Tandem Stenting of Crossed Renal Arteries with Ostial Stenosis

Marcus H Howell 1, Neil E Strickman 1
Editor: Zvonimir Krajcer1
PMCID: PMC101052  PMID: 10928505

Abstract

Ostial stenosis of renal arteries has been shown to respond favorably to stenting. Heterogeneity of renal artery anatomy is very common and often increases the technical difficulty of percutaneous treatment of stenotic lesions. We present the case of a 69-year-old man with severe hypertension who had 2 right renal arteries, both of which had ostial stenosis. We treated these stenoses with simultaneous stenting of both renal ostia. Since the procedure, the patient has had significantly improved control of his hypertension.

Key words: Angioplasty, balloon/methods; blood vessel prosthesis implantation; hypertension, renovascular/therapy; renal artery/anatomy; renal artery obstruction; stents

Atherosclerotic renal artery stenosis has been shown to be a significant cause of systemic arterial hypertension and renal insufficiency. Left untreated, most stenoses will progress, possibly resulting in ischemic nephropathy and renal failure. At the Fifth Biennial International Andreas Gruentzig Society Meeting, Dr. Richard Stack 1 reported the Duke University experience. He noted that over a 30-month period, the disease progressed in 22% of patients who had renal artery stenosis. Over the last few years, multiple prospective and retrospective studies have shown that percutaneous balloon angioplasty and stenting of stenotic renal artery lesions improves blood pressure control and may delay or stabilize renal insufficiency. 2–5

Heterogeneous anatomy of renal arteries is well described; up to 25% of kidneys have multiple arterial sources. 6 This heterogeneity can often increase the technical difficulty of percutaneous treatment of renal artery stenosis. We report the percutaneous stenting of a patient who had 2 crossed right renal arteries, both of which had calcific ostial stenosis.

Case Report

In June 1998, a 69-year-old man presented with recurrent chest tightness associated with exertion. His medical history revealed coronary and cerebral atherosclerotic vascular disease and severe hypertension. He had undergone a right carotid endar-terectomy in 1992 and a 2-vessel aortocoronary bypass in 1993. He was on 2 blood pressure medications, a beta-blocker and an angiotensin-converting enzyme inhibitor, at moderate doses, and he was taking a nitrate for anginal symptoms. Laboratory analysis revealed normal complete blood count, blood urea nitrogen (16 mg/dL; normal, 8 to 20 mg/dL), serum creatinine (1.2 mg/dL; normal, 0.5 to 1.7 mg/dL), and normal chemistries. The patient's blood pressure at admission and again in the catheterization lab was measured at 190/100 mmHg in both arms.

Cardiac catheterization revealed a 90% stenosis in the saphenous vein graft to the right coronary artery; this was treated with angioplasty and stenting. Angiography of the renal arteries revealed a single left renal artery and 2 crossed right renal arteries. The cranial right renal artery (CRN), which supplied the lower pole, had a gradient of 25 mmHg, and the caudal right renal artery (CUD), which supplied the upper pole, had a gradient of 40 mmHg (Fig. 1). The left renal artery had no significant stenosis. Due to these findings, the patient was scheduled for angioplasty and stenting.

graphic file with name 15FF1.jpg

Fig. 1 A renal arteriogram reveals 2 separate renal arteries that supply the right kidney. The cranial right renal artery (CRN) supplies the lower pole, and the caudal right renal artery (CUD) supplies the upper pole. The arteriogram displays 80% ostial stenosis in both right renal arteries (arrows).

At admission four weeks later, despite maximal doses of 3 antihypertensive agents, the patient's blood pressure remained high at 200/90 mmHg. In the cardiac catheterization laboratory, arterial access was obtained and an intravenous dose of 7,500 units of unfractionated heparin was administered. The cranial-lower pole (CRN) and caudal-upper pole (CUD) right renal arteries were engaged with 8-F Renal Double Curve guides (Cordis Corporation, a Johnson & Johnson company; Warren, NJ) via, respectively, the right and left femoral arteries. Both ostial stenoses were crossed with 0.018-inch Roadrunner® wire guides (Cook Inc.; Bloomington, Ind). The cranial right renal artery (CRN) stenosis was dilated with a 5- × 20-mm Ultra-Thin™ Diamond™ (UTD) balloon (Boston Scientific Corp.; Natick, Mass) at 8 atmospheres (ATM) for 120 seconds, followed by dilation of the caudal-upper pole (CUD) obstruction with a 6- × 20-mm UTD balloon at 6 ATM for 120 seconds. Initial inflations in the CUD caused it to compress the ostium of the CRN. Therefore, we used the “kissing balloon” technique to re-dilate the caudal lesion with the 5- × 20-mm UTD balloon at 2 ATM for 41 seconds while simultaneously dilating the cranial lesion at 8 ATM for 41 seconds (Fig. 2).

graphic file with name 15FF2.jpg

Fig. 2 The “kissing balloon” technique is demonstrated here: simultaneous inflation of 2 balloons (arrows) across the ostia of the CRN and CUD.

CRN = cranial right renal artery; CUD = caudal right renal artery

In order to avoid crushing the kissing ostium, we intended to dilate 1 lesion at maximal pressure while dilating the companion lesion at minimal pressure. A Palmaz® P154 stent (Cordis) mounted on a 6- × 20-mm UTD balloon was placed across the CRN ostium and deployed at 6 ATM for 50 seconds, while at the same time the CUD was protected using the kissing balloon technique by inflating a 6- × 20-mm UTD balloon at 2 ATM for 50 seconds. The same technique was used in reverse for the introduction of a Palmaz P154 stent into the CUD ostium while the cranial stent was protected from crush injury by in-flation of the CRN stent.

This approach did not succeed in preventing crushing of the CRN ostium (Fig. 3). It became necessary to cross the crushed CRN ostium with a Titan™ 4.0- × 18-mm coronary balloon (Cordis), because the crushed and severely disfigured stent would barely accommodate a 0.018-inch wire through its small opening. The Titan balloon was inflated to 20 ATM for 120 seconds, and the CUD was simultaneously di-lated with a 6- × 20-mm UTD balloon (Fig. 4). This achieved excellent results. Angiography showed that the gradients in both the cranial and caudal right renal arteries had resolved (Fig. 5). By the time the patient was discharged the next day, his blood pressure had dropped to 120/70 mmHg; however, he still required 2 antihypertensive medications.

graphic file with name 15FF3.jpg

Fig. 3 The stent in the ostium of the cranial right renal artery (CRN) appears crushed after balloon angioplasty (arrows).

graphic file with name 15FF4.jpg

Fig. 4 Simultaneous inflation of both renal artery ostia to alleviate stent crushing. A coronary system (4.0- × 18-mm Titan™ balloon over a 0.018-mm Roadrunner® wire guide) was necessary for entering the cranial right renal artery; a standard peripheral balloon (6- × 20-mm Ultra-Thin™ Diamond™) was used in the caudal right renal artery.

UTD = Ultra-Thin™ Diamond™

graphic file with name 15FF5.jpg

Fig. 5 Final angiographic appearance of both renal artery ostia after angioplasty and stent implantation.

CRN = cranial right renal artery; CUD = caudal right renal artery

The patient underwent follow-up coronary and renal angiography in September 1999, when he presented with a complaint of angina. Both stents remained highly patent, despite minor luminal narrowing primarily in the CRN (Fig. 6). The patient's blood pressure was mildly elevated at 160/75 mmHg, and he was still taking 2 antihypertensive medications.

graphic file with name 15FF6.jpg

Fig. 6 Angiographic follow-up at 1 year demonstrates no significant ostial stenosis (black arrows).

CRN = cranial right renal artery; CUD = caudal right renal artery

Discussion

The diagnosis and treatment of renal artery stenosis is particularly important in patients who have coronary artery disease. Harding and colleagues 7 showed that the prevalence of renal artery stenosis is 30% in these patients. There is a direct relationship between the severity of coronary disease and the presence of significant renal artery stenosis. In Harding's study, patients with 1-vessel disease had a 10% incidence of renal artery stenosis; those with 2-vessel disease had a 20% incidence; and those with 3-vessel disease had a 30% incidence. Of patients with left main disease, 40% were found to have some degree of renal artery stenosis. Richard Stack 1 also found that the survival rate was significantly worse for patients who had coronary artery disease if they had concomitant renal artery stenosis greater than 50%. Even when adjusted for the standard prognostic factors, renal artery stenosis was an independent risk factor for death in patients with coronary artery disease.

The 1st percutaneous renal angioplasty was performed on 12 December 1977. 8 Long-term follow-up studies in the early 1980s showed that this procedure could be done safely and could lead to significant improvements in blood pressure. 9,10 These studies reported the following results in regard to blood pressure: cure (normotensive on no medication) in 10%; improvement (15 mmHg decrease on the same or lower level of medication) in 55%; and no benefit in 35%. Later studies have confirmed this blood pressure benefit. Approximately 50% of patients with renal artery stenosis who undergo balloon angioplasty show improved or stabilized renal function afterwards. 11–13

Stenting of renal artery stenosis has been performed since 1986. Reporting 5-year patency rates of 84% and secondary patency rates after reintervention greater than 90%, Blum and coworkers 14 showed that stenting for ostial renal artery stenosis results in greater success rates than angioplasty alone. Numerous other studies have shown the same superior angiographic success rates and similar or slightly improved clinical success rates. 3,4,15–17

Ho and associates 18 described the successful use of the kissing balloon technique for stent placement in a bifurcating renal artery lesion that had responded poorly to balloon angioplasty. Because of this report and our experience with coronary artery interventions at bifurcations, we decided to use the kissing balloon technique for deployment of stents in our pa-tient's complex renal artery lesion.

Conclusion

Renal artery stenting has proved superior to angioplasty alone for renal artery stenosis, especially for ostial lesions. Stenting improves early technical success and reduces long-term restenosis. We have shown that the kissing balloon technique can be successful for ostial stenting, even in the presence of complex anatomy of the renal arteries.

Despite the lack of prospective randomized control studies, there is good evidence that percutaneous stent treatment of renal artery stenosis provides clinical benefit, especially in patients with uncontrolled hypertension, coronary artery disease, congestive heart failure, and renal insufficiency.

Footnotes

Address for reprints: Marcus Howell, MD, 6720 Bertner Avenue, MC 1–133, Houston, TX 77030

References

  • 1.Stack R. Renal artery stenosis: under-diagnosed and under-treated in the cardiac patient? J Invas Cardiol 1999;11:103–6. [PubMed]
  • 2.van de Ven PJ, Kaatee R, Beutler JJ, Beek FJ, Woittiez AJ, Buskens E, et al. Arterial stenting and balloon angioplasty in ostial atherosclerotic renovascular disease: a randomised trial. Lancet 1999;353:282–6. [DOI] [PubMed]
  • 3.Henry M, Amor M, Henry I, Ethevenot G, Tzvetanov K, Courvoisier A, et al. Stents in the treatment of renal artery stenosis: long-term follow-up. J Endovasc Surg 1999;6:42–51. [DOI] [PubMed]
  • 4.Iannone LA, Underwood PL, Nath A, Tannenbaum MA, Ghali MG, Clevenger LD. Effect of primary balloon expandable renal artery stents on long-term patency, renal function, and blood pressure in hypertensive and renal insufficient patients with renal artery stenosis. Cathet Cardiovasc Diagn 1996;37:243–50. [DOI] [PubMed]
  • 5.Dorros G, Jaff M, Jain A, Dufek C, Mathiak L. Follow-up of primary Palmaz-Schatz stent placement for atherosclerotic renal artery stenosis. Am J Cardiol 1995;75:1051–5. [DOI] [PubMed]
  • 6.Skandalakis JE, Gray SW. Embryology for surgeons: the embryological basis for the treatment of congenital anomalies. 2nd edition. Baltimore: Williams & Wilkins, 1994: 623–5.
  • 7.Harding MB, Smith LR, Himmelstein SI, Harrison K, Philips HR, Schawb SJ, et al. Renal artery stenosis: prevalence and associated risk factors in patients undergoing routine cardiac catheterization. J Am Soc Nephrol 1992;2: 1608–16. [DOI] [PubMed]
  • 8.Mahler F, Krneta A, Haertel M. Treatment of renovascular hypertension by transluminal renal artery dilatation. Ann Intern Med 1979;90:56–7. [DOI] [PubMed]
  • 9.Mahler F, Probst P, Haertel M, Weidmann P, Krneta A. Lasting improvement of renovascular hypertension by transluminal dilatation of atherosclerotic and nonatherosclerotic renal artery stenoses. A follow-up study. Circulation 1982;65:611–7. [DOI] [PubMed]
  • 10.Sos TA, Pickering TG, Sniderman K, Saddekni S, Case DB, Silane MF, et al. Percutaneous transluminal renal angioplasty in renovascular hypertension due to atheroma or fibromuscular dysplasia. N Engl J Med 1983;309:274–9. [DOI] [PubMed]
  • 11.Weibull H, Bergqvist D, Jonsson K, Hulthen L, Mannhem P, Bergentz SE. Long-term results after percutaneous transluminal angioplasty of atherosclerotic renal artery stenosis—the importance of intensive follow-up. Eur J Vasc Surg 1991;5:291–301. [DOI] [PubMed]
  • 12.Weibull H, Bergqvist D, Bergentz SE, Jonsson K, Hulthen L, Manhem P. Percutaneous transluminal renal angioplasty versus surgical reconstruction of atherosclerotic renal artery stenosis: a prospective randomized study. J Vasc Surg 1993; 18:841–52. [DOI] [PubMed]
  • 13.Jensen G, Zachrisson BF, Delin K, Volkmann R, Aurell M. Treatment of renovascular hypertension: one year results of renal angioplasty. Kidney Int 1995;48:1936–45. [DOI] [PubMed]
  • 14.Blum U, Krumme B, Flugel P, Gabelmann A, Lehnert T, Buitrago-Tellez C, et al. Treatment of ostial renal-artery stenoses with vascular endoprostheses after unsuccessful balloon angioplasty. N Engl J Med 1997;336:459–65. [DOI] [PubMed]
  • 15.Gross CM, Kramer J, Waigand J, Uhlich F, Olthoff H, Luft FC, et al. Ostial renal artery stent placement for atherosclerotic renal artery stenosis in patients with coronary artery disease. Cathet Cardiovasc Diagn 1998;45:1–8. [DOI] [PubMed]
  • 16.White CJ, Ramee SR, Collins TJ, Jenkins JS, Escobar A, Shaw D. Renal artery stent placement: utility in lesions difficult to treat with balloon angioplasty. J Am Coll Cardiol 1997;30:1445–50. [DOI] [PubMed]
  • 17.Dorros G, Prince C, Mathiak L. Stenting of a renal artery stenosis achieves better relief of the obstructive lesion than balloon angioplasty. Cathet Cardiovasc Diagn 1993;29:191–8. [DOI] [PubMed]
  • 18.Ho DS, Chen WH, Woo C. Stenting of a renal artery bi-furcation stenosis. Cathet Cardiovasc Diagn 1998;45:445–9. [DOI] [PubMed]