Multicenter clinical evaluation of a novel transcription-mediated amplification assay for SARS-CoV-2 molecular testing

Evaluación clínica multicéntrica de un nuevo ensayo de amplificación mediada por transcripción para la detección molecular de SARS-CoV-2

Miguel Fernández-Huerta a, Paula Salmerón b, Yolanda Hernández-Hermida a, Cristina Andrés b, Jordi Niubó a, Laura Calatayud a,c, M Ángeles Domínguez a,d,e, Tomàs Pumarola b, Carmen Ardanuy a,c,e, Andrés Antón b, Jordi Càmara a,c,*
PMCID: PMC10107077  PMID: 37076332

Abstract

Introduction

The onset and spread of COVID-19 pandemic has forced clinical laboratories to rapidly expand testing capacity for SARS-CoV-2. This study evaluates the clinical performance of the TMA Procleix SARS-CoV-2 assay in comparison to the RT-PCR assay Allplex™ SARS-CoV-2 for the qualitative detection of SARS-CoV-2 RNA.

Methods

Between November 2020 and February 2021, 610 upper-respiratory specimens received for routine SARS-CoV-2 molecular testing were prospectively collected and selected at the Hospital Universitari Vall d’Hebron and the Hospital Universitari Bellvitge in Barcelona, Spain. All samples were processed in parallel with the TMA and the RT-PCR assays, and results were compared. Discrepancies were retested by an additional RT-PCR method and the clinical history of these patients was reviewed.

Results

Overall, the level of concordance between both assays was 92.0% (κ, 0.772). Most discordant results (36/38, 94.7%) corresponded to samples testing positive with the TMA assay and negative with the RT-PCR method. Of these discrepant cases, most (28/36, 77.8%) were finally classified as confirmed or probable SARS-CoV-2 cases according to the discrepant analysis.

Conclusion

In conclusion, the TMA Procleix SARS-CoV-2 assay performed well for the qualitative detection of SARS-CoV-2 RNA in a multisite clinical setting. This novel TMA assay demonstrated a greater sensitivity in comparison to RT-PCR methods for the molecular detection of SARS-CoV-2. This higher sensitivity but also the qualitative feature of this detection of SARS-CoV-2 should be considered when making testing algorithm decisions.

Keywords: SARS-CoV-2, COVID-19, Transcription-mediated amplification

Introduction

The onset and rapid spread of COVID-19 pandemic has forced clinical laboratories to rapidly expand testing capacity for SARS-CoV-2.1, 2 As quarantine measures are gradually being relaxed and novel disturbing SARS-CoV-2 variants are emerging, there is potential risk for upsurge in cases and rates of viral transmission. Accurate and sensitive viral detection methods with high-throughput diagnostic capability are essential for rapid clinical decision-making, implementing infection prevention practices and guiding public health responses. In this regard, numerous nucleic acid amplification tests (NAATs), granted with emergency-use authorization (EUA) by the Food and Drug Administration (FDA), are available for the molecular detection of SARS-CoV-2 in upper respiratory specimens.3

Most SARS-CoV-2 NAATs are based on real-time reverse transcription PCR (RT-PCR) techniques, such as the RT-PCR assay Allplex™ SARS-CoV-2 (Seegene, Inc., South Korea). Nevertheless, in accordance with the need for pandemic-scale diagnostic testing, transcription-mediated amplification (TMA) methods like the TMA Procleix SARS-CoV-2 assay (Grifols International, S.A., Spain) also emerged as part of the diagnostic arsenal against COVID-19.4

If appropriately optimized, NAATs are highly sensitive and specific. Nevertheless, in real-life clinical performance several preanalytical and analytical issues may occasionally affect sensitivity and specificity of NAATs. Thus, the quick communication of these experiences is urgently required to better understand the potential benefits and limitations of the implementation of each technique into SARS-CoV-2 testing strategies.

Herein, we evaluate the clinical performance of the TMA Procleix SARS-CoV-2 assay in comparison to the RT-PCR assay Allplex™ SARS-CoV-2 for the qualitative detection of SARS-CoV-2 RNA. A comprehensive analysis of discrepant results between both methods is conducted to assess the relevance of these cases in the management of SARS-CoV-2 infected individuals.

Methods

Sample selection and laboratory procedures

Between November 2020 and February 2021, 154 and 456 upper-respiratory specimens received for routine SARS-CoV-2 molecular testing were prospectively collected and randomly selected for the current evaluation at the Hospital Universitari Vall d’Hebron (HUVH) and the Hospital Universitari Bellvitge (HUB) in Barcelona, Spain; respectively. The selection included nasopharyngeal and nasal swabs, placed in viral transport medium (VTM) from various manufacturers, collected from symptomatic and asymptomatic individuals, with no bias towards age or gender.

The detailed methodology for the current evaluation was conducted as follows (Fig. S1). VTM aliquots from selected fresh specimens were manually transferred into sterile tubes containing lysis buffer in a 1:1 proportion. If available, surplus specimen material was stored at −80 for subsequent analysis. SARS-CoV-2 molecular testing was first performed with the TMA Procleix SARS-CoV-2 assay. Using the same inactivated aliquot, nucleic acid extraction was performed with automated systems recommended by the manufacturer and SARS-CoV-2 detection was sequentially executed using the RT-PCR assay Allplex™ SARS-CoV-2.

In case of discrepant results, stored samples were thawed, inactivated and processed with an additional RT-PCR technique: the Alinity m SARS-CoV-2 assay (Abbott Molecular, US) or the RT-PCR cobas® SARS-CoV-2 test (Roche Molecular Systems, US). Also, discrepant cases were further analyzed through an exhaustive review of the medical record.

All technical procedures were performed as described by the manufacturers and results were interpreted according to the manufacturer's criteria.

Description of the TMA Procleix SARS-CoV-2 assay

The TMA Procleix SARS-CoV-2 assay is a novel multiplex NAAT based on TMA methodology for the qualitative molecular detection of SARS-CoV-2 virus.4 The assay, EUA granted by the FDA, incorporates an exogenous internal control, and specifically targets the viral N gene. The test is ready to be run on the fully automated all-in-one Procleix Panther system (Grifols International, S.A.) using 750 μL material from upper-respiratory swabs. In this regard, the assay provides a ratio value that corresponds to an analytical parameter based on chemiluminescense with a qualitative interpretation. Values ≥1.00 are considered reactive/positive, according to the manufacturer's criteria.

Data analyses

Statistical analyses were performed using GraphPad Prism version 8.0.1 (GraphPad Software Inc., USA). The Cohen's kappa statistic (κ) was used to evaluate the agreement between both assays, and 95% confidence intervals (CI) were calculated by exact methods.

Minor disagreement was defined as a positive test with one assay and an inconclusive test with the other. In the current comparison, only the RT-PCR assay Allplex™ SARS-CoV-2 could yield inconclusive results, defined by the manufacturer as those cases testing positive for the E target, but negative for both the N and the RdRP/S targets. Major disagreement was considered that reverting from positive to negative depending on the study assay. If available, discrepant specimens were retested with an additional RT-PCR method and managed as follows. A confirmed case of SARS-CoV-2 was considered that discrepancy testing positive with the additional RT-PCR assay. A probable case was considered that testing negative with the additional RT-PCR technique, but having epidemiological or clinical data suggestive and compatible with recent history of SARS-CoV-2 infection. Unresolved cases were reported as inconclusive.

Institutional Review Board approval (PR(AG)259/2020) was obtained from the HUVH Clinical Research Ethics Committee.

Results

The qualitative comparison of the 610 study samples is displayed in Table 1 . Overall, the level of concordance between both assays was 92.0% (95% CI, 89.5–94.0%), with a κ value of 0.772 (0.715–0.830). Of the 38 major discrepancies, 36 (94.7%) corresponded to samples testing positive with the TMA Procleix SARS-CoV-2 assay and negative with the RT-PCR assay Allplex™ SARS-CoV-2.

Table 1.

Comparison between TMA and RT-PCR assays (n = 610).

TMA
Procleix SARS-CoV-2
RT-PCR
Allplex™ SARS-CoV-2
Level of concordance
% (95% CI)
Kappa value
(95% CI)
Positive Inconclusiveb Negative
Positivea 106 11 36 92.0 (89.5–94.0) 0.772 (0.715–0.830)
Negative 2 0 455
a

The instrument provides a ratio value that corresponds to an analytical parameter based on chemiluminescense with a qualitative interpretation. Values ≥1.00 are considered reactive/positive, according to the manufacturer's criteria. In this regard, all positive results yielded ratio values ≥2.00.

b

Inconclusive cases are considered those testing positive for the E target, but negative for both the N and the RdRP/S targets with the RT-PCR Allplex™ SARS-CoV-2 assay, according to the manufacturer's criteria.

Abbreviations: TMA: transcription-mediated amplification; RT-PCR: real-time reverse-transcriptase PCR; CI: confidence interval.

A comprehensive analysis of discrepant results is detailed in Table 2A, Table 2B . While Table 2A focuses on the analytical approach of this evaluation, Table 2B provides an in-depth analysis of discrepant cases by incorporating some key epidemiological features from the medical record. The main conclusions of this assessment are summarized in Table 3 . Overall, from the 36 discrepancies mentioned above, 28 (77.8%) were finally classified as confirmed (11/36) or probable (17/36) SARS-CoV-2 cases. On the other hand, both major discrepancies testing positive with the RT-PCR assay Allplex™ SARS-CoV-2 but negative with the TMA Procleix SARS-CoV-2 assay remained inconclusive. Finally, eight of the 11 minor disagreements (72.7%), were at last reported as confirmed (6/11) or probable (2/11) SARS-CoV-2 cases.

Table 2A.

Analytical evaluation of discrepant results (n = 49).

Case TMA Procleix
RT-PCR Allplex™
Type of
disagreement
RT-PCR Alinity m
Analytical
result
Ratioa Ct
Result Ctc
Result N Result E RdRP N RdRP & N
1 P 4.18 N NA NA NA Major N NA N
2 N NA P NA NA 37.59 Major NDd ND
3 P 3.92 N NA NA NA Major P 40.99 P
4 P 3.81 N NA NA NA Major P 39.53 P
5 P 3.81 Ib 37.96 NA NA Minor P 40.94 P
6 P 4.00 N NA NA NA Major N NA N
7 P 4.24 N NA NA NA Major N NA N
8 P 4.24 Ib 38.02 NA NA Minor P 40.03 P
9 P 4.18 N NA NA NA Major NDd ND
10 P 4.20 Ib 36.91 NA NA Minor NDd ND
11 P 3.73 N NA NA NA Major N NA N
12 P 4.15 N NA NA NA Major P 40.90 P
13 P 4.36 N NA NA NA Major P 39.60 P
14 P 4.21 N NA NA NA Major P 37.21 P
15 P 4.16 N NA NA NA Major P 40.90 P
16 P 4.05 Ib 37.88 NA NA Minor NDd ND
17 N NA P NA NA 37.63 Major NDd ND
18 P 4.21 N NA NA NA Major N NA N
19 P 4.23 Ib 38.06 NA NA Minor P 40.99 P
20 P 4.50 N NA NA NA Major P 40.99 P
21 P 4.46 N NA NA NA Major N NA N
22 P 3.95 N NA NA NA Major N NA N
23 P 3.97 N NA NA NA Major N NA N
24 P 4.00 N NA NA NA Major N NA N
25 P 4.12 I 38.23 NA NA Minor P 40.99 P
Case TMA
Procleix SARS-CoV-2
RT-PCR
Allplex™ SARS-CoV-2
Type of disagreement RT-PCR cobas® SARS-CoV-2
Analytical result
Result Ratioa Result Ct
Result Ct
N E RdRP N ORF-1ab E
26 P 4.17 N NA NA NA Major N NA NA N
27 P 3.78 N NA NA NA Major Ib NA 35.63 Ib
28 P 3.84 N NA NA NA Major Ib NA 37.73 Ib
29 P 4.16 N NA NA NA Major N NA NA N
30 P 3.93 N NA NA NA Major P 32.4 34.98 P
31 P 3.90 N NA NA NA Major N NA NA N
32 P 4.22 N NA NA NA Major Ib NA 38.47 Ib
33 P 4.02 N NA NA NA Major N NA NA N
34 P 3.95 N NA NA NA Major N NA NA N
35 P 4.28 N NA NA NA Major P 36.19 37.22 P
36 P 3.92 Ib 36.58 NA NA Minor P 33.15 35.01 P
37 P 4.05 N NA NA NA Major N NA NA N
38 P 4.59 Ib 38.15 NA NA Minor P 33.35 34.51 P
39 P 4.03 N NA NA NA Major N NA NA N
40 P 4.25 N NA NA NA Major Ib NA 38.01 Ib
41 P 4.22 N NA NA NA Major P 35.58 35.87 P
42 P 3.99 N NA NA NA Major N NA NA N
43 P 3.84 N NA NA NA Major N NA NA N
44 P 4.04 Ib 38.18 NA NA Minor N NA NA Ib
45 P 4.46 Ib 38.26 NA NA Minor N NA NA Ib
46 P 4.04 N NA NA NA Major Ib NA 37.94 Ib
47 P 4.35 N NA NA NA Major P 30.80 31.60 P
48 P 4.15 Ib 38.14 NA NA Minor N NA NA Ib
49 P 4.33 N NA NA NA Major N NA NA N
a

The ratio provided by the instrument corresponds to an analytical parameter based on chemiluminescense with a qualitative interpretation. Values ≥1.00 are considered reactive/positive, according to the manufacturer's criteria.

b

Inconclusive cases are considered: (i) those testing positive for the E target, but negative for both the N and the RdRP/S targets with the RT-PCR Allplex™ SARS-CoV-2 assay, according to the manufacturer's criteria, (ii) those testing positive for the E target, but negative for the ORF-1ab target with the RT-PCR cobas® SARS-CoV-2 assay, according to the manufacturer's criteria, and (iii) those unresolved after the discrepant analytical testing.

c

The two SARS-CoV-2-specific probes are labelled with the same fluorophore, so the test provides a unique Ct value for both RdRP and N targets.

d

Specimens unavailable for discrepant analytical testing.

A minor disagreement is considered that discrepancy testing positive with one assay but inconclusive with the other study technique. Contrary, a major disagreement is considered that reverting from positive to negative depending on the study assay.

Abbreviations: TMA: transcription-mediated amplification; RT-PCR: real-time reverse-transcriptase PCR; Ct: cycle-threshold; P: positive; N: negative; I: inconclusive; NA: not amplified; ND: not determined.

Table 2B.

Comprehensive analysis of discrepant results (n = 49).

Case Age Gender Centre Analytical result SC2 serostatusa SC2 NAAT last month SC2 NAAT subsequent week Relevant clinical history Conclusion
1 39.7 F HUB N ND P ND History of SC2 infection Probable
2 41.2 F HUB ND ND N ND Unknown I
3 65.5 F HUB P P P ND History of SC2 infection Confirmed
4 55.1 M HUB P N N ND Pneumonia Confirmed
5 23.5 F HUB P ND ND P Contact of SC2 infected individual Confirmed
6 47.9 F HUB N ND P ND History of SC2 infection Probable
7 21.1 F HUB N ND P ND History of SC2 infection Probable
8 68.9 M HUB P ND P P History of SC2 infection Confirmed
9 56.7 M HUB ND Nb P ND History of SC2 infection Probable
10 20.0 M HUB ND P P P History of SC2 infection Probable
11 8.0 F HUB N ND ND ND Unknown I
12 85.2 M HUB P ND ND P Contact of SC2 infected individual Confirmed
13 56.8 M HUB P P ND ND History of SC2 infectionc Confirmed
14 25.4 M HUB P ND ND ND Unknown Confirmed
15 45.2 F HUB P P P ND History of SC2 infection Confirmed
16 39.2 F HUB ND ND P ND History of SC2 infection Probable
17 66.3 M HUB ND ND ND ND Symptomatic contact of SC2 infected individual I
18 34.1 F HUB N P P ND History of SC2 infection Probable
19 70.4 F HUB P ND ND ND History of SC2 infectionc Confirmed
20 46.8 F HUB P P P ND History of SC2 infection Confirmed
21 45.9 M HUB N Nb P ND History of SC2 infection Probable
22 31.8 M HUB N ND ND ND History of SC2 infectionc Probable
23 5.0 M HUB N ND ND ND Unknown I
24 61.0 F HUB N ND N ND Unknown I
25 2.0 M HUB P ND ND P Unknown Confirmed
26 36.1 M HUVH N ND ND ND History of SC2 infectiond Probable
27 1.8 M HUVH I ND ND ND Asymptomatic screening Probable
28 31.6 F HUVH I P P ND History of SC2 infection Probable
29 57.7 F HUVH N ND ND ND History of SC2 infectiond Probable
30 14.9 M HUVH P ND ND ND Symptomatic contact of SC2 infected individual Confirmed
31 37.8 M HUVH N ND ND ND Contact of SC2 infected individual I
32 37.4 F HUVH I ND ND ND Symptomatic contact of SC2 infected individual Probable
33 26.5 F HUVH N ND ND ND Mild symptoms compatible with COVID-19 I
34 54.1 F HUVH N P P ND History of SC2 infection Probable
35 94.8 F HUVH P P ND ND History of SC2 infection Confirmed
36 59.1 M HUVH P ND ND ND Mild symptoms compatible with COVID-19 Confirmed
37 44.5 F HUVH N P P ND History of SC2 infection Probable
38 42.7 M HUVH P ND ND ND Mild symptoms compatible with COVID-19 Confirmed
39 19.4 F HUVH N ND N ND History of SC2 infectionc Probable
40 57.1 F HUVH I ND ND ND Asymptomatic screening I
41 18.1 F HUVH P P N ND History of SC2 infectionc Confirmed
42 57.5 F HUVH N P ND ND History of SC2 infectionc Probable
43 56.3 M HUVH N ND ND ND Contact of SC2 infected individual I
44 11.9 M HUVH I ND ND ND Asymptomatic screening I
45 50.6 M HUVH I ND ND ND Contact of SC2 infected individual I
46 58.4 F HUVH I P P ND History of SC2 infection Probable
47 63.6 F HUVH P P ND ND History of SC2 infectiond Confirmed
48 30.0 F HUVH I ND ND ND Mild symptoms compatible with COVID-19 I
49 16.4 M HUVH N ND ND ND Contact of SC2 infected individual I
a

SC2 serostatus refers to the presence or absence of antibodies suggestive and compatible with history of SARS-CoV-2 infection, disregarding vaccination.

b

These cases correspond to non-reactive results near the positivity threshold. Of note, this phenomenon may suggest the initial immune response against SARS-CoV-2 with the incipient production of specific antibodies.

c

History of positive SARS-CoV-2 NAAT before the last month.

d

History of positive SARS-CoV-2 antigen test in the last month.

Abbreviations: SC2: SARS-CoV-2; NAAT: nucleic acid amplification test; M: male; F: female; P: positive; HUB: Hospital Universitari Bellvitge; HUVH: Hospital Universitari Vall d’Hebron; N: negative; ND: not determined; I: inconclusive.

Table 3.

Summary of the discrepant analysis described in Table 2A, Table 2B (n = 49).

Type of disagreement
Analytical resultb
N. (%)
Conclusionc
N. (%)
Positive Negative Inconclusived Not determinede Confirmed Probable Inconclusive
TMA Procleix Positive/RT-PCR Allplex™ Negative (n = 36) Major 11 (30.6) 19 (52.8) 5 (13.9) 1 (2.8) 11 (30.6) 17 (47.2) 8 (22.2)
TMA Procleix Negative/RT-PCR Allplex™ Positive (n = 2) Major 2 (100.0) 2 (100.0)
TMA Procleix Positive/RT-PCR Allplex™ Inconclusivea (n = 11) Minor 6 (54.6) 3 (27.3) 2 (18.2) 6 (54.6) 2 (18.2) 3 (27.3)
a

Inconclusive cases are considered those testing positive for the E target, but negative for both the N and the RdRP/S targets with the RT-PCR Allplex™ SARS-CoV-2 assay, according to the manufacturer's criteria.

b

The statements correspond to the analytical result of discrepant of cases, presented in Table 2A. In this assessment, discrepancies were further studied with an additional RT-PCR technique for SARS-CoV-2 testing.

c

The statements correspond to the comprehensive analysis of discrepant results, presented in Table 2B. In this assessment, discrepant cases were further studied through a comprehensive clinical and epidemiological review of the medical record.

d

Unresolved cases after the discrepant analytical testing (identified as I-inconclusive in Table 2A, Table 2B).

e

Specimens unavailable for discrepant analytical testing (identified as ND-not determined in Table 2A, Table 2B).

A minor disagreement is considered that discrepancy testing positive with one assay but inconclusive with the other study technique. Contrary, a major disagreement is considered that reverting from positive to negative depending on the study assay.

Discussion

The current wide availability of commercial assays to address the critical need for large-scale SARS-CoV-2 testing also requires comprehensive analytical, clinical and real-life experiences to better understand the strengths and limitations of each diagnostic technique. The present multicenter study compares the clinical performance of the TMA Procleix SARS-CoV-2 assay to a commercial RT-PCR test for the qualitative detection of SARS-CoV-2 RNA.

Overall, the TMA Procleix SARS-CoV-2 assay performed well for the detection of SARS-CoV-2 compared to the RT-PCR assay AllplexTM SARS-CoV-2, with a substantial agreement between assays (κ, 0.772). Most major discrepancies (95%) corresponded to samples testing positive with the TMA Procleix SARS-CoV-2 assay, but reverting to negative results with the RT-PCR test. When pondering the discrepant analyses, many of these cases (31%) were analytically confirmed to be positive for SARS-CoV-2, while many others (47%) were classified as probable positive according to the epidemiological and clinical history. These results suggest that the TMA Procleix SARS-CoV-2 assay may be more sensitive for the detection of SARS-CoV-2 RNA compared to RT-PCR methods, similar to what has been reported for this and other TMA tests such as the Aptima SARS-CoV-2 assay (Hologic, Inc., US).4, 5, 6, 7 On the other hand, most minor discrepancies, these were specimens testing positive with the TMA test but inconclusive with the RT-PCR assay Allplex™ SARS-CoV-2, were also reported as confirmed or probable SARS-CoV-2 cases in accordance with the hypothesis of an inherent higher sensitivity of the TMA assay. In this regard, findings from an in vitro supplemental experiment, using 10-fold dilutions in five SARS-CoV-2 positive samples, analytically reinforce this conclusion (Fig. S2).

Our study evidences that the TMA Procleix SARS-CoV-2 assay is highly sensitive. The methodology used strongly suggests that this feature, compared to RT-PCR methods, is not a matter of pre-analytical variability rather the intrinsic chemistry behind TMA technology.8, 9 Despite the presumptively high specificity of this technique, the increased sensitivity and the lack of quantitative/semi-quantitative parameters as a rough measure of SARS-CoV-2 RNA load (such as the cycle-threshold [Ct] value in RT-PCR methods) may hinder the clinical interpretation of some positive results, particularly among asymptomatic individuals, thus overestimating the true prevalence/incidence of active and transmissible infections. In low-prevalence settings, over-reporting positive cases may lead to unnecessary community isolation and contact tracing.10 Nevertheless, in this scenario the low incidence of COVID-19 may also allow confirmation of positive TMA results with RT-PCR methods to limit this inherent deviation and optimize the clinical specificity of SARS-CoV-2 testing.10 Moreover, pooling strategies, nowadays common for SARS-CoV-2 testing in clinical laboratories,11, 12 may clearly benefit from this highly sensitive TMA method.4

The study does require some considerations. First, information is missing in some discrepant cases due to the retrospective nature of the epidemiological and clinical data collection, and some specimens were unavailable for discrepant analytical testing. Also, at some point of the study, the TMA Procleix SARS-CoV-2 assay was used to actively select positive results, thus biasing the true positivity rate during that period. Furthermore, we did not strictly included acute SARS-CoV-2 infections and, as a consequence, some cases may correspond to the final stages of COVID-19 disease with uncertain clinical relevance. Finally, although chemiluminescense ratios ≥1.00 with the TMA Procleix SARS-CoV-2 assay are considered reactive/positive for SARS-CoV-2 according to the manufacturer's criteria, all positive results in this evaluation yielded ratio values ≥2.00. Of note, in our experience, positive ratios between 1.00 and 2.00 are rarely confirmed with RT-PCR methods.

In conclusion, the TMA Procleix SARS-CoV-2 assay performed well for the qualitative detection of SARS-CoV-2 RNA in a multisite clinical setting. Furthermore, this novel TMA assay demonstrated greater sensitivity in comparison to RT-PCR tests for the molecular detection of SARS-CoV-2, and this performance characteristic should be considered when making testing algorithm decisions. Nevertheless, additional experiences are required to fully refine the applicability of this methodology to the volatile dynamics of COVID-19 pandemic.

Funding

This study was supported by a grant from Centro de Investigación Biomédica en Red (CIBER) de Enfermedades Respiratorias (CIBERES, CB06/06/0037), an initiative of the Instituto de Salud Carlos III, Madrid, Spain. Co-funded by the European Regional Development Fund/European Social Fund (ERDF/ESF, “Investing in your future”).

Conflicts of interest

None.

Acknowledgments

We wish to thank all the staff of the Microbiology Departments of Hospital Universitari de Bellvitge and Hospital Vall d’Hebron who contributed to this project on a daily basis. We thank CERCA Programme/Generalitat de Catalunya for institutional support.

Footnotes

Appendix A

Supplementary data associated with this article can be found, in the online version, at doi:10.1016/j.eimce.2022.01.007.

Appendix A. Supplementary data

The following are the supplementary data to this article:

mmc1.pdf (377.7KB, pdf)

References

  • 1.Zhu N., Zhang D., Wang W., Li X., Yang B., Song J., et al. A novel coronavirus from patients with pneumonia in China, 2019. N Engl J Med. 2020;382:727–733. doi: 10.1056/NEJMoa2001017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Reusken C.B.E.M., Broberg E.K., Haagmans B., Meijer A., Corman V.M., Papa A., et al. Laboratory readiness and response for novel coronavirus (2019-nCoV) in expert laboratories in 30 EU/EEA countries January 2020. Euro Surveill. 2020;25:2000082. doi: 10.2807/1560-7917.ES.2020.25.6.2000082. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Food and Drug Administration (FDA). Individual EUAs for molecular diagnostic tests for SARS-CoV-2. Available from: https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19-emergency-use-authorizations-medical-devices/in-vitro-diagnostics-euas-molecular-diagnostic-tests-sars-cov-2 [consulted 22.4.21].
  • 4.Sauleda S., Palacios L., Brès V., Piñana M., Alonso-Hernandez L., Bes M., et al. Clinical evaluation of the Procleix SARS-CoV-2 assay, a sensitive, high-throughput test that runs on an automated system. Diagn Microbiol Infect Dis. 2021 doi: 10.1016/j.diagmicrobio.2021.115560. Epub ahead of print. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Gorzalski A.J., Tian H., Laverdure C., Morzunov S., Verma S.C., VanHooser S., et al. High-throughput transcription-mediated amplification on the Hologic Panther is a highly sensitive method of detection for SARS-CoV-2. J Clin Virol. 2020;129:104501. doi: 10.1016/j.jcv.2020.104501. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Pham J., Meyer S., Nguyen C., Williams A., Hunsicker M., McHardy I., et al. Performance characteristics of a high-throughput automated transcription-mediated amplification test for SARS-CoV-2 detection. J Clin Microbiol. 2020;58:e01669–e1720. doi: 10.1128/JCM.01669-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Smith E., Zhen W., Manji R., Schron D., Duong S., Berry G.J. Analytical and clinical comparison of three nucleic acid amplification tests for SARS-CoV-2 detection. J Clin Microbiol. 2020;58:e01134–e1220. doi: 10.1128/JCM.01134-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Hatzakis A., Papachristou H., Nair S.J., Fortunko J., Foote T., Kim H., et al. Analytical characteristics and comparative evaluation of Aptima HIV-1 Quant Dx assay with Ampliprep/COBAS TaqMan HIV-1 test v2.0. Virol J. 2016;13:176. doi: 10.1186/s12985-016-0627-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Hofmann W.P., Dries V., Herrmann E., Gärtner B., Zeuzem S., Sarrazin C. Comparison of transcription mediated amplification (TMA) and reverse transcription polymerase chain reaction (RT-PCR) for detection of hepatitis C virus RNA in liver tissue. J Clin Virol. 2005;2:289–293. doi: 10.1016/j.jcv.2004.08.011. [DOI] [PubMed] [Google Scholar]
  • 10.Skittrall J.P., Wilson M., Smielewska A.A., Parmar S., Fortune M.D., Sparkes D., et al. Specificity and positive predictive value of SARS-CoV-2 nucleic acid amplification testing in a low-prevalence setting. Clin Microbiol Infect. 2021;27:469e9–469e15. doi: 10.1016/j.cmi.2020.10.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Lohse S., Pfuhl T., Berkó-Göttel B., Rissland J., Geißler T., Gärtner B., et al. Pooling of samples for testing for SARS-CoV-2 in asymptomatic people. Lancet Infect Dis. 2020;20:1231–1232. doi: 10.1016/S1473-3099(20)30362-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Mutesa L., Ndishimye P., Butera Y., Souopgui J., Uwineza A., Rutayisire R., et al. A pooled testing strategy for identifying SARS-CoV-2 at low prevalence. Nature. 2021;589:276–280. doi: 10.1038/s41586-020-2885-5. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

mmc1.pdf (377.7KB, pdf)

Articles from Enfermedades Infecciosas Y Microbiologia Clinica (English Ed.) are provided here courtesy of Elsevier

RESOURCES