Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2024 May 2.
Published in final edited form as: Biomater Sci. 2023 May 2;11(9):3297–3307. doi: 10.1039/d3bm00167a

In vivo evaluation of compliance mismatch on intimal hyperplasia formation in small diameter vascular grafts

Yuan Yao 1,, Grace Pohan 1,, Marie FA Cutiongco 2,3,^, YeJin Jeong 1, Joshua Kunihiro 1, Aung Moe Zaw 1, Dency David 1, Hanyue Shangguan 4,5,6, Alfred C H Yu 4,5,6, Evelyn KF Yim 1,2,5,6,*
PMCID: PMC10160004  NIHMSID: NIHMS1884914  PMID: 36943136

Abstract

Small diameter synthetic vascular grafts have high failure rate due to the thrombosis and intimal hyperplasia formation. Compliance mismatch between the synthetic graft and native artery has been speculated to be one of the main causes of intimal hyperplasia. However, changing the compliance of synthetic materials without altering material chemistry remains a challenge. Here, we used poly(vinyl alcohol) (PVA) hydrogel as a graft material due to its biocompatibility and tunable mechanical properties to investigate the role of graft compliance in the development of intimal hyperplasia and in vivo patency. Two groups of PVA small diameter grafts with low compliance and high compliance were fabricated by dip casting method and implanted in a rabbit carotid artery end-to-side anastomosis model for 4 weeks. We demonstrated that the grafts with compliance that more closely matched with rabbit carotid artery had lower anastomotic intimal hyperplasia formation and higher graft patency compared to low compliance grafts. Overall, this study suggested that reducing the compliance mismatch between the native artery and vascular grafts is beneficial for reducing intimal hyperplasia formation.

1. Introduction

Cardiovascular disease remains the leading cause of death around the globe, representing 32% of deaths worldwide.1 Vascular reconstruction surgeries are common surgical interventions to restore blood perfusion. Autologous grafts, such as internal thoracic arteries and saphenous veins from patients, are common vascular graft choices; however, autologous grafts have limited availability.2 Vascular grafts made of synthetic materials, such as expanded polytetrafluoroethylene (ePTFE) or polyethylene terephthalate (PET), are currently used as alternative to autologous grafts, and have shown satisfactory patency when replacing large diameter (>6mm) vessels.3, 4 However, currently available small diameter synthetic vascular grafts have low patency rate due to thrombosis and intimal hyperplasia.5 The absence of endothelial cell coverage on the luminal surfaces of synthetic grafts and the compliance mismatch between the synthetic vascular graft and native artery have been proposed as the main causes of graft failure.6, 7

Compliance mismatch has long been hypothesized to promote intimal hyperplasia formation,8, 9 which is detrimental to the patency of vascular grafts. Intimal hyperplasia is frequently observed at the anastomotic regions, and biophysical and biomechanical factors, such as anastomotic geometry and mechanical properties of vascular grafts, are important contributors.1012 Computational fluid dynamics studies have demonstrated that implantation of non-compliant vascular grafts disturbs blood flow patterns, creating flow separation, recirculation and stagnation zones near the distal anastomoses.13 Compliance mismatch has also been shown to decrease wall shear stress, create oscillating wall shear stress, and increase suture line stresses.14, 15 All these effects could cause endothelial injury and activation, smooth muscle cell migration and proliferation, and ultimately intimal hyperplasia formation.16, 17 Although many simulation studies have demonstrated the potential role of compliance mismatch in the development of intimal hyperplasia, most simulation studies were developed based on simplified assumptions and do not accurately represent in vivo flow conditions. Additionally, contradictory in vivo results were reported. Some studies found that compliance mismatch significantly promoted intimal hyperplasia,1820 while others suggested that compliance alone had minimal effect.21 In vivo studies demonstrating compliance matching grafts could reduce intimal hyperplasia were mainly comparing synthetic grafts with autologous grafts. Few studies directly relate compliance-mismatch and the resulting changes in the development of intimal hyperplasia in vivo because of confounding factors of graft properties.22 Thus, clinically relevant in vivo studies using synthetic vascular grafts with tunable compliance without affecting material chemistry are needed to elucidate the effect of compliance mismatch on intimal hyperplasia formation and graft patency.

Poly(vinyl alcohol) (PVA) is a non-toxic polymer and has been widely used in various biomedical applications.23 PVA-based hydrogels have also shown potential in tissue engineering.24 Our previous studies demonstrated that the PVA hydrogels crosslinked by sodium trimetaphosphate (STMP) have good biocompatibility and low platelet adhesion and thrombin generation,2527 suggesting the STMP-crosslinked PVA hydrogel is a promising graft material. Additionally, we developed a dip cast method to fabricate small diameter PVA vascular grafts, allowing us to manipulate the mechanical properties of the PVA vascular grafts by changing the number of dip coating layers.27 Our previous study showed that by reducing the number of dip coating layers, we were able to reduce the thickness and increase the compliance of PVA vascular grafts without the need of additional chemical modification.28

In this study, we use PVA small diameter vascular grafts fabricated with different thicknesses to systematically study the effect of compliance mismatch on the intimal hyperplasia formation in an in vivo rabbit model. We hypothesize that compliance mismatch has a direct effect on intimal hyperplasia formation, and the decrease in compliance mismatch will reduce intimal hyperplasia and improve graft patency. Small diameter PVA vascular grafts were fabricated by dip casting method. We implanted PVA small diameter grafts with different compliance in a rabbit carotid artery end-to-side model and examined the graft performance and intimal hyperplasia formation. Our study demonstrated decreasing compliance mismatch between graft and native artery has a beneficial effect on graft patency.

2. Materials and methods

2.1. Fabrication of PVA tubular grafts

PVA grafts were fabricated by dip casting method as previously reported.27 Briefly, 10% (w/v) PVA (Sigma-Aldrich, 85–124kDa, 87–89% hydrolyzed) solution was prepared by dissolving PVA in deionized (DI) water at 121°C for 20 min. 30 mL 10% PVA solution was mixed thoroughly with 2.5 mL 15% (w/v) STMP (Sigma-Aldrich) solution, followed by the addition of 1 mL 30% (w/v) sodium hydroxide (NaOH, Sigma-Aldrich) solution. The mixture was stirred to ensure homogeneity and subsequently centrifuged to remove bubbles. After centrifuge, the PVA crosslinking solution was used directly for fabrication.

To fabricate tubular grafts, flexible metal cylinder rods were used as dipping molds. The molds were treated with air plasma for 1 minute. The freshly plasma-treated molds were immersed immediately in crosslinking PVA solution and dip-coated with PVA crosslinking solution repeatedly, as shown in Figure 1A. The grafts with different compliance were fabricated by adjusting the number of dips. Upon completion of the last dip, the PVA-coated molds were moved in a cabinet with controlled temperature (20 °C) and humidity (60%–70%). To fabricate curved grafts, after overnight crosslinking in the cabinet, the PVA-coated molds were bent on a cylinder rod and continued to crosslink in cabinet for 3 days.

Figure 1.

Figure 1.

Fabrication of PVA small diameter vascular grafts with low compliance (LC) and high compliance (HC). (A) Schematic diagram of PVA small diameter vascular graft fabrication procedure. (B) Representative pictures of low compliance and high compliance PVA grafts.

2.2. Wall thickness measurement

To determine the wall thickness of PVA tubular grafts, 1 mm length cross-sectional segments were made by cutting the graft with a blade at different regions along the graft. The cross-sections of the PVA ring were imaged with a camera, and the thicknesses were measured in ImageJ. For each individual graft, the wall thickness values of different regions were averaged to get the wall thickness results for the graft.

2.3. Burst pressure measurement

To determine burst pressure, the PVA graft was clamped at one end and the other end was connected to a nitrogen gas cylinder.28 A pressure regulator was used to control the gas supply to gradually increase the pressure. The pressure at which the graft burst was recorded as burst pressure.

2.4. Compliance measurement

Compliance of PVA tubular grafts was determined by measuring the outer diameter change at internal pressure of 80 mmHg and 120 mmHg.28 The PVA graft was clamped and fixed on a cutting board. The other end of the graft was connected to a saline bag through an intravenous administration tubing. The 80 mmHg and 120 mmHg internal pressure was achieved by changing the height of the saline bag. The graft was imaged with a camera, and the outer diameter of the graft when being exposed to internal pressure of 80 mmHg (D80) and 120 mmHg (D120) was measured in ImageJ. The compliance was calculated using the equation: Compliance (%/40 mmHg) = (D120-D80)/D80 × 100%. Each graft was measured at 9 regions along the graft and the average compliance value was recorded as the compliance of the graft.

2.5. Implantation of PVA vascular grafts in rabbit right common carotid artery end-to-side anastomosis model

PVA grafts with low compliance and high compliance were sterilized with gamma irradiation,29 and subsequently implanted in rabbit right common carotid artery with end-to-side anastomosis technique, as previously reported.30 Studies were done in accordance with approved protocol (AUPP 18–10 and AUPP 16–09) by the Animal Care Committee according to the Canadian Council on Animal Care’s Guidelines. Briefly, 9 male New Zealand White rabbits (Charles River laboratories, body weight of 3.5–4.0 kg) were used in this study. Animals were randomly separated into 2 groups: low compliance group (n=5) and high compliance group (n=4). During the surgery, the animals were anaesthetized with 2–3% gas isoflurane through endotracheal intubation. The right common carotid artery of the animal was exposed. Anticoagulant (heparin, 200 IU/kg) was given intravenously prior to arterial clamping. Five low compliance PVA graft and four high compliance PVA grafts with inner diameter of 1.7 mm, length of 2 cm, and curvature of 1.4 cm radius in the middle were implanted on the rabbit right common carotid arteries with interrupted suturing. Ultrasound imaging were done at 10–14 days (midpoint) after surgery and 27–30 days (endpoint) after surgery to detect blood flow and graft patency using SonixTouch (Analogic Ultrasound, Peabody) or Siemens Acuson (X300 ultrasound system). All scans were done with a 10.0 MHz linear array transducer. Imaging focal depth was positioned at the center of the graft lumen. Image persistence settings were enabled to yield images with reduced speckle appearance. Image-guided Doppler spectrogram measurements (10 MHz frequency, 1–10 kHz pulse repetition frequency) were also performed to assess the graft’s flow velocity profile over time. The Doppler range gate was positioned at the center of the graft lumen, and angle correction was 60–70°.

2.6. Tissue processing and histological analysis

The standard practice to investigate stenosis and intimal hyperplasia requires pre-mortal pressurized perfusion. However, due to animal ethical protocol limitation, pre-mortal perfusion could not be performed. To harvest the graft at endpoint, the animals were administrated with heparin (500 IU) and euthanized with sodium pentobarbital (100 mg/kg). The PVA grafts and the connected native arteries were explanted and fixed in 4% paraformaldehyde (PFA) for 72 h. After fixation, the samples were dehydrated and embedded in paraffin. The paraffin-embedded tissue blocks were then sectioned with a microtome at the proximal and distal anastomoses.

Hematoxylin and eosin (H&E) staining, Masson’s trichrome staining, and immunohistochemistry staining were done on the cross-sectional tissue slides from the anastomoses. The stained slides were imaged with a microscope and analyzed in ImageJ software. H&E-stained slides were used to analyze the intimal hyperplasia percentage by dividing the area of intimal hyperplasia to the total luminal area of anastomoses. Masson’s trichrome staining was done with a Trichrome staining kit (Abcam) according to the supplier’s protocol. The images of the Masson’s trichrome-stained slides were processed with a Color Deconvolution software in ImageJ. The area of collagen and muscle fibers were measured in ImageJ and the percentage was calculated by dividing the area of collagen or muscle fibers to the total luminal area of anastomoses and the area of the intimal hyperplasia. Immunohistochemistry (IHC) staining of α-smooth muscle actin (α-SMA), smooth muscle myosin heavy chain (MHC), and smooth muscle protein 22α (SM22α) was done to visualize smooth muscle cells and characterize smooth muscle cell phenotype. The IHC staining was performed with Abcam mouse HRP/DAB (ABC) detection IHC kit, with primary antibodies of mouse anti-α-SMA, mouse anti-smooth muscle MHC I, and goat anti-TAGLN/Transgelin antibody from Abcam, and biotinylated goat anti-mouse or donkey anti-goat secondary antibodies. The samples were counter-stained with hematoxylin from Abcam. The quantification of IHC staining was done in Fiji with a color deconvolution function. The percentage of positive area was calculated by dividing the area of positive region to the area of intimal hyperplasia.

2.8. Statistical analysis

Statistical analysis was performed using GraphPad Prism 8. The data were presented as mean ± Standard Deviation (SD). The number of biological replicas was indicated in the legend of each figure. Statistical analysis was done with t-test and two-way ANOVA with Sidak’s post hoc test. Statistical significance was set to be p<0.05.

2.9. Ethical statement

Rabbit end-to-side studies were approved (#AUPP 16–09 and #AUPP 18–10) by the Animal Care Committee according to the Canadian Council on Animal Care’s Guidelines, the requirements of Province of Ontario’s Animals for Research Act, and the University of Waterloo’s Guidelines for the Use of Animals in Research and Teaching. Rabbit end-to-end studies were done in accordance with approved guidelines of the Institutional Animal Care and Use Committee of the National University of Singapore.

3. Results

3.1. Fabrication of grafts with low compliance (LC) and high compliance (HC)

PVA grafts with low compliance (LC) and high compliance (HC) were fabricated as shown in Figure 1A. LC grafts were fabricated with 8 layers of dip casting while HC grafts were fabricated with 6 layers of dip casting. Photos of LC and HC grafts showed that LC grafts were thicker compared to HC grafts (Figure 1B). Both the LC and HC grafts were curved during the fabrication to avoid kinking after implantation. The measurement of wall thickness showed that LC grafts had an average wall thickness of 0.409 ± 0.047 mm, which is significantly higher than HC grafts, which has a wall thickness of 0.269 ± 0.034 mm (P=0.0133, n=3) (Figure 2A). The different wall thickness also yielded distinct burst pressure and compliance. LC and HC grafts had burst pressure of 691.6 ± 104.2 mmHg and 439.2 ± 72.0 mmHg, respectively (Figure 2B). The compliance of LC and HC grafts is 1.23 ± 0.16 %/40mmHg and 1.83 ± 0.27 %/40mmHg, respectively, representing 48% and 71% matching to the compliance of native rabbit carotid artery (2.57 ± 0.27 %/40mmHg) (Figure 2C). The results demonstrated that by using dip casting fabrication, we were able to fabricate PVA grafts with significantly different compliance without the need to alter material chemistry. The HC grafts also maintained sufficient burst pressure.

Figure 2.

Figure 2.

Mechanical properties of low compliance (LC) grafts and high compliance (HC) grafts. (A) Wall thickness, (B) burst pressure, and (C) compliance of LC and HC grafts. n=3, * indicates a significant difference using t-test with p<0.05. Red dash line indicates the compliance of native rabbit carotid artery.

3.2. In vivo performance of grafts with low compliance (LC) and high compliance (HC) in end-to-side model

In order to study the effect of compliance on graft performance in an end-to-side anastomosis model, we implanted 5 LC grafts and 4 HC grafts in rabbit right common carotid artery, as shown in Figure 3A. The grafts had inner diameter of 1.7 mm and length of 2 cm. A curvature was created during the graft fabrication to prevent kinking. Additionally, the implantation angle at the two anastomoses was approximately 45°, and the distance between the 2 anastomoses was around 1.5 cm. After 1 month implantation, we found 2 out of 5 LC grafts were patent (40% patency), while 3 out of 4 HC grafts were patent (75% patency) (Figure 3B).

Figure 3.

Figure 3.

Implantation of PVA small diameter vascular grafts with low compliance (LC) and high compliance (HC) in rabbit right common carotid artery in an end-to-side anastomotic model. (A) Representative photos of low compliance (LC) PVA grafts and high compliance (HC) PVA grafts after implantation. (B) 1-month patency rate of LC and HC grafts.

We used ultrasound imaging to detect the blood flow inside the implanted grafts at both midpoint (day 10–14) and endpoint (day 27–30). We detected stronger blood flow inside the HC grafts compared to LC grafts at both midpoint and endpoint (Figure 4A). Doppler measurement of blood flow velocity showed that flow velocity in HC grafts was higher at both midpoint and endpoint, and at the endpoint, the flow velocity in HC grafts was significantly higher than in LC grafts (Figure 4B).

Figure 4.

Figure 4.

In vivo patency of PVA small diameter vascular grafts with low compliance (LC) and high compliance (HC). (A) Representative ultrasound images of blood flow inside the grafts at midpoint (Day 10–14) and endpoint (Day 27–30). Red arrow indicates the direction of bloof flow. (B) Midpoint and endpoint flow velocity measurement. n=5 for LC grafts, and n=4 for HC grafts. * indicates a significant difference using t-test with p<0.05.

3.3. Histological analysis of explanted low compliance (LC) and high compliance (HC) grafts

Anastomotic intimal hyperplasia formation has been identified as the main cause of failure of small diameter vascular grafts.31 H&E staining was done at the proximal and distal anastomoses for both LC and HC grafts to analyze the intimal hyperplasia formation (Figure 5A and B). HC grafts showed decreased area of intimal hyperplasia compared to LC grafts at both proximal and distal anastomoses. Additionally, intimal hyperplasia was formed mainly on the arteries. No intimal hyperplasia was observed on the graft luminal surfaces regardless of graft compliance. Quantification results showed that the intimal hyperplasia on HC grafts was significantly lower than that on LC grafts at the proximal anastomosis, while slightly but not significantly lower at the distal anastomosis. A noticeable animal-to-animal variation of intimal hyperplasia area was observed at the distal anastomosis of HC grafts (Figure 5B).

Figure 5.

Figure 5.

Intimal hyperplasia formation at the anastomoses of grafts with low compliance (LC) and high compliance (HC). (A) Representative hematoxylin and eosin (H&E) staining images of proximal and distal anastomoses. Part of the PVA graft were broken during the tissue processing and staining procedure in the HC graft. (B) Schematic diagram of locations of proximal and distal anastomoses. (C) Intimal hyperplasia percentage at proximal and distal anastomoses. n=5 for LC grafts, and n=4 for HC grafts. * indicates a significant difference using t-test with p<0.05. N.S. indicates no significant difference.

Masson’s trichrome staining demonstrated the presence of collagen and muscle fibers in the intimal hyperplasia (Figure 6A). Quantification results showed that, when dividing the area of collagen and muscle fibers by the total luminal area, the percentage of collagen and the percentage of muscle fibers for LC grafts at both the proximal and distal anastomoses were higher compared to those for HC grafts, although the differences were not statistically significant (Figure 6B). We further divided the area of collagen and the area of muscle fibers by the total area of intimal hyperplasia, no significant difference was observed.

Figure 6.

Figure 6.

Composition of intimal hyperplasia formed at the anastomoses of grafts with different compliance. (A) Representative Masson’s trichrome staining images of the stenosis in proximal and distal anastomose. (B) Percentage of collagen and muscle fiber area in the lumen of the grafts. (C) Percentage of collagen and muscle fiber area in the stenosis. n=5 for LC grafts, and n=4 for HC grafts. N.S. indicates no significant difference.

Immunohistochemistry staining of α-smooth muscle actin (α-SMA) showed abundant vascular smooth muscle cells in the intimal hyperplasia region of both LC grafts and HC grafts (Figure 7). More α-SMA positive area was observed in the intimal hyperplasia of HC grafts. The quantification of α-SMA positive area is as shown in Figure 10A. HC grafts had slightly higher percentage of α-SMA positive region in proximal anastomoses and significantly higher α-SMA positive region in the distal anastomoses compared to LC grafts. Smooth muscle myosin heavy chain (SM MHC) is a specific marker of differentiated smooth muscle cells.32 The immunohistochemistry staining of LC and HC grafts at the anastomoses region is as shown in Figure 8. Higher percentage of MHC positive region was observed in the proximal anastomoses of LC grafts and both proximal and distal anastomoses of HC grafts. The quantification results showed that the MHC positive region at the distal anastomoses of HC grafts was significantly higher than that of LC grafts (Figure 10B). Smooth muscle 22-α (SM22α) has been shown to modulate SMC phenotype.33 We stained the anastomoses with SM22α (Figure 9) and quantified the percentage of positive region. Both the LC and HC grafts havd high percentage of SM22α positive area at proximal and distal anastomoses (Figure 10C).

Figure 7.

Figure 7.

Immunohistochemistry staining of α-smooth muscle actin (α-SMA) for the PVA low compliance (LC) grafts and high compliance (HC) grafts at the proximal and distal anastomoses.

Figure 10.

Figure 10.

Quantification of α-smooth muscle actin (α-SMA) (A), myosin heavy chain (MHC) (B), and smooth muscle protein 22-α (SM22α) (C) positive area. n=5 for LC grafts and n=4 for HC grafts. * indicates a significant difference of p<0.05 using two-way ANOVA with Sidak’s post hoc test.

Figure 8.

Figure 8.

Immunohistochemistry staining of smooth muscle myosin heavy chain (SM MHC) for the PVA low compliance (LC) grafts and high compliance (HC) grafts at the proximal and distal anastomoses.

Figure 9.

Figure 9.

Immunohistochemistry staining of smooth muscle protein 22-α (SM22α) for the PVA low compliance (LC) grafts and high compliance (HC) grafts at the proximal and distal anastomoses.

3.4. In vivo performance of low compliance (LC) and high compliance (HC) grafts in end-to-end model

In our previous study, we had implanted straight compliance matching PVA small diameter vascular grafts with 1mm internal diameter in rabbit left femoral artery using an end-to-end anastomosis technique.34, 35 In addition to the compliance matching grafts, we also implanted graft with lower compliance to study the effect of compliance on graft performance and patency in an end-to-end model. Two groups of PVA groups were implanted (Supplementary Figure 1). Compliance mismatch PVA graft group had average compliance of 4.917 %/40mmHg, while the compliance matching group had average compliance of 6.491 %/40mmHg, which is comparable to the compliance of rabbit femoral artery (5.9 ± 0.5 %/40mmHg).34 The angiography showed that the 2 grafts in compliance matching group were both patent, while the 2 grafts in the compliance mismatch group were occluded (Supplementary Figure 2). The results suggested that the compliance matching between artificial graft and native artery could potentially improve graft patency.

4. Discussion

The patency of synthetic small diameter vascular grafts remains disappointing due to fast occlusion. Intimal hyperplasia formation at the anastomosis region has been identified as one of the main causes of failure.36 Vascular grafts made of synthetic materials typically have a significantly lower compliance than native arteries, which causes localized flow disturbance, adverse shear stress, and altered suture-line stress,15, 37 thus leading to the dysfunction of endothelial cells, activation of smooth muscle cells, and consequently intimal hyperplasia formation. Previous in vivo studies showed that graft patency can be improved by matching the compliance of synthetic graft and the artery; however, most were comparing synthetic vascular grafts with autografts or comparing synthetic grafts made of different materials. Thus, a synthetic material that can modulate graft compliance without affecting material chemistry or topography is needed to understand the effect of compliance on intimal hyperplasia formation and graft patency. Our previous studies demonstrated that STMP-crosslinked PVA is a promising vascular graft material and dip casting method can be used to fabricate PVA tubular grafts.26, 35 We found that the mechanical properties of fabricated PVA tubular grafts could be modulated by manipulating the fabrication parameters.28 In this study, we were able to fabricate PVA small diameter vascular grafts with different compliance by changing the number of dipping layers without the need of altering material chemistry. The HC grafts fabricated with 6 layers of dip coating had significantly lower wall thickness and significantly higher compliance compared to LC grafts, which were fabricated with 8 layers of dip coating. More importantly, although the wall thickness of HC grafts was significantly decreased, both HC and LC grafts still maintained sufficient burst pressure to withstand physiological blood pressure, thus allowing us to study the relevance between graft compliance and patency in vivo. We previously demonstrated that gamma irradiation is a suitable sterilization method for PVA grafts.29 In the previous study, we compared the mechanical properties of PVA grafts before and after gamma irradiation. The gamma irradiation did not alter the swelling, Young’s modulus, burst pressure, and suture retention of the PVA grafts. The compliance of both thick and thin PVA grafts was increased by gamma irradiation, but grafts with thicker wall still had significantly lower compliance than grafts with thinner wall. Thus, we sterilized the HC and LC grafts with gamma-irradiation prior to implantation in this study.

To study the effect of compliance on intimal hyperplasia formation and graft patency, we implanted both the HC and LC grafts in a rabbit right common carotid artery end-to-side anastomosis model, where a graft needed to be bended to connect the 2 ends of the graft to the artery wall. Graft kinking has been a concern when using a straight graft to replace curved arterial arch, which causes blood flow turbulence and increased thrombus formation.38 In our previous study, when using straight ePTFE graft in end-to-side anastomosis model, a kink was formed on the graft; while with PVA grafts that were pre-curved, the grafts were easily attached to the arterial wall without causing significant kinking.39 Thus, in this study, both HC and LC grafts were fabricated into curved shape to prevent kinking. In end-to-side anastomosis model, the anastomotic junction angle affects the shear stress and flow patterns.4042 An anastomotic junction angle of 45° has been shown to induce low flow separation and high shear stress around the anastomosis zones, thus producing low amount of intimal thickening.40 Hence, in our study, we cut the two ends of the graft prior to implantation and implanted the grafts with an junction angle of 45°. Previous simulation studies using computational modeling reported that a greater compliance mismatch between the synthetic graft and native artery has more adverse effect on blood flow profile, causing more severe flow separation and decreased wall shear stresses.22, 43 This phenomenon has been hypothesized to initiate and amplify the intimal hyperplasia formation. However, these results were yet to be verified in vivo. In this study, we used a rabbit right common carotid artery end-to-side anastomosis model to study the effect of compliance on graft patency in vivo. Although the compliance of HC grafts is still lower than native rabbit carotid artery (2.57 %/40 mmHg), HC grafts showed a higher 1-month patency rate, stronger blood flow and reduce intimal hyperplasia area compared to LC grafts (Figure 35). The results are in line with computational modeling and demonstrated that the decrease in compliance discrepancy between a synthetic graft and artery could increase the patency rate and reduce intimal hyperplasia formation. In end-to-side anastomosis model, spatial distribution of intimal hyperplasia was reported to be preferentially located around the heel and toe at the anastomotic junctions and floor of the native arteries, due to the flow separation and decreased wall shear stress around those regions.44, 45 In our studies, we also observed that the intimal hyperplasia was mainly formed at the floor of native arteries. Although many studies focused on distal anastomotic intimal hyperplasia,22, 46 intimal hyperplasia has also been reported to be formed around proximal anastomoses.47 We observed comparable amount of intimal hyperplasia at both proximal and distal anastomoses. We speculate that the hemodynamics changed due to both graft compliance mismatch and anastomotic geometry and the surgical trauma cause activation of endothelial cells and smooth muscle cells in the current end-to-side anastomosis model, which together lead to intimal hyperplasia formation. Additionally, the PVA grafts we used had higher wall thickness compared to the native artery, which has been shown to has higher risk of surgical trauma, thus will further amplifying the formation of intimal hyperplasia.43, 48 Using the same animal model, we previously implanted ePTFE small diameter vascular grafts. The compliance of the ePTFE vascular grafts was around 0.704 %/40mmHg.39 2 out 3 ePTFE grafts were patent. In patent ePTFE grafts, intimal hyperplasia was only observed on the surface of ePTFE grafts, while in occluded ePTFE graft, significant amount of intimal hyperplasia was observed both in floor of the artery and on the surface of ePTFE grafts at both anastomoses.39 Interestingly, different from ePTFE grafts and other synthetic vascular grafts, no intimal hyperplasia was formed on the surface of PVA grafts. We speculate that this phenomenon could potentially be due to the hydrophilicity and bioinert nature of PVA hydrogels.

Intimal hyperplasia formation is characterized by vascular smooth muscle cell migration, proliferation, and secretion of extracellular matrix.49 Many studies have suggested that compliance mismatch at the anastomoses leads to increased vascular smooth muscle cell proliferation and migration, and induces phenotype changes of smooth muscle cells from contractile to proliferative phenotype.22 We did histological staining and analysis to study the composition of intimal hyperplasia. In our Masson’s trichrome staining, the presence of collagen and smooth fibers were demonstrated, and the percentage of collagen and the percentage of fiber area in the lumen of the LC grafts were slightly but not significantly higher than that of HC grafts at both proximal and distal anastomoses. However, the composition percentages of collagen and of fiber in the intimal hyperplasia were comparable between LC and HC grafts. Additionally, immunohistochemistry staining of α-SMA showed abundant α-SMA+ cells in both LC and HC grafts, indicating the presence of smooth muscle cells (Figure 67). We observed stronger α-SMA expression in the anastomotic IHC section of most HC grafts compared to LC grafts. α-SMA is a marker of contractile smooth muscle cells,50, 51 and α-SMA has been reported to negatively regulate the smooth muscle cell migration and proliferation.52 Loss of α-SMA has also been shown to contribute to increased SMC hyperplasia in vivo.53, 54 We speculate that in our study, the higher compliance synthetic vascular grafts might have induced stronger α-SMA expression compared to lower compliance grafts, thus reducing intimal hyperplasia formation. The staining of smooth muscle cell contractile protein, MHC, further supported our speculation. We found that HC grafts had more MHC positive region especially at the distal anastomoses compared to LC grafts, suggesting that more contractile cells were present in the HC grafts than LC grafts at the distal anastomoses.

Many computational simulation studies have reported that compliance affects intimal hyperplasia formation not only in end-to-side anastomosis model but also in end-to-end anastomosis model.14, 22, 55 Our previous studies implanted straight PVA small diameter vascular grafts with inner diameter of approximately 1 mm in rabbit left femoral artery end-to-end anastomosis model.34, 35 As a preliminary study when we compared the 1mm PVA vascular graft with lower compliance, we showed that compliance matching grafts were patent after 14 days of implantation, while the mismatch groups were occluded. This suggests that in end-to-end anastomosis, reducing compliance mismatch is also promising in improving vascular graft patency.

5. Conclusion

In summary, in this study we successfully fabricated PVA-based small diameter vascular grafts with different mechanical properties and studied the role of compliance in graft patency and intimal hyperplasia formation in vivo. We demonstrated that compliance mismatch between the graft and artery had a direct effect on the intimal hyperplasia formation and graft patency. Additionally, our proof-of-concept in vivo study showed the grafts with higher compliance could potentially improve graft patency, increase smooth muscle cell contractility, and reduced intimal hyperplasia formation.

Supplementary Material

ESI

Acknowledgements

This work was supported by the National Institutes of Health grants [NIH R01HL130274], NSERC-CREATE Training in Global Biomedical Technology Research and Innovation at the University of Waterloo [CREATE-509950-2018], Canadian Foundation for Innovation (CFI)-JELF 35573. The rabbit femoral study was supported by the Singapore Ministry of Health’s National Medical Research Council Exploratory/Developmental Grant Scheme (NMRC/EDG/0068/2009). The work was also partially funded by Singapore A*STAR-ANR (1122703037) and the National Research Foundation, Prime Minister’s Office, Singapore under its Research Center of Excellence programme administered by the Mechanobiology Institute. We acknowledge Dr. Filip Konecny, Ms. Sabrina Mattiassi, Ms. Erica Sharp, Ms. Rebecca Mac, Dr. Adrian Chee, Dr. Billy Yat Shun Yiu, Mr Matthew Yeo, Dr. Marek Kukumberg, and Ms. Jean Flanagan from the University of Waterloo Central Animal Facility for their help with the rabbit study, and Dr. Adrian Chee, Dr. Billy Yat Shun Yiu, and Dr. Marek Kukumberg for taking the ultrasound measurements in the in vivo work.

Footnotes

Conflict of interest statement

There are no conflicts to declare.

References

  • 1.W. H. Organization, Cardiovascular diseases (CVDs), https://www.who.int/en/news-room/fact-sheets/detail/cardiovascular-diseases-(cvds), (accessed June 11, 2021).
  • 2.Wang X, Lin P, Yao Q and Chen C, World Journal of Surgery, 2007, 31, 682–689. [DOI] [PubMed] [Google Scholar]
  • 3.Roina Y, Auber F, Hocquet D and Herlem G, Materials Today Chemistry, 2021, 20, 100412. [Google Scholar]
  • 4.Pashneh-Tala S, MacNeil S and Claeyssens F, Tissue Engineering Part B: Reviews, 2015, 22, 68–100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Walpoth BH and Bowlin GL, Expert Review of Medical Devices, 2005, 2, 647–651. [DOI] [PubMed] [Google Scholar]
  • 6.Jeong Y, Yao Y and Yim EKF, Biomaterials Science, 2020, 8, 4383–4395. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Zhuang Y, Zhang C, Cheng M, Huang J, Liu Q, Yuan G, Lin K and Yu H, Bioactive Materials, 2021, 6, 1791–1809. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Baird RN, Kidson IG, L’Italien GJ and Abbott WM, American Journal of Physiology-Heart and Circulatory Physiology, 1977, 233, H568–H572. [DOI] [PubMed] [Google Scholar]
  • 9.Baird R and Abbott W, The Lancet, 1976, 308, 948–950. [DOI] [PubMed] [Google Scholar]
  • 10.Li L, Terry CM, Shiu Y-TE and Cheung AK, Kidney International, 2008, 74, 1247–1261. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Haruguchi H and Teraoka S, Journal of Artificial Organs, 2003, 6, 227–235. [DOI] [PubMed] [Google Scholar]
  • 12.Remuzzi A and Ene-Iordache B, Clinical Journal of the American Society of Nephrology, 2013, 8, 2186. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Acuna A, Berman AG, Damen FW, Meyers BA, Adelsperger AR, Bayer KC, Brindise MC, Bungart B, Kiel AM, Morrison RA, Muskat JC, Wasilczuk KM, Wen Y, Zhang J, Zito P and Goergen CJ, Journal of Biomechanical Engineering, 2018, 140. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Owida AA, Do H and Morsi YS, Computer Methods and Programs in Biomedicine, 2012, 108, 689–705. [DOI] [PubMed] [Google Scholar]
  • 15.Ballyk PD, Walsh C, Butany J and Ojha M, Journal of Biomechanics, 1997, 31, 229–237. [DOI] [PubMed] [Google Scholar]
  • 16.Sottiurai VS, Lim Sue S, Feinberg EL II, Bringaze WL, Tran AT and Batson RC, European Journal of Vascular Surgery, 1988, 2, 245–256. [DOI] [PubMed] [Google Scholar]
  • 17.Sottiurai VS, Yao JST, Batson RC, Sue SL, Jones R and Nakamura YA, Annals of Vascular Surgery, 1989, 3, 26–33. [DOI] [PubMed] [Google Scholar]
  • 18.Brossollet LJ, 1992.
  • 19.Abbott WM, Megerman J, Hasson JE, L’Italien G and Warnock DF, Journal of vascular surgery, 1987, 5, 376–382. [PubMed] [Google Scholar]
  • 20.Stewart SFC and Lyman DJ, Annals of biomedical engineering, 2004, 32, 991–1006. [DOI] [PubMed] [Google Scholar]
  • 21.Wu MH-D, Shi Q, Sauvage LR, Kaplan S, Hayashida N, Patel MD, Wechezak AR and Walker MW, Annals of Vascular Surgery, 1993, 7, 156–168. [DOI] [PubMed] [Google Scholar]
  • 22.Post A, Diaz-Rodriguez P, Balouch B, Paulsen S, Wu S, Miller J, Hahn M and Cosgriff-Hernandez E, Acta Biomater, 2019, 89, 84–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Paradossi G, Cavalieri F, Chiessi E, Spagnoli C and Cowman MK, Journal of Materials Science: Materials in Medicine, 2003, 14, 687–691. [DOI] [PubMed] [Google Scholar]
  • 24.Jiang S, Liu S and Feng W, Journal of the Mechanical Behavior of Biomedical Materials, 2011, 4, 1228–1233. [DOI] [PubMed] [Google Scholar]
  • 25.Cutiongco MF, Anderson DE, Hinds MT and Yim EK, Acta Biomater, 2015, 25, 97–108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Chaouat M, Le Visage C, Baille WE, Escoubet B, Chaubet F, Mateescu MA and Letourneur D, Advanced Functional Materials, 2008, 18, 2855–2861. [Google Scholar]
  • 27.Cutiongco MF, Goh SH, Aid-Launais R, Le Visage C, Low HY and Yim EK, Biomaterials, 2016, 84, 184–195. [DOI] [PubMed] [Google Scholar]
  • 28.Jeong Y, Yao Y, Mekonnen TH and Yim EKF, Frontiers in Materials, 2021, 7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Pohan G, Mattiassi S, Yao Y, Zaw AM, Anderson DEJ, Cutiongco MFA, Hinds MT and Yim EKF, Tissue Engineering Part A, 2020, DOI: 10.1089/ten.TEA.2020.0002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Yao Y, Jeong Y, Zaw AM, Kukumberg M and Yim EKF, in Vascular Tissue Engineering: Methods and Protocols, eds. Zhao F and Leong KW, Springer; US, New York, NY, 2022, DOI: 10.1007/978-1-0716-1708-3_15, pp. 177–189. [DOI] [Google Scholar]
  • 31.Keynton RS, Evancho MM, Sims RL, Rodway NV, Gobin A and Rittgers SE, Journal of Biomechanical Engineering, 2001, 123, 464–473. [DOI] [PubMed] [Google Scholar]
  • 32.Nguyen AT, Gomez D, Bell RD, Campbell JH, Clowes AW, Gabbiani G, Giachelli CM, Parmacek MS, Raines EW, Rusch NJ, Speer MY, Sturek M, Thyberg J, Towler DA, Weiser-Evans MC, Yan C, Miano JM and Owens GK, Circulation Research, 2013, 112, 17–22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Feil S, Hofmann F and Feil R, Circulation Research, 2004, 94, 863–865. [DOI] [PubMed] [Google Scholar]
  • 34.Cutiongco MF, Kukumberg M, Peneyra JL, Yeo MS, Yao JY, Rufaihah AJ, Le Visage C, Ho JP and Yim EK, Front Bioeng Biotechnol, 2016, 4, 44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Cutiongco MFA, Choo RKT, Shen NJX, Chua BMX, Sju E, Choo AWL, Le Visage C and Yim EKF, Frontiers in Bioengineering and Biotechnology, 2015, 3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Obiweluozor FO, Emechebe GA, Kim D-W, Cho H-J, Park CH, Kim CS and Jeong IS, Cardiovascular Engineering and Technology, 2020, 11, 495–521. [DOI] [PubMed] [Google Scholar]
  • 37.Frauenfelder T, Boutsianis E, Schertler T, Husmann L, Leschka S, Poulikakos D, Marincek B and Alkadhi H, Biomed Eng Online, 2007, 6, 35-35. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Misfeld M, Scharfschwerdt M and Sievers H-H, The Annals of Thoracic Surgery, 2004, 78, 1060–1063. [DOI] [PubMed] [Google Scholar]
  • 39.Yao Y, Zaw AM, Anderson DEJ, Hinds MT and Yim EKF, Biomaterials, 2020, 249, 120011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Keynton RS, Rittgers SE and Shu MCS, Journal of Biomechanical Engineering, 1991, 113, 458–463. [DOI] [PubMed] [Google Scholar]
  • 41.Jackson ZS, Ishibashi H, Gotlieb AI and Langille BL, Journal of Vascular Surgery, 2001, 34, 300–307. [DOI] [PubMed] [Google Scholar]
  • 42.Kim JT, Kim H and Ryou HS, Applied Sciences, 2021, 11. [Google Scholar]
  • 43.Bouchet M, Gauthier M, Maire M, Ajji A and Lerouge S, Materials Science and Engineering: C, 2019, 100, 715–723. [DOI] [PubMed] [Google Scholar]
  • 44.Migliavacca F and Dubini G, Biomechanics and modeling in mechanobiology, 2005, 3, 235–250. [DOI] [PubMed] [Google Scholar]
  • 45.Bassiouny HS, White S, Glagov S, Choi E, Giddens DP and Zarins CK, Journal of Vascular Surgery, 1992, 15, 708–717. [DOI] [PubMed] [Google Scholar]
  • 46.Trubel W, Moritz A, Schima H, Raderer F, Scherer R, Ullrich R, Losert U and Polterauer P, ASAIO J, 1994, 40, M273–278. [DOI] [PubMed] [Google Scholar]
  • 47.Baek I, Hwang J, Park J, Kim H, Park J-S and Kim DJ, Journal of Vascular Surgery, 2015, 61, 1575–1582.e1571. [DOI] [PubMed] [Google Scholar]
  • 48.Furdella KJ, Higuchi S, Behrangzade A, Kim K, Wagner WR and Vande Geest JP, Acta Biomater, 2021, 123, 298–311. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Lemson MS, Tordoir JHM, Daemen MJAP and Kitslaar PJEHM, European Journal of Vascular and Endovascular Surgery, 2000, 19, 336–350. [DOI] [PubMed] [Google Scholar]
  • 50.Rensen SSM, Doevendans PAFM and van Eys GJJM, Netherlands Heart Journal, 2007, 15, 100–108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Xie S-A, Zhang T, Wang J, Zhao F, Zhang Y-P, Yao W-J, Hur SS, Yeh Y-T, Pang W, Zheng L-S, Fan Y-B, Kong W, Wang X, Chiu J-J and Zhou J, Biomaterials, 2018, 155, 203–216. [DOI] [PubMed] [Google Scholar]
  • 52.Chen L, DeWispelaere A, Dastvan F, Osborne WRA, Blechner C, Windhorst S and Daum G, PLOS ONE, 2016, 11, e0155726. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Papke CL, Cao J, Kwartler CS, Villamizar C, Byanova KL, Lim S-M, Sreenivasappa H, Fischer G, Pham J, Rees M, Wang M, Chaponnier C, Gabbiani G, Khakoo AY, Chandra J, Trache A, Zimmer W and Milewicz DM, Human Molecular Genetics, 2013, 22, 3123–3137. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Zhang Y.-n., Xie B.-d., Sun L, Chen W, Jiang S-L, Liu W, Bian F, Tian H and Li R-K, Journal of Cellular and Molecular Medicine, 2016, 20, 1049–1061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Stewart SFC and Lyman DJ, Annals of Biomedical Engineering, 2004, 32, 991–1006. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ESI

RESOURCES