease diabetes-related physiological distress,6 and decrease fear of hypoglycemia.13 A recent qualitative study of Pickup and colleagues showed that sensor accuracy and reliability, (false) alarms, and other technical issues remain barriers of RT-CGM.14 Theoretically, improving these barriers will enhance adherence to RT-CGM, which might result in better glycemic outcomes and quality of life.5 Therefore, improving the technical capabilities of RT-CGM systems remains essential.

In this issue of Journal of Diabetes Science and Technology, Nakamura and Balo present a prospective, open-labeled, nonrandomized study evaluating the accuracy and efficacy of the fourth generation RT-CGM system, the Dexcom G4 Platinum (DG4P) system. This system is replaces the Dexcom SEVEN PLUS CGM system and received the CE mark and FDA approval in 2012. The DG4P system uses an advanced version of the glucose oxidase-based sensor which can be used for up to 7 days and must be calibrated twice-daily. Like its predecessor, the DG4P system displays “actual” glucose values every 5 minutes.

Seventy-two T1DM and T2DM patients were enrolled to participate in a DG4P system session for up to 7 days. Thirty-six participants wore 2 sensors simultaneously for the purpose of analyzing sensor precision. All participants were invited to participate in 3 in-clinic sessions, on days 1, 4, and 7, for the purpose of obtaining venous blood samples for YSI blood glucose determination. During home use, participants were asked to perform at least 7 finger pricks per day.

Adequate accuracy assessment methods were used to evaluate the accuracy of the DG4P system.15 Mean absolute relative difference (MARD) was 13% when calibrated with YSI. Clarke error grid (CEG) analysis showed 81% of values falling within CEG zone A and 98% of values falling in zone A+B. Clinical accuracy was analyzed using continuous glucose error grid analysis, although this method has been criticized for being too forgiving.16 Of the values, 80% were accurate within the hypoglycemia range, 98% of values were accurate within the euglycemia range and 96% of values were accurate within hyperglycemia range. Accuracy assessments of the DG4P system were quite similar when compared with either YSI or SMBG values. The authors also demonstrated that the DG4P system reached a high level of reliability, with 94% of sensors lasting 7 days and good accuracy on the last day. A longer life span of sensor could also, in theory, improve adherence to CGM. These results are in line with other studies evaluating the accuracy of the DG4P system17-19 and show that the DG4P system is among the best of commercially available CGM systems. However, by the time accuracy and reliability data get reported, the next generation CGM system is often already launched.20

Although accuracy of CGM systems is increasing, it will likely never become perfect because of the compartment problem (assessing glucose values in the interstitium). However, how low do we need to go with accuracy? To date, CGM systems are not approved to be used as a stand-alone device. Patients still need to perform finger pricks to confirm hypo- and hyperglycemic episodes and make insulin dosing decisions. Being able to use sensor values in diabetes self-management will possibly increase treatment satisfaction with RT-CGM. Recently, Kovatchev and colleagues were the first to evaluate, in silico, what level of accuracy is needed for using RT-CGM values in diabetes self-management and demonstrated that further improvement of MARD below 10% did not contribute to better glycemic outcomes.21 More studies are needed to verify these findings. On the other hand, the FDA approval of threshold suspension is in fact a judgment that confirmatory finger pricks are not always needed to make a treatment decision based on CGM glucose values, which thereby are deemed accurate enough.

Furthermore, it was suggested that patient satisfaction with RT-CGM accuracy is 1 of the key determinants of quality-of-life-related benefits.22 Therefore, improving perceived accuracy of RT-CGM systems is an important goal. Perceived accuracy can be improved by enhancing the absolute level of accuracy of RT-CGM systems, but another key element in improving satisfaction is education and expectation management. Therefore, besides improving technical capabilities of RT-CGM systems, educational programs should be developed for the purpose of improving confidence in using RT-CGM data.23

In the presented study, the authors demonstrated good diagnostic features of the DG4P system in the hyperglycemic range, with hyperglycemia detection rates and true alert rates over 90%, which could reduce another main barrier of CGM, alarm, and alert fatigue.14 However, the diagnostic feature of the DG4P system in hypoglycemic range, with approximately 1 in 5 false detection and alert rates, remains unsatisfactory and needs further improvement.

Results of the study also show a reduction in time spent in hypoglycemia during 1 sensor session. However, the clinical relevance of this outcome should be questioned. Reduction in hypoglycemia was evaluated in the very short term, without a control group and participants performed at least 7 finger pricks per day, which could also contribute to improved glycemic control.

As stated before, the effect of RT-CGM on HbA1c reduction, without increase in hypoglycemia, has been established.1-6 RT-CGM trials focusing on hypoglycemia prevention also demonstrated decrease in mild-to-moderate hypoglycemia.7,8 However, an effect on severe hypoglycemia has never been established in an randomized controlled trial evaluating the effect of conventional RT-CGM (without threshold suspension) on glycemic control.7,8,24 Only 1 observational study showed a reduction in severe hypoglycemia.9 Recently, Ly and colleagues demonstrated a reduction in severe hypoglycemia in patients with impaired awareness of hypoglycemia when using RT-CGM with threshold suspension.10 Although most trials were not powered to show differences in severe hypoglycemia, it may be suggested that conventional RT-CGM (without threshold suspension) is not enough to prevent severe hypoglycemia and that closing the loop is essential for that purpose.

Footnotes

Abbreviations: CE, Communauté Européenne; CEG, Clarke error grid; DG4P, Dexcom G4 Platinum; FDA, Food and Drug Administration; MARD, mean absolute relative difference; RT-CGM, real-time continuous glucose monitoring; SMBG, self-monitoring of blood glucose; T1DM, type 1 diabetes mellitus; T2DM, type 2 diabetes mellitus; YSI, Yellow Springs Instrument.

Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

References

  • 1. Deiss D, Bolinder J, Riveline JP, et al. Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring. Diabetes Care. 2006;29(12):2730-2732. [DOI] [PubMed] [Google Scholar]
  • 2. Battelino T, Conget I, Olsen B, et al. The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial. Diabetologia. 2012;55(12):3155-3162. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Bergenstal RM, Tamborlane WV, Ahmann A, et al. Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Engl J Med. 2010;363(4):311-320. [DOI] [PubMed] [Google Scholar]
  • 4. Raccah D, Sulmont V, Reznik Y, et al. Incremental value of continuous glucose monitoring when starting pump therapy in patients with poorly controlled type 1 diabetes: the RealTrend study. Diabetes Care. 2009;32(12):2245-2250. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Tamborlane WV, Beck RW, Bode BW, et al. Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med. 2008;359(14):1464-1476. [DOI] [PubMed] [Google Scholar]
  • 6. Hermanides J, Norgaard K, Bruttomesso D, et al. Sensor-augmented pump therapy lowers HbA(1c) in suboptimally controlled type 1 diabetes: a randomized controlled trial. Diabet Med. 2011;28(10):1158-1167. [DOI] [PubMed] [Google Scholar]
  • 7. Battelino T, Phillip M, Bratina N, Nimri R, Oskarsson P, Bolinder J. Effect of continuous glucose monitoring on hypoglycemia in type 1 diabetes. Diabetes Care. 2011;34(4):795-800. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Beck RW, Hirsch IB, Laffel L, et al. The effect of continuous glucose monitoring in well-controlled type 1 diabetes. Diabetes Care. 2009;32(8):1378-1383. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Choudhary P, Ramasamy S, Green L, et al. Real-time continuous glucose monitoring significantly reduces severe hypoglycemia in hypoglycemia-unaware patients with type 1 diabetes. Diabetes Care. 2013;36(12):4160-4162. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Ly TT, Nicholas JA, Retterath A, Lim EM, Davis EA, Jones TW. Effect of sensor-augmented insulin pump therapy and automated insulin suspension vs standard insulin pump therapy on hypoglycemia in patients with type 1 diabetes: a randomized clinical trial. JAMA. 2013;310(12):1240-1247. [DOI] [PubMed] [Google Scholar]
  • 11. Markowitz JT, Pratt K, Aggarwal J, Volkening LK, Laffel LM. Psychosocial correlates of continuous glucose monitoring use in youth and adults with type 1 diabetes and parents of youth. Diabetes Technol Ther. 2012;14(6):523-526. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Tansey M, Laffel L, Cheng J, et al. Satisfaction with continuous glucose monitoring in adults and youths with Type 1 diabetes. Diabet Med. 2011;28(9):1118-1122. [DOI] [PubMed] [Google Scholar]
  • 13. Halford J, Harris C. Determining clinical and psychological benefits and barriers with continuous glucose monitoring therapy. Diabetes Technol Ther. 2010;12(3):201-205. [DOI] [PubMed] [Google Scholar]
  • 14. Pickup JC, Ford HM, Samsi K. Real-time continuous glucose monitoring in type 1 diabetes: a qualitative framework analysis of patient narratives. Diabetes Care. 2015;38(4):544-550. [DOI] [PubMed] [Google Scholar]
  • 15. Wentholt IM, Hart AA, Hoekstra JB, DeVries JH. How to assess and compare the accuracy of continuous glucose monitors? Diabetes Technol Ther. 2008;10(2):57-68. [DOI] [PubMed] [Google Scholar]
  • 16. Wentholt IM, Hoekstra JB, DeVries JH. A critical appraisal of the continuous glucose-error grid analysis. Diabetes Care. 2006;29(8):1805-1811. [DOI] [PubMed] [Google Scholar]
  • 17. Christiansen M, Bailey T, Watkins E, et al. A new-generation continuous glucose monitoring system: improved accuracy and reliability compared with a previous-generation system. Diabetes Technol Ther. 2013;15(10):881-888. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Matuleviciene V, Joseph JI, Andelin M, et al. A clinical trial of the accuracy and treatment experience of the Dexcom G4 sensor (Dexcom G4 system) and Enlite sensor (guardian REAL-time system) tested simultaneously in ambulatory patients with type 1 diabetes. Diabetes Technol Ther. 2014;16(11):759-767. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Kropff J, Bruttomesso D, Doll W, et al. Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions. Diabetes Obes Metab. 2015;17(4):343-349. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Garcia A, Rack-Gomer AL, Bhavaraju NC, et al. Dexcom G4AP: an advanced continuous glucose monitor for the artificial pancreas. J Diabetes Sci Technol. 2013;7(6):1436-1445. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Kovatchev BP, Patek SD, Ortiz EA, Breton MD. Assessing sensor accuracy for non-adjunct use of continuous glucose monitoring. Diabetes Technol Ther. 2015;17(3):177-186. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Polonsky WH, Hessler D. What are the quality of life-related benefits and losses associated with real-time continuous glucose monitoring? A survey of current users. Diabetes Technol Ther. 2013;15(4):295-301. [DOI] [PubMed] [Google Scholar]
  • 23. Polonsky WH, Hessler D. Perceived accuracy in continuous glucose monitoring: understanding the impact on patients. J Diabetes Sci Technol. 2015;9(2):339-341. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Bergenstal RM, Klonoff DC, Garg SK, et al. Threshold-based insulin-pump interruption for reduction of hypoglycemia. N Engl J Med. 2013;369(3):224-232. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society

RESOURCES