Abstract
Background
Cardiac MRI plays an important role in the diagnosis and follow‐up of patients with congenital heart disease (CHD). Gadolinium‐based contrast agents are often needed to overcome flow‐related and off‐resonance artifacts that can impair the quality of conventional noncontrast 3D imaging. As serial imaging is often required in CHD, the development of robust noncontrast 3D MRI techniques is desirable.
Purpose
To assess the clinical utility of noncontrast enhanced magnetization transfer and inversion recovery prepared 3D free‐breathing sequence (MTC‐BOOST) compared to conventional 3D whole heart imaging in patients with CHD.
Study type
Prospective, image quality.
Population
A total of 27 adult patients (44% female, mean age 30.9 ± 14.8 years) with CHD.
Field Strength/Sequence
A 1.5 T; free‐breathing 3D MTC‐BOOST sequence.
Assessment
MTC‐BOOST was compared to diaphragmatic navigator‐gated, noncontrast T2 prepared 3D whole‐heart imaging sequence (T2prep‐3DWH) for comparison of vessel dimensions, lumen‐to‐myocardium contrast ratio (CR), and image quality (vessel wall sharpness and presence and type of artifacts) assessed by two experienced cardiologists on a 5‐point scale.
Statistical Tests
Mann–Whitney test, paired Wilcoxon signed‐rank test, Bland–Altman plots. P < 0.05 was considered statistically significant.
Results
MTC‐BOOST significantly improved image quality and CR of the right‐sided pulmonary veins (PV): (CR: right upper PV 1.06 ± 0.50 vs. 0.58 ± 0.74; right lower PV 1.32 ± 0.38 vs. 0.81 ± 0.73) compared to conventional T2prep‐3DWH imaging where the PVs were not visualized in some cases due to off‐resonance effects. MTC‐BOOST demonstrated resistance to degradation of luminal signal (assessed by CR) secondary to accelerated or turbulent flow conditions. T2prep‐3DWH had higher image quality scores than MTC‐BOOST for the aorta and coronary arteries; however, great vessel dimensions derived from MTC‐BOOST showed excellent agreement with standard T2prep‐3DWH imaging.
Data Conclusion
MTC‐BOOST allows for improved contrast‐free imaging of pulmonary veins and regions characterized by accelerated or turbulent blood flow compared to standard T2prep‐3DWH imaging, with excellent agreement of great vessel dimensions.
Evidence Level
1
Technical Efficacy
Stage 2
Keywords: magnetization Transfer, MTC‐BOOST, 3D whole heart imaging, congenital heart disease
Congenital heart disease (CHD) affects approximately 8–10:1000 live births globally with approximately 90% of this group surviving into adulthood. 1 , 2 With the adult CHD population continuing to grow, life‐long follow‐up and surveillance are often required to assess for potential late‐stage complications. Due to the heterogeneous and complex anatomy of patients with CHD, cardiac MRI plays a fundamental role in diagnosis, preprocedural planning, and follow‐up. 3 , 4 , 5
Whole heart imaging using ECG‐triggered noncontrast enhanced (non‐CE) T2 prepared (T2prep) balanced steady‐state free precession (bSSFP) (T2prep‐3DWH) is commonly employed for anatomical assessment in CHD. 6 Limitations of this approach include heterogeneous blood pool contrast in regions of turbulent flow and suboptimal visualization of pulmonary and coronary venous vasculature, in part due to off‐resonance effects. 7 , 8 This is an important consideration for CHD patients where valvular and vessel abnormalities are prevalent. 1 Contrast‐enhanced MR angiography (CE‐MRA) with intravenous gadolinium‐based contrast agents can be used to improve imaging of vascular structures affected by luminal signal loss. 3 , 9 , 10 , 11 However, CE‐MRA is performed without the use of cardiac gating, which can impede the accurate assessment of vessel dimensions due to the presence of motion‐related artifacts that are most evident at the aortic root and coronary arteries. 12 , 13 Furthermore, as administration of gadolinium‐based contrast agents has known risks, such as nephrogenic systemic fibrosis in patients with significant renal dysfunction, and undetermined risks related to potential gadolinium retention, noncontrast imaging approaches are desirable. 14 Black‐blood cardiac MRI is commonly employed and allows for noncontrast visualization of myocardium and vessel wall; however, the conventional approach using double inversion recovery is limited to 2D acquisitions with nonisotropic resolution. A recent study demonstrated the feasibility of 3D black‐blood fast spin echo techniques for whole heart CHD imaging; however, this technique failed to consistently visualize the coronary arteries, representing an important limitation of this approach. 15
This proof‐of‐concept study aims to investigate the potential clinical utility of a recently developed noncontrast magnetization transfer (MTC) and inversion recovery (IR) prepared sequence, “MTC‐BOOST,” that simultaneously generates co‐registered 3D bright‐ and black‐blood whole heart images, for anatomical assessment in patients with CHD.
Materials and Methods
The study was approved by the National Research Ethics Service (1/11/12 and 15/NS/0030) and written informed consent was obtained from each participant according to institutional guidelines. Twenty‐seven adult patients (58% male, mean age 30.9 ± 14.8 years) with a spectrum of CHD who were scheduled for a clinically indicated cardiac MRI were recruited to undergo additional imaging with MTC‐BOOST from 2018 to 2019.
MTC‐BOOST Sequence
The data were acquired with an ECG‐triggered bSSFP 3D whole‐heart MTC‐BOOST imaging sequence, 16 schematically shown in Fig. 1. MTC preparation in combination with an IR pulse is used in the odd heartbeats, whereas in the even heartbeats, only MTC preparation is performed. The bright blood MTC‐IR prepared acquisition (odd heartbeats) allows bright‐blood visualization of the heart anatomy and thoracic vascular structures including pulmonary arteries and veins. A phase sensitive inversion recovery (PSIR)‐like reconstruction enables the generation of a fully co‐registered 3D black blood dataset. 17 In odd heartbeats, a short TI IR approach 18 is used to suppress signal from epicardial fat (Fig. 1a), whereas frequency‐selective presaturation 19 is used in even heartbeats for fat suppression (Fig. 1b). Data acquisition is performed using a 3D Cartesian trajectory with spiral‐like order 20 and segmented over multiple heartbeats. For respiratory motion correction, low‐resolution 2D image navigators (iNAVs) are acquired in each heartbeat by spatially encoding ramp‐up pulses of the bSSFP sequences enabling beat‐to‐beat translational respiratory motion estimation and correction without data rejection (100% respiratory scan efficiency). 21
Motion Correction and Image Reconstruction
The MTC‐BOOST data were reconstructed offline using MATLAB (The MathWorks, Natick, MA). The MTC‐IR and MTC prepared volumes were independently reconstructed in end‐expiration using a previously described nonrigid motion correction framework. 22 Briefly, right–left (RL) and superior–inferior (SI) displacements were estimated from the iNAVs, SI estimation was used to group the data in four to six equally populated bins, whereas both RL and SI estimations were used to correct for intrabin translational motion. 23 Each bin was reconstructed with soft‐gated iterative SENSE. The bin corresponding to end‐expiration was then used as the reference bin to estimate bin‐to‐bin 3D nonrigid deformation fields via image registration. 24 Successively the 3D nonrigid motion fields are incorporated in a generalized matrix description reconstruction pipeline where the motion‐corrected 3D volume is reconstructed using a linear conjugate gradient method. 25 The motion‐corrected MTC‐IR and MTC prepared volumes were then combined in a PSIR‐like reconstruction to obtain a co‐registered black‐blood volume.
Data Acquisition
Consecutive patients with CHD (n = 27) were scheduled for a clinically indicated cardiac MRI examination. Patients with intracardiac devices were excluded. Images were acquired on a 1.5 T MRI system (Magnetom Aera, Siemens Healthcare, Erlangen, Germany).
Data were acquired with an 18‐channel chest coil and a 32‐channel spine coil. The MTC‐BOOST imaging sequence was acquired precontrast administration with the following imaging parameters: coronal orientation, fully sampled with acquired spatial resolution of 1.4 × 1.4 × 2.8 mm, reconstructed isotropic resolution of 1.4 mm3, subject‐specific FOV = 320 × 320 × 90–120 mm3, TE = 1.5 msec, TR = 3.2 msec, flip angle for both odd and even heartbeats 90°, MT preparation: 15 Gaussian pulses, flip‐angle 800°, frequency offset 3000 Hz, duration 20.5 msec. In this study, a high‐frequency offset (3000 Hz) was chosen for MTC preparation to minimize image artifacts in regions with imperfect B0 and to minimize on‐resonance frequencies. Cine imaging was acquired in transverse orientation to determine the subject specific mid‐diastolic resting period (cardiac acquisition window ~120 msec, corresponding to 28–35 k‐space lines acquired per heartbeat). The established T2prep‐3DWH imaging was acquired using the following imaging parameters: spatial resolution 1.4 × 1.4 × 1.4 mm, T2prep of 40 msec, acceleration GRAPPA 2×, flip angle 90°. The T2prep‐3DWH sequence was combined with a conventional diaphragmatic (dNAV) navigator for respiratory motion compensation, with a gating window of ±3.5 mm in end‐expiration. Acquisition window was manually set to ~120 msec, depending on the subject‐specific mid‐diastolic resting period.
Image Analysis
Image analysis was performed using Osirix v12.0 (Pixmeo, Geneva, Switzerland). The visualization of cardiac and vascular anatomy in CHD patients using bright‐blood MTC‐BOOST was compared to standard bright‐blood T2prep‐3DWH imaging. To this end, the lumen‐to‐myocardium contrast ratio (CR) was assessed for both the MTC‐BOOST bright blood and T2prep‐3DWH sequences.
CR measurements were taken in the following anatomical structures: right and left upper pulmonary vein (RUPV and LUPV), right and left lower pulmonary vein (RLPV and LLPV), left ventricle (LV), ascending aorta (Asc Ao), inferior vena cava (IVC), superior vena cava (SVC), right atrium (RA), right ventricle (RV), main pulmonary artery (MPA), left anterior descending coronary artery (LAD), left circumflex coronary artery (LCx) and the right coronary artery (RCA).
Image quality of the bright‐blood images of the MTC‐BOOST and T2prep‐3DWH imaging for each patient were assessed by two cardiologists (A.Fr. and I.V., Society for Cardiovascular Magnetic Resonance level III training). Each reader was blinded to the acquisition type they were grading and specific predefined anatomical structures were assessed: Asc Ao, MPA, right and left pulmonary veins (RPV and LPV), left main stem coronary artery (LM) and RCA. The image quality assessment was based on a 5‐point scoring system and was divided on sharpness of vessel borders (1 = nondiagnostic, 2 = poor, 3 = adequate, 4 = good, 5 = excellent) and presence of residual artifacts (1 = severe, 2 = moderate–severe, 3 = moderate, 4 = mild, and 5 = minimal or no artifact, where grades 1 and 2 represent nondiagnostic image quality). Overall diagnostic confidence of the MTC‐BOOST dataset was compared to the established T2prep‐3DWH. Diagnostic information was related to: 1) exclusion of anomalies, 2) suspected abnormalities, and 3) unsuspected abnormalities relevant to the patient's assessment regarding CHD. Level of confidence (1 = low, 2 = moderate, 3 = high, 4 = excellent).
Additionally, vessel dimensions were derived from both the bright‐blood and black‐blood MTC‐BOOST datasets and compared to the established T2prep‐3DWH. Vessel dimensions were measured using bright blood MTC‐BOOST, T2prep‐3DWH, and black blood MTC‐BOOST datasets at the level of the Asc Ao and proximal descending aorta (DAo), MPA, left (LPA) and right (RPA) branch pulmonary arteries. Blinded co‐axial measurements (max diameter) were performed using multiplanar reformats by a cardiologist (A.Fo, European Association of Cardiovascular Imaging level II training). For interobserver comparison, an additional cardiologist (I.R., Society for Cardiovascular Magnetic Resonance level III training) assessed aortic and pulmonary artery vessel dimensions derived from bright blood MTC‐BOOST imaging.
Statistical Analysis
Statistical analysis was performed using IBM SPSS Statistics for Windows version 28.0.1.1 (IBM Corp, Armonk, N.Y., USA). Intraclass correlation coefficients and Bland–Altman plots were used to assess agreement of quantitative measurements of the previously defined anatomical structures using both MTC‐BOOST and T2prep‐3DWH imaging techniques. Image quality scores for vessel sharpness, presence of artifact and diagnostic confidence were assessed and compared using a paired Wilcoxon signed‐rank test. The total acquisition time of MTC‐BOOST and T2prep‐3DWH was compared using a Mann–Whitney test. Bland Altman plots were used to assess intersequence and interobserver agreement of vessel dimensions. A P‐value of <0.05 was considered statistically significant.
Results
Patient Characteristics and Cardiac MRI Study
The patient characteristics are described in Table 1. Figure 2 shows the co‐registered bright‐ and black‐blood images acquired with the MTC‐BOOST imaging technique for three representative CHD patients with: subclavian flap repair of aortic coarctation (subject 1), dextrocardia situs inversus (subject 2) and repaired Tetralogy of Fallot (rTOF, subject 3), respectively. Supplementary Figure S1 depicts images from the MTC‐BOOST (bright and black blood datasets) vs. the conventional sequence in five representative views of the heart. The average scan time for MTC‐BOOST, that utilizes iNAV‐based respiratory motion compensation and generates both dark‐ and bright‐blood images, was lower than the standard T2prep‐3DWH that employs a diaphragmatic navigator to yield bright‐blood images only (9 ± 1.1 minutes vs. 11.1 ± 2.6 minutes). However, the slice thickness of MTC‐BOOST was double T2prep‐3DWH, and T2prep‐3DWH was undersampled (GRAPPA 2x) compared to fully sampled MTC‐BOOST, limiting direct comparisons of acquisition time.
TABLE 1.
Patient Characteristics | |
---|---|
Aortopulmonary window repair | 1 |
Anomalous pulmonary venous drainage | 4 |
Pulmonary stenosis, pulmonary valve atresia, tetralogy of fallot | 5 |
Single ventricle Fontan | 4 |
Aortic coarctation | 5 |
Bicuspid aortic valve | 5 |
PDA/ASD repair | 1 |
Situs inversus dextrocardia | 1 |
Left‐sided superior vena cava | 1 |
Number of males (total number of patients) | 15 (27) |
Mean Age ± SD (year) | 31 ± 15 |
PDA = patent ductus arteriosus; ASD = atrial septal defect; LSVC = left‐sided superior vena cava; SD = standard deviation.
MTC‐BOOST Improved Pulmonary Vein Visualization Compared to T2prep‐3DWH Imaging
MTC‐BOOST yielded marked improvements in pulmonary vein (PV) imaging compared to T2prep‐3DWH imaging sequences, in which luminal PV signal was reduced and, in some cases, could not be visualized. This was observed in a wide variety of anatomical variants with and without PV pathology (Fig. 3). This was clearly demonstrated in a patient with prior surgical correction of anomalous pulmonary venous drainage, where individual PVs were either not or only partially visualized adjacent to their left atrial insertion points using conventional T2prep‐3DWH imaging (Fig. 3a). In contrast, MTC‐BOOST in the same patient resulted in clear delineation of the PVs with uniform luminal signal (Fig. 3b). Similarly, in a patient with Scimitar syndrome, the left PV insertion points to the left atrium (LA) were not seen with standard T2prep‐3DWH imaging (Fig. 3c) but were clearly visualized in MTC‐BOOST (Fig. 3d). In the same patient, the anomalous right‐sided PV was detected in both T2prep‐3D whole heart and MTC‐BOOST acquisitions (Fig. 3e,f). Improved PV imaging with MTC‐BOOST was also clearly observed in a patient with pulmonary artery hypoplasia, where standard T2prep‐3DWH imaging showed bilateral PV luminal signal loss that was mitigated by MTC‐BOOST (Fig. 3g,h).
MTC‐BOOST Conferred Resistance to Flow‐Induced Artifacts Compared to T2prep‐3DWH Imaging
Many patients with CHD have vascular or valvular abnormalities that can result in accelerated or turbulent blood flow that often results in luminal signal loss on standard T2prep‐3DWH imaging due to dephasing of the MRI signal. Examples of such luminal signal loss were observed following T2prep‐3DWH imaging in patients with hypoplastic PAs associated with accelerated blood flow (Fig. 4a,c). In contrast, corresponding MTC‐BOOST images (Fig. 4b,d) yielded a higher luminal CR compared to T2prep‐3DWH (CR: 1.3 vs. 0.8, and 1.9 vs. 0.8 for subjects 5 and 7, respectively). MTC‐BOOST also provided a reduction in luminal signal loss compared to T2prep‐3DWH due to turbulent flow from stenotic valves as demonstrated in a patient with recurrent aortic stenosis following aortic valve replacement (CR: 1.4 vs. 0.9; Fig. 4e,f).
Caval Imaging in Fontan Patients Using MTC‐BOOST
Comparison of MTC‐BOOST and conventional T2prep‐3DWH imaging was also performed in single ventricle Fontan patients (Fig. 5). Here, MTC‐BOOST showed good visualization of the superior and inferior vena cava, which are characterized by low‐velocity venous blood flow, to the branch PAs in patients who have undergone Fontan palliation (Fig. 5b,d). The luminal caval signal was similar to the standard T2prep‐3DWH imaging (Fig. 5a,c) performed in the same patients.
Visualization of Stents
Neither bright‐ nor black‐blood MTC‐BOOST offers an advantage in comparison to the established T2prep‐3DWH imaging for artifact‐free intraluminal stent visualization. Artifact from a branch PA stent had a similar appearance in both MTC‐BOOST and T2prep‐3DWH acquisitions (CR: 0.53 vs. 0.58, respectively), (Fig. 5 red arrow, Supplementary Fig. S2).
Assessment of Vessel Dimensions by MTC‐BOOST Compared to T2prep‐3DWH
Aortic dimensions demonstrate excellent agreement between bright blood MTC‐BOOST and T2prep‐3DWH sequences for the assessment of vessel dimensions at different anatomical regions (Fig. 6) including the mid ascending aorta (mean difference: 0.074 mm, 95% confidence interval [CI]: −1.5 to 1.6), the descending thoracic aorta (mean difference: 0.2 mm, 95% CI:−1.4 to 1.8), the main pulmonary artery (mean difference: −0.09 mm, 95% CI: −2.0 to 1.8) and branch PAs. Interobserver comparison of great vessel dimensions (A.Fo. and I.R.) showed excellent agreement between readers for the ascending aorta (mean difference: 0.1 mm, 95% CI: −1.8 to 2) and main pulmonary artery (mean difference: −0.2 mm, 95% CI: −2.2 to 1.8) (Supplementary Fig. S3). Additionally, comparison of vessel dimensions derived from black blood MTC‐BOOST imaging and the T2prep‐3DWH using Bland–Altman analysis demonstrated excellent agreement between both sequences across all measured vessel diameters (Supplementary Fig. S4).
Luminal to Myocardial CR of MTC‐BOOST Compared to Standard T2prep‐3DWH Imaging
The lumen‐to‐myocardium CR results of the great vessels and cardiac chambers demonstrate significantly higher right upper (1.06 ± 0.50 vs. 0.58 ± 0.74) and right lower (1.32 ± 0.38 vs. 0.81 ± 0.73) pulmonary vein contrast ratios for MTC‐BOOST compared to T2prep‐3DWH imaging, respectively. The T2prep‐3DWH sequence yielded lumen to myocardium contrast ratios lower than 1, consistent with a marked degradation of luminal signal (Supplementary Fig. S5). There was no significant difference in the CR of the left upper and left lower PV between the two sequences (P = 0.09 and P = 0.08, respectively). The contrast ratios for the aorta, inferior and superior vena cava, cardiac chambers and left anterior descending coronary artery were higher using the T2prep‐3DWH sequence.
MTC‐BOOST Image Quality Assessment
MTC‐BOOST resulted in significantly superior image quality compared to T2prep‐3DWH with regard to both vessel sharpness (Table 2) and the presence of artifact (Table 3) for the right pulmonary veins (vessel sharpness, mean ± SD, reviewer 1: 4 ± 0.6 vs. 3.3 ± 0.8, reviewer 2: 4.5 ± 0.9 vs. 1.4 ± 0.7; presence of artifact, reviewer 1: 4 ± 0.7 vs. 3.3 ± 0.8, reviewer 2: 3.3 ± 0.8 vs. 1.6 ± 0.6) and left pulmonary veins (vessel sharpness, reviewer 1: 4.1 ± 0.6 vs. 3.1 ± 0.6, reviewer 2: 4.4 ± 1 vs. 1.7 ± 0.8; presence of artifact, reviewer 1: 4.1 ± 0.6 vs. 3.1 ± 0.6, reviewer 2: 4 ± 1 vs. 1.7 ± 0.7). There was no significant difference in vessel sharpness (P = 0.6 and 0.5) or presence of artifact (P = 0.9 and 0.7) as assessed by image quality scores for the main pulmonary artery for either reader (Tables 2 and 3). Image quality scores for the coronary arteries were higher using the conventional T2prep‐3DWH imaging compared to MTC‐BOOST (Tables 2 and 3); however, mean quality scores for MTC‐BOOST ranged from 2.7 to 3.2 for vessel sharpness and 2.7 to 4 for the presence of artifact, indicating that the majority of MTC‐BOOST acquisitions achieved diagnostic image quality. The diagnostic confidence following analysis of MTC‐BOOST bright‐ and black‐blood datasets was compared to T2prep‐3DWH and showed significant superiority of the proposed sequence with a mean quality score of 3.7 ± 0.5 for both the bright‐ and black‐blood datasets versus 3.2 ± 0.8 for the T2prep‐3DWH sequence. In 7 out of the 27 datasets, the T2prep‐3DWH sequence had a lower quality score because of poorer delineation of the pulmonary veins and in six additional datasets, this was combined with artifacts in the pulmonary arteries. In three datasets, the bright‐ and black‐blood MTC‐BOOST sequences had a lower score due to poorer delineation of the coronary arteries.
TABLE 2.
Vessel Sharpness | Clinical (T2 prepared 3D whole heart) | MTC‐BOOST | P value |
---|---|---|---|
LM (R1) | 3.9 ± 1.0 | 3.2 ± 0.7 | 0.07 |
LM (R2) | 3.9 ± 1.5 | 3.0 ± 1.6 | <0.05 |
RCA (R1) | 3.7 ± 1.3 | 2.7 ± 1.1 | <0.05 |
RCA (R2) | 3.8 ± 1.4 | 2.8 ± 1.4 | <0.05 |
Aorta (R1) | 4.6 ± 0.6 | 4.2 ± 0.7 | <0.05 |
Aorta (R2) | 4.6 ± 0.8 | 4.3 ± 0.9 | 0.1 |
MPA (R1) | 4.3 ± 0.9 | 4.2 ± 0.8 | 0.6 |
MPA (R2) | 3.8 ± 1.4 | 4.0 ± 1.1 | 0.5 |
RPV (R1) | 3.3 ± 0.8 | 4.0 ± 0.6 | <0.05 |
RPV (R2) | 1.4 ± 0.7 | 4.5 ± 0.9 | <0.05 |
LPV (R1) | 3.1 ± 0.6 | 4.1 ± 0.6 | <0.05 |
LPV (R2) | 1.7 ± 0.8 | 4.4 ± 1.0 | <0.05 |
Image quality scores—vessel wall sharpness. LM = left main coronary artery; RCA = right coronary artery; MPA = main pulmonary artery; RPV = right‐sided pulmonary veins; LPV = left‐sided pulmonary veins. Two blinded reviewers (R1, R2) assessed the clinical sequence (T2 prepared 3D whole heart) and MTC‐BOOST bright blood datasets to grade the sharpness of vessel borders (1 = nondiagnostic, 2 = poor, 3 = adequate, 4 = good, 5 = excellent). For each observer, the scores were compared using a paired Wilcoxon signed‐rank test, with statistical significance difference set to P < 0.05.
TABLE 3.
Presence of Artifact | Clinical | MTC‐BOOST | P value |
---|---|---|---|
LM (R1) | 4.1 ± 1.0 | 3.3 ± 0.7 | <0.05 |
LM (R2) | 3.7 ± 1.5 | 4.0 ± 1.0 | 0.5 |
RCA (R1) | 4 ± 1.2 | 2.9 ± 1.1 | <0.05 |
RCA (R2) | 3.9 ± 1.4 | 2.7 ± 1.3 | <0.05 |
Aorta (R1) | 4.7 ± 0.6 | 4.2 ± 0.5 | <0.05 |
Aorta (R2) | 4.3 ± 1.0 | 3.7 ± 0.7 | <0.05 |
MPA (R1) | 4.2 ± 0.9 | 4.2 ± 0.8 | 0.9 |
MPA (R2) | 3.8 ± −1.4 | 3.8 ± 0.8 | 0.7 |
RPV (R1) | 3.3 ± 0.8 | 4.0 ± 0.7 | <0.05 |
RPV (R2) | 1.6 ± 0.6 | 3.3 ± 0.8 | <0.05 |
LPV (R1) | 3.1 ± 0.6 | 4.1 ± 0.6 | <0.05 |
LPV (R2) | 1.7 ± 0.7 | 4.0 ± 1.0 | <0.05 |
Image quality scores—presence of artifact. LM = left main coronary artery; RCA = right coronary artery; MPA = main pulmonary artery; RPV = right‐sided pulmonary veins; LPV = left‐sided pulmonary veins. Two blinded reviewers (R1, R2) assessed the clinical sequence (T2 prepared 3D whole heart) and MTC‐BOOST bright blood datasets to grade presence of artifact (1 = severe artifact, 2 = significant artifact, 3 = moderate artifact, 4 = mild artifact, and 5 = minimal artifact) where grades 1 and 2 delineate nondiagnostic datasets. For each observer, the scores were compared using a paired Wilcoxon signed‐rank test, with statistical significance difference set to P < 0.05.
Discussion
In this study, we demonstrated the feasibility of free‐breathing 3D whole heart MT‐IR prepared BOOST (MTC‐BOOST) for contrast‐free imaging of CHD patients. MTC‐BOOST simultaneously yields a bright‐blood and a complementary co‐registered black‐blood 3D whole‐heart volume. This proof‐of‐concept study compared MTC‐BOOST to the established bright‐blood T2prep‐3DWH imaging technique for the visualization of cardiac and vascular anatomy in CHD patients, including patients with obstructive right and left heart disease, single ventricle pathology and anomalous pulmonary venous drainage. This broad case mix allowed for an assessment of the performance of MTC‐BOOST compared to the conventional T2prep‐3DWH imaging sequence in an anatomically diverse population. The results demonstrate excellent agreement between the two sequences in terms of vessel dimensions with improved visualization in MTC‐BOOST of the pulmonary venous vasculature and regions of accelerated or turbulent flow, which is likely conferred by resistance to off‐resonance artifacts. In addition, a wide spectrum of CHD patients was represented in the current study, including complex CHD, single ventricle palliation, pulmonary venous abnormalities, valvular and great artery disease, supporting the applicability of MTC‐BOOST to the broader CHD population.
Conventional T2prep‐3DWH sequences often demonstrated severe artifacts in the PV and occasionally LA. The reduction in luminal signal observed in PVs with conventional T2prep‐3DWH imaging is likely mediated by off‐resonance effects as previously described by Hu et al, who demonstrated that PV blood off‐resonance effects were responsible for signal void artifacts in the PV vasculature when imaged with 3D noncontrast enhanced T2 prepared SSFP sequences. 7 The same study also demonstrated that PV blood flow, in addition to off‐resonance, can cause signal void artifacts in the LA. In contrast, the MT‐IR bright blood component of the MTC‐BOOST sequence yielded clear and consistent delineation of PVs associated with significantly higher image quality scores for the right PVs with a corresponding increase in luminal‐to‐myocardial CR. The contrast‐to‐noise ratio was not used for comparison due to different reconstruction methods employed leading to significant variation in background noise between datasets. The exact cause of the higher susceptibility of the right‐sided pulmonary veins to off‐resonance and luminal signal reduction using conventional T2prep‐3DWH sequences remains unclear. It may be related to their proximity to the lung parenchyma or to the use of pencil‐beam navigators placed over the right hemidiaphragm. Furthermore, the higher and more uniform luminal signal in regions of accelerated or turbulent blood flow observed with MTC‐BOOST support the resistance of MT preparation to flow‐mediated off‐resonance effects compared to standard T2 prepared acquisitions. For other structures including the aorta, inferior and superior vena cava, cardiac chambers and left anterior descending coronary artery, T2prep‐3DWH yielded higher lumen to myocardium contrast ratios. However, these structures were clearly visualized using both sequences and the observed differences in CR for the great vessels did not significantly impact image quality.
The improved noncontrast imaging of regions of accelerated or turbulent blood flow and the pulmonary venous vasculature observed with MTC‐BOOST has multiple potential applications in CHD patients. CHD is characterized by regions of turbulent blood flow (eg, hypoplastic vessels or coarctation, vessel anastomoses, and stenotic or regurgitant valves) and MTC‐BOOST may allow for a detailed anatomical assessment in these regions without the need for contrast administration. Furthermore, improved imaging of the pulmonary venous vasculature has several potential advantages in CHD, including patients with Fontan circuits, where elevated central venous pressures can promote the development of venovenous collateral vessels. As the pulmonary veins are a common site of terminal drainage of venous collateral vessels with potential for right‐to‐left shunt, clear imaging of the pulmonary vasculature is needed to accurately identify these shunts, which can be a target for intervention. The clear delineation of the pulmonary veins also allows for the assessment of partial anomalous pulmonary venous drainage (PAPVD) and pulmonary venous stenosis, which is limited using standard noncontrast‐enhanced T2prep‐3DWH sequences. The identification of left‐to‐right shunt on flow quantification studies often necessitates exclusion of anomalous pulmonary venous drainage and the proposed sequence is able to achieve this task without the administration of a contrast agent. The reduction in gadolinium utilization could lower costs and the risk of contrast‐related complications. Additionally, the higher diagnostic confidence associated with MTC‐BOOST may reduce the need for additional imaging techniques such as CT.
The MTC‐BOOST framework, as previously demonstrated, 16 was integrated with image‐based navigation. This allowed for data acquisition with nearly 100% respiratory scan efficiency and with predictable acquisition time. The use of a diaphragmatic navigator (dNAV) gating to account for respiratory motion introduces problems, such as prolonged and unpredictable scan times due to the fact that data acquired outside a pre‐defined respiratory window (usually end‐expiration) is rejected and needs to be reacquired. Furthermore, dNAV provides only an indirect estimation of the respiratory motion of the heart, as it tracks the SI displacement of the dome of the right hemidiaphragm. In addition, the intersection between the dNAV and the PVs may cause inflow artifacts and lead to reduced sharpness of such vessels. 26 , 27 MTC‐BOOST relies on the acquisition of two differently weighted bright‐blood volumes (over two heart beats) for two reasons. First, to allow recovery from the IR prepulse and improve visualization of the vessels in the bright‐blood volume and second, to provide a complementary co‐registered black‐blood PSIR volume without time penalty. This study has demonstrated excellent agreement in vascular and cardiac measurements derived from both the bright‐ and black‐blood MTC‐BOOST datasets with T2prep‐3DWH based dimensions. No particular advantage was observed when measuring vessel dimensions using black‐blood MTC‐BOOST images compared to the bright‐blood dataset. Unlike 2D spin echo black‐blood imaging, the 3D black‐blood MTC‐BOOST dataset does not allow for in‐stent luminal assessment. 28 However, there is supporting evidence that PSIR carries important diagnostic information for thrombus visualization that is relevant for Fontan and baffle (eg Mustard procedure) assessment. 29 Future studies should further investigate the clinical value of the black‐blood MTC‐BOOST volume in the CHD population in comparison to conventional 2D spin echo black‐blood and recently proposed 3D black‐blood approaches. 29
Limitations
First, we compared noncontrast‐enhanced MTC‐BOOST against standard noncontrast T2prep‐3DWH but not against CE‐MRA. Future studies comparing MTC‐BOOST and CE‐MRA are required to assess whether the need for contrast administration can be avoided. The number of subjects in the current study was limited and a larger study would allow for inclusion of a broader representation of CHD patients. Finally, MTC‐BOOST was acquired using anisotropic resolution, which is suboptimal for 3D multiplanar reformat assessment and may explain the lower image quality scores for coronary imaging with MTC‐BOOST in comparison with conventional (isotropic) T2 prepared 3D imaging. Forthcoming studies might benefit from an isotropic acquisition with higher spatial resolution combined with image acceleration techniques.
Conclusion
MTC‐BOOST allows for single acquisition noncontrast assessment of cardiac, arterial, and venous anatomy in adult patients with CHD. It provides clear delineation of the pulmonary veins and regions of accelerated blood flow, which was not possible using conventional T2prep‐3DWH imaging techniques that are susceptible to signal voids due to turbulent flow and off‐resonance effects. MTC‐BOOST had lower image quality scores for the coronary arteries and aorta but yielded excellent agreement for great vessel dimensions. For the same spatial resolution, scan times for fully sampled MTC‐BOOST may be longer than T2prep3DWH (undersampled by factor of 2) and assessment of intravascular stent patency was not possible with either of these noncontrast sequences. Overall, as MTC‐BOOST improves visualization of the pulmonary veins and provides accurate great vessel dimensions compared to established clinical sequences, it holds promise for the assessment of CHD.
Supporting information
Imran Rashid and Giulia Ginami contributed equally to this work.
This research was supported by the National Institute for Health Research (NIHR), Cardiovascular Health Technology Cooperative (HTC), and the Biomedical Research Centre based at Guy's and St. Thomas' NHS Foundation Trust and King's College London.
References
- 1. Liu Y, Chen S, Zühlke L, et al. Global birth prevalence of congenital heart defects 1970–2017: Updated systematic review and meta‐analysis of 260 studies. Int J Epidemiol 2019;48(2):455‐463. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Buber J, Valente AM. Predicting survival in adults with congenital heart disease: What are the odds? Heart (British Cardiac Society) 2018;104(20):1643‐1644. [DOI] [PubMed] [Google Scholar]
- 3. Fratz S, Chung T, Greil GF, et al. Guidelines and protocols for cardiovascular magnetic resonance in children and adults with congenital heart disease: SCMR expert consensus group on congenital heart disease. J Cardiovasc Magn Reson 2013;15(1):51. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Greil G, Tandon AA, Silva Vieira M, Hussain T. 3D whole heart imaging for congenital heart disease. Front Pediatr 2017;5:36. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Valsangiacomo Buechel ER, Grosse‐Wortmann L, Fratz S, et al. Indications for cardiovascular magnetic resonance in children with congenital and acquired heart disease: An expert consensus paper of the imaging working group of the AEPC and the cardiovascular magnetic resonance section of the EACVI. Eur Heart J Cardiovasc Imaging 2015;16(3):281‐297. [DOI] [PubMed] [Google Scholar]
- 6. Botnar RM, Stuber M, Danias PG, Kissinger KV, Manning WJ. Improved coronary artery definition with T2‐weighted, free‐breathing, three‐dimensional coronary MRA. Circulation 1999;99(24):3139‐3148. [DOI] [PubMed] [Google Scholar]
- 7. Hu P, Stoeck CT, Smink J, et al. Noncontrast SSFP pulmonary vein magnetic resonance angiography: Impact of off‐resonance and flow. J Magn Reson Imaging 2010;32(5):1255‐1261. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Roujol S, Foppa M, Basha TA, et al. Accelerated free breathing ECG triggered contrast enhanced pulmonary vein magnetic resonance angiography using compressed sensing. J Cardiovasc Magn Reson 2014;16:91. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Sorensen TS, Korperich H, Greil GF, et al. Operator‐independent isotropic three‐dimensional magnetic resonance imaging for morphology in congenital heart disease: A validation study. Circulation 2004;110(2):163‐169. [DOI] [PubMed] [Google Scholar]
- 10. Greil GF, Powell AJ, Gildein HP, Geva T. Gadolinium‐enhanced three dimensional magnetic resonance angiography of pulmonary and systemic venous anomalies. J Am Coll Cardiol 2002;39:335‐341. [DOI] [PubMed] [Google Scholar]
- 11. Geva T, Greil GF, Marshall AC, Landzberg M, Powell AJ. Gadolinium‐enhanced 3‐dimensional magnetic resonance angiography of pulmonary blood supply in patients with complex pulmonary stenosis or atresia: Comparison with x‐ray angiography. Circulation 2002;106:473‐478. [DOI] [PubMed] [Google Scholar]
- 12. Potthast S, Mitsumori L, Stanescu LA, et al. Measuring aortic diameter with different MR techniques: Comparison of three‐dimensional (3D) navigated steady‐state free‐precession (SSFP), 3D contrast‐enhanced magnetic resonance angiography (CE‐MRA), 2D T2 black blood, and 2D cine SSFP. J Magn Reson Imaging 2010;31(1):177‐184. [DOI] [PubMed] [Google Scholar]
- 13. Gebker R, Gomaa O, Schnackenburg B, Rebakowski J, Fleck E, Nagel E. Comparison of different MRI techniques for the assessment of thoracic aortic pathology: 3D contrast enhanced MR angiography, turbo spin echo and balanced steady state free precession. Int J Cardiovasc Imaging 2007;23(6):747‐756. [DOI] [PubMed] [Google Scholar]
- 14. McDonald RJ, McDonald JS, Kallmes DF, et al. Gadolinium deposition in human brain tissues after contrast‐enhanced MR imaging in adult patients without intracranial abnormalities. Radiology 2017;285(2):546‐554. [DOI] [PubMed] [Google Scholar]
- 15. Henningsson M, Zahr R, Dyer A, et al. Feasibility of 3D black‐blood variable refocusing angle fast spin echo cardiovascular magnetic resonance for visualization of the whole heart and great vessels in congenital heart disease. J Cardiovasc Magn Reson 2018;20(1):76‐86. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Ginami G, Lopez K, Mukherjee RK, et al. Non‐contrast enhanced simultaneous 3D whole‐heart bright‐blood pulmonary veins visualization and black‐blood quantification of atrial wall thickness. Magn Reson Med 2019;81(2):1066‐1079. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Kellman P, Arai AE, McVeigh ER, Aletras AH. Phase‐sensitive inversion recovery for detecting myocardial infarction using gadolinium‐delayed hyperenhancement. Magn Reson Med 2002;47(2):372‐383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Atlas SW, Grossman RI, Hackney DB, Goldberg HI, Bilaniuk LT, Zimmerman RA. STIR MR imaging of the orbit. AJR Am J Roentgenol 1988;151:1025‐1030. [DOI] [PubMed] [Google Scholar]
- 19. Haase A, Frahm J, Hanicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol 1985;30:341‐344. [DOI] [PubMed] [Google Scholar]
- 20. Prieto C, Doneva M, Usman M, et al. Highly efficient respiratory motion compensated free‐breathing coronary MRA using golden‐step Cartesian acquisition. J Magn Reson Imaging 2015;41(3):738‐746. [DOI] [PubMed] [Google Scholar]
- 21. Henningsson M, Koken P, Stehning C, Razavi R, Prieto C, Botnar RM. Whole‐heart coronary MR angiography with 2D self‐navigated image reconstruction. Magn Reson Med 2012;67:437‐445. [DOI] [PubMed] [Google Scholar]
- 22. Cruz G, Atkinson D, Henningsson M, Botnar RM, Prieto C. Highly efficient nonrigid motion‐corrected 3D whole‐heart coronary vessel wall imaging. Magn Reson Med 2017;77:1894‐1908. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Henningsson M, Smink J, Razavi R, Botnar RM. Prospective respiratory motion correction for coronary MR angiography using a 2D image navigator. Magn Reson Med 2013;69(2):486‐494. [DOI] [PubMed] [Google Scholar]
- 24. Modat M, Ridgway GR, Taylor ZA, et al. Fast free‐form deformation using graphics processing units. Comput Methods Prog Biomed 2010;98(3):278‐284. [DOI] [PubMed] [Google Scholar]
- 25. Batchelor PG, Atkinson D, Irarrazaval P, Hill DL, Hajnal J, Larkman D. Matrix description of general motion correction applied to multishot images. Magn Reson Med 2005;54(5):1273‐1280. [DOI] [PubMed] [Google Scholar]
- 26. Keegan J, Drivas P, Firmin DN. Navigator artifact reduction in three‐dimensional late gadolinium enhancement imaging of the atria. Magn Reson Med 2014;72:779‐785. [DOI] [PubMed] [Google Scholar]
- 27. Moghari MH, Peters DC, Smink J, et al. Pulmonary vein inflow artifact reduction for free‐breathing left atrium late gadolinium enhancement. Magn Reson Med 2011;66:180‐186. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Henningsson M, Malik S, Botnar R, Castellanos D, Hussain T, Leiner T. Black‐blood contrast in cardiovascular MRI. J Magn Reson Imaging 2022;55:61‐80. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Ginami G, Neji R, Phinikaridou A, Whitaker J, Botnar RM, Prieto C. Simultaneous bright‐ and black‐blood whole‐heart MRI for noncontrast enhanced coronary lumen and thrombus visualization. Magn Reson Med 2018;79:1460‐1472. [DOI] [PMC free article] [PubMed] [Google Scholar]