Skip to main content
PLOS One logoLink to PLOS One
. 2021 Apr 29;16(4):e0250258. doi: 10.1371/journal.pone.0250258

Biologically transparent illumination is a safe, fast, and simple technique for detecting the correct position of the nasogastric tube in surgical patients under general anesthesia

Hirofumi Hirano 1, Hanayo Masaki 1, Teppei Kamada 1, Yoshie Taniguchi 1, Eiji Masaki 1,*
Editor: Benedikt Ley2
PMCID: PMC8084215  PMID: 33914808

Abstract

The aim of this study was to evaluate the effectiveness of using biologically transparent illumination to detect the correct position of the nasogastric tube in surgical patients. This prospective observational study enrolled 102 patients undergoing general surgeries. In all cases, a nasogastric tube equipped with a biologically transparent illumination catheter was inserted after general anesthesia. The identification of biologically transparent light in the epigastric area either with or without finger pressure indicated that the tube had been successfully inserted into the stomach. X-ray examination was performed to ascertain the tube position and was compared with the findings of the biologically transparent illumination technique. Biologically transparent light was detected in 72 of the 102 patients. In all of these 72 patients, the position of the nasogastric tube in the stomach was confirmed by X-ray examination. The light was not detected in the other 30 patients; X-ray examination showed that the nasogastric tube was positioned in the stomach in 21 of these 30 patients but not in the other 9. The sensitivity and specificity of the illumination were 77.4% and 100%, respectively. The results suggest that biologically transparent illumination is a useful and safe technique for detecting the correct position of the nasogastric tube in surgical patients under general anesthesia. When the BT light cannot be identified, X-ray examination is mandatory to confirm the position of the nasogastric tube.

Introduction

Nasogastric tube (NGT) insertion is a general procedure that allows access to the stomach for diagnostic and therapeutic purposes in various clinical settings, including emergency rooms, operating rooms, intensive care units, and nursing homes. Malposition of the NGT in the respiratory tract or esophagus can lead to serious complications such as aspiration pneumonia, pneumothorax, and even death [1,2]. Several methods have been employed to aid the correct positioning of NGT. Auscultation of gurgling sounds in the epigastrium is most common method but lacks accuracy [3]. Measuring the pH of aspirates or CO2 from the NGT are alternative techniques, but these techniques have lower sensitivity and specificity than X-ray examination [4,5]. Ultrasonography is another alternative but requires special training for NGT visualization [6]. Because of the abovementioned limitations of these methods, X-ray examination is recognized as the gold standard for confirming the correct position of the NGT; however, X-ray examination also has problems, including delayed verification, radiation exposure, and the burden on medical staff [7].

Biologically transparent (BT) light is highly bio-permeable red light from a light-emitting diode (LED). BT light emitted in the abdominal digestive tract can be visualized from outside the body. Identification of BT light emitted from the tip of a special catheter (BT catheter) confirms the tip position. BT illumination has previously been utilized to determine the site for colonoscopy-assisted percutaneous sigmoidopexy [8].

The purpose of this study was to investigate whether the identification of BT light in the epigastric area from outside the body can aid in detecting the correct position of the NGT in the stomach. In other words, are the sensitivity and specificity for verifying the correct position of the NTG with BT illumination close to 100% when X-ray examination is used as the reference standard? To test this hypothesis, an NGT equipped with a BT catheter was inserted into surgical patients under general anesthesia and the BT light was checked in the epigastric area. We also performed conventional epigastric auscultation and visual inspection of gastric aspirate. X-ray examination was performed as the reference standard.

Materials and methods

This study was approved by the ethics committee of the International University of Health and Welfare Hospital (No. 20-B-446) and conformed to the Ethical Guidelines for Clinical Studies in Japan by the Ministry of Health, Labour and Welfare as well as the principles expressed in the Declaration of Helsinki. This prospective observational study was conducted at a local university hospital with 400 beds. Verbal informed consent, which was approved by the university’s institutional review board, was obtained from all patients and recorded in our database.

BT light and BT catheter

BT light is highly bio-permeable red light emitted from an LED. The light is delivered from the light source (32037000; Neuroceuticals Inc., Tokyo, Japan; Fig 1A), which is connected to the BT catheter (Fig 1B). The BT catheter (70231000; Neuroceuticals Inc.) consists of plastic optical fibers and has a diameter of 1.5 or 2.0 mm. Because BT light is transferred to the tip of the BT catheter (Fig 1C), the NGT tip position can be identified by introducing the BT catheter into the NGT. The wavelength of the BT light is 660 nm, which passes through soft tissue but not hard tissue such as bone and cartilage. Red light similar to this wavelength is reported to pass through the human cheek but does not penetrate the skull and surrounding soft tissue [9]. BT light transferred to the tip of the BT catheter in the abdominal digestive tract can be visually confirmed from outside the body. Therefore, it seems to be possible to identify the correct tip position of the NGT in the stomach by inserting the BT catheter into the NGT and identifying the BT light emitted in the epigastric area.

Fig 1.

Fig 1

(A) The biologically transparent (BT) light source. (B) The BT catheter. (C) The BT light is transferred to the tip of BT catheter.

Patients and procedures

This prospective observational study enrolled patients aged 18 years or older who needed gastric decompression because of mask-bag ventilation after induction of general anesthesia for general surgeries. The decision to insert the NGT (Salem Sump/225AABZX00046000; Covidien Japan, Tokyo, Japan) was made by the anesthesiologist. The anesthesiologist chose among three NGT sizes (14, 16, and 18 Fr) and the first 5–10 cm of the tip was lubricated. After the BT catheter was inserted, the NGT was gently advanced from the nostril to a length determined by the nose-ear-xiphoid method. The anesthesiologist did not force the insertion. If the nasal passage was difficult, the approach was changed to the oral route. Identification of BT light was performed by the anesthesiologist either with or without finger pressure. Pressure was applied using the index finger 3–4 cm below the tenth rib on the midclavicular line in the epigastric area. After identification of the BT light, the BT catheter was removed from the NGT. Next, auscultation for gurgling sounds during injection of 10 mL of air was performed, followed by aspiration and visual inspection of gastric fluid. Finally, X-ray examination was carried out to confirm the position of the NGT. Both the anesthesiologist and the surgeon interpreted the X-ray image, and the sensitivity and specificity of BT light were calculated using the X-ray examination results as the reference standard. We also examined sensitivity and specificity in obese patients, and thickness of the abdominal wall (just below the tenth rib on the midclavicular line) was measured to distinguish the influence of obesity. Obesity was defined as BMI >27.5.

Statistical analysis

All variables are expressed as the mean ± standard deviation. Sensitivity, specificity, positive predictive value, and negative predictive value were calculated to evaluate the diagnostic effectiveness of BT light.

We hypothesized that the NGT would always be placed in the stomach when the BT light is detected in the epigastric area from outside the body, resulting in a sensitivity and positive predictive value of 100%. The required sample size was calculated by the Clopper-Pearson method [10] on the basis that the lower limit of the 95% confidence interval exceeds positive predictive value of 95% [11], using SAS software, version 9.4 (SAS Institute, Cary, NC). Accordingly, the procedure was performed until BT light was positively identified in 72 patients.

Results

A total of 102 patients were enrolled. The patient characteristics are summarized in Table 1. Mean age was 63.6 ± 16.0 years and male to female ratio was 62:40. There were 19 obese patients. Mean BMI and male-to-female ratio in the obese patient group were 29.2 ± 2.5 and 14:5, respectively.

Table 1. Patient characteristics (N = 102).

Height (cm) 161 ± 9
Weight (kg) 62 ± 14
Age (years) 64 ± 16
BMI (kg/m2) 24 ± 4
Percentage of obese patients 18.6

Data are expressed as the mean ± standard deviation.

Of the 102 patients, BT light was identified in the epigastric area in 72 (Fig 2) but not in 30. X-ray examination confirmed the presence of the NGT in the stomach in 93 patients but not in the other 9. In addition, the position of the NGT in all 72 BT light-positive patients was confirmed to be in the stomach by X-ray examination. In 21 of the 30 BT light-negative patients, the presence of the NGT in the stomach was confirmed, whereas in the other 9, the NGT was found to be in the trachea (n = 1) or in the esophagus or esophago-columnar junction (n = 8) (Table 2). The sensitivity and specificity of the BT light test were 77.4% and 100%, respectively. Because the sensitivity for detecting the correct position of the NTG with BT light was not close to 100% when X-ray examination was used as the reference standard, the usefulness of BT light in identifying the correct position of the NGT was not considered equivalent to that of X-ray examination.

Fig 2. Positive sign of biologically transparent (BT) light, which showed several patterns.

Fig 2

The BT light was detected in the epigastric area of the center site (A), the upper site (B), the lateral site (C), and using finger pressure (D).

Table 2. Results of diagnostic tests.

Evaluation test Reference test (X-ray) Sensitivity (%) Specificity (%) PPV (%) NPV (%)
Correct position (n = 93) Incorrect position (n = 9)
BT light
Positive (+) 72 0 77 100 100 30
Negative (−) 21 9
Auscultation
Positive (+) 91 3 97 66 96 75
Negative (−) 2 6
Gastric aspirate
Positive (+) 82 4 88 56 95 31
Negative (−) 11 5

PPV, positive predictive value; NPV, negative predictive value.

The abdominal thickness of true-positive and false-negative patients with BT light were 21.0 ± 8.1 (mean ± S.D.; n = 72) and 20.1 ± 5.4 (mean ± S.D.; n = 21), respectively.

The sensitivity and specificity of the auscultation test were 97.8% and 66.7%, respectively. Likewise, the sensitivity and specificity of visual inspection of the gastric aspirate were 88.1% and 55.6%, respectively.

The sensitivity and specificity of the BT light test in obese patients were 73.3% and 100%, comparable to the results in patients as a whole (Table 3).

Table 3. Results of diagnostic tests in obese patient.

Evaluation test Reference test (X-ray) Sensitivity (%) Specificity (%) PPV (%) NPV (%)
Correct position (n = 15) Incorrect position (n = 4)
BT light
Positive (+) 11 0 73 100 100 50
Negative (−) 4 4
Auscultation
Positive (+) 15 2 100 50 88 100
Negative (−) 0 2
Gastric aspirate
Positive (+) 13 3 87 25 81 33
Negative (−) 2 1

PPV, positive predictive value; NPV, negative predictive value.

Discussion

The key results of this study are as follows: 1) the NGT was always positioned in the stomach when the BT light was identified in the epigastric area; 2) the BT light was not identified when the NGT was not positioned in the stomach; 3) the BT light was not always detected in the epigastric area when the NGT was positioned in the stomach. These results suggest that using BT light to identify the NGT position might be useful and safe, although further improvements in the BT illumination technique, including the quality of the equipment, are needed to reduce the false-negative rate. Currently, when BT light is not identified, X-ray examination is mandatory to confirm the position of the NGT.

Using BT light to identify the correct position of the NGT is safe, fast, and simple. The safety of the BT illumination method was ascertained by the fact that there were no false-positive cases in the present study; that is, the specificity of this method was 100%. The NTG was always positioned in the stomach when BT light was detected in the epigastric area. Using NGT for administration of nutrition or drugs after detection of BT light in the epigastric area might avoid serious complications due to malposition of the NGT [12], although we did not specifically investigate this matter. Furthermore, no harmful events related to insertion of the BT catheter were observed in the present study. BT light was immediately identified as the NGT equipped with the BT catheter was advanced into the stomach. Little extra time (up to 60 s) is required to insert the BT catheter into the NGT, connect the BT catheter to its light source, and identify the correct position. BT light can be identified by personnel without special training or knowledge. Moreover, this method can be used in any medical setting where a BT catheter and its light source are available. This safe, fast, and simple method has the potential to solve many problems associated with NGT insertion.

This study demonstrated unsatisfactory sensitivity (77.4%) with the BT illumination technique. There are several possible reasons for this relatively low sensitivity. First, the BT light might not have been in the direction of the abdominal wall in the false-negative cases. The intensity is most powerful in the direction perpendicular to the cross-section of the BT catheter tip. Therefore, the light intensity might have been too low to be detectable from outside the body if the cross-section of the BT catheter tip was pointed away from the abdominal wall. Second, the contents of the stomach might interfere with the transmission of the BT light. Although oral intake was restricted before the procedure for all patients, a certain amount of gastric fluid was aspirated in some patients. Gastric fluid might change the transmission of BT light, thereby resulting in false-negative cases. Finally, many obese patients were included in this study. Although BT light is highly bio-permeable, the extent to which it can pass through thick tissues has not been investigated. The abdominal thickness did not differ between true-positive and false-negative patients [21.0 ± 8.1 (mean ± S.D.; n = 72) and 20.1 ± 5.4 (mean ± S.D.; n = 21), respectively]. In addition, the sensitivity and specificity of the BT light test in obese patients were comparable to the results in patients as a whole. Statistical analysis of this point cannot be performed, but it seems that abdominal thickness might not be associated with the detection of BT light in this study.

Previous studies have reported that the correct diagnostic rate (i.e., sensitivity) for the position of the NGT in the stomach is 86%-97% with ultrasonography [1315], 78%-91% with pH measurement of the aspirates [13,15], and 86% with CO2 measurement from the NGT [4]. In contrast, when the NGT is not in the stomach, the accurate diagnostic rate (i.e., specificity) for the respective techniques was 67%-100% [13,14], 67%-86% [13,16], and 22% [4]. The present study revealed that the sensitivity and specificity when using BT illumination were 77% and 100%, respectively. These results suggest that BT illumination might be inferior to ultrasonography and pH measurement of the aspirates in terms of sensitivity. Despite the better sensitivity of these two techniques, ultrasonography and pH measurement of the aspirates have their own drawbacks, which the BT illumination method does not share. As mentioned above, special training is needed for NGT visualization with ultrasonography and it also takes more time (about 24 min compared with 1 min for BT illumination) [15]. The low specificity of pH measurement of the aspirates compared with BT illumination is a definite limitation. The present study also demonstrated that the specificity of the gastric aspirate test was quite low. Therefore, the BT illumination method might be a useful technique for identifying the correct position of the NGT. However, further improvement is needed to increase sensitivity for detecting the position of the NGT in the stomach.

There are several limitations to this study. First, the NGT was inserted under general anesthesia in the supine position. NGTs are inserted in a variety of clinical settings for various purposes. In particular, NGTs are often used in elderly nursing home residents for feeding in the sitting position without the use of sedatives. The results of the BT illumination method reported in this study might differ from cases without general anesthesia in non-supine positions. Second, we used only one type of NGT with three different diameters (14, 16, and 18 Fr). Many kinds of NGT are available in many different sizes and NGTs are made from different kinds of materials, which might affect the biological transparency of NGTs. Finally, we evaluated the usefulness of BT illumination for identifying the correct position of the NTG in only 102 patients. The number of patients was determined in order to estimate the sensitivity and positive predictive value of BT illumination compared with X-ray examination. The high specificity of BT illumination (100%) suggests that it is safe to use; however, the small number of cases in which the NGT was not positioned in the stomach (9 cases) was insufficient to confirm its complete safety. Therefore, a study involving a larger number of patients will be necessary to assess the safety of BT illumination for detecting the correct position of the NGT.

In conclusion, the NGT was always positioned in the stomach when the BT light was identified in the epigastric area, and the BT light was not identified when the NGT was not positioned in the stomach. These results suggest that using BT illumination is a safe, fast, and simple technique for identifying the position of the NGT and might be useful in surgical patients under general anesthesia. When BT light is not identified, X-ray examination is mandatory to confirm the position of the NGT.

Supporting information

S1 File. Study of NG tube with BT.

(XLSX)

Acknowledgments

We thank the operating room staff at the International University of Health and Welfare Hospital for assistance with data collection.

Data Availability

All relevant data are within the manuscript and its Supporting information files.

Funding Statement

The authors received no specific funding for this work.

References

  • 1.Roubenoff R, Ravich W. Pneumothorax due to nasogastric feeding tubes: Report of four cases, review of the literature, and recommendations for prevention. Arch Intern Med. 1989;149:184–188. 10.1001/archinte.149.1.184 [DOI] [PubMed] [Google Scholar]
  • 2.Bankier AA, Wiesmayr MN, Henk C. Radiographic detection of intrabronchial malpositions of nasogastric tubes and subsequent complications in intensive care unit patients. Intensive Care Med. 1997;23:406–410. 10.1007/s001340050348 [DOI] [PubMed] [Google Scholar]
  • 3.Harrison AM, Clay B, Grant MJ, Sanders SV, Webster HF, Reading JC, et al. Nonradiographic assessment of enteral feeding tube position. Crit Care Med. 1997;25:2055–2059. 10.1097/00003246-199712000-00026 [DOI] [PubMed] [Google Scholar]
  • 4.Mordiffi SZ, Goh ML, Phua J, Chan YH. Confirming nasogastric tube placement: Is the colorimeter as sensitive and specific as X-ray? A diagnostic accuracy study. Int J Nurs Stud. 2016;61:248–257. 10.1016/j.ijnurstu.2016.06.011 [DOI] [PubMed] [Google Scholar]
  • 5.Rowat AM, Graham C, Dennis M. Study to determine the likely accuracy of pH testing to confirm nasogastric tube placement. BMJ Open Gastroenterol. 2018;5:e000211. 10.1136/bmjgast-2018-000211 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Mak MY, Tam G. Ultrasonography for nasogastric tube placement verification: An additional reference. Br J Community Nurs. 2020;25:328–334. 10.12968/bjcn.2020.25.7.328 [DOI] [PubMed] [Google Scholar]
  • 7.Tho PC, Mordiffi S, Ang E, Chen H. Implementation of the evidence review on best practice for confirming the correct placement of nasogastric tube in patients in an acute care hospital. Int. J. Evid. Based Healthc. 2011;9:51–60. 10.1111/j.1744-1609.2010.00200.x [DOI] [PubMed] [Google Scholar]
  • 8.Imakita T, Suzuki Y, Ohdaira H, Urashima. Colonoscopy-assisted percutaneous sigmoidopexy: A novel, simple, safe, and efficient treatment for inoperable sigmoid volvulus (with videos). Gastrointest Endosc. 2019;90:514–520. 10.1016/j.gie.2019.04.246 [DOI] [PubMed] [Google Scholar]
  • 9.Jagdeo JR, Adams LE, Brody NI, Siegel DM. Transcranial red and near infrared light transmission in a cadaveric model. PLoS One. 2012;7:e47460. 10.1371/journal.pone.0047460 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Clopper CJ, Pearson ES. The use of confidence or fiducial limits illustrated in the case of the binomial. Biometrika. 1934;26:404–413. [Google Scholar]
  • 11.Andersson M, López-Vega JM, Petit T, Zamagni C, Easton V, Kamber J, et al. Efficacy and Safety of Pertuzumab and Trastuzumab Administered in a Single Infusion Bag, Followed by Vinorelbine: VELVET Cohort 2 Final Results. Oncologist. 2017;22:1160–1168. 10.1634/theoncologist.2017-0079 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Metheny NA, Meert KL, Clouse RE. Complications related to feeding tube placement. Curr Opin Gastroenterol. 2007;23:178–82. 10.1097/MOG.0b013e3280287a0f [DOI] [PubMed] [Google Scholar]
  • 13.Kim HM, So BH, Jeong WJ, Choi SM, Park KN. The effectiveness of ultrasonography in verifying the placement of a nasogastric tube in patients with low consciousness at an emergency center. Scand J Trauma Resusc Emerg Med. 2012;12;20:38. 10.1186/1757-7241-20-38 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Brun PM, Chenaitia H, Lablanche C, Pradel AL, Deniel C, Bessereau J, et al. 2-point ultrasonography to confirm correct position of the gastric tube in prehospital setting. Mil Med. 2014;179:959–963. 10.7205/MILMED-D-14-00044 [DOI] [PubMed] [Google Scholar]
  • 15.Vigneau C, Baudel JL, Guidet B, Offenstadt G, Maury E. Sonography as an alternative to radiography for nasogastric feeding tube location. Intensive Care Med. 2005;31:1570–1572. 10.1007/s00134-005-2791-1 [DOI] [PubMed] [Google Scholar]
  • 16.Boeykens K, Steeman E, Duysburgh I. Reliability of pH measurement and the auscultatory method to confirm the position of a nasogastric tube. Int J Nurs Stud. 2014;51:1427–1433. 10.1016/j.ijnurstu.2014.03.004 [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Benedikt Ley

1 Mar 2021

PONE-D-21-02589

Biologically transparent illumination is a safe, fast, and simple technique for detecting the correct position of the nasogastric tube in surgical patients under general anesthesia

PLOS ONE

Dear Dr. Masaki,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Kindly address the comments made by reviewer 1 and upload all relevant data in format that will not allow to identify any of the study participants. If data can not be attached directly to the publication, kindly provide information on how and where data can be accessed by interested parties.

Please submit your revised manuscript by Apr 10 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Benedikt Ley

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. PLOS requires an ORCID iD for the corresponding author in Editorial Manager on papers submitted after December 6th, 2016. Please ensure that you have an ORCID iD and that it is validated in Editorial Manager. To do this, go to ‘Update my Information’ (in the upper left-hand corner of the main menu), and click on the Fetch/Validate link next to the ORCID field. This will take you to the ORCID site and allow you to create a new iD or authenticate a pre-existing iD in Editorial Manager. Please see the following video for instructions on linking an ORCID iD to your Editorial Manager account: https://www.youtube.com/watch?v=_xcclfuvtxQ

3. Please clarify whether the IRB approved the use of verbal consent, and how consent was recorded.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The paper would benefit from including data to support the conclusions made, including the % of obese women in the cohort; stratification of test performance by obesity; an evaluation of how BT sensitivity could be improved by combining it with auscultation and gastric pH (other point of care tools to ascertain NGT location); and a breakdown of + BT Light by anatomical site. I would recommend publication once these issues are addressed.

Minor comments include:

Page 6 – define ‘with or without finger pressure’. Where and how this is done?

Page 7 – ‘on the basis that the lower limited of the 95% confidence interval exceeds 95%’.. please clarify exceeds 95% of what?

Page 8 – Table 1. Can you provide range and % of overweight and obese patients in the cohort. One would expect performance BTL to suffer the more obese the patients are.

Page 8: ‘the power of BT light to identify the correct position of the NGT was not equivalent to that of X-ray examination’ – can you rephrase this? It’s not entirely clear what is referred to here.

Page 9: Remove ‘ the latter of which was quite low’ and expand on this in the discussion.

Page 9: Figure 2. This figure caption mentions anatomical areas where BT light was detected. Is there a breakdown of this available per anatomical site, and how this affected test performance? You are alluding to this in the first paragraph of your discussion but you are not showing the data.

Abdominal wall = epigastric area? Use consistent descriptors for your anatomical sites in the paper.

Page 11: I suggest to move the section on BTL and abdominal wall thickness to the results part of the paper, including methodology of the measurements. As above, include more details of the proportion of obese patients etc in your results. Potentially stratify results BMI.

Page 13 – discuss safety of BT light (here or in background section)

Page 12 – can a combination of BT, auscultation, and gastric aspirate (can be done in at bedside) be used to improve sensitivity?

Reviewer #2: I think it’s a clever approach.maybe more patients could have been found.

Nicely presented.Helpful for clinical practice.Simple techniques.Also nice to see the pictures at the end.

I have no more remarks about it

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Apr 29;16(4):e0250258. doi: 10.1371/journal.pone.0250258.r002

Author response to Decision Letter 0


5 Mar 2021

We responded to all comments of editor and corrected the revised manuscript according to these responses.

The responses to reviewer are in the attachment file, "Response to Reviewers"

Attachment

Submitted filename: Response to Reviewers.docx

Decision Letter 1

Benedikt Ley

22 Mar 2021

PONE-D-21-02589R1

Biologically transparent illumination is a safe, fast, and simple technique for detecting the correct position of the nasogastric tube in surgical patients under general anesthesia

PLOS ONE

Dear Dr. Masaki,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by May 06 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Benedikt Ley

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Additional Editor Comments (if provided):

On page 5 You have added the sentence: "We will open all data underlying the findings described in the present study without restriction according to the request" and this sentence is suitable for a reply to the reviewers, but should not be added to the manuscript. Kindly include all underlying data never the less.

Kindly have a native English speaker revise the introduced changes, the language in most cases will require some modification. Specifically reviewer 1 had earlier asked to revise the sentence on page 10, which you kindly revised to: "Because the sensitivity for verifying the correct position of the NTG with BT light did not close to 100% when X-ray examination is used as the reference standard, the usefulness of BT light to identify the correct position of the NGT was not equivalent to that of X-ray examination." Unfortunately this sentence is still hard to understand, kindly revise together with a native English speaker.

[Note: HTML markup is below. Please do not edit.]

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Apr 29;16(4):e0250258. doi: 10.1371/journal.pone.0250258.r004

Author response to Decision Letter 1


30 Mar 2021

Reply to the Editor.

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

We reviewed all the references in our manuscript. We can not find any retracted papers in our references. If you have any information about the retracted paper in our references, please let me know.

Additional Editor Comments (if provided):

On page 5 You have added the sentence: "We will open all data underlying the findings described in the present study without restriction according to the request" and this sentence is suitable for a reply to the reviewers, but should not be added to the manuscript. Kindly include all underlying data never the less.

According to the pointing out of the Editor, we delete the sentence (page 5). All underling data are included in the manuscript.

Kindly have a native English speaker revise the introduced changes, the language in most cases will require some modification. Specifically reviewer 1 had earlier asked to revise the sentence on page 10, which you kindly revised to: "Because the sensitivity for verifying the correct position of the NTG with BT light did not close to 100% when X-ray examination is used as the reference standard, the usefulness of BT light to identify the correct position of the NGT was not equivalent to that of X-ray examination." Unfortunately this sentence is still hard to understand, kindly revise together with a native English speaker.

We are apologize for our limited fluency in English. The manuscript has been professionally edited by a native English speaker familiar with this area of research. The sentence "Because the sensitivity for verifying the correct position of the NTG with BT light did not close to 100% when X-ray examination is used as the reference standard, the usefulness of BT light to identify the correct position of the NGT was not equivalent to that of X-ray examination." was revised to “Because the sensitivity for detecting the correct position of the NTG with BT light was not close to 100% when X-ray examination was used as the reference standard, the usefulness of BT light in identifying the correct position of the NGT was not considered equivalent to that of X-ray examination. (Page 10).

The certification of the editing is enclosed.

Additional correction.

We are deeply sorry, but we found the simple mistakes in Table 2. The numbers of patients in Auscultation test were wrong (accordingly, sensitivity, and NPV). We corrected the numbers. Those corrections did not change any other contents in the manuscript including the conclusions at all. We are sorry again.

Attachment

Submitted filename: Response to the Editor.docx

Decision Letter 2

Benedikt Ley

5 Apr 2021

Biologically transparent illumination is a safe, fast, and simple technique for detecting the correct position of the nasogastric tube in surgical patients under general anesthesia

PONE-D-21-02589R2

Dear Dr. Masaki,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Benedikt Ley

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Benedikt Ley

12 Apr 2021

PONE-D-21-02589R2

Biologically transparent illumination is a safe, fast, and simple technique for detecting the correct position of the nasogastric tube in surgical patients under general anesthesia

Dear Dr. Masaki:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Benedikt Ley

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 File. Study of NG tube with BT.

    (XLSX)

    Attachment

    Submitted filename: Response to Reviewers.docx

    Attachment

    Submitted filename: Response to the Editor.docx

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES