The effects of metal nanoparticles incorporation on the mechanical properties of denture base acrylic resin

Secil Ozkan Ata 1,*, Canan Akay 2, Emre Mumcu 3
PMCID: PMC10069800  PMID: 37020640

Abstract

Purpose:

The aim of this study was to examine the flexural strength of acrylic resin base material incorporated with iron, copper, and titanium nanoparticles.

Materials and methods:

Seventy bars of samples (65x10x2.5 mm3) were divided into seven groups. Acrylic samples were prepared according to the manufacturer’s instructions. Fe2O3, CuO and TiO2 nanoparticles were manually added in a proportion of 1wt% and 3wt% to the heat-polymerized acrylic resin. The Universal Testing Machine was used for 3-point flexural test of 5 mm/min force. ANOVA and Weibull analyses were used for the statistical analyses.

Results:

A statistical difference was found between the nanoparticle-added group and the control group. The highest mean value was observed for the 1wt% TiO2 added group, (84.99 MPa) and the lowest value was for the 3wt% CuO added group (71.32 MPa) (p<0,001). The 3wt% Fe2O3 and CuO added groups showed lower values than the control group.

Conclusion:

The incorporation of TiO2 nanoparticles into acrylic resin in a proportion of 1wt% increased the flexural strength values of the resins. Within the limitations, the nanoparticle addition to acrylic resins could improve the mechanical properties; however, when the percentage of nanoparticle addition increases, the flexural strength values of the acrylic resins decrease.

Keywords: Acrylic resin , nanoparticle, flexural strength , prosthodontics, weibull analysis

Introduction

Acrylic resins are commonly used denture base materials. It is preferred especially due to its easy laboratory processes, low weight, reasonable price as well as convenient aesthetic and color matching properties (1, 2). However, high thermal expansion coefficient, low thermal conductivity and denture fractions are among the disadvantages of this material (1,3). Especially denture fractions that occur by mistakenly dropping, or fatigue due to recurring stretches caused by chewing activities, are still serious problems (4, 5). There are studies focusing on the methods to prevent acrylic resin fractures and improve its mechanical features by reinforcing with different amounts of metal, metal oxide particles, carbon and fibers such as glass aramid (6, 7).

Recently, the use of nanoparticles has become increasingly common because of their anti-corrosion features, biocompatibility and resistance to fatigue and rupture. There are some studies in the literature examining the mechanical effects of denture base acrylic resins incorporated with nanoparticles such as copper oxide (CuO), iron oxide (Fe2O3), zinc oxide (ZnO), titanium dioxide (TiO2), aluminum oxide (Al2O3) and silicon oxide (SiO2) (7, 8, 9). According to these studies, the concentration, size, and shape of nanoparticles might affect the endurance and the fracture resistance of materials (8). It was reported that adding nanoparticles in a proper concentration might improve the mechanical properties of a material (9).

Based on the literature review, few studies examined iron and copper nanoparticle incorporation into heat-polymerized acrylic resins. Therefore, the aim of this study was to examine the flexural strength of heat-cure acrylic resins incorporated with 1wt% and 3wt% CuO, Fe2O3 and TiO2 nanoparticles. The null hypothesis was that the nanoparticle added groups would not have higher flexural strength values.

Materials and methods

Sample preparation

Materials used in this study were listed in Table 1. For sample preparation, stainless steel molds were used according to the ISO 1567 standards (2.5 mm thick, 10 mm wide and 65 mm long) (10). Pink Wax (Cavex Set Up wax; Cavex, Holland) was put into the molds, after 70 wax samples were prepared, they invested in dental stone (Fujirock EP; GC) by using a metal denture mold. Wax was eliminated by conventional methods.

Table 1.

Experimental materials used in this study.

Material Name Manufacturer Composition
Panacryl Heat-cure acrylic resin Arma Dental, İstanbul, Turkey 95% Methyl Methacrylate (MMA), 5% Ethilenglicoldimethilacrylate (EGDMA)
Fe2O3 nanoparticles Nanografi Nanotechnology, Ankara, Turkey 99.9 % purity 50 nm particle size
CuO nanoparticles Nanografi Nanotechnology, Ankara, Turkey 99.9 % purity 20 nm particle size
TiO2 nanoparticles Nanografi Nanotechnology, Ankara, Turkey 99.9 % purity 30 nm particle size

The heat-cure acrylic resin samples (Panacryl, Arma Dental, İstanbul, Turkey) were prepared by mixing powder and liquid at a 25 gr/10 ml ratio according to the manufacturer’s recommendation. As for the groups with nanoparticle incorporation in a proportion of 1wt% and 3wt%, Fe2O3 nanoparticles (99.9 % purity 50 nm particle size), CuO nanoparticles (99.9 % purity 20 nm particle size) and TiO2 nanoparticles (99.9 % purity 30 nm particle size) were added to acrylic resin using a digital precision scale (Nanografi Nanotechnology, Ankara, Turkey). The acrylic resin powder with nanoparticles (1 wt% and 3 wt%) were thoroughly homogenized in a mixer (President Dental, Germany) in a 2900 rpm cycle for 30 seconds.

For polymerization, the resins were put into the mold, were kept for 8 h at 74±1 ºC in water, after 8 h, and they were boiled for 2 h. After the resin samples were retrieved from the molds, they were polished with 200, 400 and 600 grits respectively to obtain a standard surface (Waterproof silicon carbide paper, English Abrasives Ltd., London, England) for 5 minutes under the water-cooling (Figure 1). The sizes of the samples were measured to ensure standardization for the test processes. The samples were kept in distilled water for one week.

Figure 1.

Figure 1.

Experimental design of the study (A: 70 waxes, B: Waxes placed in muffles, C: Control and modified heat-cure acrylic resins, D: Flexural strength test).

Flexural strength test

Universal Testing Machine (Lloyd-LRX, Lloyd Instruments, Fareham, UK) was used to determine the flexural resistance. A load was applied in the center of the samples at the crosshead speed of 5 mm/min until a fracture occurs. The maximum force (N) and the flexural strength values (mm) of each sample were automatically recorded in the processing unit of the machine.

Statistical analysis

Statistical package software (SPSS Version 24.0; SPSS Inc., Chicago, IL, USA) was used for data analysis. The data obtained in the study displayed a statistically normal distribution. Since the data had normal distribution within the groups, the means and the variations among the groups were examined by using the One-Way Variance Analysis (ANOVA) and the post-hoc Tukey test. The confidence interval was set to 95% and p <0.05 was considered statistically significant. Variations within groups were examined by calculating the Weibull module (m). The fracture values were classified from minimum to maximum, and line graphics were obtained by using the median regression method.

Results

The values obtained from the flexural strength tests were compared by using ANOVA and Tukey test. The means, standard deviation values, minimum -maximum and Weibull analysis results are shown in Table 2. Boxplot graph and Weibull graph are shown in Figure 2 and Figure 3. According to the flexural strength test results in this study, the significantly highest flexural strength value was found for 1wt% TiO2 added acrylic resin. 3wt% CuO added acrylic resin had lower flexural strength values than those of the control group. Similarly, flexural resistance values, Weibull modulus and Weibull characteristics of 3wt% Fe2O3, CuO and TiO2 added acrylic resins were lowest than those of the other groups.

Table 2.

The mean flexural strength values, SDs, min and max values of each group.

Groups Mean ± SD Min Max Weibull Modulus Weibull Characteristics
1% CuO 76.58±1.40 69.35 82.56 18.31 78.68
3% CuO 71.32±1.99 60.24 79.41 12.26 74.16
1% Fe2O3 74.91±1.37 68.11 81.26 18.88 76.90
3% Fe2O3 73.12±1.56 61.81 78.40 15.38 73.99
1% TiO2 84.99±1.66 79.12 93.35 16.53 87.52
3% TiO2 72.07±1.84 61.34 79.11 13.38 74.72
Control 74.20±1.43 67.79 79.90 17.33 76.35

Figure 2.

Figure 2.

Boxplot graph of flexural stress values and statistical significant differences.

Figure 3.

Figure 3.

Weibull plot of failure probability against stress to failure (MPa) for each group.

Discussion

In the present study, 1wt% TiO2 added group had significantly higher flexural strength value than all other groups, while 3wt% Fe2O3, CuO and TiO2 added groups had lower flexural strength value than the other groups. Thus, the null hypothesis of the study was rejected.

Some recent studies in the literature examined the effects of adding nanoparticles in different concentrations to enhance the resistance of denture base material against fractures and improve its mechanical properties. According to these studies, the addition of nanoparticles in various amounts ranging between 1wt% and 5wt% strengthens the mechanical properties of materials; however, these properties weaken when the amount exceeds 5wt% (11, 12). Therefore, 1wt% and 3wt% nanoparticle incorporation were examined in the present study. The flexural strength values obtained in this study also showed a significant decrease in flexural strength values for 3wt% concentration added groups when compared to other groups. Compared to monomers and polymers, nanoparticles might negatively affect the properties of a material by clustering in large sizes (13, 14). When nanoparticles are inappropriately distributed in an acrylic resin matrix, monomer reaction decreases, and the amount of non-reactive monomers increases (15). It is also possible that the stress concentrations caused by filler particles alter the elasticity module and crack the growth mode of resin (14). Therefore, if the nanoparticle can be uniformly distributed in the resin, the stress concentration can be avoided, and the mechanical properties of the resin can be improved (16). Therefore, in this study, the nanoparticles were incorporated into resin with a mixing machine for 30 minutes to ensure homogenization.

Toodehzaeim et al. (17) added 0.01wt, 0.5wt and 1wt% CuO nanoparticles to nanocomposites It was reported that the shear bond strength values did not reveal a significant difference between the control group and CuO particle added groups. The results of the present study showed a decrease in strength values when compared to the control group in the case of CuO particle addition. Poosti et al. (18) examined the mechanical properties of TiO2 added resins by using the shear bond test and did not find a difference between TiO2 added group and the control group. The difference in our study might be explained by using different material and mechanical tests as well as different nanoparticle proportions.

The 3-point bending test is a commonly used and well-recognized method to measure the bending resistance of denture base polymers in conformity with international ISO 1567 standards (10). The samples were applied 5 mm/min force as suggested in Barbosa’s study (19). According to ISO 1567, the bending resistance of an acrylic resin polymerized by using any known methods should not be lower than 65 MPa (10). The group with the lowest bending resistance in our study was the 3% CuO added group (71.32 Mpa), which might be explained by the fact that acrylic resin failed to complete the chain reaction and the full polymerization due to metal nanoparticles. Similarly, the literature review showed that the effect of incorporating nanoparticles such as Fe2O3, TiO2 and SiO2 on the mechanical properties of acrylic resins is directly related to the nature of “concentration”. The strength of a material decreases as nanoparticle concentration increases (7, 11). In contrast, a study found that the addition of 1wt% TiO2 decreases the micro hardness of an acrylic resin while adding 5wt% TiO2 causes an increase in this type of hardness. Since such an increase causes higher fatigue resistance in acrylic resin, it can be used in certain equipment restorations such as occlusal splints (20, 21). On the other hand, since higher concentrations (5wt% TiO2) are likely to deteriorate the impact resistance of a resin, any attempt to add nanoparticles after reaching saturation during polymerization might lead to interruptions in the sustainability of resin chains (22). In addition, it was observed that nanometals added in high concentrations might cause great changes in the color of the acrylic resin.

Weibull analysis was included in the present study because it provides information about the variations in the results as well as the reliability and endurance of the material used and makes a more accurate clustering of fracture tension values and indicates. The highest Weibull characteristic value in this study was found in 1wt% TiO2 added group. A high Weibull characteristic value means that the material is durable and displays better mechanical properties. On the other hand, the lowest Weibull characteristic values were calculated for the groups in which 3wt% nanoparticles were added. These findings imply that acrylic resins might get weaker and have lower fracture resistance in time. The Weibull results in our study are consistent with the flexural resistance test results. Thus, Weibull distribution can be considered an alternative method in evaluating the fracture possibility of materials (23).

This invitro study is limited, as it does not accurately simulate real conditions of the oral cavity. When it is considered the fact that the acrylic resin might be affected by factors such as temperature and Ph changes in the oral cavity, it might be difficult to predict the material’s original clinical reactions by interpreting the in vitro reactions. Obtaining microscopic views of fractured surfaces might provide more detailed information about fatigue resistance by allowing an effective examination of surface structure and some possible voids inside the resin interface. Due to the mentioned limitations, it is necessary to conduct more advanced clinical and in vitro studies examining the effects of incorporating various nanoparticles into acrylic resin.

Conclusion

It can be concluded that adding 1wt%TiO2 nanoparticles to acrylic resin improved the mechanical properties of acrylic resins. Despite some limitations, nanoparticle addition to acrylic resins might improve its mechanical features; however, flexural resistance values of acrylic resin decreases as the percentage of incorporated nanoparticles increases.

Footnotes

Ethics committee approval:Not required.

Informed consent: Not required.

Peer review:Externally peer-reviewed.

Author contributions: SOA, CA, EM participated in designing the study. SOA, CA, EM participated in generating the data for the study. SOA, CA, EM participated in gathering the data for the study. SOA, CA, EM participated in the analysis of the data. SOA, CA, EM wrote the majority of the original draft of the paper. SOA, CA, EM participated in writing the paper. SOA, CA, EM have had access to all of the raw data of the study. SOA, CA, EM have reviewed the pertinent raw data on which the results and conclusions of this study are based. SOA, CA, EM have approved the final version of this paper. SOA, CA, EM guarantee that all individuals who meet the Journal’s authorship criteria are included as authors of this paper.

Conflict of interest: The author had no conflict of interest to declare

Financial disclosure: The author declared that this study has received no financial support.

References

  • 1.Vikram S, Chander NG. Effect of zinc oxide nanoparticles on the flexural strength of polymethylmethacrylate denture base resin. Eur Oral Res. 2020. Jan;54(1):31–5. 10.26650/eor.20200063 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Salman TA, Khalaf HA. The influence of adding of modified ZrO2-TiO2 nanoparticles on certain physical and mechanical properties of heat polymerized acrylic resin. J Bagh College Dentistry. 2015;27(3):33–9. 10.12816/0015032 [DOI] [Google Scholar]
  • 3.Jasim BS, Ismail IJ. The effect of silanized alumina nano-fillers addition on some physical and mechanical properties of heat cured polymethyl methacrylate denture base material. J Bagh College of Dentistry. 2014;26(2):18–23. 10.12816/0015190 [DOI] [Google Scholar]
  • 4.Kim SH, Watts DC. The effect of reinforcement with woven E-glass fibers on the impact strength of complete dentures fabricated with high-impact acrylic resin. J Prosthet Dent. 2004. Mar;91(3):274–80. 10.1016/j.prosdent.2003.12.023 [DOI] [PubMed] [Google Scholar]
  • 5.Jagger DC, Harrison A. The effect of chopped poly(methyl methacrylate) fibers on some properties of acrylic resin denture base material. Int J Prosthodont. 1999;12(6):542–6. [PubMed] [Google Scholar]
  • 6.Oyar P, Sana FA, Nasseri B, Durkan R. Effect of green gold nanoparticles synthesized with plant on the flexural strength of heat-polymerized acrylic resin. Niger J Clin Pract. 2018. Oct;21(10):1291–5. [DOI] [PubMed] [Google Scholar]
  • 7.Moslehifard E, Robati Anaraki M, Shirkavand S. Effect of adding TiO2 nanoparticles on the SEM morphology and mechanical properties of conventional heat-cured acrylic resin. J Dent Res Dent Clin Dent Prospect. 2019;13(3):234–40. 10.15171/joddd.2019.036 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Sodagar A, Akhoundi MS, Bahador A, Jalali YF, Behzadi Z, Elhaminejad F, et al. Effect of TiO2 nanoparticles incorporation on antibacterial properties and shear bond strength of dental composite used in Orthodontics. Dental Press J Orthod. 2017;22(5):67–74. 10.1590/2177-6709.22.5.067-074.oar [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Agnihotri R, Gaur S, Albin S. Nanometals in dentistry: applications and toxicological implications-a systematic review. Biol Trace Elem Res. 2020. Sep;197(1):70–88. 10.1007/s12011-019-01986-y [DOI] [PubMed] [Google Scholar]
  • 10.International Organization for Standardization (CH) . International Standard ISO 1567, Dentistry-Denture base polymers. Geneve (Washington): International Organization for Standardization; 1999. [Google Scholar]
  • 11.Sodagar A, Bahador A, Khalil S, Shahroudi AS, Kassaee MZ. The effect of TiO2 and SiO2 nanoparticles on flexural strength of poly (methyl methacrylate) acrylic resins. J Prosthodont Res. 2013. Jan;57(1):15–9. 10.1016/j.jpor.2012.05.001 [DOI] [PubMed] [Google Scholar]
  • 12.Shi JM, Bao YZ, Huang ZM, Weng ZX. Preparation of poly (methyl methacrylate)/nanometer calcium carbonate composite by in-situ emulsion polymerization. J Zhejiang Univ Sci. 2004. Jun;5(6):709–13. 10.1631/jzus.2004.0709 [DOI] [PubMed] [Google Scholar]
  • 13.Burunkov J, Denisiuk I, Vorzobova N, Daroczi L. Hegedus Cs, Charnovych S, Kokenyesi S. Fabrication and characterization of gold/acrylic polymer nanocomposites. Eur Polym J. 2013;49(10):3072–7. 10.1016/j.eurpolymj.2013.05.024 [DOI] [Google Scholar]
  • 14.Ellakwa AE, Morsy MA, El-Sheikh AM. Effect of aluminum oxide addition on the flexural strength and thermal diffusivity of heat-polymerized acrylic resin. J Prosthodont. 2008. Aug;17(6):439–44. 10.1111/j.1532-849X.2008.00318.x [DOI] [PubMed] [Google Scholar]
  • 15.Xia Y, Zhang F, Xie H, Gu N. Nanoparticle-reinforced resin-based dental composites. J Dent. 2008. Jun;36(6):450–5. 10.1016/j.jdent.2008.03.001 [DOI] [PubMed] [Google Scholar]
  • 16.Fujishima M, Takatori H, Tada H. Interfacial chemical bonding effect on the photocatalytic activity of TiO2-SiO2 nanocoupling systems. J Colloid Interface Sci. 2011. Sep;361(2):628–31. 10.1016/j.jcis.2011.06.024 [DOI] [PubMed] [Google Scholar]
  • 17.Toodehzaeim MH, Zandi H, Meshkani H, Hosseinzadeh Firouzabadi A. The Effect of CuO nanoparticles on antimicrobial effects and shear bond strength of orthodontic adhesives. J Dent (Shiraz). 2018. Mar;19(1):1–5. [PMC free article] [PubMed] [Google Scholar]
  • 18.Poosti M, Ramazanzadeh B, Zebarjad M, Javadzadeh P, Naderinasab M, Shakeri MT. Shear bond strength and antibacterial effects of orthodontic composite containing TiO2 nanoparticles. Eur J Orthod. 2013. Oct;35(5):676–9. 10.1093/ejo/cjs073 [DOI] [PubMed] [Google Scholar]
  • 19.Barbosa DB, de Souza RF, Pero AC, Marra J, Compagnoni MA. Flexural strength of acrylic resins polymerized by different cycles. J Appl Oral Sci. 2007. Oct;15(5):424–8. 10.1590/S1678-77572007000500010 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Vojdani M, Bagheri R, Khaledi AA. Effects of aluminum oxide addition on the flexural strength, surface hardness, and roughness of heat-polymerized acrylic resin. J Dent Sci. 2012;7(3):238–44. 10.1016/j.jds.2012.05.008 [DOI] [Google Scholar]
  • 21.Lee SY, Lai YL, Hsu TS. Influence of polymerization conditions on monomer elution and microhardness of autopolymerized polymethyl methacrylate resin. Eur J Oral Sci. 2002. Apr;110(2):179–83. 10.1034/j.1600-0722.2002.11232.x [DOI] [PubMed] [Google Scholar]
  • 22.Asopa V, Suresh S, Khandelwal M, Sharma V, Asopa SS, Kaira LS. A Comparative evaluation of properties of zirconia reinforced high impact acrylic resin with that of high impact acrylic resin. The Saudi Journal for Dental Research. 2015;6(2):146–51. 10.1016/j.sjdr.2015.02.003 [DOI] [Google Scholar]
  • 23.Viwattanatipa N, Jermwiwatkul W, Chintavalakorn R, Kanchanavasita W. Weibull analysis of bond strength of orthodontic buccal tubes bonded to resin composite surface with various techniques. Orthod Waves. 2010;69(2):66–74. 10.1016/j.odw.2009.12.013 [DOI] [Google Scholar]

Articles from European Oral Research are provided here courtesy of Istanbul University Faculty of Dentistry

RESOURCES