Abstract
Objectives
To compare the effects of two types of titratable mandibular advancement devices (MADs), namely MAD-H (allowing limited vertical opening) and MAD-S (allowing free vertical opening), on respiratory parameters and upper airway dimensions in patients with mild to moderate obstructive sleep apnea (OSA).
Materials and methods
Patients with mild to moderate OSA (5 ≤ apnea–hypopnea index (AHI) < 30 /h) were randomly assigned to two parallel MAD groups. All MADs were subjectively titrated according to a standardized protocol during a 3-month follow-up. Every patient underwent two polysomnographic recordings, and two cone beam computed tomography scans in supine position: one at baseline and another one after 3 months with the MAD in situ. The primary outcome variables were the AHI in supine position (AHI-supine) and the minimal cross-sectional area of the upper airway in supine position (CSAmin-supine).
Results
A total of 49 patients were recruited, and 31 patients (21 men and 10 women) with a mean (± SD) age of 48.5 (± 13.9) years and a mean AHI of 16.6 (± 6.7) /h completed the study. In the per-protocol analysis, there was no significant difference between MAD-H (n = 16) and MAD-S (n = 15) in their effects on AHI-supine (P = 0.14) and CSAmin-supine (P = 0.59). Similar results were found in the intention-to-treat analysis (P = 0.47 and 0.57, respectively).
Conclusions
Within the limitations of this study, we conclude that there is no significant difference in the effects of an MAD allowing limited vertical opening and an MAD allowing free vertical opening on respiratory parameters and upper airway dimensions in patients with mild to moderate OSA.
Clinical relevance
MADs allowing limited vertical opening and allowing free vertical opening have similar effects on respiratory parameters and upper airway dimensions in patients with mild to moderate OSA.
Trial registration: ClinicalTrials.gov Identifier: NCT02724865. https://clinicaltrials.gov/ct2/show/NCT02724865
Keywords: Obstructive sleep apnea, Mandibular advancement device, Freedom of mandibular vertical opening, Cone beam computed tomography, Upper airway dimensions
Introduction
Obstructive sleep apnea (OSA) is characterized by recurrent obstructions of the upper airway, often resulting in oxygen desaturations and arousals from sleep [1]. Excessive daytime sleepiness and fatigue, lack of concentration, and loud snoring reported by the patient’s bed partner are frequently reported complaints [2, 3]. Furthermore, patients with untreated OSA are at increased risk of, among others, hypertension, stroke, heart failure, diabetes, and involvement in car accidents [4–7]. The diagnosis of OSA relies on the combination of symptoms, clinical signs, and objective assessment of obstructive respiratory events. The gold standard for assessing respiration during sleep in patients with OSA is full-night polysomnography (PSG). One commonly used indicator of OSA severity is the number of respiratory events (apneas and/or hypopneas) per hour of sleep during PSG recording, viz., apnea–hypopnea index (AHI). An adult showing an AHI of at least 5 events/hour is diagnosed with OSA [1]. Based on the AHI, OSA severity for adults is classified as mild (5 ≤ AHI < 15 events/hour), moderate (15 ≤ AHI ≤ 30 events/hour), or severe (AHI > 30 events/hour) [1].
Continuous positive airway pressure (CPAP) is the most efficacious therapy for OSA [8]. However, the equipment is relatively cumbersome, which results in low compliance of patients [9]. Alternatively, mandibular advancement device (MAD) therapy is recommended as a primary treatment option in patients with mild and moderate OSA, and in patients with severe OSA who refuse or are unable to tolerate CPAP therapy [10–12].
There is a large variety of commercially available customized MADs. Over the years, the customized MAD has evolved from the “mono-bloc” type of appliance, which consists of a single piece giving the mandible a fixed position, towards the current “bi-block titratable” type, which consists of two separate pieces that are dynamically interconnected allowing different degree of lateral and/or vertical movements. Understanding the treatment efficacy and working mechanism of MADs with different design features has an ongoing interest in the clinic [13, 14]. Previous studies [15–19] suggested that mono-block MADs may have higher effectiveness than bi-block titratable MADs because the mandible is protruded firmly. However, a bi-block titratable MAD is more comfortable for patients as it allows some degree of mandibular movement and has higher compliance [19]. Besides, it is easier and takes less time to titrate a bi-block titratable MAD to the optimal mandibular position. Therefore, bi-block titratable MADs are increasingly recommended [20].
Bi-block titratable MADs may differ in the freedom of mandibular vertical opening. There are two types of bi-block titratable MADs that are commonly used in the treatment of OSA: the Herbst appliance (MAD-H) and the SomnoDent appliance (MAD-S). They predominantly differ in the freedom of vertical opening: the MAD-H only allows limited vertical opening, while the MAD-S allows free vertical opening of the mandible during sleep. A retrospective study suggested no significant difference between MAD-H and MAD-S in changing the AHI [21]. However, that study was not randomized, and therefore the outcomes may be biased. As most previous studies focused on the comparisons between mono-bock MADs and bi-block MADs, comparisons of bi-block MADs with different freedom of mandibular vertical opening are limited. Therefore, more evidence is needed on the comparison between these two types of MAD.
The working mechanism of MAD is related to the improvement of the upper airway dimensions through mandibular protrusion [22–24]. On the other hand, mouth opening is unavoidable with MAD in situ, which may decrease the beneficial effects on the volume and the cross-sectional area (CSA) of the upper airway [25–27]. A systematic review study [28] has suggested that a small minimal cross-sectional area of the upper airway (CSAmin) is the most relevant anatomical characteristic of the upper airway related to the pathogenesis of OSA. However, when comparing different MADs, few studies have provided insights into their effects on the upper airway dimensions using three-dimensional images. The investigations on upper airway dimensions may allow better understanding of the working mechanism of MAD and help to explain the different treatment efficacy between MADs.
We hypothesized that MAD-H (allowing limited vertical opening) would lead to better improvement in respiratory parameters and upper airway dimensions compared to MAD-S (allowing free vertical opening), especially in supine position. This hypothesis was based primarily on the notion that gravity may promote more mandibular opening in the MAD-S group in the supine position, which subsequently results in less improvement of the upper airway CSAmin [25, 26] and AHI-supine [29, 30] compared with the MAD-H group. Therefore, the aim of this randomized controlled trial (RCT) was to compare the effects of MAD-H and MAD-S on respiratory parameters and upper airway dimensions in patients with mild to moderate OSA.
Material and methods
Overview
This study was a multi-center RCT with a parallel design, in which the effects of two types of titratable MADs on respiratory parameters and upper airway dimensions were compared. The allocation sequence was automatically generated through an electronic data capture system (Castor EDC), using random block size of either 4, 6, or 8 with an allocation ratio of 1:1. The Medical Research Ethics Committee of the Academic Medical Center Amsterdam (AMC) approved this study (#: NL44085.018.13). The study was registered at clinicaltrials.gov (ClinicalTrials.gov identifier: NCT02724865). Written informed consent was obtained from all participants.
Participants
Eligible patients, diagnosed with OSA at four sleep centers in the Netherlands (Onze Lieve Vrouwe Gasthuis Ziekenhuis (OLVG), Nederlands Slaap Instituut, Medisch Centrum Jan van Goyen, and AMC), were referred to the department of Oral and Maxillofacial Surgery of AMC to participate in the present study. The inclusion criteria were: 1. ≥ 18 years old; 2. ability to speak, read, and write Dutch; 3. ability to follow-up; 4. ability to use a computer with internet connection for online questionnaires; 5. diagnosis with symptomatic mild or moderate OSA (5 ≤ apnea–hypopnea index (AHI) ˂ 30 events/hour) with at least two OSA symptoms (e.g., snoring, fragmented sleep, witnessed apneas, and/or excessive daytime sleepiness determined by Epworth Sleepiness Scale (ESS) [1]); and 6. expected to maintain current lifestyle (e.g., sports, medicine, diet, etc.). The exclusion criteria were: 1. untreated periodontal problems, dental pain, and/or a lack of retention possibilities for an MAD; 2. medication usage that could influence respiration or sleep; 3. evidence of respiratory/sleep disorders other than OSA ( e.g., central sleep apnea syndrome); 4. systematic disorders based on medical history and examination (e.g., rheumatoid arthritis); 5. severe temporomandibular disorders based on a functional examination of the masticatory system; 6. coexistence of non-respiratory sleep disorders (e.g., insomnia, periodic limb movement disorder, or narcolepsy); 7. known medical history of mental retardation, memory disorders, or psychiatric disorders; 8. reversible morphological upper airway abnormalities (e.g., enlarged tonsils, deviated nasal septum, and/or inferior nasal turbinate hypertrophy); 9. inability to provide informed consent; 10. simultaneous use of other modalities to treat OSA; and/or 11. previous treatment with an MAD.
MADs
The two types of MAD compared in this study were MAD-H (Herbst appliance; 4Dental labs, Amsterdam, the Netherlands) and MAD-S (SomnoDent appliance; SomnoDent Flex, SomnoMed, Sydney, Australia). The MAD-H consisted of two splints connected to each other with adjustable iron-bars for titration. The vertical opening was limited by these bars (Fig. 1A). The MAD-S consisted of two separate splints allowing free vertical opening. It had a screw mechanism on the upper splint which was used for titration (Fig. 1B).
The standardized titration [31] of the MAD was performed at the department of Orofacial Pain and Dysfunction of the Academic Centre for Dentistry Amsterdam (ACTA) for the MAD-S, and the department of Oral and Maxillofacial Surgery of AMC for the MAD-H. All MAD providers were trained for the standardized titration protocol before the start of the study. After adequate assessment of the central relation and maximum protrusion using the George Gauge system with a standard 5-mm vertical dimension (Great Lakes Orthodontics, Tonawanda, NY), MADs were set at 60% of the maximal mandibular protrusion at baseline [31]. The patient returned to the clinic at 4, 8, and 12 weeks after placement of the MAD for titration. At each consecutive visit, the MADs were evaluated and advanced to 75% and 90% if subjective improvement (e.g., perceived reduction of snoring or excessive daytime sleepiness) of OSA was not reached. On the other hand, if side effects were not acceptable for the patient (e.g., tooth pain or signs of temporomandibular disorders), the advancement was adjusted backwards in steps of 15% per visit to 75%, 60%, or 45%, depending on the actual position at the time of the clinical visit. No adjustments were made when the patient reported a sufficient efficacy without side effects.
MAD side-effects and compliance
The self-reported side effects were recorded at each clinical visit, including the following: 1. sensitive teeth in the morning; 2. painful jaw muscles; 3. painful temporomandibular complaints; and 4. changed occlusion in the morning. The compliance information was collected by a telephone survey, which consisted of four questions: 1. number of hours of MAD use per night; 2. number of hours of total sleep per night; 3. number of days of MAD use per week; and 4. the overall level of satisfaction of the MAD usage. Level of satisfaction was based on a visual analogue scale of 0 to 100, where 0 was unsatisfied, and 100 was very satisfied. The compliance data were expressed as percentage of hours of MAD use per total sleep time, and as percentage of days of MAD use per week.
Polysomnography (PSG) recordings
Every patient underwent a baseline PSG recording and a 3-month follow-up PSG recording with MAD in situ at one of the afore-mentioned sleep centers. A digital PSG system (Embla A10, Broomfield, CO, USA) was used and recorded electroencephalogram (EEG) (FP2-C4/C4-O2), electrooculogram (EOG), electrocardiogram (ECG), and submental and anterior tibial electromyogram (EMG). Nasal airflow was measured by a nasal pressure cannula, and blood oxygen saturation was measured by finger pulse oximetry. Straps containing piezoelectric transducers recorded thoracoabdominal motion, and a position sensor (Sleepsense, St Charles, IL, USA) attached to the midline of the abdominal wall was used to differentiate between supine, prone, right lateral, left lateral, and upright positions [32, 33].
Sleep and respiration were analyzed following the criteria of the American Academy of Sleep Medicine (AASM) Task Force [34]. All polysomnographic variables were scored manually by scorers blinded to the MAD type. Sleep outcome variables included total sleep time (TST), time spent stage N1, stage N2, stage N3, and stage rapid-eye-movement (REM), time spent in supine position, and arousal index. An apnea was defined as the cessation of nasal airflow of more than 90% for a period of ≥ 10 s in the presence of respiratory efforts. A hypopnea was scored whenever there was a > 30% reduced oronasal airflow for at least 10 s, accompanied by ≥ 3% oxygen desaturation from pre-event baseline or an arousal. Respiratory outcome variables included AHI, AHI-supine, AHI-non-supine, and oxygen desaturation index (ODI). Responders were defined as ≥ 50% reduction in baseline AHI with a residual AHI < 10 events/h at the time of therapy evaluation; otherwise, patients were regarded as non-responders [35]. In this study, the AHI-supine was considered as the primary outcome variable for the respiration.
Cone beam computed tomography (CBCT)
All patients underwent two CBCT scans (NewTom 5G, QR systems, Italy) at the department of Oral Radiology of ACTA: a baseline scan and a follow-up scan with the MAD in situ (in the same protrusion position as during the follow-up PSG recording). The exposure settings were 110 kV, 4 mA, 0.3 mm voxel size, 3.6-s exposure time (pulsed radiation), and 18–36-s scanning time, depending on the size of the patient [36]. At baseline, patients were instructed to maintain light contact between the molars in natural occlusion, while at the follow-up scan, patients were instructed to relax their masticatory muscles with the MAD in situ. CBCT scans were performed in supine position while patients were awake. The head of the patient was positioned with the Frankfort horizontal plane (a plane joining the anatomical landmarks porion (Po) and orbitale (Or)) being perpendicular to floor. After scanning, further standardization of head position was performed, during which the palatal plane (anterior nasal spine (ANS)-posterior nasal spine (PNS)) was adjusted to be parallel to the horizontal plane in the sagittal view and perpendicular to the horizontal plane in the axial view [36].
Using Amira® software (v4.1, Visage Imaging Inc., Carlsbad, CA, USA), upper airway segmentation was performed by applying the superior and inferior boundaries of the upper airway. The superior boundary of the upper airway was the palatal plane, and the inferior boundary was the horizontal plane across the base of the epiglottis (parallel to the palatal plane) (Fig. 2A). After upper airway segmentation, the upper airway volume (V) and the cross-sectional area (CSA) of each slice were calculated automatically in the software. Based on the results of CSA, the minimum CSA(CSAmin) could be identified and located (Fig. 2B). On the specific slice where the CSAmin was located, the anterior–posterior dimension (A-P) and lateral dimension (La) of CSAmin were measured by the observer, using the linear measuring tool integrated in the software (Fig. 2C). The length of upper airway (L) was calculated by multiplying the slice numbers of the upper airway with 0.3 mm (the thickness of every slice) [36]. As a small CSAmin is the most relevant anatomical characteristic of the upper airway related to the pathogenesis of OSA [28], the CSAmin was selected as the primary outcome variable of upper airway dimensions.
All of the upper airway variables were measured by an experienced examiner. To test the intra-rater reliability of the assessment of the upper airway, 10 CBCT scans were randomly selected and re-measured after a 1-month interval of the original measurements.
Based on baseline and follow-up CBCT images, the vertical opening with the MAD in situ was determined in two steps in 3Diagnosys ® software (v5.3.1, 3diemme, Cantu, Italy). Firstly, the overbite was measured based on baseline CBCT images (Fig. 3A). Secondly, the vertical distance between the tip of upper and lower incisor with the MAD in situ was measured based on follow-up CBCT images (Fig. 3B). All vertical distances were measured as the perpendicular distance to palatal plane. By adding these two values, the mandibular vertical opening with MAD in situ was determined.
Sample size
According to the guidelines of Cohen, an effect size d, which is defined as a standardized difference in the means of an outcome variable between two groups, of 0.20 is small, one of 0.5 is medium, and one of 0.8 is large [37]. In our study, we assumed a large effect size (i.e., a large difference between both groups) based on our clinical experience. To detect the effect size d of 0.8 between two MAD groups with a power of 80% and a significance level of 5%, a sample size of about 25 patients per group is needed. Therefore, we planned to randomize 25 patients per group.
Statistical analysis
Normality of continuous data was tested by Shapiro–Wilk test. Independent t test (for normally distributed variables), Mann–Whitney U test (for non-normally distributed variables), and Chi-squared test or Fisher’s exact test (for nominal variables) were used to compare the baseline characteristics, MAD titration outcomes, MAD response, MAD side-effect reports, and MAD compliance between the MAD-H group and the MAD-S group.
The intra-rater reliability for the upper airway variables was assessed using a two-way mixed, absolute agreement, single measures intraclass correlation coefficient (ICC). Reliability was defined as poor (ICC < 0.5), moderate (ICC = 0.5–0.75), good (ICC = 0.75–0.9), or excellent (ICC > 0.9) [38].
Two-way analyses of variance (ANOVA) were used to compare the mean differences of all outcome variables separately for within-subjects factor (i.e., baseline without MAD versus follow-up with MAD in situ), for between-subjects factor (i.e., the mean of the outcome variables of MAD-H versus MAD-S), and to assess the interaction effect between the two factors (i.e., treatment effect between baseline and follow-up of MAD-H versus MAD-S) on the outcome variables. Any significantly different baseline characteristics between both MAD groups were controlled as covariates in the assessment of the interaction effect [39]. Bonferroni-Holm method was used for the secondary outcome variables to correct for the increased risk of Type I error due to multiple statistical comparisons [40].
A per-protocol (PP) analysis included patients who completed the entire treatment and was performed for all primary and secondary outcomes variables. An intention-to-treat (ITT) analysis included all patients who underwent randomization and was performed only for the primary outcome variables. In ITT analysis, the missing data were imputed by a multiple imputation procedure (five imputed datasets) in SPSS software (SPSS version 20, Chicago, IL, USA). A post hoc power analysis was conducted for the primary outcome variables using the software G*power (version 3.1.9, Franz Faul, Universität Kiel, Germany).
Results
Flowchart of participants
The flow-chart of 49 patients with OSA who were recruited for this study is shown in Fig. 4. In total, 49 patients underwent MAD randomization and were included in the ITT analysis: 26 patients in MAD-H group and 23 patients in MAD-S group. Of those patients, 10 patients in MAD-H group and 8 patients in MAD-S group were considered as dropouts for the reasons, such as lost contact, quit study, refused the second PSG recordings, and had incomplete dataset. Finally, 16 patients in MAD-H group and 15 patients in MAD-S group completed the entire study and were included in the PP analysis.
Participant characteristics
The baseline characteristics of the MAD-H group, the MAD-S group, and dropouts are shown in Table 1. There were no significant differences between dropouts and the participants who completed the study in age, gender distribution, body mass index (BMI), neck circumference, and Epworth Sleepiness Scale (ESS) (P = 0.30–0.98). The MAD-H group and the MAD-S group did not significantly differ regarding gender distribution, BMI, neck circumference, and ESS (P = 0.08–1.00). However, there was a tendency that the MAD-S group (53.3 ± 15.1 years) was significantly older than the MAD-H group (44.0 ± 11.4 years) (t = − 1.94, P = 0.06). Therefore, age was controlled as covariate in the later analysis of comparing outcome variables between both MAD groups.
Table 1.
MAD-H (n = 16) | MAD-S (n = 15) | Dropouts (n = 18) | t/FET | P | |
---|---|---|---|---|---|
Age (years) | 44.0 ± 11.4 | 53.3 ± 15.1 | 47.8 ± 12.6 | -1.94 (t) | 0.06 |
Gender (men vs women) | 11 vs 5 | 10 vs 5 | 11 vs 7 | FET | 1.00 |
BMI (kg/m2) | 27.8 ± 4.6 | 26.5 ± 3.3 | 25.3|27.2|32.5 | 0.89 (t) | 0.38 |
Neck circumference (cm) | 40.2 ± 3.8 | 37.7 ± 3.8 | 39.3 ± 3.3 | 1.83 (t) | 0.08 |
Epworth Sleepiness Scale a | 8.6 ± 3.7 | 9.1 ± 3.9 | 8.9 ± 5.1 | -0.35 (t) | 0.73 |
Normally distributed data are shown as means ± standard deviations (SD); non-normally distributed data are presented as 25th percentile| median|75th percentile; t independent t test; FET Fisher’s exact test; a: two patients of MAD-H group, one patient of MAD-S group, and 5 patients of dropouts with incomplete ESS score were excluded from the analysis
MAD-H Herbst appliance; MAD-S SomnoDent appliance; BMI body mass index
MAD use
The titration results, treatment response, side-effects, and compliance of the MAD-H and MAD-S groups are shown in Table 2. No significant difference was found in the mandibular protrusion and mandibular vertical opening after titration between both groups (P = 0.54–0.90). Besides, no significant difference was found in the treatment response between both groups (χ2 = 0.03, P = 0.85). There was no significant difference between both MAD groups in the number of each side-effect (P = 0.33–1.00). Seventeen patients responded to the compliance telephone survey, and a few patients counted the wearing hours of MAD prior to their sleep, resulting in wearing hours of more than 100% of their sleep. There was no significant difference between both groups in MAD compliance and satisfaction (P = 0.08–0.40).
Table 2.
MAD-H (n = 16) | MAD-S (n = 15) | t/Z/χ2/FET | P | |
---|---|---|---|---|
Titration results | ||||
Protrusion percentage (%) | 75.0|75.0|75.0 | 60.0|75.0|90.0 | -0.62 (Z) | 0.54 |
Protrusion amount (mm) | 8.4 ± 2.1 | 8.7 ± 2.3 | -0.43 (t) | 0.67 |
Vertical opening (mm) | 11.3 ± 1.5 | 11.4 ± 1.6 | -0.13 (t) | 0.90 |
Treatment response | ||||
Responders vs non-responders | 8 vs 8 | 7 vs 8 | 0.03 (χ2) | 0.85 |
Side-effects | ||||
Sensitive teeth in the morning | 4 | 1 | FET | 0.33 |
Painful jaw muscles | 3 | 1 | FET | 0.60 |
Painful temporomandibular complaints | 3 | 3 | FET | 1.00 |
Changed occlusion in the morning | 3 | 2 | FET | 1.00 |
MAD compliance | (n = 10)a | (n = 7)a | ||
Wearing hour (%) | 100.0|107.1|127.1 | 100.0|100.0|100.0 | -1.75 (Z) | 0.08 |
Wearing day (%) | 100.0|100.0|100.0 | 100.0|100.0|100.0 | -0.84 (Z) | 0.40 |
Level of satisfaction (%) | 68.3 ± 24.0 | 77.9 ± 15.8 | -0.92 (t) | 0.37 |
Normally distributed data are shown as means ± standard deviations (SD); non-normally distributed data are presented as 25th percentile| median|75th percentile; a:10 patients in MAD-H group and 7 patients in MAD-S group responded to compliance telephone survey; t independent t test; Z Mann–Whitney U test; χ2: Chi-squared test; FET Fisher’s exact test
MAD-H Herbst appliance; MAD-S SomnoDent appliance; Wearing hour percentage of hours of MAD use per total sleep time, and a few patients counted the wearing hours of MAD prior to their sleep, resulting in wearing hours of more than 100% of their sleep; Wearing day percentage of days of MAD use per week; Level of satisfaction based on a visual analogue scale of 0 to 100, where 0 was unsatisfied, and 100 was very satisfied
Sleep and respiration
The sleep and respiratory variables without and with MAD in situ of the MAD-H group and the MAD-S group are presented in Table 3. For the primary outcome variable AHI-supine, both PP and ITT analyses indicated that there was no significant difference between MAD-H and MAD-S in improving the AHI-supine (P = 0.14 and P = 0.47, respectively). The individual effects of both MADs on AHI-supine are illustrated in Fig. 5. According to the post hoc power analysis, the effect size f of the AHI-supine was 0.29 in the PP analysis and was 0.10 in the ITT analysis (partial η2 = 0.08 and 0.01, respectively), which is qualified as small to medium. Besides, both PP analysis and ITT analysis indicated that there was no significant change in AHI-supine with MAD in situ in the total group (P = 0.06 and P = 0.12, respectively). For the secondary variables, there was no significant difference between the effects of MAD-H and MAD-S (all P > 0.05). However, the AHI, AHI-non-supine and ODI reduced significantly with MAD in situ in the total group (P < 0.01, P = 0.01, and P = 0.02, respectively).
Table 3.
MAD-H group (n = 16) | MAD-S group (n = 15) | Baseline MAD-H vs MAD-S | Within-subjects effect (Baseline vs follow-up in the total group) | Interaction effect a (MAD-H vs MAD-S in treatment effect) | |||||
---|---|---|---|---|---|---|---|---|---|
Baseline | follow-up | Baseline | follow-up | P | F | P | F | P | |
Primary outcome | |||||||||
AHI-supine (/h) | 25.5 ± 11.6 | 21.8 ± 24.6 | 35.0 ± 17.9 | 23.3 ± 18.5 | 0.09 | 3.85 | 0.06 | 2.26 | 0.14 |
Secondary outcomes | |||||||||
Sleep variables | |||||||||
Total sleep time (min) | 420.5 ± 56.6 | 415.6 ± 67.1 | 378.0 ± 63.7 | 388.5 ± 75.9 | 0.06 | 0.04 | 0.85 | 0.16 | 0.69 |
Stage N1 (%)b | 9.3 ± 5.4 | 9.1 ± 6.6 | 8.2 ± 4.8 | 4.8 ± 3.3 | 0.56 | 2.37 | 0.14 | 2.92 | 0.10 |
Stage N2 (%)b | 50.9 ± 7.8 | 48.9 ± 11.1 | 53.3 ± 9.6 | 46.9 ± 12.3 | 0.97 | 2.67 | 0.11 | 0.55 | 0.47 |
Stage N3 (%)b | 18.6 ± 6.7 | 19.6 ± 7.3 | 17.9 ± 6.8 | 23.9 ± 9.8 | 0.79 | 4.14 | 0.05 | 2.10 | 0.16 |
Stage REM (%)b | 21.1 ± 5.7 | 22.5 ± 6.6 | 20.6 ± 4.7 | 24.4 ± 6.8 | 0.78 | 2.33 | 0.14 | 0.65 | 0.43 |
Supine sleep time (%) | 53.6 ± 33.2 | 48.9 ± 26.6 | 33.5 ± 29.8 | 30.0 ± 26.1 | 0.06 | 0.60 | 0.45 | 0.88 | 0.36 |
Arousal index (/h)b | 7.6 ± 6.9 | 14.2 ± 14.0 | 12.5 ± 15.7 | 11.1 ± 10.6 | 0.50 | 0.62 | 0.44 | 1.30 | 0.27 |
Respiratory variables | |||||||||
AHI (/h) | 16.2 ± 6.1 | 11.4 ± 7.7 | 17.1 ± 7.5 | 11.1 ± 8.7 | 0.69 | 12.55 | < 0.01* | 1.82 | 0.19 |
AHI-non-supine (/h) | 7.1 ± 7.1 | 5.5 ± 4.4 | 12.4 ± 8.9 | 4.8 ± 3.3 | 0.07 | 8.49 | 0.01* | 0.85 | 0.37 |
ODI (/h) | 16.5 ± 9.7 | 11.0 ± 6.9 | 15.6 ± 10.7 | 12.3 ± 7.4 | 0.91 | 6.13 | 0.02* | 0.00 | 0.95 |
Data are shown as means ± standard deviations (SD); a: all outcome variables were controlled for age, and the variables of total sleep time, supine sleep time, and AHI-non-supine were controlled additionally for their baseline values; b:two patients in MAD-S group with incomplete sleep dataset were excluded from the analysis of sleep variables; *: significant difference after Bonferroni-Holm correction
AHI apnea–hypopnea index; AHI-supine AHI in supine position; AHI-non-supine AHI in positions other than supine position; ODI: oxygen desaturation index; REM: rapid eye movement
Reliability of the upper airway assessment
The intra-rater reliability for the upper airway assessment was excellent, with ICC = 0.96 for the CSAmin and ICC = 0.90 to 0.94 for the secondary outcome variables.
Upper airway dimensions
The upper airway dimensions without and with MAD in situ of the MAD-H group and the MAD-S group are presented in Table 4. For the primary outcome variable CSAmin, both PP analysis and ITT analysis showed similar results: there was no significant difference between MAD-H and MAD-S in increasing the CSAmin (P = 0.59 and P = 0.57, respectively). According to the post hoc power analysis, the effect size f of the CSAmin was 0.10 in the PP analysis and was 0.08 in the ITT analysis (partial η2 = 0.01 and 0.007, respectively), which is qualified as small. Besides, both PP analysis and ITT analysis indicated that the CSAmin increased significantly with MAD in situ in the total group (P = 0.01 and P < 0.01, respectively). For the secondary outcome variables, there was no significant difference between the effects of MAD-H and MAD-S (all P > 0.05). However, the lateral dimension of the CSAmin (La) increased significantly with MAD in situ in the total group (P < 0.01).
Table 4.
MAD-H group (n = 16) | MAD-S group (n = 15) | Baseline MAD-H vs MAD-S | Within-subjects effect (Baseline vs follow-up in the total group) | Interaction effect a (MAD-H vs MAD-S in treatment effect) | |||||
---|---|---|---|---|---|---|---|---|---|
Baseline | follow-up | Baseline | follow-up | P | F | P | F | P | |
Primary outcome | |||||||||
CSAmin (mm2) | 54.8 ± 36.9 | 75.6 ± 50.4 | 58.5 ± 29.3 | 87.8 ± 59.1 | 0.50 | 9.13 | 0.01* | 0.30 | 0.59 |
Secondary outcomes | |||||||||
A-P (mm) | 4.6 ± 2.1 | 5.0 ± 2.4 | 4.4 ± 1.7 | 5.3 ± 1.9 | 0.77 | 3.37 | 0.08 | 0.35 | 0.56 |
La (mm) | 11.4 ± 4.7 | 15.4 ± 5.1 | 13.1 ± 4.6 | 16.2 ± 5.8 | 0.32 | 16.74 | < 0.01* | 0.12 | 0.74 |
L (mm) | 67.0 ± 9.6 | 65.6 ± 9.9 | 65.9 ± 8.0 | 64.4 ± 8.8 | 0.59 | 4.30 | 0.05 | 0.00 | 0.99 |
V (cm3) | 10.9 ± 3.4 | 12.2 ± 5.5 | 11.4 ± 5.4 | 12.8 ± 5.7 | 0.75 | 3.42 | 0.08 | 0.10 | 0.75 |
Data are shown as means ± standard deviations (SD); a: all outcome variables were controlled for age; *: significant difference after Bonferroni-Holm correction
CSAmin minimum cross-sectional area of the upper airway; A-P anterior–posterior dimension of the CSAmin; La lateral dimension of the CSAmin; L length of the upper airway; V upper airway volume
Discussion
The aim of this RCT was to compare the effects of MAD-H (allowing limited vertical opening) and MAD-S (allowing free vertical opening) on respiratory parameters and upper airway dimensions in patients with mild to moderate OSA. The results showed that despite differences in the freedom of mandibular vertical opening, there was no significant difference between MAD-H and MAD-S in improving the respiratory parameters and upper airway dimensions.
MAD-H vs MAD-S
In the present study, there was no significant difference between the MAD-H and MAD-S in affecting the AHI-supine, as well as AHI and AHI-non-supine. According to the post hoc power analysis, the effect size f of the AHI-supine was 0.29 in the PP analysis and was 0.10 in the ITT analysis, which is qualified as small to medium (an effect size f = 0.10 is small, one = 0.25 is medium, and one = 0.40 is large [37]). With this effect size, an enlargement of our total sample size to around 100–800 patients, is needed to find a statistically significant difference between both groups (with power 0.8; 5% significance level), which makes the clinical relevance of such a finding questionable. Regarding the clinical relevance of two interventions, the concept of the number needed to treat (NNT) is often used. In this study, 50% in MAD-H group and 47% in MAD-S group were treated successfully. The NNT is 33, implying that about 33 patients with mild to moderate OSA need to be treated with MAD-H to get one more successfully treated patient as compared to treatment with MAD-S, which is not clinically relevant [41]. Although the freedom of vertical opening is different between MAD-H and MAD-S, it seems that the respiratory outcomes were not affected by this design feature. The same protrusion and vertical opening of the mandible in awake state may explain the similarity in treatment efficiency between MAD-H and MAD-S. However, as we did not measure the vertical opening in sleep, future studies are warranted to verify this hypothesis. Similar to our study, a retrospective study of Verburg et al. [21] also compared MAD-H and MAD-S and found no significant difference in improving the AHI. Thus, an MAD allowing limited vertical opening and an MAD allowing free vertical opening may result in similar respiratory outcomes.
No significant difference was found between both MADs in changing the upper airway dimensions. With the same protrusion position and vertical opening, the same effect of MAD-H and MAD-S was applied to the tongue and mandible in the awake state, and thus no difference of upper airway changes could be found between both groups. Our results are similar to the outcomes of two previous studies based on cephalometric images [16, 42], in which no significantly different effects on the upper airway dimensions were found when two MADs with similar protrusion and vertical opening were compared. However, limited to 2-D radiology images, those studies only compared the upper airway in anterior–posterior direction [16, 42]. By using CBCT images, our study has the advantage of providing additional information on lateral direction dimension, CSAmin, and volume of the upper airway. The current CBCT scanning and measurement protocols were standardized, and have been proven to have an excellent reliability in the upper airway assessment in the present study (ICC = 0.90 to 0.96) as well as in our previous study [36]. However, a study of Ryan et al. [43] has suggested that different CBCT scanning timings of the same patient with same scanning and patient positioning protocols can result in different upper airway volumetric data, suggesting the intra-individual variability in CBCT assessments. However, this intra-individual variability was randomly present in both groups, and therefore may not explain our non-significant findings in the comparison between groups.
The self-reported side effects were not significantly different between both MAD groups. However, sensitive teeth and painful jaw muscles were 3–4 times more frequent in the MAD-H group compared to MAD-S group, which might be due to the different design feature. However, it seems that the side effects of MAD in most cases were of a minor nature and probably improve in the longer term [44, 45]. A study with a long-term is needed to confirm the clinical relevance of our short-term finding.
Overall MAD effects
For respiratory parameters, there was significant decrease in the AHI and AHI-non-supine but no significant decrease in the AHI-supine with MAD in situ in the total group, which indicated that the MADs used in this study were less effective in reducing the AHI in supine position. As MADs used in this study allow more or less freedom of vertical opening, gravity would favor mouth opening and weaken the beneficial effects of mandibular protrusion in supine position for both MADs. A study of Milano et al. [46] compared a group using an MAD allowing free vertical opening and a group using the same MAD with elastics to restrict the vertical opening, and suggested that the AHI-supine was improved significantly in both groups but that the improvement of AHI-supine was significantly greater in the group MAD with elastics. However, the patients in their study were positional OSA, which represents a different study sample as compared to our study and may explain the different effects of MAD allowing free vertical opening on AHI-supine. Inserting an MAD did not significantly change the time spent in supine position and REM stage, which is consistent with other studies [18, 47]. Given that sleeping in supine position and REM stage may worsen the respiratory events [48], the treatment effects of MAD in the present study were not related to the time spent in supine position and/or REM stage at therapy evaluation.
The treatment response rate in the present study (48%) is in line with other studies reporting MAD response rate between 43 and 77% [49]. Using the same titration protocol and response rate criteria as the present study, the study of de Ruiter et al. [31] reported a 44.4% treatment response rate, which is in line with our results as well.
The CSAmin increased significantly with MAD in situ in the total group. Further, the increase of CSAmin was mainly due to the enlargement in the lateral dimension, which is similar to previous studies [50–52]. Although an MAD works primarily by protruding the mandible and tongue, the enlargement of the upper airway was predominantly in the lateral dimensions. The precise mechanism for this observation is not completely understood, but it has been suggested that this is related to the soft tissue connections between the tongue, soft palate, and lateral pharyngeal walls through the palatopharyngeal and palatoglossal arches [51]. Mandibular advancement possibly stretches the lateral pharyngeal walls through these soft tissue connections, and results in lateral enlargement of the CSAmin.
Limitations
The present study has several limitations. Firstly, the information of the patients who were initially screened for this study was missing. With these data missing, we cannot be certain that our sample represents the total population of OSA patients that are being seen in the clinics in Amsterdam. Although the dropout rate (36%) seemed high in our study, the dropout rate was comparable to our previous studies [33, 53]. Importantly, there were no significant differences between dropouts and the participants in baseline characteristics. Further, the intention-to-treat analyses showed similar results as the per-protocol analyses. Secondly, the upper airway images were taken in awake state, and therefore the upper airway morphology may not be identical to the sleep state. Besides, CBCT assessment while awake do not reflect the actual effects of MAD on the upper airway in sleep as both types of MADs allow more or less mouth opening in sleep. However, it is currently challenging to scan the upper airway during sleep. As to resemble the sleep state as much as possible, patients were instructed to relax their masticatory muscles with the MAD in situ when performing the follow-up CBCT scans. Importantly, since the CBCT scans at baseline and at therapy evaluation were both performed in awake supine condition, the effects of MADs on upper airway dimensions were unbiased. Thirdly, only a part of patients responded to the compliance telephone survey (response rate: 55%). It could be possible that patients who responded to the telephone survey in the present study were more willing to use MAD, resulting an overestimation of MAD compliance. Besides, the intensive follow-up titration protocol may also have contributed to a high compliance. Finally, subjective treatment outcomes, including snoring, daytime sleepiness (Epworth Sleepiness Scale), and quality of life, are also important therapeutic targets in patients with OSA, and therefore future studies should involve these aspects. Besides, as the present study investigated the treatment effect of both MADs in a short-term, a long-term follow-up study is needed to confirm our short-term findings.
Strengths
The standardized titration protocol used for both types of MAD is one of the major strengths of the present study. To the best of our knowledge, when comparing different MADs, few previous studies have reported a standardized titration protocol. In the present study, a detailed standardized titration protocol was applied, and no significant difference was found in finial protrusion position of both MAD groups, which enables an unbiased comparison. The standardized stepwise titration protocol for MAD, developed by one of the co-authors of this paper (GA), has been proved to have good efficacy, good tolerance, and good adherence [31], and is therefore recommended for use in clinical studies and practices.
Inserting an MAD intra-orally induces a certain amount of mandibular vertical opening. A study of Mayoral et al. [54] has indicated that as the vertical dimension increases, the mandible rotates posteriorly and places itself in a more retrusive location (0.3 mm for every 1 mm of vertical increase). Therefore, it is recommended to limit the vertical opening of MADs [54, 55]. In this study, the vertical opening for MAD-H and for MAD-S in awake state were similar. To guarantee an accurate measurement of the vertical opening, we added the overbite and inter-incisor distance with MAD in situ together. As most previous studies only used inter-incisor distance to estimate the vertical opening, the accurate measurement of the vertical opening is another strength of the present study.
Conclusions
Within the limitations of this study, we conclude that there is no significant difference between the effects of an MAD allowing limited vertical opening and an MAD allowing free vertical opening on respiratory parameters and upper airway dimensions in patients with mild to moderate OSA.
Acknowledgements
The authors gratefully acknowledge Dr. Naichuan Su, Department of Oral Public Health, Academic Centre for Dentistry Amsterdam (ACTA), for his assistance with the statistical analyses of this study; and the staff of the OLVG, Netherlands Slaap Instituut, Medisch Centrum Jan van Goyen and AMC for their assistance with this work.
Author contribution
Xiaoxin Shi: Conceptualization, Methodology, Validation, Formal analysis, Investigation, Data curation; Writing—Original Draft, Visualization; Frank Lobbezoo: Project administration, Resources, Writing—Review & Editing, Supervision; Hui Chen: Conceptualization, Methodology, Writing—Review & Editing, Supervision; Boudewijn R.A.M. Rosenmöller: Methodology, Investigation, Writing—Review & Editing; Erwin Berkhout: Software, Methodology, Resources, Writing—Review & Editing, Supervision; Jan de Lange: Conceptualization, Writing—Review & Editing, Funding acquisition; Ghizlane Aarab: Conceptualization, Methodology, Resources, Writing—Review & Editing, Supervision. All authors reviewed and approved the final manuscript.
Funding
Xiaoxin Shi has received a scholarship from the China Scholarship Council.
Data availability
The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.
Declarations
Ethics approval
This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Medical Research Ethics Committee of the Academic Medical Center Amsterdam (AMC) (#: NL44085.018.13).
Informed consent
Informed consent was obtained from all individual participants included in the study.
Conflict of interest
Xiaoxin Shi declares that she has no conflicts of interest. Frank Lobbezoo is a member of the Academic Advisory Boards for Grind Care and Oral Function of Sunstar Suisse S.A. and receives research grants from Sunstar Suisse S.A., SomnoMed, Vivisol, Health Holland, and Airway Management. Hui Chen declares that she has no conflicts of interest. Boudewijn R.A.M. Rosenmöller declares that he has no conflicts of interest. Erwin Berkhout declares that he has no conflicts of interest. Jan de Lange declares that he has no conflicts of interest. Ghizlane Aarab is a member of the Academic Advisory Board for Oral Function of Sunstar Suisse S.A. and receives research grants from Sunstar Suisse S.A., SomnoMed, Vivisol, and Health Holland.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.American Academy of Sleep Medicine Task Force Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. Sleep. 1999;22(5):667–689. doi: 10.1093/sleep/22.5.667. [DOI] [PubMed] [Google Scholar]
- 2.Young T, Palta M, Dempsey J, Skatrud J, Weber S, Badr S. The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med. 1993;328(17):1230–1235. doi: 10.1056/NEJM199304293281704. [DOI] [PubMed] [Google Scholar]
- 3.Gottlieb DJ, Whitney CW, Bonekat WH, Iber C, James GD, Lebowitz M, Nieto FJ, Rosenberg CE. Relation of sleepiness to respiratory disturbance index: the Sleep Heart Health Study. Am J Respir Crit Care Med. 1999;159(2):502–507. doi: 10.1164/ajrccm.159.2.9804051. [DOI] [PubMed] [Google Scholar]
- 4.Peppard PE, Young T, Palta M, Skatrud J. Prospective study of the association between sleep-disordered breathing and hypertension. N Engl J Med. 2000;342(19):1378–1384. doi: 10.1056/NEJM200005113421901. [DOI] [PubMed] [Google Scholar]
- 5.Yaggi HK, Concato J, Kernan WN, Lichtman JH, Brass LM, Mohsenin V. Obstructive sleep apnea as a risk factor for stroke and death. N Engl J Med. 2005;353(19):2034–2041. doi: 10.1056/NEJMoa043104. [DOI] [PubMed] [Google Scholar]
- 6.Redline S, Yenokyan G, Gottlieb DJ, Shahar E, O'Connor GT, Resnick HE, Diener-West M, Sanders MH, Wolf PA, Geraghty EM. Obstructive sleep apnea–hypopnea and incident stroke: the sleep heart health study. Am J Respir Crit Care Med. 2010;182(2):269–277. doi: 10.1164/rccm.200911-1746OC. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Kendzerska T, Gershon AS, Hawker G, Tomlinson G, Leung RS. Obstructive sleep apnea and incident diabetes. A historical cohort study. Am J Respir Crit Care Med. 2014;190(2):218–225. doi: 10.1164/rccm.201312-2209OC. [DOI] [PubMed] [Google Scholar]
- 8.Giles TL, Lasserson TJ, Smith BH, White J, Wright J, Cates CJ. Continuous positive airways pressure for obstructive sleep apnoea in adults. Cochrane Database Syst Rev. 2006;3:CD001106. doi: 10.1002/14651858.CD001106.pub3. [DOI] [PubMed] [Google Scholar]
- 9.Richard W, Venker J, den Herder C, Kox D, van den Berg B, Laman M, van Tinteren H, de Vries N. Acceptance and long-term compliance of nCPAP in obstructive sleep apnea. Eur Arch oto-rhino-laryngology: Off J Eur Fed Oto-Rhino-Laryngological Soc (EUFOS): Affiliated Ger Soc Oto-Rhino-Laryngology - Head Neck Surg. 2007;264(9):1081–1086. doi: 10.1007/s00405-007-0311-3. [DOI] [PubMed] [Google Scholar]
- 10.Sharples LD, Clutterbuck-James AL, Glover MJ, Bennett MS, Chadwick R, Pittman MA, Quinnell TG. Meta-analysis of randomised controlled trials of oral mandibular advancement devices and continuous positive airway pressure for obstructive sleep apnoea-hypopnoea. Sleep Med Rev. 2016;27:108–124. doi: 10.1016/j.smrv.2015.05.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Epstein LJ, Kristo D, Strollo PJ, Jr, Friedman N, Malhotra A, Patil SP, Ramar K, Rogers R, Schwab RJ, Weaver EM, Weinstein MD. Clinical guideline for the evaluation, management and long-term care of obstructive sleep apnea in adults. J Clin Sleep Med: JCSM: Off Publ Am Acad Sleep Med. 2009;5(3):263–276. [PMC free article] [PubMed] [Google Scholar]
- 12.Randerath W, Verbraecken J, de Raaff CAL, Hedner J, Herkenrath S, Hohenhorst W, Jakob T, Marrone O, Marklund M, McNicholas WT, Morgan RL, Pepin JL, Schiza S, Skoetz N, Smyth D, Steier J, Tonia T, Trzepizur W, van Mechelen PH, Wijkstra P (2021) European Respiratory Society guideline on non-CPAP therapies for obstructive sleep apnoea. Eur Respir Rev: Off J Eur Respir Soc 30(162). 10.1183/16000617.0200-2021 [DOI] [PMC free article] [PubMed]
- 13.Ahrens A, McGrath C, Hagg U. A systematic review of the efficacy of oral appliance design in the management of obstructive sleep apnoea. Eur J Orthod. 2011;33(3):318–324. doi: 10.1093/ejo/cjq079. [DOI] [PubMed] [Google Scholar]
- 14.Alsufyani NA, Al-Saleh MA, Major PW. CBCT assessment of upper airway changes and treatment outcomes of obstructive sleep apnoea: a systematic review. Sleep Breathing. 2013;17(3):911–923. doi: 10.1007/s11325-012-0799-7. [DOI] [PubMed] [Google Scholar]
- 15.Ghazal A, Sorichter S, Jonas I, Rose EC. A randomized prospective long-term study of two oral appliances for sleep apnoea treatment. J Sleep Res. 2009;18(3):321–328. doi: 10.1111/j.1365-2869.2009.00738.x. [DOI] [PubMed] [Google Scholar]
- 16.Zhou J, Liu YH. A randomised titrated crossover study comparing two oral appliances in the treatment for mild to moderate obstructive sleep apnoea/hypopnoea syndrome. J Oral Rehabil. 2012;39(12):914–922. doi: 10.1111/joor.12006. [DOI] [PubMed] [Google Scholar]
- 17.Umemoto G, Toyoshima H, Yamaguchi Y, Aoyagi N, Yoshimura C, Funakoshi K. Therapeutic Efficacy of Twin-Block and Fixed Oral Appliances in Patients with Obstructive Sleep Apnea Syndrome. J Prosthodont. 2019;28(2):e830–e836. doi: 10.1111/jopr.12619. [DOI] [PubMed] [Google Scholar]
- 18.Bloch KE, Iseli A, Zhang JN, Xie X, Kaplan V, Stoeckli PW, Russi EW. A randomized, controlled crossover trial of two oral appliances for sleep apnea treatment. Am J Respir Crit Care Med. 2000;162(1):246–251. doi: 10.1164/ajrccm.162.1.9908112. [DOI] [PubMed] [Google Scholar]
- 19.Lee WH, Wee JH, Lee CH, Kim M-S, Rhee C-S, Yun P-Y, Yoon I-Y, Kim J-W. Comparison between mono-bloc and bi-bloc mandibular advancement devices for obstructive sleep apnea. Eur Arch Otorhinolaryngol. 2013;270(11):2909–2913. doi: 10.1007/s00405-013-2417-0. [DOI] [PubMed] [Google Scholar]
- 20.Ramar K, Dort LC, Katz SG, Lettieri CJ, Harrod CG, Thomas SM, Chervin RD. Clinical practice guideline for the treatment of obstructive sleep apnea and snoring with oral appliance therapy: an update for 2015: an American Academy of Sleep Medicine and American Academy of Dental Sleep Medicine clinical practice guideline. J Clin Sleep Med. 2015;11(7):773–827. doi: 10.5664/jcsm.4858. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Verburg FE, Bollen KHA, Donker H-J, Kramer GJC. The effectiveness of two types of MADS for OSA therapy. Clin Oral Invest. 2018;22(5):1995–2003. doi: 10.1007/s00784-017-2290-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Yoshida K. Effect of a prosthetic appliance for treatment of sleep apnea syndrome on masticatory and tongue muscle activity. J Prosthet Dent. 1998;79(5):537–544. doi: 10.1016/S0022-3913(98)70175-1. [DOI] [PubMed] [Google Scholar]
- 23.Ng AT, Gotsopoulos H, Qian J, Cistulli PA. Effect of oral appliance therapy on upper airway collapsibility in obstructive sleep apnea. Am J Respir Crit Care Med. 2003;168(2):238–241. doi: 10.1164/rccm.200211-1275OC. [DOI] [PubMed] [Google Scholar]
- 24.Tsuiki S, Lowe AA, Almeida FR, Kawahata N, Fleetham JA. Effects of mandibular advancement on airway curvature and obstructive sleep apnoea severity. Eur Respir J. 2004;23(2):263–268. doi: 10.1183/09031936.04.00094304. [DOI] [PubMed] [Google Scholar]
- 25.Kim EJ, Choi JH, Kim KW, Kim TH, Lee SH, Lee HM, Shin C, Lee KY, Lee SH. The impacts of open-mouth breathing on upper airway space in obstructive sleep apnea: 3-D MDCT analysis. Eur Arch Otorhinolaryngol. 2011;268(4):533–539. doi: 10.1007/s00405-010-1397-6. [DOI] [PubMed] [Google Scholar]
- 26.Vroegop AV, Vanderveken OM, Van de Heyning PH, Braem MJ. Effects of vertical opening on pharyngeal dimensions in patients with obstructive sleep apnoea. Sleep Med. 2012;13(3):314–316. doi: 10.1016/j.sleep.2011.08.005. [DOI] [PubMed] [Google Scholar]
- 27.Barbero M, Flores-Mir C, Blanco JC, Nuño VC, Casellas JB, Girado JLC, Amezaga JA, De Carlos F. Tridimensional upper airway assessment in male patients with OSA using oral advancement devices modifying their vertical dimension. J Clin Sleep Med. 2020;16(10):1721–1729. doi: 10.5664/jcsm.8666. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Chen H, Aarab G, de Ruiter MH, de Lange J, Lobbezoo F, van der Stelt PF. Three-dimensional imaging of the upper airway anatomy in obstructive sleep apnea: a systematic review. Sleep Med. 2016;21:19–27. doi: 10.1016/j.sleep.2016.01.022. [DOI] [PubMed] [Google Scholar]
- 29.Meurice J-C, Marc I, Carrier G, Sériès F. Effects of mouth opening on upper airway collapsibility in normal sleeping subjects. Am J Respir Crit Care Med. 1996;153(1):255–259. doi: 10.1164/ajrccm.153.1.8542125. [DOI] [PubMed] [Google Scholar]
- 30.Isono S, Tanaka A, Tagaito Y, Ishikawa T, Nishino T. Influences of head positions and bite opening on collapsibility of the passive pharynx. J Appl Physiol. 2004;97(1):339–346. doi: 10.1152/japplphysiol.00907.2003. [DOI] [PubMed] [Google Scholar]
- 31.De Ruiter M, Aarab G, De Vries N, Lobbezoo F, de Lange J. A stepwise titration protocol for oral appliance therapy in positional obstructive sleep apnea patients: proof of concept. Sleep Breathing. 2020;24(3):1229–1236. doi: 10.1007/s11325-020-02045-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Benoist LBL, Verhagen M, Torensma B, van Maanen JP, de Vries N. Positional therapy in patients with residual positional obstructive sleep apnea after upper airway surgery. Sleep Breathing. 2017;21(2):279–288. doi: 10.1007/s11325-016-1397-x. [DOI] [PubMed] [Google Scholar]
- 33.de Ruiter MHT, Benoist LBL, de Vries N, de Lange J. Durability of treatment effects of the Sleep Position Trainer versus oral appliance therapy in positional OSA: 12-month follow-up of a randomized controlled trial. Sleep Breathing. 2018;22(2):441–450. doi: 10.1007/s11325-017-1568-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Berry RB, Budhiraja R, Gottlieb DJ, Gozal D, Iber C, Kapur VK, Marcus CL, Mehra R, Parthasarathy S, Quan SF. Rules for scoring respiratory events in sleep: update of the 2007 AASM manual for the scoring of sleep and associated events: deliberations of the sleep apnea definitions task force of the American Academy of Sleep Medicine. J Clin Sleep Med. 2012;8(5):597–619. doi: 10.5664/jcsm.2172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Camañes-Gonzalvo S, Bellot-Arcís C, Marco-Pitarch R, Montiel-Company JM, García-Selva M, Agustín-Panadero R, Paredes-Gallardo V, Puertas-Cuesta FJ. Comparison of the phenotypic characteristics between responders and non-responders to obstructive sleep apnea treatment using mandibular advancement devices in adult patients: Systematic review and meta-analysis. Sleep Med Rev. 2022;64:101644. doi: 10.1016/j.smrv.2022.101644. [DOI] [PubMed] [Google Scholar]
- 36.Chen H, Aarab G, Parsa A, de Lange J, van der Stelt PF, Lobbezoo F. Reliability of three-dimensional measurements of the upper airway on cone beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;122(1):104–110. doi: 10.1016/j.oooo.2016.04.005. [DOI] [PubMed] [Google Scholar]
- 37.Cohen J (1988) Statistical power analysis for the behavioral sciences, 2nd edn. Routledge, New York. 10.4324/9780203771587
- 38.Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15(2):155–163. doi: 10.1016/j.jcm.2016.02.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Kim HY (2018) Statistical notes for clinical researchers: analysis of covariance (ANCOVA). Restor Dent Endod 43(4):e43. 10.5395/rde.2018.43.e43 [DOI] [PMC free article] [PubMed]
- 40.Armstrong RA. When to use the Bonferroni correction. Ophthalmic Physiol Opt. 2014;34(5):502–508. doi: 10.1111/opo.12131. [DOI] [PubMed] [Google Scholar]
- 41.Kalso E, Edwards J, McQuay HJ, Moore RA. Five easy pieces on evidence-based medicine (3) Eur J Pain (London, England) 2001;5(2):227–230. doi: 10.1053/eujp.2001.0242. [DOI] [PubMed] [Google Scholar]
- 42.Geoghegan F, Ahrens A, McGrath C, Hagg U. An evaluation of two different mandibular advancement devices on craniofacial characteristics and upper airway dimensions of Chinese adult obstructive sleep apnea patients. Angle Orthod. 2015;85(6):962–968. doi: 10.2319/040314-245.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Ryan DPO, Bianchi J, Ignácio J, Wolford LM, Gonçalves JR. Cone-beam computed tomography airway measurements: Can we trust them? Am J Orthod Dentofac Orthop. 2019;156(1):53–60. doi: 10.1016/j.ajodo.2018.07.024. [DOI] [PubMed] [Google Scholar]
- 44.Lawton H, Battagel J, Kotecha B. A comparison of the Twin Block and Herbst mandibular advancement splints in the treatment of patients with obstructive sleep apnoea: a prospective study. Eur J Orthod. 2005;27(1):82–90. doi: 10.1093/ejo/cjh067. [DOI] [PubMed] [Google Scholar]
- 45.Pantin CC, Hillman DR, Tennant M. Dental side effects of an oral device to treat snoring and obstructive sleep apnea. Sleep. 1999;22(2):237–240. doi: 10.1093/sleep/22.2.237. [DOI] [PubMed] [Google Scholar]
- 46.Milano F, Mutinelli S, Sutherland K, Milioli G, Scaramuzzino G, Cortesi A, Siciliani G, Lombardo L, Cistulli P. Influence of vertical mouth opening on oral appliance treatment outcome in positional obstructive sleep apnea. J Dent Sleep Med. 2018;5(1):17–23. doi: 10.15331/jdsm.6918. [DOI] [Google Scholar]
- 47.Dieltjens M, Vroegop AV, Verbruggen AE, Wouters K, Willemen M, De Backer WA, Verbraecken JA, Van de Heyning PH, Braem MJ, de Vries N. A promising concept of combination therapy for positional obstructive sleep apnea. Sleep Breathing. 2015;19(2):637–644. doi: 10.1007/s11325-014-1068-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Eiseman NA, Westover MB, Ellenbogen JM, Bianchi MT. The impact of body posture and sleep stages on sleep apnea severity in adults. J Clin Sleep Med. 2012;8(6):655–666. doi: 10.5664/jcsm.2258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Dieltjens M, Vanderveken OM, Van de Heyning PH, Braem MJ. Current opinions and clinical practice in the titration of oral appliances in the treatment of sleep-disordered breathing. Sleep Med Rev. 2012;16(2):177–185. doi: 10.1016/j.smrv.2011.06.002. [DOI] [PubMed] [Google Scholar]
- 50.Chan AS, Sutherland K, Schwab RJ, Zeng B, Petocz P, Lee RW, Darendeliler MA, Cistulli PA. The effect of mandibular advancement on upper airway structure in obstructive sleep apnoea. Thorax. 2010;65(8):726–732. doi: 10.1136/thx.2009.131094. [DOI] [PubMed] [Google Scholar]
- 51.Sutherland K, Deane SA, Chan AS, Schwab RJ, Ng AT, Darendeliler MA, Cistulli PA. Comparative effects of two oral appliances on upper airway structure in obstructive sleep apnea. Sleep. 2011;34(4):469–477. doi: 10.1093/sleep/34.4.469. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Ryan C, Love L, Peat D, Fleetham J, Lowe A. Mandibular advancement oral appliance therapy for obstructive sleep apnoea: effect on awake calibre of the velopharynx. Thorax. 1999;54(11):972–977. doi: 10.1136/thx.54.11.972. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Benoist L, de Ruiter M, de Lange J, de Vries N. A randomized, controlled trial of positional therapy versus oral appliance therapy for position-dependent sleep apnea. Sleep Med. 2017;34:109–117. doi: 10.1016/j.sleep.2017.01.024. [DOI] [PubMed] [Google Scholar]
- 54.Mayoral P, Lagravère M, Míguez-Contreras M, Garcia M. Antero-posterior mandibular position at different vertical levels for mandibular advancing device design. BMC Oral Health. 2019;19(1):1–8. doi: 10.1186/s12903-019-0783-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Nikolopoulou M, Naeije M, Aarab G, Hamburger H, Visscher C, Lobbezoo F. The effect of raising the bite without mandibular protrusion on obstructive sleep apnoea. J Oral Rehabil. 2011;38(9):643–647. doi: 10.1111/j.1365-2842.2011.02221.x. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.