Abstract
Objectives
Spatial–temporal biodistribution of antiretroviral drugs (ARVs) can now be achieved using MRI by utilizing chemical exchange saturation transfer (CEST) contrasts. However, the presence of biomolecules in tissue limits the specificity of current CEST methods. To overcome this limitation, a Lorentzian line-shape fitting algorithm was developed that simultaneously fits CEST peaks of ARV protons on its Z-spectrum.
Case presentation
This algorithm was tested on the common first line ARV, lamivudine (3TC), that has two peaks resulting from amino (–NH2) and hydroxyl (–OH) protons in 3TC. The developed dual-peak Lorentzian function fitted these two peaks simultaneously, and used the ratio of –NH2 and –OH CEST contrasts as a constraint parameter to measure 3TC presence in brains of drug-treated mice. 3TC biodistribution calculated using the new algorithm was compared against actual drug levels measured using UPLC-MS/MS. In comparison to the method that employs the –NH2 CEST peak only, the dual-peak Lorentzian fitting algorithm showed stronger correlation with brain tissue 3TC levels, signifying estimation of actual drug levels.
Conclusions
We concluded that 3TC levels can be extracted from confounding CEST effects of tissue biomolecules resulting in improved specificity for drug mapping. This algorithm can be expanded to measure a variety of ARVs using CEST MRI.
Keywords: antiretroviral drug, chemical exchange saturation transfer, lamivudine, magnetic resonance imaging
Introduction
Chemical exchange saturation transfer (CEST) contrast arises when an exchangeable proton of a molecule is magnetically saturated then transferred to water by chemical exchange. This incites a reduction in the magnetic resonance imaging (MRI) water signal, reflecting the molecule’s concentration. Continuous proton exchange leads to the buildup of water signal reductions enabling detection of the molecule in tissue. Unlike traditional drug imaging, which requires exogenous imaging contrast agents, CEST effect is an endogenous property of a molecule [1–10]. Indeed, we previously reported CEST effectiveness in detecting brain lamivudine (3TC) biodistribution [11]. The current study serves to advance CEST approach by ameliorating specificity and accuracy of determining antiretroviral drug (ARV) brain tissue biodistribution (BD).
CEST-based data analysis strategies have been examined in previous studies, such as the calculation of symmetric magnetization transfer ratio (MTRasym) from a Z-spectrum [12–16]. A Z-spectrum is the plot of water MRI signals at varying saturation frequencies. MTRasym is calculated by subtracting downfield water signals (at offset frequencies) from upfield water signals (at frequencies lower than water). However, MTRasym is confounded by magnetization transfer contrasts (MTC) from semi-solid macromolecules along with relayed nuclear Overhauser effects (rNOE) of macromolecules. To improve CEST specificity, a variety of quantification methods have been developed that included Lorentzian line-shape fitting [17–20], AREX [21–23] and chemical exchange rotation transfer [24–26]. Each method is able to differentiate MTC, rNOE and water spillover that is also called as water direct saturation (DS). Our previous works used 5-pool Lorentzian fitting to measure CEST effect of amino (–NH2) protons of 3TC to detect drug biodistribution [11]. Nonetheless, some tissue metabolites such as creatine and glutamine can, also, generate CEST peaks at about the same saturation frequency offset (2 ppm) as the 3TC–NH2 proton [19, 27], [28], [29], [30], [31], [32]. Therefore, the 2 ppm peak reflects the combined CEST effects of 3TC, creatine, glutamine and other biomolecules that have amine or amino protons. To improve the specificity of 3TC detections, we proposed a dual-peak Lorentzian line-shape function. This fitting function is used to account for the CEST peaks of –NH2 and –OH protons simultaneously. The inclusion of the –OH CEST effect adds a constraint for the fitting of –NH2 proton, and thus improves 3TC detection specificity. The proposed algorithm uses a 2-step fitting strategy to fit the in vivo Z-spectrum. In this approach background signal from direct saturation, MTC, and other biomolecules is fitted using a polynomial function. The peaks of –NH2 and –OH protons are then fitted using a dual-peak Lorentzian function. The method is now validated against drug levels measured using ultra performance liquid chromatography – tandem mass spectrometer (UPLC-MS/MS).
Materials and methods
CEST MRI of 3TC-treated mice
All animal studies were approved by the University of Nebraska Medical Center Institutional Animal Care and Use Committee (IACUC) in accordance with the standards incorporated in the Guide for the Care and Use of Laboratory Animals (National Research Council of the National Academies, 2011). Complete details about animal administration with 3TC and MRI acquisitions were previously described [11]. Briefly, male C57BL/6 mice (14–16 weeks old) were treated by oral gavage for five days with 3TC (250 mg/kg) or vehicle (0.2% hydroxypropylmethyl cellulose and 0.1% Tween 80 in sterile water; control group). CEST MRI was performed on a 7 Tesla MRI scanner (Bruker BioSpec 70/20, Billerica, MA). Experiments were performed using a RARE sequence (TR/TE = 1600/16 ms, RARE factor = 8) with a continuous RF for saturation with the power = 2 µT, duration = 1 s, saturation frequencies = −5 to 5 ppm in steps of 0.2 ppm. A library of prior CEST data sets were used for comparison with the dual-peak Lorentzian fitting algorithm.
Theory
A dual-peak Lorentzian fitting method was deployed to simultaneously analyze CEST effects of –NH2 and –OH protons of 3TC (Figure 1A). The 3TC CEST MTRasym analyses are illustrated in Figure 1B. The CEST effects of –NH2 and –OH protons were proportional to 3TC concentrations with correlation coefficients for –NH2 and –OH protons at R2=0.97 and 0.92, respectively (Figure 1C). Increasing rates of CEST contrasts with 3TC concentrations (slope of the line) were 0.0052 mM−1 for –NH2 proton, and 0.0026 mM−1 for –OH proton, respectively. The slope was defined as “CESTivity” reflecting the CEST effect change of a proton with drug concentration. The CESTivity of –NH2 proton was twice that of the –OH proton (Figure 1D). Dual-peak Lorentzian functions were developed to fit the CEST effects of –NH2 and –OH protons. This was achieved by converting Z-spectrum data into a linear combination of water relaxation rates of the –NH2 and –OH protons in the rotating frame under RF saturation by
(1) |
where Z (%) is the normalized Z-spectrum that is the original Z-spectrum normalized by unsaturated water signal, R1 is the longitudinal relaxation rate of water, θ is the tilt angle of the effective magnetization with respect to Z-axis, and R Iρ is the water relaxation rate [29–36]. For animal data,
(2) |
where RNH2 and ROH are the longitudinal relaxation rates of –NH2 and –OH protons in the rotating frame, respectively. R bg accounts for the rotating frame rate of water and all other effects including MTC of semi-solid macromolecules and CEST effects of biomolecules other than 3TC. R bg is represented by a polynomial function for DS and MTC, and a Lorentzian function for the CEST effect at 3 ppm from the amine and amide protons in biomolecules such as glutamate and mobile proteins:
(3) |
where C0, C1 and C2 are the 0th to the 2nd order polynomial coefficients, Δω is the frequency offset relative to water at 0 ppm refers to the apparent relaxation rate of the 3 ppm component, w3ppm is the full width at half maximum (FWHM) of the 3 ppm Lorentzian curve, and Δω3ppm is the chemical shift. RNH2 and ROH are fitted using Lorentzian functions:
(4) |
and
(5) |
where wNH2 and wOH are the FWHM of the peaks of amino and hydroxyl Lorentzian curves, respectively. ΔωNH2 and ΔωOH are the chemical shifts of –NH2 and –OH protons, respectively. and refer to the relaxation rates of –NH2 and –OH protons, respectively, and they are linked by
(6) |
where the initial value of r a/h is set to 0.5. This is estimated from the CESTivity ratio of –NH2 and –OH protons, which is 2 (Figure 1C) using Eq. (1). The background (Eq. (3)) is first fitted on 0.2–5 ppm that excluded data points between 0.5 and 2.5 ppm. In the second step, the data points in 0.5–2.5 ppm are fitted using Eqs. (4) and (5). The algorithm was first tested on averaged Z-spectra across brain image pixels of mice administered with 3TC (Figure 1D and E), in which the in vivo data in the frequency offset range 0.2–5 ppm are shown as dots. The background estimated using Eq. (4) is shown in a solid line in Figure 1D, and the solid line in Figure 1E represents the fitting result of the –NH2 and –OH effects using Eqs. (4) and (5). The average goodness-of-fit is higher than 0.9 for all data sets. The inclusion of hydroxyl effect adds a constraint for the fitting of amino proton via r a/h , and thus improves the accuracy of the fitting result. The algorithm is implemented in Matlab (MathWorks, Natick, MA) using “lsqcurvefit” function for both polynomial and Lorentzian fittings.
Statistical Analysis
Statistical analysis was performed using Matlab. Data were presented as mean ± standard error of the mean (SEM). For comparisons of two groups, Student’s t-tests (one-tailed) were used. Significant differences were determined at p<0.05, and trends of changes were defined as p<0.1. The associations of MRI CEST results and drug levels determined by UPLC-MS/MS were evaluated using Pearson correlations in Matlab.
Results
3TC-treated mice
The dual-peak fitting algorithm was tested on CEST MRI data previously acquired [11]. CEST effects of 3TC dosing were measured on brain sub-regions of C57BL/6 male mice. 3TC was administered orally for 5 days (250 mg/kg/day), in accordance to animal equivalent dose (AED) conversions [37], with dosing matched to the equivalent of five times the human dose. Control animals received 0.2% hydroxypropylmethyl cellulose and 0.1% Tween 80 in sterile water as a vehicle by equivalent administration routes. T2-weighted images were used as an anatomical reference for brain region-of-interest (ROI) evaluations (Figure 2A and D). CEST 3TC effects were measured in five brain sub-regions, including the hippocampus (HIP), cortex (CTX), piriform cortex (PIR), thalamus (TH) and hypothalamus (HY). The CEST maps on brain regions of 3TC and vehicle-treated mice are shown in Figures 2A–F. Stronger CEST effects of –NH2 and –OH protons were noted in the brain of drug treated mice (Figure 2E and F) compared to the vehicle-treated mice (Figure 2B and C). CEST effect of –NH2 proton was significantly higher in 3TC-treated mice (n=7) compared to controls (n=8) on thalamus (p=0.022) and piriform cortex (p=0.049) (Figure 2G). The trend of significant increase was observed on brain regions, cortex (p=0.070), hippocampus (p=0.059) and hypothalamus (p=0.078) in drug-treated animals. The CEST effect of –OH proton was higher in the thalamus (p=0.018) and piriform cortex (p=0.029) in 3TC-treated mice. Trends of significant increase were noted for –OH proton effects on hippocampus (p=0.076) and hypothalamus (p=0.058) in the drug treated group.
Next, we correlated CEST MRI results obtained by the dual-peak fitting algorithm with drug levels measured using UPLC-MS/MS in the 3TC-administered group (n=5) in the cortex and hippocampus regions (Figure 2I–N). In the cortex, –NH2 and –OH CEST effects were correlated with drug levels at coefficients R2=0.75 and 0.60, respectively (Figure 2I and J), whereas the correlation coefficient of the combined CEST effects of –NH2 and –OH protons with drug levels was R2=0.75 (Figure 2K). Within the hippocampus, the correlation coefficients of drug levels with –NH2 proton and with –OH proton were R2=0.44, and R2=0.10, respectively, (Figure 2L–M). Combined CEST effects of –NH2 and –OH protons with measured tissue drug levels showed a correlation coefficient of R 2 =0.56 (Figure 2N).
–NH2 proton -based algorithms
Results from the dual-peak fitting algorithm for 3TC biodistribution in brain regions were compared to results from a prior algorithm based only on –NH2 proton- CEST effect (Table 1). In the –NH2-only-based algorithm, the correlation coefficient of CEST effect in the cortex with drug levels was R2=0.62 (Table 1) [11]. Whereas, the constrained –NH2 CEST effects with the new algorithm showed higher correlation with drug levels, R2=0.75 (Figure 2, Table 1). These data demonstrated a high accuracy of dual-peak fitting algorithm for drug detection. The correlation coefficient of drug levels with combined –NH2 and –OH proton CEST effects was similar to that of constrained –NH2 (R2=0.75). In hippocampus region, accuracy of drug detection was improved after employing the dual-peak fitting algorithm. The correlation of constrained –NH2 effect with drug levels was R2=0.44 compared to R2=0.20 by the prior algorithm. Also, the correlation coefficient of combined –NH2 and –OH effects showed higher correlation coefficient R2=0.56. Taken together, the –NH2 CEST effect fitted with constraints correlated closely to actual drug levels. The correlation was further improved when the combined CEST effect of constrained –NH2 and –OH was deployed.
Table 1:
–NH2 only | Constrained –NH2 | Constrained –OH | Combined –NH2 and –OH | |
---|---|---|---|---|
CTX | 0.62a | 0.75 | 0.60 | 0.75 |
HIP | 0.20a | 0.44 | 0.10 | 0.56 |
aFrom [11].
Discussion
In traditional drug imaging modalities, drug molecules are tagged with imaging contrast agents or loaded with imaging contrast agents into nanoparticles. Usually paramagnetic metals are used for MRI (2–4) and radioactive materials for PET and SPECT (2–8) tests. Similar techniques were deployed to assay the biodistribution of nanoformulated ARVs [38–41]. The limitations of the prior methods are significant [1–10]. First, the loading rates of nanoparticles are usually limited to achieve effective therapy and/or imaging sensitivity [42]. Second, toxicity can be associated with imaging agents and nanoparticles [43–45]. Third, blood-brain barrier (BBB) penetration needs to be considered when designing traditional drug imaging methods for the CNS. Previously, we showed that the CEST effect can be used for in vivo imaging of 3TC [11]. Unlike traditional methods, the CEST contrast comes as a consequence of the drug’s chemical structure and therefore does not need extrinsic imaging agents. Thus, CEST imaging eliminates the limitations associated with extrinsic imaging agents used in traditional techniques. In this study, we developed the analysis algorithm to improve the accuracy and specificity of in vivo CEST MRI mapping of ARVs.
The 2-step fitting strategy, in which fitting of the background Z-spectrum is followed by the fitting of CEST peaks of interest, was used in previous studies for detection of creatine and phosphocreatine [29, 30, 32, 46]. The innovation of our algorithm is the simultaneous fitting of peaks of –NH2 and –OH protons that are linked and constrained via the CESTivity ratio (r a/h ) in Eq. (6). This extracts the –NH2 effect of 3TC from the peak that can contain effects from other tissue biomolecules such as creatine and glutamine serving to improves the specificity of 3TC detection. In summary, we successfully developed a new algorithm that uses the CEST effects of both amino (–NH2) and hydroxyl (–OH) protons for in vivo 3TC detection. The new algorithm shows high specificity for 3TC biodistribution measurements which can be extended to other ARVs.
Acknowledgments
The authors would thank Bioimaging (MRI) core facility for the help on data acquisition and processing.
Footnotes
Research funding: The study was partially supported by NIH R21MH128123, U54GM115458, R01MH121402, R21HD106842, P30GM127200, P20GM130447, and Nebraska Research Initiative.
Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.
Competing interests: Authors state no conflict of interest.
Informed consent: Not applicable.
Ethical approval: All animal studies were approved by the University of Nebraska Medical Center Institutional Animal Care and Use Committee (IACUC) in accordance with the standards incorporated in the Guide for the Care and Use of Laboratory Animals (National Research Council of the National Academies, 2011).
References
- 1.Dadfar SM, Roemhild K, Drude NI, von Stillfried S, Knüchel R, Kiessling F, et al. Iron oxide nanoparticles: diagnostic, therapeutic and theranostic applications. Adv Drug Deliv Rev. 2019;138:302–25. doi: 10.1016/j.addr.2019.01.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lux F, Sancey L, Bianchi A, Crémillieux Y, Roux S, Tillement O. Gadolinium-based nanoparticles for theranostic MRI-radiosensitization. Nanomedicine. 2015;10:1801–15. doi: 10.2217/nnm.15.30. [DOI] [PubMed] [Google Scholar]
- 3.Zhu L, Zhou Z, Mao H, Yang L. Magnetic nanoparticles for precision oncology: theranostic magnetic iron oxide nanoparticles for image-guided and targeted cancer therapy. Nanomedicine. 2017;12:73–87. doi: 10.2217/nnm-2016-0316. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Bailly C, Cléry PF, Faivre-Chauvet A, Bourgeois M, Guérard F, Haddad F, et al. Immuno-PET for clinical theranostic approaches. Int J Mol Sci. 2016;18:57. doi: 10.3390/ijms18010057. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Bodet-Milin C, Bailly C, Touchefeu Y, Frampas E, Bourgeois M, Rauscher A, et al. Clinical results in medullary thyroid carcinoma suggest high potential of pretargeted immuno-PET for tumor imaging and theranostic approaches. Front Med. 2019;6:124. doi: 10.3389/fmed.2019.00124. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Lenzo NP, Meyrick D, Turner JH. Review of gallium-68 PSMA PET/CT imaging in the management of prostate cancer. Diagnostics. 2018;8:16. doi: 10.3390/diagnostics8010016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Pruis IJ, van Dongen G, van Zanten SEMV. The added value of diagnostic and theranostic PET imaging for the treatment of CNS tumors. Int J Mol Sci. 2020;21:1029. doi: 10.3390/ijms21031029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Angelakeris M. Magnetic nanoparticles: a multifunctional vehicle for modern theranostics. Biochim Biophys Acta Gen Subj. 2017;1861:1642–51. doi: 10.1016/j.bbagen.2017.02.022. [DOI] [PubMed] [Google Scholar]
- 9.Li K, Nejadnik H, Daldrup-Link HE. Next-generation superparamagnetic iron oxide nanoparticles for cancer theranostics. Drug Discov Today. 2017;22:1421–9. doi: 10.1016/j.drudis.2017.04.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Mitchell MJ, Billingsley MM, Haley RM, Wechsler ME, Peppas NA, Langer R. Engineering precision nanoparticles for drug delivery. Nat Rev Drug Discov. 2021;20:101–24. doi: 10.1038/s41573-020-0090-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Bade AN, Gendelman HE, McMillan J, Liu Y. Chemical exchange saturation transfer for detection of antiretroviral drugs in brain tissue. Aids. 2021;35:1733–41. doi: 10.1097/qad.0000000000002960. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Pankowska A, Kochalska K, Lazorczyk A, Dyndor K, Koziol P, Zienczuk B, et al. Chemical exchange saturation transfer (CEST) as a new method of signal obtainment in magnetic resonance molecular imaging in clinical and research practice. Pol J Radiol. 2019;84:e147–52. doi: 10.5114/pjr.2019.84242. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Liu G, Song X, Chan KWY, McMahon MT. Nuts and bolts of chemical exchange saturation transfer MRI. NMR Biomed. 2013;26:810–28. doi: 10.1002/nbm.2899. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Vinogradov E, Sherry AD, Lenkinski RE. CEST: from basic principles to applications, challenges and opportunities. J Magn Reson. 2013;229:155–72. doi: 10.1016/j.jmr.2012.11.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Ward KM, Aletras AH, Balaban RS. A new class of contrast agents for MRI based on proton chemical exchange dependent saturation transfer (CEST) J Magn Reson. 2000;143:79–87. doi: 10.1006/jmre.1999.1956. [DOI] [PubMed] [Google Scholar]
- 16.Zaiss M, Bachert P. Chemical exchange saturation transfer (CEST) and MR Z-spectroscopy in vivo: a review of theoretical approaches and methods. Phys Med Biol. 2013;58:R221–69. doi: 10.1088/0031-9155/58/22/r221. [DOI] [PubMed] [Google Scholar]
- 17.Desmond KL, Moosvi F, Stanisz GJ. Mapping of amide, amine, and aliphatic peaks in the CEST spectra of murine xenografts at 7 T. Magn Reson Med. 2014;71:1841–53. doi: 10.1002/mrm.24822. [DOI] [PubMed] [Google Scholar]
- 18.Cui J, Afzal A, Zu Z. Comparative evaluation of polynomial and Lorentzian lineshape-fitted amine CEST imaging in acute ischemic stroke. Magn Reson Med. 2022;87:837–49. doi: 10.1002/mrm.29030. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Cai K, Singh A, Poptani H, Li W, Yang S, Lu Y, et al. CEST signal at 2 ppm (CEST@2 ppm) from Z-spectral fitting correlates with creatine distribution in brain tumor. NMR Biomed. 2015;28:1–8. doi: 10.1002/nbm.3216. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Zhou IY, Wang E, Cheung JS, Zhang X, Fulci G, Sun PZ. Quantitative chemical exchange saturation transfer (CEST) MRI of glioma using image downsampling expedited adaptive least-squares (IDEAL) fitting. Sci Rep. 2017;7:84. doi: 10.1038/s41598-017-00167-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Goerke S, Soehngen Y, Deshmane A, Zaiss M, Breitling J, Boyd PS, et al. Relaxation-compensated APT and rNOE CEST-MRI of human brain tumors at 3 T. Magn Reson Med. 2019;82:622–32. doi: 10.1002/mrm.27751. [DOI] [PubMed] [Google Scholar]
- 22.Zaiss M, Schuppert M, Deshmane A, Herz K, Ehses P, Fullbier L, et al. Chemical exchange saturation transfer MRI contrast in the human brain at 9.4T. Neuroimage. 2018;179:144–55. doi: 10.1016/j.neuroimage.2018.06.026. [DOI] [PubMed] [Google Scholar]
- 23.Zaiss M, Xu J, Goerke S, Khan IS, Singer RJ, Gore JC, et al. Inverse Z-spectrum analysis for spillover-MT-and T1 -corrected steady-state pulsed CEST-MRI–application to pH-weighted MRI of acute stroke. NMR Biomed. 2014;27:240–52. doi: 10.1002/nbm.3054. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Zu Z, Janve VA, Li K, Does MD, Gore JC, Gochberg DF. Multi-angle ratiometric approach to measure chemical exchange in amide proton transfer imaging. Magn Reson Med. 2012;68:711–9. doi: 10.1002/mrm.23276. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Zu Z, Janve VA, Xu J, Does MD, Gore JC, Gochberg DF. A new method for detecting exchanging amide protons using chemical exchange rotation transfer. Magn Reson Med. 2013;69:637–47. doi: 10.1002/mrm.24284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Zu Z, Louie EA, Lin EC, Jiang X, Does MD, Gore JC, et al. Chemical exchange rotation transfer imaging of intermediate-exchanging amines at 2 ppm. NMR Biomed. 2017;30:3756. doi: 10.1002/nbm.3756. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Lee JS, Xia D, Jerschow A, Regatte RR. In vitro study of endogenous CEST agents at 3 T and 7 T. Contrast Media Mol Imaging. 2016;11:4–14. doi: 10.1002/cmmi.1652. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Khlebnikov V, van der Kemp WJM, Hoogduin H, Klomp DWJ, Prompers JJ. Analysis of chemical exchange saturation transfer contributions from brain metabolites to the Z-spectra at various field strengths and pH. Sci Rep. 2019;9:1089. doi: 10.1038/s41598-018-37295-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Chen L, Barker PB, Weiss RG, van Zijl PCM, Xu J. Creatine and phosphocreatine mapping of mouse skeletal muscle by a polynomial and Lorentzian line-shape fitting CEST method. Magn Reson Med. 2019;81:69–78. doi: 10.1002/mrm.27514. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Chen L, Zeng H, Xu X, Yadav NN, Cai S, Puts NA, et al. Investigation of the contribution of total creatine to the CEST Z-spectrum of brain using a knockout mouse model. NMR Biomed. 2017;30:3834. doi: 10.1002/nbm.3834. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Chen L, Schär M, Chan KWY, Huang J, Wei Z, Lu H, et al. In vivo imaging of phosphocreatine with artificial neural networks. Nat Commun. 2020;11:1072. doi: 10.1038/s41467-020-14874-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Chen L, Wei Z, Cai S, Li Y, Liu G, Lu H, et al. High-resolution creatine mapping of mouse brain at 11.7 T using non-steady-state chemical exchange saturation transfer. NMR Biomed. 2019;32:e4168. doi: 10.1002/nbm.4168. [DOI] [PubMed] [Google Scholar]
- 33.Jin T, Autio J, Obata T, Kim SG. Spin-locking versus chemical exchange saturation transfer MRI for investigating chemical exchange process between water and labile metabolite protons. Magn Reson Med. 2011;65:1448–60. doi: 10.1002/mrm.22721. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Trott O, Palmer AG. R1rho relaxation outside of the fast-exchange limit. J Magn Reson. 2002;154:157–60. doi: 10.1006/jmre.2001.2466. [DOI] [PubMed] [Google Scholar]
- 35.Jin T, Wang P, Zong X, Kim S-G. Magnetic resonance imaging of the Amine-Proton EXchange (APEX) dependent contrast. Neuroimage. 2012;59:1218–27. doi: 10.1016/j.neuroimage.2011.08.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Sui R, Chen L, Li Y, Huang J, Chan KWY, Xu X, et al. Whole-brain amide CEST imaging at 3T with a steady-state radial MRI acquisition. Magn Reson Med. 2021;86:893–906. doi: 10.1002/mrm.28770. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Nair AB, Jacob S. A simple practice guide for dose conversion between animals and human. J Basic Clin Pharm. 2016;7:27–31. doi: 10.4103/0976-0105.177703. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Singh A, Kutscher HL, Bulmahn JC, Mahajan SD, He GS, Prasad PN. Laser ablation for pharmaceutical nanoformulations: multi-drug nanoencapsulation and theranostics for HIV. Nanomedicine. 2020;25:102172. doi: 10.1016/j.nano.2020.102172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Kevadiya BD, Ottemann B, Mukadam IZ, Castellanos L, Sikora K, Hilaire JR, et al. Rod-shape theranostic nanoparticles facilitate antiretroviral drug biodistribution and activity in human immunodeficiency virus susceptible cells and tissues. Theranostics. 2020;10:630–56. doi: 10.7150/thno.39847. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Kevadiya BD, Woldstad C, Ottemann BM, Dash P, Sajja BR, Lamberty B, et al. Multimodal theranostic nanoformulations permit magnetic resonance bioimaging of antiretroviral drug particle tissue-cell biodistribution. Theranostics. 2018;8:256–76. doi: 10.7150/thno.22764. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Ottemann BM, Helmink AJ, Zhang W, Mukadam I, Woldstad C, Hilaire JR, et al. Bioimaging predictors of rilpivirine biodistribution and antiretroviral activities. Biomaterials. 2018;185:174–93. doi: 10.1016/j.biomaterials.2018.09.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Dreifuss T, Betzer O, Shilo M, Popovtzer A, Motiei M, Popovtzer R. A challenge for theranostics: is the optimal particle for therapy also optimal for diagnostics? Nanoscale. 2015;7:15175–84. doi: 10.1039/c5nr03119b. [DOI] [PubMed] [Google Scholar]
- 43.Gupta N, Rai DB, Jangid AK, Kulhari H. A review of theranostics applications and toxicities of carbon nanomaterials. Curr Drug Metabol. 2019;20:506–32. doi: 10.2174/1389200219666180925094515. [DOI] [PubMed] [Google Scholar]
- 44.Bruna Galdorfini C-A, Marina Paiva A, Eloísa Berbel M, Bruna Lallo da S, Nathalia Cristina R, João Augusto O-J, et al. Drug delivery using theranostics: an overview of its use, advantages and safety assessment. Curr Nanosci. 2020;16:3–14. doi: 10.2174/1573413715666190618162321. [DOI] [Google Scholar]
- 45.Li M, Sagastume E, Zepeda-Orozco D, Wen H, Graves S, Lee D, et al. Preclinical evaluation of 203/212Pb-based theranostics-dosimetry and renal toxicity.<strong/>. J Nucl Med. 2020;61:289. [Google Scholar]
- 46.Chung JJ, Jin T, Lee JH, Kim SG. Chemical exchange saturation transfer imaging of phosphocreatine in the muscle. Magn Reson Med. 2019;81:3476–87. doi: 10.1002/mrm.27655. [DOI] [PMC free article] [PubMed] [Google Scholar]