Skip to main content
Lippincott Open Access logoLink to Lippincott Open Access
. 2023 Feb 22;76(5):660–666. doi: 10.1097/MPG.0000000000003748

Elevations in End-Tidal CO2 With CO2 Use During Pediatric Endoscopy With Airway Protection: Is This Physiologically Significant?

Chinenye R Dike *,†,, Andrew Huang Pacheco *, Elizabeth Lyden , David Freestone *, Ojasvini Choudhry *, Warren P Bishop §, Mohanad Shukry
PMCID: PMC10097482  PMID: 36821847

Background:

Inflation of the gastrointestinal lumen is vital for proper visualization during endoscopy. Air, insufflated via the endoscope, is gradually being replaced with carbon dioxide (CO2) in many centers, with the intention of minimizing post-procedural discomfort due to retained gas. Recent studies suggest that the use of CO2 during pediatric esophagogastroduodenoscopy (EGD) with an unprotected airway is associated with transient elevations in exhaled CO2 (end-tidal CO2, EtCO2), raising safety concerns. One possible explanation for these events is eructation of insufflation gas from the stomach.

Objectives:

To distinguish eructated versus absorbed CO2 by sampling EtCO2 from a protected airway with either laryngeal mask airway (LMA) or endotracheal tube (ETT), and to observe for changes in minute ventilation (MV) to exclude hypoventilation events.

Methods:

Double-blinded, randomized clinical trial of CO2 versus air insufflation for EGD with airway protection by either LMA or ETT. Tidal volume, respiratory rate, MV, and EtCO2 were automatically recorded every minute. Cohort demographics were described with descriptive characteristics. Variables including the percent of children with peak, transient EtCO2 ≥ 60 mmHg were compared between groups.

Results:

One hundred ninety-five patients were enrolled for 200 procedures. Transient elevations in EtCO2 of ≥60 mmHg were more common in the CO2 group, compared to the air group (16% vs 5%, P = 0.02), but were mostly observed with LMA and less with ETT. Post-procedure pain was not different between groups, but flatulence was reported more with air insufflation (P = 0.004).

Conclusion:

Transient elevations in EtCO2 occur more often with CO2 than with air insufflation during pediatric EGD despite protecting the airway with an LMA or, to a lesser degree, with ETT. These elevations were not associated with changes in MV. Although no adverse clinical effects from CO2 absorption were observed, these findings suggest that caution should be exercised when considering the use of CO2 insufflation, especially since the observed benefits of using this gas were minimal.

Keywords: carbon dioxide, end-tidal CO2, minute ventilation, pediatrics


What Is Known.

  • Adult studies have demonstrated safety of carbon dioxide (CO2) insufflation. Most have also shown fewer post-procedural gas-related symptoms when CO2 gas is used for insufflation over air. These results have been reported with both esophagogastroduodenoscopy (EGD) and colonoscopy.

  • Transient elevations of end-tidal CO2 (EtCO2) are observed in nonintubated children undergoing EGD without concurrent elevations in transcutaneous CO2 levels, suggesting that eructated insufflation gas could be responsible.

What Is New

  • Transient elevations in EtCO2 persists in children undergoing EGD with CO2 insufflation despite airway protection with laryngeal mask airway or endotracheal tube.

  • CO2 insufflation is not associated with changes in minute ventilation or other adverse physiologic or anesthesia events, suggesting that these elevations in EtCO2 are not clinically significant.

Carbon dioxide (CO2) is increasingly used over air for endoscopic insufflation in children, largely based on adult studies demonstrating its safety and, in most studies, decreased post-procedure gas-related discomfort (15). Pediatric studies remain few (611); some of these reported decreased post-procedural discomfort with CO2 use, especially following colonoscopy (911). In one (6), pain reduction was only evident following adjusting for baseline abdominal pain, and no such benefit was observed in another (8). A recent meta-analysis concluded that there is limited evidence to recommend routine use of CO2 in pediatric endoscopy (12), while another concluded that CO2 gas is preferred for pediatric colonoscopy (13). Furthermore, peak end-tidal CO2 (EtCO2) measurement in 2 of the pediatric studies showed higher EtCO2 in the CO2 group, raising concerns about systemic absorption of CO2 (6,8).

Our previously published randomized clinical trial at the University of Iowa, comparing CO2 and air insufflation for esophagogastroduodenoscopy (EGD), revealed that the use of CO2 in children resulted in an increased frequency of transient (usually ≤ 1 minute) elevations in EtCO2 (≥ 60 mmHg) (8). These children did not have a protected airway, and we suspected systemic absorption of the insufflation gas. However, in a follow-up nested study, these peaks in EtCO2 were not associated with elevations in transcutaneous CO2 (TCO2), providing reassurance that systemic absorption was not occurring (7). We concluded that the observed peaks in EtCO2 during EGD were likely due to eructation of insufflated CO2. Some evidence of absorption of CO2 during colonoscopy has been reported by Dharmaraj et al (6), who noted EtCO2 elevations during this procedure. Indeed, the reason for using CO2 insufflation during colonoscopy is the expectation of gradual systemic absorption of the gas, reducing intestinal distention. However, use of CO2 in upper endoscopy carries the potential additional hazard of inhalation of concentrated CO2 escaping into the upper airway. Our prior observations of brief spikes of EtCO2 > 60 mmHg without elevated systemic (transcutaneous) CO2 suggests this to be a potential hazard.

To better examine the hypothesis that eructated CO2 is responsible for transient elevations in EtCO2 during EGD in nonintubated children, we designed this study to discover if frequent spikes in EtCO2 would still be detected with a protected airway, sampling directly from an endotracheal tube (ETT) or laryngeal mask airway (LMA) to minimize contamination by eructated CO2. We also sought to determine if changes in ventilation were associated with any observed elevations of EtCO2 in these children, as decreased gas exchange could also cause increased systemic CO2 and thus ETCO2.

MATERIALS AND METHODS

Design

This is a double-blinded, prospective, randomized clinical trial that compared CO2 versus air insufflation in children with a protected airway undergoing endoscopic procedures. The study was registered as a phase III clinical trial with Clinicaltrials.gov (NCT04541667) and approved by the Institutional Review Board (IRB) at University of Nebraska Medical Center.

Subject Enrollment

Children between 6 months to 19 years of age who were about to undergo an endoscopic procedure at the Children’s Hospital and Medical Center Omaha were recruited. In all cases, these were previously planned to be performed with a protected airway; either LMA or ETT. Patients were enrolled after obtaining age-appropriate consent, or consent with assent, as required by the IRB. Procedures included mostly diagnostic EGD-related procedures, including consecutive EGD and colonoscopy, EGD only, and enteroscopy. One patient who had colonoscopy only was enrolled but was not included in the outcome analysis. Patients with American Society of Anesthesiology (ASA) Physical Status Classification System of >2, children with chronic lung disease, and children who are wards of the state were excluded. Spanish-speaking families were enrolled with the help of interpreters. Families requiring the services of other language interpreters were excluded.

Randomization

Participants were randomized 1:1 for air versus CO2 by the endoscopic nurse or surgical technologist using a sealed envelope system. The endoscopy gas delivery system was set up to deliver either air or CO2 after randomization and was then covered to maintain blinding of the endoscopist. The anesthesia team, patients, and endoscopists were all blinded to gas being used. Additionally, the biostatistician was blinded to the randomization gas during analysis. The method of airway protection used (ETT or LMA) was left to the discretion of the anesthesiologist and was not randomized.

Data Collection

Data collected included patient demographics, patient’s medical problems or comorbidities, ASA status, indication for procedure, procedure duration, presence of a fellow trainee, year of training, and whether there were any procedure-related complications. Anesthesia-related characteristics were also recorded, including anesthesia duration, agents and doses used, and any anesthesia-related complications, such as hypotension, hypertension, or hypoxia. Anesthesia was provided by a pediatric anesthesiologist.

Ventilation was recorded continuously via direct connection to the LMA or ETT, including respiratory rate (RR) and tidal volume (TV). EtCO2 levels were also continuously monitored. All these values were recorded every minute. Oxygenation was monitored continuously by pulse oximetry. Data was entered in a secure institutional database, REDCap.

Outcomes

Primary

  • Minute ventilation (MV) = TV × RR

  • Hypercapnia; defined as EtCO2 level ≥ 60 mmHg, at any point during the procedure.

Secondary

  • Abdominal discomfort as assessed by the nurse or reported by the patient using FACES scale or Visual Analogue scale both pre- and post-procedure.

  • Patient-reported pre- and post-procedure headache, nausea, emesis, abdominal bloating, flatulence, and belching.

  • Procedure or anesthesia-related complications.

Post-procedure outcomes were reported or assessed in the recovery room when the participants awakened from anesthesia.

Statistical Analysis

Cohort characteristics were summarized using counts and percentages for categorical data, and medians and range for continuous data. The Mann-Whitney test was used to compare continuous variables between the groups, while Fisher exact test was used to compare categorical variables. The peak value of EtCO2 over the duration of the procedure was determined for each patient. The percent of children with at least 1 elevation of EtCO2 ≥ 60 mmHg was compared between groups. For MV, adjustment for the patient’s weight was done by dividing the patient’s mean MV by the patient’s weight and compared between groups. SAS software version 9.4 was used for data analysis (SAS Institute Inc., Cary, NC) and a P value <0.05 was considered statistically significant.

Power Calculation

When designing the study, we planned to enroll a sample size large enough to detect small differences in mean EtCO2 between groups. A sample size of 75 participants per group was estimated to achieve 80% power to detect a small difference in mean EtCO2 between groups assuming a significance level (alpha) of 0.05 using a 2-sided 2-sample t test. We aimed to enroll 100 patients in each group to accommodate for a 25% drop out rate.

RESULTS

A total of 195 patients were enrolled for 200 procedures from November 18, 2019 to July 6, 2021. One hundred two procedures were randomized to the CO2 group, while 98 procedures were randomized to the air group. One patient had a colonoscopy-only procedure and was excluded from analysis. A procedure involving both an enteroscopy and a colonoscopy was excluded from the primary outcome analysis because the procedure duration was the longest in the cohort and did not match any other procedure in the other group. A procedure that did not have any recorded MV values was excluded from the MV analysis. Thirteen patients who were approached for the study declined participation and 1 family consented for the study but withdrew consent before study procedures were started. Therefore, a total of 197 procedures were included in all the analysis.

Characteristics of the Cohort (Table 1)

TABLE 1.

Characteristics of cohort

CO2 Air P value
Age, y, median (IQR) 0.92
Race/ethnicity (%) 12 (7) 12 (7) 0.83
 Caucasian 82 86
 African American/African 1 1
 Hispanic 6 6
 Other 1 1
 Two or more races 10 6
Sex (%) 0.57
 Male 45 41
 Female 55 59
Weight, kg, median (IQR) 53.2 (40.8) 47 (31.7) 0.38
BMI/weight for length z score, median (IQR) 0.61 (2.08) 0.23 (2.05) 0.29
Duration of procedure, min, median (IQR) 32 (28) 31 (34) 0.77
EGD only + EGD portion of combined procedures 10 (7) 10 (7) 0.33
Colonoscopy 30 (16) 28 (16) 0.77
Anesthesia duration, min, median (IQR) 55.5 (30) 52.5 (36) 0.49
Primary GI symptom (%) 0.77
 Abdominal pain 63 60 0.6
 Diarrhea 20 23 0.64
 Feeding or growth concerns 9 11 0.33
 Dysphagia 7 11 0.38
 Other 66 59
Procedure type (%) 0.97
 EGD only 29 32
 EGD/colonoscopy 67 68
 Enteroscopy 1 0
 Colonoscopy only 1 0
Personnel involved (%) 0.69
 Faculty only 47.1 53.1
 Faculty + 1st year fellow 17.7 16.3
 Faculty + 2nd year fellow 18.6 19.4
 Faculty + 3rd year fellow 16.7 11.2
ASA status (%) 0.32
 ASA 1 21 28
 ASA 2 79 72
Airway protection type (%) 0.07
 LMA 71 83
 ETT 29 17

Other symptoms include bloody stools, fatigue. BMI z score (children ≥ 2 years); weight for length z score (children < 2 years).

ASA = American Society of Anesthesiology; BMI = body mass index; EGD = esophagogastroduodenoscopy; ETT = endotracheal tube; GI = gastrointestinal; IQR = interquartile range; LMA = laryngeal mask airway.

Patient characteristics, including age, body mass index z score, weight (in kilograms), gender, race/ethnicity, ASA status, procedure, and anesthesia duration were similar in both groups. About 57% of the cohort were females and 84% Caucasians. Seventy-six percent of the cohort had an ASA status of 2. Around half of the procedures in each group involved a trainee while half was done by faculty alone. LMA was used to protect the airway in 77% of the procedures and ETT in the remainder. The median age was 12 years in both groups and about 9% of the cohort was below 4 years.

Outcomes

Primary

The percentage of participants who experienced any episode of hypercapnia (EtCO2 ≥ 60 mmHg) was higher in the CO2 group when compared to the air group (21% vs 9%, P= 0.02; see Fig. 1). Out of the 136 participants who had consecutive EGD and colonoscopy, 29% experienced hypercapnia at some point during their procedure, 23% in the CO2 group, versus 8% in the air group (P = 0.02). Hypercapnia events were more common in the upper endoscopy portion of the procedure in the CO2 group (21%) when compared to the air group (5%) (P = 0.02) but were not significantly different during the colonoscopy portion of procedures (10% vs 5%, P= 0.33). Episodes of hypercapnia in the CO2 group were significantly more common than air with the use of LMA (22% vs 10 %, P = 0.045). This was less apparent in the ETT group (17% in the CO2 group vs 0% in the air group, P = 0.14). The median value of MV was similar in both groups; 75.9 mL/kg/min in the CO2 group and 74.3 mL/kg/min in the air group (P = 0.45). Median MV measured over 5 minutes before peak EtCO2 was 53.28 mL/kg/min in the CO2 group versus 40.23 mL/kg/min in the air group (P = 0.20) and was 68.5 mL/kg/min in the CO2 group during the 5 minutes after episodes of hypercapnia when compared to 49.5 mL/kg/min in the air group (P = 0.067). Multiple spikes of EtCO2 ≥ 60 mmHg were observed more often in the CO2 group (P = 0.03) and, although many of the EtCO2 spikes lasted no more than 1 minute, 2 patients in the CO2 group had prolonged episodes of elevated EtCO2, lasting up to 10 minutes.

FIGURE 1.

FIGURE 1.

Percent of procedures with transient spike in EtCO2 level ≥ 60 mmHg. *: P value <0.05; ns: P value is not significant. EtCO2 = end-tidal carbon dioxide.

Secondary

Patient-Reported or Nursing-Assessed Outcomes

Table 2 lists the percentages of pre- and post-procedure patient-reported or nursing-assessed outcomes in both groups. Although most of the pre- and post-procedure patient-reported symptoms were similar between groups, patient reported pre-procedure bloating was higher in the air group when compared to the CO2 group (9% vs 2%, P = 0.02). Additionally, more patients in the air group reported flatulence post-procedure (42% vs 20%, P = 0.004). Furthermore, the percent change in flatulence from pre- to post-procedure remained significantly higher in the air group (33.8%) when compared to the CO2 group (17.5%) with P= 0.034 for all procedures and 42.6% (air) versus 19.4% (CO2) for EGD/colonoscopy procedures with P = 0.013. There was no significant change in flatulence between groups from pre- to post-procedure in EGD-only procedures. Patient reported pre-procedure abdominal pain was associated with post-procedure abdominal pain but was not different between groups.

TABLE 2.

Patient reported clinical outcomes (pre- and post-procedure)

Post-procedure patient reported outcomes and hypercapnia
% procedures
CO2 Air P value
Belching pre-procedure (%)
 All procedures 5 1 0.38
 EGD only 0 0
 EGD + colonoscopy 6 2 0.37
Belching post-procedure (%)
 All procedures 14 12 0.81
 EGD only 0 0
 EGD + colonoscopy 18 15 0.80
Bloating pre-procedure (%)
 All procedures 2 13 0.02
 EGD only 0 0
 EGD + colonoscopy 3 16 0.02
Bloating post-procedure (%)
 All procedures 11 16 0.47
 EGD only 6 0 1
 EGD + colonoscopy 13 20 0.32
Flatulence pre-procedure (%)
 All procedures 4 9 0.3
 EGD only 6 0 1
 EGD+ colonoscopy 3 11 0.14
Flatulence post-procedure (%)
 All procedures 20 42 0.004
 EGD only 19 0 0.23
 EGD+ colonoscopy 21 53 0.0005
Nausea pre-procedure (%)
 All procedures 13 6 0.15
 EGD only 7 10 1
 EGD + colonoscopy 16 5 0.05
Nausea post-procedure (%)
 All procedures 7 8 1
 EGD only 4 3 1
 EGD + colonoscopy 7 9 0.76
Emesis pre-procedure (%)
 All procedures 4 2 0.68
 EGD only 3 3 1
 EGD + colonoscopy 3 2 1
Emesis post-procedure (%)
 All procedures 1 1 1
 EGD only 3 0 0.48
 EGD + colonoscopy 0 2 0.48
Headache pre-procedure (%)
 All procedures 7 5 0.77
 EGD only 7 3 0.61
 EGD + colonoscopy 7 6 1
Headache post-procedure (%)
 All procedures 13 13 1
 EGD only 7 17 0.42
 EGD + colonoscopy 16 11 0.46
Abdominal pain pre-procedure (%)
 All procedures 24 23 1
 EGD only 13 21 0.65
 EGD + colonoscopy 27 24 0.83
Abdominal pain post-procedure (%)
 All procedures 19 21 0.84
 EGD only 13 0 0.48
 EGD + colonoscopy 19 26 0.5
Hypercapnia (EtCO2) ≥ 60 mmHg %
 All procedures 21 8 0.016
 EGD only 16 5 0.02
 EGD + colonoscopy 9 5 0.33

EGD = esophagogastroduodenoscopy; EtCO2 = end-tidal carbon dioxide.

Procedure and Anesthesia-Related Outcomes

There were no procedural complications during this study. Anesthesia-related complications were minimal and transient and were similar in both groups. A few patients (3 in the CO2 group and 2 in the air group) experienced hypotension, 2 patients in both groups had hypertension, and 1 patient in the air group developed tachypnea. Procedure duration was similar in both groups.

DISCUSSION

The findings from this clinical trial add to existing knowledge and fill a gap created by earlier pediatric studies of endoscopic insufflation using CO2 (611). Our previous study (8) revealed that transient EtCO2 spikes (≥60 mmHg) are seen when CO2 is used for insufflation during pediatric upper endoscopy. We now show that these also occur with the use of a protected airway, although more common with LMA than ETT. Observed peaks in EtCO2 were not associated with changes in MV or adverse events. The only observed benefit of CO2 use was decreased post-procedure flatulence.

Although EtCO2 has been shown to be unreliable in the detection of changes in ventilation in nonintubated patients (1416), it provides an accurate estimate of ventilation in intubated patients (17) during routine anesthesia. Nevertheless, the use of CO2 as an insufflation gas during endoscopy is an additional variable that must be considered during EtCO2 monitoring. The LMA has been shown to be as effective for ventilation and gas exchange as the ETT for healthy children undergoing EGD (18). Therefore, we theorized that protecting the airway with either an LMA or ETT should minimize the possibility of contamination of sampled end-tidal gas by eructated CO2 as we had proposed to be the cause of elevated EtCO2 with CO2 endoscopic insufflation (7,8).

The persistence of brief episodes of elevated EtCO2 despite use of airway protection in this study needs explanation. Reduced ventilation of the patients was likely not responsible, as MV was not diminished prior to events. Although we did not monitor systemic absorption in this study, our prior published work (7) showed no systemic hypercarbia, measured by TCO2 monitoring, in association with increased EtCO2 events. Thus, it is possible that there is no significant absorption of CO2 from the upper gut, or that it is so gradual as to not be detectible by that technique. Another possible explanation is that some of the insufflation gas may leak into the trachea despite the presence of an ETT or LMA. Since fraction of CO2 gas in the exhaled breath is very small, any small trace contamination from exogenous source (leakage of insufflated CO2) causes significant elevation in ETCO2. The brevity (<1 minute) of these events, and the apparent higher incidence of EtCO2 elevations in the LMA group, compared to the ETT group, suggests this hypothesis. It is certainly possible that this device is more likely than the ETT to allow eructated CO2 to enter the lower airway. However, the selection of ETT versus LMA was not randomized, and further studies comparing ETT versus LMA, would be required to confirm or refute this hypothesis. Regardless of the actual mechanism, it is clear that transient elevations of EtCO2 occur during CO2 insufflation. Fortunately, no patient in either group experienced any adverse events associated with the use of CO2.

One can question that, even in the absence of adverse effects, are the benefits of CO2 insufflation worth any associated respiratory risk? Adult studies report improved patient comfort following CO2 insufflation for a variety of procedures (1,1921), including upper endoscopy (21). Many of these studies found only subjectively reported benefits of CO2, such as decreased patient-reported bloating or discomfort, without objective findings by observers (22). Findings from pediatric studies are not consistent. Some reported decreased patient complaints of discomfort with CO2 insufflation for colonoscopy (6,9,11). Kresz et al (10) found a higher need for additional narcotics in the air group during the procedure, a highly unusual occurrence in our practice, regardless of the insufflation gas used. Dharmaraj et al (6) only found a decreased post-procedural abdominal discomfort in the CO2 group on multivariate analysis after adjusting for a pre-operative diagnosis of abdominal pain. In this study, we did not find a difference in patient-reported or nursing-assessed abdominal pain, even in patients who underwent consecutive EGD/colonoscopy, and after adjusting for a pre-operative diagnosis of abdominal pain. Although patients in the air group reported more flatulence after the procedure than the CO2 group, they also reported more bloating pre-procedure.

CONCLUSIONS

In conclusion, we observed a significantly higher number of brief EtCO2 spikes in children receiving CO2 instead of air for endoscopic insufflation, a difference that occurs even with airway protection, and more commonly with LMAs than with ETTs (8). These EtCO2 spikes were not caused by reduced MV and were not physiologically significant, given the lack of any associated adverse events. In addition, our previous study (7) showed that these transient spikes of EtCO2 were not associated with any detectable elevation of transcutaneous (systemic) CO2 levels. Nevertheless, given the lack of substantial observed benefits of CO2 insufflation for upper endoscopy in children, the increased costs of its use (23), and the possible safety concerns suggested by transient spikes in EtCO2 levels of ≥60 mmHg, the routine use of CO2 for all pediatric upper endoscopy procedures must be questioned. For more complex or prolonged procedures, such as gastrostomy placement, enteroscopy, or use of electrocautery, its potential benefits may be more substantial. In these cases, minimizing the amount of peritoneal gas accumulation or the risk spark-induced combustion may be significant.

Acknowledgments

We would like to thank the pediatric gastroenterology faculty and fellows, pediatric anesthesiology faculty and fellows, pediatric endoscopy suite staff, and recovery nurses at Children’s Hospital and Medical Center particularly Nancy Nielsen RN who led the study randomization, the study research coordinators: Evan Roberts MBA, Kym Abraham RN, Denise Hoover BS (Pediatric Research Office), data management team: Alicia Bremer MSN APRN and Cynthia Hug BS (Children’s Hospital and Medical Center Omaha), and Elizabeth Patterson GI administrative assistant for all their help with this study.

Footnotes

The authors report no conflicts of interest.

W.P.B. and M.S. are co-senior authors.

This study is a double-blinded randomized clinical trial and was registered with clinical trials.gov (NCT04541667).

Sources of Funding: This study was funded by the University of Nebraska Medical Center Diversity Grant (Dike: PI).

REFERENCES

  • 1.Chen SW, Hui CK, Chang JJ, et al. Carbon dioxide insufflation during colonoscopy can significantly decrease post-interventional abdominal discomfort in deeply sedated patients: a prospective, randomized, double-blinded, controlled trial. J Gastroenterol Hepatol 2016;31:808–13. [DOI] [PubMed] [Google Scholar]
  • 2.Lenz P, Meister T, Manno M, et al. CO2 insufflation during single-balloon enteroscopy: a multicenter randomized controlled trial. Endoscopy 2014;46:53–8. [DOI] [PubMed] [Google Scholar]
  • 3.Sajid MS, Caswell J, Bhatti MI, Sains P, Baig MK, Miles WFA. Carbon dioxide insufflation vs conventional air insufflation for colonoscopy: a systematic review and meta-analysis of published randomized controlled trials. Colorectal Dis 2015;17:111–23. [DOI] [PubMed] [Google Scholar]
  • 4.Seo EH, Kim TO, Park MJ, et al. The efficacy and safety of carbon dioxide insufflation during colonoscopy with consecutive esophagogastroduodenoscopy in moderately sedated outpatients: a randomized, double-blind, controlled trial. J Clin Gastroenterol 2013;47:e45–9. [DOI] [PubMed] [Google Scholar]
  • 5.Shi H, Chen S, Swar G, Wang YG, Ying MG. Carbon dioxide insufflation during endoscopic retrograde cholangiopancreatography: a review and meta-analysis. Pancreas 2013;42:1093–100. [DOI] [PubMed] [Google Scholar]
  • 6.Dharmaraj R, Dunn R, Fritz J, et al. Efficacy and safety of carbon dioxide versus air insufflation for colonoscopy in deeply sedated pediatric patients. J Pediatr Gastroenterol Nutr 2020;71:34–9. [DOI] [PubMed] [Google Scholar]
  • 7.Dike CR, Bishop WP, Titler SS, et al. Transient end-tidal carbon dioxide elevation during pediatric upper endoscopy with carbon dioxide insufflation: is it true hypercapnia? J Pediatr Gastroenterol Nutr 2022;74:413–18. [DOI] [PubMed] [Google Scholar]
  • 8.Dike CR, Rahhal R, Bishop WP. Is carbon dioxide insufflation during endoscopy in children as safe and as effective as we think? J Pediatr Gastroenterol Nutr 2020;71:211–5. [DOI] [PubMed] [Google Scholar]
  • 9.Homan M, Mahkovic D, Orel R, Mamula P. Randomized, double-blind trial of CO2 versus air insufflation in children undergoing colonoscopy. Gastrointest Endosc 2016;83:993–7. [DOI] [PubMed] [Google Scholar]
  • 10.Kresz A, Mayer B, Zernickel M, et al. Carbon dioxide versus room air for colonoscopy in deeply sedated pediatric patients: a randomized controlled trial. Endosc Int Open 2019;7:E290–e97. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Thornhill C, Navarro F, Alabd Alrazzak B, et al. Insufflation with carbon dioxide during pediatric colonoscopy for control of postprocedure pain. J Clin Gastroenterol 2018;52:715–20. [DOI] [PubMed] [Google Scholar]
  • 12.Ji C, Liu X, Huang P. Carbon dioxide vs. air insufflation for pediatric gastrointestinal endoscopy: a systematic review and meta-analysis of randomized controlled trials. Front Pediatr 2021;9:610066. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Guacho JAL, Moura DTH, Ribeiro IB, et al. Insufflation of carbon dioxide versus air during colonoscopy among pediatric patients: a systematic review and meta-analysis of randomized controlled trials. Clin Endosc 2021;54:242–49. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Jabre P, Jacob L, Auger H, et al. Capnography monitoring in nonintubated patients with respiratory distress. Am J Emerg Med 2009;27:1056–9. [DOI] [PubMed] [Google Scholar]
  • 15.Mehta JH, Williams GW, 2nd, Harvey BC, Grewal NK, George EE. The relationship between minute ventilation and end tidal CO2 in intubated and spontaneously breathing patients undergoing procedural sedation. PLoS One 2017;12:e0180187. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Mehta PP, Kochhar G, Albeldawi M, et al. Capnographic monitoring in routine EGD and colonoscopy with moderate sedation: a prospective, randomized, controlled trial. Am J Gastroenterol 2016;111:395–404. [DOI] [PubMed] [Google Scholar]
  • 17.Razi E, Moosavi GA, Omidi K, Saebi A, Razi A. Correlation of end-tidal carbon dioxide with arterial carbon dioxide in mechanically ventilated patients. Arch Trauma Res 2012;1:58–62. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Acquaviva MA, Horn ND, Gupta SK. Endotracheal intubation versus laryngeal mask airway for esophagogastroduodenoscopy in children. J Pediatr Gastroenterol Nutr 2014;59:54–6. [DOI] [PubMed] [Google Scholar]
  • 19.Aquino JCM, Bernardo WM, de Moura DTH, et al. Carbon dioxide versus air insufflation enteroscopy: a systematic review and meta-analysis based on randomized controlled trials. Endosc Int Open 2018;6:E637–e45. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Dellon ES, Hawk JS, Grimm IS, Shaheen NJ. The use of carbon dioxide for insufflation during GI endoscopy: a systematic review. Gastrointest Endosc 2009;69:843–9. [DOI] [PubMed] [Google Scholar]
  • 21.Jowhari F, Hookey L. Gastroscopy should come before colonoscopy using CO2 insufflation in same day bidirectional endoscopies: a randomized controlled trial. J Can Assoc Gastroenterol 2019;3:120–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Wideman TH, Edwards RR, Walton DM, Martel MO, Hudon A, Seminowicz DA. The multimodal assessment model of pain: a novel framework for further integrating the subjective pain experience within research and practice. Clin J Pain 2019;35:212–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Lo SK, Fujii-Lau LL, Enestvedt BK, et al. The use of carbon dioxide in gastrointestinal endoscopy. Gastrointest Endosc 2016;83:857–65. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Pediatric Gastroenterology and Nutrition are provided here courtesy of Wolters Kluwer Health

RESOURCES