Abstract
Background
Mechanical ventilation is an essential component in the treatment of patients with acute respiratory distress syndrome. Prompt adaptation of the settings of a ventilator to the variable needs of patients is essential to ensure personalised and protective ventilation. Still, it is challenging and time-consuming for the therapist at the bedside. In addition, general implementation barriers hinder the timely incorporation of new evidence from clinical studies into routine clinical practice.
Results
We present a system combing clinical evidence and expert knowledge within a physiological closed-loop control structure for mechanical ventilation. The system includes multiple controllers to support adequate gas exchange while adhering to multiple evidence-based components of lung protective ventilation. We performed a pilot study on three animals with an induced ARDS. The system achieved a time-in-target of over 75 % for all targets and avoided any critical phases of low oxygen saturation, despite provoked disturbances such as disconnections from the ventilator and positional changes of the subject.
Conclusions
The presented system can provide personalised and lung-protective ventilation and reduce clinician workload in clinical practice.
Keywords: Protective ventilation, Acute respiratory distress syndrome, Physiological closed-loop control
Background
Positive pressure mechanical ventilation remains the cornerstone of respiratory support and treatment for patients with acute respiratory distress syndrome (ARDS). Here, the mechanical ventilator supplies supplementary oxygen, allows respiratory rate and ventilation adjustment, and keeps the lung open by applying positive pressures (at inspiration and end-expiration). However, while mechanical ventilation is invaluable for keeping patients alive, it remains a double-edged sword. Namely, it bears the risk of damaging the lung further due to overdistention, cyclic alveolar opening and closing, and inflammation ensuing from mechanical stress and strain of lung tissue. Such injurious ventilation can lead to ventilator-induced lung injury (VILI) [1]. Hence, the application of lung-protective ventilator settings is crucial.
Choosing the correct ventilator settings for personalised and lung-protective ventilation requires experienced personnel with sufficient time to constantly observe and respond to a patient’s changing state. Yet, for many reasons, achieving this degree of therapy with continuous observation and instantaneous response for each patient is rarely possible [2, 3]. In addition, the number of experienced clinicians may be limited, especially in smaller, rural hospitals or during a pandemic like the recent Covid-19 pandemic. Computerised and closed-loop control systems are one possibility to mitigate these challenges [4].
The main therapeutic goals of mechanical ventilation are to provide adequate gas exchange while considering protective limits for pressure and volume. Adequate gas exchange includes oxygenation and alveolar ventilation, which are reflected by arterial partial pressures of oxygen (PaO) and carbon dioxide (PaCO), respectively. In addition, continuously available measurements, including peripheral oxygen saturation (SpO), end-tidal partial pressure of CO (PETCO), as surrogates of PaO and PaCO, and pressure and flow measured at the airway opening (p and ), are being used to guide ventilator settings. Available ventilator settings include the fraction of inspired oxygen (FIO), the positive end-expiratory pressure (PEEP), respiratory rate (), tidal volume (V) and the driving pressure (). Tidal volume is often scaled to predicted body weight (pbw), resulting in V.
Several important clinical trials evaluated the correlation of different ventilator settings on the primary outcome measure of mortality rate (ARMA [5], ALVEOLI [6], LOVS [7], ExPress [8]) and secondary measures, such as ventilator-free days, organ failure-free days and oxygenation. Based on randomised controlled trials and extensive meta-analysis, strong evidence exists for using V for ARDS patients. These recommendations have been included in local, such as the German S3 guidelines on invasive ventilation and use of extracorporeal procedures in acute respiratory insufficiency [9], and international guidelines, such as the Surviving Sepsis Campaign [10] or the ARDSNet protocol [5]. Clear evidence also exists for using a PEEP of at least 5 mbar and even higher PEEP [9–12]. However, no consensus on the optimal method to choose the PEEP exists [13, 14]. Amato et al. found the greatest correlation between the relative risk of death in the hospital and driving pressure () [15], leading to the recommendation of limiting [9]. Application of the lowest necessary FIO to keep patients within the SpO targets is well-accepted. Oxygen toxicity and possible increased mortality with liberal oxygen targeting strategies have also been reported [16].
These recommendations and guidelines provide targets and upper limits for ventilator settings; however, continuous and personalised adherence is rarely possible in the clinical environment for several reasons. Computerised decision support systems (CDSS) are an option to improve adherence and personalise the ventilator settings by recommending ventilator settings to the clinician. The clinician must, however, be present to apply the change. Systems using heuristic rules [17], applying model-based optimisations using physiological models [18, 19], or a combination of both [20] have been presented. A CDSS can provide a deeper physiological representation of a patient’s state and improve the treatment strategy [18]. However, the requirement for a clinician’s presence remains a limitation in reducing the clinician’s workload.
A closed-loop system, which automatically, i.e., without needing a clinician’s presence, adjusts settings based on targets, feedback of measured values and a control algorithm, would be advantageous in this case. A CDSS or clinical protocol can be programmed to run in a closed-loop form. Examples would be the automatic ARDSNet protocol system [21] or the automation of the open lung concept [22]. Automatic closed-loop control are often focused on either oxygenation [23, 24] or ventilation (PETCO) [25–27]. Highly automated systems which combine oxygen and carbon dioxide controllers have also been presented, see for example [28], or the commercially available system INTELLiVENT®-ASV (Hamilton Medical AG, Switzerland) [29]. We recently reviewed currently available physiological closed-loop control (PCLC) systems [4]. Lung-protective ventilation, as described above, typically played a secondary role for many of these systems compared to achieving the physiological setpoints. The clinical evidence is often more recent than the above-mentioned algorithms and, therefore, not included in their design.
In this paper, we present our System for automatic Lung-protective Ventilation (SOLVe) with the aim to couple evidence-based protective guidelines with closed-loop control of mechanical ventilation. The system has defined protective operating ranges for ventilator settings, including adaptive limits, uses multiple closed-loop controllers and incorporates clinical knowledge into the controllers.
Results
In this pilot study, the SOLVe system was evaluated in three animal subjects to demonstrate algorithmic performance. Table 1 shows the details of each experiment, during which the automated system ran for 6 h.
Table 1.
Subject | Respiratory failure model type | Number of disconnection | Tilting | Number of PEEP titrations | Compromise operating range allowed? |
---|---|---|---|---|---|
A | II | 0 | No | 2 | No |
B | I | 4 | Yes | 2 | No |
C | II | 2 | Yes | 1 | Yes |
The percentage of time-in-target for each of the goals is shown in Fig. 1. The SpO target was met for almost 80 % of the time in all three subjects. The PETCO was met for subject C for more than 95 % of the time, while subjects A and B spent considerably less time within the target region. Subjects A and B met the protective target region for over 80 % of the time, while subject C was in the protective compromise region for almost 80 % of the time.
Subject A had a of (mean ± standard deviation), which meant it was on the edge of the possible protective target region. This limitation on the minute ventilation made it impossible for the PETCO target to be met but validated the hierarchy defined in the controller structure that limitation trumps PETCO. The of subject B was the largest at , which allowed SOLVe to spend time in the acceptable target region and therefore achieve the PETCO target for more than 60 % while staying within the protective ranges for V and . For subject C, the was , but the compromise target region was activated at the clinicians’ discretion. The PETCO target was met but with a more aggressive range.
The effect of tilting the subjects is shown for the exemplary case of subject B in Fig. 2. Tilting the subject up caused the to decrease significantly. Tilting the subject’s head down resulted in a large increase in and subsequently returning to the normal position increased even further.
Automated PEEP titration
An automated PEEP titration was initiated at the beginning of each experiment. For subjects A and B, a second PEEP titration was initiated by the clinician after 4 h of closed-loop mechanical ventilation. In all cases, the best PEEP was identified automatically, and no clinician interaction was required. The change in compliance between before and after the automated PEEP titration is given in Table 2. After the titration, the identified best PEEP was higher for two trials, lower for two trials and the same for one trial compared to before the titration. In all cases, the compliance was the same or higher after the PEEP titration.
Table 2.
Subject | Trial | Before titration | After PEEP titration | Difference | |||
---|---|---|---|---|---|---|---|
PEEP | PEEP | PEEP | |||||
[mbar] | [ml/mbar] | [mbar] | [ml/mbar] | [mbar] | [%] | ||
A | 1 | 15 | 15 | 16 | 15 | +1 | 0.0 |
A | 2 | 18 | 13 | 14 | 16 | – 4 | 23.1 |
B | 1 | 12 | 13 | 14 | 16 | +2 | 23.1 |
B | 2 | 14 | 17 | 14 | 24 | 0 | 41.2 |
C | 1 | 15 | 17 | 14 | 18 | – 1 | 5.9 |
SpO control results
In addition to the time-in-target results, the dynamic response of the SpO was evaluated. As an illustrative example, Fig. 3 shows a section where SpO controller was particularly active in subject A. After the initial phase, where the FIO was titrated down from 1.0, the base values were found. Small FIO increases occurred continuously to keep the SpO within the target range. At the FIO reached the base limit and the controller moved to the second region. This also included a PEEP increase, which stabilised the SpO within the target region again. The new FIO limit is also shown.
Brief disconnections (15 s) between the subject and the mechanical ventilator were performed and three such cases are shown in Fig. 4.
In each case, the disconnection causes a significant decrease in SpO. For case (a), the lowest SpO value was 85.6 % and the SpO was below the target for 100 s. In case (b), no intervention of the SpO controller was needed because the SpO recovered in 10 s after reconnection. The SpO subsequently overshot the target, to which the SpO controller responded by decreasing the FIO. Case (c) represents the worst-case, as the SpO drops below the critical value of 85 %. Here, the controller correctly responded by setting the FIO to 1.0. The lowest SpO value was 63.4 %, the SpO was below the critical value for 30 s and below the target for 40 s.
PETCO control and PRVC results
For illustration, the dynamic responses of the cascaded PETCO and PRVC control are shown for selected situations in Fig. 5.
The clinician adapted the PETCO target range at for subject B. Since the PETCO was below the new target range and the current acceptable operating range was active, SOLVe automatically decreased the V first. The PRVC controller adapted such that the new V was achieved, thereby also moving into the protective target region. The was also automatically decreased since the PETCO was still below the target even after the V change. This brought the PETCO into the target range.
For subject C, the PETCO exceeded the target range after , to which the controller responded by increasing the rate twice. Subsequently, the V dropped below 4 ml/kg which alerted the clinician to consider activating the compromise region. After the confirmation, the PRVC controller increased the to ensure V= 4 ml/kg. At , the PETCO again left the target range, but the rate was already at the maximum. Therefore, the PETCO controller increased the V to 5 ml/kg.
Discussion
We designed a closed-loop control expert system that automatically adapts all ventilator settings to achieve the SpO, PETCO, and lung protective targets recommended for mechanical ventilation in ARDS patients. A pilot study for the SOLVe system showcases its accuracy, robustness and reliability. The system achieved especially good time-in-target results for the SpO control and avoided any critical phases of de-saturation despite provoked disturbances. The recommended protective limits were adhered to at all times, and, if possible, the PETCO was also regulated. As such, the system can reliably help the clinician keep mechanical ventilation safe and could drastically reduce the workload of clinical staff responsible for observing and adjusting the mechanical ventilator settings. In each case, the ventilator settings were also highly personalised and continuously adapted to respond to variable disease states and conditions and to the various artificial disturbances.
Automated PEEP titrations
Decremental PEEP titrations finding the maximum compliance, as applied by the SOLVe system, are only one of many methods proposed for optimising PEEP. However, no consensus on the optimal method currently exists [13, 14]. An early landmark study by Suter et al. proposed titrating PEEP so that oxygen delivery is maximised [30]. The calculation of oxygen delivery, however, requires invasive measurement of cardiac output and oxygen tension. Respiratory system compliance was proposed as a substitute, since it correlated well with oxygen delivery. Suarez-Sipmann et al. showed that compliance could be used to identify the beginning of collapse after recruitment and confirmed these findings with oxygenation and computed tomography scans [31]. Therefore, the chosen approach for the SOLVe system aligns with the current clinical practice. However, unlike current manual PEEP titrations, which often go way past the best PEEP to identify it, the SOLVe system stops automatically at the best PEEP. Therefore, no additional recruitment manoeuvres are necessary after the PEEP titration, as was done in a large clinical trial [32]. As a secondary benefit, automatically titrating PEEP to maximise enlarges the operating region for V and and remains invaluable for the subsequent protective control.
SpO control
The SpO controller was shown to be dynamic, safe and highly personalised and achieved a time-in-target of over 80 % and avoided critical de-saturation events.
Recent publications [16, 33] highlight the importance of oxygen control and avoiding both hyperoxemia and hypoxemia. Excessive oxygen over long periods can be toxic, and links between liberal FIO and increased mortality have been presented [16]. Hypoxemia can lead to irreversible tissue damage, neurological damage or death and must be avoided at all costs [33].
Achieving a tight oxygen control requires many adaptions of FIO, as shown in Fig. 3, which would place a high burden on clinical staff. The presented SpO controller performed this task reliably and robustly, even during disturbances. Different sample times for phases above and below the target and small changes of FIO in steps of 0.02 allowed for fast and dynamic yet robust feedback control of SpO. The small FIO changes differ from the ARDSNet table recommendation of steps of 0.1. These large FIO changes could lead to oscillatory behaviour around the target—or having to accept remaining above the target with excessive FIO.
The approach to defining controller states for coupling the FIO and PEEP incorporates important clinical knowledge into the controller. Previous computerised systems have also included a coupling between these variables, such as evaluating the PEEP/FIO ratio to determine if PEEP changes should occur [34] or using the ARDSNet protocol as guidance (INTELLiVENT®-ASV) [29]. Importantly, however, the SOLVe system bases the level of PEEP first on optimising compliance and second on oxygenation. Furthermore, the pairing of FIO and PEEP is personalised by automatically finding a suitable base pairing of FIO and PEEP. No limits exist for this initial base pairing, meaning it is individualised to fit the diverse patient and disease populations in the clinical environment. All further pairings are then based on this initial base pairing.
PETCO and PRVC control
The default target for the PETCO was chosen to be 30 mmHg to 60 mmHg, which is within the limits given in German S3 guidelines for permissive hypercapnia [9]. This target was only partially achievable for subjects A and B without infringing on the lung-protective limits of . Once the target for PETCO could not be met without infringing the limit, the clinician was alerted of the conflict. For subjects A and B, the clinician decided not to increase the PRVC controller’s operating region leading to PETCO greater than 60 mmHg and hypercapnia outside the target range. This was tolerated in the animal experiments at the discretion of the attending clinician, and there was no hard cut-off for PETCO in this pilot study. Instead, the goal was to test whether the higher-priority lung-protective ventilator settings were adhered to. This limitation within the algorithm was a core consideration during the design of the rule-based control law. In a comparable clinical situation, the clinical team would evaluate other therapeutic options, such as prone positioning, recruitment manoeuvres, or extra-corporeal CO removal, before increasing the aggressiveness/invasiveness of mechanical ventilation.
In any case, the SOLVe system always ensures that the applied is within the protective limits and only goes beyond this limit after clinical confirmation—thus allowing the clinician to consider other therapeutic options first. Through visual messages, the clinician is aware of this conflict and can make better decisions. Hence, the SOLVe system can be used as an automated tool to assist the clinician in providing protective ventilation and reaching the therapeutic targets without the automation interfering with the clinical decisions.
The PETCO control in the SOLVe system is based on simple rules and clinical expertise. Nonetheless, the controller held the PETCO within the target most of the time without infringing on protective limits. Importantly, this highlights the hierarchy that is the most critical parameter, and the simplicity of the rules makes the controller more transparent and understandable for clinical staff.
We note that PETCO is used as a surrogate for PaCO which would be the ideal measurement for CO control. However, PaCO is only available with discrete invasive blood gas analysis. While the difference between PETCO and PaCO is non-linear and time-varying, studies have shown that PETCO and PaCO correlate to a certain degree [35]. Our application used every available PaCO from an hourly arterial blood gas analysis to confirm that the current PETCO target range was acceptable. Changes to the PETCO target could have been made if the discrepancy became too large. Adjusting alveolar ventilation to the arterial pH, as recommended by the ARDSNet protocol, was deemed impractical due to the discontinuity of the parameter and the lack of an adequate surrogate.
The upper limit of 45 breaths per minute for the respiration rate set by the clinicians at the beginning of all experiments is above the common 35 breaths per minute [5]. This value was allowed to be higher due to the experiments being performed on adolescent pigs with a healthy resting breathing rate of 37±12 breaths per minute [36]. This limit would be set for future clinical applications to a value appropriate for the patient. Due to the stiff respiratory mechanics (small ), the time constant of the respiratory system () was never greater than 0.22 s, and the upper limit for remained at 45 for all experiments. All subjects had complete expiration, and no build-up of intrinsic PEEP was observed.
An active research area is the mechanical power and respiratory work applied during mechanical ventilation. A system by Becher et al. automatically determined the respiratory rate and tidal volume to minimise mechanical power [37]. However, the clinician had to choose the required minute ventilation in their system. The ASV mode (Hamilton Medical AG, Switzerland) and the FLEX system [20] continuously update the breathing rate to minimize the respiratory work to expedite weaning. Van der Staay et al. have recently provided a review of these concepts [38]. In the future, the PRVC and PETCO controllers of the SOLVe system could be extended to optimise for minimum mechanical power.
Limitations
The experimental validation has the following limitations. Firstly, the system was only tested in a pilot study of three animal subjects. An experimental ARDS was induced in the animals, mimicking ARDS’s lung mechanical and functional characteristics [39]. Nonetheless, respiratory system compliance was significantly lower than reported in most literature on ARDS patients. This mostly led to the subjects being ventilated with very low tidal volumes (4 ml/kg) and in the range of 13 mbar to 20 mbar. Additionally, the respiratory frequency was also in the upper region of to keep the subjects within the PETCO target for as long as possible. All animals required low to moderate FIO levels to correct the hypoxaemia. Despite these challenges, the SOLVe system was still able to ventilate the subjects protectively and safely. The system is expected to act equally well in subjects with larger respiratory system compliance and more severe hypoxaemia.
Conclusion
The SOLVe system is a closed-loop control system which automatically adapts all relevant ventilator settings to achieve the SpO, PETCO, and protective targets recommended for mechanical ventilation in ARDS patients. An automatic PEEP titration algorithm finds the best PEEP and optimises the operating region for the subsequent protective ventilation. The system was evaluated in vivo and was shown to be safe and reliable. It is predicted that the system would (1) lead to a better application of the current clinical evidence for mechanical ventilation, (2) provide optimal adherence to current guidelines on protective ventilation, (3) provide a personalised therapy that is continuously adapted to the patient’s needs and (4) reduce clinicians’ workload to implement their therapeutic strategy for each patient.
Materials and methods
SOLVe addresses the two main targets of mechanical ventilation strategies: gas exchange and protective ventilation. Since this is a multiple-input and multiple-output problem, four controllers and various continuous physiological measurements are included in the SOLVe system. An estimator provides online parameter estimates for a respiratory mechanics model (resistance and compliance ). A block diagram representation of the complete system is given in Fig. 6, showing the controllers, sensors and actuator, and the bi-directional user interface with the clinician. Table 3 shows the pairing of target, measurement and actuation.
Table 3.
Target | Measurement | Controller | Actuation |
---|---|---|---|
Oxygenation | SpO | SpO controller | FIO, PEEP |
Ventilation | PETCO | PETCO controller | V, |
Protective volume and pressure | V, p, PEEP | Pressure-regulated volume control | , V |
Best PEEP | , p, | Automatic PEEP titration | PEEP |
The phases of operation of the SOLVe system and the interaction with the clinician are given in Fig. 7. Importantly, interaction with the clinician remains vital for the SOLVe system, even if the goal is to reduce this to a minimum. The clinician initially inputs targets and limits and observes the PEEP titration. During the closed-loop control, the clinician does not have to be present. Three controllers run in parallel at different update frequencies during this phase, as described later. The SOLVe system has a set of degrees and limits of freedom while keeping ventilation protective; if the system cannot achieve adequate gas exchange and ventilation within protective limits, the system alerts the clinician. They may be able to increase the operating space of SOLVe, or SOLVe may have reached a limit, at which point the clinician takes over control.
System setup
Our prototype system consists of a real-time PC (MicroLabBox, dSPACE GmbH, Paderborn, Germany) connected to a medical panel PC (, MCD Medical Computers Deutschland GmbH, Mönchengladbach, Germany) running MATLAB 2017b (The MathWorks Inc., Natick, USA) and dSPACE Control Desk version 7.1 (dSPACE GmbH, Paderborn, Germany). A modified mechanical ventilator (EVE, Fritz Stephan GmbH, Gackenbach, Germany) receives remote commands and sends all measurement data via a custom RS232 protocol to the real-time PC. The ventilator features a built-in pulse oximeter (MASIMO Rainbow, Irvine, USA) and a proximal mainstream capnograph (MASIMO IRMA CO2, Irvine, USA). All ventilation data (pressure, flow, and ventilation phase) are sampled at 100,Hz, while the other variables, such as SpO, pulse rate, and PETCO are sampled at 1 Hz. All data processing and control algorithms are executed on the real-time PC at a sampling rate of 100 Hz.
Defining protective operating ranges
Achieving both protective ventilator settings and sufficient gas exchange may not be possible for all critically ill patients with ARDS. Therefore, operating ranges defining the limits and permitted degrees of freedom for the ventilator settings are mandatory and have been included in the SOLVe system.
SOLVe cannot replace the standard multimodal therapy approach for patients with ARDS, which, e.g., includes antibiotic treatment and fluid management. Other supplementary therapeutic options beyond ventilation, such as prone positioning or extra-corporeal membrane oxygenation, should always be considered according to the patient’s status, especially if a patient does not fall within the SOLVe operating range.
A first operating range is based on V and and divided into three regions: target, acceptable and compromise as shown in Fig. 8.
The target region spans the area defined by clinical recommendations for tidal volume and driving pressure, V and [9, 10], respectively. A lower limit of V ensures minimum alveolar ventilation of the patient, and is a ventilator device limitation. To prevent consequences of severe hypercapnia (excessive CO), the acceptable region allows for larger yet safe tidal volumes of up to 8 ml/kg and still strictly limits the . The compromise region allows to facilitate minimum alveolar ventilation even for patients with very stiff respiratory mechanics. However, unlike the target and acceptable region, explicit activation is required by the attending clinician since it falls outside of clinical recommendations [9, 15].
SOLVe applies a fast method to determine the probable operating point for each patient and situation by considering the respiratory mechanics. The passive respiratory system (without spontaneous breathing) can be modelled as a linear first-order system given by the equation of motion:
1 |
where is the airway pressure, is the airway flow, is the volume, is the respiratory resistance, the respiratory compliance and is the total pressure at end expiration.
and V are proportionally linked by respiratory compliance since , if airway flow returns to zero at the end of inspiration. Since tidal volume is normally scaled to the patient’s predicted body weight, compliance can also be scaled to the predicted body weight (). This linear relationship is also plotted in Fig. 8 and shows the minimum or maximum span for the operating regions outlined above. Patients with a cannot be ventilated within the defined regions and do not qualify for application of the SOLVe system, and other therapeutic options should be considered.
The operating range for the respiratory rate is defined to prevent a build-up of intrinsic PEEP and possible barotrauma. Initially, the clinician defines an upper and lower limit for . Thereafter, the SOLVe system observes the time constant of the respiratory system () and reduces the upper limit () to prevent intrinsic PEEP. Assuming an exponential pressure decay during the passive expiration, the computed is used to estimate the required expiration time. An exponential decay is almost complete after a time of , meaning complete expiration has occurred and the expiration flow is very close to zero. Keeping the inspiration to expiration ratio constant at 1:1, as recommended by [9], expiration must be complete within half a breath period. Therefore, an upper limit for can be calculated by:
2 |
Furthermore, automatic checks for the build-up of intrinsic PEEP use are included. First, intrinsic PEEP is identified using the online estimation of the value (see eq. 1) according to the method by Nucci et al. [40]. Second, complete expiration was observed by ensuring the expiratory flow was zero before the next breath started. If either method detects an intrinsic PEEP, the system automatically reduces the by 5 bpm and alerts the clinician.
The respiratory system parameters ( and ) and total pressure at end expiration () are calculated on a breath-by-breath basis using the equation of motion (eq. 1) by means of least squares estimation [41]. The linear first-order model was robust and sufficiently accurate for our application.
Automated PEEP titration
Different methods of finding an adequate PEEP level have been proposed without consensus on the optimal method [13]. One such method is finding the PEEP at which the is greatest [13]. Setting adequate PEEP to avoid derecruitment can significantly improve and maintain the compliance of the lungs. The procedure for finding the maximum compliance involves first increasing the PEEP level to a certain value (e.g., 24 mbar) and then step-wise reducing (titrating) the PEEP and observing the compliance at every level. This procedure is graphically shown in Fig. 9. Importantly, decremental instead of incremental titrations should be used due to the hysteresis of the lungs [42].
This procedure was automated in the SOLVe system. To start the procedure and avoid a single large step, the system automatically increases the PEEP in steps of 4 mbar until the starting pressure 24 mbar is reached. Each PEEP level is held for 2 min, while the median is calculated and evaluated. A logistic regression classifier evaluates whether the current PEEP level is the best. It has previously been presented in detail [43]. Briefly, the relative change in compared to the previous PEEP level is calculated and classified into one of three classes:
3 |
If the current PEEP level is classified as above best PEEP, the PEEP is reduced by 2 mbar and is re-evaluated. This procedure is repeated until either best PEEP or below best PEEP are found and the titration procedure ends. If best PEEP is reached, the relative compliance change is small enough, meaning the peak of the PEEP versus curve has probably been reached. Reaching the below best PEEP means the compliance has decreased and the best PEEP was on the previous level. PEEP is increased to the previous level before the titration ends.
The titration is performed in a pressure-controlled mode with a of 14 mbar. A volume guarantee supervisor ensures that the V remains between 4 ml/kg and 8 ml/kg, and can vary the if needed. Finally, a lower limit of 6 mbar for the PEEP exists, as recommended in [9]. To prevent hypoxemia, FIO is kept at 1.0 during the PEEP titration. In the case of a rapidly falling SpO, a safety routine is initiated to stabilise the patient.
SpO controller
The SpO controller utilises both FIO adjustments and PEEP increases to achieve the SpO target, which the clinician provides as a range between target high and target low.
Before starting the closed-loop control, an automatic initialisation phase finds a base pairing of PEEP and FIO. The PEEP is found using the automated PEEP titration to maximise compliance. Following this, the FIO is decreased from 1.0 until the SpO target is reached, as shown in Fig. 10.
Subsequently, automatic adjustments in FIO depend on measured SpO as shown in Fig. 10 and Table 4. Small FIO adjustments ensure the target can be reached without excessive overshoot. While above the target, the reduced update rate (100 s) prevents limit cycles or undershooting the target. Below the target, a faster response is required; hence a faster update rate of 20 s is used. If the SpO falls below the lower limit (85%), the FIO is immediately set to 1.0 to prevent hypoxemia. Once a patient has remained stable within the upper target range for a predefined period (20 min for our case), with no increases in FIO or PEEP, the expert system slowly decreases the FIO until the patient is at the lower end of the target. Thereby, the FIO is reduced to a minimum, preventing risks of an unnecessarily high FIO.
Table 4.
SpO | Description | Update rule | Update time |
---|---|---|---|
> high target + 4 | Greatly above target | Decrease FIO by 0.1 | 100 s |
> high target + 2 | Above target | Decrease FIO by 0.05 | 100 s |
> high target | Slightly above target | Decrease FIO by 0.02 | 100 s |
> target middle | Upper target region | If stable: decr. 0.02, else: no change | 20 min |
> low target | Lower target region | No change | – |
< low target | Slightly below target | Increase FIO by 0.02 | 40 s |
< low target -2 | Below target | Increase FIO by 0.05 | 20 s |
85 | SpO low state | Set FIO to 1.0 | Immediately |
In addition to increasing FIO, increasing PEEP is a valid therapeutic measure if the SpO becomes low. A higher PEEP increases the lung’s functional residual capacity and can prevent alveolar derecruitment, reduce atelectasis and improve gas exchange.
Therefore, if a significant decline in the gas exchange capability of the lungs has occurred since the base pairing of PEEP and FIO was found, the current PEEP may be insufficient. To avoid conflicts between the best PEEP based on maximum compliance (automated PEEP titration above) and increasing PEEP based on oxygenation, a set of conditions for PEEP increases were designed:
No increase in PEEP for the first 60 min after a PEEP titration manoeuvre.
PEEP can only increase in steps of 2 mbar.
Only two increases are allowed per 24 h interval.
There should be 5 min between subsequent PEEP increases to allow for the change to have an effect [44].
The coupling and the evaluation of whether a PEEP increase should occur are also included in the flowchart in Fig. 10. If the above conditions are not satisfied, the SOLVe algorithm has reached a limit and issues an alarm. Clinical staff must evaluate the patient’s condition at the bedside and take appropriate therapeutic actions.
For better visualisation to the user, the three operating states of the controller are given names as shown in Table 5. The automatic change between states is sequential, i.e. base second final, and non-reversible. The system does not automatically move down to a lower state because PEEP reductions bear the risk of alveolar derecruitment and deterioration of gas exchange. Therefore, a decision to reduce PEEP requires the clinician’s attention.
Table 5.
State | Upper FIO limit | PEEP |
---|---|---|
Base | FIO | PEEP |
Second | FIO | PEEP |
Final | FIO | PEEP |
Movement between states is sequential and non-reversible
Pressure-regulated volume control and PETCO control
A cascaded control loop is used for the control of PETCO and V as shown in Fig. 11. The clinician inputs a target high and target low to define a PETCO target range. Note that the V is calculated automatically by the system and cannot be adapted by the clinician.
The inner loop contains the pressure-regulated volume control, which adjusts to achieve the V. Here, a proportional-integral (PI) controller with feed-forward is used. The inverse of the estimated respiratory system model is used for the feed-forward element, while the PI controller eliminates any steady-state error. The is updated on a breath-by-breath basis and limited to a maximum change of 3 mbar per breath.
The outer loop regulates the PETCO by adapting the and V. The actual controller is rule-based and contains the rules as shown in Fig. 12. This outer loop runs with a sampling time of 1 min. The upper limit for the respiration rate () is defined based on the operating range explained in the earlier section.
In vivo experiments
A pilot study used the SOLVe system to ventilate adolescent pigs with respiratory failure (German landrace, male, approx. 40 kg). The study was approved by the Institutional Animal Care and Use Committee (Tierversuchskommission, Landesamt für Gesundheit und Soziales, Berlin, Germany; approval number G 0229/18), and all animal procedures were conducted complying with national regulations and institutional animal care committee guidelines.
All animals were placed in the horizontal supine position at the beginning of the experiment. They were put under general anaesthesia and instrumented before a baseline phase was started. Two different models for respiratory failure induction were used [39]. For model I, surfactant depletion by saline-based lung lavages was used. For model II the combination of surfactant depletion and injurious ventilation was used. The Berlin definition of severe ARDS, i.e. a PaO/FIO 100 mmHg with PEEP 5 cmHO, was used to quantify the respiratory failure before starting the SOLVe system. First, the clinical targets and limits of ventilation were set according to clinical reasoning. Then, 6 h of automated ventilation using SOLVe were started. The animals remained deeply sedated throughout the experiment and had no spontaneous breathing.
The goal for the system was to remain within target ranges for SpO, PETCO and protective ventilation for as long as possible without any interaction from the clinician. The default targets for the system are given in Table 6. The SpO target was also considered met if the SpO was above the target, but the FIO was at the lower limit of 0.3. Furthermore, it was considered critical if the SpO dropped below 85 %. The targets were adapted during the experiments to test the controllers.
Table 6.
Target | Unit | Target low | Target high | |
---|---|---|---|---|
SpO | % | 91 | 93 | |
PETCO | mmHg | 30 | 60 | |
Protective target region | V | ml/kg | 4 | 6 |
mbar | 5 | 14 | ||
Protective acceptable region | V | ml/kg | 6 | 8 |
mbar | 5 | 14 | ||
Protective compromise region | V | ml/kg | 4 | 6 |
mbar | 14 | 20 |
To test the robustness of the SOLVe system, two kinds of disturbances were introduced during the experiment. First, a brief disconnection (15 s) from the ventilator, simulating routine mucus clearance or an accidental disconnection, causes a rapid de-saturation and lung collapse. Second, the subjects were tilted away from the horizontal position (head-up tilting of and a head-down tilting of ). This manoeuvre provokes complex changes to blood pressure, ventilation-to-perfusion matching within the lung and large changes in respiratory compliance. A graphical representation is shown in Fig. 13.
Abbreviations
- ARDS
Acute respiratory distress syndrome
- bpm
Breaths per minute
Respiratory system compliance
Respiratory system compliance scaled to predicted body weight
- CDSS
Computerised decision support system
- FIO
Fraction inspired oxygen
Respiratory rate
- PCLC
Physiological closed-loop control
- PEEP
Positive end-expiratory pressure
- PETCO
End-tidal partial pressure of carbon dioxide
- pbw
Predicted body weight
- PRVC
Pressure regulated volume control
Respiratory system resistance
- SpO
Oxygen saturation measured using pulse oximetry
- SOLVe
System for automatic Lung-protective Ventilation
- VILI
Ventilator-induced lung injury
- V
Tidal volume
- V
Tidal volume scaled to predicted body weight
Driving pressure
Author contributions
Algorithm design: PvP, AP, WB, LH, RK, PAP, MR, RCEF. Animal experiments and data acquisition: MR, MT, PAP, LH, PvP. Analysis and interpretation of data: PvP, AP, RCEF, MW. Drafting of the manuscript: PvP, AP, RCEF, MW. Critical revision: WB, RK, SL. All authors read and approved the final manuscript.
Funding
Open Access funding enabled and organized by Projekt DEAL. This work has been funded by the Federal Ministry of Education and Research (BMBF, Germany) and is part of the project SOLVe with grant number 13GW0240.
Availability of data and materials
The data sets used and analysed during the current study are available from the corresponding author on reasonable request.
Declarations
Ethical approval and consent to participate
All experiments were in accordance with the guidelines of the European and German Society of Laboratory Animal Science and were approved by the state authority for the care and use of animals (LAGeSo, Berlin, Germany; Approval G0229/18).
Consent for publication
Not applicable
Competing interests
RK is an employee of EKU Elektronik GmbH, Germany. LH and WB are employees of Fritz Stephan GmbH, Germany. Otherwise, the authors have no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Philip von Platen, Email: platen@hia.rwth-aachen.de.
Anake Pomprapa, Email: pomprapa@hia.rwth-aachen.de.
Steffen Leonhardt, Email: leonhardt@hia.rwth-aachen.de.
Marian Walter, Email: walter@hia.rwth-aachen.de.
References
- 1.Slutsky AS, Ranieri VM. Ventilator-induced lung injury. N Engl J Med. 2013;369:2126–2136. doi: 10.1056/NEJMra1208707. [DOI] [PubMed] [Google Scholar]
- 2.Rubenfeld GD, Cooper C, Carter G, Thompson BT, Hudson LD. Barriers to providing lung-protective ventilation to patients with acute lung injury. Crit Care Med. 2004;32:1289–1293. doi: 10.1097/01.CCM.0000127266.39560.96. [DOI] [PubMed] [Google Scholar]
- 3.Kaier K, Heister T, Wolff J, Wolkewitz M. Mechanical ventilation and the daily cost of ICU care. BMC Health Serv Res. 2020;20:267. doi: 10.1186/s12913-020-05133-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.von Platen P, Pomprapa A, Lachmann B, Leonhardt S. The dawn of physiological closed-loop ventilation–a review. Crit Care. 2020;24:121. doi: 10.1186/s13054-020-2810-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Brower RG, Matthay MA, Morris A, Schoenfeld D, Thompson BT, Wheeler A. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med. 2000;342:1301–1308. doi: 10.1056/NEJM200005043421801. [DOI] [PubMed] [Google Scholar]
- 6.Brower RG, Lanken PN, MacIntyre N, Matthay MA, Morris A, Ancukiewicz M, et al. Higher versus lower positive end-expiratory pressures in patients with the acute respiratory distress syndrome. N Engl J Med. 2004;351:327–336. doi: 10.1056/NEJMoa032193. [DOI] [PubMed] [Google Scholar]
- 7.Meade MO, Cook DJ, Guyatt GH, Slutsky AS, Arabi YM, Cooper DJ, et al. Ventilation strategy using low tidal volumes, recruitment maneuvers, and high positive end-expiratory pressure for acute lung injury and acute respiratory distress syndrome: a randomized controlled trial. JAMA. 2008;299:637. doi: 10.1001/jama.299.6.637. [DOI] [PubMed] [Google Scholar]
- 8.Mercat A, Richard JCM, Vielle B, Jaber S, Osman D, Diehl JL, et al. Positive end-expiratory pressure setting in adults with acute lung injury and acute respiratory distress syndrome: a randomized controlled trial. JAMA. 2008;299:646. doi: 10.1001/jama.299.6.646. [DOI] [PubMed] [Google Scholar]
- 9.Deutsche Gesellschaft für Anästhesiologie und Intensivmedizin (DGAI). S3-Leitlinie: Invasive Beatmung und Einsatz extrakorporaler Verfahren bei akuter respiratorischer Insuffizienz; 2017. https://www.awmf.org/leitlinien/detail/ll/001-021.html.
- 10.Rhodes A, Evans LE, Alhazzani W, Levy MM, Antonelli M, Ferrer R, et al. Surviving sepsis campaign: international guidelines for management of sepsis and septic shock: 2016. Intensive Care Med. 2017;43:304–377. doi: 10.1007/s00134-017-4683-6. [DOI] [PubMed] [Google Scholar]
- 11.Putensen C. Meta-analysis: ventilation strategies and outcomes of the acute respiratory distress syndrome and acute lung injury. Ann Intern Med. 2009;151:566. doi: 10.7326/0003-4819-151-8-200910200-00011. [DOI] [PubMed] [Google Scholar]
- 12.Oba Y, Thameem DM, Zaza T. High levels of PEEP may improve survival in acute respiratory distress syndrome: a meta-analysis. Respir Med. 2009;103:1174–1181. doi: 10.1016/j.rmed.2009.02.008. [DOI] [PubMed] [Google Scholar]
- 13.Hess DR. Recruitment maneuvers and PEEP titration. Respir Care. 2015;60:1688–1704. doi: 10.4187/respcare.04409. [DOI] [PubMed] [Google Scholar]
- 14.Henderson WR, Chen L, Amato MBP, Brochard LJ. Fifty years of research in ARDS. Respiratory mechanics in acute respiratory distress syndrome. Am J Respir Crit Care Med. 2017;196:822–833. doi: 10.1164/rccm.201612-2495CI. [DOI] [PubMed] [Google Scholar]
- 15.Amato MBP, Meade MO, Slutsky AS, Brochard L, Costa ELV, Schoenfeld DA, et al. Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med. 2015;372:747–755. doi: 10.1056/NEJMsa1410639. [DOI] [PubMed] [Google Scholar]
- 16.Cumpstey AF, Oldman AH, Martin DS, Smith A, Grocott MPW. Oxygen targets during mechanical ventilation in the ICU: a systematic review and meta-analysis. Crit Care Explor. 2022;4:e0652. doi: 10.1097/CCE.0000000000000652. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Kwok HF, Linkens DA, Mahfouf M, Mills GH. Rule-base derivation for intensive care ventilator control using ANFIS. Artif Intell Med. 2003;29:185–201. doi: 10.1016/S0933-3657(02)00074-X. [DOI] [PubMed] [Google Scholar]
- 18.Rees SE, Allerød C, Murley D, Zhao Y, Smith BW, Kjaergaard S, et al. Using physiological models and decision theory for selecting appropriate ventilator settings. J Clin Monit Comput. 2006;20:421–429. doi: 10.1007/s10877-006-9049-5. [DOI] [PubMed] [Google Scholar]
- 19.Tehrani FT, Abbasi S. A model-based decision support system for critiquing mechanical ventilation treatments. J Clin Monit Comput. 2012;26:207–215. doi: 10.1007/s10877-012-9362-0. [DOI] [PubMed] [Google Scholar]
- 20.Tehrani FT, Roum JH. Flex: a new computerized system for mechanical ventilation. J Clin Monit Comput. 2008;22:121–130. doi: 10.1007/s10877-008-9113-4. [DOI] [PubMed] [Google Scholar]
- 21.Pomprapa A, Schwaiberger D, Pickerodt P, Tjarks O, Lachmann B, Leonhardt S. Automatic protective ventilation using the ARDSNet protocol with the additional monitoring of electrical impedance tomography. Crit Care. 2014;18:R128. doi: 10.1186/cc13937. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Schwaiberger D, Pickerodt PA, Pomprapa A, Tjarks O, Kork F, Boemke W, et al. Closed-loop mechanical ventilation for lung injury: a novel physiological-feedback mode following the principles of the open lung concept. J Clin Monit Comput. 2018;32:493–502. doi: 10.1007/s10877-017-0040-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Tehrani F, Rogers M, Lo T, Malinowski T, Afuwape S, Lum M, et al. Closed-loop control if the inspired fraction of oxygen in mechanical ventilation. J Clin Monit Comput. 2002;17:367–376. doi: 10.1023/A:1024261021473. [DOI] [PubMed] [Google Scholar]
- 24.Johannigman JA, Muskat P, Barnes S, Davis K, Beck G, Branson RD. Autonomous control of oxygenation. J Trauma. 2008;64:S295–S301. doi: 10.1097/TA.0b013e31816bce54. [DOI] [PubMed] [Google Scholar]
- 25.Martinoni EP, Pfister CA, Stadler KS, Schumacher PM, Leibundgut D, Bouillon T, et al. Model-based control of mechanical ventilation: design and clinical validation. Br J Anaesth. 2004;92:800–807. doi: 10.1093/bja/aeh145. [DOI] [PubMed] [Google Scholar]
- 26.Jin-Oh H, Dumont GA, Ansermino JM. System identification and closed-loop control of end-tidal CO2 in mechanically ventilated patients. IEEE Trans Inform Technol Biomed. 2012;16:1176–1184. doi: 10.1109/TITB.2012.2204067. [DOI] [PubMed] [Google Scholar]
- 27.Pomprapa A, Misgeld B, Sorgato V, Stollenwerk A, Walter M, Leonhardt S. Robust control of end-tidal CO2 using the hinf loop-shaping approach. Acta Polytech. 2013;53:895–900. doi: 10.14311/AP.2013.53.0895. [DOI] [Google Scholar]
- 28.Tehrani F, Rogers M, Lo T, Malinowski T, Afuwape S, Lum M, et al. A dual closed-loop control system for mechanical ventilation. J Clin Monit Comput. 2003;18:111–129. doi: 10.1023/B:JOCM.0000032744.99885.38. [DOI] [PubMed] [Google Scholar]
- 29.Hamilton Medical AG. INTELLiVENT-ASV operator’s manual HAMILTON-C3; 2021.
- 30.Suter PM, Fairley B, Isenberg MD. Optimum end-expiratory airway pressure in patients with acute pulmonary failure. N Engl J Med. 1975;292:284–289. doi: 10.1056/NEJM197502062920604. [DOI] [PubMed] [Google Scholar]
- 31.Suarez-Sipmann F, Böhm SH, Tusman G, Pesch T, Thamm O, Reissmann H, et al. Use of dynamic compliance for open lung positive end-expiratory pressure titration in an experimental study. Crit Care Med. 2007;35:214–221. doi: 10.1097/01.CCM.0000251131.40301.E2. [DOI] [PubMed] [Google Scholar]
- 32.Writing Group for the Alveolar Recruitment for Acute Respiratory Distress Syndrome Trial (ART) Investigators. Effect of lung recruitment and titrated positive end-expiratory pressure (PEEP) vs low peep on mortality in patients with acute respiratory distress syndrome: A Randomized Clinical Trial. JAMA. 2017;318:1335–1345. [DOI] [PMC free article] [PubMed]
- 33.Munshi L, Ferguson ND. Evolving issues in oxygen therapy in acute care medicine. JAMA. 2020;323:607–608. doi: 10.1001/jama.2019.22029. [DOI] [PubMed] [Google Scholar]
- 34.Tehrani FT. A closed-loop system for control of the fraction of inspired oxygen and the positive end-expiratory pressure in mechanical ventilation. Comput Biol Med. 2012 doi: 10.1016/j.compbiomed.2012.09.007. [DOI] [PubMed] [Google Scholar]
- 35.Wang J, Zhang J, Liu Y, Shang H, Peng L, Cui Z. Relationship between end-tidal carbon dioxide and arterial carbon dioxide in critically ill patients on mechanical ventilation. Medicine (Baltimore) 2021;100:e26973. doi: 10.1097/MD.0000000000026973. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Sipos W, Wiener S, Entenfellner F, Sipos S. Physiological changes of rectal temperature, pulse rate and respiratory rate of pigs at different ages including the critical peripartal period. Wien Tierarztl Monatsschr. 2013;100:93–98. [Google Scholar]
- 37.Becher T, Adelmeier A, Frerichs I, Weiler N, Schädler D. Adaptive mechanical ventilation with automated minimization of mechanical power–a pilot randomized cross-over study. Crit Care. 2019;23:338. doi: 10.1186/s13054-019-2610-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.van der Staay M, Chatburn RL. Advanced modes of mechanical ventilation and optimal targeting schemes. Intensive Care Med Exp. 2018;6:30. doi: 10.1186/s40635-018-0195-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Boerger E, Russ M, von Platen P, Taher M, Hinken L, Pomprapa A, et al. Induction of severe hypoxemia and low lung recruitability for the evaluation of therapeutic ventilation strategies: a translational model of combined surfactant-depletion and ventilator-induced lung injury. ICMx. 2022;10:32. doi: 10.1186/s40635-022-00456-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Nucci G, Mergoni M, Bricchi C, Polese G, Cobelli C, Rossi A. On-line monitoring of intrinsic PEEP in ventilator-dependent patients. J Appl Physiol. 2000;89:985–995. doi: 10.1152/jappl.2000.89.3.985. [DOI] [PubMed] [Google Scholar]
- 41.Bates JHT. Lung mechanics: an inverse modeling approach. Cambridge University Press; 2009.
- 42.Hickling KG. Best compliance during a decremental, but not incremental, positive end-expiratory pressure trial is related to open-lung positive end-expiratory pressure: a mathematical model of acute respiratory distress syndrome lungs. Am J Respir Crit Care Med. 2001;163:69–78. doi: 10.1164/ajrccm.163.1.9905084. [DOI] [PubMed] [Google Scholar]
- 43.von Platen P, Pomprapa A, Lohse A, Leonhardt S, Pickerodt PA, Russ M, et al. Automated positive end-expiratory pressure titration during mechanical ventilation. IFAC-PapersOnLine. 2021;54:412–417. doi: 10.1016/j.ifacol.2021.10.291. [DOI] [Google Scholar]
- 44.Chiumello D, Coppola S, Froio S, Mietto C, Brazzi L, Carlesso E, et al. Time to reach a new steady state after changes of positive end expiratory pressure. Intensive Care Med. 2013;39:1377–1385. doi: 10.1007/s00134-013-2969-x. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data sets used and analysed during the current study are available from the corresponding author on reasonable request.