. 2023 Apr 24;13(9):1530. doi: 10.3390/diagnostics13091530

Evolution of the Probe-Based Loop-Mediated Isothermal Amplification (LAMP) Assays in Pathogen Detection

Xiaoling Zhang 1, Yongjuan Zhao 1, Yi Zeng 1, Chiyu Zhang 1,*
Editor: Chan Yean Yean1
PMCID: PMC10177487  PMID: 37174922

Abstract

Loop-mediated isothermal amplification (LAMP), as the rank one alternative to a polymerase chain reaction (PCR), has been widely applied in point-of-care testing (POCT) due to its rapid, simple, and cost-effective characteristics. However, it is difficult to achieve real-time monitoring and multiplex detection with the traditional LAMP method. In addition, these approaches that use turbidimetry, sequence-independent intercalating dyes, or pH-sensitive indicators to indirectly reflect amplification can result in false-positive results if non-specific amplification occurs. To fulfill the needs of specific target detection and one-pot multiplex detection, a variety of probe-based LAMP assays have been developed. This review focuses on the principles of these assays, summarizes their applications in pathogen detection, and discusses their features and advantages over the traditional LAMP methods.

Keywords: LAMP, probe, specificity, pathogen detection, sensitivity

1. Introduction

Pathogen detection is essential for the prevention and control of infectious diseases. The development of rapid, sensitive, specific, and cost-effective assays for the detection of pathogens (e.g., bacteria and viruses) remains a big challenge in the field of diagnosis [1,2,3]. This is largely due to the vast number and variety of pathogens, especially fast-mutating viruses [4].

Molecular diagnostics are commonly used in pathogen detection [5,6]. Currently, molecular diagnostic techniques can be broadly classified into the following categories: polymerase chain reaction (PCR), digital PCR, clustered regularly interspaced short palindromic repeats (CRISPR) test, isothermal acid amplification, biochip technique, and gene sequencing, etc. [5,6,7,8,9,10,11,12]. Since the revolutionary development of PCR in the 1980s, especially the development of quantitative (or real-time) PCR (qPCR) in the 1990s, nucleic acid amplification tests (NAATs) have become indispensable tools throughout the entire field of life science [13]. Currently, qPCR-based diagnostic procedures, with consistently reliable accuracy, have grown to be the gold standard for nucleic acid analysis and are widely used in laboratories and diagnostic facilities [7,14]. However, qPCR is not cost-effective because sophisticated instruments and technicians are required. In addition, qPCR is relatively time-consuming as it is cycled under strictly defined durations and temperature conditions [14,15]. Due to the limitations of qPCR mentioned above, a remarkable development trend of isothermal amplification methods has been observed since 1995. In recent years, isothermal NAATs have shown good performance in the rapid detection of pathogens [16]. This technology can be simply performed in low-cost devices and has a fast processing time compared to qPCR, making it particularly suitable for point-of-care testing (POCT) [17,18,19]. After the pioneering development of the rolling circle amplification (RCA), various alternative approaches were developed to enable isothermal amplification, mainly including loop-mediated isothermal amplification (LAMP), nucleic acid sequence-based amplification (NASBA), recombinase polymerase amplification (RPA), helicase-dependent amplification (HDA), and so forth [20,21,22,23,24,25,26]. LAMP is one of the most frequently used isothermal amplification methods [16].

LAMP was first described by Notomi et al. in 2000 [23]. It utilizes two pairs of primers, including two inner primers (forward: FIP (F1c + F2) and backward: BIP (B1c + B2)) and two outer primers (F3 and B3), to initiate isothermal DNA amplification at a stable temperature of 60–65 °C by Bst DNA polymerase with a strand displacement activity (Figure 1). The inner primers contain two target sequences specific to two different regions in plus and minus strands of the template DNA. LAMP amplification contains two phases, an initial amplification phase, and an exponential amplification phase. In the initial amplification phase, the inner primers hybridize with the target DNA and start complementary strand synthesis (Figure 1A). At the same time, the outer primers also initiate DNA synthesis to displace the newly synthesized DNA strands initiated by the inner primers (e.g., FIP) (Figure 1B). The released single-stranded (ss) DNA further serves as a template for DNA synthesis with the corresponding inner primers (e.g., BIP) (Figure 1C) and outer primers (e.g., B3) (Figure 1D). The newly released ssDNA can fold into a dumbbell DNA molecule not only to initiate the self-primed DNA synthesis but also to serve as a template for new amplification cycles using the inner primers (Figure 1E–G).

Figure 1.

Figure 1

The principle of the traditional LAMP. F3 and B3: Forward and backward outer primers, respectively; FIP (F1c + F2) and BIP (B1c + B2): Forward and backward inner primers, respectively; LF and LB: Forward and backward loop primers, respectively.

In 2002, Nagamine et al. further optimized the LAMP method to accelerate amplification by adding additional loop primers [27]. Loop primers (forward: LF and backward: LB) bind to the loop regions of the complex dumbbell structure and promote new DNA synthesis to accelerate isothermal amplification (Figure 1H). As a result, the number of LAMP products can increase even up to a billion copies in less than an hour [23,27]. LAMP has been demonstrated to have high sensitivity and good tolerance to PCR-inhibiting substances [15,16]. LAMP also exhibits the versatility of result reading and the ability to expand detection in combination with other technologies. Among the methods most widely used for the detection of LAMP signal are turbidity caused by precipitated magnesium pyrophosphate, sequence-independent intercalating dyes like metal indicators for calcium, intercalating fluorescent dyes such as SYBR green and STYO 9, and pH-sensitive indicators (e.g., Cresol red) [15,28,29]. In addition, LAMP can be combined with other different technologies to further expand its detection capacity. For example, magnetic beads or gold nanoparticles (AuNPs) can be used to create easily manageable lateral flow tests in combination with LAMP assays [30,31]. Furthermore, LAMP amplicon detection can also be combined with various sequencing techniques, such as nanopore sequencing [32], pyrosequencing [33], or high-throughput sequencing [34], to realize multiplex detection of different pathogens (e.g., Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) and other respiratory viruses).

Despite the advantage in sensitivity and amplification speed, the traditional LAMP methods have some limitations. The major limitation is that LAMP frequently yields spurious amplicons if the primers interact with each other [35], which can result in false positive results when sequence-independent intercalating dyes or pH-sensitive indicators are used [35,36,37,38]. Another limitation is that it is difficult for LAMP to achieve single-tube multiplex detection, and as a consequence, most traditional LAMP methods only detect one target in a single reaction [15,35,39]. To fulfill the needs of specific target detection and one-pot multiplex reaction, several probe-based LAMP assays have been developed (Figure 2). These probe-based LAMP assays not only guarantee the accurate detection of the target without being affected by non-specific products but also enable the multiplex detection of various targets in a single reaction [16].

Figure 2.

Figure 2

Evolution of probe-based LAMP assays.

Herein, we review the evolution of probe-based LAMP assays, describe the principles of these methods, and summarize their applications in pathogen detection. Furthermore, the review discusses the advantages of probe-based LAMP assays over the traditional LAMP methods and highlights their superiority and flexibility in clinical application.

2. The Principles of the Probe-Based LAMP Assays

In the probe-based LAMP assays, different probes were designed for LAMP assays, including the modification of one of the core LAMP primers and/or the introduction of additional probes or additives to the reaction [16,40].

2.1. Assimilating Probe-Based LAMP Assay

In 2011, Kubota et al. developed the assimilating probe-based LAMP assay [41]. The assimilating probes consist of two partially complementary oligonucleotides: a fluorescence probe and a quenching probe. The fluorescence probe was designed by adding a fluorophore (denoted by F, red)-labeled universal oligonucleotide (F strand) to the 5′-end of the loop primer (LF or LB). The quenching probe (Q strand) complements the F strand of the fluorescence probe and is labeled with a quencher (denoted by Q, blue) at its 3′ end. In the absence of target DNA, the fluorescence is quenched since the hybridization of the fluorescent probe and the quenching probe brings the quencher and the fluorophore together (Figure 3A). In the presence of target DNA, LAMP is initiated. Along the DNA strands initiated by the fluorescent probe, the newly synthesized DNA strand displaces the quenching probe to release the fluorescent signal of the fluorophore (Figure 3B).

Figure 3.

Figure 3

The principle of the assimilating probe-based LAMP assay. F: fluorophore; Q: quencher.

2.2. Detection of Amplification Using Release of Quenching (DARQ)

The DARQ was first reported by Tanner in 2012 [42]. Similar to the assimilating probe-based LAMP assay, a partially double-stranded linear DNA probe was also used in DARQ. The probe was designed by labeling a quencher at the 5′ end of the FIP and adding an additional 3′ fluorophore-labeled probe that is complementary to F1c of FIP (Figure 4A). In the absence of target DNA, no fluorescent signal is generated from the double-stranded linear DNA probe. During the LAMP amplification, the newly synthesized DNA strand displaces the 3′ fluorophore-labeled probe to release a fluorescent signal (Figure 4B).

Figure 4.

Figure 4

The principle of the Detection of amplification using the release of quenching (DARQ). F: fluorophore; Q: quencher.

2.3. Quenching Probe (Q Probe)-Based LAMP Assay

In 2018, Shiratoa et al. developed a reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay using a quenching probe (Q probe) [43]. The Q probe that targets the loop regions of the LAMP amplicons was simply designed by lengthening the 3′ end of one loop primer by 10–15 nt and featured by a 3′ fluorophore (e.g., BODIPY)-labeled cytosine (Figure 5A). When the Q probe hybridizes with the target sequence, the fluorescence is quenched by the guanine residue present in the target sequences (Figure 5B). In 2019, Takayama et al. further simplified the design of the Q probe by directly labeling the fluorophore at the 5′ cytosine of one of the loop primers [44]. Distinct from the double-stranded linear DNA probes described above, the fluorescent signal of the Q probe is released in the absence of target DNA, and the fluorescence is quenched when the Q probe hybridizes with the targets. As a result, the fluorescent signal decreases, accompanied by the accumulation of amplicons during LAMP amplification. The LAMP amplicons can be real-time monitored using a real-time fluorescent PCR instrument, and exponential amplification results in an inverted S curve with continuous quenched fluorescent signals.

Figure 5.

Figure 5

The principle of the Quenching probe (Q probe)-based LAMP assay. F: fluorophore; Q: quencher; C: cytosine.

2.4. Enzyme-Mediated Probe-Based LAMP Assays

Enzyme-mediated probe-based LAMP assays utilize an extra enzyme (e.g., endonuclease, high fidelity DNA polymerase) or Bst 5.0/5.1 DNA polymerase to recognize and cut the probes to release fluorescent signals.

2.4.1. Multiple Endonuclease Restriction Real-Time Loop-Mediated Isothermal Amplification (MERT-LAMP)

Wang et al. developed MERT-LAMP in 2015 [45]. In MERT-LAMP, the probe was designed by linking an additional oligonucleotide to the 5′ end of an inner primer (e.g., FIP), and the oligonucleotide carried a 5′ fluorophore and a 3′ quencher, both of which were separated by a restriction endonuclease recognition site (Figure 6A). In the initial amplification phase, the forward inner primer FIP with endonuclease recognition site binds to the F2c region to promote the synthesis of DNA. The newly synthesized strand is displaced by upstream DNA synthesis primed using the outer primer F3. The BIP annealing to the B2c sequence synthesizes the complementary sequence of the oligonucleotide probe. When the double-stranded DNA containing the endonuclease recognition site is generated, Nb.BsrDI recognizes the restriction endonuclease site and digests the double-stranded DNA. This process separates the fluorophore and the quencher, resulting in a fluorescent signal (Figure 6B). Accompanied by the exponential amplification of LAMP, fluorescence signals are continuously generated.

Figure 6.

Figure 6

The principle of the MERT-LAMP. F: fluorophore; Q: quencher.

2.4.2. Tth Endonuclease Cleavage Loop-Mediated Isothermal Amplification (TEC-LAMP)

Higgins et al. developed TEC-LAMP in 2018 [46]. Instead of an endonuclease recognition site, an abasic site that can be cleaved with Tth endonuclease IV was used to link the fluorophore and the quencher. The reaction process is similar to that of MERT-LAMP (Figure 6).

2.4.3. High-Fidelity DNA Polymerase-Mediated LAMP

To overcome the shortcomings of existing LAMP methods for detecting SARS-CoV-2, three high-fidelity DNA polymerase-mediated real-time LAMP methods with similar principles have been developed since 2021 [36,37,38]. The release of a fluorescent signal was termed “proofreading” enzyme-mediated probe cleavage (Proofman probe) [36], high-fidelity DNA polymerase mediated probe (HFman probe) [37], and Luminescence from ANticipated Target due to Exonuclease Removal of Nucleotide mismatch (LANTERN) [38]. In these assays, one of the loop primers with or without a 3′-mismatched base is selected as the probe by labeling a fluorophore and a quencher at the 3′ end and 5′ end, or the 5′ end and 3′ end, respectively (Figure 7A). To perform real-time monitoring, a small amount of high-fidelity DNA polymerase is introduced into the LAMP system. During LAMP amplification, the probes specifically bind to the loop regions of the dumbbell structure, and then high-fidelity DNA polymerase recognizes and cuts 3′-matched and/or mismatched base labeled with a fluorophore to release fluorescent signal (Figure 7B) [36,37,38]. Previous studies showed that high-fidelity DNA polymerase could well recognize and cut the probes that specifically bind to the target sequence, regardless of the presence or absence of a mismatched base at the 3′ end of the probes or primers [37,47,48]. Furthermore, after removing the 3′-labeled base, the probe exposes a 3′-OH and then is used as a primer to initiate new DNA synthesis using Bst DNA polymerase, which facilitates the amplification.

Figure 7.

Figure 7

The principle of High-fidelity DNA polymerase-mediated LAMP. F: fluorophore; Q: quencher.

2.4.4. LAMP Coupled with TaqMan Probe and a New Generation Bst DNA Polymerase (Taqman-LAMP)

In 2021, Liang et al. reported the development of a TaqMan-LAMP, which used an innovative Bst 5.0/5.1 DNA polymerase (Figure 8A) [49]. Differing from Bst 4.0 DNA polymerase and earlier Bst DNA polymerases, the innovative Bst 5.0/5.1 DNA polymerase does not only have a DNA synthesis activity and a stronger strand displacement capacity but also a 5′-3′ exonuclease activity. The sequence located between F1 and F2 is selected as a target for the design of a TaqMan probe that is labeled with fluorophore and quencher at its 5′ end and 3′ end, respectively. The inner primer FIP is annealed to F2c to promote DNA synthesis. When the Taqman probe specifically binds to the target, Bst 5.0/5.1 DNA polymerase can cleave the TaqMan probe to release the fluorescent signal (Figure 8B). During the exponential amplification phase, ongoing binding and cleavage of the TaqMan probes result in a continuous generation of fluorescent signals.

Figure 8.

Figure 8

The principle of the Taqman-LAMP. F: fluorophore; Q: quencher.

3. Applications of the Probe-Based LAMP Methods in Pathogen Detection

The urgent demand for specific and accurate POCT detection of emerging and reemerging infectious diseases has promoted the development of various probe-based LAMP methods [17,18,19,50]. The applications of various probe-based LAMP assays in pathogen detection are summarized in Table 1.

Kubota et al. developed the assimilating probe-based LAMP and subsequently designed a simple, low-power, portable handheld device to enable real-time detection of the LAMP amplicons [41,51]. This assay was mainly used for singleplex detection of viruses, and the sensitivities of this assay in the detection of foot-and-mouth disease virus (FMDV) and porcine circovirus 3 (PCV3) were 102 copies per µL and 50 copies per reaction, respectively [52,53].

The DARQ was first reported by Tanner et al. for singleplex and quadruplex detection of E. coli, Caenorhabditis elegans, the human cystic fibrosis transmembrane conductance regulator, and human BRCA1 DNA [42]. Since its development, the assay has been applied in singleplex or multiplex detection of other pathogens [54,55,56,57,58,59,60]. For example, Fan et al. used DARQ to develop a visual multiplex fluorescent LAMP assay for the detection of FMDV, vesicular stomatitis virus (VSV), and bluetongue virus (BTV) [59]. The detection limit was as low as 526–2477 copies/reaction for plasmid standards of these viruses, indicating that the multiplex LAMP assay can simultaneously detect 1–3 targets in a single reaction tube. The same team further developed a multiplex fluorescence-based LAMP assay for detecting chicken parvovirus (ChPV), chicken infectious anemia virus (CIAV), and fowl adenovirus serotype 4 (FAdV-4), with detection limits of 307, 749 and 648 copies per reaction, respectively [60].

Takayama et al. developed a reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay using a Q probe for the detection of the influenza virus (IV), respiratory syncytial virus (RSV), and rhinovirus [44,61]. Shirato et al. used the Q probe to develop a real-time LAMP assay for the detection of the Middle East respiratory syndrome coronavirus (MERS-CoV) [43]. The Q probe-based LAMP was also applied to detect Genus Phytophthora, Species Phytophthora, and severe fever with thrombocytopenia syndrome virus (SFTSV) [62,63]. However, all the above-mentioned Q probe-based LAMP assays were developed in a singleplex format.

In order to achieve single-tube multiplex detection of Listeria monocytogenes and Listeria ivanovii, Wang et al. developed MERT-LAMP [45]. As low as 250 fg of DNA per reaction of tested Listeria DNA could be detected. Higgins et al. developed TEC-LAMP for the simultaneous detection of Streptococcus pneumoniae, Neisseria meningitidis, and Haemophilus influenzae [46]. The limit of detection (LOD) of the LEC-LAMP assay was 3.1 genome copies per reaction for the singleplex detection of N. meningitidis and 102 genome copies per reaction for multiplex detection of all three bacteria [46].

In the high-fidelity DNA polymerase-mediated LAMP assay, Ding et al. first introduced a thermostable proofreading DNA polymerase (Pfu) to trigger the cleavage of the Proofman probe [36]. The N and Orf1ab genes of SARS-CoV-2 and human glyceraldehyde 3-phosphate dehydrogenase (GAPDH) gene were simultaneously detected with different fluorophores, and the detection limit was reported to be 100 SARS-CoV-2 RNA copies per reaction [36]. However, the performance of the SARS-CoV-2 RT-LAMP assay was not evaluated using clinical samples. Soon, a more sensitive SARS-CoV-2 RT-HF-LAMP assay was developed using an HFman probe [37]. Subsequently, HFman probe-based LAMP (HF-LAMP) was used to develop various assays for singleplex or multiplex detection of human immunodeficiency virus-1 (HIV-1), Hantaan virus (HTNV) and Seoul virus (SEOV), Monkeypox virus (MPXV) and Monkey B virus (BV), and BK virus (BKV), and the detection sensitivity can achieve three copies of viral RNA per reaction [64,65,66,67]. Evaluation with clinical samples showed an excellent consistency of HF-LAMP assays to commercial RT-qPCR assays. Importantly, the HF-LAMP method was demonstrated to have high detection potential for highly variable viruses and can be developed to an extraction-free format for the detection of various human viruses directly using clinical samples [64,65,66,67].

Liang et al. developed a TaqMan-LAMP and used this method for rapid and specific detection of the F gene of Pigeon paramyxovirus-1 (PPMV-1) [49]. The assay could be completed within 25 min, and the LOD was estimated to be 10 copies per µL for PPMV-1 cDNA and 0.1 ng for PPMV-1 RNA [49].

More application examples of these probe-based LAMP methods in the detection of various pathogens can be found in Table 1.

Table 1.

Applications of the probe-based LAMP methods in pathogen detection.

Methods Year Singleplex or Multiplex Targets Reaction Condition Sensitivity Limitation of Detection Clinical Evaluation Tt Values Ref.
Assimilating probe 2011 singleplex Ralstonia solanacearum 65 °C for 60 min 50 fg 137 copies per reaction NE * consistently around 20 min regardless of copy number (102 to 106 gene copies) [41]
Assimilating probe 2011 singleplex Salmonella enterica 65 °C for 60 min 5 pg 15 copies per reaction NE NA ** [51]
Assimilating probe 2015 multiplex Salmonela enterica and Phage λ 65 °C for 30 min 50 fg 9.8 × 103 and 1000 copies per reaction for Salmonella enterica and Phage λ NE NA [68]
Assimilating probe 2020 singleplex foot-and-mouth disease virus 62 °C for 30 min 100 copies 2.5 × 103 copies per reaction 69 clinical samples (28 blood, 28 oropharyngeal fluid, and 13 tissues) 4.19–15.98 (9.95 ± 3.08) min [52]
Assimilating probe 2020 singleplex Porcine circovirus type 3 62 °C for 30 min 50 copies NA 326 pig samples (136 tissues and 190 sera) 8.42–27.73 (17.34 ± 4.45) min [53]
DARQ 2012 multiplex Bacteriophage λ, HeLa, Escherichia coli, and Caenorhabditis elegans genomic DNA 65 °C for 60 min 5 ng, 10 pg, 5 ng, 82.5 ng for Bacteriophage λ, HeLa, Escherichia coli, and Caenorhabditis elegans genomic DNA, respectively. NA NE 11.8 ± 0.03 min [42]
DARQ 2016 singleplex Salmonella 65 °C for 40 min and inactivated at 85 °C for 5 min. 10 copies NA 312 fecal samples NA [54]
DARQ 2017 singleplex avian reovirus 65 °C for 60 min 10 copies NA 98 clinical tendon tissue samples NA [55]
DARQ 2019 multiplex methicillin-resistant Staphylococcus aureus (MRSA) 65 °C for 60 min 103 copies, 103 copies, 104 copies for singleplex detection of femB, spa, and mecA of MRSA genes.104 copies for duplex and triplex detection. NA NE NA [56]
DARQ 2019 singleplex Brucella 65 °C for 45 min and terminated at 85 °C for 5 min. 20 copies NA 250 samples NA [57]
DARQ 2021 multiplex SARS-CoV-2, influenza A, influenza B, human RNA 60 °C, 15 s for 108 cycles 50 copies of SARS-CoV-2 RNA, 1:10000 diluted influenza A RNA (VR-1737D) and approximately 21 copies of influenza B RNA (VR-1885DQ). NA NE NA [58]
DARQ 2022 multiplex foot-and-mouth disease, vesicular stomatitis, and bluetongue viruses 63 °C for 75 min, 80 °C for 5 min for termination 1000 copies for FMDV, 100 copies for VSV, BTV-4 2477copies/reaction for FMDV, 526 copies/reaction for VSV, and 913 copies/reaction for BTV 111 clinical samples, including 12 vesicular fluid samples, 30 esophageal–pharyngeal samples, 42 whole-blood samples, 6 vesicular skin samples, and 21 oral swabs collected from cattle NA [59]
DARQ 2023 multiplex Chicken parvovirus (ChPV), chicken infectious anemia virus (CIAV), and fowl adenovirus serotype 4 (FAdV-4) 63 °C for 75 min for amplification and 80 °C for 5 min for termination 66,106,95 copies for ChPV, CIAV, and FAdV-4, respectively. 307 copies/µL for ChPV, 749 copies/µL for CIAV and 648 copies/µL for FAdV-4 342 samples (cloacal swab, heart, liver, and kidney) collected from chicken farms NA [60]
Q Probe 2017 singleplex Fusarium oxysporum f. sp. Lycopersici 66 °C for 60 min, and subsequent melting curve analysis from 30 or 35 °C to 95 °C with a decrement of 0.2 °C per second. 3 ng NA 4 Soil DNA NA [69]
Q Probe 2018 singleplex Middle East respiratory syndrome coronavirus 63 °C for 30 min 20 copies NA 19 nasal aspirates, secretions, or swabs 20.2–30.9 min [43]
QProbe 2020 singleplex Genus Phytophthora and Species Phytophthora preheat 5 min at 68 °C, 68 °C for 60 min, and annealing curve analysis at 98 °C to 80 °C, ramping at 0.05 °C per sec. 100 fg NA 161 taxa, including subspecies, varieties, and hybrids) of phytophthora, 12 species (12 isolates) of Pythium, 17 species (17 isolates) of Phytopythium, and one isolate of each of the soil-borne pathogens. NA [62]
Q Probe 2021 singleplex severe fever with thrombocytopenia syndrome virus (SFTSV) 63 °C for 30 min 10–100 copies NA 12 SFTSV strains NA [63]
Enzyme-mediated probe-based LAMP assays (MERT-LAMP) 2015 multiplex L. monocytogenes and L. ivanovii 64 °C for 60 min and then at 80 °C for 5 min 250 fg NA 70 raw meat samples (including pork, beef, lamb, and chicken samples) NA [45]
Enzyme-mediated probe-based LAMP assays (TEC-LAMP) 2018 multiplex Streptococcus pneumoniae, Neisseria meningitidis, Haemophilus influenzae, and Internal Amplification Control 67 °C for 60 min 100 copies for S. pneumoniae, N. meningitidis, and H. influenzae, 50 copies for Internal Amplification Control 39.5, 17.3, and 25.9 genome copies per reaction for S. pneumoniae, N. meningitidis, and H. influenzae, respectively. 65 samples, including 34 blood, 5 blood culture, 17 CSF, 5 pleural fluid, 1 knee fluid, and 3 other body fluids NA [46]
Enzyme-mediated probe-based LAMP assays (Proofman) 2021 multiplex N and Orf1ab genes SARS-CoV-2, human glyceraldehyde-3-phosphate dehydrogenase (GAPDH) 60 °C for 60 min 100 copies of N gene RNA 100 copies per reaction NE NA [36]
Enzyme-mediated probe-based LAMP assays (HFman) 2022 multiplex ORF and E genes of SARS-CoV-2, human β-actin gene 64 °C for 50 min 30 copies of ORF and E gene RNA 78 and 115 copies per reaction for ORF gene and E gene, respectively. 190 nasopharyngeal swabs (NP) samples 4.0–33.5 min (ORF gene),4.0–45.0 min (E gene) [37]
Enzyme-mediated probe-based LAMP assays (HFman) 2022 singleplex HIV-1 subtypes, including CRF01_AE, CRF07_BC, CRF08_BC, CRF55_01B, and unique recombinant forms (URFs). 64 °C for 50 min 3 copies 89 copies per reaction 101 plasma sample 9.67–29.98 min [64]
Enzyme-mediated probe-based LAMP assays (HFman) 2022 multiplex Hantaan virus (HTNV) and Seoul virus (SEOV) 64 °C for 50 min 3 copies 41 and 73 copies per reaction for HTNV and SEOV, respectively. 46 serum samples NA [65]
Enzyme-mediated probe-based LAMP assays (HFman) 2023 multiplex Monkeypox virus (MPXV) and Monkey B virus (BV) 64 °C for 50 min 3 copies 28.7 and 27.8 copies per reaction for MPXV and BV, respectively. simulated 6 serum samples collected from monkeys NA [66]
Enzyme-mediated probe-based LAMP assays (HFman) 2023 singleplex BK virus (BKV) 64 °C for 50 min 3 copies 12 copies/reaction 132 urine samples from HIV-1 infected individuals, 20 BKV positive samples, and 10 BKV negative samples 3.9 to 15.4 min [67]
Enzyme-mediated probe-based LAMP assays (LANTERN) 2022 multiplex S gene of SARS-CoV-2 and human ACTB 65 °C for 40 min 2 copies 8 copies per reaction 52 COVID-19 positive samples and 22 COVID-19 negative samples NA [38]
Enzyme-mediated probe-based LAMP assays (TaqMan-LAMP) 2021 singleplex pigeon paramyxovirus type 1 65 °C for 10 s and 65 °C for 50 s with 25 cycles. 10 copies NA 108 fecal samples NA [49]

* NE, not evaluated. ** NA, not available.

4. Comparison of the Probe-Based LAMP Methods with the Traditional LAMP Methods

The probe-based LAMP methods were developed from the traditional LAMP methods. Because of the use of a probe in the reaction system, the instruments and detection system for the probe-based LAMP methods are different from those for the traditional LAMP methods (Table 2). Real-time thermal cycler or small portable isothermal fluorimeters are required for the real-time monitoring of LAMP amplicons, or regular thermal cycler or low-cost simple heating block (e.g., dry incubator) for end-point fluorescence detection of the products. Therefore, the facility for the probe-based LAMP methods is more likely to be more costly than the traditional LAMP methods. It is worth noting that some portable fluorescence analyzers and isothermal fluorimeters with minimal laboratory resources are currently available or under development (Supplementary Table S1) [70,71,72,73,74,75]. Furthermore, end-point fluorescence detection requires much less instrumentation than real-time monitoring. Using a simple, low-cost device, such as a UV transilluminator [76] or light-boxes using cardboard and filter paper [38], the results can be easily achieved post-amplification.

Table 2.

Comparison of the probe-based LAMP methods with the traditional LAMP methods.

Traditional LAMP Probe-Based LAMP
Reaction instrument Thermal cycler/small portable isothermal devices/simple, low-cost device Real-time thermal cycler/small portable isothermal fluorimeters/simple, low-cost device
Reaction time about 60 min less than 30 min
Detection method Agarose gel electrophoresis/Turbidity/Fluorescence detection by non-specific fluorescent dyes or pH indicators Real-time monitoring/end-point fluorescence detection
Specificity (non-specific signal) Relatively low High
Sensitivity (copies/reaction) High Slightly higher
Single-tube multiplex detection No Yes
The possibility of false positive results Relatively high Low
Operation Difficult Relatively easy
Testing cost Slightly low Low
Potential for commercial and/or clinical application Low High

Since the first description of the LAMP method in 2000, a few LAMP assays have been commercially used [77]. The major reason is the low specificity and accuracy of the traditional LAMP methods, which are largely ascribed to the inevitable introduction of non-specific amplification during the reaction and the use of sequence-independent double-stranded DNA binding dyes or pH-sensitive indicators [15,77]. The use of probes can largely improve the specificity of NAATs since the specific hybridization of a probe with its target is a prerequisite for enzymes to recognize and cut the probe and further release fluorescence signals [16]. Therefore, like probe-based qPCR assays, the probe-based LAMP assays are highly specific and accurate for the detection of various pathogens compared to the traditional LAMP assays, and meanwhile, the use of different probes confers the single tube multiplex detection capacity of the probe-based LAMP assays (Table 2) [16]. Furthermore, the sensitivities of the probe-based LAMP methods are slightly higher, and the process is slightly easier than the traditional LAMP methods.

5. Comparison of Various Probe-Based LAMP Methods

The various probe-based LAMP methods exhibit distinct features due to being derived from different principles (Table 3). Of these methods, the probe designs for assimilating probe-LAMP, Q probe-LAMP, MERT-LAMP/TEC-LAMP, and Taqman-LAMP are relatively difficult. The Assimilating probe-LAMP requires an additional oligonucleotide probe, and the TaqMan-LAMP requires the design of a probe between F1 and F2 or B1 and B2. Furthermore, a terminal cytosine and an endonuclease recognition site are required for the probe of the Q probe-LAMP and the MERT-LAMP/TEC-LAMP, respectively. Although the probe-based LAMP methods largely improve the specificity of the assays, the specificity of the Assimilating probe-LAMP, DARQ, and MERT-LAMP/TEC-LAMP are relatively lower compared to the other three methods. The main reason is that their probes are involved in the inner primers that more likely yield non-specific amplification and the release of the fluorescent signal depends on the strand displacement of the double-stranded linear probes.

Table 3.

Comparison of various probe-based LAMP assays.

Methods Probe Design Specificity Sensitivity (LOD, Copies per Reaction) * One-pot Multiplexing Variant-Tolerance Amplification Speed (Tt Values) *
Assimilating probe Difficult. Two partially complementary oligonucleotides need to be designed. Relatively low. The probe is involved in one of the inner primers that more likely result in non-specific amplification, and the release of fluorescent signal depends on strand displacement of double-stranded probe. Low
(15–9.8 × 103)
Yes No Relatively fast
(4.19–27.73 min)
DARQ Easy Relatively low. The probe is involved in one of the inner primers that more likely result in non-specific amplification, and the release of fluorescent signal depends on strand displacement of double-stranded probe. High
(526–2477)
Yes No Relatively fast
(11.8 ± 0.03 min)
Q Probe Difficult. The 3′ or 5′ end of at least one loop primer is required to be cytosine for labeling a fluorescent dye. High High
(ND)
Yes $ No Relatively slow (20.2–30.9 min)
MERT-LAMP/TEC-LAMP Difficult. An endonuclease recognition site is required to be included in the probe. Relatively low. The probe is involved in one of the inner primers that more likely result in non-specific amplification, and the release of fluorescent signal depends on strand displacement of double-stranded probe. High
(17.3–39.5)
Yes No Relatively slow (ND)
Proofman/HFman/LANTERN Slightly easier. Any one of the loop primers can be selected as the probe by labeling a fluorophore and a quencher at both ends. High High
(8–100)
Yes Yes Relatively fast
(4.0–45.0 min)
TaqMan-LAMP Difficult. The probe needs to be designed between F1 and F2 or B1 and B2. High High
(ND)
Yes $ No Relatively fast
(ND)

* Based on self-reported data. $ According to the principles, it enables multiplex detection. However, there is no data to support this. “ND”, No data in the original paper.

According to the principles, all these probe-based methods have the capacity for multiplex detection. However, it is relatively difficult for these LAMP methods to realize single-tube high-sensitive detection of three or more targets (triple or quadruple) since the presence of too many primers might cause complex primer–primer interaction and therefore reduce amplification efficiency (detection sensitivity) (Table 1 and Table 3). The MERT-LAMP/TEC-LAMP and the Proofman/HFman/LANTERN-based methods appear to have higher detection sensitivity than other probe-based methods. All these methods have faster amplification than the traditional LAMP assays, and most of these methods have Tt values of less than 30 min.

On the other hand, most emerging and remerging infectious diseases are caused by viruses (especially RNA viruses such as the recent SARS-CoV-2 and influenza viruses), which are more variable. The presence of mismatched between primers/probes and viral genomic sequence may reduce LAMP amplification efficiency, resulting in lower detection sensitivity and even a failure in detection [28]. The introduction of high-fidelity DNA polymerase enables Proofman/HFman/LANTERN-based LAMP assays with high potential to detect various viral variants (Table 3) [28,30,37,78].

6. Conclusions and Future Perspectives

As a promising method for nucleic acid detection, LAMP has attracted more and more attention in the last two decades for POCT detection of animal and plant viruses, bacteria, fungi, mycoplasma, parasites, and other pathogens due to its high sensitivity, rapidity, and simplicity [6,19,26,79,80]. The major reason making the traditional LAMP methods inferior to qPCR assays and limiting their commercial and clinical application is their low specificity and accuracy, which are ascribed to non-specific amplification and/or the use of sequence-independent interferents (e.g., non-specific double-stranded DNA binding dyes) [15,35,77]. The developments of various probe-based LAMP methods not only resolve the problems of non-specific detection and low accuracy but also improve the amplification speed and detection sensitivity and enable single-pot multiplex detection [37,38,40]. In view of its advantages over the traditional LAMP methods and its comparability to qPCR in specificity, sensitivity, multiplex detection capacity, and faster speed than the latter, the probe-based LAMP methods show high commercial and clinical application potential to be readily used, especially in the detection of emerging and reemerging infectious diseases.

Despite the promising advancements of the probe-based LAMP methods, two underlying questions remain to be further addressed for future POCT applications at home and in the field. First, purified nucleic acids are often used in various NAATs. Although several extraction-free probe-based LAMP assays were developed for SARS-CoV-2 detection [81,82,83,84], the extraction-free assays largely depend on the sample types. For example, because of the relatively low inhibition of nasopharyngeal swabs (NPs) on amplification reactions, NPs are more suitable for extraction-free nucleic acids detection regardless of qPCR and LAMP than other sample types (e.g., plasma/serum and tissues) [67,84]. Therefore, additives that can mitigate the inhibition of clinical samples on LAMP amplification deserve to be developed. Furthermore, a suitable buffer system for the collection of NPs and other samples also needs to be developed since currently used commercial viral transport media often contain substances inhibiting LAMP amplification [37,67].

Second, the results of the probe-based LAMP methods can be real-time monitored using a real-time (thermal or isothermal) fluorimeter or be observed at the end-point using a regular fluorimeter [70,71,73]. A small, portable, cheap, and battery-powered isothermal fluorimeter should be developed. In particular, a small portable device that integrates rapid nucleic acid extraction, isothermal amplification, and fluorescence monitoring is strongly encouraged.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics13091530/s1, Table S1: Information list of portable fluorescence analyzers.

Author Contributions

Conceptualization, C.Z.; investigation, X.Z., Y.Z. (Yi Zeng), and Y.Z. (Yongjuan Zhao); resources, X.Z.; writing—original draft preparation, X.Z.; writing—review and editing, C.Z.; visualization, X.Z.; supervision, C.Z.; funding, C.Z. All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data sharing is not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Funding Statement

This research was funded by Shanghai Science& Technology Innovation Action Program (21Y11900500) and Shanghai Public Health Clinical Center hospital-level scientific research project (KY-GW-2023-32).

Footnotes

Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

References

  • 1.Brister J.R., Bao Y., Bao Y., Zhdanov S.A., Ostapchuck Y., Chetvernin V., Kiryutin B., Zaslavsky L., Kimelman M., Tatusova T.A. Virus Variation Resource—Recent updates and future directions. Nucleic Acids Res. 2014;42:D660–D665. doi: 10.1093/nar/gkt1268. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Hatcher E.L., Zhdanov S.A., Bao Y., Blinkova O., Nawrocki E.P., Ostapchuck Y., Schaffer A.A., Brister J.R. Virus Variation Resource—Improved response to emergent viral outbreaks. Nucleic Acids Res. 2017;45:D482–D490. doi: 10.1093/nar/gkw1065. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Sanders R.W.d.J., Menno D. Pandemic moves and countermoves: Vaccines and viral variants. Lancet. 2021;397:1326–1327. doi: 10.1016/S0140-6736(21)00730-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Kaya H.O., Cetin A.E., Azimzadeh M., Topkaya S.N. Pathogen detection with electrochemical biosensors: Advantages, challenges and future perspectives. J. Electroanal. Chem. 2021;882:114989. doi: 10.1016/j.jelechem.2021.114989. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Islam K.U., Iqbal J. An Update on Molecular Diagnostics for COVID-19. Front. Cell. Infect. Microbiol. 2020;10:560616. doi: 10.3389/fcimb.2020.560616. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Habibzadeh P., Mofatteh M., Silawi M., Ghavami S., Faghihi M.A. Molecular diagnostic assays for COVID-19: An overview. Crit. Rev. Clin. Lab. Sci. 2021;58:385–398. doi: 10.1080/10408363.2021.1884640. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Gupta N., Augustine S., Narayan T., O’Riordan A., Das A., Kumar D., Luong J.H.T., Malhotra B.D. Point-of-Care PCR Assays for COVID-19 Detection. Biosensors. 2021;11:141. doi: 10.3390/bios11050141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Kojabad A.A., Farzanehpour M., Galeh H.E.G., Dorostkar R., Jafarpour A., Bolandian M., Nodooshan M.M. Droplet digital PCR of viral DNA/RNA, current progress, challenges, and future perspectives. J. Med. Virol. 2021;93:4182–4197. doi: 10.1002/jmv.26846. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Zhang J., Lv H.L., Li L.X., Chen M.J., Gu D.Y., Wang J., Xu Y. Recent Improvements in CRISPR-Based Amplification-Free Pathogen Detection. Front. Microbiol. 2021;12:751408. doi: 10.3389/fmicb.2021.751408. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Mustafa M.I., Makhawi A.M. SHERLOCK and DETECTR: CRISPR-Cas Systems as Potential Rapid Diagnostic Tools for Emerging Infectious Diseases. J. Clin. Microbiol. 2021;59:e00745-20. doi: 10.1128/JCM.00745-20. [DOI] [PMC free article] [PubMed] [Google Scholar] [Retracted]
  • 11.Kaminski M.M., Abudayyeh O.O., Gootenberg J.S., Zhang F., Collins J.J. CRISPR-based diagnostics. Nat. Biomed. Eng. 2021;5:643–656. doi: 10.1038/s41551-021-00760-7. [DOI] [PubMed] [Google Scholar]
  • 12.Datta M., Singh D.D., Naqvi A.R. Molecular Diagnostic Tools for the Detection of SARS-CoV-2. Int. Rev. Immunol. 2021;40:143–156. doi: 10.1080/08830185.2020.1871477. [DOI] [PubMed] [Google Scholar]
  • 13.Sritong N., Sala de Medeiros M., Basing L.A., Linnes J.C. Promise and perils of paper-based point-of-care nucleic acid detection for endemic and pandemic pathogens. Lab Chip. 2023;23:888–912. doi: 10.1039/D2LC00554A. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Zhu H., Zhang H., Xu Y., Laššáková S., Korabečná M., Neužil P. PCR past, present and future. Biotechniques. 2020;69:317–325. doi: 10.2144/btn-2020-0057. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Soroka M., Wasowicz B., Rymaszewska A. Loop-Mediated Isothermal Amplification (LAMP): The Better Sibling of PCR? Cells. 2021;10:1931. doi: 10.3390/cells10081931. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Becherer L., Borst N., Bakheit M., Frischmann S., Zengerle R., von Stetten F. Loop-mediated isothermal amplification (LAMP)—Review and classification of methods for sequence-specific detection. Anal. Methods. 2020;12:717–746. doi: 10.1039/C9AY02246E. [DOI] [Google Scholar]
  • 17.Bodulev O.L., Sakharov I.Y. Isothermal Nucleic Acid Amplification Techniques and Their Use in Bioanalysis. Biochemistry. 2020;85:147–166. doi: 10.1134/S0006297920020030. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Pumford E.A., Lu J., Spaczai I., Prasetyo M.E., Zheng E.M., Zhang H., Kamei D.T. Developments in integrating nucleic acid isothermal amplification and detection systems for point-of-care diagnostics. Biosens. Bioelectron. 2020;170:112674. doi: 10.1016/j.bios.2020.112674. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Dinnes J., Deeks J.J., Adriano A., Berhane S., Davenport C., Dittrich S., Emperador D., Takwoingi Y., Cunningham J., Beese S., et al. Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection. Cochrane Database Syst. Rev. 2021;3:CD013705. doi: 10.1002/14651858.CD013705. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Compton J. Nucleic acid sequence-based amplification. Nature. 1991;350:91–92. doi: 10.1038/350091a0. [DOI] [PubMed] [Google Scholar]
  • 21.Malek L., Sooknanan R., Compton J. Protocols for Nucleic Acid Analysis by Nonradioactive Probes. Volume 28. Humana Press; Clifton, NJ, USA: 1994. Nucleic acid sequence-based amplification (NASBA) pp. 253–260. (Methods in Molecular Biology™ Book Series). [DOI] [PubMed] [Google Scholar]
  • 22.Baner J., Nilsson M., Mendel-Hartvig M., Landegren U. Signal amplification of padlock probes by rolling circle replication. Nucleic Acids Res. 1998;26:5073–5078. doi: 10.1093/nar/26.22.5073. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Notomi T., Okayama H., Masubuchi H., Yonekawa T., Watanabe K., Amino N., Hase T. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 2000;28:E63. doi: 10.1093/nar/28.12.e63. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Vincent M., Xu Y., Kong H. Helicase-dependent isothermal DNA amplification. EMBO Rep. 2004;5:795–800. doi: 10.1038/sj.embor.7400200. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Piepenburg O., Williams C.H., Stemple D.L., Armes N.A. DNA detection using recombination proteins. PLoS Biol. 2006;4:e204. doi: 10.1371/journal.pbio.0040204. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Gao D., Guo X., Yang Y., Shi H., Hao R., Wang S., Li Z.J., Zhao R., Song H. Microfluidic chip and isothermal amplification technologies for the detection of pathogenic nucleic acid. J. Biol. Eng. 2022;16:33. doi: 10.1186/s13036-022-00312-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Nagamine K., Hase T., Notomi T. Accelerated reaction by loop-mediated isothermal amplification using loop primers. Mol. Cell. Probes. 2002;16:223–229. doi: 10.1006/mcpr.2002.0415. [DOI] [PubMed] [Google Scholar]
  • 28.Zhou Y., Wan Z., Yang S., Li Y., Li M., Wang B., Hu Y., Xia X., Jin X., Yu N., et al. A Mismatch-Tolerant Reverse Transcription Loop-Mediated Isothermal Amplification Method and Its Application on Simultaneous Detection of All Four Serotype of Dengue Viruses. Front. Microbiol. 2019;10:1056. doi: 10.3389/fmicb.2019.01056. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Tanner N.A., Zhang Y., Evans T.C., Jr. Visual detection of isothermal nucleic acid amplification using pH-sensitive dyes. Biotechniques. 2015;58:59–68. doi: 10.2144/000114253. [DOI] [PubMed] [Google Scholar]
  • 30.Li Y., Chen X., Zhao Y., Wan Z., Zeng Y., Ma Y., Zhou L., Xu G., Reboud J., Cooper J.M., et al. A rapid variant-tolerant reverse transcription loop-mediated isothermal amplification assay for the point of care detection of HIV-1. Analyst. 2021;146:5347–5356. doi: 10.1039/D1AN00598G. [DOI] [PubMed] [Google Scholar]
  • 31.Zhu X., Wang X., Han L., Chen T., Wang L., Li H., Li S., He L., Fu X., Chen S., et al. Multiplex reverse transcription loop-mediated isothermal amplification combined with nanoparticle-based lateral flow biosensor for the diagnosis of COVID-19. Biosens. Bioelectron. 2020;166:112437. doi: 10.1016/j.bios.2020.112437. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Peto L., Rodger G., Carter D.P., Osman K.L., Yavuz M., Johnson K., Raza M., Parker M.D., Wyles M.D., Andersson M., et al. Diagnosis of SARS-CoV-2 Infection with LamPORE, a High-Throughput Platform Combining Loop-Mediated Isothermal Amplification and Nanopore Sequencing. J. Clin. Microbiol. 2021;59:e03271-20. doi: 10.1128/JCM.03271-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Liang C., Chu Y., Cheng S., Wu H., Kajiyama T., Kambara H., Zhou G. Multiplex loop-mediated isothermal amplification detection by sequence-based barcodes coupled with nicking endonuclease-mediated pyrosequencing. Anal. Chem. 2012;84:3758–3763. doi: 10.1021/ac3003825. [DOI] [PubMed] [Google Scholar]
  • 34.Warneford-Thomson R., Shah P.P., Lundgren P., Lerner J., Morgan J., Davila A., Abella B.S., Zaret K., Schug J., Jain R., et al. A LAMP sequencing approach for high-throughput co-detection of SARS-CoV-2 and influenza virus in human saliva. eLife. 2022;11:e69949. doi: 10.7554/eLife.69949. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Rolando J.C., Jue E., Barlow J.T., Ismagilov R.F. Real-time kinetics and high-resolution melt curves in single-molecule digital LAMP to differentiate and study specific and non-specific amplification. Nucleic Acids Res. 2020;48:e42. doi: 10.1093/nar/gkaa099. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Ding S., Chen G., Wei Y., Dong J., Du F., Cui X., Huang X., Tang Z. Sequence-specific and multiplex detection of COVID-19 virus (SARS-CoV-2) using proofreading enzyme-mediated probe cleavage coupled with isothermal amplification. Biosens. Bioelectron. 2021;178:113041. doi: 10.1016/j.bios.2021.113041. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Dong Y., Zhao Y., Li S., Wan Z., Lu R., Yang X., Yu G., Reboud J., Cooper J.M., Tian Z., et al. Multiplex, Real-Time, Point-of-care RT-LAMP for SARS-CoV-2 Detection Using the HFman Probe. ACS Sens. 2022;7:730–739. doi: 10.1021/acssensors.1c02079. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Ooi K.H., Liu M.M., Moo J.R., Nimsamer P., Payungporn S., Kaewsapsak P., Tan M.H. A Sensitive and Specific Fluorescent RT-LAMP Assay for SARS-CoV-2 Detection in Clinical Samples. ACS Synth. Biol. 2022;11:448–463. doi: 10.1021/acssynbio.1c00538. [DOI] [PubMed] [Google Scholar]
  • 39.Shirato K. Detecting amplicons of loop-mediated isothermal amplification. Microbiol. Immunol. 2019;63:407–412. doi: 10.1111/1348-0421.12734. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Becherer L., Bakheit M., Frischmann S., Stinco S., Borst N., Zengerle R., von Stetten F. Simplified Real-Time Multiplex Detection of Loop-Mediated Isothermal Amplification Using Novel Mediator Displacement Probes with Universal Reporters. Anal. Chem. 2018;90:4741–4748. doi: 10.1021/acs.analchem.7b05371. [DOI] [PubMed] [Google Scholar]
  • 41.Kubota R., Alvarez A.M., Su W.W., Jenkins D.M. FRET-Based Assimilating Probe for Sequence-Specific Real-Time Monitoring of Loop-Mediated Isothermal Amplification (LAMP) Biol. Eng. Trans. 2011;4:81–100. doi: 10.13031/2013.38509. [DOI] [Google Scholar]
  • 42.Tanner N.A., Zhang Y., Evans T.C., Jr. Simultaneous multiple target detection in real-time loop-mediated isothermal amplification. Biotechniques. 2012;53:81–89. doi: 10.2144/0000113902. [DOI] [PubMed] [Google Scholar]
  • 43.Shirato K., Semba S., El-Kafrawy S.A., Hassan A.M., Tolah A.M., Takayama I., Kageyama T., Notomi T., Kamitani W., Matsuyama S., et al. Development of fluorescent reverse transcription loop-mediated isothermal amplification (RT-LAMP) using quenching probes for the detection of the Middle East respiratory syndrome coronavirus. J. Virol. Methods. 2018;258:41–48. doi: 10.1016/j.jviromet.2018.05.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Takayama I., Nakauchi M., Takahashi H., Oba K., Semba S., Kaida A., Kubo H., Saito S., Nagata S., Odagiri T., et al. Development of real-time fluorescent reverse transcription loop-mediated isothermal amplification assay with quenching primer for influenza virus and respiratory syncytial virus. J. Virol. Methods. 2019;267:53–58. doi: 10.1016/j.jviromet.2019.02.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Wang Y., Wang Y., Lan R., Xu H., Ma A., Li D., Dai H., Yuan X., Xu J., Ye C. Multiple Endonuclease Restriction Real-Time Loop-Mediated Isothermal Amplification: A Novel Analytically Rapid, Sensitive, Multiplex Loop-Mediated Isothermal Amplification Detection Technique. J. Mol. Diagn. 2015;17:392–401. doi: 10.1016/j.jmoldx.2015.03.002. [DOI] [PubMed] [Google Scholar]
  • 46.Higgins O., Clancy E., Cormican M., Boo T.W., Cunney R., Smith T.J. Evaluation of an Internally Controlled Multiplex Tth Endonuclease Cleavage Loop-Mediated Isothermal Amplification (TEC-LAMP) Assay for the Detection of Bacterial Meningitis Pathogens. Int. J. Mol. Sci. 2018;19:524. doi: 10.3390/ijms19020524. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Hao W., Fan L., Chen Q., Chen X., Zhang S., Lan K., Lu J., Zhang C. Modified Proofreading PCR for Detection of Point Mutations, Insertions and Deletions Using a ddNTP-Blocked Primer. PLoS ONE. 2015;10:e0123468. doi: 10.1371/journal.pone.0123468. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Zhang M., Liu K., Hu Y., Lin Y., Li Y., Zhong P., Jin X., Zhu X., Zhang C. A novel quantitative PCR mediated by high-fidelity DNA polymerase. Sci. Rep. 2017;7:10365. doi: 10.1038/s41598-017-10782-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Liang R., Liang L., Ren X., Jia Y., Han K., Zhao J., Song C., Cui S. Development of a TaqMan loop-mediated isothermal amplification assay for the rapid detection of pigeon paramyxovirus type 1. Arch. Virol. 2021;166:1599–1605. doi: 10.1007/s00705-021-04963-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Thompson D., Lei Y. Mini review: Recent progress in RT-LAMP enabled COVID-19 detection. Sens. Actuators Rep. 2020;2:100017. doi: 10.1016/j.snr.2020.100017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Jenkins D.M., Kubota R., Dong J., Li Y., Higashiguchi D. Handheld device for real-time, quantitative, LAMP-based detection of Salmonella enterica using assimilating probes. Biosens. Bioelectron. 2011;30:255–260. doi: 10.1016/j.bios.2011.09.020. [DOI] [PubMed] [Google Scholar]
  • 52.Lim D.R., Kim H.R., Chae H.G., Ku B.K., Nah J.J., Ryoo S., Wee S.H., Lee C., Lyoo Y.S., Park C.K. Probe-based real-time reverse transcription loop-mediated isothermal amplification (RRT-LAMP) assay for rapid and specific detection of foot-and-mouth disease virus. Transbound. Emerg. Dis. 2020;67:2936–2945. doi: 10.1111/tbed.13669. [DOI] [PubMed] [Google Scholar]
  • 53.Kim H.R., Lim D.R., Chae H.G., Park J.Y., Kim S.H., Lee K.K., Lee C., Lyoo Y.S., Park C.K. Advanced target-specific probe-based real-time loop-mediated isothermal amplification assay for the rapid and specific detection of porcine circovirus 3. Transbound. Emerg. Dis. 2020;67:2336–2344. doi: 10.1111/tbed.13671. [DOI] [PubMed] [Google Scholar]
  • 54.Mashooq M., Kumar D., Niranjan A.K., Agarwal R.K., Rathore R. Development and evaluation of probe based real time loop mediated isothermal amplification for Salmonella: A new tool for DNA quantification. J. Microbiol. Methods. 2016;126:24–29. doi: 10.1016/j.mimet.2016.04.014. [DOI] [PubMed] [Google Scholar]
  • 55.Kumar D., Chauhan T.K., Agarwal R.K., Dhama K., Goswami P.P., Mariappan A.K., Tiwari A.K., Mishra B.P. A double-stranded probe coupled with isothermal amplification for qualitative and quantitative detection of avian reovirus. Arch. Virol. 2017;162:979–985. doi: 10.1007/s00705-016-3184-1. [DOI] [PubMed] [Google Scholar]
  • 56.Nanayakkara I.A., White I.M. Demonstration of a quantitative triplex LAMP assay with an improved probe-based readout for the detection of MRSA. Analyst. 2019;144:3878–3885. doi: 10.1039/C9AN00671K. [DOI] [PubMed] [Google Scholar]
  • 57.Bhat I.A., Mashooq M., Kumar D., Varshney R., Rathore R. Development of probe-based real-time loop-mediated isothermal amplification for detection of Brucella. J. Appl. Microbiol. 2019;126:1332–1339. doi: 10.1111/jam.13938. [DOI] [PubMed] [Google Scholar]
  • 58.Zhang Y., Tanner N.A. Development of multiplexed reverse-transcription loop-mediated isothermal amplification for detection of SARS-CoV-2 and influenza viral RNA. Biotechniques. 2021;70:167–174. doi: 10.2144/btn-2020-0157. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Fan Q., Xie Z., Wei Y., Zhang Y., Xie Z., Xie L., Huang J., Zeng T., Wang S., Luo S., et al. Development of a visual multiplex fluorescent LAMP assay for the detection of foot-and-mouth disease, vesicular stomatitis and bluetongue viruses. PLoS ONE. 2022;17:e0278451. doi: 10.1371/journal.pone.0278451. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Fan Q., Xie Z., Zhang Y., Xie Z., Xie L., Huang J., Zeng T., Wang S., Luo S., Li M. A multiplex fluorescence-based loop-mediated isothermal amplification assay for identifying chicken parvovirus, chicken infectious anaemia virus, and fowl aviadenovirus serotype 4. Avian Pathol. 2023;52:128–136. doi: 10.1080/03079457.2022.2159326. [DOI] [PubMed] [Google Scholar]
  • 61.Nakauchi M., Takayama I., Takahashi H., Semba S., Saito S., Kubo H., Kaida A., Oba K., Nagata S., Odagiri T., et al. Development of real-time fluorescent reverse transcription loop-mediated isothermal amplification assays for rhinovirus detection. J. Med. Virol. 2019;91:1232–1238. doi: 10.1002/jmv.25427. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Hieno A., Li M., Afandi A., Otsubo K., Suga H., Kageyama K. Detection of the Genus Phytophthora and the Species Phytophthora nicotianae by LAMP with a QProbe. Plant Dis. 2020;104:2469–2480. doi: 10.1094/PDIS-12-19-2523-RE. [DOI] [PubMed] [Google Scholar]
  • 63.Sano S., Fukushi S., Yamada S., Harada S., Kinoshita H., Sugimoto S., Yoshikawa T., Kurosu T., Takamatsu Y., Shimojima M., et al. Development of an RT-LAMP Assay for the Rapid Detection of SFTS Virus. Viruses. 2021;13:693. doi: 10.3390/v13040693. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Zhang X., Li H., Liu Z., Zhao Y., Zeng Y., Dong Y., Li L., Zhang C. An HFman probe-based reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay for HIV-1 detection. Mol. Cell. Probes. 2022;64:101834. doi: 10.1016/j.mcp.2022.101834. [DOI] [PubMed] [Google Scholar]
  • 65.Zeng Y., Feng Y., Zhao Y., Zhang X., Yang L., Wang J., Gao Z., Zhang C. An HFman Probe-Based Multiplex Reverse Transcription Loop-Mediated Isothermal Amplification Assay for Simultaneous Detection of Hantaan and Seoul Viruses. Diagnostics. 2022;12:1925. doi: 10.3390/diagnostics12081925. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Zeng Y., Zhao Y., Ren X., Zhou X., Zhang C., Wan Z., Kuang Y.Q. Rapid detection of monkeypox virus and monkey B virus by a multiplex loop-mediated isothermal amplification assay. J. Infect. 2023;86:E114–E116. doi: 10.1016/j.jinf.2023.02.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Zhao Y., Zeng Y., Lu R., Wang Z., Zhang X., Wu N., Zhu T., Wang Y., Zhang C. Rapid point-of-care detection of BK virus in urine by an HFman probe-based loop-mediated isothermal amplification assay and a finger-driven microfluidic chip. PeerJ. 2023;11:e14943. doi: 10.7717/peerj.14943. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Kubota R., Jenkins D.M. Real-time duplex applications of loop-mediated AMPlification (LAMP) by assimilating probes. Int. J. Mol. Sci. 2015;16:4786–4799. doi: 10.3390/ijms16034786. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Ayukawa Y., Hanyuda S., Fujita N., Komatsu K., Arie T. Novel loop-mediated isothermal amplification (LAMP) assay with a universal QProbe can detect SNPs determining races in plant pathogenic fungi. Sci. Rep. 2017;7:4253. doi: 10.1038/s41598-017-04084-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Natoli M.E., Kundrod K.A., Chang M.M., Smith C.A., Paul S., Coole J.B., Butlin N.G., Tanner N.A., Baker E., Schmeler K.M., et al. Improving Performance of a SARS-CoV-2 RT-LAMP Assay for Use With a Portable Isothermal Fluorimeter: Towards a Point-of-Care Molecular Testing Strategy. J. Biomol. Tech. JBT. 2021;32:180–185. doi: 10.7171/jbt.21-3203-013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Higgins O., Chueiri A., O’Connor L., Lahiff S., Burke L., Morris D., Pfeifer N.M., Santamarina B.G., Berens C., Menge C., et al. Portable Differential Detection of CTX-M ESBL Gene Variants, bla(CTX-M-1) and bla(CTX-M-15), from Escherichia coli Isolates and Animal Fecal Samples Using Loop-Primer Endonuclease Cleavage Loop-Mediated Isothermal Amplification. Microbiol. Spectr. 2023;11:e0331622. doi: 10.1128/spectrum.03316-22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72.Crego-Vicente B., Fernandez-Soto P., Garcia-Bernalt Diego J., Febrer-Sendra B., Muro A. Development of a Duplex LAMP Assay with Probe-Based Readout for Simultaneous Real-Time Detection of Schistosoma mansoni and Strongyloides spp.—A Laboratory Approach to Point-of-Care. Int. J. Mol. Sci. 2023;24:893. doi: 10.3390/ijms24010893. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Coole J., Kortum A., Tang Y., Vohra I., Maker Y., Kundrod K., Natoli M., Richards-Kortum R. Open-Source Miniature Fluorimeter to Monitor Real-Time Isothermal Nucleic Acid Amplification Reactions in Resource-Limited Settings. J. Vis. Exp. 2021;168:e62148. doi: 10.3791/62148. [DOI] [PubMed] [Google Scholar]
  • 74.Choi G., Guan W. Biomedical Engineering Technologies. Volume 2393. Humana; New York, NY, USA: 2022. An Ultracompact Real-Time Fluorescence Loop-Mediated Isothermal Amplification (LAMP) Analyzer; pp. 257–278. (Part of the Methods in Molecular Biology Book Series). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Choi G., Guan W. Biomedical Engineering Technologies. Volume 2393. Humana; New York, NY, USA: 2022. Sample-to-Answer Microfluidic Nucleic Acid Testing (NAT) on Lab-on-a-Disc for Malaria Detection at Point of Need; pp. 297–313. (Part of the Methods in Molecular Biology Book Series). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.de Oliveira K.G., Estrela P.F.N., Mendes G.M., Dos Santos C.A., Silveira-Lacerda E.P., Duarte G.R.M. Rapid molecular diagnostics of COVID-19 by RT-LAMP in a centrifugal polystyrene-toner based microdevice with end-point visual detection. Analyst. 2021;146:1178–1187. doi: 10.1039/D0AN02066D. [DOI] [PubMed] [Google Scholar]
  • 77.Choi G., Moehling T.J., Meagher R.J. Advances in RT-LAMP for COVID-19 testing and diagnosis. Expert Rev. Mol. Diagn. 2023;23:9–28. doi: 10.1080/14737159.2023.2169071. [DOI] [PubMed] [Google Scholar]
  • 78.Li Y., Zhou Y., Ma Y., Xu R., Jin X., Zhang C. A Mismatch-tolerant RT-LAMP Method for Molecular Diagnosis of Highly Variable Viruses. Bio Protoc. 2019;9:e3415. doi: 10.21769/BioProtoc.3415. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79.Gadkar V.J., Goldfarb D.M., Gantt S., Tilley P.A.G. Real-time Detection and Monitoring of Loop Mediated Amplification (LAMP) Reaction Using Self-quenching and De-quenching Fluorogenic Probes. Sci. Rep. 2018;8:5548. doi: 10.1038/s41598-018-23930-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Shen H., Wang S., Huang J., Lin Q., Zhang C., Liu Z., Zhang J., Liao M. A Novel, Cleaved Probe-Based Reverse Transcription Loop-Mediated Isothermal Amplification Method for Specific and Sensitive Detection of Porcine Deltacoronavirus. Front. Vet. Sci. 2022;9:896416. doi: 10.3389/fvets.2022.896416. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 81.Bhatt A., Fatima Z., Ruwali M., Misra C.S., Rangu S.S., Rath D., Rattan A., Hameed S. CLEVER assay: A visual and rapid RNA extraction-free detection of SARS-CoV-2 based on CRISPR-Cas integrated RT-LAMP technology. J. Appl. Microbiol. 2022;133:410–421. doi: 10.1111/jam.15571. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 82.Lalli M.A., Langmade J.S., Chen X., Fronick C.C., Sawyer C.S., Burcea L.C., Wilkinson M.N., Fulton R.S., Heinz M., Buchser W.J., et al. Rapid and Extraction-Free Detection of SARS-CoV-2 from Saliva by Colorimetric Reverse-Transcription Loop-Mediated Isothermal Amplification. Clin. Chem. 2021;67:415–424. doi: 10.1093/clinchem/hvaa267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 83.Schellenberg J.J., Ormond M., Keynan Y. Extraction-free RT-LAMP to detect SARS-CoV-2 is less sensitive but highly specific compared to standard RT-PCR in 101 samples. J. Clin. Virol. 2021;136:104764. doi: 10.1016/j.jcv.2021.104764. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Alafeef M., Moitra P., Dighe K., Pan D. RNA-extraction-free nano-amplified colorimetric test for point-of-care clinical diagnosis of COVID-19. Nat. Protoc. 2021;16:3141–3162. doi: 10.1038/s41596-021-00546-w. [DOI] [PubMed] [Google Scholar]