Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2023 Mar 16;18(3):e0281127. doi: 10.1371/journal.pone.0281127

Artificial intelligence-based point-of-care lung ultrasound for screening COVID-19 pneumoniae: Comparison with CT scans

Yumi Kuroda 1, Tomohiro Kaneko 2, Hitomi Yoshikawa 1, Saori Uchiyama 2, Yuichi Nagata 1, Yasushi Matsushita 3,4, Makoto Hiki 2,3, Tohru Minamino 2, Kazuhisa Takahashi 1, Hiroyuki Daida 2,5, Nobuyuki Kagiyama 2,5,*
Editor: Juan Antonio Valera-Calero6
PMCID: PMC10019704  PMID: 36928805

Abstract

Background

Although lung ultrasound has been reported to be a portable, cost-effective, and accurate method to detect pneumonia, it has not been widely used because of the difficulty in its interpretation. Here, we aimed to investigate the effectiveness of a novel artificial intelligence-based automated pneumonia detection method using point-of-care lung ultrasound (AI-POCUS) for the coronavirus disease 2019 (COVID-19).

Methods

We enrolled consecutive patients admitted with COVID-19 who underwent computed tomography (CT) in August and September 2021. A 12-zone AI-POCUS was performed by a novice observer using a pocket-size device within 24 h of the CT scan. Fifteen control subjects were also scanned. Additionally, the accuracy of the simplified 8-zone scan excluding the dorsal chest, was assessed. More than three B-lines detected in one lung zone were considered zone-level positive, and the presence of positive AI-POCUS in any lung zone was considered patient-level positive. The sample size calculation was not performed given the retrospective all-comer nature of the study.

Results

A total of 577 lung zones from 56 subjects (59.4 ± 14.8 years, 23% female) were evaluated using AI-POCUS. The mean number of days from disease onset was 9, and 14% of patients were under mechanical ventilation. The CT-validated pneumonia was seen in 71.4% of patients at total 577 lung zones (53.3%). The 12-zone AI-POCUS for detecting CT-validated pneumonia in the patient-level showed the accuracy of 94.5% (85.1%– 98.1%), sensitivity of 92.3% (79.7%– 97.3%), specificity of 100% (80.6%– 100%), positive predictive value of 95.0% (89.6% - 97.7%), and Kappa of 0.33 (0.27–0.40). When simplified with 8-zone scan, the accuracy, sensitivity, and sensitivity were 83.9% (72.2%– 91.3%), 77.5% (62.5%– 87.7%), and 100% (80.6%– 100%), respectively. The zone-level accuracy, sensitivity, and specificity of AI-POCUS were 65.3% (61.4%– 69.1%), 37.2% (32.0%– 42.7%), and 97.8% (95.2%– 99.0%), respectively.

Interpretation

AI-POCUS using the novel pocket-size ultrasound system showed excellent agreement with CT-validated COVID-19 pneumonia, even when used by a novice observer.

Introduction

The novel coronavirus disease 2019 (COVID-19) emerged in China in 2019, and has since spread rapidly worldwide [1]. This highly infectious disease, which often causes fatal pneumonia, has caused 6 million deaths as of March 2022 [24]. This rapid increase in the number of patients with COVID-19 has strained medical resources in many countries [5, 6]. In such situations, detailed screening in medical facilities for all patients is not realistic; as such, there is an increasing demand for portable, easy, and cost-effective screening tools.

Point-of-care lung ultrasound (POCUS) is a simple diagnostic test with high diagnostic accuracy for lung disease [7, 8]. A previous systematic review reported that the presence of multiple B-lines could identify the presence of pneumonia with a sensitivity of 90.4%, and a specificity of 88.4% [9]. Although this examination has the potential to serve as a useful screening tool in COVID-19 patients, lung POCUS is not sufficiently widespread, mainly because of technical difficulties in image interpretation.

Artificial intelligence, especially machine learning including deep learning, is an emerging computer technology that automatically recognizes images with excellent accuracy [1012]. This technology may have the potential to help interpret medical images and enable such examinations, even by novice observers. In this study, we investigated the effectiveness of a novel artificial intelligence-based automated B-line counting system for detecting COVID-19 pneumonia using a pocket-size ultrasound device [13].

Methods

Patient enrollment

We enrolled consecutive patients with COVID-19 who were admitted to our hospital and underwent computed tomography (CT) between August and September 2021. Lung POCUS using a pocket-size ultrasound device (Lumify; Philips Ultrasound, Inc., Bothell, WA) with a novel machine learning-based application (guided B-line quantification; Philips Ultrasound, Inc., Bothell, WA) was performed in all patients within 24 h of the CT scan (artificial intelligence-based point-of-care lung ultrasound; AI-POCUS). This machine learning program was developed by the company and is now commercially available. Detailed development process (training, validation, model architecture, etc.) is not publicly open. Patients aged <20 years and pregnant women were excluded. We further performed AI-POCUS in 15 consecutive subjects without lung diseases who underwent chest CT for the purpose of medical checkup, as controls.

AI-POCUS and CT protocols

AI-POCUS was performed in a 12-zone manner: eight zones for the ventral and four zones for the dorsal chest wall using a 1.0–4.0 Hz sector probe (S4-1; Philips Ultrasound, Inc., Bothell, WA), with a depth setting of 6–12 cm [8, 14, 15]. As shown in Fig 1, the ventral chest wall was segmented by the midline and anterior axillary lines, and further divided into the upper and lower zones. The dorsal chest wall was divided into the left-right and upper-lower zones. For those who were too sick to change their positions (such as intubated patients), only ventral eight zones were scanned [15]. The AI-POCUS movies were automatically analyzed using a real-time automated B-line counting application. All AI-POCUS scans were performed by a blinded novice observer who had experienced fewer than 10 cases of lung POCUS before the start of the study. This machine-learning-based application automatically counts the number of B-lines in the movie, and records the maximum number of B-lines. Three or more B-lines in each zone were considered abnormal (Fig 2).

Fig 1. Lung segmentations in 12-zone lung POCUS.

Fig 1

The chest was divided into eight zones for the ventral wall and four zones for the dorsal chest wall, divided by the posterior axillary line (PAL). The parasternal line (PSL) and paravertebral line (PVL) to two hemithorax. The ventral chest wall was segmented by the midline and anterior axillary lines (AAL) and further divided into upper and lower zones. The dorsal chest wall was divided into left-right and upper-lower zones at the height of the tip of the scapula.

Fig 2. Ultrasound B-lines.

Fig 2

B-lines were automatically analyzed using the real-time deep learning-based application, which counted the number of B-lines (indicated by the white lines in the upper panel) in the movie, and recorded the maximum number of B-lines. Three or more B-lines in each zone were considered abnormal (red zone in lower panel). Patients with one or more abnormal zones were considered positive for AI-POCUS.

Non-contrast chest CT scans were acquired using commercially available equipment (Aquilion ONE, Canon, Japan) at inspiration, and reconstructed as axial images with a 5-mm slice thickness, 5-mm interval, and 120 peak kilovoltage. A board-certified pulmonologist, who was blinded to patient characteristics and AI-POCUS findings, read the CT and diagnosed the presence of pneumonia in each of the 12 zones of the lung that corresponded with the lung POCUS segmentation.

Zone-level concordance was assessed between AI-POCUS and CT results for each zone. In contrast, patient-level AI-POCUS results were defined as negative when the patient had no sign of pneumonia in any of the acquired zones, and patient-level concordance was assessed for each patient.

Statistical analyses

Data are presented as the mean ± standard deviation or median [1st and 3rd quartile] for continuous variables, and as the frequency (%) for categorical variables. Differences in patient characteristics between groups were tested using the Welch t-test and Mann-Whitney test for normally and non-normally distributed variables, respectively, and Fisher’s exact test for categorical variables. The primary endpoint of the study was patient-level accuracy, sensitivity, and specificity of AI-POCUS in comparison with CT-validated pneumonia. Wilson’s method was used to calculate the 95% confidence interval for each metric [16]. All statistical analyses were performed with R version 4.0.2 (The R Foundation for Statistical Computing, Vienna, Austria). A two-tailed p value of <0.05 indicated statistical significance.

Ethics

The study protocol complied with the Declaration of Helsinki and was approved by the institutional review board (IRB) at Juntendo University Hospital (#E21-0197). Written informed consent was waived due to its purely observational nature by the IRB based on the "Ethical Guidelines for Medical and Health Research Involving Human Subjects" issued by Japanese Ministry of Health, Labor and Welfare.

Results

Study population

During the study period, 45 patients with COVID-19 were admitted to the hospital. After excluding 4 patients who did not undergo lung POCUS due to admission on weekends, 41 patients were enrolled in the study. Including 15 control subjects, a total of 56 subjects (41 patients and 15 controls) were included in the study (Fig 3). The baseline patient characteristics are summarized in Table 1. Overall, the mean age was 59.4 ± 14.8 years, 23% were female, and the mean body mass index was 25.3 ± 4.7 kg/m2. The patients were significantly younger than the controls. The median time from the onset to the CT scan was 9 [5 – 12] days, and 20% of the patients were maintained on mechanical ventilation at the time of CT and AI-POCUS.

Fig 3. Patient inclusion chart.

Fig 3

Only 4 patients (8.9%) of cases were excluded and 56 patients (41 cases and 15 controls) were included in the final analysis.

Table 1. Patient characteristic.

Overall (N = 56) Case (N = 41) Control (N = 15) p value
Age, years old 59.4 ± 14.8 56.3 ± 14.6 67.6 ± 12.5 0.008
Female 13 (23%) 8 (20%) 5 (33%) 0.30
Body mass index, kg/m2 25.3 ± 4.7 25.6 ± 5.1 24.7 ± 3.5 0.50
Days from onset 9.0 [5.0–12.0]
Mechanical Ventilation 8 (20%)

AI-POCUS and CT availability

CT evaluation was successful for all 12 zones for all subjects (492 zones for cases and 180 for controls), whereas lung POCUS was not available for the dorsal chest zones in 15 patients who were unable to change their positions. One patient was placed in the prone position, and only the dorsal zones were scanned. Clear images were not available in the left lower inside zone in the ventral chest of 26 patients because the heart was close to the chest wall, and the lung was therefore not visible from this position. Finally, lung POCUS movies of 397 zones for patients and 180 zones for controls were analyzed.

Concordance of pneumonia identification

A CT scan identified pneumonia in 395 zones (58.7%) of 40 subjects, whereas no pneumonia was detected in the other 16 subjects. Pneumonia was present almost evenly in each zone, with a slightly higher prevalence in the dorsal zones (Fig 4).

Fig 4. Distribution of pneumonia.

Fig 4

The X-axis represents the zone of the lung (R1, right anterior superior zone; R2, right anterior inferior zone; R3, right lateral superior zone; R4, right lateral inferior zone; R5, right superior dorsal zone; R6, right inferior dorsal zone; Ls are left zones corresponding to Rs), and the Y-axis represents the number of patients with pneumonia in each zone.

Panel A in Fig 5 demonstrates the patient-level concordance between the AI-POCUS and CT-validated pneumonia. One subject was excluded from this analysis because AI-POCUS was not possible for the dorsal zones, although he only had pneumonia there. AI-POCUS was able to diagnose the presence or absence of pneumonia in 52 of 55 subjects, while pneumonia was overlooked in three patients. The overall accuracy, sensitivity, specificity, PPV, NPT, and Kappa of the AI-POCUS were 94.5% (85.1%– 98.1%), 92.3% (79.7%– 97.3%), 100% (80.6%– 100%), 100% (90.4%– 100%), 84.2% (62.4–94.5%), and 0.89 (0.70–0.98), respectively.

Fig 5. Accuracy of AI-POCUS versus CT.

Fig 5

Each panel is a confusion matrix showing the concordance between the AI-POCUS results and CT-validated pneumonia diagnosis.

The screening performance of AI-POCUS using only the ventral eight zones was evaluated as a simpler approach that can be used even for those who cannot change their position. With only 8-zone AI-POCUS, pneumonia was overlooked in 9 patients, whereas 31 of 40 patients with pneumonia were still classified as positive (Fig 5, Panel B). The accuracy, sensitivity, NPV, and Kappa decreased to 83.9% (72.2%– 91.3%), 77.5% (62.5%– 87.7%), 64.0% (44.5–79.8%), and 0.68 (0.43–0.85), respectively; however, the specificity and PPV remained at 100% for both (80.6%– 100% and 89.0%– 100%).

In the zone-level analysis, AI-POCUS was correct in 377 of 577 zones (Fig 5, Panel C). The zone-level accuracy, sensitivity, specificity, PPV, NPV, and Kappa of the AI-POCUS were 65.3% (61.4%– 69.1%), 37.2% (32.0%– 42.7%), 97.8% (95.2%– 99.0%), 95.5% (89.6%– 97.7%), 57.5% (52.9%– 61.9%), and 0.31 (0.23–0.38), respectively.

Finally, we assumed that the prevalence of pneumonia in COVID-19 positive patients is 5% based on the public data from the Japanese Ministry of Health, Labour and Welfare and calculated the estimated PPV and NPV in the real-world. Given the sensitivity of 92.3% and specificity of 100%, and 5% of prior probability, PPV was 100% (99.9%– 100%) and NPV was 99.6% (99.6%– 99.6%).

Discussion

The main findings of our study were as follows: (1) AI-POCUS had excellent performance at detecting patients with COVID-19 pneumonia (sensitivity 92.3% and specificity 100%), and (2) the zone-level sensitivity was moderate (37.3%), although specificity was very high (97.8%). Even with moderate sensitivity in each zone, pneumonia in COVID-19 usually spreads to multiple lung zones, leading to a high sensitivity for each patient. Notably, these excellent results were achieved with AI-POCUS by a novice observer who had minimal experience in lung-POCUS.

With the recent miniaturization of ultrasound devices and advancements in image quality, lung-POCUS is becoming a popular examination, especially in intensive care [8, 17]. Although the lung itself, which is filled with air, cannot usually be observed by ultrasound beams, the noise and artifacts generated by ultrasound beams provide useful clinical information in lung-POCUS. Ultrasound B-lines are one of the most useful artifacts that present with pulmonary congestion, either by pneumonia or cardiogenic pulmonary edema. These reverberation artifacts originate at the pleura, reflecting an air-fluid mixture, which occurs when the subpleural interlobular septa surrounded by subpleural air-filled alveoli become edematous. Previous studies have reported that three or more B-lines visible in a single ultrasound plane are fully sensitive and specific to demonstrate subpleural thickened interlobular septa and/or ground-glass areas with a CT scan as a reference [18].

Diagnosis of pneumonia using lung-POCUS techniques, including the B-line, is expected to be an effective tool in the ongoing COVID-19 pandemic. Reports have already shown that B-lines are not only sensitive, but also associated with disease severity, future deterioration, and treatment effects in COVID-19 [15, 1923]. However, chest radiography and CT remain the leading examinations used in the management of COVID-19 worldwide, and lung-POCUS has not been sufficiently used. Technical difficulties and problems in interpreting images are major concerns when using lung-POCUS; the number of experts is not large enough to teach and supervise the use of lung-POCUS by clinicians, including general practitioners, although this technique is relatively new and may be difficult for novice observers.

In the present study, we demonstrated that with AI technology, lung-POCUS can be effective to diagnose pneumonia, with excellent accuracy, even by a novice observer. Recent advancements in AI technology, more specifically machine learning, have enabled automated image recognition with similar or even higher accuracy compared to expert clinicians, and the application of such technologies to POCUS has been enthusiastically studied [10, 11, 24, 25]. The automated B-line counting application that we used in this study was also developed using machine learning technology, and showed excellent agreement with expert readings [26, 27].

The moderate zone-level accuracy observed herein may be due to the limited ability to detect slight pneumonia in the central area of the lung far from the body surface. Such slight pneumonia in a deep area of the lung may not affect the pleura, and would therefore not generate a B line. However, pneumonia caused by COVID-19 usually spreads to multiple regions of the lungs, appearing in the form of acute respiratory distress syndrome [23]. Therefore, diffuse pneumonia caused by COVID-19 is a suitable target for AI-POCUS.

COVID-19 is expected to persist for some time. AI-POCUS allows non-specialist general practitioners to diagnose pneumonia, a problem in viral infections, in an access-free manner, which may be useful in the treatment of pneumonia during a pandemic, and is expected to become more widespread.

Limitations

This study has some limitations. First, this was a single-center study conducted in a university hospital, and included a small number of cases. Larger studies involving various settings are necessary before the current results can be widely implemented. Second, we did not check the accuracy of AI-POCUS compared with traditional lung POCUS by an expert. However, a previous study showed that AI-POCUS has over 90% sensitivity and specificity for detecting expert-annotated B-lines [26]. Next, we studied only hospitalized patients who were considered sicker than those who were cared for at home or at other facilities. The performance of the present application should be tested in less intensively sick patients to uncover it’s applicability for screening purposes.

Conclusions

AI-POCUS showed excellent performance in diagnosing patients with CT-validated COVID-19 pneumonia, even by a novice observer.

Abbreviations

COVID-19

coronavirus disease 2019

POCUS

point-of-care ultrasound

AI-POCUS

artificial intelligence-based point-of-care lung ultrasound

CT

computed tomography

Data Availability

The data are not publicly open due to privacy concerns. Sharing the present data with researchers/companies outside the author’s research group is not approved by the IRB at Juntendo University, and IRB application and approval upon an appropriate request are necessary before sharing the data (contact information: Nobuyuki Kagiyama; kgnb_27_hot@yahoo.co.jp and Data Coordinator of Department of Cardiovascular Biology and Medicine; junkanki@juntendo.ac.jp).

Funding Statement

Kagiyama and Daida are affiliated with a department funded by Philips Japan, Asahi KASEI Corporation, Inter Reha Co., Ltd, and Toho Holdings Co., Ltd., based on collaborative research agreements. Other authors have no conflict of interest to declare. This does not alter our adherence to PLOS ONE policies on sharing data and materials. This work was partially supported by the Japan Society for Promotion of Science KAKENHI (by the Japanese government), with a Grant Number 21K18086. There was no additional external funding received for this study.

References

  • 1.Organization WH. WHO Coronavirus Disease (COVID-19) Dashboard. 2022. [cited 2022 April 26]. Available from: https://covid19.who.int/. [Google Scholar]
  • 2.Carvalho T, Krammer F, Iwasaki A. The first 12 months of COVID-19: a timeline of immunological insights. Nat Rev Immunol. 2021;21(4):245–56. Epub 2021/03/17. doi: 10.1038/s41577-021-00522-1 ; PubMed Central PMCID: PMC7958099. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Hu B, Guo H, Zhou P, Shi ZL. Characteristics of SARS-CoV-2 and COVID-19. Nat Rev Microbiol. 2021;19(3):141–54. Epub 2020/10/08. doi: 10.1038/s41579-020-00459-7 ; PubMed Central PMCID: PMC7537588. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Wiersinga WJ, Rhodes A, Cheng AC, Peacock SJ, Prescott HC. Pathophysiology, Transmission, Diagnosis, and Treatment of Coronavirus Disease 2019 (COVID-19): A Review. Jama. 2020;324(8):782–93. Epub 2020/07/11. doi: 10.1001/jama.2020.12839 . [DOI] [PubMed] [Google Scholar]
  • 5.Ji Y, Ma Z, Peppelenbosch MP, Pan Q. Potential association between COVID-19 mortality and health-care resource availability. Lancet Glob Health. 2020;8(4):e480. Epub 2020/02/29. doi: 10.1016/S2214-109X(20)30068-1 ; PubMed Central PMCID: PMC7128131. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Kavanagh MM, Erondu NA, Tomori O, Dzau VJ, Okiro EA, Maleche A, et al. Access to lifesaving medical resources for African countries: COVID-19 testing and response, ethics, and politics. Lancet. 2020;395(10238):1735–8. Epub 2020/05/11. doi: 10.1016/S0140-6736(20)31093-X ; PubMed Central PMCID: PMC7252104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Mojoli F, Bouhemad B, Mongodi S, Lichtenstein D. Lung Ultrasound for Critically Ill Patients. Am J Respir Crit Care Med. 2019;199(6):701–14. Epub 2018/10/30. doi: 10.1164/rccm.201802-0236CI . [DOI] [PubMed] [Google Scholar]
  • 8.Volpicelli G, Elbarbary M, Blaivas M, Lichtenstein DA, Mathis G, Kirkpatrick AW, et al. International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012;38(4):577–91. Epub 2012/03/07. doi: 10.1007/s00134-012-2513-4 . [DOI] [PubMed] [Google Scholar]
  • 9.Xia Y, Ying Y, Wang S, Li W, Shen H. Effectiveness of lung ultrasonography for diagnosis of pneumonia in adults: a systematic review and meta-analysis. Journal of Thoracic Disease. 2016;8(10):2822–31. doi: 10.21037/jtd.2016.09.38 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Liu Y, Jain A, Eng C, Way DH, Lee K, Bui P, et al. A deep learning system for differential diagnosis of skin diseases. Nat Med. 2020;26(6):900–8. Epub 2020/05/20. doi: 10.1038/s41591-020-0842-3 . [DOI] [PubMed] [Google Scholar]
  • 11.Kagiyama N, Shrestha S, Farjo PD, Sengupta PP. Artificial Intelligence: Practical Primer for Clinical Research in Cardiovascular Disease. J Am Heart Assoc. 2019;8(17):e012788. Epub 2019/08/28. doi: 10.1161/JAHA.119.012788 ; PubMed Central PMCID: PMC6755846. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Roy S, Menapace W, Oei S, Luijten B, Fini E, Saltori C, et al. Deep Learning for Classification and Localization of COVID-19 Markers in Point-of-Care Lung Ultrasound. IEEE Trans Med Imaging. 2020;39(8):2676–87. Epub 2020/05/15. doi: 10.1109/TMI.2020.2994459 . [DOI] [PubMed] [Google Scholar]
  • 13.Baloescu C, Chen A, Raju B, Evans N, Moore CL. 19 Automated Quantification Of B-Lines in Lung Ultrasound On COVID-19 Patients. Ann Emerg Med. 2021;78(2):S9–S10. Epub 2021/08/04. doi: 10.1016/j.annemergmed.2021.07.020 PMC8335439. [DOI] [Google Scholar]
  • 14.Levy Adatto N, Preisler Y, Shetrit A, Shepshelovich D, Hershkoviz R, Isakov O. Rapid 8-Zone Lung Ultrasound Protocol is Comparable to a Full 12-Zone Protocol for Outcome Prediction in Hospitalized COVID-19 Patients. J Ultrasound Med. 2021. Epub 2021/10/27. doi: 10.1002/jum.15849 ; PubMed Central PMCID: PMC8661589. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Haurylenka D, Damantsevich V, Filustsin A, Damantsevich A. Diagnostic value of 12-zone lung ultrasound protocol for diagnosing COVID-19-associated pneumonia in outpatients. J Ultrason. 2021;21(87):e271–e6. Epub 2022/01/01. doi: 10.15557/JoU.2021.0046 ; PubMed Central PMCID: PMC8678640. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Wilson EB. Probable Inference, the Law of Succession, and Statistical Inference. Journal of the American Statistical Association. 1927;22(158):209–12. doi: 10.1080/01621459.1927.10502953 [DOI] [Google Scholar]
  • 17.Bhagra A, Tierney DM, Sekiguchi H, Soni NJ. Point-of-Care Ultrasonography for Primary Care Physicians and General Internists. Mayo Clin Proc. 2016;91(12):1811–27. Epub 2016/11/09. doi: 10.1016/j.mayocp.2016.08.023 . [DOI] [PubMed] [Google Scholar]
  • 18.Lichtenstein D, Mézière G, Biderman P, Gepner A, Barré O. The comet-tail artifact. An ultrasound sign of alveolar-interstitial syndrome. Am J Respir Crit Care Med. 1997;156(5):1640–6. Epub 1997/12/31. doi: 10.1164/ajrccm.156.5.96-07096 . [DOI] [PubMed] [Google Scholar]
  • 19.Zieleskiewicz L, Markarian T, Lopez A, Taguet C, Mohammedi N, Boucekine M, et al. Comparative study of lung ultrasound and chest computed tomography scan in the assessment of severity of confirmed COVID-19 pneumonia. Intensive Care Med. 2020;46(9):1707–13. Epub 2020/07/31. doi: 10.1007/s00134-020-06186-0 ; PubMed Central PMCID: PMC7388119. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Rubio-Gracia J, Giménez-López I, Garcés-Horna V, López-Delgado D, Sierra-Monzón JL, Martínez-Lostao L, et al. Point-of-care lung ultrasound assessment for risk stratification and therapy guiding in COVID-19 patients: a prospective noninterventional study. Eur Respir J. 2021;58(3). Epub 2021/02/13. doi: 10.1183/13993003.04283-2020 ; PubMed Central PMCID: PMC7877323 interest: I. Giménez-López has nothing to disclose. Conflict of interest: V. Garcés-Horna has nothing to disclose. Conflict of interest: D. López-Delgado has nothing to disclose. Conflict of interest: J.L. Sierra-Monzón has nothing to disclose. Conflict of interest: L. Martínez-Lostao has nothing to disclose. Conflict of interest: C. Josa-Laorden has nothing to disclose. Conflict of interest: F. Ruiz-Laiglesia has nothing to disclose. Conflict of interest: J.I. Pérez-Calvo has nothing to disclose. Conflict of interest: S. Crespo-Aznarez has nothing to disclose. Conflict of interest: J. García-Lafuente has nothing to disclose. Conflict of interest: N. Peña-Fresneda has nothing to disclose. Conflict of interest: B. Amores Arriaga has nothing to disclose. Conflict of interest: B. Gracia-Tello has nothing to disclose. Conflict of interest: M. Sánchez-Marteles has nothing to disclose. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Volpicelli G, Gargani L, Perlini S, Spinelli S, Barbieri G, Lanotte A, et al. Lung ultrasound for the early diagnosis of COVID-19 pneumonia: an international multicenter study. Intensive Care Med. 2021;47(4):444–54. Epub 2021/03/21. doi: 10.1007/s00134-021-06373-7 ; PubMed Central PMCID: PMC7980130. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Wang M, Luo X, Wang L, Estill J, Lv M, Zhu Y, et al. A Comparison of Lung Ultrasound and Computed Tomography in the Diagnosis of Patients with COVID-19: A Systematic Review and Meta-Analysis. Diagnostics (Basel). 2021;11(8). Epub 2021/08/28. doi: 10.3390/diagnostics11081351 ; PubMed Central PMCID: PMC8394642. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Zheng Y, Wang L, Ben S. Meta-analysis of chest CT features of patients with COVID-19 pneumonia. J Med Virol. 2021;93(1):241–9. Epub 2020/06/25. doi: 10.1002/jmv.26218 ; PubMed Central PMCID: PMC7361361. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Sengupta PP, Shrestha S, Berthon B, Messas E, Donal E, Tison GH, et al. Proposed Requirements for Cardiovascular Imaging-Related Machine Learning Evaluation (PRIME): A Checklist: Reviewed by the American College of Cardiology Healthcare Innovation Council. JACC Cardiovasc Imaging. 2020;13(9):2017–35. Epub 2020/09/12. doi: 10.1016/j.jcmg.2020.07.015 ; PubMed Central PMCID: PMC7953597. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Seetharam K, Kagiyama N, Sengupta PP. Application of mobile health, telemedicine and artificial intelligence to echocardiography. Echo Res Pract. 2019;6(2):R41–r52. Epub 2019/03/08. doi: 10.1530/ERP-18-0081 ; PubMed Central PMCID: PMC6432977. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Baloescu C, Toporek G, Kim S, McNamara K, Liu R, Shaw MM, et al. Automated Lung Ultrasound B-Line Assessment Using a Deep Learning Algorithm. IEEE Trans Ultrason Ferroelectr Freq Control. 2020;67(11):2312–20. Epub 2020/08/04. doi: 10.1109/TUFFC.2020.3002249 . [DOI] [PubMed] [Google Scholar]
  • 27.Moore CL, Wang J, Battisti AJ, Chen A, Fincke J, Wang A, et al. Interobserver Agreement and Correlation of an Automated Algorithm for B-Line Identification and Quantification With Expert Sonologist Review in a Handheld Ultrasound Device. J Ultrasound Med. 2021. Epub 2021/12/30. doi: 10.1002/jum.15935 . [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Juan Antonio Valera-Calero

26 Sep 2022

PONE-D-22-21871Artificial Intelligence-based Point-of-care Lung Ultrasound for Screening COVID-19 Pneumoniae: Comparison with CT ScansPLOS ONE

Dear Dr. Kagiyama,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Nov 10 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Juan Antonio Valera-Calero

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at 

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf.

2. Thank you for stating in your Funding Statement: 

"This work was partially supported by JSPS KAKENHI, with a Grant Number 21K18086."

Please provide an amended statement that declares *all* the funding or sources of support (whether external or internal to your organization) received during this study, as detailed online in our guide for authors at http://journals.plos.org/plosone/s/submit-now.  Please also include the statement “There was no additional external funding received for this study.” in your updated Funding Statement. 

Please expand the acronym “JSPS KAKENHI” (as indicated in your financial disclosure) so that it states the name of your funders in full.

Please include your amended Funding Statement within your cover letter. We will change the online submission form on your behalf.

3. Thank you for stating the following in the Competing Interests section: 

"Kagiyama and Daida are affiliated with a department funded by Philips Japan, Asahi KASEI Corporation, Inter Reha Co., Ltd, and Toho Holdings Co., Ltd., based on collaborative research agreements. Other authors have no conflict of interest to declare."

Please confirm that this does not alter your adherence to all PLOS ONE policies on sharing data and materials, by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests).  If there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared. 

Please include your updated Competing Interests statement in your cover letter; we will change the online submission form on your behalf.

4. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. 

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. Please see http://www.bmj.com/content/340/bmj.c181.long for guidelines on how to de-identify and prepare clinical data for publication. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

Additional Editor Comments:

Please address the comments provided by the reviewers.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: No

Reviewer #2: I Don't Know

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Abstract:

The CI 95% of the performance indices are missing. Moreover, the PPV, NPV, and Kappa must be added as important performance indices to the abstract and paper.

The percentage of CT-validated pneumonia must be added to the subject (65) and frame data (577 lung zones).

If random sampling and proper sample size estimation are not used, they must be mentioned as important issues that avoid the generalization to the entire population in the abstract and discussion.

Material and Methods:

The flowchart of the experimental design (including the inclusion and exclusion criteria, etc.) must be added to the paper.

Statistical analysis:

The CI 95% must be added to the entire indices based on the TRIPOD/STARD guideline. The Kappa (agreement rate) and PPV (the probability of having pneumonia if the test result is positive) must be added.

The authors mentioned that the AI method's training, validation, and further details are unavailable. A scientific work must provide the required information to repeat the work. For example, it is unclear which validation framework (hold-out, cross-validation, etc.) was used and whether the results are based on the cross-validated confusion matrix. Moreover, the percentage of the training, validation, and test sets must be provided.

The AI method must be briefly described in the paper, and providing the citation is insufficient.

Results:

For the entire indices (plus PPV, NPV, and Kappa), the CI 95% must be provided, showing the precision of the indices.

The objective comparison with the state-of-the-art using proper statistical methods is missing.

A major issue is how the proposed method performs in practice. The authors must use the sensitivity and specificity of the proposed system and the prevalence of pneumonia in the population (based on the literature) and use the Bayesian theorem to predict the PPV and NPV of the system in the population.

Section 2 of the following paper could be easily used by the authors for this estimation:

https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8170576/

Discussion:

And finally, is the proposed diagnosis system clinically reliable?

Reviewer #2: Dear author,

The strength of the paper is well structured, the description of the related work is well done and that results are compared to results of the similar research. At all the work is of certain significance. The paper provides a very powerful message about the "Artificial Intelligence-based Point-of-care Lung Ultrasound for Screening COVID-19 Pneumoniae: Comparison with CT Scans".

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Hamid Reza Marateb

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2023 Mar 16;18(3):e0281127. doi: 10.1371/journal.pone.0281127.r002

Author response to Decision Letter 0


16 Dec 2022

Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1:

Abstract:

The CI 95% of the performance indices are missing. Moreover, the PPV, NPV, and Kappa must be added as important performance indices to the abstract and paper.

Response: Thank you for the valuable suggestion. We calculated the confidence intervals using the Wilson’s method (1), which is recommended by several papers (2, 3), as shown below. We were not able to include all metrics for all examinations (i.e. for full-zone patient-level, 8-zone patient-level, and zone-level) in the abstract, so PPV and Kappa for the secondary outcomes were included only in the manuscript text.

1 E.B. Wilson. Probable inference, the law of succession, and statistical inference. J Am Stat Assoc 1927;22:209–212.

2 A. Agresti and B.A. Coull. Approximate is better than "exact" for interval estimation of binomial proportions. Am Statistician, 1998;52:119-126.

3 L.D. Brown, T.T. Cai and A. Dasgupta Interval estimation for a binomial proportion Statistical Science, 2001;16(2):101-133

[Original in Abstract]

The patient-level accuracy of 12-zone AI-POCUS for detecting CT-validated pneumonia was 94.5%, sensitivity was 92.3%, and specificity was 100%. When simplified with 8-zone scan, the accuracy, sensitivity, and sensitivity were 83.9%, 77.5%, and 100%, respectively. The zone-level accuracy, sensitivity, and specificity of AI-POCUS were 65.3 %, 37.2%, and 97.8 %, respectively.

[Revised]

The 12-zone AI-POCUS for detecting CT-validated pneumonia in the patient-level showed the accuracy of 94.5% (85.1% – 98.1%), sensitivity of 92.3% (79.7% – 97.3%), specificity of 100% (80.6% – 100%), positive predictive value of 95.0% (89.6% - 97.7%), and Kappa of 0.33 (0.27 – 0.40). When simplified with 8-zone scan, the accuracy, sensitivity, and sensitivity were 83.9% (72.2% – 91.3%), 77.5% (62.5% – 87.7%), and 100% (80.6% – 100%), respectively. The zone-level accuracy, sensitivity, and specificity of AI-POCUS were 65.3% (61.4% – 69.1%), 37.2% (32.0% – 42.7%), and 97.8 % (95.2% – 99.0%), respectively.

The percentage of CT-validated pneumonia must be added to the subject (65) and frame data (577 lung zones).

Response: We included them in the manuscript.

[Added in Abstract]

The CT-validated pneumonia was seen in 71.4% of patients at total 577 lung zones (53.3%).

If random sampling and proper sample size estimation are not used, they must be mentioned as important issues that avoid the generalization to the entire population in the abstract and discussion.

Response: We added the following limitation in the abstract as well as discussion.

[Added in Abstract]

The sample size calculation was not performed given the retrospective all-comer nature of the study.

Material and Methods:

The flowchart of the experimental design (including the inclusion and exclusion criteria, etc.) must be added to the paper.

Response: We added the following flow chart as new Figure 1.

[Original in Results]

A total of 56 subjects (41 patients and 15 controls) were enrolled in the study.

[Added in Results]

During the study period, 45 patients with COVID-19 were admitted to the hospital. After excluding 4 patients who did not undergo lung POCUS due to admission on weekends, 41 patients were enrolled in the study. Including 15 control subjects, a total of 56 subjects (41 patients and 15 controls) were included in the study (Figure 1).

Statistical analysis:

The CI 95% must be added to the entire indices based on the TRIPOD/STARD guideline. The Kappa (agreement rate) and PPV (the probability of having pneumonia if the test result is positive) must be added.

Response: Throughout the manuscript, we added the 95% confidence intervals for the performance metrics and PPV and Kappa values. The changes in the results are shown below. The additional statistical analysis was explained as shown below.

[Added in the statistical analysis]

Wilson’s method was used to calculate the 95% confidence interval for each metric (1).

1 E.B. Wilson. Probable inference, the law of succession, and statistical inference. J Am Stat Assoc 1927;22:209–212.

[Original (overall patient-level analysis)]

The overall accuracy, sensitivity, and specificity of the AI-POCUS were 94.5%, 92.3%, and 100%, respectively.

[Revised]

The overall accuracy, sensitivity, specificity, PPV, NPT, and Kappa of the AI-POCUS were 94.5% (85.1% – 98.1%), 92.3% (79.7% – 97.3%), 100% (80.6% – 100%), 100% (90.4% – 100%), 84.2% (62.4 – 94.5%), and 0.89 (0.70 – 0.98), respectively.

[Original (patient-level analysis using only 8 zones)]

The accuracy and sensitivity decreased to 83.9% and 77.5%, respectively; however, the specificity remained at 100%.

[Revised]

The accuracy, sensitivity, NPV, and Kappa decreased to 83.9% (72.2% – 91.3%), 77.5% (62.5% – 87.7%), 64.0% (44.5 – 79.8%), and 0.68 (0.43 – 0.85), respectively; however, the specificity and PPV remained at 100% for both (80.6% – 100% and 89.0% – 100%).

[Original (zone-level)]

The zone-level accuracy, sensitivity, and specificity of the AI-POCUS were 65.3%, 37.2%, and 97.8%, respectively.

[Revised]

The zone-level accuracy, sensitivity, specificity, PPV, NPV, and Kappa of the AI-POCUS were 65.3% (61.4% – 69.1%), 37.2% (32.0% – 42.7%), 97.8% (95.2% – 99.0%), 95.5% (89.6% – 97.7%), 57.5% (52.9% – 61.9%), and 0.31 (0.23 – 0.38), respectively.

The authors mentioned that the AI method's training, validation, and further details are unavailable. A scientific work must provide the required information to repeat the work. For example, it is unclear which validation framework (hold-out, cross-validation, etc.) was used and whether the results are based on the cross-validated confusion matrix. Moreover, the percentage of the training, validation, and test sets must be provided.

The AI method must be briefly described in the paper, and providing the citation is insufficient.

Response: Thank you for letting us clarify this important point. Although we used an AI application in this study, this development process was performed by the company (Philips) and is not publicly open. In this study, we applied this already-developed commercially available software to our population that is completely different from what the company used to develop the model. In this term, this study can be considered as an external or real-world validation of the model, not the development of the model. As a validation study, we believe that the pipeline and study flow of our study are in line with other studies (e.g., references 1-3) and are in a scientific manner.

1. C. Dung-Hung, et al. Crit Care 2022;26(1):215

2. IZ Attia, at al. Int J Cardiol 2021;329:130-135.

3. N, Kagiyama, et al. Echocardiography. 2016;33(5):756-63

Results:

For the entire indices (plus PPV, NPV, and Kappa), the CI 95% must be provided, showing the precision of the indices.

Response: We added 95% confidence indices and PPV, NPV and Kappa in results as shown above in the response to the previous comment.

The objective comparison with the state-of-the-art using proper statistical methods is missing.

Response: As we explained in the previous response. This is not a study in which we developed a machine-learning model, but we performed an external validation of the machine learning-based application that was already established and fixed. The gold standard was the CT results, and there is no state-of-the-art to be compared with.

A major issue is how the proposed method performs in practice. The authors must use the sensitivity and specificity of the proposed system and the prevalence of pneumonia in the population (based on the literature) and use the Bayesian theorem to predict the PPV and NPV of the system in the population.

Section 2 of the following paper could be easily used by the authors for this estimation:

https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8170576/

Response: We thank the reviewer for this valuable suggestion. The hospitalization rate of COVID-19 in the elderly people in Japan is expected to be 5 – 15% by the Japanese Ministry of Health, Labour and Welfare (https://www.mhlw.go.jp/content/10900000/000964719.pdf). Based on this data, we assumed that the prevalence of pneumonia is 5% in COVID-19 positive patients. Given the following parameters, we calculated the PPV and NPV.

Given the sensitivity of 92.3% and specificity of 100%, and 5% of prior probability;

PPV: 100% (99.9% – 100%)

NPV: 99.6% (99.6% – 99.6%)

We added these results to the manuscript.

[Added in Results]

Finally, we assumed that the prevalence of pneumonia in COVID-19 positive patients is 5% based on the public data from the Japanese Ministry of Health, Labour and Welfare and calculated the estimated PPV and NPV in the real-world. Given the sensitivity of 92.3% and specificity of 100%, and 5% of prior probability, PPV was 100% (99.9% – 100%) and NPV was 99.6% (99.6% – 99.6%).

Discussion:

And finally, is the proposed diagnosis system clinically reliable?

Response: Yes. The present study included consecutive patients with the exclusion of less than 10%. We believe that the population represents the real-world COVID-19 population, and the present results are generalizable in patients hospitalized with COVID-19. The good patient-level accuracy is clinically meaningful. However, as we stated in the manuscript as below, there are limitations and further studies are warranted.

[Original in the limitation]

First, this was a single-center study conducted in a university hospital, and included a small number of cases. Larger studies involving various settings are necessary before the current results can be widely implemented. ~~ Next, we studied only hospitalized patients who were considered sicker than those who were cared for at home or at other facilities. The performance of the present application should be tested in less intensively sick patients to uncover it’s applicability for screening purposes.

Reviewer #2:

The strength of the paper is well structured, the description of the related work is well done and that results are compared to results of the similar research. At all the work is of certain significance. The paper provides a very powerful message about the "Artificial Intelligence-based Point-of-care Lung Ultrasound for Screening COVID-19 Pneumoniae: Comparison with CT Scans".

Response: Thank you for your very encouraging comment. We are very happy that you stated that our paper is of certain significance and provides a very powerful message. We’d like to express our appreciation.

Attachment

Submitted filename: Reponse to Reviewers AI_LungPOCUS2.docx

Decision Letter 1

Juan Antonio Valera-Calero

16 Jan 2023

Artificial Intelligence-based Point-of-care Lung Ultrasound for Screening COVID-19 Pneumoniae: Comparison with CT Scans

PONE-D-22-21871R1

Dear Dr. Kagiyama,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Juan Antonio Valera-Calero

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The authors responded to the entire issues addressed by the reviewer. The paper is now suitable for publication.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Hamid Reza Marateb

**********

Acceptance letter

Juan Antonio Valera-Calero

30 Jan 2023

PONE-D-22-21871R1

Artificial Intelligence-based Point-of-care Lung Ultrasound for Screening COVID-19 Pneumoniae: Comparison with CT Scans

Dear Dr. Kagiyama:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Juan Antonio Valera-Calero

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    Attachment

    Submitted filename: Reponse to Reviewers AI_LungPOCUS2.docx

    Data Availability Statement

    The data are not publicly open due to privacy concerns. Sharing the present data with researchers/companies outside the author’s research group is not approved by the IRB at Juntendo University, and IRB application and approval upon an appropriate request are necessary before sharing the data (contact information: Nobuyuki Kagiyama; kgnb_27_hot@yahoo.co.jp and Data Coordinator of Department of Cardiovascular Biology and Medicine; junkanki@juntendo.ac.jp).


    Articles from PLOS ONE are provided here courtesy of PLOS

    RESOURCES