Abstract.
Significance
Cerebrovascular reactivity (CVR), i.e., the ability of cerebral vasculature to dilate or constrict in response to vasoactive stimuli, is a biomarker of vascular health. Exogenous administration of inhaled carbon dioxide, i.e., hypercapnia (HC), remains the “gold-standard” intervention to assess CVR. More tolerable paradigms that enable CVR quantification when HC is difficult/contraindicated have been proposed. However, because these paradigms feature mechanistic differences in action, an assessment of agreement of these more tolerable paradigms to HC is needed.
Aim
We aim to determine the agreement of CVR assessed during HC, breath-hold (BH), and resting state (RS) paradigms.
Approach
Healthy adults were subject to HC, BH, and RS paradigms. End tidal carbon dioxide (EtCO2) and cerebral blood flow (CBF, assessed with diffuse correlation spectroscopy) were monitored continuously. CVR (%/mmHg) was quantified via linear regression of CBF versus EtCO2 or via a general linear model (GLM) that was used to minimize the influence of systemic and extracerebral signal contributions.
Results
Strong agreement (; ) among CVR paradigms was demonstrated when utilizing a GLM to regress out systemic/extracerebral signal contributions. Linear regression alone showed poor agreement across paradigms (; ).
Conclusions
More tolerable experimental paradigms coupled with regression of systemic/extracerebral signal contributions may offer a viable alternative to HC for assessing CVR.
Keywords: cerebrovascular reactivity, diffuse correlation spectroscopy, cerebral blood flow
1. Introduction
Cerebrovascular reactivity (CVR), defined as the ability of cerebral vasculature to dilate or constrict in response to a vasoactive stimulus, is an integral mechanism in brain homeostasis. Interest in CVR as a diagnostic and prognostic biomarker has accelerated in recent years because impaired CVR has been observed in numerous disease states, including cerebrovascular disease,1–7 stroke,8–13 cardiac arrest,14 and traumatic brain injury.15–23 Quantification of CVR coupled with interventions aimed at restoring abnormal CVR has the potential to reduce morbidity and mortality.
Several well-established experimental paradigms exist to study CVR non-invasively in humans. These methods typically involve quantification of the cerebral blood flow (CBF) response to changes in arterial carbon dioxide content via acetazolamide infusion, inhalation of carbon dioxide (CO2), or breath-holding. Intravenous acetazolamide administration has historically served as the preferred method to quantify CVR in clinical settings.24 Although the single dose injection is relatively simple, differences in pharmacokinetic/dynamic profiles among subjects lead to poor repeatability, and it is associated with a high incidence of side effects that preclude its modern use in many patient cohorts (e.g., children).6,7,10,13,24,25 In research settings, the most common method to quantify CVR is inhalation of CO2 to induce hypercapnia (HC), which is considered by many to be the “gold-standard” experimental paradigm. A standard HC challenge involves a limited duration inhalation of medical-grade (e.g., 5% , 21% oxygen, and balance nitrogen) coupled with monitoring of CBF and end tidal CO2 (EtCO2). This approach is increasingly used due to the potency of CO2 as a vasodilator, rapid onset and cessation of the CBF response, and repeatability.24,26–35 However, HC requires complex gas delivery equipment, time, and expertise to perform the procedure, and it can induce anxiety in some subjects. Moreover, HC may be clinically contraindicated in patients with acute illness and/or compromised hemodynamics for whom the benefits of CVR assessment do not outweigh the risks of rapid or prolonged hyperemia (e.g., sickle cell disease).36 A breath-hold (BH) challenge, wherein a subject briefly holds their breath, offers a simpler alternative for manipulating arterial CO2. This approach offers many advantages over CO2 inhalation, including a lack of complex gas delivery equipment and improved patient tolerability given the self-directed nature of the BH.37–40 However, the BH approach requires an individual’s cooperation, which limits applicability in children and in patients with severe conditions and entirely excludes its use in other cohorts (e.g., patients on a mechanical ventilator).
To bypass the limitations of existing experimental paradigms, recent investigations have explored the potential for measuring CVR using the intrinsic vasoactive stimulus of natural arterial CO2 fluctuations that occur when a subject is freely breathing.29,40–44 This resting state (RS) approach is attractive because it does not require patient compliance or intentional manipulation of arterial CO2. Thus, it has potential in those who have lost consciousness or who are sedated. However, the approach is thought to be susceptible to poor signal-to-noise due to the relatively small fluctuations in EtCO2 that occur during spontaneous breathing. Thus, CVR estimations at rest may be unreliable.45
Although the use of more tolerable experimental paradigms to assess CVR is clinically appealing, there has been limited investigation into the agreement of CVR measured by these approaches. Liu et al. recently demonstrated poor agreement of HC versus BH and of HC versus RS.45 The other handful of studies that have compared CVR across HC/BH/RS paradigms are limited in interpretation. These studies either compare relative changes in blood flow without accounting for differences in elicited EtCO2 response,35,45,46 which can vary appreciably across subjects and/or paradigms, or they focus on the spatial correlation of cohort-averaged, qualitative, regional CVR maps attained from two experimental paradigms.40,47,48 Given that these paradigms act on the vasculature in different ways and that each has relative merit in terms of robustness of response and ease of implementation, a systematic and rigorous comparison of CVR across paradigms is warranted.
Herein we quantify the agreement of CVR across HC, BH, and RS paradigms in healthy adults. To quantify CBF, we employ a non-invasive optical technique known as diffuse correlation spectroscopy (DCS). DCS offers many strategic advantages over more traditional modalities used to assess CBF (e.g., computed tomography, perfusion magnetic resonance imaging, and transcranial Doppler ultrasound), including portability, low cost, high temporal resolution, direct measure of microvascular blood flow, and ease of use in both adults and children in a wide array of clinical indications and environments. Indeed, several studies42,49–57 have explored the use of DCS to quantify the CBF response to a vasoactive stimulus (ACZ, HC). In this study, we use DCS to quantify CVR, and we hypothesize that CVR measured during HC will agree strongly with CVR assessed with more tolerable and more practical BH and RS paradigms.
2. Methods
Healthy adult subjects without a prior history of cardiovascular, neurological, respiratory, or hematological disorders were recruited at Emory University. Additional exclusion criteria included a history of major head injury within the past 2 years or a history of significant acute illness within one month of the study. Subjects were instructed to abstain from stimulants (e.g., caffeine) or depressants (e.g., alcohol) for at least 12 h prior to the study. Written informed consent was obtained for all participants. All protocols were approved by the Emory University Institutional Review Board.
Twenty-seven healthy adults (13 male) ranging in age from 22 to 37 years were enrolled in this study. A subset of four subjects was measured on multiple occasions, yielding 34 total measurement sessions. Of these sessions, one dataset was discarded due to substantial drift in optical signal attributed to poor sensor contact, and two datasets were discarded due to an improper execution of the HC exam, yielding a total of 31 available datasets.
2.1. Experimental Protocol
The CVR was quantified via three consecutive experimental paradigms: RS, BH, and HC (Fig. 1). First, for RS, the subject was directed to breath naturally (i.e., unpaced) for 10 min while minimizing bodily movements. Next, for BH, the subject was asked to complete 5 BH challenges, each of which consisted of 40 to 60 s paced breathing followed by a 20 to 30 s end-inspiration BH. Paced breathing at the subject’s natural respiration rate (determined during RS) was facilitated by a custom graphical user interface written in MATLAB (Mathworks Inc., Natick, Massachusetts, United States). End-inspiration BHs, which have been demonstrated to yield similar CVR magnitudes when compared to end-expiration BHs,38 were selected to facilitate subject comfort during the exam. Subjects were instructed to breathe in approximately half capacity before initiation of the BH and to exhale entirely at the conclusion of the BH. Finally, for HC, the subject was instructed to breathe naturally for 3 min, followed by 6 min of 5% CO2/balance room air inhalation (Nexair, Memphis, Tennessee, United States) and 6 min of room air inhalation recovery. Unpaced breathing was used throughout the HC challenge.
Subjects sat upright for the duration of the study protocol. Continuous (20 Hz) DCS measurements of CBF were made by securing an optical sensor over the right or left forehead. Given that CVR is a global phenomenon,58 laterality of the sensor placement was determined based on the hemisphere that yielded the highest detected light intensities to maximize the DCS signal-to-noise ratio. Trends in peripheral oxygen saturation (SpO2, LifeSense II, Nonin Medical, Inc.) and arterial blood pressure (ABP, CNAP Monitor 500, NIPD1000D-1, Biopac Systems) were also continuously monitored (1 Hz for SpO2 and 125 Hz for ABP, Fig. S1A in the Supplementary Material). In postprocessing, beat-to-beat mean ABP (MAP) was determined from the ABP waveform using BP_annotate in MATLAB.59 Further, a gas mask was placed over the mouth and nose to facilitate continuous monitoring of expelled (4 Hz, LifeSense II, Nonin Medical Inc., Plymouth, Minnesota, United States) as well as to enable delivery of carbon dioxide during HC. For RS and BH paradigms, the mask was configured such that the subject breathed room air. For the HC challenge, the mask was attached to a reservoir of 5% CO2. One-way, non-rebreathing valves were used to maintain a constant inhaled CO2 concentration (Figure S1B and Table ST1 in the Supplementary Material). In postprocessing, a peak detection algorithm in MATLAB (findpeaks) was used to estimate end tidal CO2 (EtCO2) from the expelled CO2 waveform. The EtCO2 time series was then visually inspected to remove obvious outliers attributed to irregular breathing or incomplete exhalation.
2.2. Diffuse Optical Instrumentation
DCS measures of blood flow were acquired using a custom-built device consisting of a 852-nm-long-coherence near-infrared laser (iBeam Smart, TOPTICA Photonics, Farmington, New York, United States), two four-channel single photon counting modules (SPCM AQ4C-IO, Perkin-Elmer, Quebec, Canada), and an 8 channel counter/timer data acquisition board (PCIe6612, National Instruments, Austin, Texas, United States) that allowed for fast (20 Hz) quantification of the intensity autocorrection function, , at time, t, and delay time, .60
Frequency-domain near-infrared spectroscopy (fdNIRS) was also employed to assess baseline tissue optical properties (namely, the wavelength-dependent absorption and scattering coefficients, and , respectively), as well as changes in over time.53 The fdNIRS system is a customized, commercially available device with 8 rapidly (20 Hz) multiplexed sources (690, 730, 750, 775, 785, 800, 825, and 830 nm) that are modulated at 110 MHz and four photomultiplier tube detectors with a gain modulation of 110 MHz + 5 kHz for heterodyne detection at 5 kHz (Imagent, ISS, Urbana Champaign, Illinois, United States). To assess baseline optical properties, measurements were made over the right and left frontal hemispheres with a rigid fdNIRS sensor containing one source and four detectors spaced at 2.0, 2.5, 3.0, and 3.5 cm.53 Resulting values were averaged to yield a global estimate of and .
Continuous DCS + fdNIRS data were acquired with a flexible sensor containing one source fiber bundle and two detector bundles spaced 1.0 and 2.5 cm from the source ( and , respectively). The source bundle contained a multimode fiber for DCS (FT1000EMT, Thorlabs, Newton, New Jersey, United States) and three multimode fibers for fdNIRS acquisition at 690, 800, and 830 nm (FT400EMT, Thorlabs). The detector bundle contained one single mode fiber for DCS (780HP, Thorlabs) and one multimode fiber for fdNIRS (FT600EMT, Thorlabs). The detector bundle contained seven single mode fibers for DCS (780HP, Thorlabs) and three multimode fibers for fdNIRS (FT600EMT, Thorlabs). Each source and detection fiber bundle was mechanically coupled to a 5-mm right-angle prism mirror (MRA05-E03, Thorlabs) using a custom 3D-printed prism/fiber holder. Prism/fiber holders were cast within a polyurethane mold (Vytaflex30, Smooth-On Inc., Macungie, Pennsylvania, United States colored black with PearlEx Carbon Black #640, Jacquard, Healdsburg, California, United States). The sensor was designed to be compliant with ANSI maximum permissible exposure standards of skin to laser radiation of at 852 nm.
2.3. DCS Data Analysis
Measured were first averaged across all seven detectors. Next, at and the averaged at were downsampled from 20 to 1 Hz to enhance the signal-to-noise ratio. These averaged and downsampled curves were then fit for an index of blood flow using two separate models. First, curves were fit to the semi-infinite homogenous solution to the correlation diffusion equation to extract an index of blood flow (BFi(t), ) at each source detector separation. For , fits were restricted to to enhance the sensitivity to cortical blood flow.49 Note that, for our experimental configuration, ranged from 1 to . We opted to use 1.2 to improve the cortical sensitivity while still ensuring sufficient data points for fitting. For all fits, we assumed a fixed index of refraction of 1.4, and we incorporated fdNIRS-estimated, subject-specific , , and , as described elsewhere,53 to avoid the confounding influence of optical properties on BFi estimation.61 When RS fdNIRS measurements were not available (), group averaged optical properties were used. For each source detector separation, data were fit simultaneously for the coherence factor () and BFi using fminsearchbnd in MATLAB with bounds for both and BFi set to [0.2, 0.55] and [1e-12, 1e-3] , respectively.51
Next, to minimize extracerebral signal contributions, we also fit measured data from and simultaneously to a three-layer model62 (assuming scalp, skull, and brain) for scalp and brain blood flow ( and , respectively) using a single cost function.63
(1) |
where is the number of detectors (2) and is the number of delay times. We minimize using fminsearchbnd in MATLAB with bounds for both CBFi and SBFi set to [1e-15, 1e-3] . For these fits, we assumed negligible (0) flow in the skull and fixed values for layer optical properties and thicknesses according to the literature (Table 1).64 To improve the stability of fits, for each separation was estimated by averaging the first five points of the autocorrelation curves.
Table 1.
Layer | () | () | L (cm) |
---|---|---|---|
Scalp | 0.10 | 15 | 0.4 |
Skull | 0.10 | 9 | 0.8 |
Brain | 0.16 | 3 | Inf |
Assumed optical properties for absorption and scattering coefficients ( and ) at 852 nm, as well as layer thicknesses () for scalp, skull, and brain.
For each experimental paradigm, the relative change in blood flow as a function of time (rCBF(t)) was calculated as for the semi-infinite model and as for the 3-layer model. Here, the subscript 0 denotes the mean flow index during the baseline of the given paradigm. For RS, the baseline was chosen as the 2 min at the start of the RS period. For BH, the baseline was chosen as the 30 s prior to the first BH epoch. For HC, the baseline was chosen as the 3-min period of room air inhalation immediately prior to CO2 initiation. The change in EtCO2 as a function of time was estimated as using the same baseline periods as rCBF. To account for the gas transit time from the mask to the capnogram and the physiological delay between alveolar diffusion of CO2 in the lungs and arterial CO2 reaching the cerebrovasculature, we time-aligned and via cross correlation. For this analysis, the EtCO2 timeseries was incrementally shifted at 0.01s intervals up to 10s. For each shift, was interpolated to the time axis, and Pearson’s correlation coefficient between and was estimated. Finally, was shifted by the time lag that yielded the highest correlation coefficient between and .
2.4. Estimation of CVR
We employed 2 numerical methods to estimate CVR (%/mmHg) for each experimental paradigm: (1) linear regression and (2) a general linear model (GLM) that regresses out systemic contributions to the DCS measured blood flow. The latter approach was included to minimize the known confounding extracerebral contributions to the DCS signal50,65 as well as the influence of heart rate and blood pressure variability.
2.4.1. Linear regression
Linear regression is commonly used to assess CVR by modeling a simple linear relationship between and :
(2) |
where reflects CVR (%/mmHg), is the intercept, and is the fit residual. For each experimental paradigm, linear regression (regress, MATLAB) was applied between the cross-correlated, time-aligned and signals to obtain an estimate of CVR. To ensure that the model explains a significant amount of variability in the data, we discarded CVR estimates for which the model p-value for was . We further discarded any CVR value .
2.4.2. General linear model
By the nature of the DCS measurement, detected light carries information about both cerebral and extracerebral (i.e., scalp and skull) hemodynamics. To minimize extracerebral contributions, we took a cue from best practices in functional near-infrared spectroscopy.66–68 We employed a GLM to regress out the relative changes in BFi at (assumed to reflect scalp hemodynamics) from the relative changes in BFi at (assumed to be sensitive to both superficial and brain layers), which is given as
(3) |
where reflects CVR (which we call , %/mmHg) and represents the relative influence that the rBFi at has on rBFi measured at .
We also explored the use of a GLM to regress out systemic contributions to the signal that are reflected in . Here the GLM took the form of
(4) |
where reflects CVR (which we call ) and represents the relative influence that the scalp perfusion () has on measured CBF index ().
CVR was estimated for each experimental paradigm using the cross-correlated, time-aligned signals by solving the system of linear equations utilizing the Moore-Penrose pseudoinverse (pinv MATLAB function). To ensure that the model explains a significant amount of variability in the data, we discarded CVR estimates when the model p-value for was . We further discarded any CVR value .
2.5. Statistical Analysis
Data are reported as median (interquartile range) unless stated otherwise. Paired Wilcoxon sign-rank tests were used to assess whether MAP, HR, and EtCO2 were statistically different from baseline for BH and HC paradigms. Wilcoxon rank-sum tests were used to assess differences in the change of these systemic parameters between HC and BH paradigms. To compare CVR across experimental paradigms, we performed Wilcoxon rank-sum tests between all combinations of paradigms for a given numerical method (e.g., linear-regression CVR compared between HC and BH), between all combinations of numerical methods for a given paradigm (e.g., HC CVR compared between linear regression and GLM methods), and between DCS analytical models for a given paradigm (e.g., linear regression HC CVR compared between semi-infinite and three-layer models). To quantify agreement of CVR across experimental paradigms, we calculate Pearson’s correlation coefficient (R) as well as Lin’s concordance correlation coefficient (CCC), which is a measure of how well a set of bivariate data compares to a “gold-standard” measurement (HC CVR, in our case).69 We define the following agreement thresholds for CCC: poor (0 to 0.5), moderate (0.5 to 0.7), strong (0.7 to 0.9), and excellent () agreements. We further visualized agreement through Bland-Altman plots, and we quantified the mean bias (95% confidence interval).70 All statistical analyses were performed in MATLAB. Statistical significance was assessed at the 0.05 confidence level.
3. Results
As expected, EtCO2 significantly increased during both BH and HC paradigms (both , Table 2). EtCO2 changes were more pronounced during HC (13.0 versus 7.5 mmHg in HC versus BH; ). During the RS paradigm, EtCO2 varied appreciably with a median range (i.e., max – min over the 10 min period) of 6.5 mmHg. MAP significantly increased by 8.0 (5.6, 11.6) mmHg during BH and 4.9 (1.9, 10.3) mmHg during HC (both ) (Table 2). Heart rate significantly increased by 11.5 (9.1, 19.2) bpm during BH (); no changes were observed during HC.
Table 2.
EtCO2 (mmHg) | MAP (mmHg) | HR (beats/minute) | |
---|---|---|---|
Delta from baseline | |||
HC | 13.0 (7.80, 19.0)+* | 4.9 (1.9, 10.3)+ | 2.1 (-3.1, 4.7)* |
BH | 7.5 (5.9, 9.2)+ | 8.0 (5.6, 11.6)+ | 11.5 (9.1, 19.2)+ |
Range: median (IQR) | |||
Resting state | 6.5 (5.3, 8.8) | 1.9 (1.4, 2.3) | 3.3 (2.9, 5.3) |
Median (interquartile range) change in , mean ABP (MAP) and heart rate (HR) for each experimental paradigm. For HC, this change reflects the steady state change from room-air inhalation, and for BH, this change reflects the peak increase during BH. For RS, the median (IQRs) of the range (max – min over the 10-min period) is reported. + denotes significant difference from baseline (); * denotes significant difference between HC and BH ().
Of the 31 available datasets, a subset of CVR estimates for each experimental paradigm was discarded due to a failure to meet quality control criteria. The exact number of datasets included for each paradigm, along with median (IQR) CVR for each experimental paradigm and numerical method, is included in Table 3. The CVR was statistically indistinguishable when compared across experimental paradigms for a given numerical method (e.g., linear regression) or for a given DCS analytical model (e.g., semi-infinite, Table 3, Fig. 2). Within a given paradigm, the use of the three-layer model increased CVR compared with the semi-infinite model. Further, the use of a GLM with short separation regression depressed the median CVR and constrained the range compared with the linear regression approach.
Table 3.
Linear regression | GLM | |||
---|---|---|---|---|
Semi-infinite | Three-layer | Semi-infinite | Three-layer | |
HC | 1.95 (0.92, 3.38)+, 29 | 3.14 (1.48, 5.11), 24 | 1.30 (0.49, 2.06)+, 26 | 3.27 (1.55, 5.48)*, 21 |
BH | 2.15 (1.24, 5.68)+, 28 | 4.86 (2.70, 8.34)*, 21 | 1.17 (0.73, 2.48)+, 19 | 2.73 (1.87, 5.03)*, 20 |
RS | 1.89 (1.36, 2.44), 20 | 3.51 (2.37, 4.33)*, 18 | 1.67 (0.98, 2.02), 18 | 3.33 (1.95, 4.40)*, 19 |
CVR (%/mmHg) for HC, BH, and RS experimental paradigms estimated with linear regression and GLMs using the semi-infinite and three-layer models for DCS analysis. Data are reported as median (IQR), N datasets (out of 31) that passed quality control metrics. * denotes significant difference between semi-infinite and three-layer for a given CVR paradigm and numerical method (). + denotes significant difference between linear regression and the GLM estimate of CVR for a given experimental paradigm ().
The relationship between CVR during HC versus BH is shown in Fig. 3 and summarized in Table 4. When using linear regression to estimate CVR, the correlation between HC and BH was only statistically significant for the three-layer model [, , Fig. 3(c)]; however, agreement was poor (). No association between HC and BH was observed with linear regression for the semi-infinite model [Fig. 3(a)]. The strength of the HC versus BH CVR relationship greatly improved when using a GLM-SS to estimate CVR [Figs. 3(e) and 3(g)]. A strong correlation and agreement were observed using the GLM on the semi-infinite data [, , Fig. 3(e)], where the mean bias from the line of equivalence of was not significant [Fig. 3(f)]. Further, a modest correlation and moderate agreement were observed using the GLM on 3-layer data with scalp flow regression [, , Fig. 3(g)], with a non-significant mean bias of [Fig. 3(h)].
Table 4.
HC versus BH | ||||
---|---|---|---|---|
Linear regression | General linear model | |||
Semi-infinite | Three-layer | Semi-infinite | Three-layer | |
N | 25 | 18 | 17 | 16 |
R | 0.15 | 0.58 | 0.80 | 0.69 |
p-value | 0.48 | 0.0087 | 0.00011 | 0.0021 |
CCC | 0.11 (−0.19, 0.39) | 0.45 (0.14, 0.68) | 0.75 (0.49, 0.88) | 0.52 (0.24, 0.72) |
Mean bias | 1.28 (0.08, 2.49) | 1.36 (−0.53, 3.25) | −0.12 (−0.57, 0.33) | 0.76 (−1.28, 2.80) |
Relationship between CVR estimated during hypercapnia versus BH using each numerical method and DCS analytical model corresponding to Fig. 3. N denotes the number of datasets that met quality inclusion criteria for both hypercapnia and BH, R denotes Pearson’s correlation coefficient and its corresponding p-value (p), CCC denotes Lin’s CCC with 95% confidence intervals, and mean bias (95% confidence interval).
We next compared CVR agreement between HC and RS paradigms (Fig. 4, Table 5). When using linear regression to estimate CVR, the correlation between HC and RS was statistically significant, but the agreement was poor for the semi-infinite model [, , and , Fig. 4(a)]; no association between RS and HC was observed for the three-layer model [Fig. 4(c)]. In a similar manner as the HC versus BH comparison, the strength of the HC versus RS CVR relationship greatly improved when using a GLM to estimate CVR [Figs. 4(e) and 4(g)]. A strong, statistically significant correlation and borderline strong agreement were observed using the GLM with short separation regression on the semi-infinite data [, , , Fig. 4(e)] with a non-significant mean bias of [Fig. 4(f)]. A significant correlation and moderate agreement were observed using the GLM on three-layer data with scalp signal regression [, , Fig. 4(g)], with a non-significant mean bias of [Fig. 4(h)].
Table 5.
Hypercapnia versus resting state | ||||
---|---|---|---|---|
Linear regression | General linear model | |||
Semi-infinite | Three-layer | Semi-infinite | Three-layer | |
N | 20 | 17 | 16 | 17 |
R | 0.45 | 0.31 | 0.76 | 0.65 |
p-value | 0.045 | 0.23 | 0.00058 | 0.0048 |
CCC | 0.35 (0.02, 0.61) | 0.21 (−0.13, 0.51) | 0.69 (0.40, 0.86) | 0.55 (0.20, 0.77) |
Mean bias | −0.46 (−1.17, 0.26) | −0.89 (−2.26, 0.47) | −0.12 (−0.50, 0.25) | −0.68 (−1.55, 0.19) |
Relationship between CVR estimated during hypercapnia versus RS using each numerical method and DCS analytical model corresponding to Fig. 4. N denotes the number of datasets that met quality inclusion criteria, R denotes Pearson’s correlation coefficient and its corresponding p-value (p), CCC denotes Lin’s CCC with 95% confidence intervals, and mean bias (95% confidence interval).
4. Discussion
Herein we used DCS to quantify the agreement between CVR measured during HC compared with more tolerable and more practical BH and RS paradigms. Although these experimental paradigms are often used interchangeably in the literature as vehicles to assess CVR, we found surprisingly poor agreement between CVR assessed with each of these paradigms when using a simple linear regression to estimate CVR. This poor agreement was observed when using both the semi-infinite and three-layer models for DCS blood flow estimation. However, by employing a GLM to regress out extracerebral contributions, we found strong agreement between the “gold-standard” HC challenge and both BH and RS. This result suggests that the more CVR tolerable paradigms may serve as a substantially equivalent substitution for the HC challenge, thereby enabling bedside assessment of microvascular CVR with DCS in patient cohorts for whom HC is contraindicated or not feasible.
The use of a GLM with short separation regression (GLM-SS) is designed to remove extracerebral and systemic contributions to the long separation blood flow signal and, by extension, to enhance sensitivity to changes in brain-layer blood flow caused by dissolved CO2 in the bloodstream. Although commonly employed by the functional near-infrared spectroscopy (fNIRS) community to improve detection of evoked brain activity,66,68,71 this approach is novel in the assessment of CVR with DCS. Furthermore, this approach offers a unique advantage over other CVR-quantifying modalities, such as TCD, that lack easy access to information regarding systemic changes without the use of supplemental monitoring of ABP, heart rate, etc. Functional magnetic resonance imaging (MRI) commonly employs a global signal regression to account for systemic influences, among other contributors;72 however, when applied to the assessment of CVR, this approach yields only qualitative estimates that complicate inter-subject comparisons and is non-ideal for use in BH or HC.73 By employing short separation regression with DCS, we regress out extracerebral and systemic contributions to the signal in a manner that is comparatively more feasible than TCD and that can provide quantitative assessment of CVR across a wider range of experimental paradigms. We note that our implementation of the GLM-SS did not employ the use of a pre-whitening filter to minimize the influence of serially correlated (i.e., colored or heteroscedastic) noise.74 This filter was not used because (1) of subjects exhibited oscillatory behavior in the temporal autocorrelation of error term, (2) prewhitening did not significantly change the estimation of CVR (e.g., for HC CVR with versus without pre-whitening, , with an average difference of ), and (3) prewhitening reduced the number of usable datasets (e.g., for HC, 23 datasets passed with pre-whitening vs. 26 without pre-whitening).
We have demonstrated a strong agreement among experimental paradigms using the GLM-SS approach; however, more work is needed to validate that CVR estimated with this approach reflects the true CVR. For reference, the literature commonly reports hypercapnic CVR in healthy adults to be to 8%/mmHg.35,55,58 Our median (range) of 1.3 (0.5 to 4%/mmHg) is comparatively lower, possibly because the systemic influences that we regress out are known to be caused by factors that change CBF in tandem with the effects of dissolved CO2 (e.g., blood pressure and cardiac output). Alternatively, the blunted values may be due to the choice of source detector separations, which may regress out too much of the desired brain-layer BFi signal due to a lack of optimal short separation or which may not have sufficient brain sensitivity due to the 2.5-cm-long separation. Finally, the blunted values could reflect that our assessments were made while the subjects were seated. Many studies, particularly with MRI, are performed with the subject in a supine posture, which is a known factor that increases the magnitude of CVR.75 Regardless, validation studies against other modalities (e.g., MRI) and clinical studies demonstrating the ability to detect abnormalities in CVR with this approach are warranted.
Although the strong agreement between HC and BH/RS with the GLM-SS approach is promising, we note that the overall rejection rate is high (23, 28, and 42% for HC, BH, and RS, respectively). We explored several explanations for this high rejection rate. For a subset of subjects (), we observed that baseline BFi () < BFi (), which may suggest a lack of brain sensitivity. For the RS paradigm, the ratio of BFi ()/BFi () was significantly lower in rejected vs. passing datasets (0.81 versus 1.40; ); however, this trend was not observed for HC or BH paradigms. Further, in a subset of 6 subjects, we briefly applied pressure to the sensor to reduce scalp flow. Although the sample size was small, we observed that the relative change in BFi at compared with during this pressure manipulation was higher but not significantly different in rejected data (2/6) across all paradigms (e.g., for RS, 0.88 versus 0.67). Taken together, these observations may suggest that the high rejection rate may be partially due to either regressing out too much brain signal with the 1 cm separation or not enough brain sensitivity with the 2.5-cm separation. The literature suggests that the latter is more likely as limited brain sensitivity is a well-known challenge in DCS measurements. Multiple ongoing, long-term efforts aim to address this brain sensitivity limitation;76 they include novel hardware developments (e.g., time-domain77,78 and interferometric DCS,79,80 along with moving to the second optical window81) and improved analytical approaches.50,62,82 As a mitigating near-term strategy, computational simulations are warranted to optimize source-detector separations, akin to what has been done with functional NIRS.66,83 For the RS paradigm, another explanation that we investigated for the high rejection rate was that the natural fluctuations in EtCO2 observed during unpaced breathing (typically a difference 1 to 3 mmHg on a breath-by-breath basis) may be too small to elicit reliable fluctuations in CBF, as has been previously suggested.26,40,45 However, median RS EtCO2 IQRs and beat-to-beat EtCO2 changes between accepted and rejected data were not statistically different (), contradicting this previously offered explanation. Future work should be directed to better understand the factors that contribute to this high rejection rate to increase clinical viability of the CVR assessment with the GLM-SS approach.
Quantifying CVR with DCS is a particularly attractive opportunity for technology transfer as a diagnostic medical device for use in clinical settings given the unique economic, ergonomic, and clinical workflow integration benefits conferred by this optical neuromonitoring modality. Notable advantages include microvascular sensitivity in contrast to TCD, bedside portability in contrast to MRI, and ease of use in all patient cohorts in contrast to both TCD and MRI. Indeed, given these positive attributes, several studies42,49–57 have utilized DCS to quantify the CBF response to a vasoactive stimulus (e.g., ACZ, HC). Nevertheless, despite the advantages of DCS as a CVR-quantifying platform, drawbacks remain. The penetration depth of DCS is limited to the superficial cortex, and the spatial sensitivity is poor, unlike MRI, and as previously discussed, the measurement can be significantly confounded by the influence of extracerebral layers (i.e., skull, scalp, and cerebrospinal fluid). With regards to the latter limitation, we explored both the GLM-SS and the use of the three-layer model. With the three-layer model, we found that the magnitude of CVR significantly increased for most experimental paradigms and numerical approaches compared with the semi-infinite model, suggesting that the model did increase brain sensitivity. However, agreement among experimental paradigms was strongest with the semi-infinite model using the GLM-SS. The weaker agreement with the 3-layer model could be due to the sensitivity of this model to errors in the model’s assumed input parameters (Supplementary Material). For example, it has been shown to be sensitive to errors in both skull and scalp thickness, such that an underestimation of scalp thickness by could induce errors in the estimation of rCBF.63 To minimize these sources of error, future studies could benefit from concurrent anatomical scans (e.g., MRI or CT) to assess layer thickness and/or from the implementation of a realistic, multilayered Monte Carlo-based fitting process coupled with a pressure modulation paradigm to optimize model parameters.50
This study is not without limitations. First, the order of the experimental paradigms was not randomized by subject. Although we have no indication that order should influence agreement, the best practice to avoid unintended systematic error in experimental design would have been to randomize the order. Second, our estimation of CVR assumes a linear relationship between blood flow and end tidal CO2. This first-order approximation may break down outside the autoregulatory range of EtCO2 ( to 45 mmHg), wherein the relationship between macrovascular blood flow velocity and EtCO2 has been shown to be logistic.84,85 Future work could benefit from quantifying the hemodynamic response function to step changes in EtCO2 to better account for possible non-linearity between CBF and EtCO2.39,55,71,83,86 Third, we did not rigorously control for environmental stimuli. Although the exact influence of the environment on our estimations of CVR is not well delineated, there is evidence to suggest that environmental stimuli can influence the brain’s response to changes in EtCO2. For example, Peng et al. showed that the time delay between the gray matter BOLD signal change and EtCO2 is larger with eyes-open versus eyes-closed states.87 The effect of environmental influences should be the subject of rigorous further investigation. Finally, there are numerous assumptions/approximations made when fitting for BFi with the semi-infinite model and for CBFi with the three-layer model (Supplementary Material). For example, with the semi-infinite model, we incorporate FDNIRS-estimated optical properties into the fitting procedure. The estimation of optical properties is predicated on several assumptions, e.g., does not change with time.51 For the three-layer model, we assumed that optical properties do not change with time, and we also used fixed, literature defined scalp and skull thicknesses for all subjects. Although the magnitude of the errors induced by each of these approximations in isolation may be small, the cumulative effect could be significant. Future work that targets this issue by minimizing the number of assumptions needed is merited.
5. Conclusions
This work provided the first systematic comparison of CVR across experimental paradigms using DCS. We demonstrated that using a GLM to regress out extracerebral and systemic contributions to the DCS-measured blood flow yielded a strong agreement between the “gold-standard” HC and the more tolerable and feasible paradigms of BH and RS. These findings suggest that BH and RS paradigms may offer feasible experimental alternatives to HC for practical clinical evaluation of CVR with DCS.
Supplementary Material
Acknowledgements
We acknowledge financial support for the following authors: EB: the National Institutes of Health (Grant Nos. R01 NS115994 and R01-HL152322) and ES: the National Institutes of Health (Grant No. 1F31HL154703-01A1). We thank Wesley Mantigou, Nguyen Nguyen, Melina Maleknia, and Michella Obialor for their auxiliary support with optical sensor instrumentation and hardware development. We also thank Hongting Zhao for helpful discussions on multilayer DCS modeling.
Biographies
Kyle R. Cowdrick received his master’s degree in bioengineering innovation and design at the Johns Hopkins University and is a founding member (preincorporation) of SpineAlign Surgical, LLC (acquired by Globus Medical). Currently, he is a doctoral candidate at Georgia Tech and Emory University. Also, he is the new partnerships liaison for Georgia Tech’s BME Capstone Design. His research interests include early-stage feasibility studies to demonstrate the translational clinical value proposition of diffuse optical spectroscopies in pediatric medicine and critical care.
Tara Urner received her bachelor’s degree in physics and astronomy with a minor in computer science from Earlham College in 2016. Currently, she is a biomedical engineering doctoral candidate at Georgia Institute of Technology and Emory University. Her research interests include expanding the clinical value of diffuse optical techniques for noninvasive, low-cost bedside monitoring of hemodynamics in brain and other tissues.
Eashani Sathialingam received her bachelor’s degree in biomedical engineering at the University of California, Irvine. She received her PhD in biomedical engineering from Georgia Institute of Technology and Emory University in 2022 and was a NIH F31 predoctoral fellow. Her PhD studies focused on developing algorithms and applying diffuse optical technologies for in vitro and in vivo studies.
Zhou Fang received his bachelor of science degree in biomedical engineering from Georgia Institute of Technology and bachelor of science in engineering science from Emory University. Currently, he is pursuing a PhD in machine learning at Georgia Institute of Technology. His research focuses on applying spatial genomics technologies and computational techniques to study stem cell heterogeneity.
Ayesha Quadri received her bachelor’s degree in neuroscience from Georgia Institute of Technology in 2021. She is a first-year medical student at Morehouse School of Medicine. After graduating from Georgia Tech, she served as a clinical research specialist for the Buckley Lab at Emory University School of Medicine.
Katherine Turrentine: Biography is not available.
Seung Yup Lee received his PhD in biomedical engineering at the University of Michigan, Ann Arbor. Currently, he is an assistant professor in the Department of Electrical and Computer Engineering at Kennesaw State University and an adjunct assistant professor in the Wallace H. Coulter Department of Biomedical Engineering at Georgia Institute of Technology and Emory University. His research interests include the development of a low-cost, miniaturized tissue optical sensing platform for clinical applications.
Erin M. Buckley is an assistant professor in the Wallace H. Coulter Department of Biomedical Engineering at Georgia Institute of Technology and Emory University and the Department of Pediatrics at Emory University. Her research interests include the development of translational diffuse optical spectroscopies for noninvasive, low-cost, bedside monitoring of hemodynamics in the brain and other tissues.
Disclosures
The authors of this manuscript report no relevant financial interests or other potential conflicts of interest to disclose.
Contributor Information
Kyle R. Cowdrick, Email: kylecowdrick@gatech.edu.
Tara Urner, Email: urner@gatech.edu.
Eashani Sathialingam, Email: eashsathi@gmail.com.
Zhou Fang, Email: zfang38@gatech.edu.
Ayesha Quadri, Email: qayesha43@gmail.com.
Katherine Turrentine, Email: kturrentine1@gmail.com.
Seung Yup Lee, Email: slee274@kennesaw.edu.
Erin M. Buckley, Email: erin.buckley@emory.edu.
References
- 1.De Vis J. B., et al. , “Calibrated MRI to evaluate cerebral hemodynamics in patients with an internal carotid artery occlusion,” J. Cereb. Blood Flow Metab. 35, 1015–1023 (2015). 10.1038/jcbfm.2015.14 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Donahue M. J., et al. , “Relationships between hypercarbic reactivity, cerebral blood flow, and arterial circulation times in patients with Moyamoya Disease,” J. Magn. Reson. Imaging JMRI 38, 1129–1139 (2013). 10.1002/jmri.24070 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Gupta A., et al. , “Cerebrovascular reserve and stroke risk in patients with carotid stenosis or occlusion: a systematic review and meta-analysis,” Stroke J. Cereb. Circ. 43, 2884–2891 (2012). 10.1161/STROKEAHA.112.663716 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Mandell Daniel M., et al. , “Mapping cerebrovascular reactivity using blood oxygen level-dependent MRI in patients with arterial steno-occlusive disease,” Stroke 39, 2021–2028 (2008). 10.1161/STROKEAHA.107.506709 [DOI] [PubMed] [Google Scholar]
- 5.Mikulis D. J., et al. , “Preoperative and postoperative mapping of cerebrovascular reactivity in moyamoya disease by using blood oxygen level—dependent magnetic resonance imaging,” J. Neurosurg. 103, 347–355 (2005). 10.3171/jns.2005.103.2.0347 [DOI] [PubMed] [Google Scholar]
- 6.Marstrand J. R., et al. , “Cerebral perfusion and cerebrovascular reactivity are reduced in white matter hyperintensities,” Stroke 33, 972–976 (2002). 10.1161/01.STR.0000012808.81667.4B [DOI] [PubMed] [Google Scholar]
- 7.Fujiwara Y., et al. , “Simultaneous impairment of intracranial and peripheral artery vasoreactivity in CADASIL patients,” Cerebrovasc. Dis. 33, 128–134 (2012). 10.1159/000334185 [DOI] [PubMed] [Google Scholar]
- 8.Geranmayeh F., et al. , “Measuring vascular reactivity with breath-holds after stroke: a method to aid interpretation of group-level BOLD signal changes in longitudinal fMRI studies,” Hum. Brain Mapp. 36, 1755–1771 (2015). 10.1002/hbm.22735 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Alexandre K., et al. , “Regional impairment of cerebrovascular reactivity and BOLD signal in adults after stroke,” Stroke 36, 1146–1152 (2005). 10.1161/01.STR.0000166178.40973.a7 [DOI] [PubMed] [Google Scholar]
- 10.Neu P., et al. , “Cerebrovascular reactivity in major depression: a pilot study,” Psychosom. Med. 66, 6–8 (2004). 10.1097/01.PSY.0000107880.03026.54 [DOI] [PubMed] [Google Scholar]
- 11.Markus H., Cullinane M., “Severely impaired cerebrovascular reactivity predicts stroke and TIA risk in patients with carotid artery stenosis and occlusion,” Brain 124, 457–467 (2001). 10.1093/brain/124.3.457 [DOI] [PubMed] [Google Scholar]
- 12.Mauro S., et al. , “Transcranial doppler assessment of cerebrovascular reactivity in symptomatic and asymptomatic severe carotid stenosis,” Stroke 27, 1970–1973 (1996). 10.1161/01.STR.27.11.1970 [DOI] [PubMed] [Google Scholar]
- 13.Webster M. W., et al. , “Compromised cerebral blood flow reactivity is a predictor of stroke in patients with symptomatic carotid artery occlusive disease,” J. Vasc. Surg. 21, 338–345 (1995). 10.1016/S0741-5214(95)70274-1 [DOI] [PubMed] [Google Scholar]
- 14.Lee J. K., et al. , “A pilot study of cerebrovascular reactivity autoregulation after pediatric cardiac arrest,” Resuscitation 85, 1387–1393 (2014). 10.1016/j.resuscitation.2014.07.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Mathieu F., et al. , “Continuous near-infrared spectroscopy monitoring in adult traumatic brain injury: a systematic review,” J. Neurosurg. Anesthesiol. 32(4), 288–299 (2019). 10.1097/ANA.0000000000000620 [DOI] [PubMed] [Google Scholar]
- 16.Zeiler F. A., et al. , “Intracranial and extracranial injury burden as drivers of impaired cerebrovascular reactivity in traumatic brain injury,” J. Neurotrauma 35, 1569–1577 (2018). 10.1089/neu.2017.5595 [DOI] [PubMed] [Google Scholar]
- 17.Zeiler F. A., et al. , “Critical thresholds of intracranial pressure-derived continuous cerebrovascular reactivity indices for outcome prediction in noncraniectomized patients with traumatic brain injury,” J. Neurotrauma 35, 1107–1115 (2017). 10.1089/neu.2017.5472 [DOI] [PubMed] [Google Scholar]
- 18.Zeiler F. A., et al. , “Transcranial doppler systolic flow index and ICP-derived cerebrovascular reactivity indices in traumatic brain injury,” J. Neurotrauma 35, 314–322 (2018). 10.1089/neu.2017.5364 [DOI] [PubMed] [Google Scholar]
- 19.Kenney K., et al. , “Cerebral vascular injury in traumatic brain injury,” Exp. Neurol. 275, 353–366 (2016). 10.1016/j.expneurol.2015.05.019 [DOI] [PubMed] [Google Scholar]
- 20.Diedler J., et al. , “The limitations of near-infrared spectroscopy to assess cerebrovascular reactivity: the role of slow frequency oscillations,” Anesth. Analg. 113, 849–857 (2011). 10.1213/ANE.0b013e3182285dc0 [DOI] [PubMed] [Google Scholar]
- 21.Lavinio A., et al. , “Cerebrovascular reactivity and autonomic drive following traumatic brain injury,” Acta Neurochir. Suppl. 102, 3–7 (2008). 10.1007/978-3-211-85578-2_1 [DOI] [PubMed] [Google Scholar]
- 22.Svaldi D. O., et al. , “Cerebrovascular reactivity alterations in asymptomatic high school football players,” Dev. Neuropsychol. 40, 80–84 (2015). 10.1080/87565641.2014.973959 [DOI] [PubMed] [Google Scholar]
- 23.Chan S., et al. , “A case study of magnetic resonance imaging of cerebrovascular reactivity: a powerful imaging marker for mild traumatic brain injury,” Brain Inj. 29, 403–407 (2015). 10.3109/02699052.2014.974209 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Fierstra J., et al. , “Measuring cerebrovascular reactivity: what stimulus to use?” J. Physiol. 591, 5809–5821 (2013). 10.1113/jphysiol.2013.259150 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Settakis G., et al. , “Acetazolamide as a vasodilatory stimulus in cerebrovascular diseases and in conditions affecting the cerebral vasculature,” Eur. J. Neurol. 10, 609–620 (2003). 10.1046/j.1468-1331.2003.00675.x [DOI] [PubMed] [Google Scholar]
- 26.Liu P., De Vis J. B., Lu H., “Cerebrovascular reactivity (CVR) MRI with CO2 challenge: a technical review,” NeuroImage 187, 104–115 (2018). 10.1016/j.neuroimage.2018.03.047 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Ziyeh S., et al. , “Blood oxygen level–dependent MRI of cerebral CO2 reactivity in severe carotid stenosis and occlusion,” Stroke 36, 751–756 (2005). 10.1161/01.STR.0000157593.03470.3d [DOI] [PubMed] [Google Scholar]
- 28.Lu H., et al. , “Alterations in cerebral metabolic rate and blood supply across the adult lifespan,” Cereb. Cortex 21, 1426–1434 (2011). 10.1093/cercor/bhq224 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Prokopiou P. C., et al. , “Modeling of dynamic cerebrovascular reactivity to spontaneous and externally induced CO2 fluctuations in the human brain using BOLD-fMRI,” NeuroImage 186, 533–548 (2019). 10.1016/j.neuroimage.2018.10.084 [DOI] [PubMed] [Google Scholar]
- 30.Lu H., et al. , “MRI mapping of cerebrovascular reactivity via gas inhalation challenges,” J. Vis. Exp. 94, e52306 (2014). 10.3791/52306 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Tancredi F. B., Hoge R. D., “Comparison of cerebral vascular reactivity measures obtained using breath-holding and Inhalation,” J. Cereb. Blood Flow Metab. 33, 1066–1074 (2013). 10.1038/jcbfm.2013.48 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Spano V. R., et al. , “ blood oxygen level–dependent MR mapping of cerebrovascular reserve in a clinical population: safety, tolerability, and technical feasibility,” Radiology 266, 592–598 (2013). 10.1148/radiol.12112795 [DOI] [PubMed] [Google Scholar]
- 33.Wise R. G., et al. , “Dynamic forcing of end-tidal carbon dioxide and oxygen applied to functional magnetic resonance imaging,” J. Cereb. Blood Flow Metab. 27, 1521–1532 (2007). 10.1038/sj.jcbfm.9600465 [DOI] [PubMed] [Google Scholar]
- 34.Yezhuvath U. S., et al. , “On the assessment of cerebrovascular reactivity using hypercapnia BOLD MRI,” NMR Biomed. 22, 779–786 (2009). 10.1002/nbm.1392 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Kastrup A., et al. , “Assessment of cerebrovascular reactivity with functional magnetic resonance imaging: comparison of and breath holding,” Magn. Reson. Imaging 19, 13–20 (2001). 10.1016/S0730-725X(01)00227-2 [DOI] [PubMed] [Google Scholar]
- 36.Moreton F. C., et al. , “Respiratory challenge MRI: practical aspects,” NeuroImage Clin. 11, 667–677 (2016). 10.1016/j.nicl.2016.05.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Bright M. G., Murphy K., “Reliable quantification of BOLD fMRI cerebrovascular reactivity despite poor breath-hold performance,” Neuroimage 83, 559–568 (2013). 10.1016/j.neuroimage.2013.07.007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Urback A. L., MacIntosh B. J., Goldstein B. I., “Cerebrovascular reactivity measured by functional magnetic resonance imaging during breath-hold challenge: a systematic review,” Neurosci. Biobehav. Rev. 79, 27–47 (2017). 10.1016/j.neubiorev.2017.05.003 [DOI] [PubMed] [Google Scholar]
- 39.Murphy K., Harris A. D., Wise R. G., “Robustly measuring vascular reactivity differences with breath-hold: normalising stimulus-evoked and resting state BOLD fMRI data,” NeuroImage 54, 369–379 (2011). 10.1016/j.neuroimage.2010.07.059 [DOI] [PubMed] [Google Scholar]
- 40.Liu P., et al. , “Cerebrovascular reactivity mapping without gas challenges,” NeuroImage 146, 320–326 (2017). 10.1016/j.neuroimage.2016.11.054 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Liu P., et al. , “A comparison of physiologic modulators of fMRI signals,” Hum. Brain Mapp. 34, 2078–2088 (2013). 10.1002/hbm.22053 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Busch D. R., et al. “Cerebral blood flow response to hypercapnia in children with obstructive sleep apnea syndrome,” Sleep 39, 209–216 (2016). 10.5665/sleep.5350 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Tung K.-C., et al. , “Alterations in resting functional connectivity due to recent motor task,” NeuroImage 78, 316–324 (2013). 10.1016/j.neuroimage.2013.04.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Wise R. G., et al. , “Resting fluctuations in arterial carbon dioxide induce significant low frequency variations in BOLD signal,” NeuroImage 21, 1652–1664 (2004). 10.1016/j.neuroimage.2003.11.025 [DOI] [PubMed] [Google Scholar]
- 45.Liu P., et al. , “Cerebrovascular reactivity mapping using intermittent breath modulation,” NeuroImage 215, 116787 (2020). 10.1016/j.neuroimage.2020.116787 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Palazzo P., et al. , “Lack of correlation between cerebral vasomotor reactivity and flow-mediated dilation in subjects without vascular disease,” Ultrasound Med. Biol. 39, 10–15 (2013). 10.1016/j.ultrasmedbio.2012.08.022 [DOI] [PubMed] [Google Scholar]
- 47.Kannurpatti S. S., Biswal B. B., “Detection and scaling of task-induced fMRI-BOLD response using resting state fluctuations,” NeuroImage 40, 1567–1574 (2008). 10.1016/j.neuroimage.2007.09.040 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Lipp I., et al. , “Wise agreement and repeatability of vascular reactivity estimates based on a breath-hold task and a resting state scan,” Neuroimage 113, 387–396 (2015). 10.1016/j.neuroimage.2015.03.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Selb J., et al. , “Sensitivity of near-infrared spectroscopy and diffuse correlation spectroscopy to brain hemodynamics: simulations and experimental findings during hypercapnia,” Neurophotonics 1, 015005 (2014). 10.1117/1.NPh.1.1.015005 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Wu M. M., et al. , “Improved accuracy of cerebral blood flow quantification in the presence of systemic physiology cross-talk using multi-layer Monte Carlo modeling,” Neurophotonics 8, 015001 (2021). 10.1117/1.NPh.8.1.015001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Buckley E. M., et al. , “Validation of diffuse correlation spectroscopic measurement of cerebral blood flow using phase-encoded velocity mapping magnetic resonance imaging,” J. Biomed. Opt. 17, 037007 (2012). 10.1117/1.JBO.17.3.037007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Durduran T., et al. , “Diffuse optics for tissue monitoring and tomography,” Rep. Prog. Phys. 73, 076701 (2010). 10.1088/0034-4885/73/7/076701 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Jain V., et al. , “Cerebral oxygen metabolism in neonates with congenital heart disease quantified by MRI and optics,” J. Cereb. Blood Flow Metab. 34, 380–388 (2014). 10.1038/jcbfm.2013.214 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Durduran T., et al. , “Optical measurement of cerebral hemodynamics and oxygen metabolism in neonates with congenital heart defects,” J. Biomed. Opt. 15, 037004 (2010). 10.1117/1.3425884 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Milej D., et al. , “Characterizing dynamic cerebral vascular reactivity using a hybrid system combining time-resolved near-infrared and diffuse correlation spectroscopy,” Biomed. Opt. Express 11, 4571 (2020). 10.1364/BOE.392113 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Zirak P., et al. , “Microvascular versus macrovascular cerebral vasomotor reactivity in patients with severe internal carotid artery stenosis or occlusion,” Acad. Radiol. 21, 168–174 (2014). 10.1016/j.acra.2013.10.010 [DOI] [PubMed] [Google Scholar]
- 57.Zirak P., et al. , “Effects of acetazolamide on the micro- and macro-vascular cerebral hemodynamics: a diffuse optical and transcranial doppler ultrasound study,” Biomed. Opt. Express 1, 1443 (2010). 10.1364/BOE.1.001443 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Zhou Y., Rodgers Z. B., Kuo A. H., “Cerebrovascular reactivity measured with arterial spin labeling and blood oxygen level dependent techniques,” Magn. Reson. Imaging 33, 566–576 (2015). 10.1016/j.mri.2015.02.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Laurin A., “BP_annotate (MATLAB Central File Exchange),” (2021).
- 60.Wang D., et al. , “Fast blood flow monitoring in deep tissues with real-time software correlators,” Biomed. Opt. Express 7, 776–797 (2016). 10.1364/BOE.7.000776 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Irwin D., et al. , “Influences of tissue absorption and scattering on diffuse correlation spectroscopy blood flow measurements,” Biomed. Opt. Express 2, 1969 (2011). 10.1364/BOE.2.001969 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Li J., et al. , “Noninvasive detection of functional brain activity with near-infrared diffusing-wave spectroscopy,” J. Biomed. Opt. 10, 044002 (2005). 10.1117/1.2007987 [DOI] [PubMed] [Google Scholar]
- 63.Zhao H., Sathialingam E., Buckley E. M., “Accuracy of diffuse correlation spectroscopy measurements of cerebral blood flow when using a three-layer analytical model,” Biomed. Opt. Express 12, 7149–7161 (2021). 10.1364/BOE.438303 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Farina A., et al. , “In-vivo multilaboratory investigation of the optical properties of the human head,” Biomed. Opt. Express 6, 2609–2623 (2015). 10.1364/BOE.6.002609 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Verdecchia K., et al. , “Assessment of a multi-layered diffuse correlation spectroscopy method for monitoring cerebral blood flow in adults,” Biomed. Opt. Express 7, 3659 (2016). 10.1364/BOE.7.003659 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Gagnon L., et al. , “Short separation channel location impacts the performance of short channel regression in NIRS,” NeuroImage 59, 2518–2528 (2012). 10.1016/j.neuroimage.2011.08.095 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Saager R. B., Telleri N. L., Berger A. J., “Two-detector corrected near infrared spectroscopy (C-NIRS) detects hemodynamic activation responses more robustly than single-detector NIRS,” NeuroImage 55, 1679–1685 (2011). 10.1016/j.neuroimage.2011.01.043 [DOI] [PubMed] [Google Scholar]
- 68.Yücel M. A., et al. , “Short separation regression improves statistical significance and better localizes the hemodynamic response obtained by near-infrared spectroscopy for tasks with differing autonomic responses,” Neurophotonics 2, 035005 (2015). 10.1117/1.NPh.2.3.035005 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Lin L. I., “A concordance correlation coefficient to evaluate reproducibility,” Biometrics 45, 255–268 (1989). 10.2307/2532051 [DOI] [PubMed] [Google Scholar]
- 70.Giavarina D., “Understanding bland Altman analysis,” Biochem. Medica 25, 141–151 (2015). 10.11613/BM.2015.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.von Lühmann A., et al. , “Using the general linear model to improve performance in fNIRS Single trial analysis and classification: a perspective,” Front. Hum. Neurosci. 14, 30 (2020). 10.3389/fnhum.2020.00030 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Liu T. T., Nalci A., Falahpour M., “The global signal in fMRI: nuisance or information?” NeuroImage 150, 213–229 (2017). 10.1016/j.neuroimage.2017.02.036 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Chen J. J., Gauthier C. J., “The role of cerebrovascular-reactivity mapping in functional MRI: calibrated fMRI and resting-state fMRI,” Front. Physiol. 12, 657362 (2021). 10.3389/fphys.2021.657362 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Huppert T. J., “Commentary on the statistical properties of noise and its implication on general linear models in functional near-infrared spectroscopy,” Neurophotonics 3, 010401 (2016). 10.1117/1.NPh.3.1.010401 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Favre M. E., et al. , “Cerebrovascular reactivity and cerebral autoregulation are improved in the supine posture compared to upright in healthy men and women,” PLOS ONE 15, e0229049 (2020). 10.1371/journal.pone.0229049 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Ayaz H., et al. , “Optical imaging and spectroscopy for the study of the human brain: status report,” Neurophotonics 9, S24001 (2022). 10.1117/1.NPh.9.S2.S24001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Sutin J., et al. , “Time-domain diffuse correlation spectroscopy,” Optica 3, 1006 (2016). 10.1364/OPTICA.3.001006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Samaei S., et al. , “Time-domain diffuse correlation spectroscopy (TD-DCS) for noninvasive, depth-dependent blood flow quantification in human tissue in vivo,” Sci. Rep. 11, 1817 (2021). 10.1038/s41598-021-81448-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79.Kholiqov O., et al. , “Time-of-flight resolved light field fluctuations reveal deep human tissue physiology,” Nat. Commun. 11, 1–15 (2020). 10.1038/s41467-019-14228-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Zhou W., et al. , “Functional interferometric diffusing wave spectroscopy of the human brain,” Sci. Adv. 7, eabe0150 (2021). 10.1126/sciadv.abe0150 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Carp S. A., et al. , “Diffuse correlation spectroscopy measurements of blood flow using 1064 nm light,” J. Biomed. Opt. 25, 097003 (2020). 10.1117/1.JBO.25.9.097003 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Baker W. B., et al. , “Modified Beer-Lambert law for blood flow,” Biomed. Opt. Express 5, 4053–4075 (2014). 10.1364/BOE.5.004053 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Gagnon L., et al. , “Further improvement in reducing superficial contamination in NIRS using double short separation measurements,” NeuroImage 85, 127–135 (2014). 10.1016/j.neuroimage.2013.01.073 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Battisti‐Charbonney A., Fisher J., Duffin J., “The cerebrovascular response to carbon dioxide in humans,” J. Physiol. 589, 3039–3048 (2011). 10.1113/jphysiol.2011.206052 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Claassen J. A. H. R., et al. , “Transcranial Doppler estimation of cerebral blood flow and cerebrovascular conductance during modified rebreathing,” J. Appl. Physiol. 102, 870–877 (2007). 10.1152/japplphysiol.00906.2006 [DOI] [PubMed] [Google Scholar]
- 86.Taylor A. J., Kim J. H., Ress D., “Characterization of the hemodynamic response function across the majority of human cerebral cortex,” NeuroImage 173, 322–331 (2018). 10.1016/j.neuroimage.2018.02.061 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Peng T., et al. , “The effects of respiratory CO2 fluctuations in the resting-state BOLD signal differ between eyes open and eyes closed,” Magn. Reson. Imaging 31, 336–345 (2013). 10.1016/j.mri.2012.06.013 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.