We would like to thank Krieter and Wanner for their kind remarks on our recently published study [1]. The authors have raised some interesting issues, and we are grateful for the opportunity to address them.
First, we agree with the authors that, with a molecular weight of 45 kDa and a Stokes’ radius of 2.8 nm, the lambda free light chain (λFLC) dimer is a particularly interesting biomarker, because it is ideally suited to characterize the typical cut-off range of medium cut-off (MCO) dialysis membranes. Both kappa and lambda light chains, 22 and 45 kDa, respectively, could be considered a good differential marker of depuration efficacy, especially λFLC in the 40–45 kDa molecular weight range, where there are no clear differentiating markers.
Second, Krieter and Wanner wonder whether the λFLC determinations used in our study were based on an assay with monoclonal or polyclonal antibodies. Indeed, we should have specified the assay used to determine FLCs in our study. There are two main FLC assays: N Latex FLC (Siemens, Marburg, Germany), based on the use of a monoclonal specific antibody, and Freelite® (The Binding Site, Birmingham, UK), based on polyclonal specific antibodies and used for nephelometric or turbidimetric instrument platforms. In our study, serum concentrations of κ and λFLC were measured by nephelometry on a BN™ II System analyzer (Siemens Healthineers), using the Freelite® Human Lambda and Kappa Free kits for use on the Siemens BNTM II, references LK018.T and LK016.T, respectively (The Binding Site). This test is a sensitive latex-enhanced immunoassay based on polyclonal antisera. We agree with the authors that the λFLC polyclonal assay would be better suited to detect differences between high-flux and MCO dialyzers.
Assuming that only the polyclonal assay is appropriate for discriminating between the hemodialysis efficacy of different MCO dialyzers, it must be asked whether our results are comparable and coherent. A study of six patients by Krieter et al. [2] found a λFLC reduction ratio (RR) of 28 ± 4% with ELISIO HX17 (surface area 1.7 m2) and 39 ± 13% with Theranova 400 (surface area 1.7 m2). Our study of 23 patients [1] obtained a λFLC RR of 44.0 ± 8.0% with ELISIO HX19 (surface area 1.9 m2) and 48.3 ± 7.4% with Theranova 400. These slightly higher values can be partly explained by differences in the prescription parameters of the dialysis treatment: dialyzer surface area, blood flow (300 vs 439 ± 26 mL/min) and dialysis duration (240 vs 288 ± 17 min). Results consistent with these two studies have recently been published by Martínez-Miguel et al. [3]. Their study, with 14 patients, blood flow 384 mL/min and dialysis duration 227 min, obtained a λFLC RR of 35.9 ± 10.1% with ELISIO HX21 (surface area 2.1 m2) and 45.5 ± 10.0% with Theranova 500 (surface area 2.0 m2).
Considering that the values of free λFLC RR varied between 20% and 60% in our study [1], there were no significant differences in the λFLC RRs between the four MCO dialyzer treatments, and all of them were significantly higher than those obtained with high-flux HD treatments and significantly lower than those obtained with hemodiafiltration. Therefore, λFLC could be considered a good differential marker of dialysis treatment efficacy, as it allows clear discrimination between high-flux HD, MCO dialyzer hemodialysis and high-volume hemodiafiltration treatments.
We would like to thank Krieter and Wanner again for their interest in our study.
Contributor Information
Francisco Maduell, Department of Nephrology, Hospital Clínic Barcelona, Barcelona, Spain.
José Jesús Broseta, Department of Nephrology, Hospital Clínic Barcelona, Barcelona, Spain.
Maria del Carmen Salgado, Department of Biochemistry, Hospital Clínic Barcelona, Barcelona, Spain.
Nayra Rico, Department of Biochemistry, Hospital Clínic Barcelona, Barcelona, Spain.
CONFLICT OF INTEREST STATEMENT
The authors declare no financial support for the project. F.M. has received consultancy and lecture fees from Baxter, Fresenius Medical Care, Medtronic, Nipro, Toray and Vifor. The other authors declare no conflicts of interest.