Skip to main content

This is a preprint.

It has not yet been peer reviewed by a journal.

The National Library of Medicine is running a pilot to include preprints that result from research funded by NIH in PMC and PubMed.

ArXiv logoLink to ArXiv
[Preprint]. 2023 May 11:arXiv:2305.07173v1. [Version 1]

Simulation of a low-Z-medium detector for low-dose high-resolution TOF-PET

Kepler Domurat-Sousa, Cameron M Poe, Maya S McDaniel, Eric Spieglan, Joao F Shida, Evan Angelico, Bernhard W Adams, Patrick J La Riviere, Henry J Frisch, Allison H Squires
PMCID: PMC10197732  PMID: 37214132

Abstract

Two major challenges in time-of-flight positron emission tomography (TOF-PET) are low spatial resolution and high radioactive dose to the patient, both of which result from limitations in detection technology rather than fundamental physics. A new type of TOF-PET detector employing low-atomic number (low-Z) scintillation media and large-area, high-resolution photodetectors to record Compton scattering locations in the detector has been proposed as a promising alternative, but neither a direct comparison to state-of-the-art TOF-PET nor the minimum technical requirements for such a system have yet been established. Here we present a simulation study evaluating the potential of a proposed low-Z detection medium, linear alkylbenzene (LAB) doped with a switchable molecular recorder, for next-generation TOF-PET detection. We developed a custom Monte Carlo simulation of full-body TOF-PET using the TOPAS Geant4 software package. By quantifying contributions and tradeoffs for energy, spatial, and timing resolution of the detector, we show that a reasonable combination of specifications improves TOF-PET sensitivity by more than 5x, with comparable or better spatial resolution and 40-50% enhanced contrast-to-noise as compared to state-of-the-art scintillating crystal materials. These improvements enable clear imaging of a brain phantom simulated at less than 1% of a standard radiotracer dose, which could enable expanded access and new clinical applications for TOF-PET.

Full Text Availability

The license terms selected by the author(s) for this preprint version do not permit archiving in PMC. The full text is available from the preprint server.

20 pages, 7 figures


Articles from ArXiv are provided here courtesy of arXiv

RESOURCES