Skip to main content
BMJ Open Access logoLink to BMJ Open Access
. 2021 Oct 16;78(4):326–334. doi: 10.1136/thoraxjnl-2020-216509

Automatic oxygen titration versus constant oxygen flow rates during walking in COPD: a randomised controlled, double-blind, crossover trial

Tessa Schneeberger 1,2,, Inga Jarosch 1,2, Daniela Leitl 1,2, Rainer Gloeckl 1,2, Wolfgang Hitzl 3,4,5, Clancy John Dennis 6, Tatjana Geyer 7, Carl-Peter Criée 8, Andreas Rembert Koczulla 1,2,9,10, Klaus Kenn 1,2,9
PMCID: PMC10086462  PMID: 34656996

Abstract

Rationale

In patients with COPD, oxygen (O2)-supplementation via a constant flow oxygen system (CFOS) can result in insufficient oxygen saturation (SpO2 <90%) during exercise. An automatically titrating O2-system (ATOS) has been shown to be beneficial compared with an untitrated CFOS, however, it is unknown if ATOS is superior to CFOS, titrated during exercise as stipulated by guidelines. The aim was to investigate the effects of ATOS compared with titrated CFOS on walking capacity in people with hypoxaemic COPD.

Methods

Fifty participants completed this prospective randomised controlled, double-blind, crossover trial. Participants performed two endurance shuttle walk tests (ESWTs) with: (1) exercise titrated CFOS (ESWTCFOS) and (2) ATOS targeting an SpO2 of 92% (ESWTATOS). Primary outcome measure was walking time. Secondary measures were SpO2, transcutaneous-PCO2 (TcPCO2), respiratory rate (RR), heart rate (HR) at isotime (end of shortest ESWT) with blood gases and dyspnoea at rest and end exercise.

Results

Participants (median (IQR): age 66 (59, 70) years, FEV1 28.8 (24.8, 35.1) % predicted, PO2 54.7 (51.0, 57.7) mm Hg, PCO2 44.2 (38.2, 47.8) mm Hg) walked significantly longer with ESWTATOS in comparison to ESWTCFOS (median effect (95% CI) +144.5 (54 to 241.5) s, p<0.001). At isotime, SpO2 was significantly higher (+3 (95% CI 1 to 4) %, p<0.001) with ATOS while TcPCO2, RR and HR were comparable. End exercise, PO2 (+8.85 (95% CI 6.35 to 11.9) mm Hg) and dyspnoea (−0.5 (95% CI −1.0 to −0.5) points) differed significantly in favour of ATOS (each p<0.001) while PCO2 was comparable.

Conclusion

In patients with hypoxaemia with severe COPD the use of ATOS leads to significant, clinically relevant improvements in walking endurance time, SpO2, PO2 and dyspnoea with no impact on PCO2.

Trial registration number

NCT03803384.

Keywords: ambulatory oxygen therapy, COPD pathology, exercise


Key messages.

What is the key question?

  • Does an automatic oxygen flow system perform better than a constant flow system with flows titrated for exercise as recommended by guidelines and is there an effect on carbon dioxide (CO2) retention?

What is the bottom line?

  • In patients with hypoxaemia with severe COPD the use of an automatic oxygen flow system leads to an increase in walking endurance time over the minimal important difference and despite higher oxygen (O2)-flows, has no impact on PCO2.

Why read on?

  • We show that during walking exercise, despite constant flows being titrated according to guidelines, an automatically titrating oxygen system provides more O2, maintains saturation and results in greatly improved walking durations and lower dyspnoea without adversely affecting CO2.

Introduction

Oxygen (O2) therapy is commonly used to treat people with hypoxic chronic obstructive pulmonary disease (COPD).1 During exercise O2 can increase O2-transport,2 delay muscle fatigue,2 alter breathing mechanics,3 reduce dyspnoea3 and improve exercise endurance.3 To be effective, prescribed flows are recommended to be titrated to an oxygen saturation (SpO2) of ≥90%.4 However, during demanding situations like exercise, titration of O2-flow to maintain adequate SpO2 is challenging.

In practice, contrary to guidelines suggesting titration via walking test,2 5 exercise O2-flows are frequently titrated at rest or prescribed as a fixed addition to the titrated resting flow.6

Given this, it is common to see continuous O2-flows, while physically active, which are inadequate for varying physiological demands.7 8 Further, in the study setting it has been seen that a single fixed flow is insufficient in keeping adequate oxygenation in some patients with COPD.9

In clinical practice there are concerns that providing excessive O2 in patients with COPD may induce a reduction in minute ventilation and thus lead to potentially dangerous hypercapnia.10 Possibly this thinking carries over to exercise where there is a fear that using high flow rates and therefore high O2-levels in people with severe to very severe COPD may have a deleterious effect on already elevated carbon dioxide (CO2) levels. For this reason it is recommended that in all patients, especially those with baseline hypercapnia, blood gases should be checked after each titration of flow for signs of respiratory acidosis and worsening hypercapnia.2 5

To overcome the challenge of varying metabolic demands during exercise and physical activity an automatically titrating oxygen system (ATOS), which regulates O2 flow to maintain a predefined SpO2-target, has been proposed as a solution to optimise the effects of O2-therapy.1 To date, only two studies have examined walking-exercise with ATOS in patients with COPD. These studies have shown that ATOS is better at maintaining oxygenation compared with a constant flow oxygen system (CFOS) during exercise. In general, hypoxaemia as well as hyperoxia occurred less with longer walking durations when using ATOS. These trials however had small sample sizes and were potentially biased towards ATOS as they had inappropriate CFOS O2-flow rates as a control comparison, and high SpO2-targets.6 11

Considering that ATOS can provide flows above what is normally used by patients during exercise, ATOS may exacerbate any underlying hypercapnic tendencies. Since constant blood gas monitoring in exercise is impractical, the continuous monitoring of transcutaneous carbon-dioxide partial pressure (TcPCO2) is logical. However, to date, TcPCO2 during exercise with ATOS has not been examined. Overall, stronger evidence to support or refute the use of ATOS during exercise for people with COPD is needed. Given this, the aims of this trial were to determine, in people with COPD who were hypoxaemic at rest and/or during exercise, whether an ATOS was more effective than individually titrated CFOS at: improving endurance exercise capacity (primary outcome), oxygenation, respiratory rate, heart rate and reducing dyspnoea/leg fatigue. Additionally, CO2-levels were monitored for potential side effects of using ATOS during walking exercise.

We hypothesised that ATOS would be superior to CFOS at increasing walking exercise capacity, but with higher CO2-levels.

Methods

This study was a prospective, single centre, randomised controlled crossover trial with blinding of participants, investigators and statistician. Randomisation of test order was achieved by an independent person prior to the study using a computer-generated random sequence. Test order was concealed in sequentially numbered sealed opaque envelopes. Participants between 40 and 80 years with a confirmed diagnosis of severe or very severe COPD (GOLD stage III to IV) with hypoxaemia or an indication for O2-therapy during exercise (PO2 <55 mm Hg at rest or during exercise or nadir SpO2 <88% during exercise) were recruited. Excluded were participants with an acute exacerbation of COPD or those who had cardiovascular medical conditions or orthopaedic restrictions limiting the ability to perform walking tests.

All participants were recruited and tested within an inpatient pulmonary rehabilitation programme over a period of 11 months (Schoen Klinik Berchtesgadener Land, Germany).

Informed written consent was obtained from all participants.

After an initial incremental shuttle walk test, participants performed on consecutive days (24-hour wash-out period), in a randomised order, two endurance shuttle walk tests (ESWTs) at 85% of maximum pace with12: (1) individually titrated constant oxygen-flow rates (ESWTCFOS) and (2) automatically titrated oxygen-flow rates (ESWTATOS). The FreeO2 (OxyNov, Canada), is a device which uses physiological data (primarily SpO2) in a closed loop algorithm to control an O2-flow from 0 to 20 L/min (flow accuracy ±0.1 L/min) to maintain SpO2 to a predefined target.13 14 The FreeO2 can also provide a constant flow. Therefore, to blind participants and investigators, the FreeO2 device was used for both ESWTs with the display of the device covered and an independent person other than the study investigator configured the settings before the test. The FreeO2 device, including the O2-cylinder was attached to a cart (online supplemental figure S1) and pushed by the investigator.

Supplementary data

thoraxjnl-2020-216509supp001.pdf (292KB, pdf)

For the ESWTATOS, SpO2-target was set at 92% to maintain participants SpO2 ≥90%. For the ESWTCFOS an individually exercise titrated O2-flow was used (see online supplemental for more information about O2-flow rate titration).

Outcome measures

Primary outcome was change in endurance exercise capacity as measured by ESWT. Secondary outcomes were time to SpO2 <90% and differences in SpO2, TcPCO2, heart rate (measured continuously via ear lobe sensor; SenTec, Switzerland), breathing frequency (measured continuously by respiratory inductance plethysmography; ApneaLink, ResMed, Australia) at rest, end exercise and 25%, 50%, 75% and 100% of isotime (end of shortest ESWT). O2-partial pressure, CO2-partial pressure, pH, base excess, hydrogen carbonate, lactate (measured by blood gas analysis; RAPIDPoint, Siemens, Germany) and sensation of dyspnoea and leg fatigue (10-point Borg scale15) were taken at rest and at the end of ESWTs. Blood gas samples were taken after 10 min at rest breathing with CFOS or ATOS and immediately at ESWT termination by an independent person.

At the end of each test, participants were asked to rate their perception of oxygenation, comfort of O2-delivery and possibility of using the O2-supplementation in everyday life on a standardised Likert-scale. Finally, after all tests, participants were asked to rate their preferred O2-system.

Statistical methods

Sample size calculation

In a retrospective study (n=12 patients with COPD) at our clinic, mean ESWT duration (primary endpoint) was compared in a CFOS and ATOS group. The mean difference in ESWT duration between both groups was 234.9 s and a SD of 392 s. To achieve a power of 90% at a significance level of 5% in a 2×2 crossover design for testing the effect one-sided, a sample size of 50 subjects was calculated. Assuming a drop-out rate of 10%, 55 subjects were enrolled in this study.

Data analysis

Data was checked for consistency and normality. Spearman’s and Pearson’s correlation coefficients were computed to analyse relations between variables. Data deviated from normality and hence non-parametric 2×2 crossover models were applied, 95% CI for medians and median effects were computed. No carry-over effects were found in the primary outcome. Two multivariable regression models with forward and backward variable selection algorithms were set up and tested to find models with high prediction accuracy, one model to predict the PCO2 post walk in the ATOS group and another for the CFOS group. Multiple R, coefficients of determination were computed, regression coefficients were tested and residuals were analysed for normality. Observed versus predicted values were illustrated.

All reported tests were two-sided, and p values<0.05 were considered statistically significant. NCSS (NCSS 10, NCSS, Kaysville, Utah, USA) and STATISTICA 13 (Hill, T. & Lewicki, P. Statistics: Methods and Applications, StatSoft, Tulsa, Oklahoma, USA) were used to analyse data descriptively, for testing crosstabulations tables, all two-sample tests and were also used to test and illustrate multiple regression models. StatXact (2013), V.10.0.0, Cytel Software Corporation (Cambridge, Massachusetts, USA) was used for sample size computations and for testing treatment and carry-over effects in non-parametric crossover analyses. P values were computed based on Monte-Carlo simulation methods instead of using asymptotic p values.

PASW 24 (IBM SPSS Statistics for Windows, V.21.0.) was used to compute correlations as well as descriptive analyses.

Results

Trial flow and participant characteristics

Fifty-five participants were recruited with five participants not completing the study (figure 1). All had severe to very severe COPD and were hypoxaemic (table 1). The median (IQR) pace for the ESWTs was 3.1 (2.6, 4.2) km/hour. ESWTCFOS was performed with a median O2-flow of 3 L/min while ESWTATOS resulted in a median flow of 4.5 L/min (table 2, figure 2A).

Figure 1.

Figure 1

Consolidated Standards of Reporting Trials—flow diagram. PR, pulmonary rehabilitation.

Table 1.

Participant characteristics

Variable COPD, n=50
Age, years 66 (59, 70)
Gender, female/male, n 23/27
BMI, kg/m2 23.9 (21.3, 27.0)
Waist–hip ratio
 Females 0.89 (0.83, 0.94)
 Males 1.0 (0.94, 1.04)
Pulmonary function
FEV1, L 0.78 (0.59, 1.0)
FEV1, % predicted 28.8 (24.8, 35.1)
FVC, L 1.69 (1.19, 2.28)
FVC, % predicted 50.1 (41.6, 63.7)
FEV1/FVC, % 45.8 (40.8, 53.7)
TLC, L 7.7 (7.0, 9.1)
TLC, % predicted 137.0 (121.8, 154.8)
FRCpleth, L 6.8 (5.9, 7.6)
FRCpleth, % predicted 213.6 (182.0, 251.8)
RV, L 5.9 (5.0, 7.0)
RV, % predicted 269.9 (218.1, 335.6)
RV/TLC, % 77.1 (68.6, 82.7)
RV/TLC, % predicted 188.5 (172.1, 206.9)
Raw, kPa s L-1 0.87 (0.63, 1.10)
Raw, % predicted 289.6 (209.6, 367.3)
sRaw, kPa s 6.0 (4.4, 8.9)
sRaw, % predicted 575.8 (397.4, 844.0)
GOLD grade
 III, n 23 (46)
 IV, n 27 (54)
Blood gases, room air—rest
pH 7.41 (7.39, 7.42)
PO2, mm Hg 54.7 (51.0, 57.7)
PCO2, mm Hg 44.2 (38.2, 47.8)
Exercise capacity
6-minute walk distance, m 303.5 (245.0, 371.5)
6-minute walk distance, % predicted 48.4 (38.5, 59.5)
Incremental shuttle walk test, m 205.0 (130.0, 325.0)
Long-term oxygen therapy
Period of application, months 47 (18, 79)
O2-flow rates, L
 Rest 1.75 (1.0, 2.9)
 Night 2.0 (1.0, 2.9)

Data presented as median (interquartile range (IQR)) or number.

%, per cent; BMI, body mass index; FEV1, forced expiratory volume in 1 s; FRC, functional residual capacity; FVC, forced vital capacity; kg, kilograms; kg/m2, kilograms per metre squared; L, litre; m, metres; mm Hg, millimetres of mercury; n, number; PCO2, partial pressure of carbon dioxide; pH, potential of hydrogen; PO2, partial pressure of oxygen; Raw, airway resistance; RV, residual volume; sRaw, specific airway resistance; TLC, total lung capacity.

Table 2.

Results

Constant O2-flow, CFOS, n=50 Automatic O2-flow, ATOS, n=50 P valuex) Median effect (95% CI) automatic—constant
Primary outcome
ESWT Time, s 333.50 (214, 581) 522.5 (277, 1200) 1.203E–04 144.5 (54 to 241.5)
Secondary outcomes
ESWT Distance, m 310 (200, 620) 465 (200, 1030) 2.602E–04 150 (60 to 31)
Oxygen Mean O2-flow rate 3.0 (3.0, 4.0) 4.5 (3.2, 6.1) 1.000E–05 1.34 (0.68 to 2.14)
Blood gas analyses
PO2rest, mm Hg 80.2 (73.1, 90.6) 65.35 (62.6, 68.1) 1.582E–14 –17.8 (–22.1 to –13.8)
PO2post, mm Hg 61.35 (55.0, 64.5) 71.5 (64.2, 75.8) 3.304E–08 8.85 (6.35 to 11.9)
PCO2rest, mm Hg 42.5 (39.8, 46.1) 40.85 (39.2, 48.2) 0.87
PCO2post, mm Hg 50.45 (46.7, 54.0) 50.45 (46.8, 54.8) 0.41
Phrest 7.41 (7.39, 7.43) 7.42 (7.40, 7.43) 0.64
Phpost 7.35 (7.32, 7.38) 7.34 (7.32, 7.37) 0.19
BErest, mmol/L 1.75 (0.30, 3.20) 1.20 (0.10, 3.60) 0.68
BEpost, mmol/L 0.80 (–0.80, 2.10) 0.50 (–1.10, 2.40) 0.52
HCO3rest, mmol/L 26.0 (24.70, 27.30) 25.4 (24.50, 27,60) 0.97
HCO3post, mmol/L 25.0 (23.60, 26.10) 24.80 (23.40, 26.50) 0.78
SaO2rest, % 95.0 (94.60, 96.60) 93.08 (91.70, 93.50) 1.582E–14 -3.0 (-3.45 to -2.6)
SaO2post, % 89.19 (86.50, 91.30) 92.75 (91.30, 94.04) 1.450E–08 3.17 (2.15 to 4.37)
Lactaterest, mmol/L 1.19 (0.99, 1.67) 1.09 (0.88, 1.45) 0.23
Lactatepost, mmol/L 1.90 (1.50, 2.62) 1.85 (1.14, 2.40) 4.134E–02 –0.135 (–0.29 to –0.005)
SenTec digital monitor
SpO2rest, % 96.0 (95.0, 98.0) 93.0 (93.0, 95.0) 2.286E–11 –3 (–3.5 to –2.5)
SpO2isotime_100%, % 89.0 (86.0, 93.0) 92.0 (90.0, 94.0) 5.275E–04 3 (1 to 4)
SpO2post, % 89.0 (86.0, 93.0) 93.0 (92.0, 94.0) 1.758E–04 3 (1.5 to 5)
SpO2min, % 88.0 (85, 91) 87.5 (83.0, 89.0) 0.05
SpO2max, % 97.0 (96.0, 99.0) 95.0 (94.0, 96.0) 1.122E–09 –2 (–2.5 to –1.5)
SpO2mean, % 92.0 (89.0, 94.0) 92.0 (91.0, 93.0) 0.77
Time to SpO2 <90%, s n=32
102.0 (73.0, 158.0)
n=39
55.0 (19.0, 74.0)
6.937E–06 –71.8 (–101 to –47)
Time to SpO2 <85%, s n=11
152.0 (112.0, 194.0)
n=16
81.5 (61.0, 137.5)
0.11
TcPCO2rest, mm Hg 43.0 (39.8, 46.1) 42.6 (39.4, 46.3) 0.19
TcPCO2isotime_100%, mm Hg 47.7 (45.4, 50.9) 47.7 (44.6, 53.7) 0.93
TcPCO2post, mm Hg 47.75 (45.4, 50.9) 48.0 (45.5, 53.6) 0.57
TcPCO2max, mm Hg 48.9 (45.8, 51.75) 49.2 (46.4, 55.2) 0.42
Heart raterest, b/min 83.0 (78.0, 88.0) 84.0 (79.0, 93.0) 3.859E–02 2 (0 to 4.5)
Heart rateisotime_100%, b/min 100.5 (95.0, 106.0) 100.5 (95.0, 108.0) 0.17
Heart ratepost, b/min 101.0 (95.0, 107.0) 103.0 (96.0, 112.0) 4.381E–03 3.5 (1 to 5.5)
Heart ratemax, b/min 104.5 (99.0, 116.0) 106.0 (99.0, 117.0) 0.06
Respiratory rate
Respiratory raterest, 1 /min 18.0 (14.0, 24.0) 20 (17.0, 24.0) 1.079E–03 2 (1 to 3.5)
Respiratory rateisotime_100%, 1 /min 31.0 (24.0, 45.5) 31.0 (23.5, 44.0) 0.25
Respiratory ratepost, 1 /min 29.0 (24.0, 34.0) 29.0 (21.0, 34.0) 0.40
Borg scale
Dyspnoearest, points 1.0 (0.0, 2.0) 1.0 (0.0, 2.0) 0.96
Dyspnoeapost, points 6.0 (5.0, 7.0) 5.0 (4.0, 7.0) 1.914E–04 –0.5 (–1 to –0.5)
Leg fatiguerest, points 1.0 (0.0, 2.0) 1.0 (0.0, 2.0) 0.79
Leg fatiguepost, points 4.0 (3.0, 6.0) 4.0 (3.0, 5.0) 0.10

Differences with p values<0.01 are presented in bold.

Note: Median effects are not necessarily medians of differences; x)non-parametric 2×2 crossover models based on Monte Carlo simulations using Wilcoxon (mid rank) test. P-values are unadjusted for multiple comparisons.

Data presented as median (IQR).

ATOS, automatically titrating oxygen system; BE, base excess; CFOS, constant flow oxygen system; ESWT, endurance shuttle walk test; HCO3, hydrogen carbonate; O2, oxygen; PCO2, partial pressure of carbon dioxide; pH, potential of hydrogen; PO2, partial pressure of oxygen; SaO2, oxygen saturation measured by blood gas analyses; SpO2, oxygen saturation; TcPCO2, transcutaneous carbon dioxide.

Figure 2.

Figure 2

(A) Oxygen flow rates during the endurance shuttle walk test (ESWT). (B) Oxygen saturation (SpO2) during ESWT. Dashed line at 92% SpO2. (C) Transcutaneous carbon dioxide (TcPCO2) during ESWT. Shaded area in grey between 45 and 55 mm Hg. (D) Heart rate (HR) during ESWT. Data is presented as median; error bars: IQR; p values were calculated via non-parametric 2×2 cross-over models based on Monte Carlo simulations using Wilcoxon (mid rank) test. P values are unadjusted for multiple comparisons.

Primary outcome

Participants walked significantly (p<0.001) longer in ESWTATOS compared with ESWTCFOS (table 2); 68% (n=34) of participants walked longer in ESWTATOS, 20% (n=10) walked longer in ESWTCFOS and 12% (n=6) walked equally. A longer duration than the minimal important difference (MID) of 65 s for an ESWT12 was achieved by 76.5% (n=26/34) of participants walking longer in ESWTATOS and 50% (n=5/10) of those in ESWTCFOS.

Reasons for ESWT termination significantly differed between the two tests (p=0.001). Dyspnoea was reported as the main reason for stopping the ESWT in 70% (n=35/50) of participants with CFOS while only 48% (n=24/50) of the participants stopped due to breathlessness with ATOS (p=0.02). Participants with a higher residual volume /total lung capacity per cent predicted showed a greater rest to end exercise change in dyspnoea during walking (ESWTCFOS: r=0.34, p=0.01; ESWTATOS: r=0.31, p=0.03) and more breathlessness at end exercise (ESWTCFOS: r=0.43, p=0.002; ESWTATOS: r=0.40, p=0.004) (online supplemental figure S2 and table S1).

As a subgroup analysis, the cohort was divided into two groups: (1) ATOS responders (‘participants walking ≥MID (65 s) during ESWTATOS in comparison to ESWTCFOS’) and (2) ATOS non-responders (participants walking less than 65 s during the ESWTATOS compared with ESWTCFOS’). Baseline characteristics were comparable. ATOS responders showed a significantly higher oxygenation and felt less breathlessness in comparison to ATOS non-responders at the end of ESWTATOS and received a significantly higher mean O2-flow rate during ESWTATOS (table 3, online supplemental figure S3A, B). Respiratory rate and TcPCO2 during ESWTATOS were comparable at each time point (online supplemental figure S3C, D), however mean TcPCO2 over the complete ESWTATOS duration was significantly different between ATOS responders and non-responders: TcPCO2mean, mm Hg 47.1 (45.3, 52.7) versus 45.0 (41.6, 48.8); median effect (95% CI) 3.5 mm Hg (0.1 to 7.1), p=0.037.

Table 3.

Subgroup analyses (ATOS responders vs non-responders)—results. Data presented as median (IQR)

Non-responder (n=24) Responder (n=26) P valuex) Median effect (95% CI)
responder—non-responder
Baseline characteristics
FEV1, % predicted 29.4 (25.6, 36.9) 30.1 (24.1, 33.4) 0.67
RV,/TLC, % predicted 182.8 (168.0, 203.0) 199.3 (178.7, 209.6) 0.18
FRC, % predicted 207.7 (176.5, 240.3) 228.7 (188.5, 252.7) 0.42
PO2—rest with room air, mm Hg 53.6 (52.0, 57.2) 54.7 (50.4, 57.3) 0.86
PCO2—rest with room air, mm Hg 44.4 (39.5, 46.8) 44.6 (38.1, 48.6) 0.72
O2 flow rate—exercise, L/min 3.5 (3.0, 4.0) 3.0 (2.5, 4.0) 0.42
Age, years 66.5 (61.0, 73.0) 62.0 (59.0, 68.0) 0.20
BMI, kg/m2 24.3 (21.3, 26.4) 23.7 (21.3, 26.1) 0.45
Waist circumference, cm 96.5 (89.5, 107.0) 89.5 (83.0, 100.5) 0.13
Waist–hip ratio 0.97 (0.89, 1.0) 0.91 (0.86, 0.98) 0.14
ESWTATOS
Mean O2-flow rate, L 3.9 (2.6, 4.7) 5.2 (3.9, 6.7) 8.768E–03 1.6 (0.49 to 2.7)
Time, s 280.0 (199.0, 857.0) 751.5 (419.0, 1200.0) 2.664E–03 268 (75 to 549)
Distance, m 265.0 (155.0, 925.0) 675.0 (350.0, 1210.0) 1.287E–02 225 (50 to 530)
Blood gas analyses
PO2rest, mm Hg 65.4 (62.7, 67.9) 65.1 (61.0, 68.6) 0.75
PO2post, mm Hg 66.3 (60.7, 73.2) 74.1 (71.0, 76.6) 3.364E–03 7.25 (2.4 to 11.6)
SaO2rest, mm Hg 93.1 (91.4, 93.7) 92.9 (91.7, 93.3) 0.62
SaO2post, mm Hg 91.8 (88.9, 93.7) 93.5 (92.0, 94.2) 4.141E–02 1.24 (0.04 to 2.8)
PCO2rest, mm Hg 40.8 (39.9, 47.0) 41.8 (39.0, 48.3) 0.74
PCO2post, mm Hg 48.3 (46.2, 53.6) 51.6 (48.6, 56.1) 0.23
Phrest 7.41 (7.39, 7.43) 7.41 (7.40, 7.43) 0.68
Phpost 7.36 (7.33, 7.38) 7.34 (7.31, 7.36) 0.05
Lactaterest, mmol/L 1.2 (1.0, 1.6) 1.0 (0.8, 1.2) 0.14
Lactatepost, mmol/L 1.9 (1.6, 2.4) 1.4 (1.0, 2.2) 0.08
Borg scale
Dyspnoeapost, points 6.0 (4.0, 7.5) 5.0 (4.0, 5.0) 1.107E–02 –1 (–2 to 0)
ΔDyspnoea, points 4.0 (3.0, 6.5) 3.0 (2.0, 4.0) 1.614E–03 –1 (–2 to 0)
Leg fatiguepost, points 6.0 (4.0, 7.5) 4.0 (3.0, 5.0) 0.35
ΔLeg fatigue, points 4.0 (3.0, 6.5) 2.0 (1.0, 3.0) 0.58

Differences with p values<0.01 are presented in bold. Note: Median effects are not necessarily medians of differences; x) Mann-Whitney U test. P values are unadjusted for multiple comparisons.

ATOS, automatically titrating oxygen system; BMI, body mass index; ESWT, endurance shuttle walk test; FEV1, forced expiratory volume in 1 s; FRC, functional residual capacity; PCO2, partial pressure of carbon dioxide; pH, potential of hydrogen; PO2, partial pressure of oxygen; RV, residual volume; SaO2, oxygen saturation measured by blood gas analyses; TLC, total lung capacity.

Secondary outcomes

SpO2 was significantly higher at rest and 25% isotime with CFOS compared with ATOS, whereas at 100% isotime and at end exercise the reverse was seen (figure 2B). Time to SpO2 <90% was significantly shorter with ATOS (table 2).

In line with this, PO2 was significantly higher with CFOS at rest and significantly higher at end exercise with ATOS. Except for lactate, there was no statistical difference in pH, base excess, hydrogen carbonate and PCO2 between the two interventions at rest and end exercise (table 2). The magnitude of change (Δ) in PCO2 from rest to end exercise varied in both tests (ESWTCFOS: range, −3.7 to 20.3 mm Hg; ESWTATOS: range, −1.7 to 19.6 mmHg). 62% of the participants had a comparable PCO2 response between the two ESWTs.

There was a weak to medium correlation between the mean O2-flow and the change in PCO2 (ΔPCO2) during ESWTCFOS and ESWTATOS, independent of O2-system (ESWTCFOS: r=0.29, p=0.04; ESWTATOS: r=0.39, p=0.007; online supplemental table S2).

A multivariable regression model for PCO2 post each ESWT, based on pre walking measures, achieved a multiple correlation coefficient of R=0.78, R2=0.60 (ESWTCFOS) and R=0.72, R2=0.52 (ESWTATOS). The relation between observed and predicted variables is illustrated in figure 3A, B.

Figure 3.

Figure 3

(A) Illustration of relation between observed and predicted PCO2 values (mm Hg) post endurance shuttle walk test (ESWT) with constant oxygen flows of the regression model. The model based on body mass index (p=0.0002), waist circumference (p=0.00008), residual volume in litres (p=0.008) and PCO2 at rest prior to walking test (p=0.00002). Multivariable regression model equation: PCO2 post walk predicted=22.358 + 1.325*Body Mass Index – 0.395*waist circumference + 1.382*residual volume + 0.574*PCO2 pre walking test. (B) Illustration of relation between observed and predicted PCO2 values (mm Hg) post ESWT with automatic oxygen flows of the regression model. The model is based on body mass index (p=0.011), waist circumference (p=0.028) and PCO2 in rest prior walking test (p=0.028). Multivariable regression model equation: PCO2 post walk predicted=26.462 + 0.628*PCO2 pre walking test +0.693*Body Mass Index – 0.203*waist circumference. ATOS, automatically titrating oxygen system; CFOS, constant flow oxygen system; PCO2, partial pressure of carbon dioxide.

No differences were found for TcPCO2 at any time point between the ESWTCFOS and ESWTATOS (table 2, figure 2C). Heart rate was statistically, significantly different between the two interventions at rest, 25%, 50% isotime and end exercise (table 2, figure 2D). Respiratory rate differed significantly at rest and dyspnoea differed at end exercise while leg fatigue showed no difference at any time point (table 2).

When questioned, a significantly (p=0.0005) higher number of participants strongly agreed that perceived oxygenation was ‘satisfactory’ while using ATOS (n=28) in comparison to CFOS (n=9). After completion of all study-related measurements a significantly (p<0.0001) higher number of participants (n=37) preferred ATOS over a CFOS. No other question differed significantly (online supplemental figure S4A–D).

Discussion

An automatically titrating O2-system used during walking resulted in a significantly improved exercise capacity and was associated with better oxygenation and less dyspnoea at end exercise compared with a constant flow O2-system. The change in both TcPCO2 and PCO2 during walking was comparable between systems. However, significant correlations between the change in PCO2 and the mean O2-flow, independent of delivery system, were observed.

Comparison to other trials

Our results augment the data supporting the use of ATOS compared with CFOS in walking exercise by applying methodological rigour to the study design, using a robust sample size calculation, a more realistic SpO2 target for ATOS (92% vs 94%6 11 SpO2) and a constant flow titration during exercise (as recommended by clinical guidelines)2 5 instead of a resting titration or standardised constant O2-flow (eg, resting O2-flow+1 L/min6; 2 L/min O2 11 ). This is also the first trial to measure CO2-levels continuously during exercise with ATOS and the first to present an independent measurement of the partial pressure of oxygen after walking exercise. Further, this is the first time a comparison of isotime measures during the ESWT have been shown and this is also the first study to report participants perception and preference due to the O2-delivery systems during walking exercise.

Exercise capacity

Even using individually exercise titrated CFOS, the increase in walking distance with ATOS was comparable or even greater to that seen in previous studies with a resting-titrated or standardised CFOS (change in walking duration: 43% vs 33%6 and 17%11). This may be due to the greater number of participants in our study as well more being hypoxaemic at rest (PO2: 54.7 vs 626 and 7211 mm Hg) making the effects of supplemental oxygen potentially more consistent.16

No differences in lung function or anthropometric parameters were found between ATOS responders (change in walking duration >MID) and non-responders. In our study, ATOS responders tended to have lower lactate values as well as less leg fatigue at ESWTATOS end compared with non-responders (table 3). Further, while responders walked a significantly longer duration and reported less dyspnoea, a significantly higher average O2-flow rate with ATOS was used. We might assume that the higher airflow may have contributed to a reduced sensation of breathlessness,17 however, no correlation between O2-flow rates and dyspnoea was found.

In addition to an increased walking capacity in 68% of the study cohort, significantly fewer participants had to stop ESWTATOS due to dyspnoea. Possibly this could be due to a better correction of hypoxaemia during the later stages (from 100% isotime) of the ESWT with ATOS. It has been shown that acute O2-supplementation improves exercise performance by reducing ventilation, dynamic hyperinflation and the perception of dyspnoea3 18 and those effects were potentially greater with ATOS. Somfay et al showed that supplemental oxygen during exercise induced a dose-dependent improvement in endurance capacity and symptom perception, which they attributed to decreased hyperinflation and slower breathing pattern.19 Participants in our trial were also given increasing doses of oxygen as SpO2 declined during exercise, so we might expect similar changes, however, when comparing ATOS to CFOS, we saw similar 100% isotime and end exercise respiratory rate. As inspiratory capacity was not measured in our study, we cannot report if a reduction in dynamic hyperinflation occurred.

Oxygenation

Oxygen saturation was superior at 100% isotime (Δ3%) and at end exercise (Δ3%) with ATOS compared with CFOS by a statistically and clinically relevant amount.20 Vivodtzev et al reported that minimum SpO2-values occurred with CFOS (CFOS: 83.6% vs ATOS: 89.5%; p<0.001)6 which is contrary to the present study where minimum SpO2-values during walking were comparable although a greater number of patients desaturated and the time to desaturation (SpO2 <90%) was significantly shorter with ATOS. Compared with Lellouche et al and Vivodtzev et al, individually titrated CFOS may have been better suited to the participants physiological needs during exercise6 11 However, as CFOS was exercise titrated in the present study, flows may have been greater than needed at rest which is confirmed by SpO2-profiles where a significantly higher SpO2 at rest and ESWT 25% isotime with CFOS was seen compared with ATOS (figure 2). The decline in SpO2 at the beginning of walking with ATOS could indicate that the automatic increase in O2-flows were slower than the physiological decrease during early exercise. The lower SpO2-target used in the current trial (92% vs 94%6 11) may have also slowed the response of ATOS compared with previous trials.

CO2

In clinical practice, increasing O2-flows in patients with hypercapnia has ongoing concerns due to the risk of hyperoxia-induced hypercapnia and is therefore regularly avoided.21 22 However, in agreement with PCO2 results observed by Lellouche et al,11 carbon dioxide levels measured via blood gas sample and transcutaneously were comparable between the two ESWTs, even with higher O2-flows from 50% isotime on with ATOS. When comparing ATOS-responders to non-responders, responders received on average 1.6 L/min more oxygen with ATOS and mean TcPCO2 during walking was significantly higher. Overall, for both systems a significant correlation between change in PCO2 and the mean O2-flow was found.

Exercise-induced hypercapnia

Some patients with COPD, while normocapnic at rest, retain CO2 during increased activity and may develop exercise-induced hypercapnia (EIH: defined as PCO2 ≥45 mm Hg post ESWT).23 24 To examine a possible difference in the effects between participants who were normocapnic at rest (PCO2 <45 mm Hg breathing room air; n=27) and those who were hypercapnic at rest (PCO2 ≥45 mm Hg breathing room air; n=23) a subgroup analysis was performed (online supplemental table S3). We found, in the normocapnic subgroup, signs of EIH in 77.8% (n=21/27) of the participants with CFOS and 74.1% (n=20/27) with ATOS. Development of EIH in both ESWTs was observed in 74.1% (n=20) of participants. This is greater than previously seen by Andrianopoulos et al where only 31% of participants developed EIH during a 6-minute walk test.25 This could be explained by participants being less hypoxaemic at rest with better lung function and only 79% having a prescription of O2 during exercise compared with 100% in the present study.25

In this study we developed a multivariable regression model to predict the PCO2 post walking exercise. As exercise-induced changes in PCO2 are highly dependent on several pathophysiological mechanisms in COPD23 the ability to predict the PCO2 post walking may help clinicians better select patients for ATOS. However, we suggest evaluating this in a larger sample in order to readjust the models and further improve accuracy. Of note, independent of O2-system, PCO2 was on average normocapnic pre-exercise (CFOS: 42.5 mm Hg; ATOS: 40.85 mm Hg) and was a median of 50.45 mm Hg post exercise. This is reinforced by a moderate negative correlation between resting PCO2 and change in ESWT PCO2 showing participants with a lower starting PCO2 had the greatest changes during exercise. Simard et al found, that in some patients with COPD, EIH is a precursor to chronic hypercapnia.26 The large number of participants with EIH in the current study might underline the importance of observing PCO2-levels in patients with severe and very severe COPD during exercise to consider using other aids (eg, non-invasive ventilation).27 28

Limitations

Some limitations of the present study must be considered. Surprisingly, 14 patients reached the maximal exercise duration of 20 min with ATOS and 5 of those with CFOS also. Had this limit been longer, the effect size may have been different, however, given the present study reached an effect of statistical significance greater than the MID, the final outcome is unaffected. The ISWT was conducted once, but 90% of the participants did either not reach the maximum duration or reached it only in one test. We therefore assume, that in the majority of the study population the speed was appropriately chosen. Second the use of the 6-minute walk test to obtain an O2-titration is not optimal for an ESWT, however it is the practice suggested in clinical guidelines and is superior to using the same O2-flow for the entire group.

Further, this study demonstrates the immediate effects of O2-therapy during walking exercise and might not reflect longer usage scenarios or during different exercise modalities. Also of note, no adjustment for multiple comparisons was done on the familywise error, however, p values with high precision (in scientific notation) are presented for use in a correction such as a Bonferroni-Holm. Finally, the regression models are based on learning samples only and the generalisability of the models should be evaluated in independent validation data sets.

Current generation ATOS devices are limited though as they are large, assume an infinite O2-supply and rely completely on a finger sensor. Future studies should determine whether improved ATOS with faster O2-flow regulation have the potential to improve results. They should also focus on medium-term and long-term effects of using ATOS in exercise training or daily life as well as economic factors like O2-usage (eg, overall use during exercise training), device, maintenance and staff costs (eg, reduced number of O2-titration assessments). Further, other pathologies, such as interstitial lung disease29 or COVID-1930 may desaturate in exercise and an ATOS could be a promising approach.

Conclusion

As shown by this randomised, double-blinded, crossover trial, we found that the use of an automatically titrating supplemental O2-system, capable of adjusting O2-flows during exercise in response to SpO2, leads to significantly and clinically relevant improvements in walking endurance time, oxygenation and dyspnoea compared with an individually, exercise titrated constant O2-flow system as commonly used in practice.

The clinical implications are that ATOS set to keep a SpO2-target of 92% in patients with hypoxaemia with severe and very severe COPD has an immediate positive effect on exercise capacity with no difference in CO2. This result disputes the ‘one flow rate fits all’ mentality for O2-supplementation during exercise and suggests that automatic systems may be a promising method for improved, individually tailored treatment. Finally, participants preferred an automatic O2-system over constant and future work should be invested in making a more portable version which reacts faster to changing O2-levels.

Acknowledgments

Consumable materials for the SenTec device were kindly supported by SenTec AG. SenTec AG had no influence on the study design, execution or analyses.

Footnotes

Contributors: TS, KK and IJ conducted the study design. TS, DL, TG, IJ and RG performed the data collecting. WH and TS performed the data analyses. TS prepared the manuscript. KK, CJD, C-PC, ARK, IJ, WH, RG, DL, TG and TS critically revised the manuscript. All authors have read and approved the manuscript. TS takes the responsibility for the integrity.

Funding: The authors have not declared a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

Competing interests: None declared.

Provenance and peer review: Not commissioned; externally peer reviewed.

Supplemental material: This content has been supplied by the author(s). It has not been vetted by BMJ Publishing Group Limited (BMJ) and may not have been peer-reviewed. Any opinions or recommendations discussed are solely those of the author(s) and are not endorsed by BMJ. BMJ disclaims all liability and responsibility arising from any reliance placed on the content. Where the content includes any translated material, BMJ does not warrant the accuracy and reliability of the translations (including but not limited to local regulations, clinical guidelines, terminology, drug names and drug dosages), and is not responsible for any error and/or omissions arising from translation and adaptation or otherwise.

Data availability statement

Data are available upon reasonable request. Because of a data privacy statement in the ethics proposal, general raw data sharing is not permitted. The study protocol and informed consent form (German language) can be shared.

Ethics statements

Patient consent for publication

Not applicable.

Ethics approval

The study was approved by an Ethic Committee (Philipps-University of Marburg).

References

  • 1. Branson RD. Oxygen therapy in COPD. Respir Care 2018;63:734–48. 10.4187/respcare.06312 [DOI] [PubMed] [Google Scholar]
  • 2. Hardinge M, Annandale J, Bourne S, et al. British thoracic Society guidelines for home oxygen use in adults. Thorax 2015;70 Suppl 1:i1–43. 10.1136/thoraxjnl-2015-206865 [DOI] [PubMed] [Google Scholar]
  • 3. O'Donnell DE, D'Arsigny C, Webb KA. Effects of hyperoxia on ventilatory limitation during exercise in advanced chronic obstructive pulmonary disease. Am J Respir Crit Care Med 2001;163:892–8. 10.1164/ajrccm.163.4.2007026 [DOI] [PubMed] [Google Scholar]
  • 4. Global Strategy for the Diagnosis, Management, and Prevention of Chronic Obstructive Pulmonary Disease - 2020 Report, 2020. Available: https://goldcopd.org/wp-content/uploads/2019/12/GOLD-2020-FINAL-ver1.2-03Dec19_WMV.pdf [Accessed 27 May 2020].
  • 5.et alHaidl P, Jany B, Geiseler J. Guidelines for long-term oxygen therapy S2k-Guideline published by the German respiratory Society, 2020. Available: https://www.awmf.org/uploads/tx_szleitlinien/020-002l_S2k_Langzeit_Sauerstofftherapie_2020-08.pdf [Accessed 02 Nov 2020]. [DOI] [PubMed]
  • 6. Vivodtzev I, L'Her E, Vottero G, et al. Automated O2 titration improves exercise capacity in patients with hypercapnic chronic obstructive pulmonary disease: a randomised controlled cross-over trial. Thorax 2019;74:298–301. 10.1136/thoraxjnl-2018-211967 [DOI] [PubMed] [Google Scholar]
  • 7. Sliwiński P, Lagosz M, Górecka D, et al. The adequacy of oxygenation in COPD patients undergoing long-term oxygen therapy assessed by pulse oximetry at home. Eur Respir J 1994;7:274–8. 10.1183/09031936.94.07020274 [DOI] [PubMed] [Google Scholar]
  • 8. Morrison D, Skwarski KM, MacNee W. The adequacy of oxygenation in patients with hypoxic chronic obstructive pulmonary disease treated with long-term domiciliary oxygen. Respir Med 1997;91:287–91. 10.1016/S0954-6111(97)90032-9 [DOI] [PubMed] [Google Scholar]
  • 9. Jarosch I, Gloeckl R, Damm E, et al. Short-Term effects of supplemental oxygen on 6-Min walk test outcomes in patients with COPD: a randomized, placebo-controlled, single-blind, crossover trial. Chest 2017;151:795–803. 10.1016/j.chest.2016.11.044 [DOI] [PubMed] [Google Scholar]
  • 10. Abdo WF, Heunks LMA. Oxygen-induced hypercapnia in COPD: myths and facts. Crit Care 2012;16:323. 10.1186/cc11475 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Lellouche F, L'Her E, Bouchard P-A, et al. Automatic oxygen titration during walking in subjects with COPD: a randomized crossover controlled study. Respir Care 2016;61:1456–64. 10.4187/respcare.04406 [DOI] [PubMed] [Google Scholar]
  • 12. Singh SJ, Puhan MA, Andrianopoulos V, et al. An official systematic review of the European respiratory Society/American thoracic Society: measurement properties of field walking tests in chronic respiratory disease. Eur Respir J 2014;44:1447–78. 10.1183/09031936.00150414 [DOI] [PubMed] [Google Scholar]
  • 13. Lellouche F, L'her E. Automated oxygen flow titration to maintain constant oxygenation. Respir Care 2012;57:1254–62. 10.4187/respcare.01343 [DOI] [PubMed] [Google Scholar]
  • 14. Lellouche F, Bouchard P-A, Roberge M, et al. Automated oxygen titration and weaning with FreeO2 in patients with acute exacerbation of COPD: a pilot randomized trial. Int J Chron Obstruct Pulmon Dis 2016;11:1983–90. 10.2147/COPD.S112820 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Borg G. Perceived exertion as an indicator of somatic stress. Scand J Rehabil Med 1970;2:92–8. [PubMed] [Google Scholar]
  • 16. Ameer F, Carson KV, Usmani ZA. Ambulatory oxygen for people with chronic obstructive pulmonary disease who are not hypoxaemic at rest. Cochrane Database Syst Rev 2014;2014:CD000238. 10.1002/14651858.CD000238.pub2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Swan F, Newey A, Bland M, et al. Airflow relieves chronic breathlessness in people with advanced disease: an exploratory systematic review and meta-analyses. Palliat Med 2019;33:618–33. 10.1177/0269216319835393 [DOI] [PubMed] [Google Scholar]
  • 18. Maltais F, Simon M, Jobin J, et al. Effects of oxygen on lower limb blood flow and O2 uptake during exercise in COPD. Med Sci Sports Exerc 2001;33:916–22. 10.1097/00005768-200106000-00010 [DOI] [PubMed] [Google Scholar]
  • 19. Somfay A, Porszasz J, Lee SM, et al. Dose-response effect of oxygen on hyperinflation and exercise endurance in nonhypoxaemic COPD patients. Eur Respir J 2001;18:77–84. 10.1183/09031936.01.00082201 [DOI] [PubMed] [Google Scholar]
  • 20. Stoller JK, Panos RJ, Krachman S, et al. Oxygen therapy for patients with COPD: current evidence and the long-term oxygen treatment trial. Chest 2010;138:179–87. 10.1378/chest.09-2555 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Aubier M, Murciano D, Fournier M, et al. Central respiratory drive in acute respiratory failure of patients with chronic obstructive pulmonary disease. Am Rev Respir Dis 1980;122:191–9. 10.1164/arrd.1980.122.2.191 [DOI] [PubMed] [Google Scholar]
  • 22. Aubier M, Murciano D, Milic-Emili J, et al. Effects of the administration of O2 on ventilation and blood gases in patients with chronic obstructive pulmonary disease during acute respiratory failure. Am Rev Respir Dis 1980;122:747–54. 10.1164/arrd.1980.122.5.747 [DOI] [PubMed] [Google Scholar]
  • 23. O'Donnell DE, D'Arsigny C, Fitzpatrick M, et al. Exercise hypercapnia in advanced chronic obstructive pulmonary disease: the role of lung hyperinflation. Am J Respir Crit Care Med 2002;166:663–8. 10.1164/rccm.2201003 [DOI] [PubMed] [Google Scholar]
  • 24. Light RW, Mahutte CK, Brown SE. Etiology of carbon dioxide retention at rest and during exercise in chronic airflow obstruction. Chest 1988;94:61–7. 10.1378/chest.94.1.61 [DOI] [PubMed] [Google Scholar]
  • 25. Andrianopoulos V, Vanfleteren LEGW, Jarosch I, et al. Transcutaneous carbon-dioxide partial pressure trends during six-minute walk test in patients with very severe COPD. Respir Physiol Neurobiol 2016;233:52–9. 10.1016/j.resp.2016.08.003 [DOI] [PubMed] [Google Scholar]
  • 26. Simard AA, Maltais F, LeBlanc P. Functional outcome of patients with chronic obstructive pulmonary disease and exercise hypercapnia. Eur Respir J 1995;8:1339–44. 10.1183/09031936.95.08081339 [DOI] [PubMed] [Google Scholar]
  • 27. Gloeckl R, Andrianopoulos V, Stegemann A, et al. High-pressure non-invasive ventilation during exercise in COPD patients with chronic hypercapnic respiratory failure: a randomized, controlled, cross-over trial. Respirology 2019;24:254–61. 10.1111/resp.13399 [DOI] [PubMed] [Google Scholar]
  • 28. Menadue C, Piper AJ, van 't Hul AJ. Non-invasive ventilation during exercise training for people with chronic obstructive pulmonary disease. Cochrane Database Syst Rev 2014;5:CD007714. 10.1002/14651858.CD007714.pub2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Lama VN, Flaherty KR, Toews GB, et al. Prognostic value of desaturation during a 6-minute walk test in idiopathic interstitial pneumonia. Am J Respir Crit Care Med 2003;168:1084–90. 10.1164/rccm.200302-219OC [DOI] [PubMed] [Google Scholar]
  • 30. Spruit MA, Holland AE, Singh SJ, et al. COVID-19: interim guidance on rehabilitation in the hospital and Post-Hospital phase from a European respiratory Society and American thoracic Society-coordinated international Task force. Eur Respir J 2020. doi: 10.1183/13993003.02197-2020. [Epub ahead of print: 13 Aug 2020]. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary data

thoraxjnl-2020-216509supp001.pdf (292KB, pdf)

Data Availability Statement

Data are available upon reasonable request. Because of a data privacy statement in the ethics proposal, general raw data sharing is not permitted. The study protocol and informed consent form (German language) can be shared.


Articles from Thorax are provided here courtesy of BMJ Publishing Group

RESOURCES