An integrated augmented reality surgical navigation platform using multi-modality imaging for guidance

Harley H L Chan 1,*, Stephan K Haerle 2, Michael J Daly 1, Jinzi Zheng 1, Lauren Philp 3,4, Marco Ferrari 1,5,6, Catriona M Douglas 1,5,7, Jonathan C Irish 1,5,7
Editor: Domokos Máthé8
PMCID: PMC8087077  PMID: 33930063

Abstract

An integrated augmented reality (AR) surgical navigation system that potentially improves intra-operative visualization of concealed anatomical structures. Integration of real-time tracking technology with a laser pico-projector allows the surgical surface to be augmented by projecting virtual images of lesions and critical structures created by multimodality imaging. We aim to quantitatively and qualitatively evaluate the performance of a prototype interactive AR surgical navigation system through a series of pre-clinical studies. Four pre-clinical animal studies using xenograft mouse models were conducted to investigate system performance. A combination of CT, PET, SPECT, and MRI images were used to augment the mouse body during image-guided procedures to assess feasibility. A phantom with machined features was employed to quantitatively estimate the system accuracy. All the image-guided procedures were successfully performed. The tracked pico-projector correctly and reliably depicted virtual images on the animal body, highlighting the location of tumour and anatomical structures. The phantom study demonstrates the system was accurate to 0.55 ± 0.33mm. This paper presents a prototype real-time tracking AR surgical navigation system that improves visualization of underlying critical structures by overlaying virtual images onto the surgical site. This proof-of-concept pre-clinical study demonstrated both the clinical applicability and high precision of the system which was noted to be accurate to <1mm.

Introduction

Imaging-based surgical navigation (SN) systems are routinely used to guide surgical procedures in anatomically complex areas such as the head and neck [1]. Previous studies have demonstrated that the use of SN can improve efficiency and safety in these challenging areas [2]. The added value of SN is twofold: firstly, it facilitates the identification of critical anatomical structures to avoid unnecessary complications; and furthermore, it helps to delineate tumor boundaries during oncologic ablations with the intent to improve adequacy of margins [35].

Currently, research is focused on augmented reality (AR) methods such as video-computed tomography (CT) augmentation [6, 7] and intraoperative imaging [8] to further improve the usefulness of SN. Optical “see-through” techniques, such as video-CT, consist of generating virtual anatomical structures from cross-sectional images (e.g. CT or magnetic resonance imaging [MRI]) that are overlaid on the endoscopic image [9]. Literatures reveal that augmented reality in medical research had been very active in past decade [10]. However, majority of proposed approaches are AR system independently operate and not integrate into surgical navigation system. Nowadays, the surgical navigation system has been routinely used in the interventional procedure. The integration of SN with AR would synergistically improve the performance of technology based guidance. In fact, an integrated AR-SN system would provide a precise, real-time topographical localization of the surgical field by means of intuitive visual enhancement.

Following this line of research, the Guided Therapeutics (GTx) group (TECHNA Institute, University Health Network, Toronto, Ontario, Canada) has developed a handheld AR device integrated into a SN system, which is capable of surgical site augmentation using both medical images and computer-generated virtual images. Briefly, this augmentation is achieved by using a tracked pico-projector which superimposes pre-contoured structures (i.e. the volume occupied by the tumour, or critical neural and vascular structures) onto the surgical field. The aim of this was to assess the feasibility of this novel imaging system (AR-SN integrated) through pre-clinical mice models.

Materials and methods

System architecture

The prototype AR-SN system consists of pico-projector (L1 v2, AAXA Technology Inc., Santa Ana, California, United States), an infrared (IR) real-time optical tracking system (Polaris Spectra, NDI, Waterloo, Ontario, Canada), a universal serial bus (USB) 2.0M pixels generic camera, and a laptop computer. The pico-projector employed in the prototype is light (170g) and physically small (54mm x 170mm x 21mm).

The enclosure of the AR device is fabricated with acrylic and designed to encase the pico-projector, the IR reference marker mount, and the USB camera. The signal transmission of pico-projector is via High-Definition Multimedia Interface (HDMI) connector. Data processing is performed using a laptop computer (M4500, Precision laptop, Dell, Round Rock, Texas). Fig 1 illustrates the spatial relationship and interaction between individual components in the augmented reality surgical navigation platform.

Fig 1. Prototype augment reality surgical navigation platform consists of optical tracking system and tracked pico-projector.

Fig 1

A SN platform, named “GTxEyes” (in-house development) was used in combination with the prototype as it allows for image display, fusion, and overlay of multiple imaging modalities including standard CT, cone beam CT, MRI, single photon emission CT (SPECT), positron emission tomography (PET), and video-endoscopy. The GTxEyes platform [11] was developed using open-source, cross-platform libraries included IGSTK [12], ITK [13], and VTK [14].

Surgical plans including the target lesion, ablation margins, critical structures, and safety margin (i.e. voxel contouring of structures to be spared during ablation) were created pre-operatively using ITK-SNAP [15] and built on the SN system. Real-time tracking detects the location of the surgical instruments in 3D space, thus guiding the surgeon throughout the ablation and can alert the surgeon when the navigated instrumentation enters a pre-determined safety margin volume [16]. The AR-SN platform supports fully automatic bony and soft tissue digital segmentation based on voxel intensity threshold value (e.g. Hounsfield Unit for CT imaging) and all surgical plans can represented by either 3D surface or volume rendering and overlaid onto the surgical field with adjustable opacity. The AR-SN platform allows the surgeon to scroll through image slices projected on the operative field with a tracked pointer and to accordingly decide the depth of images through which to augment the surgical view.

Image overlay and system operation

The AR-SN system was registered into a single coordinate system by pairing correspondent landmarks using fiducial markers identifiable in both the image and subject. Once registration was completed, the AR-SN platform could track multiple surgical instruments simultaneously, as described in a previous study [17, 18].

To facilitate real-time tracking of the AR device in 3D space, an optical sensor attachment was mounted to the enclosure of the pico-projector. Pre-operative calibration consecutively performs to define the spatial relationship between the sensor and the centre of the pico-projector, which elaborates by a transformation matrix stored in the AR-SN platform. Additionally, calibration allows tracking of the spatial position of the projector and synchronization of a virtual camera in the AR-SN platform. This calibration procedure ensures the pico-projector correctly illuminates the surgical field providing a reliable image of the pre-contoured structures (i.e. those delineated in the surgical plans). Further details on the calibration and registration procedure are provided in the S1 File. By virtue of the real-time tracking, the pico-projector can be repositioned during the surgical procedure according to intraoperative requirements without compromising projection accuracy. Moreover, the integrated AR-SN can project multiple virtual objects and render multimodality fused imaging. Contours from external software such as ITK-SNAP [15] and 3D slicer [18, 19] can also be imported into the AR-SN platform for sophisticated delineation of anatomical structures.

Preclinical animal studies

The integrated AR-SN system is evaluated on four preclinical mice models. All animal studies are performed in the animal care resource facility at the Princess Margaret Cancer Research Centre in accordance with protocols approved by the animal care committee at University Health Network. Four independent studies are performed to evaluate system performance and to investigate multimodality imaging in intraoperative AR-SN-based guidance. The first two studies investigated PET/CT for AR image guidance, whereas the third and fourth studies investigated PET/MRI and SPECT/CT, respectively.

In the first study, 64Cu isotope (10–13 MBq per mouse) was administered to a healthy, immunocompetent CD-1 mouse (Charles River, Wilmington, Massachusetts, United States) intravenously via a lateral tail vein one hour before micro-PET imaging (Focus 220, Siemens, Malvern, Pennsylvania, United States). A whole body image acquisition time took 15 minutes per mouse, the image resolution was 0.146 x 0.146 x 0.796 mm3. CT images were acquired after PET-imaging using a micro-CT scanner (Locus Ultra, GE Healthcare, Pittsburgh, Pennsylvania, United States) with imaging parameters set at 80 kVp and 50 mA. During CT scanning, multiple 3D printed polycarbonate fiducial markers were introduced to surround the mouse bed to facilitate the subsequent AR-SN system registration. The resulting image volume was 366 x 196 x 680 voxels (366 x 196 matrix with 680 slice images), with isotropic voxel size of 0.154 mm3. PET/CT images were then co-registered using Inveon Research Workplace (IRW) software (Siemens Healthcare, Malvern, Pennsylvania, United States). CT images were down-sampled to 0.3 mm3 isotopic voxel size to minimize computational intensity.

In the second study, 5 million BT474-M3 cells (HER2-overexpressing breast carcinoma) suspended in 100μL PBS were injected subcutaneously into the right mammary fat pad of athymic female CD-1 nude mouse. The animal was then monitored bi-weekly for tumour growth. Once the tumor reached a size of ~400-500mm3, the animal underwent the PET/CT as previously described. The tumour was preoperatively contoured based on available imaging using ITK-SNAP [15], and subsequently imported into the AR-SN platform.

Third study was investigating PET/MRI AR image guidance, a female CD-1 nude mouse was inoculated with 5 million MDA-MB-231 breast cancer cells through a subcutaneous injection in the bilateral upper mammary fat pads (100μL on each side). The mouse was then monitored as previously described. Once tumours reached a size of approximately 200-250mm3, liposomes loaded with 64Cu were administrated intravenously 24 hours prior to imaging at a dose of 10–13 MBq/mouse and 20 μmol phospholipid/kg [20]. The mouse then underwent micro-PET imaging with a longer acquisition time of 60 minutes. Whole thorax-abdomen MRI was then performed using a 7T micro-MRI scanner (M2, Aspect Imaging, Shoham, Israel). The fusion of the MR and PET data sets was registered using IRW software with a rigid body algorithm based on normalized mutual information.

The fourth study experimented SPECT/CT image modality for AR image guidance. A female athymic CD-1 nude mouse were injected 10 million 231-H2N cells (human breast cancer cells) into the left thigh. The mouse was monitored as previously described until the tumour reached a volume of 125 mm3. The mouse was then injected intravenously with 37 MBq of 111In-Fab fragments of anti-HEGF (human epidermal growth factor) antibody via lateral tail vein injection 48 hours prior to imaging [21]. SPECT/CT imaging (Bioscan, Washington DC, Washington, United States), was performed with dual-modality machine. Photons were accepted from the 10% windows centered on indium two photo-peaks at 171 and 245 keV. The SPECT projections were acquired in a 256 x 256 x 16 matrix for 85 minutes. Voxel size was isotropic 0.3mm3. Images were reconstructed using an ordered-subset expectation maximization algorithm (9 iterations) [22]. Cone-beam CT images were acquired (180 projections, 45 kVp) immediately before the micro-SPECT images. Eight 3D printed markers were attached around the scan bed for tracking registration. Co-registration of SPECT and CT images was performed using pertinent software (InvivoScope, Bioscan Inc, Washington DC, Washington, United States).

Upon completion of all multimodality imaging studies, mice were euthanized using an overdose of 5% inhaled isoflurane and their body position was rigidly maintained in the scan bed. For each experiment, the AR device was mounted on an articulating arm located at approximately 250–350 mm above the supine mouse.

AR-SN system accuracy measurement

The accuracy of the AR-SN system was quantitatively evaluated by overlaying the projection image on a checkerboard phantom with known dimensions. The phantom was composed of 25 rectangular grids (10 x 10 mm2 each) with central divots (5 rows and 5 columns). The distance between the central divots was 20mm. The acrylic phantom was fabricated by a high precision computer numerical control (CNC) machine. The guidance images for the phantom navigation were acquired with a prototype cone-beam CT C-arm [23]. CT images were 256 x 256 x 192 voxels by volume with an isotopic voxel size of 0.78 mm3. The accuracy of the AR-SN system was evaluated by comparing landmark localization on the phantom with and without the use of AR guidance. The central divot of each grid was localized manually using the tracked pointer three separate times per grid. The represented location of the divot was calculated from the mean location of the tracked pointer over the three measurements. These localization exercises were conducted over a range of distances between the phantom and AR projector including 200, 300, and 400 mm. The uncertainty of the projected location was calculated as follows: (xi, yi, zi)real represents as “real” location of the divot center acquired manually without AR image guidance and (xi, yi, zi)virtual represents as “virtual” location of the divot generated from the AR projection image, the estimation of error is the distance between the “real” and “virtual” location of the central divot such that Err = ‖(xi,yi,zi)real−(xi,yi,zi)virtual‖.

Results

Preclinical animal study

First and second study: AR-SN with PET/CT

In both the first and second studies, the AR-SN system provided reliable augmentation of 3D virtual skeleton (in gray) and a semi-opaque coronal slice of PET image. Fig 2 demonstrates the experiment setup and AR projection images indicate skeleton and uptake of isotope from PET image in the liver and bladder.

Fig 2. Augmented reality guided procedure with immunocompetent CD-1 healthy mouse model using PET/CT image.

Fig 2

(A) experimental setup (B), (C) and (D) tri-planer view of coronal, sagittal and axial imaging slices respectively. d) virtual image containing 3D skeleton (gray) and coronal slice of PET image. (F) showing the image projection on the abdominal surface of dissected mouse indicate the accumulated isotope in the liver and bladder.

In the second study, the right mammary fat pad tumour was delineated with ITK-SNAP [15] and imported to the AR-SN system in addition to the PET/CT images. The content of AR projection images (Fig 3A) included fusion with semi-opaque coronal slices of both CT (in gray) and PET (in hot colour scale) images. Additionally, a surface rendering of 3D bone structures (in gray) was also projected on the surgical field. The 3D virtual tumour on the right mammary fat pad (Fig 3A, in green) distinctively highlighted the tumour location relative to the other anatomy on PET/CT slice. Fig 3B demonstrates the virtual image created through projection of a PET/CT coronal slice onto the mouse body, with the virtual tumour seen in green and the mouse skeleton seen in gray. The surgical dissection of the mouse (Fig 3C) demonstrated that the anatomical tumor location matched the location identified by the augmented image overlay on the mouse body (Fig 3D). This confirmed the correlation of the virtual tumor projection and the anatomical findings at the time of surgery and demonstrated the precision of the AR instrument in physical space.

Fig 3. Augmented reality guided procedure with the breast carcinoma xenograft model.

Fig 3

(A) projection image includes semi-opaque fused micro-CT and PET images with contour of tumor in green and skeleton of mouse in gray. (B) projection image overlaid on top of the mouse indicates the tumor location (green) and highlights isotope uptake in the liver. (C) mouse dissection demonstrating the tumor location. (D) mouse dissection with the overlaid projection image demonstrating the correct anatomical localization of the virtual contoured tumor on the mammary fat pad tumour.

Third study: AR-SN with PET/MRI

The registered PET/MRI was imported into the AR-SN system. The content of the augmented reality projection image included a fused PET/MRI coronal slice (Fig 4). The projected PET image displayed increased signal intensity signifying 64Cu-isotope uptake by the bilateral fat-pad tumour which was precisely overlaid onto the anatomical location of the tumours. The projection image also highlighted the high-level lung uptake of the 64Cu-radioisotope (Fig 4).

Fig 4. Augmented reality guided procedure with the breast carcinoma xenograft model demonstrating breast tumors in the left and right mammary fat pads.

Fig 4

(A) A registered PET/MRI image coronal slice for AR guide procedure. (B) The AR PET image projected on the mouse’s body surface highlights the tumor location. (C) The dissected mouse reveals the tumor location on the right in the same location. (D) AR projection on the mouse demonstrates the tumor location on the left side and the 64Cu isotope uptake in the lungs and liver.

Fourth study: AR-SN with SPECT/CT

SPECT/CT DICOM data was imported to the AR-SN navigation platform and fiducial markers were used to register the images with the AR platform. The content of the AR projection image included a semi-opaque fused SPECT (in rainbow) and CT (in gray) Coronal slice (Fig 5B). The axial slice and sagittal slice of the same mouse is showing in Fig 5A and 5C, respectively. In addition, the surface rendering of the full mouse skeleton (Fig 5D) created using bone segmentation was virtually projected in purple. Fig 5E also highlights the high-level isotope uptake in the liver.

Fig 5. Augmented reality guided procedure with the tumor bearing athymic CD-1 nude mouse using registered SPECT/CT image.

Fig 5

(A), (B) and (C) axial, coronal and sagittal images of fused SPECT/CT images respectively. SPECT signal is seen in rainbow and CT in gray. (D) an augmented reality projection image showing surface rendering of bone in purple combined with SPECT/CT coronal slice. (E) mouse dissection mouse demonstrating augmented reality projection SPECT-CT images overlaid on the mouse body also highlighting isotope uptake by the liver.

AR-SN system accuracy measurement

The accuracy measurement proceeds with determining the divots 3D location with and without AR guidance in a range of projection distance from 200mm to 400mm. The fiducial registration error (FRE) with 4 divots is 0.45mm. After tracker-image registration, a tracked pointer is used to localize the central of divot per described in the method section. Fiducial localization error (FLE) represent by root-mean-square (RMS) reveal the discrepancy between real divot and virtual divot location. At 200mm, 300mm and 400mm projection distance, the RMS of measured divots is 0.42 ± 0.25 mm, 0.53 ± 0.31mm and 0.7 ± 0.37 mm, respectively. The overall system RMS across various projection distances was 0.55 ± 0.33 mm. Fig 6A illustrates the checkerboard phantom partially augmented by the projection image of 3D virtual checkerboard generated from CBCT. The projection demonstrated remarkable spatial accuracy and correlation between the physical location of the phantom central divots and the projected AR image (Fig 6B). The FLE for each projection distance is summarized in Fig 6C. The AR system performed best at projection distance of 200mm and progressively worsened increasing the distance to 300 and 400 mm (Fig 6C).

Fig 6. Accuracy estimation of augmented reality navigation system.

Fig 6

(A) CBCT generated virtual checkerboard overlaid on a real checkerboard, (B) graphical representation of localization for each divot from various projection distances, (C) projection error at varies projection distance.

Discussion

This study demonstrates the feasibility and accuracy of our novel AR-SN system prototype. This system is composed of AR device fully integrated with a real-time SN system and is capable of providing target anatomical localization with an accuracy of 0.55 mm. Our implementation of AR technology distinct from existing approaches that are published in the literatures where most of the existing projection based AR device operates separately from the navigation system and incapable to track surgical instruments [2426]. Firstly, our AR system is fully migrated to surgical navigation system. Secondly, our AR system use of surgical tracking device which provide real-time localization therefore displacement of AR device does not require re-registration. S1 Video depicts the core technology of AR-SN system through artificial chest wall phantom with delineated tumour, the AR device is projecting virtual tumour and CBCT image slice according to tracked pointer location. Pointer is gradually moved to expose tumour location underneath the skin surface.

The rationale behind this novel system is that the projection of virtual images on the surgical field can provide a useful visual guide to localize the extent of tumour intraoperatively and to alert surgeons to the presence of critical structures including vasculature or nerves that may not be readily visible. The principle of the surgical safety alert is discussed in our published manuscript regarding skull base surgery [6, 16]. In developing this AR-SN system we hope to provide a tool to aid in challenging surgical ablations at high risk of incomplete resection and major complications due to complex anatomy.

Several AR devices have previously been reported in the surgical literature. The majority of these devices have been designed as HMDs, as first described in 1969 by Land and Sutherland in collaboration with the Department of Defence [27]. In 1994, advances in static and dynamic registration of an optical see-through HMD were reported [28]. Since then, optical see-through AR has been successfully implemented in a number of non-medical applications as well as more recent medical applications in the field of neurosurgery and head and neck surgery [2931]. However, HMDs have several drawbacks including the indirect view of the surgical field and extra equipment that crosses the surgeon’s sightline and/or restricts head movement due to device wiring. The add-on weight of HMDs on the surgeon head could potentially reduce strength and concentration with prolonging usage of HMDs during the course of surgery.

To address these limitations, scientists proposed using image overlay systems, which provide an alternative approach to enhance surgical visualization. Weiss et al. developed an inexpensive image overlay system composed of a LCD monitor and a half-silvered mirror mounted on a MRI scanner, which could be adapted to various angles and procedures [32]. With this method, the surgical field is augmented by superimposing the images from the translucent mirror on to the operative field. However, with this system the operator is required to be stationary and the procedure must by accomplished close to the MRI scanner which may not be clinically practical. Baum et al. recently improved the versatility of this system by miniaturizing the monitor with a tablet device capable of operating independently from the scanner [33].

The next logical step forward for surgical AR is directly projecting the desired visual information onto the operative field. However, few papers have reported the application of this concept in the field of head and neck surgery [31, 34, 35]. Our preliminary results demonstrate that our AR-SN system can reliably and accurately augment visualization intraoperatively on both phantoms and animal models, while overcoming some disadvantages of existing systems. Our system was created using light equipment that can be easily adapted to the operative setting without any interference with the surgeons’ sightline and working space. The capability of visually enhancing the position of both the tumour and critical structures on the anatomical planes could allow for both accurate tumour delineation and the prevention of major complications. Moreover, our preclinical analysis demonstrated how the prototype could be adapted to several potential imaging sources (i.e. PET/CT, PET/MRI, SPECT/CT). This aspect is critically important as it provides the surgeon with a visual image based on radiologic and nuclear medicine semiotics (i.e. contrast agent uptake, inflammation- vs neoplastic-induced changes) and enables seamless incorporation of patient-specific medical information into an image-guided procedure. Meanwhile, the multi-image modality guidance can be very helpful in surgical navigation system with AR capability.

The checkerboard phantom study provided a quantitative proof of system accuracy, with over 225 points being localized with minimal error at various projection distances (0.55 ± 0.33 mm). This sub-millimeter accuracy is consistent with navigation system requirements and comparable to reported literature [36]. Our prototype system is based on a low cost video camera that serves as a sensor for the pico-projector calibration. This calibrated camera could be additionally used as a video see-through AR device to stream surgical video to the navigation platform. This potential to combine video into the pico-projector system could further enhance the functionality of our platform as a dual AR device.

There are a wide range of potential clinical applications for this type of AR-SN system. This technology may be beneficial for a broad spectrum of surgical procedures requiring sophisticated surgical planning, precise resection, and sparing of critical structures (e.g. spine surgery, chest wall surgery, orthopedic oncological surgery, and head and neck surgery). In the field of head and neck oncology, this technology could be applied to guide complex resections especially in areas where the bony framework substantially limits the motion of soft tissues during the surgical procedure. Furthermore, this technology could help identify small volume mass such as intraparenchymal lesions. The added clinical value of this AR-SN system is currently under investigation at the hybrid preclinical/clinical Guided Therapeutics Operating Room (GTxOR–TECHNA Institute, University Health Network, Toronto, Ontario, Canada) at our institution.

Our present study does have some limitations that we hope to overcome during future technology development. Firstly, our data is based on a preclinical study with a limited number of animals. Consequently, the highly controlled operative environment of animal models may not accurately reflect the clinical setting. Meanwhile, several literatures [37] had already demonstrated the value of AR technology in clinical setting. Secondly, the surgical site and anatomical structures in the pre-clinical studies are relatively motionless. This led us to assume that this system would be applicable only to areas where soft tissue morphological changes are limited by the bony framework (i.e. maxillofacial skeleton and surrounding spaces). In fact, our system does not currently have the capacity to account for tissue deformation induced by cauterization, manipulation, and resection. Therefore, procedures involving mainly soft tissues areas that are prone to significant deformation (i.e. lung) may not benefit from this AR-SN system. Finally, image distortion due to projection on non-planar surfaces is a further limitation of projection-based AR techniques. Despite these potential limitations, we are currently in the process of translating this technology to patient studies in key surgical applications such as head and neck and orthopedic oncology to evaluate system performance under clinical conditions.

Conclusions

We have reported the development of a novel, integrated AR-SN system. This proof-of-concept study demonstrated the feasibility of our AR-SN system for multi-modality image-guided surgery in a preclinical setting. The accuracy demonstrated from the phantom study was within acceptable uncertainty. Our AR-system was found to be highly precise and capable of sub-millimeter accuracy, which is in keeping with existing commercially available SN systems. These preliminary results represent a promising framework for future technology development and eventual clinical translation.

Supporting information

S1 Fig

(TIF)

S1 Video

(MP4)

S1 File

(DOCX)

Acknowledgments

Thanks for Deborah A. Scollard, Kristin McLarty and Raymond M. Reilly’s Lab for their assistant in microSPECT/CT study.

Data Availability

All relevant data are within the manuscript and its Supporting Information files.

Funding Statement

This work is funded by the Guided Therapeutics (GTx) Program-TECHNA Institute, University Health Network, Kevin and Sandra Sullivan Chair in Surgical Oncology, Hatch Engineering Fellowship Fund, and Princess Margaret Hospital Foundation. IDEAS grant by Radiation Medicine Program, Princess Margaret Cancer Centre.

References

  • 1.Keschner D, Lee J (2010) Use of surgical navigation during endoscopic skull base surgery. Operative Techniques in Otolaryngology-Head and Neck Surgery 21: 44–50. [Google Scholar]
  • 2.Eliashar R, Sichel JY, Gross M, Hocwald E, Dano I, et al. (2003) Image guided navigation system-a new technology for complex endoscopic endonasal surgery. Postgrad Med J 79: 686–690. [PMC free article] [PubMed] [Google Scholar]
  • 3.Feichtinger M, Pau M, Zemann W, Aigner RM, Karcher H (2010) Intraoperative control of resection margins in advanced head and neck cancer using a 3D-navigation system based on PET/CT image fusion. J Craniomaxillofac Surg 38: 589–594. 10.1016/j.jcms.2010.02.004 [DOI] [PubMed] [Google Scholar]
  • 4.Catanzaro S, Copelli C, Manfuso A, Tewfik K, Pederneschi N, et al. (2017) Intraoperative navigation in complex head and neck resections: indications and limits. Int J Comput Assist Radiol Surg 12: 881–887. 10.1007/s11548-016-1486-0 [DOI] [PubMed] [Google Scholar]
  • 5.Tarsitano A, Ricotta F, Baldino G, Badiali G, Pizzigallo A, et al. (2017) Navigation-guided resection of maxillary tumours: The accuracy of computer-assisted surgery in terms of control of resection margins—A feasibility study. J Craniomaxillofac Surg 45: 2109–2114. 10.1016/j.jcms.2017.09.023 [DOI] [PubMed] [Google Scholar]
  • 6.Haerle SK, Daly MJ, Chan H, Vescan A, Witterick I, et al. (2015) Localized intraoperative virtual endoscopy (LIVE) for surgical guidance in 16 skull base patients. Otolaryngol Head Neck Surg 152: 165–171. 10.1177/0194599814557469 [DOI] [PubMed] [Google Scholar]
  • 7.Lee CY, Chan H, Ujiie H, Fujino K, Kinoshita T, et al. (2018) Novel Thoracoscopic Navigation System With Augmented Real-Time Image Guidance for Chest Wall Tumors. Ann Thorac Surg 106: 1468–1475. 10.1016/j.athoracsur.2018.06.062 [DOI] [PubMed] [Google Scholar]
  • 8.Siewerdsen JH, Daly MJ, Chan H, Nithiananthan S, Hamming N, et al. (2009) High-performance intraoperative cone-beam CT on a mobile C-arm: an integrated system for guidance of head and neck surgery. Proc SPIE 7261: 72610J. [Google Scholar]
  • 9.Rolland JP, Fuchs H (2000) Optical Versus Video See-Through Head-Mounted Displays in Medical Visualization. Presence: Teleoperators and Virtual Environments 9: 287–309. [Google Scholar]
  • 10.Ayoub A, Pulijala Y (2019) The application of virtual reality and augmented reality in Oral & Maxillofacial Surgery. BMC Oral Health 19: 238. 10.1186/s12903-019-0937-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Daly MJ, Chan H, Nithiananthan S, Qiu J, Barker E, et al. (2011) Clinical implementation of intraoperative cone-beam CT in head and neck surgery. Medical Imaging 2011: Visualization, Image-Guided Procedures, and Modeling 7964: 796426–796428. [Google Scholar]
  • 12.Enquobahrie A, Cheng P, Gary K, Ibanez L, Gobbi D, et al. (2007) The Image-Guided Surgery Toolkit IGSTK: An Open Source C++ Software Toolkit. Journal of Digital Imaging 20: 21–33. 10.1007/s10278-007-9054-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Ibanez L, Schroeder W, Ng L, Cates J The ITK software guide.
  • 14.Schroeder W, Martin K, Lorensen B The Visualization Toolkit: An Object Oriented Approach to 3D Graphics. Kitware Inc., 2004. ISBN 1-930934-19-X. [Google Scholar]
  • 15.Yushkevich PA, Piven J, Hazlett HC, Smith RG, Ho S, et al. (2006) User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage 31: 1116–1128. 10.1016/j.neuroimage.2006.01.015 [DOI] [PubMed] [Google Scholar]
  • 16.Dixon BJ, Daly MJ, Chan H, Vescan A, Witterick IJ, et al. (2014) Augmented real-time navigation with critical structure proximity alerts for endoscopic skull base surgery. Laryngoscope 124: 853–859. 10.1002/lary.24385 [DOI] [PubMed] [Google Scholar]
  • 17.Erovic BM, Daly MJ, Chan HH, James AL, Papsin BC, et al. (2013) Evaluation of intraoperative cone beam computed tomography and optical drill tracking in temporal bone surgery. Laryngoscope 123: 2823–2828. 10.1002/lary.24130 [DOI] [PubMed] [Google Scholar]
  • 18.Sternheim A, Daly M, Qiu J, Weersink R, Chan H, et al. (2015) Navigated pelvic osteotomy and tumor resection: a study assessing the accuracy and reproducibility of resection planes in Sawbones and cadavers. J Bone Joint Surg Am 97: 40–46. 10.2106/JBJS.N.00276 [DOI] [PubMed] [Google Scholar]
  • 19.Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin JC, et al. (2012) 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging 30: 1323–1341. 10.1016/j.mri.2012.05.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Lee H, Zheng J, Gaddy D, Orcutt KD, Leonard S, et al. (2015) A gradient-loadable (64)Cu-chelator for quantifying tumor deposition kinetics of nanoliposomal therapeutics by positron emission tomography. Nanomedicine 11: 155–165. 10.1016/j.nano.2014.08.011 [DOI] [PubMed] [Google Scholar]
  • 21.Chan C, Scollard DA, McLarty K, Smith S, Reilly RM (2011) A comparison of 111In- or 64Cu-DOTA-trastuzumab Fab fragments for imaging subcutaneous HER2-positive tumor xenografts in athymic mice using microSPECT/CT or microPET/CT. EJNMMI Res 1: 15. 10.1186/2191-219X-1-15 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Hudson HM, Larkin RS (1994) Accelerated image reconstruction using ordered subsets of projection data. Medical Imaging, IEEE Transactions on 13: 601–609. 10.1109/42.363108 [DOI] [PubMed] [Google Scholar]
  • 23.Chan Y, Siewerdsen JH, Rafferty MA, Moseley DJ, Jaffray DA, et al. (2008) Cone-beam computed tomography on a mobile C-arm: novel intraoperative imaging technology for guidance of head and neck surgery. J Otolaryngol Head Neck Surg 37: 81–90. [PubMed] [Google Scholar]
  • 24.Besharati Tabrizi L, Mahvash M (2015) Augmented reality-guided neurosurgery: accuracy and intraoperative application of an image projection technique. J Neurosurg 123: 206–211. 10.3171/2014.9.JNS141001 [DOI] [PubMed] [Google Scholar]
  • 25.Marmulla R, Hoppe H, Muhling J, Eggers G (2005) An augmented reality system for image-guided surgery. Int J Oral Maxillofac Surg 34: 594–596. 10.1016/j.ijom.2005.05.004 [DOI] [PubMed] [Google Scholar]
  • 26.Qu M, Hou Y, Xu Y, Shen C, Zhu M, et al. (2015) Precise positioning of an intraoral distractor using augmented reality in patients with hemifacial microsomia. J Craniomaxillofac Surg 43: 106–112. 10.1016/j.jcms.2014.10.019 [DOI] [PubMed] [Google Scholar]
  • 27.Land RI, Sutherland IE (1969) Real-time, color, stereo, computer displays. Appl Opt 8: 721–723. 10.1364/AO.8.000721 [DOI] [PubMed] [Google Scholar]
  • 28.Azuma R, Bishop G (1994) Improving static and dynamic registration in an optical see-through HMD. Proceedings of the 21st annual conference on Computer graphics and interactive techniques: ACM. pp. 197–204. [Google Scholar]
  • 29.Bajura M, Neumann U. Dynamic registration correction in augmented-reality systems; 1995 11–15 March 1995. pp. 189–196. [Google Scholar]
  • 30.Khamene A, Wacker F, Vogt S, Azar F, Wendt M, et al. (2003) An Augmented Reality system for MRI-guided needle biopsies. Stud Health Technol Inform 94: 151–157. [PubMed] [Google Scholar]
  • 31.Wong K, Yee HM, Xavier BA, Grillone GA (2018) Applications of Augmented Reality in Otolaryngology: A Systematic Review. Otolaryngol Head Neck Surg 159: 956–967. 10.1177/0194599818796476 [DOI] [PubMed] [Google Scholar]
  • 32.Weiss CR, Marker DR, Fischer GS, Fichtinger G, Machado AJ, et al. (2011) Augmented reality visualization using Image-Overlay for MR-guided interventions: system description, feasibility, and initial evaluation in a spine phantom. AJR Am J Roentgenol 196: W305–307. 10.2214/AJR.10.5038 [DOI] [PubMed] [Google Scholar]
  • 33.Baum Z, Ungi T, Lasso A, Fichtinger G (2017) Usability of a real-time tracked augmented reality display system in musculoskeletal injections: SPIE. [Google Scholar]
  • 34.Chand M, Keller DS, Devoto L, McGurk M (2018) Furthering Precision in Sentinel Node Navigational Surgery for Oral Cancer: a Novel Triple Targeting System. J Fluoresc 28: 483–486. 10.1007/s10895-018-2211-x [DOI] [PubMed] [Google Scholar]
  • 35.Marmulla Rd, Hoppe H, Mühling J, Hassfeld S (2005) New Augmented Reality Concepts for Craniofacial Surgical Procedures. Plastic and Reconstructive Surgery 115: 1124–1128. 10.1097/01.prs.0000156215.06282.76 [DOI] [PubMed] [Google Scholar]
  • 36.Gavaghan KA, Peterhans M, Oliveira-Santos T, Weber S (2011) A portable image overlay projection device for computer-aided open liver surgery. IEEE Trans Biomed Eng 58: 1855–1864. 10.1109/TBME.2011.2126572 [DOI] [PubMed] [Google Scholar]
  • 37.Eckert M, Volmerg JS, Friedrich CM (2019) Augmented Reality in Medicine: Systematic and Bibliographic Review. JMIR Mhealth Uhealth 7: e10967. 10.2196/10967 [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Domokos Máthé

2 Mar 2021

PONE-D-20-36797

An integrated augmented reality surgical navigation platform using multi-modality imaging for guidance

PLOS ONE

Dear Dr. Chan,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please consider the minor revisions required by Reviewer 1 and with those responses in view, I very much look forward to receiving the revision again.

Please submit your revised manuscript by Apr 10 2021 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Domokos Máthé

Academic Editor

PLOS ONE

Journal Requirements:

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following in the Acknowledgments Section of your manuscript:

"This work is funded by the Guided Therapeutics (GTx) Program-TECHNA Institute, University Health

347 Network, Kevin and Sandra Sullivan Chair in Surgical Oncology, Hatch Engineering Fellowship Fund, and

348 Princess Margaret Hospital Foundation. IDEAS grant by Radiation Medicine Program, Princess Margaret

349 Cancer Centre. Thanks for Deborah A. Scollard, Kristin McLarty and Raymond M. Reilly for their assistant

350 in microSPECT/CT study."

We note that you have provided funding information that is not currently declared in your Funding Statement. However, funding information should not appear in the Acknowledgments section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form.

Please remove any funding-related text from the manuscript and let us know how you would like to update your Funding Statement. Currently, your Funding Statement reads as follows:

"The author(s) received no specific funding for this work."

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

Additional Editor Comments:

Thank you for bearing with us in your continued patience as so many reviewers could not cope with the multi-disciplinary nature of your manuscript. I myself, having been able to participate in previous similar medical device developments also very much appreciate the manuscript and its results. Please kindly refer to the requests of the one reviewer requiring minor, albeit important improvements, too.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: I Don't Know

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: As I know this is the firts proof-of-concept study which demonstrated the feasibility of a system for multi-modality image-guided surgery in a preclinical setting. This methods may be a very important testing platform of the development of image-guided surgery especially the spatial precision which will be very helpful in par example neurosurgery.

I have some questions:

1) Fig 3. shows the result of projection of the PET/SPEC/CT image of the tumor on the skin and tomour of the mouse. But the image (see Fig 6 B-D) is obviously mistaken because the image or tumour is projected to the bed. How could you develop the projector to reduce this mistake?

2) Based on the description, the spatial resolution of the system is not clear. But Fig 6. try to visualise the arrangement but the mistake of measurement along the x-y-z axis is not shown. How could you determine these parameters?

3) How could you develope a rendering methods to reduce the real resolution problem of the molecular imaging modality? How could joint to the PET/CT images also to reduce this type problems?

4) I suggest to specify the limitation of this proof-of-concept very thoroughly and please comlete how you will solve these.

Reviewer #2: Authors demonstrated that the feasibility of their novel augmented reality system with pico-projector which was fully integrated with a real-time surgical navigation system. They also showed how their prototype could be adapted to several different imaging sources (PET/CT, SPECT/CT, PET/MRI). Their phantom study provided a quantitative proof of system accuracy with minimal error at various projection distances which might be acceptable in real clinical setting.

As they mentioned, there are some limitations. I think that major drawback would be image distortion on non-planar surface and morphological tissue change during operation. So this technology could be applied to relatively hard and non-deformable organ such as brain and bony framework.

Even if there are some limitations, this article would be readable for surgeons or physicians who are interested in image guided surgery.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2021 Apr 30;16(4):e0250558. doi: 10.1371/journal.pone.0250558.r002

Author response to Decision Letter 0


31 Mar 2021

PONE-D-20-36797

An integrated augmented reality surgical navigation platform using multi-modality imaging for guidance

Response to reviewers:

Reviewer #1: As I know this is the first proof-of-concept study which demonstrated the feasibility of a system for multi-modality image-guided surgery in a preclinical setting. This methods may be a very important testing platform of the development of image-guided surgery especially the spatial precision which will be very helpful in par example neurosurgery.

I have some questions:

1) Fig 3. shows the result of projection of the PET/SPEC/CT image of the tumor on the skin and tumor of the mouse. But the image (see Fig 6 B-D) is obviously mistaken because the image or tumour is projected to the bed. How could you develop the projector to reduce this mistake?

Thank you for your comment. We agree with the reviewers regarding the misleading information of the virtual tumor being projected onto the mouse’s bedding. Indeed, a portion of the tumor is behind the imaging bed with respect to the projection direction. The AR-SN system follows the instruction correctly projecting the virtual tumor below the projection surface. However, the obstruction in the space between the projector’s prospective to the subject’s surface may not be corrected by the AR-SN system. The attached figure (axial slice from same pre-clinical animal study) demonstrates the cause of the misleading projection on the imaging bed and fiducial marker. Additionally, the semi-translucent 3D printed fiducial marker is diffusing projection light to the surrounding area that further extends the projection error. We have recently started a comprehensive upgrade of our AR-SN system. One of the enhanced abilities will be projecting registration landmarks which will hopefully help to reduce errors in projection. The new abilities of our system allows for fine adjustments of the AR image including scale factor, horizontal and vertical shift. Prior to using the AR-SN system, the user will be able to project registration landmarks and check their position with respect to fiducial markers and makes any image adjustment if require. This fine adjustment function will able to mitigate systematic error. Regarding the surface topology-induced projection error, if we consider the size of tumors in humans as well as the clinical scenarios in which AR-SN would be used (i.e., projection of a tumor onto a rather flat surgical field to provide surgeons with an on-the-field representation of tumor extension) the image distortion will likely be very minimal.

2) Based on the description, the spatial resolution of the system is not clear. But Fig 6. try to visualize the arrangement but the mistake of measurement along the x-y-z axis is not shown. How could you determine these parameters?

Thank you for your question. We agree with the reviewers comments on error measurement. The sentence below is amended in the manuscript (lines 176 to 180) to explain further how to compute root-mean-square (RMS) from the system (x,y,z) uncertainty.

“The uncertainty of the projected location was calculated as follows: (xi, yi, zi)real represents as “real” location of the divot center acquired manually without AR image guidance and (xi, yi, zi)virtual represents as “virtual” location of the divot generated from the AR projection image, the estimation of error is the distance between the “real” and “virtual” location of the central divot such that Err=‖(x_i,y_i,z_i )_real-(x_i,y_i,z_i )_virtual ‖.”

3) How could you develop a rendering methods to reduce the real resolution problem of the molecular imaging modality? How could joint to the PET/CT images also to reduce this type problems?

Thank you for your interesting question. The PET/CT image is fused by rigid registration using normalized mutual information. The opacity of the fused image can be adjusted in order to highlight either the PET signal or CT image. To reduce real resolution problems, the voxel value is interpolated in order to mitigate pixilation due to low the resolution of the PET image. Additionally, a Gaussian smoothing filter can be applied to the surface rendering of virtual objects to improve resolution. However, these proposed solutions will only able to minimize and not completely resolve the issue of low image resolution.

4) I suggest to specify the limitation of this proof-of-concept very thoroughly and please complete how you will solve these.

Thank you for reviewer’s suggestion. In the manuscript from line 317 to 330 the limitations of our AR-SN system have been specified. The major challenge in the development of the AR-SN system is projecting the correct image onto an irregular three-dimensional surface as the image can become distorted. In machine vision, methods to correct this image distortion have been widely investigated. Camera calibration of intrinsic / extrinsic parameters is one of the common approaches to which can used to rectify distorted images. A similar approach can be applied to try to correct the projected image. However, the fundamental problem of image distortion originates from the surface topology of the subject. In order to correct this distortion, the surface topology needs to be characterized. This can be achieved by segmenting the surface from the CT image then interpolating the surface by best fitting B-spline or non-uniform rational B-splines (NURES). Once the topology of the surface is formulated, the distortion can be corrected by mathematically mapping the individual pixels to the formulated surface. This described approach is computationally intensive. Our AR-SN system has a refresh rate of 30 fps and thus the system may experience lag time issues with this approach. To minimize the lag problem, graphics processing unit (GPU) implementation can be considered.

Reviewer #2: Authors demonstrated that the feasibility of their novel augmented reality system with pico-projector which was fully integrated with a real-time surgical navigation system. They also showed how their prototype could be adapted to several different imaging sources (PET/CT, SPECT/CT, PET/MRI). Their phantom study provided a quantitative proof of system accuracy with minimal error at various projection distances which might be acceptable in real clinical setting.

As they mentioned, there are some limitations. I think that major drawback would be image distortion on non-planar surface and morphological tissue change during operation. So this technology could be applied to relatively hard and non-deformable organ such as brain and bony framework. Even if there are some limitations, this article would be readable for surgeons or physicians who are interested in image guided surgery.

Thank you for your valuable comments and feedbacks. We realize that the limitation of our prototype system is that it may underperform when the image is projected onto deformable soft tissue. We are currently investigating potential solutions to this problem to improve the utility of our system.

Attachment

Submitted filename: ResponseToReviewers.docx

Decision Letter 1

Domokos Máthé

12 Apr 2021

An integrated augmented reality surgical navigation platform using multi-modality imaging for guidance

PONE-D-20-36797R1

Dear Dr. Chan,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Domokos Máthé

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Domokos Máthé

22 Apr 2021

PONE-D-20-36797R1

An integrated augmented reality surgical navigation platform using multi-modality imaging for guidance

Dear Dr. Chan:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Domokos Máthé

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Fig

    (TIF)

    S1 Video

    (MP4)

    S1 File

    (DOCX)

    Attachment

    Submitted filename: ResponseToReviewers.docx

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting Information files.


    Articles from PLoS ONE are provided here courtesy of PLOS