Bubble devices versus other pressure sources for nasal continuous positive airway pressure in preterm infants

Raj Prakash 1, Antonio G De Paoli 2, Peter G Davis 3, Sam J Oddie 4, William McGuire 5,
Editor: Cochrane Neonatal Group
PMCID: PMC10064833  PMID: 37009665

Abstract

Background

Several types of pressure sources, including underwater bubble devices, mechanical ventilators, and the Infant Flow Driver, are used for providing continuous positive airway pressure (CPAP) to preterm infants with respiratory distress. It is unclear whether the use of bubble CPAP versus other pressure sources is associated with lower rates of CPAP treatment failure, or mortality and other morbidity. 

Objectives

To assess the benefits and harms of bubble CPAP versus other pressure sources (mechanical ventilators or Infant Flow Driver) for reducing treatment failure and associated morbidity and mortality in newborn preterm infants with or at risk of respiratory distress.

Search methods

We searched the Cochrane Central Register of Controlled Trials (CENTRAL; 2023, Issue 1); MEDLINE (1946 to 6 January 2023), Embase (1974 to 6 January 2023), Maternity & Infant Care Database (1971 to 6 January 2023), and the Cumulative Index to Nursing and Allied Health Literature (1982 to 6 January 2023). We searched clinical trials databases and the reference lists of retrieved articles.

Selection criteria

We included randomised controlled trials comparing bubble CPAP with other pressure sources (mechanical ventilators or Infant Flow Driver) for the delivery of nasal CPAP to preterm infants.

Data collection and analysis

We used standard Cochrane methods. Two review authors separately evaluated trial quality, extracted data, and synthesised effect estimates using risk ratio (RR), risk difference (RD), and mean difference. We used the GRADE approach to assess the certainty of the evidence for effects on treatment failure, all‐cause mortality, neurodevelopmental impairment, pneumothorax, moderate‐severe nasal trauma, and bronchopulmonary dysplasia.

Main results

We included 15 trials involving a total of 1437 infants. All trials were small (median number of participants 88). The methods used to generate the randomisation sequence and ensure allocation concealment were unclear in about half of the trial reports. Lack of measures to blind caregivers or investigators was a potential source of bias in all of the included trials. The trials took place during the past 25 years in care facilities internationally, predominantly in India (five trials) and Iran (four trials). The studied pressure sources were commercially available bubble CPAP devices versus a variety of mechanical ventilator (11 trials) or Infant Flow Driver (4 trials) devices. 

Meta‐analyses suggest that the use of bubble CPAP compared with mechanical ventilator or Infant Flow Driver CPAP may reduce the rate of treatment failure (RR 0.76, 95% confidence interval (CI) 0.60 to 0.95; (I² = 31%); RD −0.05, 95% CI −0.10 to −0.01; number needed to treat for an additional beneficial outcome 20, 95% CI 10 to 100; 13 trials, 1230 infants; low certainty evidence). The type of pressure source may not affect mortality prior to hospital discharge (RR 0.93, 95% CI 0.64 to 1.36 (I² = 0%); RD −0.01, 95% CI −0.04 to 0.02; 10 trials, 1189 infants; low certainty evidence). No data were available on neurodevelopmental impairment. Meta‐analysis suggests that the pressure source may not affect the risk of pneumothorax (RR 0.73, 95% CI 0.40 to 1.34 (I² = 0%); RD −0.01, 95% CI −0.03 to 0.01; 14 trials, 1340 infants; low certainty evidence). Bubble CPAP likely increases the risk of moderate‐severe nasal injury (RR 2.29, 95% CI 1.37 to 3.82 (I² = 17%); RD 0.07, 95% CI 0.03 to 0.11; number needed to treat for an additional harmful outcome 14, 95% CI 9 to 33; 8 trials, 753 infants; moderate certainty evidence). The pressure source may not affect the risk of bronchopulmonary dysplasia (RR 0.76, 95% CI 0.53 to 1.10 (I² = 0%); RD −0.04, 95% CI −0.09 to 0.01; 7 trials, 603 infants; low certainty evidence). 

Authors' conclusions

Given the low level of certainty about the effects of bubble CPAP versus other pressure sources on the risk of treatment failure and most associated morbidity and mortality for preterm infants, further large, high‐quality trials are needed to provide evidence of sufficient validity and applicability to inform context‐ and setting‐relevant policy and practice.

Keywords: Female; Humans; Infant, Newborn; Pregnancy; Bronchopulmonary Dysplasia; Bronchopulmonary Dysplasia/epidemiology; Bronchopulmonary Dysplasia/etiology; Continuous Positive Airway Pressure; Continuous Positive Airway Pressure/adverse effects; Continuous Positive Airway Pressure/methods; Dyspnea; Infant, Premature; Pneumothorax; Pneumothorax/epidemiology; Pneumothorax/etiology; Respiratory Distress Syndrome

Plain language summary

Pressure sources for nasal continuous positive airway pressure (CPAP) in preterm infants

Key messages

Bubble continuous positive airway pressure (CPAP) may reduce the risk of CPAP treatment failure when compared with CPAP delivered by mechanical ventilators or Infant Flow Driver. Bubble CPAP probably has little or no impact on the risk of death or other complications associated with premature birth but likely increases the risk of moderate‐severe nasal injury. 

What is CPAP?

CPAP is a form of breathing support that can be used to support breathing in a preterm (premature) baby with lung problems. Various types of machines can provide CPAP, including underwater bubble devices (bubble CPAP), mechanical ventilators, and Infant Flow Driver. 

What did we want to find out?

We wanted to determine whether there is evidence to favour bubble systems versus ventilator or Infant Flow Driver systems for reducing the rate of CPAP treatment failure (the baby's condition worsening or the baby needing mechanical ventilation) and reducing complications and harms.

What did we do?

We searched medical databases for randomised controlled trials (a type of study where participants are randomly assigned to one of two or more treatment groups) up to January 2023.

What did we find?

We included 15 trials that compared the use of bubble CPAP versus ventilator or Infant Flow Driver CPAP in a total of 1437 preterm babies. Trials were mostly small, and had design flaws that could put their findings at risk of bias. 

Key results

Combined analyses showed that using bubble CPAP rather than ventilator or Infant Flow Driver CPAP may reduce the risk of CPAP treatment failure, but that bubble CPAP may not affect the risk of death or other complications of prematurity. Bubble CPAP likely increases the risk of moderate‐severe nasal injury. None of the included studies looked at effects on disability or other developmental outcomes.

What are the limitations of the evidence?

We judged the certainty of the evidence for the effects of bubble versus ventilators or Infant Flow Driver for CPAP in preterm babies to be low because of concerns that the methods used in the included trials may have introduced biases, and the limited amount of data from the trials (meaning the results are less precise). Our confidence in the results is therefore limited.

Summary of findings

Summary of findings 1. Bubble versus ventilator or Infant Flow Driver nasal continuous positive airway pressure (CPAP) for preterm infants.

Bubble versus ventilator or Infant Flow Driver nasal continuous positive airway pressure (CPAP) in preterm infants
Patient or population: preterm infants receiving nasal CPAP
Setting: neonatal care facilities internationally
Intervention: bubble CPAP
Comparison: ventilator or Infant Flow Driver CPAP
Outcomes Anticipated absolute effects* (95% CI) Relative effect 
(95% CI) Absolute effect(95% CI) № of participants
(studies) Certainty of the evidence
(GRADE)
Risk with ventilator or Infant Flow Driver CPAP Risk with bubble CPAP
Treatment failure 215 per 1000 163 per 1000 
(129 to 204)
  RR 0.76
(0.60 to 0.95) 52 per 1000 fewer (11 to 86 fewer per 1000) 1230
(13 studies) ⊕⊕⊝⊝
Lowa,b
All‐cause mortality before hospital discharge 78 per 1000 72 per 1000
(50 to 106) RR 0.93
(0.64 to 1.36) 6 per 1000 fewer (28 fewer to 28 more per 1000) 1189
(10 studies) ⊕⊕⊝⊝
Lowa,b
Neurodevelopmental impairment Not assessed in any included trials
Pneumothorax 31 per 1000 23 per 1000
(13 to 42) RR 0.73
(0.40 to 1.34) 8 per 1000 fewer (18 fewer to 11 more per 1000) 1340
(14 studies) ⊕⊕⊝⊝
Lowa,b
Moderate‐severe nasal injury 48 per 1000 109 per 1000
(65 to 182) RR 2.29
(1.37 to 3.82) 61 per 1000 more (17 to 134 more per 1000) 753
(8 studies) ⊕⊕⊕⊝
Moderatea
Bronchopulmonary dysplasia 167 per 1000 127 per 1000
(89 to 184) RR 0.76
(0.53 to 1.10) 40 per 1000 fewer (78 fewer to 17 more per 1000) 603
(7 studies) ⊕⊕⊝⊝
Lowa,b
*The risk in the intervention group (and its 95% confidence interval) is based on the assumed risk in the comparison group and the relative effect (relative risk) of the intervention (and its 95% confidence interval)
CI: confidence interval; CPAP: continuous positive airway pressure; RR: risk ratio
GRADE Working Group grades of evidenceHigh certainty: We are very confident that the true effect lies close to that of the estimate of the effect.
Moderate certainty: We are moderately confident in the effect estimate: the true effect is likely to be close to the estimate of the effect, but there is a possibility that it is substantially different.
Low certainty: Our confidence in the effect estimate is limited: the true effect may be substantially different from the estimate of the effect.
Very low certainty: We have very little confidence in the effect estimate: the true effect is likely to be substantially different from the estimate of effect.

aDowngraded one level for serious study design limitations (high risk of bias due to lack of blinding of clinicians and outcome assessment) in all trials.
bDowngraded one level for serious imprecision of effect estimate (95% CI around estimate consistent with substantial harm or benefit).

Background

Nasal continuous positive airway pressure (CPAP) is a recommended and widely used method of providing non‐invasive respiratory support for spontaneously breathing preterm infants with or at risk of respiratory distress syndrome (Lissauer 2017Beltempo 2018Sweet 2019). The most common sources of pressure generation for CPAP are underwater tube 'bubble' systems, mechanical ventilators, and Infant Flow Driver. These pressure sources may differ in their effectiveness of CPAP delivery for several reasons (Green 2019). This review focused on examining whether different sources of pressure generation (bubble CPAP versus ventilator or Infant Flow Driver) affects the risk of treatment failure and associated mortality and morbidity in preterm infants. Other Cochrane Reviews have assessed the effects of different CPAP nasal interfaces and pressure levels in preterm infants (Bamat 2021De Paoli 2021), and the effects of newer forms of non‐invasive ventilation including bilevel positive airway pressure and non‐invasive positive pressure ventilation (Lemyre 2016Lemyre 2017). The use of nasal cannulae for delivering heated and humidified air or supplemental oxygen at high flow rates to generate a distending pressure (but without an intrinsic pressure monitoring or pressure relief/blow‐off system) is also the subject of a separate Cochrane Review (Wilkinson 2016).

Description of the condition

Respiratory distress syndrome (RDS) is an important cause of morbidity and mortality in preterm infants (Fraser 2004). RDS is primarily caused by deficiency of alveolar surfactant. As most surfactant is produced after about 32 weeks' gestation, very preterm infants born before then are at high risk of developing RDS. The incidence and severity of RDS increases with decreasing gestational age at birth, occurring in more than 80% of extremely preterm infants born before 28 weeks' gestation (Stoll 2015). If left untreated, the structurally immature and surfactant‐deficient lung has a tendency to segmental collapse and atelectasis, ventilation‐perfusion mismatch, and pulmonary hypertension that worsens hypoxia and hypercarbia. Consequently, infants with severe RDS can become fatigued and apnoeic and require supplemental oxygen and assisted ventilation (Sweet 2019). Mechanical ventilation via an endotracheal tube, especially if associated with high airway pressures and high concentrations of oxygen, may cause iatrogenic injuries that contribute to the pathogenesis of bronchopulmonary dysplasia (Laughton 2011). Preterm infants who experience severe RDS are at high risk of other morbidities including pneumothorax, persistent patent ductus arteriosus, severe intraventricular haemorrhage, retinopathy of prematurity, and necrotising enterocolitis, that are associated with a prolonged need for respiratory support and hospitalisation, and with mortality and neurodevelopmental impairment (Horbar 2012).

Two major advances in perinatal care ‐ antenatal corticosteroids to stimulate endogenous surfactant production and exogenous surfactant replacement ‐ have greatly improved respiratory and other outcomes for preterm infants, particularly very preterm infants (Curstedt 2015McGoldrick 2020). Following the widespread adoption of these interventions over the past several decades, the principal form of respiratory support for preterm infants with or at risk of RDS has moved from mechanical ventilation via an endotracheal tube to non‐invasive ventilation, most commonly via nasal CPAP devices (Stoll 2015Soll 2019).

Nasal CPAP maintains low pressure (typically 5 to 8 cm H₂O) distension of the lungs when infants are breathing spontaneously and thereby increases functional residual capacity and improves oxygenation (Wright 2016). Other effects include conserving surfactant and reducing alveolar fluid, dilating the larynx to reduce supraglottic airway resistance, synchronising respiratory thoraco‐abdominal movements, and enhancing the Hering‐Breuer inflation reflex following airway occlusion (Krouskop 1975Martin 1977Yu 1977Richardson 1978Miller 1985Gaon 1999De Paoli 2005). There is evidence showing that use of nasal CPAP (compared to spontaneous breathing) reduces the risk of respiratory failure, receipt of mechanical ventilation, and mortality in preterm infants with respiratory distress (Ho 2020).

Treatment failure

Nasal CPAP and other modalities of non‐invasive respiratory support aim to prevent the iatrogenic problems associated with mechanical ventilation via an endotracheal tube and minimise ventilator‐induced lung injury and other complications (Glaser 2021). Evidence from randomised controlled trials suggests that use of nasal CPAP (compared to mechanical ventilation via an endotracheal tube) for primary respiratory support reduces the risk of bronchopulmonary dysplasia in preterm infants and reduces the need for endotracheal re‐intubation in preterm infants following a period of mechanical ventilation (Davis 2003Subramaniam 2016). The effect size of these benefits, however, is limited due to the high rate of CPAP 'treatment failure' ‐ almost half of all very preterm infants treated with nasal CPAP require endotracheal intubation and mechanical ventilation during the first week after birth (Dargaville 2016Thukral 2016). Treatment failure occurs more commonly in extremely preterm infants, and prolongs the need for respiratory support and supplemental oxygen and is associated with an increased risk of death or bronchopulmonary dysplasia (Dargaville 2013).

Several factors are thought to affect the risk of treatment failure and associated complications in preterm infants, including the CPAP interface (e.g. mask versus prongs) and pressure levels ('low' (~4 to 5 cm H₂O) versus 'high' (~7 to 8 cm H₂O)). These factors are considered in separate Cochrane Reviews (Bamat 2021De Paoli 2021). This review focused on assessing the trial evidence for the effect of different pressure sources ‐ bubble CPAP versus ventilator or Infant Flow Driver ‐ on treatment failure, and mortality and morbidity in preterm infants.

Description of the intervention

Several sources of pressure generation are available and in use (with considerable variation in practice) (Pillow 2012Gupta 2016Mukerji 2017Ekhaguere 2019):

  • underwater (water‐seal) 'bubble' CPAP (continuous flow generating an expiratory resistance depending on the depth of submersion underwater of the distal end of the expiratory limb of the circuit);

  • mechanical ventilator (continuous flow generating CPAP via expiratory limb resistance or variable‐flow CPAP using the Venturi principle to generate pressure at the nasal level via a titratable valve);

  • Infant Flow Driver (variable flow CPAP with a constant pressure set at the nasal level using a flow generator attached to device‐specific short binasal prongs).

Bubble CPAP has been in use as a means of respiratory support in preterm infants since the early 1970s (Gregory 1971). In addition to commercially manufactured devices (e.g. Fisher & Paykel), less complex bubble CPAP systems can be adapted from modified oxygen cannulae connected to a bubble bottle (Welty 2016WHO 2016). Such bubble systems are the most widely used form of CPAP in low‐ and middle‐income countries due to their simplicity of design, ease of use, and low cost (Thukral 2016Won 2019).

Commercially available mechanical ventilator CPAP devices that incorporate flow resistance valves on the expiratory limb of the nasal CPAP circuit are more commonly used in high‐income settings. In other ventilator CPAP systems, pressure is generated at the nasal level by a gas‐jet device that adjusts the gas flow based on sensed pressure (Kamper 1990). 

Infant Flow Driver (Electro Medical Equipment Ltd, Brighton, Sussex, UK; SiPAP System, CareFusion, San Diego, USA), a technologically complex CPAP system used mainly in well‐resourced settings, sets a constant distending pressure using a nasal device controlled directly by gas flow (Moa 1993Moa 1998).

How the intervention might work

The constancy of the delivered distending pressure throughout the respiratory cycle determines the effectiveness of CPAP in optimising alveolar recruitment, functional residual capacity, and (therefore) gas exchange (Pillow 2012). The pressure source (bubble versus ventilator or Infant Flow Driver) is thought to be a key determinant of CPAP stability, although several other factors including the nasal interface (e.g. masks versus prongs) can contribute (De Paoli 2021). 

Bubble CPAP, in addition to maintaining a stable end‐distending pressure, generates respiratory tract vibrations that improve gas exchange by delivering low‐amplitude, high‐frequency oscillations at the alveolar level (Lee 1998). Evidence from experimental animal models of RDS suggests that bubble CPAP improves lung recruitment and gas exchange compared with constant‐pressure CPAP (Pillow 2007). It has been suggested that the oscillations generated by bubbles in the water column might decrease ventilation‐perfusion mismatch similarly to one putative mechanism of action of high‐frequency oscillatory ventilators (Gupta 2016Gupta 2016a). Uncertainty remains, however, regarding how and to what extent the bubbling affects respiratory parameters in preterm infants with RDS (Morley 2005). There is some evidence showing that bubble CPAP may increase the resistive work of breathing and respiratory cycle asynchrony, potentially increasing the risk of treatment failure compared with ventilator CPAP, and that pressure fluctuations can be accentuated to dangerous levels by accumulations of condensate in the expiratory limb of the ventilatory circuit (Liptsen 2005Youngquist 2013). Furthermore, there is concern that transmission of oscillation and vibration to the bubble CPAP interface may increase the risk of nasal injury and trauma compared with constant‐pressure CPAP (Imbulana 2018).

Why it is important to do this review

International policy statements that exist to guide practice do not make unconditional recommendations about which pressure source to use in providing CPAP for preterm infants (Committee on Fetus and Newborn 2014). Given the possibility and plausibility that the choice of pressure source for delivering CPAP may affect the risk of treatment failure and associated mortality and morbidity in preterm infants, appraising and synthesising the trial evidence could inform practice, policy, and research.

Objectives

To assess the benefits and harms of bubble CPAP versus other pressure sources (mechanical ventilators or Infant Flow Driver) for reducing treatment failure and associated morbidity and mortality in newborn preterm infants with or at risk of respiratory distress.

Methods

Criteria for considering studies for this review

Types of studies

Randomised controlled trials (including cluster‐randomised controlled trials).

Cross‐over studies were not eligible for inclusion.

Types of participants

Preterm infants (< 37 weeks' gestation) supported with nasal CPAP, either as primary treatment for respiratory distress after birth, or following a period of mechanical ventilation (postextubation).

Types of interventions

Underwater bubble CPAP devices compared with ventilator or Infant Flow Driver CPAP devices.

Types of outcome measures

We focused on infant‐ and family‐important outcomes, principally CPAP treatment failure and neonatal morbidities that plausibly affect rates of mortality or neurodevelopmental impairment. We did not include surrogate outcomes such as physiological measures of respiratory function.

Primary outcomes
  • Treatment failure indicated by recurrent apnoea, hypoxia, hypercarbia, increasing oxygen requirement, or the receipt of mechanical ventilation within 72 hours after initiation of nasal CPAP

  • All‐cause mortality prior to hospital discharge

  • Neurodevelopmental impairment assessed by a validated test after 12 months' post‐term: neurological evaluations, developmental scores, and classifications of disability, including cerebral palsy and auditory and visual impairment

Secondary outcomes
  • Pneumothorax (including pneumomediastinum, pneumopericardium) before hospital discharge

  • Moderate‐severe nasal trauma defined by trial investigators including ulceration, bleeding, septal injury, scarring (not including hyperaemia or erythema)

  • Bronchopulmonary dysplasia: oxygen or respiratory support requirement at 36 weeks' postmenstrual age (Jobe 2001; Ehrenkranz 2005)

  • Duration of CPAP use (days)

  • Duration of oxygen supplementation (days)

  • Duration of hospitalisation (days)

  • Patent ductus arteriosus receiving medical or surgical treatment

  • Necrotising enterocolitis (Bell stage 2 or greater) (Bell 1978)

  • Severe intraventricular haemorrhage (Papile 1978)

  • Severe retinopathy of prematurity (ICROP 2005)

Search methods for identification of studies

An information specialist developed the search strategies in consultation with the review authors.

Electronic searches

We searched the following databases on 6 January 2023 without language or date restrictions:

  • Cochrane Central Register of Controlled Trials (CENTRAL; 2023, Issue 1) in the Cochrane Library (Wiley);

  • MEDLINE Ovid (1946 to 6 January 2023);

  • Embase Ovid (1974 to 6 January 2023);

  • Maternity & Infant Care Database (MIDRIS) Ovid (1971 to 6 January 2023);

  • CINAHL (Cumulative Index to Nursing and Allied Health Literature) (1982 to 6 January 2023).

The search strategies combined controlled vocabulary and text words; complete search strategies are shown in Appendix 1Appendix 2Appendix 3Appendix 4, and Appendix 5. Clinical trial filters were used, as recommended in the Cochrane Handbook for Systematic Reviews of Interventions (Higgins 2020).

Searching other resources

We searched the following clinical trials registries on 6 January 2023 for ongoing or recently completed trials: 

Search strategies for the trial registries are shown in Appendix 6.

We searched the reference lists of included studies.

Data collection and analysis

We used the standard methods of Cochrane Neonatal.

Selection of studies

One review author (WM) screened the titles and abstracts of all records identified by the search, coding each record as 'order' or 'exclude'. A second team member (RP) assessed all records coded as 'order' and made the final decision as to which records were retrieved as full‐text articles. Two review authors (WM and RP or SO) read the full texts and, using a checklist, assessed the eligibility of each article for inclusion on the basis of prespecified inclusion and exclusion criteria. One review author (ADP) checked these decisions.

Data extraction and management

Two review authors (RP and WM or SO) independently extracted data on design, methods, participants, interventions, outcomes, and treatment effects from each included study using a data collection form to aid extraction. We discussed disagreements until we reached consensus. If data from the trial reports were insufficient, we contacted trialists for further information.

Assessment of risk of bias in included studies

Two review authors (WM and RP or SO) independently assessed the risk of bias (low, high, or unclear) of all included trials using the Cochrane risk of bias tool for the following domains (Higgins 2011):

  • Sequence generation (selection bias)

  • Allocation concealment (selection bias)

  • Blinding of participants and personnel (performance bias)

  • Blinding of outcome assessment (detection bias)

  • Incomplete outcome data (attrition bias)

  • Selective reporting (reporting bias)

  • Any other bias

Any disagreements were resolved by discussion or by consulting a third assessor. For a more detailed description of risk of bias for each domain, see Appendix 7.

Measures of treatment effect

We calculated risk ratio (RR) and risk difference (RD) for dichotomous data and mean difference (MD) for continuous data, with respective 95% confidence intervals (CIs). When we deemed it appropriate to combine two or more study arms, we obtained treatment effects from combined data using the methods described in the Cochrane Handbook for Systematic Reviews of Interventions (Higgins 2020). We determined the number needed to treat for an additional beneficial outcome (NNTB) or number needed to treat for an additional harmful outcome (NNTH) for outcomes with a detected RD.

Unit of analysis issues

The unit of analysis was the participating infant in individually randomised trials. For cluster‐randomised trials (had we identified any for inclusion), we planned to undertake analyses at the level of the individual whilst accounting for clustering in the data using the methods recommended in the Cochrane Handbook for Systematic Reviews of Interventions (Higgins 2020).

Dealing with missing data

We requested additional data from trial investigators when data on important outcomes were missing or were reported unclearly. When data remained missing, we planned to examine the impact on effect size estimates by performing sensitivity analyses.

Assessment of heterogeneity

We examined treatment effects in individual trials and heterogeneity between trial results by inspecting forest plots if more than one trial was included in a meta‐analysis. We calculated the I² statistic for each analysis to quantify inconsistency across studies and to describe the percentage of variability in effect estimates that may be due to heterogeneity rather than to sampling error. If we detected moderate or high (I² > 50%) levels of heterogeneity, we explored possible causes by performing prespecified subgroup analyses (Subgroup analysis and investigation of heterogeneity).

Assessment of reporting biases

We assessed reporting bias by comparing the stated primary outcomes and secondary outcomes and the reported outcomes. Where study protocols were available, we compared these to the full publications to determine the likelihood of reporting bias. Studies using the interventions in a potentially eligible infant population but not reporting on any of the primary and secondary outcomes of this review were documented in Characteristics of included studies. We planned to use funnel plots to screen for publication bias where there was a sufficient number of trials (at least 10) reporting the outcome. If publication bias was suggested by asymmetry of the funnel plot on visual assessment, we planned to assess this statistically use Harbord's modification of Egger's test (Harbord 2006).

Data synthesis

We used a fixed‐effect model inverse variance meta‐analysis for combining data where trials examined the same intervention and the populations and methods of the trials were judged to be similar.

Subgroup analysis and investigation of heterogeneity

We prespecified subgroup analyses of bubble CPAP versus:

  • ventilator CPAP;

  • Infant Flow Driver CPAP.

We planned to explore moderate or high heterogeneity (I² > 50%) in subgroup analyses stratified by:

  • timing of nasal CPAP: primary support after birth versus postextubation;

  • CPAP levels: 'low' (~4 to 5 cm H₂O) versus 'higher' (> 5 cm H₂O);

  • gestation or birthweight: preterm or low birthweight versus very preterm or very low birthweight;

  • setting: low‐ or middle‐income versus high‐income countries (World Bank 2021).

Sensitivity analysis

We planned to perform sensitivity analyses if:

  • there was unexplained moderate or high heterogeneity (I² > 50%) by removing the outlying trial or trials;

  • a trial with high risk of bias (including high level of missing outcome data) was included in the meta‐analysis of an outcome where the other studies were at low risk of bias (removing the study with high risk of bias).

Summary of findings and assessment of the certainty of the evidence

Two review authors (SO and WM) used the GRADE approach as outlined in the GRADE Handbook to assess the certainty of the evidence for the following outcomes (Schünemann 2013).

  • Treatment failure

  • All‐cause mortality prior to hospital discharge

  • Neurodevelopmental impairment

  • Moderate‐severe nasal injury

  • Pneumothorax

  • Bronchopulmonary dysplasia

We considered evidence from randomised controlled trials to be of high certainty, and downgraded one level for serious (or two levels for very serious) limitations based upon: design (risk of bias); consistency across trials; directness of the evidence; precision of estimates; and presence of publication bias (Schünemann 2013Walsh 2021). We used GRADEpro GDT software to create a summary of findings table to report the certainty of the evidence (GRADEpro GDT).

The GRADE approach results in an assessment of the certainty of a body of evidence as one of four grades, as follows.

  • High: we are very confident that the true effect lies close to that of the estimate of the effect.

  • Moderate: we are moderately confident in the effect estimate: the true effect is likely to be close to the estimate of the effect, but there is a possibility that it is substantially different.

  • Low: our confidence in the effect estimate is limited: the true effect may be substantially different from the estimate of the effect.

  • Very low: we have very little confidence in the effect estimate: the true effect is likely to be substantially different from the estimate of effect.

Results

Description of studies

See: Characteristics of included studiesCharacteristics of excluded studies.

Results of the search

Our database searches identified 10,164 references, and our searches of the trial registries identified 864 records. After removal of 4438 duplicates, 6590 records were available for screening, of which 6548 were excluded based on title/abstract review. We assessed 42 full‐texts, excluding 26 reports (Characteristics of excluded studies). We included 15 studies (Characteristics of included studies) in the quantitative synthesis. We classified one conference abstract as awaiting classification (Characteristics of studies awaiting classification). Details are provided in Figure 1.

1.

1

Study flow diagram.

Included studies

We included 15 trials involving a total of 1437 infants (Characteristics of included studies). Most trials were small (median number of participants 88). The trials were conducted during the past 25 years in neonatal centres in India (5 trials), Iran (4 trials), Brazil (2 trials), Albania (1 trial), Armenia (1 trial), the UK (1 trial), and Italy (1 trial). Individual infants were allocated randomly to intervention or control groups in all of the trials. No studies used a cluster‐randomised design.

All trials used a commercially available bubble CPAP device; none used a low‐cost, locally adapted form of bubble CPAP. 

Eleven trials compared bubble CPAP with ventilator CPAP (Tagare 2010; Bahman‐Bijari 2011; Mohammadizadeh 2011; Yagui 2011; Hosseini 2012; Yadav 2012; Tagare 2013; Bhatti 2015; Agarwal 2016; Noori Shadkam 2017; Ribeiro 2017). Four trials compared bubble CPAP with Infant Flow Driver CPAP (Mazzella 2001; Gupta 2009; Jazexhiu‐Postoli 2015; Mazmanyan 2016).

Excluded studies

We excluded 26 reports (Characteristics of excluded studies). The most common reasons for exclusion were wrong study design (non‐randomised or cross‐over) or wrong intervention (did not include bubble CPAP as a comparison group).

Risk of bias in included studies

Methodologic quality varied between the trials (Figure 2). All trials had unclear or high risk of bias in at least one domain.

2.

2

Risk of bias summary: review authors' judgements about each risk of bias item for each included study.

Allocation

The methods used to generate the random sequence and conceal allocation were not described in about half of the trials. The other trials used computer or web‐based programs to generate the random sequence and sealed, opaque envelopes to conceal allocation.

Blinding

All trials were 'open‐label' ‐ none blinded parents, clinicians, or investigators.

Incomplete outcome data

Most trials reported complete or near‐complete assessments of primary outcomes. We assessed one trial as at unclear risk of attrition bias.

Selective reporting

Most trials reported a comprehensive group of infant‐important outcomes. We assessed two trials as at unclear risk of reporting bias as they reported few clinical outcomes.

Other potential sources of bias

We did not find evidence of between‐group baseline differences in participant characteristics or demographics in most of the trials. Gestation and weight at birth differed between groups in two trials, and baseline characteristics were not presented in two trials (unclear risk).

Effects of interventions

See: Table 1

Primary outcomes

Treatment (CPAP) failure

Meta‐analysis of data from 13 trials (1230 infants) suggests that bubble (versus ventilator/Infant Flow Driver) CPAP may reduce the risk of treatment failure slightly (Analysis 1.1):

1.1. Analysis.

1.1

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 1: Treatment failure

  • risk ratio (RR) 0.76, 95% confidence interval (CI) 0.60 to 0.95 (I² = 31%);

  • risk difference (RD) −0.05, 95% CI −0.10 to −0.01;

  • number needed to treat for an additional beneficial outcome (NNTB) 20, 95% CI 10 to 100.

There was no evidence of funnel plot asymmetry sufficient to suggest publication bias (Figure 3).

3.

3

Forest plot of comparison: 1 Bubble CPAP versus ventilator or Infant Flow Driver CPAP, outcome: 1.1 Treatment failure.

We assessed the certainty of evidence as low, downgrading one level for serious study design limitations (lack of blinding) and one level for imprecision (Table 1).

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 3.38, df = 1 (P = 0.07), I² = 70.4% (Figure 4).

4.

4

Subgroup analysis for heterogeneity

Not applicable (I² = 31%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 31%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

All‐cause mortality prior to hospital discharge

Meta‐analysis of data from 10 trials (1189 infants) suggests that bubble (versus ventilator/Infant Flow Driver) CPAP may not affect the risk of mortality prior to hospital discharge (Analysis 1.2):

1.2. Analysis.

1.2

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 2: Mortality before discharge

  • RR 0.93, 95% CI 0.64 to 1.36 (I² = 0%);

  • RD −0.01, 95% CI −0.04 to 0.02.

There was no evidence of funnel plot asymmetry sufficient to suggest publication bias (Figure 5).

5.

5

Forest plot of comparison: 1 Bubble CPAP versus ventilator or Infant Flow Driver CPAP, outcome: 1.2 Mortality before discharge.

We assessed the certainty of evidence as low, downgrading one level for serious study design limitations (lack of blinding) and one level for imprecision (Table 1).

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 1.24, df = 1 (P = 0.27), I² = 19.2% (Figure 6).

6.

6

Subgroup analysis for heterogeneity

Not applicable (I² = 0%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 0%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Moderate‐severe neurodevelopmental impairment

None of the trials assessed neurodevelopmental outcomes.

Secondary outcomes

Pneumothorax

Meta‐analysis of data from 14 trials (1340 infants) suggests that bubble (versus ventilator/Infant Flow Driver) CPAP may not affect the risk of pneumothorax (Analysis 1.3):

1.3. Analysis.

1.3

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 3: Pneumothorax

  • RR 0.73, 95% CI 0.40 to 1.34 (I² = 0%);

  • RD −0.01, 95% CI −0.03 to 0.01.

There was no evidence of funnel plot asymmetry sufficient to suggest publication bias (Figure 7).

7.

7

Forest plot of comparison: 1 Bubble CPAP versus ventilator or Infant Flow Driver CPAP, outcome: 1.3 Pneumothorax.

We assessed the certainty of evidence as low, downgrading one level for serious study design limitations (lack of blinding) and one level for imprecision (Table 1).

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 1.00, df = 1 (P = 0.32), I² = 0%.

Subgroup analysis for heterogeneity

Not applicable (I² = 0%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 0%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Moderate‐severe nasal injury

Meta‐analysis of data from 8 trials (753 infants) shows that bubble (versus ventilator/Infant Flow Driver) CPAP likely increases the risk of moderate‐severe nasal injury slightly (Analysis 1.4):

1.4. Analysis.

1.4

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 4: Moderate‐severe nasal injury

  • RR 2.29, 95% CI 1.37 to 3.82) (I² = 17%);

  • RD 0.07, 95% CI 0.03 to 0.11;

  • number needed to treat for an additional harmful outcome (NNTH) 14, 95% CI 9 to 33.

There were insufficient data points to assess funnel plot asymmetry.

We assessed the certainty of evidence as moderate, downgrading one level for serious study design limitations (lack of blinding) (Table 1).

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 1.68, df = 1 (P = 0.19), I² = 40.5%.

Subgroup analysis for heterogeneity

Not applicable (I² = 17%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 17%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Bronchopulmonary dysplasia

Meta‐analysis of data from seven trials (603 infants) suggests that bubble (versus ventilator/Infant Flow Driver) CPAP may not affect the risk of bronchopulmonary dysplasia (Analysis 1.5):

1.5. Analysis.

1.5

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 5: Bronchopulmonary dysplasia

  • RR 0.76, 95% CI 0.53 to 1.10 (I² = 0%);

  • RD −0.04, 95% CI −0.09 to 0.01.

There were insufficient data points to assess funnel plot asymmetry.

We assessed the certainty of evidence as low, downgrading one level for serious study design limitations (lack of blinding), and one level for imprecision (Table 1).

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 0.46, df = 1 (P = 0.50), I² = 0%.

Subgroup analysis for heterogeneity

Not applicable (I² = 0%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 0%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Duration of CPAP use

Meta‐analysis of data from eight trials (744 infants) that reported mean and standard deviation (SD) for each group (bubble versus ventilator/Infant Flow Driver) suggests little or no effect:

  • mean difference (MD) −0.01, 95% CI −0.16 to 0.13 days (I² = 67%) (Analysis 1.6).

1.6. Analysis.

1.6

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 6: Duration of CPAP use

There were insufficient data points to assess funnel plot asymmetry.

There was evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 4.81, df = 1 (P = 0.03), I² = 79.2% (larger effect for bubble CPAP versus Infant Flow Driver CPAP than for bubble CPAP versus ventilator CPAP).

Subgroup analysis for heterogeneity
  • CPAP levels: 'low' (up to 5 cm H₂O) versus 'higher' (> 5 cm H₂O): subgroup analysis not feasible, as trials described a range of CPAP levels that typically included both of these ranges.

  • Gestation or birthweight: preterm or low birthweight versus very preterm or very low birthweight: subgroup analysis not feasible, as trials described a range of gestational age or birthweight that typically included both of these ranges.

  • Setting: low‐ and middle‐income versus high‐income countries (World Bank 2021): seven trials took place in low‐ or middle‐income countries (Albania, India, Iran), whilst one trial took place in a high‐income country (Italy). There was no evidence of a subgroup effect (Chi² = 0.59, df = 1 (P = 0.44), I² = 0%).

Sensitivity analysis for heterogeneity 

Removal of the outlying trial, Jazexhiu‐Postoli 2015, reduced the level of heterogeneity. Meta‐analysis of the seven remaining trials suggests that bubble (versus ventilator/Infant Flow Driver) CPAP may have little or no effect:

  • MD 0.05, 95% CI −0.11 to 0.20 days (I² = 39%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Six trials reported median difference in duration of CPAP use (bubble versus ventilator/Infant Flow Driver), as follows.

Primary support:
Postextubation support:

One trial did not report duration of CPAP use (Yadav 2012).

Duration of oxygen supplementation

Analysis of data from one trial (88 infants) that reported mean and SD for each group (bubble versus ventilator CPAP) suggests little or no effect (Mohammadizadeh 2011):

1.7. Analysis.

1.7

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 7: Duration of oxygen supplementation

One trial did not show a difference in the median duration of oxygen supplementation with bubble versus ventilator/Infant Flow Driver CPAP as postextubation support (Ribeiro 2017).

The other 13 trials did not report duration of oxygen supplementation.

Duration of hospitalisation

Meta‐analysis of data from five trials (591 infants) that reported mean and SD for each group suggests that bubble (versus ventilator/Infant Flow Driver) CPAP may reduce the duration of hospitalisation:

  • MD −3.27, 95% CI −4.99 to −1.56 days (I² = 32%) (Analysis 1.8).

1.8. Analysis.

1.8

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 8: Duration of hospitalisation

There were insufficient data points to assess funnel plot asymmetry.

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 2.36, df = 1 (P = 0.12), I² = 57.7%.

Subgroup analysis for heterogeneity

Not applicable (I² = 32%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 32%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Two trials did not show a difference in the median duration of hospitalisation with bubble versus ventilator/Infant Flow Driver CPAP, as follows.

The other eight trials did not report duration of hospitalisation.

Patent ductus arteriosus

Meta‐analysis of data from six trials (597 infants) suggests little or no effect of bubble versus ventilator/Infant Flow Driver CPAP (Analysis 1.9):

1.9. Analysis.

1.9

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 9: Patent ductus arteriosus

  • RR 0.96, 95% CI 0.67 to 1.38 (I² = 0%);

  • RD −0.01, 95% CI −0.06 to 0.05.

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 0.60, df = 1 (P = 0.44), I² = 0%.

Subgroup analysis for heterogeneity

Not applicable (I² = 0%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 0%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Necrotising enterocolitis

Meta‐analysis of data from six trials (693 infants) suggests little or no effect of bubble versus ventilator/Infant Flow Driver CPAP (Analysis 1.10):

1.10. Analysis.

1.10

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 10: Necrotising enterocolitis

  • RR 0.75, 95% CI 0.39 to 1.44 (I² = 0%);

  • RD −0.01, 95% CI −0.05 to 0.02.

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 0.74, df = 1 (P = 0.39), I² = 0%.

Subgroup analysis for heterogeneity

Not applicable (I² = 0%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 0%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Severe intraventricular haemorrhage

Meta‐analysis of data from six trials (562 infants) suggests little or no effect of bubble versus ventilator/Infant Flow Driver CPAP (Analysis 1.11):

1.11. Analysis.

1.11

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 11: Severe intraventricular haemorrhage

  • RR 0.81, 95% CI 0.42 to 1.58 (I² = 0%);

  • RD −0.01, 95% CI −0.05 to 0.03.

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 2.01, df = 1 (P = 0.16), I² = 50.2%.

Subgroup analysis for heterogeneity

Not applicable (I² = 0%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 0%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Severe retinopathy of prematurity

Meta‐analysis of data from six trials (642 infants) suggests little or no effect of bubble versus ventilator/Infant Flow Driver CPAP (Analysis 1.12):

1.12. Analysis.

1.12

Comparison 1: Bubble CPAP versus ventilator or Infant Flow Driver CPAP, Outcome 12: Severe retinopathy of prematurity

  • RR 0.92, 95% CI 0.54 to 1.57 (I² = 0%);

  • RD −0.01, 95% CI −0.05 to 0.03.

There was no evidence of a subgroup difference for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP: Chi² = 0.08, df = 1 (P = 0.77), I² = 0%.

Subgroup analysis for heterogeneity

Not applicable (I² = 0%).

Sensitivity analysis for heterogeneity 

Not applicable (I² = 0%).

Sensitivity analysis for risk of bias

The meta‐analysis did not contain data from a trial with high risk of bias where the other studies had low risk of bias.

Discussion

Summary of main results

This review of 15 trials, with 1437 participants in total, suggests that use of bubble compared with ventilator or Infant Flow Driver devices as the pressure source for nasal CPAP in preterm infants may reduce the rate of treatment failure by about one‐quarter. The point estimate indicates that, for every 1000 infants treated with bubble CPAP compared with ventilator or Infant Flow Driver devices, there will be 50 fewer episodes of treatment failure. 

The available data suggest that the type of pressure source may not affect mortality prior to hospital discharge or the risk of pneumothorax or bronchopulmonary dysplasia. None of the included trials reported the effect on neurodevelopmental impairment. Meta‐analysis suggests that bubble CPAP likely increases the risk of moderate‐severe nasal injury. Bubble CPAP may be associated with reduced duration of hospitalisation of about three days. Other outcomes such as major morbidities (patent ductus arteriosus, necrotising enterocolitis, intraventricular haemorrhage, retinopathy of prematurity) and duration of CPAP or supplemental oxygen use appear not to be influenced by type of pressure source. However, the numbers of participants and trials to date are low, and the estimates of effect imprecise.

Overall completeness and applicability of evidence

Most of the trials were undertaken within the past 25 years in healthcare facilities internationally, predominantly in middle‐income countries (9 of 15 trials occurred in India or Iran). No trials were conducted in sub‐Saharan Africa. Most participants were very preterm infants; few were extremely preterm or extremely low birthweight. 

The mechanism whereby bubble CPAP may reduce the rate of treatment failure is unclear. Bubble CPAP generates oscillatory waves that may transmit more effectively the prescribed pressure to the airway and alveoli than does constant‐pressure CPAP. However, it remains unclear whether or how this physiological phenomenon affects important respiratory parameters in preterm infants (Liptsen 2005). Nevertheless, there were between‐trial differences in the definition of CPAP treatment failure (including a broad range of fraction of inspired oxygen (FiO₂) thresholds) that limit the generalisability of the findings. Similarly, although transmission of oscillations and vibration is one putative mechanism by which bubble versus ventilator or Infant Flow Driver CPAP may increase the risk of moderate‐severe nasal injury, this effect may be due in part to the confounding influence of different types of nasal interfaces used with the different CPAP pressure sources (De Paoli 2021).

Relevance to resource‐limited settings

Various CPAP devices were studied across the included trials, and the findings appear broadly applicable to current care practices for preterm infants receiving bubble or ventilator or Infant Flow Driver CPAP. Prespecified analyses did not show evidence of different subgroup effects for bubble CPAP versus ventilator CPAP compared with bubble CPAP versus Infant Flow Driver CPAP. 

However, a major limitation to the applicability of these findings to resource‐limited settings is that the bubble CPAP system used in most of the trials was a commercially manufactured device (mostly Fisher & Paykel). Such devices are not affordable in many resource‐limited settings in low‐ or middle‐income countries, and low‐cost, locally manufactured or locally adapted devices are more commonly used (WHO 2016). These 'indigenous' systems, based on the original bubble CPAP device, consist of a wide‐bore inspiratory limb with a low‐resistance interface to generate stable positive end‐expiratory pressure dependent on the depth of submersion of the expiratory tubing (Baldursdottir 2020). There is some concern that these systems deliver CPAP variably depending on respiratory tubing and flow rates employed (Ettinger 2021). However, none of the included trials compared the effect of these low‐cost systems with ventilator or Infant Flow Driver CPAP. 

Consequently, this review does not address concerns about the benefits and harms of low‐cost, locally manufactured or locally adapted bubble CPAP systems in resource‐limited settings. These concerns include uncertainty about the effect of instability in airway pressure due to condensate in the expiratory limb of the respiratory circuit. Since low‐cost bubble CPAP systems typically lack pressure alarms and pressure‐release valves, potentially harmful high levels of positive end‐expiratory pressure may be undetected (Youngquist 2013). Another major concern in some settings is that locally adapted bubble CPAP systems can provide only pure (100%) oxygen due to unavailability of oxygen‐air blenders. As a result, if not monitored, infants may be exposed to hyperoxia that may contribute to the epidemic of retinopathy of prematurity in many low‐ or middle‐income countries (Vinekar 2019). 

Quality of the evidence

We used GRADE methods to assess the certainty of the evidence for effects on treatment failure, all‐cause mortality prior to hospital discharge, neurodevelopmental impairment (no data), pneumothorax, moderate‐severe nasal injury, and bronchopulmonary dysplasia (Table 1). Using this framework, we downgraded the certainty of the evidence from high to low because of methodological weaknesses (risk of bias), in particular uncertainty about the methods used to generate the random sequence and to conceal treatment allocation in about half of the trials, and lack of blinding measures for parents, caregivers, and clinical assessors in all the trials. The methodological weaknesses may have introduced selection, performance, and detection biases. For example, it is possible that bias in the (unrecorded) use of co‐interventions such as methylxanthines may have differed between treatment groups, or that surveillance and detection bias for subjective outcomes was introduced (e.g. in checking for nasal injury more often in infants allocated to one pressure source or CPAP device versus another). 

The other reason for downgrading the certainty of evidence across all outcomes was the existence of substantial imprecision in the estimates of effect. Meta‐analyses generated 95% CI that included large benefit as well as small or no benefit or harm. Although the total number of participants in the 15 included trials was more than 1400, not all trials contributed data to all outcome estimates, and as a result estimates of effect were imprecise, especially for less common outcomes, including mortality. For example, the point estimate for the NNTB for treatment failure was 20, but the upper bound of the 95% CI included an NNTB of 100. Such imprecise estimates of effect are unlikely to meaningfully inform decision‐making in this context.

Heterogeneity was not an apparent limitation in the meta‐analyses, and planned subgroup analyses to investigate sources of moderate or high heterogeneity (I² > 50%) in primary outcomes were not conducted. The absence of such levels of heterogeneity in the meta‐analyses may be related to similarity of the trial settings and populations studied, and although a variety of different pressure sources (several types of mechanical ventilator) were used, these are likely to be broadly similar in their mechanisms of action and physiological effects.  

Potential biases in the review process

An important concern with the review process is the possibility that the findings are subject to publication and other reporting biases (Hopewell 2009). Data from trials which show statistically significant or potentially important effects tend to be more readily available for inclusion in meta‐analyses (Gale 2020). Publication bias, as well as other sources of small‐study bias, is an important contributor to inflation of effect size estimates in meta‐analyses of interventions to improve outcomes in preterm infants (Walsh 2021).

We did not show funnel plot asymmetry suggesting that publication bias (or other types of small‐study biases) exaggerated the effect size in any of the meta‐analyses that contained sufficient data points (Higgins 2020). Although we attempted to minimise the threat of publication bias by screening the reference lists of included trials and related reviews, and searching the proceedings of the major international perinatal conferences to identify trial reports that were not published in full form in academic journals, we cannot be sure that other trials have been undertaken but not reported.

Agreements and disagreements with other studies or reviews

This review is in broad agreement with another recent systematic review of randomised controlled trials that assessed the effects of bubble versus other pressure sources for delivering CPAP for preterm infants (Bharadwaj 2020). Similar to our findings, Bharadwaj 2020 showed that bubble CPAP was associated with a reduced risk of CPAP failure (within seven days). There does not appear to be an effect on mortality or bronchopulmonary dysplasia, but there may be an increase in the risk of nasal injury.

Authors' conclusions

Implications for practice.

Given the low certainty of the evidence provided by these analyses, the implications for practice remain uncertain. Although the available trial data do suggest that use of bubble CPAP may reduce the risk of treatment failure compared with ventilator or Infant Flow Driver CPAP, because of design limitations (risk of bias) and the paucity of data that could be abstracted from published reports (imprecision), it remains unclear how or if this translates to effects on other important outcomes including mortality, neurodevelopmental impairment, and other major morbidities including bronchopulmonary dysplasia.

Implications for research.

Well‐designed trials evaluating this important aspect of a recommended and commonly used neonatal therapy are needed. Trials reporting infant‐important endpoints such as the primary outcomes of this review will facilitate updating of evidence syntheses. Although blinding of clinical investigators to treatment allocation is likely to be impractical, trials should aim to minimise performance or detection bias, for example by strict and consistent application of protocols for management and criteria for subjective diagnoses such as treatment failure.

In high‐income countries with well‐resourced healthcare facilities, evaluating the comparative effects of different pressure source devices for CPAP may be particularly relevant to extremely preterm or extremely low birthweight infants at high risk of treatment failure and associated complications including bronchopulmonary dysplasia. However, in these settings research priorities may already have shifted towards comparative studies with the newer forms of non‐invasive ventilation (including nasal intermittent positive pressure ventilation and humidified high‐flow nasal cannulae) that are increasingly being adopted in practice (Mukerji 2017). These devices are considered in separate Cochrane Reviews (Wilkinson 2016Lemyre 2017). Furthermore, the clinical and research context for non‐invasive ventilation, particularly in well‐resourced facilities, has been affected by other innovations including the early use of 'less‐invasive surfactant therapy', which is associated with reduced risk of death or bronchopulmonary dysplasia compared with surfactant therapy via an endotracheal tube and continued mechanical ventilation (Abdel‐Latif 2021).

In low‐income countries with fewer and scarcer healthcare resources, the infant population most likely to be affected are more mature preterm infants in whom CPAP may be life‐saving in the absence of intensive care and additional therapies including surfactant and mechanical ventilation (Kinshella 2020). Furthermore, in these settings the bubble CPAP devices of most interest are low‐cost, locally manufactured or locally adapted devices rather than commercially manufactured, technologically complex bubble CPAP machines (WHO 2016Won 2019). However, the feasibility and ethical propriety of conducting trials comparing the use of low‐cost bubble CPAP to the best available care practices (including low‐flow oxygen without CPAP) is a matter of debate (Ekhaguere 2019).

Acknowledgements

The Methods of this review are based on a standard template used by Cochrane Neonatal.

We thank the corresponding authors who provided further information about their trials.

We thank Cochrane Neonatal: Jane Cracknell and Michelle Fiander, Managing Editors, Roger Soll, Co‐ordinating Editor, and Matteo Bruschettini, Associate Editor, who provided editorial and administrative support.

We thank Melissa Harden, Information Specialist, who designed and ran the literature searches.

We thank Colleen Ovelman, previously Managing Editor Cochrane Neonatal, for logistical support and for screening the electronic search records for previous versions of this review.

We thank peer reviewers Sivam Thanigainathan, Department of Neonatology, AIIMS Jodhpur, India, and Rebecca Naples, Royal Victoria Infirmary, Newcastle upon Tyne, UK, for their constructive comments and suggestions.

We thank Lisa Winer, Central Production Service, Cochrane, for copy editing the review.

Appendices

Appendix 1. Cochrane Central Register of Controlled Trials (CENTRAL) strategy

Cochrane Central Register of Controlled Trials (CENTRAL)

via Wiley http://onlinelibrary.wiley.com/ 

Date range: Issue 12 of 12, December 2022 

Date searched: 6th January 2023

Records retrieved: 2763

 

#1 [mh "Infant, Newborn"] 17815

#2 [mh ^"Premature Birth"] 1827

#3 (neonat* or neo NEXT nat*):ti,ab,kw 26271

#4 (newborn* or new NEXT born* or newly NEXT born*):ti,ab,kw 31671

#5 (preterm or preterms or pre NEXT term*1):ti,ab,kw 15821

#6 (preemie* or premie or premies):ti,ab,kw 56

#7 (prematur* NEAR/3 (birth* or born or deliver*)):ti,ab,kw 3471

#8 (low NEAR/3 (birthweight* or birth NEXT weight*)):ti,ab,kw 5996

#9 low NEXT birthweight*:ti,ab,kw 969

#10 (LBW or VLBW or ELBW):ti,ab,kw 1845

#11 infan*:ti,ab,kw 71072

#12 (baby or babies):ti,ab,kw 10445

#13 {OR #1‐#12} 90884

#14 [mh "Positive‐Pressure Respiration"] 3068

#15 ((((continuous* or positive) NEAR/3 pressure*) or (positive NEXT pressure* or PAP)) and (airway* or air NEXT way* or breath*1 or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)):ti,ab,kw 11077

#16 (((airway* or air NEXT way*) NEAR/3 pressure*) and (breath*1 or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)):ti,ab,kw 6212

#17 ((PPV or CPAP or C NEXT PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (airway* or air NEXT way* or breath*1 or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)):ti,ab,kw 5677

#18 ((((continuous* or positive or airway* or air NEXT way*) NEAR/3 pressure*) or positive NEXT pressure*) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face NEXT mask* or headgear* or head NEXT gear or headbox or head NEXT box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth NEXT piece* or nebuli?er* or prong*1)):ti,ab,kw 4175

#19 ((PAP or PPV or CPAP or C NEXT PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face NEXT mask* or headgear* or head NEXT gear or headbox or head NEXT box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth NEXT piece* or nebuli?er* or prong*1)):ti,ab,kw 2516

#20 {OR #14‐#19} 14058

#21 #13 AND #20 in Trials 2763

 

Key:

mh = exploded indexing term (MeSH)

mh ^ = indexing term (MeSH)

* = truncation

? = one additional letter

ti,ab,kw = terms in either title or abstract or keyword fields

near/3 = terms within three words of each other (any order)

next = terms are next to each other.

Appendix 2. MEDLINE strategy

Ovid MEDLINE(R) ALL 

via Ovid http://ovidsp.ovid.com/ 

Date range searched: <1946 to January 05, 2023>

Date searched: 6th January 2023

Records retrieved: 2414

 

1 exp Infant, Newborn/ (664702)

2 Premature Birth/ (19913)

3 (neonat* or neo nat* or neo‐nat*).ti,ab,kw,kf. (306658)

4 (newborn* or new born* or new‐born* or newly born* or newly‐born*).ti,ab,kw,kf. (199194)

5 (preterm or preterms or pre term or pre terms or pre‐term or pre‐terms).ti,ab,kw,kf. (92142)

6 (preemie* or premie or premies).ti,ab,kw,kf. (217)

7 (prematur* adj3 (birth* or born or deliver*)).ti,ab,kw,kf. (18646)

8 (low adj3 (birthweight* or birth weight* or birth‐weight*)).ti,ab,kw,kf. (40109)

9 low‐birthweight*.ti,ab,kw,kf. (8462)

10 (LBW or VLBW or ELBW).ti,ab,kw,kf. (10131)

11 infan*.ti,ab,kw,kf. (541458)

12 (baby or babies).ti,ab,kw,kf. (79472)

13 or/1‐12 (1204779)

14 exp Positive‐Pressure Respiration/ (28759)

15 ((((continuous* or positive) adj3 pressure*) or (positive‐pressure* or PAP)) and (airway* or air‐way* or breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,kw,kf. (31221)

16 (((airway* or air‐way*) adj3 pressure*) and (breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,kw,kf. (14702)

17 ((PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (airway* or air‐way* or breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,kw,kf. (12732)

18 ((((continuous* or positive or airway* or air‐way*) adj3 pressure*) or positive‐pressure*) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong?)).ti,ab,kw,kf. (8333)

19 ((PAP or PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong?)).ti,ab,kw,kf. (7027)

20 or/14‐19 (54118)

21 13 and 20 (8193)

22 randomized controlled trial.pt. (583819)

23 controlled clinical trial.pt. (95151)

24 randomized.ab. (587721)

25 placebo.ab. (234504)

26 drug therapy.fs. (2559741)

27 randomly.ab. (398865)

28 trial.ab. (629833)

29 groups.ab. (2456292)

30 or/22‐29 (5549711)

31 21 and 30 (2651)

32 exp animals/ not humans.sh. (5079099)

33 31 not 32 (2454)

34 remove duplicates from 33 (2414)

 

Key:

/ or.sh. = indexing term (Medical Subject Heading: MeSH)

exp = exploded indexing term (MeSH)

$ or * = truncation

? = one additional letter

ti,ab,kw,kf = terms in either title, abstract, keyword heading or keyword heading word fields

fs = floating subheading

adj3 = terms within three words of each other (any order).

pt = publication type

Appendix 3. Embase strategy

Embase

via Ovid http://ovidsp.ovid.com/ 

Date range searched: < 1974 to 2023 January 05>

Date searched: 6th January 2023

Records retrieved: 3438

 

1 exp infant/ (1102101)

2 prematurity/ (120734)

3 (neonat* or neo nat* or neo‐nat*).ti,ab,kw,kf. (401481)

4 (newborn* or new born* or new‐born* or newly born* or newly‐born*).ti,ab,kw,kf. (225075)

5 (preterm or preterms or pre term or pre terms or pre‐term or pre‐terms).ti,ab,kw,kf. (130056)

6 (preemie* or premie or premies).ti,ab,kw,kf. (351)

7  (prematur* adj3 (birth* or born or deliver*)).ti,ab,kw,kf. (26326)

8 (low adj3 (birthweight* or birth weight* or birth‐weight*)).ti,ab,kw,kf. (51868)

9 low‐birthweight*.ti,ab,kw,kf. (10263)

10 (LBW or VLBW or ELBW).ti,ab,kw,kf. (13973)

11  infan*.ti,ab,kw,kf. (574778)

12 (baby or babies).ti,ab,kw,kf. (111989)

13 or/1‐12 (1566468)

14 exp positive pressure ventilation/ (17447)

15 ((((continuous* or positive) adj3 pressure*) or (positive‐pressure* or PAP)) and (airway* or air‐way* or breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,kw,kf. (45629)

16 (((airway* or air‐way*) adj3 pressure*) and (breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,kw,kf. (21835)

17 ((PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (airway* or air‐way* or breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,kw,kf. (23786)

18 ((((continuous* or positive or airway* or air‐way*) adj3 pressure*) or positive‐pressure*) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong?)).ti,ab,kw,kf. (13046)

19 ((PAP or PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong?)).ti,ab,kw,kf. (13399)

20 or/14‐19 (77182)

21 13 and 20 (11985)

22 randomized controlled trial/ (744605)

23 controlled clinical trial/ (467918)

24 Random$.ti,ab,ot. (1876296)

25 randomization/ (95926)

26 intermethod comparison/ (291127)

27 placebo.ti,ab,ot. (352175)

28 (compare or compared or comparison).ti,ot. (583643)

29 ((evaluated or evaluate or evaluating or assessed or assess) and (compare or compared or comparing or comparison)).ab. (2630778)

30 (open adj label).ti,ab,ot. (102877)

31 ((double or single or doubly or singly) adj (blind or blinded or blindly)).ti,ab,ot. (264696)

32 double blind procedure/ (202377)

33 parallel group$1.ti,ab,ot. (30678)

34 (crossover or cross over).ti,ab,ot. (120062)

35 ((assign$ or match or matched or allocation) adj5 (alternate or group or groups or intervention or interventions or patient or patients or subject or subjects or participant or participants)).ti,ab,ot. (396360)

36 (assigned or allocated).ti,ab,ot. (467774)

37 (controlled adj7 (study or design or trial)).ti,ab,ot. (428081)

38 (volunteer or volunteers).ti,ab,ot. (274706)

39 human experiment/ (609337)

40 trial.ti,ot. (379354)

41 or/22‐40 (6032343)

42  21 and 41 (3808)

43 (rat or rats or mouse or mice or swine or porcine or murine or sheep or lambs or pigs or piglets or rabbit or rabbits or cat or cats or dog or dogs or cattle or bovine or monkey or monkeys or trout or marmoset$).ti,ot. and animal experiment/ (1181493)

44 Animal experiment/ not (human experiment/ or human/) (2480517)

45 43 or 44 (2544025)

46 42 not 45 (3533)

47 remove duplicates from 46 (3438)

 

Key:

/ or.sh. = indexing term (Emtree Subject Heading)

exp = exploded indexing term (Emtree)

$ or * = truncation

? = one additional letter

ti,ab,kw,kf = terms in either title, abstract, keyword heading or keyword heading word fields

adj3 = terms within three words of each other (any order).

pt = publication type

ot = original title

Appendix 4. MIDIRS strategy

Maternity & Infant Care Database (MIDIRS)

via Ovid http://ovidsp.ovid.com/ 

Date range searched: <1971 to December 13, 2022>

Date searched: 6th January 2023

Records retrieved: 143

 

1 (neonat* or neo nat* or neo‐nat*).ti,ab,hw,de. (57665)

2 (newborn* or new born* or new‐born* or newly born* or newly‐born*).ti,ab,hw,de. (45135)

3 (preterm or preterms or pre term or pre terms or pre‐term or pre‐terms).ti,ab,hw,de. (32763)

4 (preemie* or premie or premies).ti,ab,hw,de. (68)

5 (prematur* adj3 (birth* or born or deliver*)).ti,ab,hw,de. (7706)

6 (low adj3 (birthweight* or birth weight* or birth‐weight*)).ti,ab,hw,de. (13479)

7 low‐birthweight*.ti,ab,hw,de. (3490)

8 (LBW or VLBW or ELBW).ti,ab,hw,de. (3643)

9 infan*.ti,ab,hw,de. (101933)

10 (baby or babies).ti,ab,hw,de. (33202)

11 or/1‐10 (152409)

12 ((((continuous* or positive) adj3 pressure*) or (positive‐pressure* or PAP)) and (airway* or air‐way* or breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,hw,de. (1465)

13 (((airway* or air‐way*) adj3 pressure*) and (breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,hw,de. (1092)

14 ((PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (airway* or air‐way* or breath? or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)).ti,ab,hw,de. (794)

15 ((((continuous* or positive or airway* or air‐way*) adj3 pressure*) or positive‐pressure*) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong?)).ti,ab,hw,de. (396)

16 ((PAP or PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong?)).ti,ab,hw,de. (259)

17 or/12‐16 (1768)

18 11 and 17 (1699)

19 limit 18 to randomised controlled trial (143)

 

Key:

/ or.sh. = indexing term (Emtree Subject Heading)

exp = exploded indexing term (Emtree)

$ or * = truncation

? = one additional letter

ti,ab,hw,de = terms in either title, abstract, heading word, or descriptor fields

adj3 = terms within three words of each other (any order).

 

Appendix 5. CINAHL strategy

Cumulative Index to Nursing and Allied Health Literature (CINAHL Complete)

via EBSCO Host http://web.b.ebscohost.com/

Date range: Inception – 6th January 2023

Date searched: 6th January 2023

Records retrieved: 1406

 

S47 S21 AND S46   (1,406)

S46 S37 OR S45   (1,639,735)

S45 S38 OR S39 OR S40 OR S41 OR S42 OR S43 OR S44   (1,294,096)

S44 TI before N3 after OR AB before N3 after   (95,991)

S43 (MH "Controlled Before‐After Studies")   (232)

S42 (multicentre* or multi‐centre* or multicenter* or multi‐center*) OR AB (multicentre* or multi‐centre* or multicenter* or multi‐center*)   (393,726)

S41 (MH "Multicenter Studies")   (356,008)

S40 TI assign* OR AB assign*   (93,757)

S39 TI (group or groups) OR AB (group or groups)  (936,230)

S38 (MH "Control Group")   (13,464)

S37 S22 OR S23 OR S24 OR S25 OR S26 OR S27 OR S28 OR S29 OR S30 OR  S31 OR S32 OR S33 OR S34 OR S35 OR S36   (999,131)

S36 AB (cluster W3 RCT)   (482)

S35 (crossover design) OR MH (comparative studies)  (475,262)

S34 AB (control W5 group)   (141,387)

S33 PT (randomized controlled trial)  (149,463)

S32 MH (placebos)  (13,922)

S31 MH (sample size) AND AB (assigned OR allocated OR control)   (4,430)

S30 TI trial  (174,514)

S29 AB random*   (391,367)

S28 TI randomised OR randomized   (315,756)

S27 MH "Cluster Sample"   (5,174)

S26 MH "Pretest‐Posttest Design"  (51,569)

S25 MH "Random Assignment"   (77,379)

S24 MH "Single‐Blind Studies"   (15,866)

S23 MH "Double‐Blind Studies"   (54,052)

S22 MH "Randomized Controlled Trials"  (136,225)

S21 S13 AND S20  (3,118)

S20 S14 OR S15 OR S16 OR S17 OR S18 OR S19   (19,499)

S19 TI ((PAP or PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong#)) OR AB ((PAP or PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong#)) (2,115)

S18 TI ((((continuous* or positive or airway* or air‐way*) N3 pressure*) or positive‐pressure*) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong#)) OR AB ((((continuous* or positive or airway* or air‐way*) N3 pressure*) or positive‐pressure*) and (source* or device* or interface* or driver* or operator* or generator* or machine* or mask* or face‐mask* or headgear* or head gear or head‐gear or headbox or head box or head‐box or helmet* or bag* or BVM or AMBU or "Infant Flow" or mouthpiece* or mouth piece* or mouth‐piece* or nebuli?er* or prong#))  (2,767)

S17 TI ((PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (airway* or air‐way* or breath# or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)) OR AB ((PPV or CPAP or C‐PAP or NCPAP or BiPAP or APRV or IPPB or IPPV) and (airway* or air‐way* or breath# or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*))   (3,825)

S16 TI (((airway* or air‐way*) N3 pressure*) and (breath# or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)) OR AB (((airway* or air‐way*) N3 pressure*) and (breath# or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*))   (4,614)

S15 TI ((((continuous* or positive) N3 pressure*) or (positive‐pressure* or PAP)) and (airway* or air‐way* or breath# OR or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*)) OR AB ((((continuous* or positive) N3 pressure*) or (positive‐pressure* or PAP)) and (airway* or air‐way* or breath# OR or breathing or ventilat* or respir* or inspir* or inhal* or expir* or exhal*))   (10,021)

S14 MH "Positive Pressure Ventilation+"  (12,707)

S13 S1 OR S2 OR S3 OR S4 OR S5 OR S6 OR S7 OR S8 OR S9 OR S10 OR  S11 OR S12   (304,732)

S12 TI (baby or babies) OR AB (baby or babies)   (38,651)

S11 TI infan* OR AB infan*   (133,167)

S10 TI (LBW or VLBW or ELBW) OR AB (LBW or VLBW or ELBW)   (3,979)

S9 TI low‐birthweight* OR AB low‐birthweight*   (3,186)

S8 TI (low N3 (birthweight* or birth weight* or birth‐weight*)) OR AB (low N3 (birthweight* or birth weight* or birth‐weight*))   (14,338)

S7 TI (prematur* N3 (birth* or born or deliver*)) OR AB (prematur* N3 (birth* or born or deliver*))   (5,547)

S6 TI (preemie* or premie or premies) OR AB (preemie* or premie or premies)   (363)

S5 TI (preterm or preterms or pre term or pre terms or pre‐term or pre‐terms) OR AB (preterm or preterms or pre term or pre terms or pre‐term or pre‐terms)   (41,148)

S4 TI (newborn* or new born* or new‐born* or newly born* or newly‐born*) OR AB (newborn* or new born* or new‐born* or newly born* or newly‐born*)  (38,153)

S3 TI (neonat* or neo nat* or neo‐nat*) OR AB (neonat* or neo nat* or neo‐nat*)  (81,066)

S2 MH "Childbirth, Premature"   (13,094)

S1 MH "Infant, Newborn+"   (158,726)

 

Key:

MH + = exploded indexing term (MeSH)

MH = indexing term (MeSH)

* = truncation

# = up to one additional letter

? = one replacement letter

TI = terms in the title

AB = terms in the abstract

N3 = terms near three words of each other (any order).

W5 = terms within three words of each other (specified order).

Appendix 6. Trial registry strategies

ClinicalTrials.gov

via https://clinicaltrials.gov/ 

Date searched: 6th January 2023

Records retrieved: 653

 

Other Terms: (infan* OR baby OR neonat* OR prematur* OR newborn* OR LBW OR VLBW OR ELBW) AND ((PAP OR PPV OR (positive AND pressure)) AND (airway* OR breathing OR ventilat* OR respir*)) 

 

International Clinical Trials Registry Platform (ICTRP) 

via https://trialsearch.who.int/ 

Date searched: 6th January 2023

Records retrieved: 212 records for 211 trials

 

Advanced Search:

Intervention: ((PAP OR PPV OR (positive AND pressure)) AND (airway OR breathing OR ventilation OR respiration))

Recruitment Status: All

Search for Clinical Trials in Children

Appendix 7. Risk of bias tool

We used the standard methods of Cochrane and Cochrane Neonatal to assess the methodological quality (to meet the validity criteria) of the trials. For each trial, we sought information regarding the method of randomisation and the blinding and reporting of all outcomes of all the infants enrolled in the trial. We assessed each criterion as low, high, or unclear risk. Two review authors separately assessed each study. Any disagreements were resolved by discussion. We added this information to Characteristics of included studies. We evaluated the following issues and entered the findings into the risk of bias table.

1. Sequence generation (checking for possible selection bias). Was the allocation sequence adequately generated?

For each included study, we categorised the method used to generate the allocation sequence as:

a. low risk (any truly random process, e.g. random number table; computer random number generator);

b. high risk (any non‐random process, e.g. odd or even date of birth; hospital or clinic record number); or

c. unclear risk.

2. Allocation concealment (checking for possible selection bias). Was allocation adequately concealed?

For each included study, we categorised the method used to conceal the allocation sequence as:

a. low risk (e.g. telephone or central randomisation; consecutively numbered, sealed, opaque envelopes);

b. high risk (open random allocation; unsealed or non‐opaque envelopes, alternation; date of birth); or

c. unclear risk.

3. Blinding of participants and personnel (checking for possible performance bias). Was knowledge of the allocated intervention adequately prevented during the study?

For each included study, we categorised the methods used to blind study participants and personnel from the knowledge of which intervention a participant had received. Blinding was assessed separately for different outcomes or class of outcomes. We categorised the methods as:

a. low risk, high risk, or unclear risk for participants;

b. low risk, high risk, or unclear risk for personnel.

4. Blinding of outcome assessment (checking for possible detection bias). Was knowledge of the allocated intervention adequately prevented at the time of outcome assessment?

For each included study, we categorised the methods used to blind outcome assessment. Blinding was assessed separately for different outcomes or class of outcomes. We categorised the methods as:

a. low risk for outcome assessors;

b. high risk for outcome assessors; or

c. unclear risk for outcome assessors.

5. Incomplete outcome data (checking for possible attrition bias through withdrawals, dropouts, protocol deviations). Were incomplete outcome data adequately addressed?

For each included study and for each outcome, we described the completeness of data including attrition and exclusions from the analysis. We noted whether attrition and exclusions had been reported, the numbers included in the analysis at each stage (compared with the total randomised participants), reasons for attrition or exclusion where reported, and whether missing data were balanced across groups or were related to outcomes. Where sufficient information was reported or supplied by the trial authors, we re‐included missing data in the analyses. We categorised the methods as:

a. low risk (< 20% missing data);

b. high risk (≥ 20% missing data); or

c. unclear risk.

6. Selective reporting bias. Are reports of the study free of the suggestion of selective outcome reporting?

For each included study, we described how we investigated the possibility of selective outcome reporting bias and what we found. We assessed the methods as:

a. low risk (where it is clear that all of the study's prespecified outcomes and all expected outcomes of interest to the review have been reported);

b. high risk (where not all the study's prespecified outcomes have been reported; one or more reported primary outcomes were not prespecified outcomes of interest and are reported incompletely and so cannot be used; the study fails to include results of a key outcome that would be expected to have been reported); or

c. unclear risk.

7. Other sources of bias. Did the study appear to be free of other problems that could put it at high risk of bias?

For each included study, we described any important concerns we had about other possible sources of bias (e.g. whether there was a potential source of bias related to the specific study design, or whether the trial was stopped early due to some data‐dependent process). We assessed whether each study was free of other problems that could put it at risk of bias as:

a. low risk;

b. high risk; or

c. unclear risk

We explored the impact of the level of bias through undertaking sensitivity analyses as needed.

Data and analyses

Comparison 1. Bubble CPAP versus ventilator or Infant Flow Driver CPAP.

Outcome or subgroup title No. of studies No. of participants Statistical method Effect size
1.1 Treatment failure 13 1230 Risk Ratio (M‐H, Fixed, 95% CI) 0.76 [0.60, 0.95]
1.1.1 Bubble versus ventilator CPAP 9 729 Risk Ratio (M‐H, Fixed, 95% CI) 0.64 [0.47, 0.86]
1.1.2 Bubble versus Infant Flow Driver CPAP 4 501 Risk Ratio (M‐H, Fixed, 95% CI) 1.00 [0.68, 1.46]
1.2 Mortality before discharge 10 1189 Risk Ratio (M‐H, Fixed, 95% CI) 0.93 [0.64, 1.36]
1.2.1 Bubble versus ventilator CPAP 6 688 Risk Ratio (M‐H, Fixed, 95% CI) 0.81 [0.52, 1.27]
1.2.2 Bubble versus Infant Flow Driver CPAP 4 501 Risk Ratio (M‐H, Fixed, 95% CI) 1.32 [0.63, 2.77]
1.3 Pneumothorax 14 1340 Risk Ratio (M‐H, Fixed, 95% CI) 0.73 [0.40, 1.34]
1.3.1 Bubble versus ventilator CPAP 10 839 Risk Ratio (M‐H, Fixed, 95% CI) 0.58 [0.27, 1.24]
1.3.2 Bubble versus Infant Flow Driver CPAP 4 501 Risk Ratio (M‐H, Fixed, 95% CI) 1.12 [0.40, 3.16]
1.4 Moderate‐severe nasal injury 8 753 Risk Ratio (M‐H, Fixed, 95% CI) 2.29 [1.37, 3.82]
1.4.1 Bubble versus ventilator CPAP 6 592 Risk Ratio (M‐H, Fixed, 95% CI) 2.60 [1.49, 4.53]
1.4.2 Bubble versus Infant Flow Driver CPAP 2 161 Risk Ratio (M‐H, Fixed, 95% CI) 0.92 [0.21, 3.99]
1.5 Bronchopulmonary dysplasia 7 603 Risk Ratio (M‐H, Fixed, 95% CI) 0.76 [0.53, 1.10]
1.5.1 Bubble versus ventilator CPAP 5 427 Risk Ratio (M‐H, Fixed, 95% CI) 0.84 [0.52, 1.35]
1.5.2 Bubble versus Infant Flow Driver CPAP 2 176 Risk Ratio (M‐H, Fixed, 95% CI) 0.65 [0.36, 1.17]
1.6 Duration of CPAP use 8 744 Mean Difference (IV, Fixed, 95% CI) ‐0.01 [‐0.16, 0.13]
1.6.1 Bubble versus ventilator CPAP 6 508 Mean Difference (IV, Fixed, 95% CI) 0.04 [‐0.12, 0.19]
1.6.2 Bubble versus Infant Flow Driver CPAP 2 236 Mean Difference (IV, Fixed, 95% CI) ‐0.56 [‐1.07, ‐0.05]
1.7 Duration of oxygen supplementation 1 88 Mean Difference (IV, Fixed, 95% CI) 0.60 [‐2.52, 3.72]
1.7.1 Bubble versus ventilator CPAP 1 88 Mean Difference (IV, Fixed, 95% CI) 0.60 [‐2.52, 3.72]
1.8 Duration of hospitalisation 5 591 Mean Difference (IV, Fixed, 95% CI) ‐3.27 [‐4.99, ‐1.56]
1.8.1 Bubble versus ventilator CPAP 4 391 Mean Difference (IV, Fixed, 95% CI) ‐2.61 [‐4.52, ‐0.70]
1.8.2 Bubble versus Infant Flow Driver CPAP 1 200 Mean Difference (IV, Fixed, 95% CI) ‐6.00 [‐9.88, ‐2.12]
1.9 Patent ductus arteriosus 6 597 Risk Ratio (M‐H, Fixed, 95% CI) 0.96 [0.67, 1.38]
1.9.1 Bubble versus ventilator CPAP 4 257 Risk Ratio (M‐H, Fixed, 95% CI) 1.20 [0.61, 2.35]
1.9.2 Bubble versus Infant Flow Driver CPAP 2 340 Risk Ratio (M‐H, Fixed, 95% CI) 0.87 [0.57, 1.35]
1.10 Necrotising enterocolitis 6 693 Risk Ratio (M‐H, Fixed, 95% CI) 0.75 [0.39, 1.44]
1.10.1 Bubble versus ventilator CPAP 4 428 Risk Ratio (M‐H, Fixed, 95% CI) 0.62 [0.28, 1.37]
1.10.2 Bubble versus Infant Flow Driver CPAP 2 265 Risk Ratio (M‐H, Fixed, 95% CI) 1.17 [0.35, 3.92]
1.11 Severe intraventricular haemorrhage 6 562 Risk Ratio (M‐H, Fixed, 95% CI) 0.81 [0.42, 1.58]
1.11.1 Bubble versus ventilator CPAP 3 321 Risk Ratio (M‐H, Fixed, 95% CI) 0.41 [0.12, 1.39]
1.11.2 Bubble versus Infant Flow Driver CPAP 3 241 Risk Ratio (M‐H, Fixed, 95% CI) 1.19 [0.53, 2.69]
1.12 Severe retinopathy of prematurity 6 642 Risk Ratio (M‐H, Fixed, 95% CI) 0.92 [0.54, 1.57]
1.12.1 Bubble versus ventilator CPAP 4 377 Risk Ratio (M‐H, Fixed, 95% CI) 0.96 [0.53, 1.72]
1.12.2 Bubble versus Infant Flow Driver CPAP 2 265 Risk Ratio (M‐H, Fixed, 95% CI) 0.78 [0.22, 2.77]

Characteristics of studies

Characteristics of included studies [ordered by study ID]

Agarwal 2016.

Study characteristics
Methods RCT
Participants 68 newborn VLBW infants with moderate respiratory distress (Silverman score 4 to 7) treated with CPAP 5 to 7 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 34
Ventilator (Newport): N = 34
Various nasal CPAP prongs (Argyle, Hudson Binasal or Fisher & Paykel) were used.
Outcomes CPAP failure (persistent hypoxia with FiO₂ > 0.6 or requiring mechanical ventilation)
Death before discharge
Receipt of surfactant
Duration of CPAP use
Pneumothorax
Nasal trauma (septal necrosis)
Notes Setting: Delhi, India (2009 to 2011)
Funding: no specific funding
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Low risk Computer generated
Allocation concealment (selection bias) Low risk Sequentially numbered, opaque, sealed, and stapled envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Bahman‐Bijari 2011.

Study characteristics
Methods RCT
Participants 50 preterm neonates (birthweight 1000 to 2000 g) with respiratory distress (Silverman score 6 to 7) treated with CPAP 5 to 7 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 25
Ventilator (Bear 750): N = 25
In both groups, CPAP was nasopharyngeal.
Outcomes CPAP failure ("severe" intercostal retractions or prolonged apnoea (> 20 seconds) or persistent hypoxia, hypercarbia or acidosis with FiO₂ > 0.6 requiring mechanical ventilation)
Death before discharge
Duration of CPAP
Duration of hospitalisation
Pneumothorax (no events)
Bronchopulmonary dysplasia
Patent ductus arteriosus (confirmed by echocardiography)
Intraventricular haemorrhage (grade III to IV)
Nasal injury ("trauma to nasal septum and nostrils")
Notes Setting: Kerman, Iran (2009 to 2010)
Funding: Kerman University of Medical Sciences, Iran
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk Not stated (likely computer generated, as minimisation for gender and birthweight applied)
Allocation concealment (selection bias) Unclear risk Not stated; "randomly assigned"
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Bhatti 2015.

Study characteristics
Methods RCT
Participants 170 preterm neonates (< 34 weeks') with respiratory distress within 6 hours after birth treated with CPAP 5 to 7 cm H₂O (as primary support)
Interventions Bubble ("stand‐alone" system with Fisher & Paykel interface): N = 90
Ventilator (Jet‐CPAP, Phoenix Medical Systems): N = 80
Short binasal prongs were the nasal interface. 
Outcomes CPAP failure within 72 hours (persistent hypoxia with FiO₂ > 0.6 requiring endotracheal intubation and mechanical ventilation or nasal intermittent positive pressure ventilation)
Receipt of surfactant
Death before discharge
Duration of CPAP use
Pneumothorax
Bronchopulmonary dysplasia
Retinopathy of prematurity
Intraventricular haemorrhage (grade II to IV)
Necrotising enterocolitis
Nasal trauma (moderate‐severe as per "Nasal Injury Score")
Notes Setting: Chandigarh, India (2011 to 2012)
Funding: not stated
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Low risk Computer generated ("web‐based")
Allocation concealment (selection bias) Low risk Serially numbered, sealed, and opaque envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Gupta 2009.

Study characteristics
Methods RCT
Participants 140 preterm neonates (< 30 weeks' or birthweight 600 to 1500 g) ventilated at birth for respiratory distress, then treated with CPAP 6 cm H₂O (as postextubation support)
Interventions Bubble (Fisher & Paykel): N = 71
Infant Flow Driver: N = 69
Nasal CPAP was delivered through short binasal prongs provided by the device manufacturers.
Outcomes CPAP failure within 72 hours (endotracheal intubation and mechanical ventilation or nasal intermittent positive pressure ventilation)
Death before discharge
Duration of CPAP use
Pneumothorax
Bronchopulmonary dysplasia
Patent ductus arteriosus (requiring treatment)
Retinopathy of prematurity (requiring treatment)
Intraventricular haemorrhage (grade III to IV)
Necrotising enterocolitis
Notes Setting: Middlesbrough, UK (2004 to 2007)
Funding: not stated
Trial registration number: ISRCTN 83339638
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk Not stated
Allocation concealment (selection bias) Low risk Serially numbered, sealed, opaque envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Hosseini 2012.

Study characteristics
Methods RCT
Participants 161 newborn infants (28 to 37 weeks' gestation) with respiratory distress (Silverman Score 5 to 6) up to 12 hours after birth treated with CPAP 5 to 6 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 71
Ventilator (Medijet): N = 90
Short binasal prongs were used in both groups.
Outcomes CPAP failure (persistent hypoxia, hypercarbia or acidosis with FiO₂ > 0.5, or recurrent apnoea with bradycardia, and need for CPAP > 8 cm H₂O)
Death before discharge
Duration of CPAP use
Duration of hospitalisation
Pneumothorax
Nasal trauma (columella necrosis)
Intraventricular haemorrhage (grade III to IV)
Feed intolerance
Necrotising enterocolitis
Notes Setting: Tabriz, Iran (2010 to 2011)
Funding: Medin Medical Innovations GMBH (Germany) and Pars Eltyam Kala (Iran)
Trial registration number: IRCT138903174113N1
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk Not stated
Allocation concealment (selection bias) Unclear risk Not stated
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Unclear risk Bubble CPAP group included fewer very low‐ and extremely low‐birthweight or small‐for‐gestation infants than ventilator CPAP group.

Jazexhiu‐Postoli 2015.

Study characteristics
Methods RCT
Participants 200 newborn infants (28 to 35 weeks' gestation) with respiratory distress up to 6 hours after birth treated with CPAP 6 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 100
Infant Flow Driver (biphasic): N = 100
Short binasal prongs were used in both groups.
Outcomes CPAP failure (need for mechanical ventilation)
Death before discharge
Duration of CPAP use
Pneumothorax
Duration of hospitalisation (intensive care)
Intraventricular haemorrhage (grade I to IV)
Patent ductus arteriosus
"Nasal lesions"
Notes Setting: Tirana, Albania (2011 to 2013)
Funding: not stated
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk Not stated (corresponding author contacted September 2021)
Allocation concealment (selection bias) Unclear risk Not stated (corresponding author contacted September 2021)
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Unclear risk Baseline characteristics data not reported.

Mazmanyan 2016.

Study characteristics
Methods RCT
Participants 125 preterm neonates (< 37 weeks') with respiratory distress treated with nasal CPAP 4 to 6 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 66
Infant Flow Driver: N = 59
Short binasal prongs were used for Infant Flow Driver CPAP, and pliable long anatomically curved prongs for bubble CPAP.
Outcomes CPAP failure (receipt of mechanical ventilation)
Death before discharge
Duration of CPAP use
Duration of supplemental oxygen
Pneumothorax
Nasal injury (excoriation)
Retinopathy of prematurity
Periventricular haemorrhage/leucomalacia
Necrotising enterocolitis
Notes Setting: Yerevan, Armenia (2012 to 2014)
Funding: not stated
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk Not stated
Allocation concealment (selection bias) Low risk Serially numbered, sealed, and opaque envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Mazzella 2001.

Study characteristics
Methods RCT
Participants 36 preterm neonates (< 36 weeks') with respiratory distress up to 12 hours after birth treated with nasal CPAP 4 to 6 cm H₂O (as primary support)
Interventions Bubble (single nasopharyngeal tube, pressure generated by underwater seal): N = 18
Infant Flow Driver: N = 18
Short binasal prongs were used for Infant Flow Driver CPAP, and a conventional single prong for bubble CPAP.
Outcomes CPAP failure (need for mechanical ventilation)
Death before discharge
Duration of CPAP use
Pneumothorax
Bronchopulmonary dysplasia
Intraventricular haemorrhage
Nasal trauma (bleeding)
Notes Setting: Genova, Italy (1997 to 1999)
Funding: not stated
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk Not stated
Allocation concealment (selection bias) Low risk Sealed, opaque, numbered envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Unclear risk Infants in bubble CPAP group were of higher average gestational age at birth than those in Infant Flow Driver CPAP group (33 versus 32 weeks').

Mohammadizadeh 2011.

Study characteristics
Methods RCT
Participants 88 preterm neonates (29 to 33 weeks') with respiratory distress after birth treated with nasal CPAP up to 8 cm H₂O (as primary support)
Interventions Bubble (manufacturer not stated): N = 44
Ventilator (Medijet): N = 44
Nasal interface not stated.
Outcomes CPAP failure ("clinical or laboratory evidence of respiratory failure")*
Surfactant administration via endotracheal tube (if FiO₂ > 0.4 with CPAP > 7 cm)
Duration of CPAP use
Duration of oxygen therapy
Pneumothorax*
Bronchopulmonary dysplasia*
Notes Setting: Isfahan, Iran (pre‐2011)
Funding: not stated
*Data not reported; author contacted September 2021.
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk "randomly assigned"
Allocation concealment (selection bias) Unclear risk "randomly assigned"
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Unclear risk Unable to determine
Selective reporting (reporting bias) Unclear risk Unable to determine
Other bias Unclear risk Unable to determine

Noori Shadkam 2017.

Study characteristics
Methods RCT
Participants 119 preterm neonates (28 to 34 weeks' and birthweight of 1000 to 2200 g) with respiratory distress (Silverman score 5 to 7) treated with nasal CPAP 5 to 7 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 58
Ventilator (Stephan): N = 61
Nasal interface not stated.
Outcomes Death before discharge
Duration of CPAP use
Duration of supplemental oxygen therapy
Duration of hospitalisation
Pneumothorax
Patent ductus arteriosus (confirmed by echocardiography)
Retinopathy of prematurity
Intraventricular haemorrhage (grade III to IV)
Bronchopulmonary dysplasia
Notes Setting: Yazd, Iran (2013 to 2015)
Funding: not stated
Trial registration: IRCT2014081918862N1
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Low risk Random number tables
Allocation concealment (selection bias) Unclear risk Not stated (author contacted September 2021)
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Ribeiro 2017.

Study characteristics
Methods RCT
Participants 65 preterm neonates (gestational age < 34 weeks and birthweight 500 to 1500 g) with respiratory distress treated surfactant replacement, then with CPAP 4 to 5 cm H₂O (as postextubation support)
Interventions Bubble (manufacturer not stated): N = 32
Ventilator (InterMed): N = 33
Short binasal prongs were used in both groups.
Outcomes CPAP failure within 48 hours of initiation (endotracheal re‐intubation and mechanical ventilation)
Death before discharge
Pneumothorax
Patent ductus arteriosus (confirmed by echocardiography)
Bronchopulmonary dysplasia
Duration of CPAP use
Duration of oxygen supplementation
Duration of hospitalisation
Retinopathy of prematurity
Intraventricular haemorrhage (grade III to IV)
Necrotising enterocolitis
Nasal injury (mild‐moderate)*
Notes Setting: São Paulo, Brazil (2012 to 2013)
Funding: none
NB. Erratum published.
*Author contacted September 2021.
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Low risk Computer generated
Allocation concealment (selection bias) Unclear risk Not described
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Tagare 2010.

Study characteristics
Methods RCT
Participants 30 preterm neonates (< 37 weeks') with respiratory distress (Silverman score 5 to 7) up to 6 hours after birth treated with nasal CPAP 5 to 7 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 15
Ventilator (Bear Cub 750): N = 15
Short binasal prongs were used in both groups.
Outcomes Treatment failure (worsening respiratory distress with FiO₂ > 0.6 and CPAP > 8 cm H₂O)
Duration of CPAP use
Pneumothorax
Nasal trauma (skin erosion/septal injury)
(Mortality not reported.)
Notes Setting: Pune, India (2007 to 2008)
Funding: not stated
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Low risk Computer generated
Allocation concealment (selection bias) Low risk Serially numbered, sealed, opaque envelopes
Blinding (performance bias and detection bias)
All outcomes Unclear risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Unclear risk Few outcomes reported (pilot study).
Other bias Low risk No evidence of an imbalance in baseline demographics

Tagare 2013.

Study characteristics
Methods RCT
Participants 114 preterm neonates (< 37 weeks') with respiratory distress (Silverman score > 3) up to 6 hours after birth treated with nasal CPAP 6 to 7 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 57
Ventilator (Bear Cub 750): N = 57
Short binasal prongs were used in both groups.
Outcomes Treatment failure (worsening respiratory distress with FiO₂ > 0.6 and CPAP > 8 cm H₂O, recurrent apnoea, "shock", requiring endotracheal intubation and mechanical ventilation)
Mortality before hospital discharge
Duration of CPAP use
Pneumothorax
Nasal trauma (skin erosion/septal injury)
Notes Setting: Pune, India (2008 to 2009)
Funding: not stated
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Low risk Computer generated
Allocation concealment (selection bias) Low risk Sealed, coded envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Yadav 2012.

Study characteristics
Methods RCT
Participants 32 preterm neonates (< 33 weeks' and birthweight < 1500 g) ventilated within 1 week after birth for respiratory distress, then treated with CPAP 4 to 6 cm H₂O (as postextubation support)
Interventions Bubble (Fisher & Paykel): N = 16
Ventilator (SLE 2000 or Dräger Babylog 8000): N = 16
Short binasal prongs were used in both groups.
Outcomes CPAP failure within 72 hours of extubation (persistent hypoxia with FiO₂ > 0.6, recurrent apnoea and bradycardia, or systemic hypoperfusion requiring vasopressor therapy, requiring endotracheal intubation and mechanical ventilation)
Pneumothorax
Bronchopulmonary dysplasia
Duration of hospitalisation
Patent ductus arteriosus (confirmed by echocardiography)
Retinopathy of prematurity
Intraventricular haemorrhage (grade II to IV)
Necrotising enterocolitis
Culture‐positive sepsis
(Mortality not reported.)
Notes Setting: New Dehli, India (2007 to 2009)
Funding: Indian Council of Medical Research (equipment was provided by Fisher & Paykel)
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Low risk Computer generated
Allocation concealment (selection bias) Low risk Serially numbered, sealed, opaque envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive)
Other bias Low risk No evidence of an imbalance in baseline demographics

Yagui 2011.

Study characteristics
Methods RCT
Participants 39 newborn infants* (birthweight > 1500 g) with respiratory distress within 24 hours after birth treated with CPAP 4 to 6 cm H₂O (as primary support)
Interventions Bubble (Fisher & Paykel): N = 20
Ventilator (Siemens Servo – I): N = 19  
Short binasal prongs were used in both groups.
Outcomes CPAP failure (persistent hypoxia despite FiO₂ > 0.4 with CPAP > 8 cm H₂O requiring endotracheal intubation and mechanical ventilation)
Pneumothorax
Duration of CPAP use
Duration of oxygen supplementation
Duration of hospitalisation
(Mortality not reported.)
Notes Setting: São Paulo, Brazil (2008 to 2010)
Funding: not stated
*Mean gestation at birth 35 weeks'.
Risk of bias
Bias Authors' judgement Support for judgement
Random sequence generation (selection bias) Unclear risk Not described
Allocation concealment (selection bias) Low risk Serially numbered, sealed, coded envelopes
Blinding (performance bias and detection bias)
All outcomes High risk Open‐label
Incomplete outcome data (attrition bias)
All outcomes Low risk Outcome reporting complete.
Selective reporting (reporting bias) Low risk Unlikely (comprehensive; mortality not reported)
Other bias Low risk No evidence of an imbalance in baseline demographics

CPAP: continuous positive airway pressure
FiO₂: fraction of inspired oxygen
RCT: randomised controlled trial
VLBW: very low birthweight

Characteristics of excluded studies [ordered by study ID]

Study Reason for exclusion
Ahluwalia 1998 Cross‐over study
Backes 2020 Randomised trial of 2 types of bubble CPAP (Seattle‐PAP vs Fisher & Paykel CPAP)
Bhandari 1996 This comparison of nasal vs nasopharyngeal CPAP was non‐randomised and retrospective.
Buettiker 2004 Assessed effect of nasopharyngeal CPAP (conventional ventilator) versus CPAP via Hudson (conventional ventilator) vs CPAP via Infant Flow Driver
Campbell 2006 Compared Infant Flow Driver CPAP with high‐flow nasal cannulae
Courtney 2001 Cross‐over study
Davis 2001 Trial compared effects of CPAP via Hudson prongs (conventional ventilator) vs CPAP via single prong (conventional ventilator).
Diala 2022 Non‐randomised study
Guerin 2016 Cross‐over study
Jonsson 1998 Cross‐over study
Kavvadia 2000 Non‐randomised study
Liptsen 2005 Cross‐over study of bubble CPAP vs ventilator CPAP in preterm infants
Massaro 2005 Non‐randomised study
Nair 2005 Assessed effect of high‐flow nasal cannula system vs bubble CPAP
Narendran 2002 Non‐randomised study (bubble CPAP vs ventilator CPAP)
Pandit 2001 Cross‐over study
Pantalischka 2009 Cross‐over study
Pelligra 2006 Non‐randomised study (historical controls to compare bubble CPAP with ventilator CPAP)
Rego 2002 Trial comparing CPAP via Hudson vs Argyle prongs (both by conventional ventilator)
Roukema 1999a Trial of Infant Flow Driver CPAP vs conventional ventilator nasopharyngeal CPAP
Roukema 1999b Non‐randomised evaluation of Infant Flow Driver CPAP vs conventional ventilator nasopharyngeal CPAP (postextubation)
Stefanescu 2003 Trial comparing Infant Flow CPAP vs ventilator CPAP (INCA binasal prongs)
Sun 1999 Randomised trial of Infant Flow Driver CPAP vs conventional ventilator CPAP
Tayler 2022 Non‐randomised study
Telenko 1999 Cross‐over study
Trevisanuto 2005 Compared Infant Flow nasal CPAP with CPAP delivered via a polycarbonate helmet

CPAP: continuous positive airway pressure

Characteristics of studies awaiting classification [ordered by study ID]

Colaizy 2004.

Methods Randomised controlled trial
Participants Unknown
Interventions "Bubble vs. conventional CPAP"
Outcomes Unknown
Notes Author contacted September 2021.

CPAP: continuous positive airway pressure

Differences between protocol and review

This review focuses on a subtopic of an earlier protocol and review (De Paoli 2002De Paoli 2008). We made the following changes to the review at the request of the World Health Organization (WHO).

  • Updating of Background section.

  • Modification of both primary and secondary outcome measures in consultation with authorship team and WHO.

  • Modification of selected subgroup analyses in consultation with authorship team and WHO.

  • Addition of risk of bias assessment.

  • Addition of GRADE assessment and summary of findings tables.

An Information Specialist wrote new search strategies which were run without date limits. 

We prespecified that infants should receive 'nasal CPAP'. We included one trial in which both intervention and control arms received "nasopharyngeal CPAP" (Bahman‐Bijari 2011). 

Contributions of authors

All authors contributed to the development and completion of this review.

Sources of support

Internal sources

  • Royal Women's Hospital Foundation, Melbourne, Australia

    Host institution

External sources

  • Vermont Oxford Network, USA

    Cochrane Neonatal Reviews are produced with support from Vermont Oxford Network, a worldwide collaboration of health professionals dedicated to providing evidence‐based care of the highest quality for newborn infants and their families.

  • World Health Organization, Switzerland

    Editorial and administrative support for this review has been provided by a grant from World Health Organization to Cochrane Neonatal.

  • National Institute for Health Research, UK

    National Institute for Health Research (NIHR) Programme Grant (NIHR133131). The views expressed in this publication are those of the authors and not necessarily those of the National Health Service, the NIHR, or the UK Department of Health.

Declarations of interest

RP has no interests to declare.

ADP has no interests to declare.

PD has no interests to declare.

SO has no interests to declare.

WM is Co‐coordinating Editor of Cochrane Neonatal but was not involved in the editorial acceptance or processes for this review.

New

References

References to studies included in this review

Agarwal 2016 {published data only}

  1. Agarwal S, Maria A, Roy MK, Verma A. A randomized trial comparing efficacy of bubble and ventilator derived nasal CPAP in very low birth weight neonates with respiratory distress. Journal of Clinical and Diagnostic Research 2016;10(9):SC09. [DOI: 10.7860/JCDR/2016/20584.8572] [DOI] [PMC free article] [PubMed] [Google Scholar]

Bahman‐Bijari 2011 {published data only}

  1. Bahman-Bijari B, Malekiyan A, Niknafs P, Baneshi MR. Bubble–CPAP vs. Ventilatory–CPAP in preterm infants with respiratory distress. Iranian Journal of Pediatrics 2011;21(2):151-8. [PMID: ] [PMC free article] [PubMed] [Google Scholar]

Bhatti 2015 {published data only}

  1. Bhatti A, Khan J, Murki S, Sundaram V, Saini SS, Kumar P. Nasal Jet-CPAP (variable flow) versus Bubble-CPAP in preterm infants with respiratory distress: an open label, randomized controlled trial. Journal of Perinatology 2015;35(11):935-40. [DOI: 10.1038/jp.2015.98] [DOI] [PubMed] [Google Scholar]
  2. Khan J, Sundaram V, Murki S, Bhatti A, Saini SS, Kumar P. Nasal injury and comfort with jet versus bubble continuous positive airway pressure delivery systems in preterm infants with respiratory distress. European Journal of Pediatrics 2017;176(12):1629-35. [DOI: 10.1007/s00431-017-3016-7] [DOI] [PubMed] [Google Scholar]

Gupta 2009 {published data only}

  1. Gupta S, Sinha SK, Tin W, Donn SM. A randomized controlled trial of post-extubation bubble continuous positive airway pressure versus Infant Flow Driver continuous positive airway pressure in preterm infants with respiratory distress syndrome. Journal of Pediatrics 2009;154(5):645-50. [DOI: 10.1016/j.jpeds.2008.12.034] [DOI] [PubMed] [Google Scholar]

Hosseini 2012 {published data only}

  1. Hosseini MB, Heidarzadeh M, Balila M, Ghojazadeh M, Janani R, Safavi-nia S, et al. Randomized controlled trial of two methods of nasal continuous positive airway pressure (N-CPAP) in preterm infants with respiratory distress syndrome: underwater bubbly CPAP vs. Medijet system device. The Turkish Journal of Pediatrics 2012;54(6):632-40. [PMID: ] [PubMed] [Google Scholar]

Jazexhiu‐Postoli 2015 {published data only}

  1. Jazexhiu-Postoli E, Vevecka E, Petrela E, Tushe E. Results of the treatment with two different nasal continuous positive airway pressure systems in preterm newborns with respiratory distress syndrome. The Albanian Journal of Medical and Health Sciences 2015;46(2):29-35. [Google Scholar]
  2. Jazexhiu-Postoli E, Vevecka E, Rapushi E, Petrela E, Tushe E. Can we predict CPAP failure in preterms with respiratory distress syndrome? In: The Albanian Journal of Natural Technical Sciences. Vol. 19. 2014.

Mazmanyan 2016 {published data only}

  1. Mazmanyan P, Mellor K, Doré CJ, Modi N. A randomised controlled trial of flow driver and bubble continuous positive airway pressure in preterm infants in a resource-limited setting. Archives of Disease in Childhood-Fetal and Neonatal Edition 2016;101(1):16-20. [DOI] [PubMed] [Google Scholar]

Mazzella 2001 {published data only}

  1. Mazzella M, Bellini C, Calevo MG, Campone F, Massocco D, Mezzano P, et al. A randomised control study comparing the Infant Flow Driver with nasal continuous positive airway pressure in preterm infants. Archives of Disease in Childhood Fetal- and Neonatal Edition 2001;85:F86-90. [DOI: 10.1136/fn.85.2.f86] [DOI] [PMC free article] [PubMed] [Google Scholar]

Mohammadizadeh 2011 {published data only}

  1. Mohammadizadeh M, Asadi AR, Sadeghnia AR. Compare the effects of continuous positive airway pressure with two different methods to treat premature infants with respiratory distress syndrome. Journal of Isfahan Medical School. 2011 Sep 12;29(146). 2011;29(146):901-11. [Google Scholar]

Noori Shadkam 2017 {published data only}

  1. Noori Shadkam M, Movahedinia M, Noori Shadkam Z, Mehrparvar AH. Comparison of the therapeutic effects of bubble CPAP and ventilator CPAP on respiratory distress syndrome in premature neonates. Iranian Journal of Neonatology 2017;8(3):1-5. [Google Scholar]

Ribeiro 2017 {published and unpublished data}

  1. Ribeiro SN, Fontes MJ, Bhandari V, Resende CB, Johnston C. Noninvasive ventilation in newborns≤ 1,500 g after tracheal extubation: randomized clinical trial. American Journal of Perinatology 2017;34(12):1190-8. [DOI: 10.1055/s-0037-1602141] [DOI] [PubMed] [Google Scholar]

Tagare 2010 {published data only}

  1. Tagare A, Kadam S, Vaidya U, Pandit A, Patole S. A pilot study of comparison of BCPAP vs. VCPAP in preterm infants with early onset respiratory distress. Journal of Tropical Pediatrics 2010;56(3):191-4. [DOI: 10.1093/tropej/fmp092] [DOI] [PubMed] [Google Scholar]

Tagare 2013 {published data only}

  1. Tagare A, Kadam S, Vaidya U, Pandit A, Patole S. Bubble CPAP versus ventilator CPAP in preterm neonates with early onset respiratory distress—a randomized controlled trial. Journal of Tropical Pediatrics 2013;59(2):113-9. [DOI: 10.1093/tropej/fmp092] [DOI] [PubMed] [Google Scholar]

Yadav 2012 {published data only}

  1. Yadav S, Thukral A, Sankar MJ, Sreenivas V, Deorari AK, Paul VK, et al. Bubble vs conventional continuous positive airway pressure for prevention of extubation failure in preterm very low birth weight infants: a pilot study. The Indian Journal of Pediatrics 2012;79(9):1163-8. [DOI: 10.1007/s12098-011-0651-2] [DOI] [PubMed] [Google Scholar]

Yagui 2011 {published data only}

  1. Yagui AC, Vale LA, Haddad LB, Prado C, Rossi FD, Deutsch AD, et al. Bubble CPAP versus CPAP with variable flow in newborns with respiratory distress: a randomized controlled trial. Jornal de pediatria 2011;87(6):499-504. [DOI: 10.2223/JPED.2145] [DOI] [PubMed] [Google Scholar]

References to studies excluded from this review

Ahluwalia 1998 {published data only}

  1. Ahluwalia JS, White DK, Morley CJ. Infant Flow Driver or single prong nasal continuous positive airway pressure: short-term physiological effects. Acta Paediatrica 1998;87:325-7. [DOI] [PubMed] [Google Scholar]

Backes 2020 {published data only}

  1. Backes CH, Cooper JN, Notestine JL, Alfred CM, Ball MK, Rivera BK, et al. A trial comparing continuous positive airway pressure (CPAP) devices in preterm infants. Journal of Perinatology 2020;40(8):1193-201. [DOI: 10.1038/s41372-020-0690-5] [DOI] [PMC free article] [PubMed] [Google Scholar]

Bhandari 1996 {published data only}

  1. Bhandari V, Rogerson S, Barfield C, Yu V, Rowe JC. Nasal versus naso-pharyngeal continuous positive airway pressure (CPAP) use in preterm neonates (abstract). Pediatric Research 1996;39 (Suppl):196A. [Google Scholar]

Buettiker 2004 {published data only}

  1. Buettiker V, Hug MI, Baenziger O, Meyer C, Frey B. Advantages and disadvantages of different nasal CPAP systems in newborns. Intensive Care Medicine 2004;30:926-30. [DOI] [PubMed] [Google Scholar]

Campbell 2006 {published data only}

  1. Campbell DM, Shah P, Shah V, Kelly E. High flow nasal cannula CPAP versus infant flow nasal CPAP in newly-extubated neonates < 1250 g (abstract). Pediatric Research 2004;56:472A. [Google Scholar]
  2. Campbell DM, Shah PS, Shah V, Kelly EN. Nasal continuous positive airway pressure from high flow cannula versus Infant Flow for preterm infants. Journal of Perinatology 2006;26(9):546-9. [DOI: 10.1038/sj.jp.7211561] [DOI] [PubMed] [Google Scholar]

Courtney 2001 {published data only}

  1. Courtney SE, Pyon KH, Saslow JG, Arnold GK, Pandit PB, Habib RH. Lung recruitment and breathing pattern during variable versus continuous flow nasal continuous positive airway pressure in premature infants: an evaluation of three devices. Pediatrics 2001;107:304-8. [DOI] [PubMed] [Google Scholar]

Davis 2001 {published and unpublished data}

  1. Davis P, Davies M, Faber B. A randomised controlled trial of two methods of delivering nasal continuous positive airway pressure after extubation to infants weighing less than 1000g: binasal (Hudson) versus single nasal prongs. Archives of Disease in Childhood Fetal Neonatal Ed 2000;85:F82-5. [DOI] [PMC free article] [PubMed] [Google Scholar]

Diala 2022 {published data only}

  1. Diala UM, Toma BO, Shwe DD, Ofakunrin AOD, Diala OO, John C. An assessment of improved outcomes using low-cost bubble CPAP in very low birthweight neonates in a Nigerian tertiary hospital. Trop Doct 2022;52(4):503-509. [DOI: 10.1177/00494755221107461] [DOI] [PubMed] [Google Scholar]

Guerin 2016 {published data only}

  1. Guerin C, Bailey SM, Mally PV, Rojas M, Bhutada A, Rastogi S. Randomized control trial comparing physiologic effects in preterm infants during treatment with nasal continuous positive airway pressure (nCPAP) generated by bubble nCPAP and ventilator nCPAP: a pilot study. Journal of Perinatal Medicine 2016;44:655-61. [DOI: 10.1515/jpm-2015-0209] [DOI] [PubMed] [Google Scholar]

Jonsson 1998 {published data only}

  1. Jonsson B, Repfennig L, Eriksson M, Lagercrantz H. The Scandanavian mode of respiratory therapy: a physiologic study of the Infant Flow nasal CPAP (abstract). Pediatric Research 1998;43:178A. [Google Scholar]

Kavvadia 2000 {published data only}

  1. Kavvadia V, Greenough A, Dimitriou G. Effect on lung function of continuous positive airway pressure administered either by Infant Flow Driver or a single nasal prong. European Journal of Pediatrics 2000;159:289-92. [DOI] [PubMed] [Google Scholar]

Liptsen 2005 {published data only}

  1. Liptsen E, Aghai ZH, Pyon KH, Saslow JG, Nakhla T, Long J, et al. Work of breathing during nasal continuous positive airway pressure in preterm infants: a comparison of bubble vs variable-flow devices. Journal of Perinatology 2005;25:453-8. [DOI] [PubMed] [Google Scholar]

Massaro 2005 {published data only}

  1. Massaro AN, Abdel-Haq I, Aly H. Underwater seal "bubble CPAP" versus ventilator derived CPAP: does mode of delivery make a difference in clinical outcome? (abstract). E-PAS [www.pas-meeting.org/] 2005;57:2054. [Google Scholar]

Nair 2005 {published data only}

  1. Nair G, Karna P. Comparison of the effects of Vapotherm and nasal CPAP in respiratory distress in preterm infants (abstract). E-PAS [www.pas-meeting.org/] 2005;57:2054. [Google Scholar]

Narendran 2002 {published data only}

  1. Narendran V, Donovan EF, Hoath SB, Warner BB, Steichen JJ, Jobe AH. Comparison between early bubble CPAP and conventional CPAP in reducing the incidence of chronic lung disease (abstract). E-PAS [www.pas-meeting.org/] 2002;51:1960. [Google Scholar]

Pandit 2001 {published data only}

  1. Pandit PB, Courtney SE, Pyon KH, Saslow JG, Habib RH. Work of breathing during constant- and variable-flow nasal continuous positive airway pressure in preterm neonates. Pediatrics 2001;108:682-5. [DOI] [PubMed] [Google Scholar]

Pantalischka 2009 {published data only}

  1. Pantalitschka T, Sievers J, Urschitz MS, Herberts T, Reher C, Poets CF. Randomised crossover trial of four nasal respiratory support systems for apnoea of prematurity in very low birthweight infants. Archives of Disease in Childhood-Fetal and Neonatal Edition 2009;94(4):F245-8. [DOI: 10.1136/adc.2008.148981] [DOI] [PubMed] [Google Scholar]

Pelligra 2006 {published data only}

  1. Pelligra G, Abdellatif MA, Lee SK. Comparison of clinical outcomes between two modes of CPAP delivery: underwater "bubble" versus conventional ventilator-derived (abstract). E-PAS [www.pas-meeting.org/] 2006;59:475. [Google Scholar]

Rego 2002 {published data only}

  1. Rego MAC, Martinez FE. Comparison of two nasal prongs for application of continuous positive airway pressure in neonates. Pediatric Critical Care Medicine 2002;3:239-43. [DOI] [PubMed] [Google Scholar]

Roukema 1999a {published data only}

  1. Roukema H, O'Brien K, Nesbitt K, Zaw W. A randomized controlled trial of Infant Flow continuous positive airway pressure (CPAP) versus nasopharyngeal CPAP in the extubation of babies ≤ 1250g (abstract). Pediatric Research 1999;45:318A. [Google Scholar]

Roukema 1999b {published data only}

  1. Roukema H, O'Brien K, Nesbitt K, Zaw W. A crossover trial of Infant Flow continuous positive airway pressure versus nasopharyngeal CPAP in the extubation of babies ≤ 1250 grams birthweight (abstract). Pediatric Research 1999;45:317A. [Google Scholar]

Stefanescu 2003 {published data only}

  1. Stefanescu BM, Murphy WP, Hansell BJ, Fuloria M, Morgan TM, Aschner JL. A randomized, controlled trial comparing two different continuous positive airway pressure systems for the successful extubation of extremely low birth weight infants. Pediatrics 2003;112:1031-8. [DOI] [PubMed] [Google Scholar]

Sun 1999 {published and unpublished data}

  1. Sun SC, Tien HC. Randomized controlled trial of two methods of nasal CPAP (NCPAP): Flow Driver vs conventional NCPAP (abstract). Pediatric Research 1999;45:322A. [Google Scholar]

Tayler 2022 {published data only}

  1. Tayler A, Ashworth H, Bou Saba G, Wadhwa H, Dundek M, Ng E, et al. Feasibility of a novel ultra-low-cost bubble CPAP (bCPAP) System for neonatal respiratory support at Muhimbili National Hospital, Tanzania. PLoS One 2022;17(12):e0269147. [DOI: 10.1371/journal.pone.0269147] [DOI] [PMC free article] [PubMed] [Google Scholar]

Telenko 1999 {published data only}

  1. Telenko T, Peliowski A, Hudson-Mason A. Continuous positive airway pressure (CPAP) in the treatment of apnea of prematurity: a comparison of two CPAP delivery systems (abstract). Pediatric Research 1999;45:228A. [Google Scholar]

Trevisanuto 2005 {published data only}

  1. Trevisanuto D, Grazzina N, Doglioni N, Ferrarese P, Marzari F, Zanardo V. A new device for administration of continuous positive airway pressure in preterm infants: comparison with a standard nasal CPAP continuous positive airway pressure system. Intensive Care Medicine 2005;31:859-64. [DOI] [PubMed] [Google Scholar]

References to studies awaiting assessment

Colaizy 2004 {published data only}

  1. Colaizy TT, McEvoy C, Crichton C, Freitag BC, Gilhooly J, Pillers DM, Smith SA, Wallen L. Bubble vs. conventional CPAP: a prospective, randomized, pilot study (abstract). Pediatric Academic Societies [www.pas-meeting.org/] 2004;55:2646. [Google Scholar]

Additional references

Abdel‐Latif 2021

  1. Abdel-Latif ME, Davis PG, Wheeler KI, De Paoli AG, Dargaville PA. Surfactant therapy via thin catheter in preterm infants with or at risk of respiratory distress syndrome. Cochrane Database of Systematic Reviews 2021, Issue 5. Art. No: CD011672. [DOI: 10.1002/14651858.CD011672.pub2] [DOI] [PMC free article] [PubMed] [Google Scholar]

Baldursdottir 2020

  1. Baldursdottir S, Falk M, Donaldsson S, Jonsson B, Drevhammar T. Basic principles of neonatal bubble CPAP: effects on CPAP delivery and imposed work of breathing when altering the original design. Arch Dis Child Fetal & Neonatal Edition 2020;105(5):550-554. [DOI: 10.1136/archdischild-2019-318073] [DOI] [PMC free article] [PubMed] [Google Scholar]

Bamat 2021

  1. Bamat N, Fierro J, Mukerji A, Wright CJ, Millar D, Kirpalani H. Nasal continuous positive airway pressure levels for the prevention of morbidity and mortality in preterm infants. Cochrane Database of Systematic Reviews 2021, Issue 12. Art. No: CD012778. [CENTRAL: CD012778] [DOI: 10.1002/14651858.CD012778.pub2] [DOI] [PMC free article] [PubMed] [Google Scholar]

Bell 1978

  1. Bell MJ, Ternberg JL, Feigin RD, Keating JP, Marshall R, Barton L, et al. Neonatal necrotizing enterocolitis. Therapeutic decisions based upon clinical staging. Annals of Surgery 1978;187(1):1-7. [DOI: 10.1097/00000658-197801000-00001] [MEDLINE: ] [DOI] [PMC free article] [PubMed] [Google Scholar]

Beltempo 2018

  1. Beltempo M, Isayama T, Vento M, Lui K, Kusuda S, Lehtonen L, et al, on behalf of the International Network for Evaluating Outcomes of Neonates. Respiratory management of extremely preterm infants: an international survey. Neonatology 2018;14(1):28-36. [DOI: 10.1159/000487987] [DOI] [PubMed] [Google Scholar]

Bharadwaj 2020

  1. Bharadwaj SK, Alonazi A, Banfield L, Dutta S, Mukerji A. Bubble versus other continuous positive airway pressure forms: a systematic review and meta-analysis. Archives of Disease in Childhood - Fetal and Neonatal Edition 2020;105:F526-531. [DOI: ] [DOI] [PubMed] [Google Scholar]

Committee on Fetus and Newborn 2014

  1. Committee on Fetus and Newborn. Respiratory support in preterm infants at birth. Pediatrics 2014;133(1):171-4. [DOI] [PubMed] [Google Scholar]

Curstedt 2015

  1. Curstedt T, Halliday HL, Hallman M, Saugstad OD, Speer CP. 30 years of surfactant research - from basic science to new clinical treatments for the preterm infant. Neonatology 2015;107(4):314-6. [DOI: 10.1159/000381160] [DOI] [PubMed] [Google Scholar]

Dargaville 2013

  1. Dargaville PA, Aiyappan A, De Paoli AG, Dalton RG, Kuschel CA, Kamlin CO, et al. Continuous positive airway pressure failure in preterm infants: incidence, predictors and consequences. Neonatology 2013;104(1):8-14. [DOI] [PubMed] [Google Scholar]

Dargaville 2016

  1. Dargaville PA, Gerber A, Johansson S, De Paoli AG, Kamlin CO, Orsini F, et al, Australian and New Zealand Neonatal Network. Incidence and outcome of CPAP failure in preterm infants. Pediatrics 2016;138(1):e20153985. [DOI: 10.1542/peds.2015-3985] [DOI] [PubMed] [Google Scholar]

Davis 2003

  1. Davis P, Henderson-Smart DJ. Nasal continuous positive airway pressure immediately after extubation for preventing morbidity in preterm infants. Cochrane Database of Systematic Reviews 2003, Issue 2. Art. No: CD000143. [DOI: 10.1002/14651858.CD000143] [DOI] [PubMed] [Google Scholar]

De Paoli 2005

  1. De Paoli AG, Lau R, Davis PG, Morley CJ. Pharyngeal pressure in preterm infants receiving nasal continuous positive airway pressure. Archives of Disease in Childhood-Fetal and Neonatal Edition 2005;90(1):F79-81. [DOI: 10.1136/adc.2004.052274] [DOI] [PMC free article] [PubMed] [Google Scholar]

De Paoli 2021

  1. De Paoli A, Prakash R, Oddie S, Davies P, McGuire W. Masks versus prongs as interfaces for nasal continuous positive airway pressure in preterm infants. Cochrane Database of Systematic Reviews 2021, Issue 12. Art. No: CD002977. [CENTRAL: CD002977] [DOI: 10.1002/14651858.CD002977.pub2] [DOI] [PMC free article] [PubMed] [Google Scholar]

Ehrenkranz 2005

  1. Ehrenkranz RA, Walsh MC, Vohr BR, Jobe AH, Wright LL, Fanaroff AA, et al, National Institutes of Child Health and Human Development Neonatal Research Network. Validation of the National Institutes of Health consensus definition of bronchopulmonary dysplasia. Pediatrics 2005;116(6):1353-60. [DOI: 10.1542/peds.2005-0249] [PMID: ] [DOI] [PubMed] [Google Scholar]

Ekhaguere 2019

  1. Ekhaguere OA, Mairami AB, Kirpalani H. Risk and benefits of bubble continuous positive airway pressure for neonatal and childhood respiratory diseases in low- and middle-income countries. Paediatric Respiratory Reviews 2019;29:31-36. [DOI: 10.1016/j.prrv.2018.04.004] [DOI] [PubMed] [Google Scholar]

Ettinger 2021

  1. Ettinger NA, Serazin N, Nguyen R, Werdenberg J, Huibers M, Torrey S. Testing positive pressure delivered from commercial and WHO-style pediatric bubble CPAP devices. BMC Pediatr 2021;21(1):534. [DOI: 10.1186/s12887-021-03006-2] [DOI] [PMC free article] [PubMed] [Google Scholar]

Fraser 2004

  1. Fraser J, Walls M, McGuire W. Respiratory complications of preterm birth 0.1136/bmj.329.7472.962. BMJ 2004;329(7472):962-965. [DOI: ] [DOI] [PMC free article] [PubMed] [Google Scholar]

Gale 2020

  1. Gale C, McGuire W, Juszczak E. Randomised controlled trials for informing perinatal care. Neonatology 2020;117(1):8-14. [DOI: 10.1159/000499881] [PMID: ] [DOI] [PubMed] [Google Scholar]

Gaon 1999

  1. Gaon P, Lee S, Hannan S, Ingram D, Milner AD. Assessment of effect of nasal continuous positive pressure on laryngeal opening using fibre optic laryngoscopy. Archives of Disease in Childhood 1999;80:F230-2. [DOI] [PMC free article] [PubMed] [Google Scholar]

Glaser 2021

  1. Glaser K, Wright CJ. Indications for and risks of noninvasive respiratory support. Neonatology 2021;118(2):235-243. [DOI: 10.1159/000515818] [DOI] [PMC free article] [PubMed] [Google Scholar]

GRADEpro GDT [Computer program]

  1. GRADEpro GDT. Version accessed 22 April 2022. Hamilton (ON): McMaster University (developed by Evidence Prime). Available at gradepro.org.

Green 2019

  1. Green EA, Dawson JA, Davis PG, De Paoli AG, Roberts CT. Assessment of resistance of nasal continuous positive airway pressure interfaces. Archives of Disease in Childhood Fetal & Neonatal Edition 2019;104(5):F535-F539. [DOI: 10.1136/archdischild-2018-315838] [DOI] [PubMed] [Google Scholar]

Gregory 1971

  1. Gregory GA, Kitterman JA, Phibbs RH, Tooley WH, Hamilton WK. Treatment of the idiopathic respiratory-distress syndrome with continuous positive airway pressure. New England Journal of Medicine 1971;284:1333-40. [DOI] [PubMed] [Google Scholar]

Gupta 2016

  1. Gupta S, Donn SM. Continuous positive airway pressure: physiology and comparison of devices. Seminars in Fetal and Neonatal Medicine 2016;21(3):204-11. [DOI] [PubMed] [Google Scholar]

Gupta 2016a

  1. Gupta S, Donn SM. Continuous positive airway pressure: to bubble or not to bubble? Clinics in Perinatology 2016;43(4):647-59. [DOI: 10.1016/j.clp.2016.07.003] [DOI] [PubMed] [Google Scholar]

Harbord 2006

  1. Harbord RM, Egger M, Sterne JA. A modified test for small-study effects in meta-analyses of controlled trials with binary endpoints. Statistics in Medicine 2006;25(20):3443-57. [DOI: 10.1002/sim.2380] [PMID: ] [DOI] [PubMed] [Google Scholar]

Higgins 2011

  1. Higgins JP, Green S, editor(s). Cochrane Handbook for Systematic Reviews of Interventions Version 5.1.0 (updated March 2011). The Cochrane Collaboration, 2011. Available from training.cochrane.org/handbook/archive/v5.1/.

Higgins 2020

  1. Higgins JP, Thomas J, Chandler J, Cumpston M, Li T, Page MJ, et al (editors). Cochrane Handbook for Systematic Reviews of Interventions version 6.1 (updated September 2020). Cochrane, 2020. Available from training.cochrane.org/handbook/archive/v6.1.

Ho 2020

  1. Ho JJ, Subramaniam P, Davis PG. Continuous positive airway pressure (CPAP) for respiratory distress in preterm infants. Cochrane Database of Systematic Reviews 2020, Issue 10. Art. No: CD002271. [DOI: 10.1002/14651858.CD002271.pub3] [DOI] [PMC free article] [PubMed] [Google Scholar]

Hopewell 2009

  1. Hopewell S, Loudon K, Clarke MJ, Oxman AD, Dickersin K. Publication bias in clinical trials due to statistical significance or direction of trial results. Cochrane Database of Systematic Reviews 2009, Issue 1. Art. No: MR000006. [DOI: 10.1002/14651858.MR000006.pub3] [PMID: ] [DOI] [PMC free article] [PubMed] [Google Scholar]

Horbar 2012

  1. Horbar JH, Carpenter JH, Badger GJ, Kenny MJ, Soll RF, Morrow KA, et al. Mortality and neonatal morbidity among infants 501 to 1500 grams from 2000 to 2009. Pediatrics 2012;129(6):1019-26. [DOI: 10.1542/peds.2011-3028] [PMID: ] [DOI] [PubMed] [Google Scholar]

ICROP 2005

  1. International Committee for the Classification of Retinopathy of Prematurity. The International Classification of Retinopathy of Prematurity revisited. Archives of Ophthalmology 2005;123(7):991-9. [DOI: 10.1001/archopht.123.7.991] [PMID: ] [DOI] [PubMed] [Google Scholar]

Imbulana 2018

  1. Imbulana DI, Manley BJ, Dawson JA, Davis PG, Owen LS. Nasal injury in preterm infants receiving non-invasive respiratory support: a systematic review. Archives of Disease in Childhood Fetal and Neonatal Edition 2018;103(1):F29-F35. [DOI: 10.1136/archdischild-2017-313418] [DOI] [PubMed] [Google Scholar]

Jobe 2001

  1. Jobe AH, Bancalari E. Bronchopulmonary dysplasia. American Journal of Respiratory and Critical Care Medicine 2001;163(7):1723-9. [PMID: ] [DOI] [PubMed] [Google Scholar]

Kamper 1990

  1. Kamper J, Ringsted C. Early treatment of idiopathic respiratory distress syndrome using binasal continuous positive airway pressure. Acta Paediatrica Scandinavica 1990;79:581-6. [DOI] [PubMed] [Google Scholar]

Kinshella 2020

  1. Kinshella MW, Walker CR, Hiwa T, Vidler M, Nyondo-Mipando AL, Dube Q, et al. Barriers and facilitators to implementing bubble CPAP to improve neonatal health in sub-Saharan Africa: a systematic review. Public Health Reviews 2020;41:6. [DOI: 10.1186/s40985-020-00124-7] [DOI] [PMC free article] [PubMed] [Google Scholar]

Krouskop 1975

  1. Krouskop RW, Brown EG, Sweet AY. The early use of continuous positive airway pressure in the treatment of idiopathic respiratory distress syndrome. Journal of Pediatrics 1975;87:263-7. [DOI] [PubMed] [Google Scholar]

Laughton 2011

  1. Laughon M, Bose C, Allred EN, O'Shea TM, Ehrenkranz RA, Van Marter LJ, et al. Antecedents of chronic lung disease following three patterns of early respiratory disease in preterm infants. Archives of Disease in Childhood Fetal & Neonatal Edition 2011;96(2):F114–20. [DOI] [PMC free article] [PubMed] [Google Scholar]

Lee 1998

  1. Lee KS, Dunn MS, Fenwick M, Shennan AT. A comparison of underwater bubble continuous positive airway pressure with ventilator-derived continuous positive airway pressure in premature neonates ready for extubation. Biology of the Neonate 1998;73:69-75. [DOI] [PubMed] [Google Scholar]

Lemyre 2016

  1. Lemyre B, Laughon M, Bose C, Davis PG. Early nasal intermittent positive pressure ventilation (NIPPV) versus early nasal continuous positive airway pressure (NCPAP) for preterm infants. Cochrane Database of Systematic Reviews 2016, Issue 12. Art. No: CD005384. [DOI: 10.1002/14651858.CD005384.pub2] [DOI] [PMC free article] [PubMed] [Google Scholar]

Lemyre 2017

  1. Lemyre B, Davis PG, De Paoli AG, Kirpalani H. Nasal intermittent positive pressure ventilation (NIPPV) versus nasal continuous positive airway pressure (NCPAP) for preterm neonates after extubation. Cochrane Database of Systematic Reviews 2017, Issue 2. Art. No: CD003212. [DOI: 10.1002/14651858.CD003212.pub3] [DOI] [PMC free article] [PubMed] [Google Scholar]

Lissauer 2017

  1. Lissauer T, Duke T, Mellor K, Molyneux L. Nasal CPAP for neonatal respiratory support in low and middle-income countries. Archives of Disease in Childhood Fetal and Neonatal Edition 2017;102(3):F194-F196. [DOI: 10.1136/archdischild-2016-311653] [DOI] [PubMed] [Google Scholar]

Martin 1977

  1. Martin RJ, Nearman HS, Katona PG, Klaus MH. The effect of a low continuous positive airway pressure on the reflex control of respiration in the preterm infant. Journal of Pediatrics 1977;90:976-81. [DOI] [PubMed] [Google Scholar]

McGoldrick 2020

  1. McGoldrick E, Stewart F, Parker R, Dalziel SR. Antenatal corticosteroids for accelerating fetal lung maturation for women at risk of preterm birth 10.1002/14651858.CD004454.pub4. Cochrane Database of Systematic Reviews 2020, Issue 12. Art. No: CD004454. [DOI: 10.1002/14651858.CD004454.pub4] [DOI] [PMC free article] [PubMed] [Google Scholar]

Miller 1985

  1. Miller MJ, Carlo WA, Martin RJ. Continuous positive airway pressure selectively reduces obstructive apnea in preterm infants. Journal of Pediatrics 1985;106:91-4. [DOI] [PubMed] [Google Scholar]

Moa 1993

  1. Moa G, Nilsson K. Nasal continuous positive airway pressure: experiences with a new technical approach. Acta paediatrica 1993;82(2):210-211. [DOI] [PubMed] [Google Scholar]

Moa 1998

  1. Moa GU, Nilsson KJ, Zetterström H, Jonsson LO. A new device for administration of nasal continuous positive airway pressure in the newborn: an experimental study. Critical Care Medicine 1988;16(12):1238-42. [DOI] [PubMed] [Google Scholar]

Morley 2005

  1. Morley CJ, Lau R, De Paoli A, Davis PG. Nasal continuous positive airway pressure: does bubbling improve gas exchange? Archives of Disease in Childhood Fetal & Neonatal Edition 2005;90(4):F343-4. [DOI: 10.1136/adc.2004.062588] [DOI] [PMC free article] [PubMed] [Google Scholar]

Mukerji 2017

  1. Mukerji A, Shah PS, Shivananda S, Yee W, Read B, Minski J, et al. Survey of noninvasive respiratory support practices in Canadian neonatal intensive care units. Acta Paediatrica 2017;106(3):387-93. [DOI: 10.1111/apa.13644] [DOI] [PubMed] [Google Scholar]

Papile 1978

  1. Papile LA, Burstein J, Burstein R, Koffler H. Incidence and evolution of subependymal and intraventricular hemorrhage: a study of infants with birth weights less than 1,500 gm. Journal of Pediatrics 1978;92(4):529-34. [PMID: ] [DOI] [PubMed] [Google Scholar]

Pillow 2007

  1. Pillow JJ, Hillman N, Moss TJ, Polglase G, Bold G, Beaumont C, Ikegami M, Jobe AH. Bubble continuous positive airway pressure enhances lung volume and gas exchange in preterm lambs. American Journal of Respiratory and Critical Care Medicine 2007;176(1):63-9. [DOI: 10.1164/rccm.200609-1368OC] [DOI] [PMC free article] [PubMed] [Google Scholar]

Pillow 2012

  1. Pillow J. Which continuous positive airway pressure system is best for the preterm infant with respiratory distress syndrome? Clinics in Perinatology 2012;39(3):483-96. [DOI: 10.1016/j.clp.2012.06.007] [DOI] [PubMed] [Google Scholar]

Richardson 1978

  1. Richardson CP, Jung AL. Effects of continuous positive airway pressure on pulmonary function and blood gases of infants with respiratory distress syndrome. Pediatric Research 1978;12:771-4. [DOI] [PubMed] [Google Scholar]

Schünemann 2013

  1. Schünemann H, Brożek J, Guyatt G, Oxman A, editor(s). Handbook for grading the quality of evidence and the strength of recommendations using the GRADE approach (updated October 2013). GRADE Working Group, 2013. Available from gdt.guidelinedevelopment.org/app/handbook/handbook.html.

Soll 2019

  1. Soll RF, Barkhuff W. Noninvasive ventilation in the age of surfactant administration. Clinics in Perinatology 2019;46(3):493-516. [DOI: 10.1016/j.clp.2019.05.002] [DOI] [PubMed] [Google Scholar]

Stoll 2015

  1. Stoll BJ, Hansen NI, Bell EF, Walsh MC, Carlo WA, Shankaran S, et al. Trends in care practices, morbidity, and mortality of extremely preterm neonates, 1993-2012. JAMA 2015;34(10):1039-51. [DOI: 10.1001/jama.2015.10244] [DOI] [PMC free article] [PubMed] [Google Scholar]

Subramaniam 2016

  1. Subramaniam P, Ho JJ, Davis PG. Prophylactic nasal continuous positive airway pressure for preventing morbidity and mortality in very preterm infants. Cochrane Database of Systematic Reviews 2016, Issue 6. Art. No: CD001243. [DOI: 10.1002/14651858.CD001243.pub3] [DOI] [PubMed] [Google Scholar]

Sweet 2019

  1. Sweet DG, Carnielli V, Greisen G, Hallman M, Ozek E, te Pas A, Plavka R, et al. European consensus guidelines on the management of respiratory distress syndrome – 2019 update. Neonatology 2019;115:432-450. [DOI: 10.1159/000499361] [DOI] [PMC free article] [PubMed] [Google Scholar]

Thukral 2016

  1. Thukral A, Sankar MJ, Chandrasekaran A, Agarwal R, Paul VK. Efficacy and safety of CPAP in low- and middle-income countries. Journal of Perinatology 2016;36(Suppl 1):S21-S28. [DOI: 10.1038/jp.2016.29] [DOI] [PMC free article] [PubMed] [Google Scholar]

Vinekar 2019

  1. Vinekar A, Dogra M, Azad RV, Gilbert C, Gopal L, Trese M. The changing scenario of retinopathy of prematurity in middle and low income countries: Unique solutions for unique problems. Indian Journal of Ophthalmology 2019;67(6):717-719. [DOI: 10.4103/ijo.IJO_496_19] [DOI] [PMC free article] [PubMed] [Google Scholar]

Walsh 2021

  1. Walsh V, McGuire W, Halliday HL. Evaluation of the quality of perinatal trials: making the GRADE. Neonatology 2021;118(3):378-83. [DOI: 10.1159/000516239] [PMID: ] [DOI] [PubMed] [Google Scholar]

Welty 2016

  1. Welty SE. Continuous positive airway pressure strategies with bubble nasal continuous positive airway pressure: not all bubbling Is the same: The Seattle positive airway pressure system. Clinics in Perinatology 2016;43(4):661-671. [DOI: 10.1016/j.clp.2016.07.004] [DOI] [PubMed] [Google Scholar]

WHO 2016

  1. World Health Organization (WHO). Oxygen therapy for children: a manual for health workers. World Health Organization (Geneva) 2016:https://apps.who.int/iris/handle/10665/204584.

Wilkinson 2016

  1. Wilkinson D, Andersen C, O'Donnell CP, De Paoli AG, Manley BJ. High flow nasal cannula for respiratory support in preterm infants. Cochrane Database of Systematic Reviews 2016, Issue 2. Art. No: CD006405. [DOI: 10.1002/14651858.CD006405.pub3] [DOI] [PMC free article] [PubMed] [Google Scholar]

Won 2019

  1. Won A, Suarez-Rebling D, Baker AL, Burke TF, Nelson BD. Bubble CPAP devices for infants and children in resource-limited settings: review of the literature. Paediatrics and International Child Health 2019;39(3):168-176. [DOI: 10.1080/20469047.2018.1534389] [DOI] [PubMed] [Google Scholar]

World Bank 2021

  1. World Bank. World Bank County and Lending Groups. datahelpdesk.worldbank.org/knowledgebase/articles/906519-world-bank-country-and-lending-groups (accessed 4 May 2021).

Wright 2016

  1. Wright CJ, Polin RA, Kirpalani H. Continuous positive airway pressure to prevent neonatal lung injury: how did we get here, and how do we improve? Journal of Pediatrics 2016;173:17–e2. [DOI] [PubMed] [Google Scholar]

Youngquist 2013

  1. Youngquist TM, Richardson CP, DiBlasi RM. Effects of condensate in the exhalation limb of neonatal circuits on airway pressure during bubble CPAP. Respiratory Care 2013;58(11):1840-6. [DOI: 10.4187/respcare.02322] [DOI] [PubMed] [Google Scholar]

Yu 1977

  1. Yu VYH, Rolfe P. Effect of continuous positive airway pressure breathing on cardiorespiratory function in infants with respiratory distress syndrome. Acta Paediatrica Scandinavica 1977;66:59-64. [DOI] [PubMed] [Google Scholar]

References to other published versions of this review

De Paoli 2002

  1. De Paoli AG, Davis PG, Faber B, Morley CJ. Devices and pressure sources for administration of nasal continuous positive airway pressure (NCPAP) in preterm neonates. Cochrane Database of Systematic Reviews 2002, Issue 4. Art. No: CD002977. [DOI: 10.1002/14651858.CD002977] [DOI] [PubMed] [Google Scholar]

De Paoli 2008

  1. De Paoli AG, Davis PG, Faber B, Morley CJ. Devices and pressure sources for administration of nasal continuous positive airway pressure (NCPAP) in preterm neonates. Cochrane Database of Systematic Reviews 2008, Issue 1. Art. No: CD002977. [DOI: 10.1002/14651858.CD002977.pub2] [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from The Cochrane Database of Systematic Reviews are provided here courtesy of Wiley

RESOURCES