Abstract
Purpose:
To develop a fast, deep-learning approach for quantitative MTC-MRF that simultaneously estimates multiple tissue parameters and corrects the effects of B0 and B1 variations.
Methods:
An only-train-once recurrent neural network was designed to perform the fast tissue parameter quantification for a large range of different MRF acquisition schedules. It enabled a dynamic scan-wise linear calibration of the scan parameters using the measured B0 and B1 maps, which allowed accurate, multiple-tissue parameter mapping. MRF images were acquired from eight healthy volunteers at 3T. Estimated parameter maps from the MRF images were used to synthesize the MTC reference signal (Zref) through Bloch equations at multiple saturation power levels.
Results:
The B0 and B1 errors in MR fingerprints, if not corrected, would impair the tissue quantification and the subsequently corrupt the synthesized MTC reference images. Bloch equation-based numerical phantom studies and synthetic MRI analysis demonstrated that the proposed approach could correctly estimate water and semisolid macromolecule parameters, even with severe B0 and B1 inhomogeneities.
Conclusion:
The only-train-once deep-learning framework can improve the reconstruction accuracy of brain tissue parameter maps and be further combined with any conventional MRF or CEST-MRF method.
Keywords: B0 correction, B1 correction, Deep-learning, Magnetization Transfer Contrast, MR fingerprinting
1. INTRODUCTION
Magnetization transfer contrast (MTC) provides an indirect measurement of semisolid macromolecular protons based on the transfer of magnetization to the surrounding free bulk water molecules, which is not directly detectable with conventional MRI sequences, because the semisolid protons have an extremely short T2 relaxation times (< 100 μs) (1–4). The cumulative saturation with repeated radiofrequency (RF) irradiation of macromolecular protons results in a decrease in the water signal through the magnetization transfer, thereby allowing the assessment of semisolid macromolecules with improved sensitivity. MTC imaging has been shown to be a powerful biomarker for the clinical diagnosis of disorders in tissues (e.g., multiple sclerosis), which provides a profound insight into the brain-tissue microstructure (5–8). A conventional MTC experiment as measured by the MT ratio (MTR) has been widely used for clinical studies (9–11). However, the MTR contrast is highly dependent on scan parameters and water tissue relaxation effects. To overcome the poor specificity, quantitative imaging techniques have been developed by fitting MTC-weighted signals to the analytical solution of the Bloch equations using a series of image acquisitions with various saturation powers and frequency offsets (12–15). However, these model-based methods sometimes suffer from long imaging times, due to the repeated acquisitions necessary with various experimental settings, and the computationally expensive fitting process.
Magnetic resonance fingerprinting (MRF) was introduced as a time-efficient quantification technique that can simultaneously estimate multiple tissue parameters (16). The time-varying signal evolution, the so-called fingerprint, is achieved by intentionally varying imaging parameters for each scan and this is then used to probe the characteristics of tissues. A conventional approach to map a fingerprint space into a tissue parameter space is to match the fingerprint with a pre-defined dictionary (17–20). Recent studies using learning-based techniques have shown their powerful ability to map the two different domains of the MR fingerprint space and the tissue parameter space, bypassing the exhaustive dictionary search and circumventing the curse of dimensionality (21–24). Furthermore, learning-based MRF reconstruction techniques have been adopted for quantitative MTC imaging with various training schemes (25–27).
B0 and B1 inhomogeneities that originate from system imperfections hinder MRF reconstruction that solves the complex inverse problem of Bloch equations (28–31). Particularly, MRF sequences for MTC and chemical exchange saturation transfer (CEST) imaging use a pulsed or continuous RF saturation schemes with various saturation powers, and are, thus, vulnerable to B1 variation. Since the MTC and CEST measurements are based on the RF saturation of specific proton pools, errors in the B0 and B1 values directly propagate to tissue quantification errors (32–36). Consequently, B0 and B1 correction are essential in MTC- and CEST-MRF imaging. Typically, experimentally acquired B0 and B1 maps can be used to calibrate an off-resonance frequency and to scale the power of an RF pulse, respectively. In MRF with a subgrouping proton exchange model (MRF-SPEM) framework (37), B0 shifts and B1 scaling factors obtained from additional scans were used to linearly compensate for the MRF schedule for the CEST and MTC quantification. However, it is challenging to apply such compensation methods to a deep-learning-based MRF technique, because neural networks are trained on a simulated dataset with numerous combinations of tissue parameters for a specific MRF schedule without considering B0 and B1 inhomogeneities. Consequently, B0 and B1 errors would result in unpredictable quantification errors from the conventional deep-learning approach.
Recently, we introduced a novel deep-learning approach for MTC-MRF, which can quantify tissue parameters for a myriad of different MRF schedules, and dubbed Only-Train-Once MRF (OTOM) (38). The OTOM scheme trained with numerous acquisition patterns and lengths of MRF sequence can be applied to a large range of MRF schedules without compromising MRF reconstruction accuracy. In this study, we developed a linear compensation scheme for B0 and B1 correction on the OTOM framework. To validate the proposed correction method, digital phantom studies and synthetic MRI analysis were performed with a two-pool Bloch equations-based simulation. The performance of the B0 and B1 correction was demonstrated in the brain of healthy volunteers.
2. THEORY
2.1. Transient-state MTC-MRF model
A two-pool MTC exchange model (w: a free bulk water pool, m: a semisolid macromolecule pool) is used to simulate MTC-MRF signals in the presence of proton exchange and RF irradiation. The magnetization of each pool can be described with the modified Bloch-McConnell equations in a matrix format as follows (39,40):
[1] |
where
[2] |
[3] |
[4] |
[5] |
[6] |
[7] |
where T1i and T2i are the longitudinal and transverse relaxation times of a pool i, respectively; Ω is the frequency offset of the RF saturation; B1 is the RF saturation power; γ is the gyromagnetic ratio (for ω1/B1= γ = 267.5 rad/(μT×s)); kij is the proton exchange rate from a pool i to a pool j; and M0i is the equilibrium magnetization of a pool i. By solving the coupled differential equations (Eq. [1]), the longitudinal magnetization of the free bulk water pool can be written as:
[8] |
where
[9] |
[10] |
[11] |
[12] |
[13] |
where Mssw is the steady-state longitudinal magnetization of the free bulk water, and Li denotes a Lorentzian line-shape function of the RF absorption rate of a pool i. According to Eq. [8], the characteristics of the RF saturation are determined by three scan parameters: RF saturation power (B1); frequency offset (Ω); and saturation time (t = Ts). A relaxation delay time (Td) is also defined to consider the relaxation of the longitudinal magnetization in the absence of RF irradiation, changing an initial value of the magnetization for the next dynamic scan. Therefore, the transient-state MTC-MRF signal evolution (SMTC-MRF) can be described as follows:
[14] |
where
[15] |
[16] |
where ptissue is a set of tissue parameters and pscan,k is a set of scan parameters of the kth scan. The analytical solution of the MTC-MRF signal model is used to understand the complex relation among the fingerprint space (SMTC-MRF), the scan parameter space (pscan), and the tissue parameter space (ptissue). The complex relation can be solved by a deep neural network.
2.2. B0 and B1 correction
A linear compensation scheme was used to compensate the B0 and B1 inhomogeneities in a pixel-wise manner. In the presence of B1 inhomogeneity, the saturation power (B1) is linearly scaled according to the relative B1 (rB1, a scaling factor of B1, not to be confused with gamma γ, the gyromagnetic ratio) map per pixel. Similarly, the frequency offset (Ω) is shifted with the ΔB0 map. Therefore, the correction can be accomplished by B0 shifting and B1 scaling as follows (41,42):
[17] |
[18] |
[19] |
where and are the corrected and nominal RF saturation power of ith scan, respectively; and are the corrected and nominal frequency offset of ith scan, respectively; rB1 is a scaling factor of B1; ΔB0 is a field inhomogeneity offset; represents the corrected scan parameters; and represents the nominal scan parameters. Note that the ΔB0 and rB1 values of each pixel were equally applied to the entire MRF schedule for B0 and B1 correction.
3. METHODS
An overview of the OTOM method for B0 and B1 correction is described in Fig. 1. In the training phase, the recurrent neural network (RNN) was trained to solve the complex inverse problem of mapping an MR fingerprint into tissue parameters in accordance with the MRF schedule. Millions of different MRF schedules were applied to the training, and thus, the trained RNN model can estimate tissue parameters for a variety of different MRF schedules in the test phase. This enables a dynamic scan-wise modification of the MRF schedule, which can be utilized to compensate for B0 and B1 errors. The corrected scan parameters (), in lieu of the nominal scan parameters (), were calibrated using experimentally obtained ΔB0 and rB1 maps and fed to the trained RNN to estimate the field inhomogeneity artifact-free tissue parameter maps.
3.1. Data preparation: Training dataset
To train the RNN model that can be successfully applied to various MRF schedules, such as acquisition pattern and length, MTC-MRF signals (SMTC-MRF) were simulated using the analytical solution of the Bloch-McConnell equations (Eq. [14]). A random MRF schedule was generated by varying the scan parameters (pscan) for each MR fingerprint. As shown in Fig. 2A, the number of dynamic scans (N) was determined and then four scan parameters (B1, Ω, Ts, and Td) were randomly sampled N times within the pre-defined ranges (Table 1). Similarly, tissue parameters were sampled for the simulation within the corresponding ranges, as shown in Table 1. In addition, T1m was chosen to be a constant value of 1 s due to its negligible contribution to the two-pool MTC signal. Note that T2w was indirectly defined from the T1w/T2w ratio, which referred to a line-shape of direct water saturation, thus, a key parameter in the two-pool model. To generate the training dataset, 80 million combinations of scan and tissue parameters were randomly chosen. Finally, white Gaussian noise (SNR = 46 dB), whose SNR was similar to the obtained in vivo images, was added to the simulated MTC-MRF signals. The Bloch simulations were performed on a 64-bit Windows operating system (12-CORE, 3.8-GHz AMD Ryzen 9 3900XT processor and 32 GB of memory) using MATLAB (MathWorks, Natick, MA).
Table 1.
Tissue parameters | Scan parameters | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
kmw (Hz) | M0m (%) | T2m (μs) | T1w (s) | T1w/T2w | B1 (μT) | Ω (ppm) | Ts (s) | Td (s) | N (#) | |
Upper bound | 100.0 | 17.0 | 100.0 | 3.0 | 30.0 | 3.0 | 50 | 2.0 | 5 | 40 |
Lower bound | 5.0 | 2.0 | 1.0 | 0.2 | 1.0 | 0.4 | 8 | 0.4 | 3.5 | 10 |
Interval | 1.0 | 0.1 | 1.0 | 0.03 | 0.2 | 0.1 | 1 | 0.1 | 0.1 | 1 |
3.2. Recurrent neural network (RNN)
The RNN architecture was designed to analyze the input data of the MTC fingerprints and the corresponding scan parameters and to output the free bulk water and semisolid MTC parameters. As shown in Fig. 2B, the architecture consisted of bi-directional LSTM (long short-term memory) (43,44) and a single fully connected layer. The RNN extracted features from each time point, not considering the input MR fingerprint all at once, and accumulated them in a hidden state. The finally updated hidden states of the forward and backward directions were concatenated, and fed to the fully connected layer to estimate the four tissue parameters of kmw, M0m, T2m, and T1w, as follows:
[20] |
[21] |
[22] |
where
[23] |
where and are the forward and backward hidden states of ith time point, respectively; X[i] is the input vector of ith time point; SMIC–MRF,i is the MRF signal of ith time point; and N is the number of dynamic scans of the MRF sequence. The LSTM consisted of three layers with 512 hidden units each, which led to 1024 hidden units after the concatenation of the forward and backward hidden states, and the fully connected layer had 1024×4 neurons followed by rectified linear units (ReLU) as an activation function. The RNN was trained by minimizing the l1-norms of difference between the label parameters, ptissue, and the estimated parameters, , as follows:
[24] |
The network was implemented using Pytorch on an NVIDIA TITAN RTX GPU (Santa Clara, CA). The network was trained for 20 epochs with the adaptive moment estimation (ADAM) optimizer (45) and a batch size of 256. The initial learning rate was 10−3 and decreased by a factor of 0.1 for every three epochs. The training dataset was randomly divided into two parts: 90% for training and 10% for validation. The RNN model that showed the lowest validation loss was saved for our experiments. In addition, the validation loss was monitored for early stopping of training. OTOM was trained for 100 hours on 80 million dataset whereas it took 6 hours to train the FCNN with 10 million dataset. OTOM consumed 2459MB and FCNN consumed 1159MB with the batch size of 256 on NVIDIA TITAN RTX GPU.
3.3. Digital phantom study: Bloch simulation
The proposed method was validated with digital phantoms simulated from the two-pool Bloch-McConnell equations. Two phantom studies were designed to investigate the effect of B0 and B1 correction on the accuracy of tissue parameter quantification (kmw, M0m, T2m, and T1w).
In the first digital phantom study, ΔB0 and rB1 values were randomly sampled to simulate the effect of field inhomogeneities. The parameter ranges were given as −1.2 to 1.2 ppm for ΔB0 and 50 to 150% for rB1. As shown in Fig. 3A, four digital phantoms (PH1, PH2, PH3, PH4) were constructed, each of which had five circular compartments to evaluate each of the four tissue parameters, while the other three tissue parameters were randomly chosen within the pre-defined ranges (Table 1). The five tissue parameter values for each phantom were: 5, 20, 40, 60, and 80 Hz for kmw; 2, 6, 10, 14, and 17% for M0m; 1, 25, 50, 75, and 100 μs for T2m; and 0.2, 0.9, 1.6, 2.3, and 3.0 s for T1w. Longitudinal magnetization evolutions of the free bulk water protons (MTC-MRF) were encoded to the four digital phantom images using a Bloch simulation with an MRF schedule of 40 dynamic scans. The simulated fingerprints were fed to the RNN in accordance with the nominal scan parameters () and the corrected scan parameters (), respectively. The estimated tissue parameters were compared to ground-truths and the normalized root mean square errors (nRMSE) were calculated. The MRF schedule was optimized from the learning-based optimization of the acquisition schedule (LOAS) method (46). In the second digital phantom study, the quantification accuracy was evaluated with various numbers of dynamic scans (#10, #20, #30, and #40). Quantification results from the nominal scan parameters were compared with those from the B0 and B1-corrected scan parameters.
3.4. In vivo MRI measurements
Eight healthy volunteers (three females and five males; age 38.1 ± 4.1) were scanned on a 3T MRI system (Achieva dStream, Philips Healthcare, Best, the Netherlands). All subjects were examined with the approval of the institutional review board, and written, informed consent was obtained prior to the MRI experiments. The 3D MTC-MRF images were acquired from a fat-suppressed (spectral pre-saturation with inversion recovery) multi-shot turbo spin echo (TSE) pulse sequence with 2×2 compressed sensing (CS) accelerations in the two phase-encoding directions (ky-kz) (47). The imaging parameters were TE = 6 ms, FOV = 212 × 186 × 60 mm3, spatial resolution = 1.8 × 1.8 × 4 mm3, slice-selective 120° refocusing pulses, turbo factor = 104, and slice oversampling factor = 1.4. The two-channel, time-interleaved, parallel RF transmission (pTX) technique through the body coil was applied to achieve pseudo-continuous RF saturation with a 100% duty-cycle. The pTX-based saturation allowed a high degree of freedom for RF sequence design and highly sensitive saturation effects (48,49). Forty dynamic scans were acquired with a PR (pseudo-random) schedule for six subjects and an LOAS schedule for two subjects. To normalize MTC-MRF images, an unsaturated image (S0) was acquired. In addition, multi-echo gradient-spin-echo (GRASE) images with five echoes (TE = 20, 40, 60, 80, and 100 ms) were acquired for the T2 map (50). For B0 mapping, the water saturation shift referencing (WASSR) method (51) was used with the following RF saturation parameters: 26 frequency offsets (from −1.5 to 1.5 ppm at intervals of 0.125 ppm); Ts = 800 ms; and B1 = 0.5 μT. For B1 mapping, the dual refocusing echo acquisition mode (DREAM) method (52) was used with the simulated echo acquisition mode (STEAM) with a flip angle of 40°.
3.5. Validation of B0 and B1 correction using synthetic MRI analysis
A synthetic MRI analysis was performed to gauge how tissue estimates can be corrupted by the influence of B0 and B1 errors and how well the proposed method corrects B0 and B1 inhomogeneities. The field inhomogeneities were simulated with synthetic ΔB0 and relative B1 (rB1) maps. Both maps were constructed to have nine constant values (60, 70, 80, 90, 100, 110, 120, 130, and 140% for rB1; −0.8, −0.6, −0.4, −0.2, 0.0, 0.2, 0.4, 0.6, and 0.8 ppm for ΔB0). The tissue parameter maps estimated from in vivo MTC-MRF images obtained with the PR schedule of forty dynamic scans were defined as the reference (namely, Ref). Corresponding 3D MTC-MRF images were synthesized by inserting the reference tissue parameters, ΔB0, rB1 maps, and the LOAS schedule into the forward Bloch transform. Then, the synthesized 3D MTC-MRF images were fed to the RNN model. The percent error maps were calculated for all tissue parameters. To further evaluate the performance of the field inhomogeneity correction using the proposed method, nRMSE was calculated as a measure of the quantification accuracy for various numbers of scans (#10, #20, #30, and #40). In addition, the estimated tissue parameter maps and the experimentally acquired water T2 map were used to synthesize the MTC reference signal intensity (Zref) through Bloch equations.
4. RESULTS
4.1. Digital phantom studies
The performance of the RNN-based B0 and B1 correction method was evaluated using digital phantoms simulated from two-pool Bloch equations, as shown in Fig. 3. The MRF schedule optimized by LOAS was encoded to each phantom. The saturation power (B1) was scaled with the relative B1 (rB1) map and the frequency offset was shifted with the ΔB0 map (See supporting Figure S1). The estimated tissue parameter maps using the corrected scan parameters () were much more accurate than those using the nominal scan parameters () for all tissue parameters. The nRMSE values with the nominal and corrected scan parameters were 32.5% and 20.1% for Kmw, 28.1% and 7.4% for M0m, 9.7% and 3.0% for T2m, and 3.9% and 0.9% for T1w, respectively. Specifically, the accuracy of the exchange rate and the concentration parameters was greatly improved by reducing more than 12% and 20% of nRMSE, respectively. Overall, the nRMSE values of free bulk water T1 and the semisolid macromolecules T2 were lower than those of the exchange rate and the concentration.
Figure 4 illustrates the performance of B0 and B1 correction for various MRF schedule lengths. MTC-MRF images were generated using the two-pool Bloch equations with four MRF acquisition schedules with different dynamic scan numbers (LOAS #10, LOAS #20, LOAS #30, and LOAS #40) and fed to the RNN model to quantify the tissue parameters. The MRF reconstruction accuracy substantially increased after B0 and B1 correction (See supporting Figure S2). While the nRMSE values decreased with the number of dynamic scans after B0 and B1 correction, no significant correlation was observed between the number of dynamic scans and nRMSE values without B0 and B1 correction.
4.2. Synthetic MRI analysis
To assess the proposed correction method with in vivo images, a simulated relative B1 map (nine values of rB1 from 60 to 140% at intervals of 10%) and reference tissue parameter maps were assigned to synthesize MTC-MRF images, as shown in Fig. 5A. The synthetic images were fed to the trained RNN with the nominal and corrected scan parameters to estimate B1-corrupted and B1-corrected tissue parameter maps, respectively. There was excellent agreement between the reference tissue parameter maps and the estimated tissue parameter maps from the corrected scan parameters (). However, a poor agreement was observed between the reference maps and the estimated maps from the nominal scan parameters (). The Pearson correlation coefficients for and were 0.742 and 0.253 for Kmw, 0.995 and 0.503 for Mom, 0.985 and 0.976 for T2m, and 0.998 and 0.923 for T1w, respectively. The B1 inhomogeneity highly impaired the quantification of the exchange rate and the concentration of the semisolid macromolecules, whereas the free bulk water T1 and the semisolid macromolecule T2 relaxation times were relatively robust to the B1 inhomogeneity. Specifically, the error maps of kmw and M0m were exacerbated by relative B1 values lower than 80% and higher than 120%. Figure 6 shows the accuracy of tissue parameters and MTC signals at 3.5 ppm (= 1 - Zref(3.5 ppm)) estimated from various numbers of dynamic scans. As the number of dynamic scans increased, the nRMSE values decreased when B1 inhomogeneity was corrected (gray bars). MTC reference signals at 3.5 ppm were relatively insensitive to the number of dynamic scans, whose nRMSE value was less than 0.8%, even with 10 dynamic scans.
In addition, the B0 correction was validated using a synthetic MRI analysis (see Supporting Fig. S3). Good agreement was observed between the reference maps and the estimated tissue parameter maps with and without B0 correction. The Pearson correlation coefficients for and were 0.900 and 0.900 for Kmw, 0.997 and 0.996 for Mom, 0.991 and 0.991 for T2m, and 0.998 and 0.998 for T1w, respectively.
4.3. In vivo experiments
The RNN network trained with the simulated MTC-MRF signals using numerous sets of scan and tissue parameters was applied to human brain images of healthy volunteers. MTC-MRF images were experimentally acquired using LOAS schedules of 40 dynamic scans (LOAS #40). The nominal scan parameters and corrected scan parameters, compensated with the acquired ΔB0 from WASSR and rB1 maps from the DREAM methods, were fed to the trained RNN with the in vivo MTC-MRF images to estimate water and MTC parameter maps. Figure 7 shows the quantitative water and MTC parameter maps obtained from the RNN-based MRF reconstruction in accordance with the nominal and corrected scan parameters and difference images between the estimated maps with and without B0 and B1 correction. The B1 inhomogeneity pattern obviously reflected the tissue parameter maps, particularly as seen in the difference images of the exchange rate and the concentration, whereas the B0 inhomogeneity had less impact on the MTC-MRF reconstruction. The field inhomogeneity error further propagated to the synthesized MTC images at 3.5 ppm, as shown in Fig. 8. A 17% error in B1 (ROI 2 in Fig. 8) produced a water saturation signal change of up to 5%. The proposed correction method appeared to mitigate the errors that originated from the field inhomogeneities and improved reconstruction accuracy.
5. DISCUSSION
The main and transmit magnetic field inhomogeneities alter MTC-MRF signal profiles, which leads to substantive errors in the estimated tissue parameters. Here, we developed a fast and effective technique to correct the field inhomogeneity-induced artifacts and errors. The linear compensation scheme was adopted to correct the quantification errors from the local B0 and B1 variations. The performance of the correction technique was evaluated using numerical phantom studies and synthetic MRI analysis, and demonstrated a high degree of accuracy in quantifying water and semisolid macromolecule parameters, even with severe B0 and B1 inhomogeneities. The proposed network enabled fast tissue parameter estimations of the human brain due to its simple feed-forward deployments in the test phase.
MRF allows direct mapping of unique magnetization evolutions into tissue parameters by solving the ill-posed, non-linear inverse problem of MR physics models with dictionary-matching (53), model-based (37), or learning-based (25,27) methods. The MR signal model involves three different spaces: fingerprint; tissue parameter; and scan parameter spaces. The dictionary-matching and learning-based MRF techniques solve the inverse-mapping problem between the fingerprint space and the tissue parameter space. Thus, the scan parameter space is fixed for the dictionary-matching and learning-based methods. Although the model-based approach fully explores MR signal models, becoming independent of the MRF sequence, high computational complexity limits its clinical usefulness. In this work, various tissue parameters, as well as scan parameters, were exploited during the training of neural networks, so that it could be applied to MTC-MRF images obtained with any scan parameters. The deep-learning approaches have been the most effective methods with which to circumvent the curse of dimensionality in MR fingerprinting (54). Our results reiterate the effectiveness of a neural network in MR fingerprinting by expanding the parameter dimension beyond tissue parameters to scan parameters. In addition, a recurrent neural network was adopted, which is suitable for time series data, i.e., signal evolution of the MR fingerprint over time (55,56). Only the hidden state is updated throughout every time point of the input sequence and used for next step analysis, allowing the input sequence of any scan number. It is noteworthy that the single RNN model was trained only once and applied to different types of MRF acquisition schedule (pseudorandom and LOAS-optimized) with various schedule lengths in this study. In addition, it is very important to define the range of scan parameters not to fall outside the range of training data after correction, because the learning-based methods could fail to generalize beyond the parameter values encountered in the training (See Supporting Fig. S4).
Several deep-learning approaches have been applied to MTC-MRF reconstruction by training networks with simulated or in vivo MRF profiles (25,26). However, the deep-learning approaches did not consider the B1 correction. It would be beneficial to develop a network that corrects B0 and B1 error without explicit image acquisition to measure B0 and rB1 maps. However, the network trained with B1-inhomogeneity-involved dataset failed to estimate accurate tissue parameters without an explicit rB1 map as an input (57), particularly the concentration of semisolid tissues, when there was moderate-to-severe B1 inhomogeneity (see Supporting Fig. S5). When the rB1 map was given as an input, neural network was able to correct B1 inhomogeneity accurately. This emphasizes the need for B1 inhomogeneity correction in MTC-MRF using the rB1 map. In this study, we utilized the OTOM framework to support a large range of different acquisition schedules, and thereby, enabled the linear compensation of the scan parameters. The proposed approach effectively corrected B0 and B1 errors, while maintaining the fast tissue parameter quantification due to the feed-forward mechanism of the neural network. The reconstruction time for an image matrix of 256 × 256 × 9 (slices) × 40 (scans) was less than two minutes, whereas the fitting method required hours.
The exchange rate and the concentration maps of the semisolid macromolecular protons were highly corrupted with severe B1 inhomogeneity. Interestingly, a linear relationship was observed between the rB1 value and the concentration estimation. Low values of rB1 (relative to nominal B1) underestimated the concentration of semisolid tissues and high values of rB1 overestimated the concentration of semisolid tissues, which is in line with a previous observation (58). It is not surprising that B1 errors can influence MRF signal intensities and be mistaken as the saturation effect from the semisolid proton concentration, which results in incorrect tissue quantification. The exchange rate map from the proposed correction method showed high error with low values of rB1, presumably due to the insufficient saturation. Therefore, the field inhomogeneity correction resulted in somewhat different MTC values compared to previous studies without correction (25,26). Especially, the estimated MTC exchange rates of gray matter (~8 Hz) were lower than those reported previously (~15 Hz) and the concentration of the gray matter (~14 %) was significantly higher with B1 correction compared to previous reported value (~10 %), which was consistent with the values before B1 correction (See Supporting Table S1). In addition, the Intrinsic water T1 relaxation times estimated in this study were longer than the observed T1 relaxation times estimated from conventional inversion-recovery studies because of the interaction between the free bulk water and semisolid macromolecules (26). Although water T1 estimation was less affected by small B1 errors, very high rB1 values (> 120%) caused a substantial error in water T1 estimation (Fig. 5B) due to small MRF signal discrimination between different T1w values. The MTC-MRF signal becomes less dependent on the longitudinal relaxation term () when the higher saturation (eλTs, where λ is negative and includes B1) is applied (Eq. [15]). On the other hand, the semisolid macromolecular T2 was relatively robust to B1 inhomogeneity. Interestingly, the effect of B1 errors on MTC-MRF reconstruction is influenced by a B1 power schedule (see Supporting Fig. S6). In addition, the MTC-MRF signal was less sensitive to the exchange rate and therefore the estimated exchange rate could easily be distorted by B1 inhomogeneity (25). This could be even exacerbated by a different MRF schedule, leading to inconsistent rRMSE values with respect to the number of dynamic scans (See Supporting Fig. S7). Thus, the robustness of tissue parameter estimation to B1 errors could be improved by optimizing the acquisition protocol (59). The optimization of an MRF acquisition schedule that is more resilient to B0 and B1 errors may be an important future study direction.
A conventional MTC-MRF approach is relatively less sensitive to B0 error than B1 error (Supporting Fig. S3 and Supporting Fig. S8) because of a very broad line-shape as a result of the microsecond T2 of semisolid macromolecules. However, B0 inhomogeneity is not trivial in CEST parameter quantification because the chemical shift of the solute protons is sufficiently close to the water frequency and the location of CEST peaks is easily shifted by B0 error. Therefore, the proposed OTOM-based B0 and B1 correction could benefit CEST-MRF reconstruction or imaging of the brainstem, frontal lobes, and temporal lobes, where severe B0 inhomogeneity in the air-tissue interfaces remains. In addition to the B0 and B1 correction, OTOM could be further extended to MRF schedule optimization by analyzing the quantification error of a given schedule (46,60,61). The feasibility was already demonstrated (62).
6. Conclusion
A fast, deep-learning approach was developed to simultaneously estimate multiple tissue parameters and correct B0 and B1 errors, validated in digital phantoms, and demonstrated in healthy volunteers. The proposed method could achieve a high degree of accuracy for tissue parameter quantification in the presence of severe B0 and B1 inhomogeneities. The only-train-once deep-learning framework can be combined with any conventional MRF or CEST-MRF method, and improve the reconstruction accuracy of tissue parameter maps.
Supplementary Material
ACKNOWLEDGMENTS
This work was supported, in part, by grants from the National Institutes of Health (R01EB029974 and R01NS112242), and by the Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health & Welfare, the Ministry of Food and Drug Safety) (Project Number: 1711138003, KMDF-RnD KMDF_PR_20200901_0041-2021-02).
Footnotes
SUPPORTING INFORMATION
Additional Supporting Information associated with this article may be found in the online version of this article.
References
- 1.Henkelman RM, Huang X, Xiang QS, Stanisz GJ, Swanson SD, Bronskill MJ. Quantitative interpretation of magnetization transfer. Magn Reson Med 1993;29(6):759–766. [DOI] [PubMed] [Google Scholar]
- 2.Morrison C, Henkelman RM. A model for magnetization transfer in tissues. Magn Reson Med 1995;33(4):475–482. [DOI] [PubMed] [Google Scholar]
- 3.van Zijl PCM, Lam WW, Xu J, Knutsson L, Stanisz GJ. Magnetization Transfer Contrast and Chemical Exchange Saturation Transfer MRI. Features and analysis of the field-dependent saturation spectrum. Neuroimage 2018;168:222–241. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sled JG. Modelling and interpretation of magnetization transfer imaging in the brain. Neuroimage 2018;182:128–135. [DOI] [PubMed] [Google Scholar]
- 5.Eckert T, Sailer M, Kaufmann J, Schrader C, Peschel T, Bodammer N, Heinze HJ, Schoenfeld MA. Differentiation of idiopathic Parkinson’s disease, multiple system atrophy, progressive supranuclear palsy, and healthy controls using magnetization transfer imaging. Neuroimage 2004;21(1):229–235. [DOI] [PubMed] [Google Scholar]
- 6.Filippi M, Rocca MA, Martino G, Horsfield MA, Comi G. Magnetization transfer changes in the normal appearing white matter precede the appearance of enhancing lesions in patients with multiple sclerosis. Ann Neurol 1998;43(6):809–814. [DOI] [PubMed] [Google Scholar]
- 7.Levesque IR, Giacomini PS, Narayanan S, Ribeiro LT, Sled JG, Arnold DL, Pike GB. Quantitative magnetization transfer and myelin water imaging of the evolution of acute multiple sclerosis lesions. Magn Reson Med 2010;63(3):633–640. [DOI] [PubMed] [Google Scholar]
- 8.Pike GB, De Stefano N, Narayanan S, Worsley KJ, Pelletier D, Francis GS, Antel JP, Arnold DL. Multiple sclerosis: magnetization transfer MR imaging of white matter before lesion appearance on T2-weighted images. Radiology 2000;215(3):824–830. [DOI] [PubMed] [Google Scholar]
- 9.Filippi M, Iannucci G, Tortorella C, Minicucci L, Horsfield MA, Colombo B, Sormani MP, Comi G. Comparison of MS clinical phenotypes using conventional and magnetization transfer MRI. Neurology 1999;52(3):588–594. [DOI] [PubMed] [Google Scholar]
- 10.Filippi M, Bozzali M, Horsfield MA, Rocca MA, Sormani MP, Iannucci G, Colombo B, Comi G. A conventional and magnetization transfer MRI study of the cervical cord in patients with MS. Neurology 2000;54(1):207–213. [DOI] [PubMed] [Google Scholar]
- 11.Rovaris M, Agosta F, Sormani MP, Inglese M, Martinelli V, Comi G, Filippi M. Conventional and magnetization transfer MRI predictors of clinical multiple sclerosis evolution: a medium-term follow-up study. Brain 2003;126(Pt 10):2323–2332. [DOI] [PubMed] [Google Scholar]
- 12.Sled JG, Pike GB. Quantitative imaging of magnetization transfer exchange and relaxation properties in vivo using MRI. Magn Reson Med 2001;46(5):923–931. [DOI] [PubMed] [Google Scholar]
- 13.Hua J, Jones CK, Blakeley J, Smith SA, van Zijl PC, Zhou J. Quantitative description of the asymmetry in magnetization transfer effects around the water resonance in the human brain. Magn Reson Med 2007;58(4):786–793. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Heo HY, Zhang Y, Jiang S, Lee DH, Zhou J. Quantitative assessment of amide proton transfer (APT) and nuclear overhauser enhancement (NOE) imaging with extrapolated semisolid magnetization transfer reference (EMR) signals: II. Comparison of three EMR models and application to human brain glioma at 3 Tesla. Magn Reson Med 2016;75(4):1630–1639. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Gochberg DF, Gore JC. Quantitative imaging of magnetization transfer using an inversion recovery sequence. Magn Reson Med 2003;49:501–505. [DOI] [PubMed] [Google Scholar]
- 16.Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold MA. Magnetic resonance fingerprinting. Nature 2013;495(7440):187–192. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.McGivney DF, Pierre E, Ma D, Jiang Y, Saybasili H, Gulani V, Griswold MA. SVD compression for magnetic resonance fingerprinting in the time domain. IEEE Trans Med Imaging 2014;33(12):2311–2322. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Cauley SF, Setsompop K, Ma D, Jiang Y, Ye H, Adalsteinsson E, Griswold MA, Wald LL. Fast group matching for MR fingerprinting reconstruction. Magn Reson Med 2015;74(2):523–528. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Hamilton JI, Jiang Y, Chen Y, Ma D, Lo WC, Griswold M, Seiberlich N. MR fingerprinting for rapid quantification of myocardial T1, T2, and proton spin density. Magn Reson Med 2017;77(4):1446–1458. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Chen Y, Panda A, Pahwa S, Hamilton JI, Dastmalchian S, McGivney DF, Ma D, Batesole J, Seiberlich N, Griswold MA, Plecha D, Gulani V. Three-dimensional MR Fingerprinting for Quantitative Breast Imaging. Radiology 2019;290(1):33–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Cohen O, Zhu B, Rosen MS. MR fingerprinting Deep RecOnstruction NEtwork (DRONE). Magn Reson Med 2018;80(3):885–894. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Chen Y, Fang Z, Hung SC, Chang WT, Shen D, Lin W. High-resolution 3D MR Fingerprinting using parallel imaging and deep learning. Neuroimage 2020;206:116329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Fang Z, Chen Y, Liu M, Xiang L, Zhang Q, Wang Q, Lin W, Shen D. Deep Learning for Fast and Spatially Constrained Tissue Quantification From Highly Accelerated Data in Magnetic Resonance Fingerprinting. IEEE Trans Med Imaging 2019;38(10):2364–2374. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Hamilton JI, Currey D, Rajagopalan S, Seiberlich N. Deep learning reconstruction for cardiac magnetic resonance fingerprinting T1 and T2 mapping. Magn Reson Med 2021;85(4):2127–2135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Kim B, Schar M, Park H, Heo HY. A deep learning approach for magnetization transfer contrast MR fingerprinting and chemical exchange saturation transfer imaging. Neuroimage 2020;221:117165. [DOI] [PubMed] [Google Scholar]
- 26.Kang B, Kim B, Schar M, Park H, Heo HY. Unsupervised learning for magnetization transfer contrast MR fingerprinting: Application to CEST and nuclear Overhauser enhancement imaging. Magn Reson Med 2021;85(4):2040–2054. [DOI] [PubMed] [Google Scholar]
- 27.Perlman O, Farrar CT, Heo HY. MR fingerprinting for semisolid magnetization transfer and chemical exchange saturation transfer quantification. NMR Biomed 2022:e4710. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Cloos MA, Knoll F, Zhao T, Block KT, Bruno M, Wiggins GC, Sodickson DK. Multiparametric imaging with heterogeneous radiofrequency fields. Nat Commun 2016;7:12445. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Chen Y, Jiang Y, Pahwa S, Ma D, Lu L, Twieg MD, Wright KL, Seiberlich N, Griswold MA, Gulani V. MR Fingerprinting for Rapid Quantitative Abdominal Imaging. Radiology 2016;279(1):278–286. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Ma D, Coppo S, Chen Y, McGivney DF, Jiang Y, Pahwa S, Gulani V, Griswold MA. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med 2017;78(5):1781–1789. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Korzdorfer G, Jiang Y, Speier P, Pang J, Ma D, Pfeuffer J, Hensel B, Gulani V, Griswold M, Nittka M. Magnetic resonance field fingerprinting. Magn Reson Med 2019;81(4):2347–2359. [DOI] [PubMed] [Google Scholar]
- 32.Ropele S, Filippi M, Valsasina P, Korteweg T, Barkhof F, Tofts PS, Samson R, Miller DH, Fazekas F. Assessment and correction of B1-induced errors in magnetization transfer ratio measurements. Magn Reson Med 2005;53(1):134–140. [DOI] [PubMed] [Google Scholar]
- 33.Gloor M, Scheffler K, Bieri O. Quantitative magnetization transfer imaging using balanced SSFP. Magn Reson Med 2008;60(3):691–700. [DOI] [PubMed] [Google Scholar]
- 34.Windschuh J, Zaiss M, Meissner JE, Paech D, Radbruch A, Ladd ME, Bachert P. Correction of B1-inhomogeneities for relaxation-compensated CEST imaging at 7 T. NMR Biomed 2015;28(5):529–537. [DOI] [PubMed] [Google Scholar]
- 35.Schuenke P, Windschuh J, Roeloffs V, Ladd ME, Bachert P, Zaiss M. Simultaneous mapping of water shift and B1 (WASABI)-Application to field-Inhomogeneity correction of CEST MRI data. Magn Reson Med 2017;77(2):571–580. [DOI] [PubMed] [Google Scholar]
- 36.Zhou J, Zaiss M, Knutsson L, Sun PZ, Ahn SS, Aime S, Bachert P, Blakeley JO, Cai K, Chappell MA, Chen M, Gochberg DF, Goerke S, Heo HY, Jiang S, Jin T, Kim SG, Laterra J, Paech D, Pagel MD, Park JE, Reddy R, Sakata A, Sartoretti-Schefer S, Sherry AD, Smith SA, Stanisz GJ, Sundgren PC, Togao O, Vandsburger M, Wen Z, Wu Y, Zhang Y, Zhu W, Zu Z, van Zijl PCM. Review and consensus recommendations on clinical APT-weighted imaging approaches at 3T: Application to brain tumors. Magn Reson Med 2022;88(2):546–574. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Heo HY, Han Z, Jiang S, Schar M, van Zijl PCM, Zhou J. Quantifying amide proton exchange rate and concentration in chemical exchange saturation transfer imaging of the human brain. Neuroimage 2019;189:202–213. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Kang B, Heo H-Y, Park H. Only-Train-Once MR Fingerprinting for Magnetization Transfer Contrast Quantification. Medical Image Computing and Computer Assisted Intervention – MICCAI 2022; 2022; Cham. Springer Nature Switzerland. p 387–396. (Medical Image Computing and Computer Assisted Intervention – MICCAI 2022). [Google Scholar]
- 39.Heo HY, Zhang Y, Lee DH, Hong X, Zhou J. Quantitative assessment of amide proton transfer (APT) and nuclear overhauser enhancement (NOE) imaging with extrapolated semi-solid magnetization transfer reference (EMR) signals: Application to a rat glioma model at 4.7 Tesla. Magn Reson Med 2016;75(1):137–149. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Quesson B, Thiaudiere E, Delalande C, Chateil JF, Moonen CT, Canioni P. Magnetization transfer imaging of rat brain under non-steady-state conditions. Contrast prediction using a binary spin-bath model and a super-lorentzian lineshape. J Magn Reson 1998;130(2):321–328. [DOI] [PubMed] [Google Scholar]
- 41.Khlebnikov V, Windschuh J, Siero JC, Zaiss M, Luijten PR, Klomp DW, Hoogduin H. On the transmit field inhomogeneity correction of relaxation-compensated amide and NOE CEST effects at 7 T. NMR Biomed 2017;30(5). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Sun PZ, Farrar CT, Sorensen AG. Correction for artifacts induced by B(0) and B(1) field inhomogeneities in pH-sensitive chemical exchange saturation transfer (CEST) imaging. Magn Reson Med 2007;58(6):1207–1215. [DOI] [PubMed] [Google Scholar]
- 43.Hochreiter S, Schmidhuber J. Long short-term memory. Neural Comput 1997;9(8):1735–1780. [DOI] [PubMed] [Google Scholar]
- 44.Graves A, Schmidhuber J. Framewise phoneme classification with bidirectional LSTM and other neural network architectures. Neural Netw 2005;18(5–6):602–610. [DOI] [PubMed] [Google Scholar]
- 45.Kingma DP, Ba J. Adam: A Method for Stochastic Optimization. 2014. p arXiv:1412.6980.
- 46.Kang B, Kim B, Park H, Heo HY. Learning-based optimization of acquisition schedule for magnetization transfer contrast MR fingerprinting. NMR Biomed 2022;35(5):e4662. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Heo HY, Zhang Y, Lee DH, Jiang S, Zhao X, Zhou J. Accelerating chemical exchange saturation transfer (CEST) MRI by combining compressed sensing and sensitivity encoding techniques. Magn Reson Med 2017;77(2):779–786. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Togao O, Hiwatashi A, Keupp J, Yamashita K, Kikuchi K, Yoshiura T, Yoneyama M, Kruiskamp MJ, Sagiyama K, Takahashi M, Honda H. Amide Proton Transfer Imaging of Diffuse Gliomas: Effect of Saturation Pulse Length in Parallel Transmission-Based Technique. PLoS One 2016;11(5):e0155925. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Heo HY, Xu X, Jiang S, Zhao Y, Keupp J, Redmond KJ, Laterra J, van Zijl PCM, Zhou J. Prospective acceleration of parallel RF transmission-based 3D chemical exchange saturation transfer imaging with compressed sensing. Magn Reson Med 2019;82(5):1812–1821. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Chu ML, Chang HC, Oshio K, Chen NK. A single-shot T2 mapping protocol based on echo-split gradient-spin-echo acquisition and parametric multiplexed sensitivity encoding based on projection onto convex sets reconstruction. Magn Reson Med 2018;79(1):383–393. [DOI] [PubMed] [Google Scholar]
- 51.Kim M, Gillen J, Landman BA, Zhou J, van Zijl PC. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn Reson Med 2009;61(6):1441–1450. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Nehrke K, Bornert P. DREAM--a novel approach for robust, ultrafast, multislice B(1) mapping. Magn Reson Med 2012;68(5):1517–1526. [DOI] [PubMed] [Google Scholar]
- 53.Cohen O, Huang S, McMahon MT, Rosen MS, Farrar CT. Rapid and quantitative chemical exchange saturation transfer (CEST) imaging with magnetic resonance fingerprinting (MRF). Magn Reson Med 2018;80(6):2449–2463. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Barbieri M, Lee PK, Brizi L, Giampieri E, Solera F, Castellani G, Hargreaves BA, Testa C, Lodi R, Remondini D. Circumventing the curse of dimensionality in magnetic resonance fingerprinting through a deep learning approach. NMR Biomed 2022;35(4):e4670. [DOI] [PubMed] [Google Scholar]
- 55.Hoppe E, Korzdorfer G, Wurfl T, Wetzl J, Lugauer F, Pfeuffer J, Maier A. Deep Learning for Magnetic Resonance Fingerprinting: A New Approach for Predicting Quantitative Parameter Values from Time Series. Stud Health Technol Inform 2017;243:202–206. [PubMed] [Google Scholar]
- 56.Oksuz I, Cruz G, Clough J, Bustin A, Fuin N, Botnar RM, Prieto C, King AP, Schnabel JA. Magnetic Resonance Fingerprinting Using Recurrent Neural Networks. I S Biomed Imaging 2019:1537–1540. [Google Scholar]
- 57.Cohen O, Yu VY, Tringale KR, Young RJ, Perlman O, Farrar CT, Otazo R. CEST MR Fingerprinting (CEST-MRF) for Brain Tumor Quantification Using EPI Readout and Deep Learning Reconstruction. Magnetic Resonance in Medicine 2023;89(1):233–249. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Boudreau M, Stikov N, Pike GB. B1 -sensitivity analysis of quantitative magnetization transfer imaging. Magn Reson Med 2018;79(1):276–285. [DOI] [PubMed] [Google Scholar]
- 59.Boudreau M, Pike GB. Sensitivity regularization of the Cramer-Rao lower bound to minimize B1 nonuniformity effects in quantitative magnetization transfer imaging. Magn Reson Med 2018;80(6):2560–2572. [DOI] [PubMed] [Google Scholar]
- 60.Perlman O, Farrar CT, Cohen O. Deep learning global schedule optimization for chemical exchange saturation transfer MR fingerprinting (CEST-MRF). 2020; 29th Annual Meeting of ISMRM Virtual Conference & Exhibition. [Google Scholar]
- 61.Bo Z, Haldar JP, Congyu L, Dan M, Yun J, Griswold MA, Setsompop K, Wald LL. Optimal Experiment Design for Magnetic Resonance Fingerprinting: Cramer-Rao Bound Meets Spin Dynamics. IEEE Trans Med Imaging 2019;38(3):844–861. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Kang B, Heo H-Y, Park H. Learning-based prediction of encoding capability for acquisition schedule of Magnetization Transfer Contrast MR fingerprinting. 2022; 31st Annual Meeting of ISMRM London, UK. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.