Accuracy of Total Hip Arthroplasty Templating Using Set Calibration Magnifications

Benjamin Schapira 1,, Suroosh Madanipour 2, Farhad Iranpour 1, Padmanabhan Subramanian 1
Editors: Alexander Muacevic, John R Adler
PMCID: PMC10011871  PMID: 36925986

Abstract

Background

Templating for total hip arthroplasty has been adopted over recent decades as a reliable and accurate method for pre-operative planning. The use of calibration markers for this process provides a recognised benefit at the expense of cost, availability and error. Many surgeons use a set magnification of 118% to account for calibration errors when templating total hip arthroplasty. This study aims to assess the accuracy of templating with standardised magnifications and assess the effect of BMI on templating accuracy.

Materials and methods

A retrospective analysis was performed using a single-surgeon series of 119 consecutive total hip arthroplasties. Anteroposterior radiographs were taken pre- or post-operatively without calibration hardware. Pre-operatively, the total hip arthroplasty was templated on TraumaCad (BrainLab Inc, Westchester, IL) using either 118% or 119% calibration magnification. Post-operative magnification was calibrated using the known femoral head diameter. Templated and implanted prostheses were compared for size.

Results

At 118%, 61.1% of cups matched those templated with 96.3% of cups within two sizes. At 119%, 52.5% of cups used matched their templates with 100% within two sizes. There was no significant difference between 118% and 119% cup size prediction (p=0.49). A trend was noticed in increasing magnification error with increasing BMI. However, BMI had no significant effect on the accuracy of templating cup size within two cup sizes (p=0.58).

Conclusion.

Templating acetabular cups using a set magnification of 118% or 119% yields accurate results and provides a reliable method to template without calibration equipment. Whilst BMI can affect magnification error, this has no significant effect on the accuracy of implanted cups and stems within two sizes.

Keywords: preoperative templating, calibration, magnification, tha, total hip arthroplasty: tha

Introduction

Templating for total hip arthroplasty (THA) has been adopted over recent decades as a reliable and accurate method for pre-operative planning [1]. Benefits have been noted in improved anatomical planning [2,3], operative risk assessment [4,5] and estimation of implant choice [6,7]; with subsequent benefits for cost and implant longevity [8,9].

Despite increasing popularity, studies show relatively poor uptake for templating THAs. In 2011, one study found that amongst NHS trusts in South East England, only 14/28 (50%) had access to digital templating software, with just one of these trusts offering training [10]. None of the trusts regularly templated or calibrated for magnification in pre-operative planning. Despite these figures, templating has been proven to enhance surgical accuracy for all levels of training, up to the most experienced surgeons [11].

Current techniques involve expensive external calibration markers (ECMs) which often present great variation from true magnification, amounting to as much as 6.8% error or a difference of +/- 3mm in acetabular cup sizes [7,12]. These results insist markers cannot accurately predict the true magnification in pre-operative templating [6].

Given the inaccuracies and ease of misplacing ECMs, some surgeons have opted to template without such equipment and elect a standard magnification for all templating (set magnification), the proponents of which claim greater estimation of magnification, reduced cost of calibration hardware and improved templating accuracy [6,13]. The primary objectives of this study were to identify whether set magnifications for digital templating could accurately enable surgeons to estimate prosthesis size. We also aim to investigate the effects of BMI on magnification and subsequent templating accuracy. This article was previously presented as a meeting abstract at the 2021 EFORT (European Federation of National Associations of Orthopaedics and Traumatology) annual virtual congress.

Materials and methods

A retrospective analysis was performed in accordance with the STROBE (Strengthening the Reporting of Observational Studies in Epidemiology) and STARD (Standards for Reporting Diagnostic Accuracy) checklists, using a single-surgeon series of 136 consecutive THAs from April 2019-November 2020. All cases took place within the same centre during the senior-most consultants’ tenure. The mean age of patients was 70.1 years (standard deviation 11.47 years) and 56.3% of patients were female. This study followed the usual protocols of investigation, intervention and retrospective analysis of our trust's Quality Governance Team with respect to ethical approval and patient consent. All data was kept anonymised and followed in accordance with the 1975 declaration of Helsinki and its 2008 revision. Indications for primary THA included osteoarthritis, rheumatoid arthritis, avascular necrosis, dysplasia and fractured neck of the femur. Exclusion criteria included revision surgery, absence of pre-operative template and where pre-operative BMI was not recorded.

Anteroposterior (AP) pelvic radiographs were taken pre-operatively and 0-2 days post-operatively using no calibration hardware (calibration balls or mats) with a standardised radiographic protocol followed in our trust. All radiographs conformed to the same protocol using the Samsung XGEO GC80 series (Samsung Electronics Co., Ltd, Yeongtong-gu, Suwon, South Korea). Patients were positioned supine with feet internally rotated to optimise symmetry. The tube-detector distance was routinely set to 100cm with the radiographic beam aimed at the level of the greater trochanters. BMI was recorded as part of the pre-operative assessment. 

The predominant prostheses used in templating were the CPT and Trilogy systems (Zimmer, Warsaw, IN, US) with a minority using the G7 dual-mobility systems (Zimmer) in those with significant pre-operative risk factors for dislocation. All cases used CPT femoral stems. Templating aimed to restore the patient’s native anatomy using the acetabular teardrop as a reference point for cup placement. The femoral polished tapered stem component was sized to allow for appropriate restoration of the patient’s native offset, leg length and caput-collum-diaphyseal angle with a 2mm cement mantle. Planned acetabular cup size, stem size and offset were recorded in retrospect from the template (Figure 1). A posterior approach was utilised using a standard incision. Intra-operatively, the template was used as a guide and the surgeon chose the final prosthesis with complete autonomy.

Figure 1. Pre-operative anteroposterior pelvic radiograph of a 45-year-old female on TraumaCad software calibrated to 118% magnification demonstrating templating technique.

Figure 1

TraumaCad (BrainLab Inc, Westchester, IL)

Pre-operatively, the THA was templated on TraumaCad (BrainLab Inc, Westchester, IL) software using either 118% or 119% calibration magnification (unless the contralateral side had a THA with a known femoral head diameter). Pre-operative template magnification was chosen at random. All cases were templated and operated upon by a single consultant hip and knee arthroplasty surgeon.

Post-operative radiographs were reviewed on TraumaCad by a single-blinded observer who was not involved with initial templating. Magnification was calibrated using the known diameter of the femoral head (true magnification) (Figure 2). Most cases with Trilogy acetabular cups used 36mm femoral heads, with three cases using 32mm and where a dual mobility component was used, a 28mm liner head was used (four cases). The true cup size and stem length were recorded from the operative notes and compared with the template choice. The true magnification was compared with the set magnification used pre-operatively to calculate the magnification error (pre-operative set magnification - post-operative true magnification).

Figure 2. Post-operative anteroposterior pelvic radiograph visualised of a 45-year-old female on TraumaCad templating software for calibration of magnification using the known size of the femoral head.

Figure 2

Statistical analysis

Statistical analyses were performed on IBM SPSS Statistics Software version 26 (IBM Corp, Armonk, NY, US). The chi-square and Fisher’s exact test were used to compare categorical data. Independent t-tests and Mann-Whitney U tests were used to compare continuous results. A p-value of <0.05 was considered statistically significant.

Results

In total, 136 THA cases were reviewed, of which 12 revision cases and five further cases with no pre-operative template were excluded. After all exclusions, 119 cases remained for analysis. 

The median cup size templated was 54mm (range 46-64mm), the median stem templated was CPT size 1 (range 0-3) and the median modular neck option templated was the extended size (+5mm additional offset). The median cup size used was 54mm (range 44-66mm) using Trilogy in 96 cases and G7 in 23 and the median stem size used was CPT size 1 (range 0-4). 

Retrospective analysis of all pre-operative radiographs found 54 cases were calibrated at 118% and 40 cases at 119% magnification. In cases templated to 118%, post-operative radiographic magnification matched the pre-operative template in 42.6% of cases. This increased to 74.1% within a 2% error. In cases templated with 119% magnification, post-operative radiographic magnification matched the pre-operative template in 12.5% of cases and 70% within 2% error.

At 118% magnification, 61.11% of cases used the same cup size templated with 96.3% within two sizes of the template. At 119% magnification, 52.5% of cases used the same cup size templated with 100% within two sizes of the template (Figure 3). There was no significant difference between 118% and 119% cup size prediction (p=0.49). 

Figure 3. A graph presenting the difference in templated and implanted acetabular cup size when templating at 118% and 119% magnification (2mm = 1 cup size).

Figure 3

The implanted femoral stem size matched the template size at 74.1% at 118% template magnification and 65.8% at 119% with both increasing to 100% within two stem sizes (Figure 4). 

Figure 4. A graph presenting the difference in templated and implanted femoral stems when templating at 118% and 119% magnification. At both template magnifications, all stems implanted were accurate to within two sizes.

Figure 4

In total, seven cases did not have pre-operative BMI recorded. The mean BMI in our study was 30.42kg/m2 with 22 cases (20.75%) above 35kg/m2 and five cases (4.72%) above 40kg/m2. A trend was noticed in increasing magnification error with increasing BMI. However, BMI had no significant effect on the accuracy of templating the acetabulum within two cup sizes (p=0.58) (Figure 5). Where implanted cup or stem size differed by more than two sizes from the template, BMI ranged from 19.5-22.8 kg/m2 (Figure 6). 

Figure 5. A plot of the difference in calibration magnification between pre and post-operative radiographs against patient BMI. The line of best fit represents the relationship between increasing BMI and increasing magnification error.

Figure 5

Figure 6. (A) A plot demonstrating the change in acetabular cup size with BMI across all cases (one cup size = 2mm). (B) A plot demonstrating the change in femoral stem size with BMI across all cases.

Figure 6

Discussion

To our knowledge, this is the very first study to research the accuracy of different set magnifications in digital THA templating. Our results proved high accuracy and low magnification error at both 118% and 119% for both acetabular cup and femoral stem template predictions. These suggest templating at a set magnification without ECMs will provide reliable and accurate pre-operative planning.

Very few studies investigate the accuracy of templating without magnification markers. One study presented exact cup size selection in 86% and 64% in stem selection when set magnification was estimated at 116% [14]. Si et al. reported exact cup prediction in 43.3% and stem prediction in 40% of cases, improving to 95.56% and 93.33% within two sizes respectively, without an ECM [5]. Our results compare favourably with studies reporting calibration errors of 1.42% using a set magnification of 121% compared with errors as high as 2.55% with ECMs [15]. At 118% there was a propensity to overestimate the cup template by one size with a non-significant yet notable underestimation of one size at 119%. This trend was also noted by Manrique et al. who reported a greater number of one-size-smaller errors when templating at 121% [16]. They concluded the optimal magnification for pre-operative templating to be between 118-119.5%, a best-fit magnification constant also noted in other research [17]. Our study found no significant difference in accuracy between 118% and 119%. 

In contrast, studies templating with ECMs report exact prosthesis matching in 32% of acetabular cups and 43% of femoral stems, improving to 92% and 96% within two sizes of the template respectively [18]. Shaarani et al. too found similar results with exact cup choice in 38% and stem choice in 36%, both increasing to 98% within two sizes [4].

These results suggest ECM calibration risks positioning inconsistency, magnification error [19] and tolerance for excessive size miscalculations [12]. Studies agree the optimal location for ECM placement to be the anatomical centre of rotation (COR) (centre of the femoral head) [20]. Given the infeasibility, there is great debate over the best replication of this position [12,19,21]. Coronal deviation towards the source/away from the plate overestimates the magnification and vice versa [6]. This was exemplified by Ramme et al. who found an ECM AP deviation of 3.5 cm to result in one full cup size change [22]. Similarly, a lateral shift from the midline in the sagittal plane introduced as much as 7% magnification error [23] or 2% for every 17.5cm deviation, dependent on vertical height [19]. These figures explain the significant variation in magnification errors with ECMs and confirm greater accuracy when calibrating at set magnification [6,13].

The effects of BMI on magnification error and subsequent accuracy in implant size estimation are under-reported in the literature. Our results suggest a significant trend in increasing magnification error with increasing BMI, trending towards over-estimating magnification. This finding is echoed by other studies describing a proportional increase in magnification with the distance between the hip and detector plate [19,24], a key confounder in the accuracy of ECM positioning [25]. However, despite the effects on magnification errors, our results prove BMI did not have a significant impact on templating implant size. Few studies investigate the effect of BMI on template accuracy but conclude no significant correlation between template size inaccuracy and BMI of normal, overweight and obese patients [9,26].

We defined accurate acetabular implant choice within two sizes based on previous investigations and our own clinical judgment. Studies debate "accurate sizing" to be within one [12] or two sizes of template [4,26] (2mm difference) in lieu of intra- and post-operative risks. Evidence suggests there is no compromise in the post-operative clinical score when selecting intra-operative implants at one or more size differences from the template [1]. Thus, we believe accurate template predictions to be within two sizes of the prosthesis used. 

We used AP radiographs filmed in the supine position using our trusts’ standardised protocol to address inaccuracies of poorly imaged anatomy [20,27] and positioning effects [3,18]. For other unknowns such as bone quality, posterior acetabular integrity and femoral torsion, we could only hypothesise on their effect. Other imaging modalities, such as 3-D computerised tomography (CT) have demonstrated superior ability in measuring these variables and templating exact implant sizes [20]. These however come at the expense of greater radiation exposure and cost [28], and therefore require strict risk-benefit analysis for justification.

The templating itself may have been influenced by the skill and experience of the consultant surgeon. There is conflict as to how much experience truly affects personal templating ability [29]. However, most studies report little or no difference in templating accuracy amongst all grades of trainee and consultant, presuming they have received adequate training [9,21].

There were several limitations in our study. Data were retrospectively reviewed. To address selection bias, all cases analysed were operated by a single surgeon consecutively and to reduce recall bias, every template and post-operative document was reviewed twice with documentation of errors. The surgeon was not blinded to their template and as such could have been biased towards the final implant choice. However, if different surgeons templated and operated, a difference in opinions/technique could have nullified the template. The operating surgeon was not bound by his template and was at liberty to make intra-operative adaptations.

Conclusions

In conclusion, this is the first study to directly compare set magnification at two different calibrations. These results confirm that templating THA surgery with a set magnification is a reliable technique for pre-operative planning without the need for calibration equipment. We believe the optimal magnification to be 118-119%; however, we understand this figure would differ between trusts and radiographic systems. Whilst BMI can affect magnification error, this has no significant effect on the accuracy of implanted cups or stems within two sizes and as such, surgeons should feel confident with pre-operative planning via this technique.

The content published in Cureus is the result of clinical experience and/or research by independent individuals or organizations. Cureus is not responsible for the scientific accuracy or reliability of data or conclusions published herein. All content published within Cureus is intended only for educational, research and reference purposes. Additionally, articles published within Cureus should not be deemed a suitable substitute for the advice of a qualified health care professional. Do not disregard or avoid professional medical advice due to content published within Cureus.

The authors have declared that no competing interests exist.

Human Ethics

Consent was obtained or waived by all participants in this study

Animal Ethics

Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.

References

  • 1.Pre-operative planning in THA. Part III: do implant size prediction and offset restoration influence functional outcomes after THA? Cech A, Kase M, Kobayashi H, et al. Arch Orthop Trauma Surg. 2020;140:563–573. doi: 10.1007/s00402-020-03342-5. [DOI] [PubMed] [Google Scholar]
  • 2.Digital templating for THA: a simple computer-assisted application for complex hip arthritis cases. Hafez MA, Ragheb G, Hamed A, Ali A, Karim S. Biomed Tech (Berl) 2016;61:519–524. doi: 10.1515/bmt-2015-0009. [DOI] [PubMed] [Google Scholar]
  • 3.Digital templating in total hip arthroplasty: Additional anteroposterior hip view increases the accuracy. Stigler SK, Müller FJ, Pfaud S, Zellner M, Füchtmeier B. World J Orthop. 2017;8:30–35. doi: 10.5312/wjo.v8.i1.30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Accuracy of digital preoperative templating in 100 consecutive uncemented total hip arthroplasties: a single surgeon series. Shaarani SR, McHugh G, Collins DA. J Arthroplasty. 2013;28:331–337. doi: 10.1016/j.arth.2012.06.009. [DOI] [PubMed] [Google Scholar]
  • 5.Accuracy of a simple digital templating in primary uncemented total hip arthroplasty. Si HB, Zeng Y, Cao F, Pei FX, Shen B. Chin Med Sci J. 2015;30:150–155. doi: 10.1016/s1001-9294(15)30039-0. [DOI] [PubMed] [Google Scholar]
  • 6.Inaccuracies in the use of magnification markers in digital hip radiographs. Archibeck MJ, Cummins T, Tripuraneni KR, Carothers JT, Murray-Krezan C, Hattab M, White RE Jr. Clin Orthop Relat Res. 2016;474:1812–1817. doi: 10.1007/s11999-016-4704-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Assessment of accuracy of marker ball placement in pre-operative templating for total hip arthroplasty. Sinclair VF, Wilson J, Jain NP, Knowles D. J Arthroplasty. 2014;29:1658–1660. doi: 10.1016/j.arth.2014.03.013. [DOI] [PubMed] [Google Scholar]
  • 8.Templating hip arthroplasty. Alnahhal A, Aslam-Pervez N, Sheikh HQ. Open Access Maced J Med Sci. 2019;7:672–685. doi: 10.3889/oamjms.2019.088. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.The accuracy of digital templating in uncemented total hip arthroplasty. Holzer LA, Scholler G, Wagner S, Friesenbichler J, Maurer-Ertl W, Leithner A. Arch Orthop Trauma Surg. 2019;139:263–268. doi: 10.1007/s00402-018-3080-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Digital templating in total hip arthroplasty in the NHS: ‘computer says no…’. Jeyaseelan L, Ward J, Eleftheriou K, Cobb J. Eur J Orthop Surg Traumatol. 2011;21:397–399. [Google Scholar]
  • 11.Hip offset and leg length equalization in direct anterior approach total hip arthroplasty without preoperative templating. Hasegawa I, Wright AR, Andrews SN, Unebasami E, Nakasone CK. https://pubmed.ncbi.nlm.nih.gov/31773108/ Hawaii J Health Soc Welf. 2019;78:26–28. [PMC free article] [PubMed] [Google Scholar]
  • 12.Influence of calibration on digital templating of hip arthroplasty. Boese CK, Wilhelm S, Haneder S, Lechler P, Eysel P, Bredow J. Int Orthop. 2019;43:1799–1805. doi: 10.1007/s00264-018-4120-7. [DOI] [PubMed] [Google Scholar]
  • 13.Prospective, randomized, surgeon-blinded comparison of standard magnification assumption vs magnification marker usage for preoperative templating in total hip arthroplasty. Archibeck MJ, Tripuraneni KR, Carothers JT, Junick DW, Munson NR, Murray-Krezan CM. J Arthroplasty. 2017;32:3061–3064. doi: 10.1016/j.arth.2017.05.016. [DOI] [PubMed] [Google Scholar]
  • 14.Digital templating facilitates accurate leg length correction in total hip arthroplasty. Jassim SS, Ingham C, Keeling M, Wimhurst JA. https://pubmed.ncbi.nlm.nih.gove/22822575/ Acta Orthop Belg. 2012;78:344–349. [PubMed] [Google Scholar]
  • 15.A comparison of four systems for calibration when templating for total hip replacement with digital radiography. Franken M, Grimm B, Heyligers I. J Bone Joint Surg Br. 2010;92:136–141. doi: 10.1302/0301-620X.92B1.22257. [DOI] [PubMed] [Google Scholar]
  • 16.Total hip templating using a marker ball (vs. 21% magnification): no difference in accuracy but increase in potential harm. Manrique J, Battenberg A, Foltz C, Goswami K, Hozack W. Curr Orthop Pract. 2019;30:307–311. [Google Scholar]
  • 17.Preoperative sizing of hip hemiarthroplasties to accurately estimate head size from non- standardised pelvic radiographs: Can it be done? Crosswell S, Akehurst H, Ramiah R, Navadgi B. Injury. 2019;50:2030–2033. doi: 10.1016/j.injury.2019.08.025. [DOI] [PubMed] [Google Scholar]
  • 18.A novel digital templating methodology for arthroplasty: experience from patients with osteonecrosis of the femoral head. Dong N, Yang C, Li SQ, Gao YH, Liu JG, Qi X. Hip Int. 2017;27:82–86. doi: 10.5301/hipint.5000427. [DOI] [PubMed] [Google Scholar]
  • 19.Calibration markers for digital templating in total hip arthroplasty. Boese CK, Lechler P, Rose L, et al. PLoS One. 2015;10:0. doi: 10.1371/journal.pone.0128529. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Variance in predicted cup size by 2-dimensional vs 3-dimensional computerized tomography-based templating in primary total hip arthroplasty. Osmani FA, Thakkar S, Ramme A, Elbuluk A, Wojack P, Vigdorchik JM. Arthroplast Today. 2017;3:289–293. doi: 10.1016/j.artd.2016.09.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Effect of training level on accuracy of digital templating in primary total hip and knee arthroplasty. Hsu AR, Kim JD, Bhatia S, Levine BR. Orthopedics. 2012;35:0–83. doi: 10.3928/01477447-20120123-15. [DOI] [PubMed] [Google Scholar]
  • 22.Scaling marker position determines the accuracy of digital templating for total hip arthroplasty. Ramme AJ, Fisher ND, Egol J, Chang G, Vigdorchik JM. HSS J. 2018;14:55–59. doi: 10.1007/s11420-017-9578-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Comparison of two scaling methods in preoperative digital templating of total hip replacement. Leung KHL, Mak J, Lee OB, Tsang WL, Khoo J. J Orthop Trauma Rehabilitation. 2015;19:31–33. [Google Scholar]
  • 24.Accurate scaling of digital radiographs of the pelvis. A prospective trial of two methods. Wimsey S, Pickard R, Shaw G. J Bone Joint Surg Br. 2006;88:1508–1512. doi: 10.1302/0301-620X.88B11.18017. [DOI] [PubMed] [Google Scholar]
  • 25.A simple and convenient method guide to determine the magnification of digital X-rays for preoperative planning in total hip arthroplasty. Heep H, Xu J, Löchteken C, Wedemeyer C. Orthop Rev (Pavia) 2012;4:0. doi: 10.4081/or.2012.e12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Effect of body mass index on digital templating for total hip arthroplasty. Sershon RA, Diaz A, Bohl DD, Levine BR. J Arthroplasty. 2017;32:1024–1026. doi: 10.1016/j.arth.2016.09.020. [DOI] [PubMed] [Google Scholar]
  • 27.Accuracy of the preoperative planning for cementless total hip arthroplasty. A randomised comparison between three-dimensional computerised planning and conventional templating. Sariali E, Mauprivez R, Khiami F, Pascal-Mousselard H, Catonné Y. Orthop Traumatol Surg Res. 2012;98:151–158. doi: 10.1016/j.otsr.2011.09.023. [DOI] [PubMed] [Google Scholar]
  • 28.Computed tomography for preoperative planning in minimal-invasive total hip arthroplasty: radiation exposure and cost analysis. Huppertz A, Radmer S, Asbach P, et al. Eur J Radiol. 2011;78:406–413. doi: 10.1016/j.ejrad.2009.11.024. [DOI] [PubMed] [Google Scholar]
  • 29.Predictive value of preoperative digital templating in THA depends on the surgical experience of the performing physician. Mittag F, Ipach I, Schaefer R, Meisner C, Leichtle U. Orthopedics. 2012;35:0–7. doi: 10.3928/01477447-20120123-14. [DOI] [PubMed] [Google Scholar]

Articles from Cureus are provided here courtesy of Cureus Inc.