Case Report: When Two is Worse Than One: Stereo Imbalance in a Case of Wavefront-Guided Scleral Lenses

Lan Chi Nguyen 1, Wei Hau Lew 1, Matthew J Kauffman 1, Jason D Marsack 1, Raymond A Applegate 1, Daniel R Coates 1
PMCID: PMC10205691  NIHMSID: NIHMS1883685  PMID: 36951821

Abstract

Significance:

Wavefront-guided scleral lenses (WGSLs) reduce visually debilitating residual higher-order aberrations (HOAs). While reduced HOAs lead to improvement in monocular high contrast visual acuity (VA), the success of the lenses in everyday life depends on additional factors such as retinal contrast, binocular balance, and stereoacuity.

Purpose:

To report a case where WGSLs provided improved monocular vision compared to scleral lenses (SLs), but reduced binocularity and stereo acuity.

Case Report:

A 48-year-old female with moderate keratoconus OD, severe OS was fitted with SLs and WGSLs. VA with best SLs was 20/20ˉ2 OD and 20/25ˉ2 OS. Residual higher-order root mean square (HORMS) wavefront error (6 mm pupil) was 0.56 μm OD and 1.38 μm OS. VA with WGSLs was 20/16ˉ2 OD and 20/25+2 OS, and residual HORMS was 0.41 μm OD and 0.98 μm OS. Monocularly, WGSLs were reported to provide better VA. However, binocularly, patient reported an “imbalanced feeling” and preferred the SLs over WGSLs. Binocular VA at distance was 20/25 with SLs and 20/25ˉ2 with WGSL. To investigate, the Worth-4-Dot test was performed and reported fusion with SLs, but suppression OS at distance with WGSLs. Stereoacuity was 160 arc seconds at near and 120 arc seconds at distance with SLs and 400 arc seconds at near and > 1200 arc seconds at distance with WGSLs. Dichoptic contrast balancing showed a balance point of 0.48 with SLs and 0.17 with WGSLs, indicating a strong preference towards OD. Simulation of the patient’s retinal image revealed a greater difference in image contrast between the two eyes with WGSLs.

Conclusions:

WGSLs reduced HORMS and improved VA compared to SLs. However, in this case, it inadvertently caused binocular imbalance. As WGSLs become more widely available, future work should include methods to optimize binocular balance to maximize overall patient satisfaction.


Wavefront-guided scleral lenses are an emerging clinical optical treatment for patients with highly aberrated eyes (keratoconus, post-surgical ectasia, etc.) and have been demonstrated to reduce visually debilitating residual higher-order aberrations that are not correctable with traditional scleral lenses, which have been shown to mask higher-order aberrations between 60 and 65%.1,2 By specifically targeting residual higher-order aberrations with wavefront-guided scleral lenses, visual symptoms such as smearing and ghosting are diminished, leading to an improvement in the individual’s perceived visual quality.3 While reduced higher-order aberrations lead to improvement in high-contrast visual acuity, the success of lenses in everyday life depends on additional factors such as retinal contrast, interaction of residual aberrations, binocular balance, and stereoacuity.4

This case report focuses on an individual whose best-corrected monocular visual acuity, after 8 weeks of adaptation for each scleral lens type, was reported to be qualitatively superior for both eyes with wavefront-guided scleral lenses compared to scleral lenses. Objectively, the residual higher-order root mean square wavefront error over a 6 mm pupil was lower with wavefront-guided scleral lenses in both eyes than with scleral lenses. However, the patient reported an “imbalanced feeling” during binocular viewing and preferred scleral lenses over wavefront-guided scleral lenses for everyday tasks. Although wavefront-guided scleral lenses have been found to allow more highly aberrated eyes to reach normal levels of optical performance, visual performance, and quality of life,1,3,5 this case demonstrates the success of wavefront-guided scleral lenses also relies on other factors, such as binocularity. The clinical timeline indicating key events relevant to this case presentation, intervention, and resolution is shown in Figure 1.

Figure 1.

Figure 1.

Clinical Timeline: 48-year-old female with keratoconus, was initially intervened with scleral lens (SL), then wavefront-guided scleral lens (WGSL) due to elevated residual higher-order root mean square wavefront error (HORMS WFE). The patient then underwent evaluation for binocular vision. RGP= Rigid Gas Permeable; VA= visual acuity; Ox= Over refraction; Unit for Randot test=arc second.

CASE REPORT

A 48-year-old female with bilateral keratoconus (moderate disease in the right, severe disease in the left; disease severity was determined using the ABCD keratoconus grading system6) was enrolled in a research study examining wavefront-guided scleral lenses, which was approved by the Institutional Review Board of the University of Houston and adhered to the Declaration of Helsinki. Written informed consent was obtained for identifiable health information; however, no identifiable health information was included in this case report. After completion of conventional scleral lens fitting, refinement of the prescription, and eight-week adaptation, the Snellen visual acuity was 20/20ˉ2 in the right eye and 20/25ˉ2 in the left eye. As displayed in Table 1 and the timeline in Study 1, the residual higher-order root mean square wavefront error through the scleral lenses (3rd through 10th Zernike radial orders) measured with a COAS HD wavefront sensor (Johnson and Johnson Vision, Santa Ana, CA) over a 6 mm pupil was 0.56 μm right eye and 1.38 μm left eye, both of which were higher than the typical values for the subjects age at this pupil diameter.7 Following the study protocol,1,5 the patient was dilated with one drop of 1% tropicamide and one drop of 2.5% neosynephrine. Conventional scleral lenses were applied, and the residual wavefront error and lens’ offsets with respect to the patient’s dilated pupils were measured. Wavefront-guided scleral lenses targeting aberrations in the 2nd to 5th Zernike radial orders were manufactured using a DAC 2X-ALM OTT ophthalmic lens lathe (DAC International, Carpinteria, CA) in Boston XO material (Bausch and Lomb, Rochester, NY). The wavefront-guided correction was designed with an offset from the geometric center of the lens such that the correction was coaxial to the center of the dilated pupil and rotationally aligned to correct the underlying wavefront error. After refinement of the prescription and eight-week adaptation period wearing the wavefront-guided scleral lenses, the patient’s Snellen visual acuity with wavefront-guided scleral lenses was 20/16ˉ2 right eye and 20/25+2 left eye, and residual higher-order root mean square was 0.41 μm right eye and 0.98 μm left eye. When judged monocularly, wavefront-guided scleral lenses were reported to be qualitatively superior to scleral lenses for both eyes. However, the patient reported an “imbalanced feeling” during binocular viewing and preferred scleral lenses over wavefront-guided scleral lenses for everyday tasks. Binocular visual acuity at a distance was 20/25 with scleral lenses and slightly reduced to 20/25ˉ2 with wavefront-guided scleral lenses. Furthermore, simulation of the patient’s retinal image by importing the residual total root mean square wavefront error into Visual Optics Laboratory software (Saver and Associates, Inc.), and convolving the corresponding point-spread function with a logMAR visual acuity chart, suggested a greater difference in retinal image contrast between the two eyes with wavefront-guided scleral lenses as compared to scleral lenses. The image from the left eye wearing the wavefront-guided scleral lenses appeared to have clearer edges and greater contrast, but also doubled (Figure 2).

Table 1.

Acuity and aberrations data (Study 1) and binocular assessment data (Study 2) for scleral lenses and wavefront-guided scleral lenses.

Scleral Lens Wavefront-guided Scleral Lens
STUDY 1
Acuity & Aberrations
Eye right eye left eye right eye left eye
Visual Acuity 20/20−2 20/25−2 20/16−2 20/25+2
Residual HORMS (6mm Pupil) 0.56 μm 1.38 μm 0.41 μm 0.98 μm
Binocular Visual Acuity 20/25 20/25−2
STUDY 2
Binocular Assessment
Worth Four Dot fusion suppression left eye
Randot Stereoacuity at Near 160 arc sec 400 arc sec
Sloan RDS Stereoacuity at Distance 120 arc sec >1200 arc sec
Dichoptic contrast-balancing test at Near .42 .41
Dichoptic contrast-balancing test at Distance .48 .17

HORMS = higher-order root mean square; RDS = random dot stereograms

Figure 2.

Figure 2.

Simulation of retinal image quality shows a greater difference in retinal image contrast between the two eyes with wavefront-guided scleral lens as compared to scleral lens. The simulated image in the left eye has greater contrast, clearer edge but also doubling with wavefront-guided scleral lens.

To further investigate the subject’s perception, a follow-up study assessing binocular vision was conducted, which was approved by the Institutional Review Board of the University of Houston and adhered to the Declaration of Helsinki, and written informed consent was obtained. The Worth Four Dot Test was performed, and the outcomes suggested fusion at distance and near with scleral lenses and only at near with wavefront-guided scleral lenses; left-eye suppression was detected at distance with wavefront-guided scleral lenses. Using the Randot test (Stereo Optical, Inc.) at near with presbyopia correction, the measured stereoacuity was 160 arc seconds with scleral lenses but reduced to 400 arc seconds with wavefront-guided scleral lenses. To test stereoacuity at distance, a computerized global stereopsis test developed in-house was employed.8 Stereo presentation was achieved using the Nvidia active shutter glasses. The target size was scaled to 20/800 logMAR (0.75cpd) at 4 m, and the task was to identify Sloan letters that “popped out” in depth from random dot stereograms. The stereo threshold was estimated using the staircase method with six reversal points, and the mean of the last four reversals was used to calculate the mean stereo threshold. The testing procedure was repeated for each condition, and the average value was calculated. With wavefront-guided scleral lenses, the stereo acuity was poorer compared to scleral lenses at near (400 vs. 160 arc secs) and more significantly at far (>1200 vs. 120 arc secs), suggesting poorer binocularity.

To determine whether stereo acuity loss is an indirect result of binocular contrast balancing, as Marella et al. (IOVS 2022;63:ARVO E-Abstract 4307), binocular contrast balancing test was administered. Studies have suggested that abnormal binocular interaction may be affected differently at different spatial frequencies;9,10 hence, the employment of a dichoptic contrast-balancing test that consists of the letter E at different sizes scaled according to five different spatial frequencies (.5, 1.5, 5, 9.52, and 15 cycles per degree). As illustrated in Figure 3, two dichoptic Sloan letters E with different contrast levels and orientations were presented simultaneously at the same location on a 3D monitor. While wearing a pair of active shutter goggles, the subject was asked to identify the orientation of the “more visible letter E.” The task was tested with multiple contrast combinations (90–10%, 80–20%, 60–40%, 40–60%, 20–80%, and 10–90%; the first value indicates right-eye contrast), four different sizes, at distance (4 m) and near (40 cm). For each condition, a psychometric sigmoidal curve was fitted, and the midpoint (50–50) response rate between the left and right eyes was estimated to determine the contrast balance point. A contrast balance point of 0.5 reflects an equal contribution between the two eyes, while a smaller value (closer to zero) suggests a bias towards the right eye, and a value closer to 1 indicates a preference for the left eye. The subject’s balance point was 0.42 at near and 0.48 at distance with scleral lenses, which was within the reported normative data,11 suggesting a balanced binocularity both at distance and near. In contrast, the subject’s balance point was 0.41 at near and 0.17 at distance with wavefront-guided scleral lenses, suggesting a strong preference towards the right eye at distance while maintaining a more balanced binocularity at near (Figure 4). The results of the binocular assessments are summarized in Table 1, Study 2.

Figure 3.

Figure 3.

Schematic of the dichoptic contrast-balancing test.

Figure 4.

Figure 4.

Contrast balance point as a function of Spatial Frequency. At near, the contrast balance points were almost the same for wavefront-guided scleral lens and scleral lens, with a slight bias towards the better eye (right eye). These two values were close to the equal contrast balance between the two eyes (dotted red line at 0.5). Data points closer to the red line indicate better binocular balance. However, there is a strong bias towards the right eye with the wavefront-guided scleral lens at distance (dashed dark gray line), suggesting degraded binocularity.

Contrast Balance Point as a Function of Spatial Frequency

At near, the contrast balance points were almost the same for wavefront-guided scleral lens and scleral lens, with a slight bias towards the better eye (right eye). These two values were close to the equal contrast balance between the two eyes (dotted red line at 0.5). Data points closer to the red line indicate better binocular balance. However, there is a strong bias towards the right eye with the wavefront-guided scleral lens at distance (dashed dark gray line), suggesting degraded binocularity.

DISCUSSION

Wavefront-guided corrections directly target the residual, uncorrected higher-order aberrations that are inevitable with conventional corrections, particularly for highly aberrated eyes.3 This technology is no longer a theory but a reality, as indicated by commercially available wavefront-guided scleral lenses and clinical trials.1113 Typically, in both laboratory and clinical settings, the improvement in visual quality with wavefront-guided scleral lenses is quantified by visual acuity, which is a high-contrast letter recognition task. As such, it does not reflect the quality of the letter, only whether the letter can be correctly identified, as illustrated by the simulation in Figure 2. Importantly, aberrations are known interact, and these interactions can lead to both an improvement and reduction in monocular visual performance that is not reflective of the quantity of the aberration present. Work on these interactions has focused on intra-eye interactions. The current case suggests future study of inter-eye interactions.1416 In addition, visual acuity is typically measured monocularly, and clinicians would expect an improvement in stereopsis with better visual acuity. However, the visual scenes and stereo demands in the real world are far more complex and differ from the targets used in clinical tests. They vary in contrast, color, size, and most importantly, are not static. This study highlights the discrepancy between clinical tests and visual performance in the real world from the perspective of the subject. Furthermore, in this study, a contrast-balancing paradigm was incorporated as part of the binocular assessment to assess the depth of interocular suppression as a result of monocular dominance and the potential to improve stereo acuity at the balance point. Similar methods have revealed analogous deficiencies in the amblyopic population.17 Marella et al. (IOVS 2021;62:ARVO E-Abstract 15) have also reported that in the keratoconus population, the better eye dominates and suppresses the worse eye, leading to an imbalanced contrast balance point. This could potentially explain the loss of stereoacuity observed in our study subject. The subject had a lower contrast balance point with the wavefront-guided scleral lenses than with the scleral lenses. This suggests more binocular rivalry, stronger suppression, and ultimately, degraded stereopsis.

While the results were rather unexpected, given the slight improvement of monocular visual acuity following wavefront-guided scleral lenses, particularly in the right eye, this behavior has been reported in the stereopsis literature. Interocular differences in blur have been reported to affect stereoacuity more than binocular blur.18,19 Although wavefront-guided scleral lenses improved visual quality by reducing (but not to age-and-pupil-matched typical levels) higher-s in this case, stereoacuity in the presence of residual higher-order aberrations during wavefront-guided scleral lenses wear was reduced compared to that with scleral lenses. Depth perception requires precise motor vergence.20 Patients with longstanding keratoconus may have altered binocular fixation behaviors or individualized strategies to minimize the impact of higher-order aberrations on binocularity. Wavefront data reported in the current study were recorded monocularly and not in a manner in which ocular alignment or first-order prismatic effects can be assessed.

While the future is bright for wavefront-guided optics, continuing education on wavefront-guided corrections, their strengths, limitations, and other requirements is essential in utilizing wavefront-guided scleral lenses to meet the binocular visual needs of highly aberrated eyes. As wavefront-guided scleral lenses become more widely available, the next step should include binocular wavefront measurement and emphasis on binocularity to maximize overall patient satisfaction. This may include an algorithm to simulate and predict stereo thresholds with different combinations of higher-order aberration corrections that include different orders and magnitudes and a derivation of a metric to optimize retinal image similarity in both eyes.

ACKNOWLEDGMENTS

Presented as a poster at the American Academy of Optometry (AAO) annual meeting (virtual), October 2020. Funded by grants NIH/NEI R01EY019105 (JDM and RAA), and NIH/NEI P30EY0755; UHCO Start up grant (DRC).

REFERENCES

  • 1.Hastings GD, Applegate RA, Nguyen LC, et al. Comparison of Wavefront-Guided and Best Conventional Scleral Lenses after Habituation in Eyes with Corneal Ectasia. Optom Vis Sci 2019;96:238–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Kosaki R, Maeda N, Bessho K, et al. Magnitude and Orientation of Zernike Terms in Patients with Keratoconus. Invest Ophthalmol Vis Sci 2007;48:3062–8. [DOI] [PubMed] [Google Scholar]
  • 3.Nguyen LC, Kauffman MJ, Hastings GD, et al. Case Report: What Are We Doing for Our “20/20 Unhappy” Scleral Lens Patients? Optom Vis Sci 2020;97:826–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Metlapally S, Bharadwaj SR, Roorda A, et al. Binocular Cross-correlation Analyses of the Effects of High-Order Aberrations on the Stereoacuity of Eyes with Keratoconus. J Vis 2019;19:1–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Marsack JD, Ravikumar A, Nguyen C, et al. Wavefront-guided Scleral Lens Correction in Keratoconus. Optom Vis Sci 2014;91:1221–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Belin MW, Duncan JK. Keratoconus: The ABCD Grading System. Klin Monbl Augenheilkd 2016;233:701–07. [DOI] [PubMed] [Google Scholar]
  • 7.Applegate RA, Donnelly WJ, Marsack JD, et al. Three-Dimensional Relationship between High-Order Root-Mean-Square Wavefront Error, Pupil Diameter, and Aging. J Opt Soc Am (A) 2007;24:578–87. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Lew WH, Coates DR. Assessment of Depth Perception with a Comprehensive Disparity Defined Letter Test: A Pilot Study. PLoS One 2022;17:e0271881. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Kwon M, Wiecek E, Dakin SC, Bex PJ. Spatial-Frequency Dependent Binocular Imbalance in Amblyopia. Sci Rep 2015;5:1–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Ding J, Klein SA, Levi DM. Binocular Combination in Abnormal Binocular Vision. J Vis 2013;13:1–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.ClinicalTrials.gov. Wavefront Guided Scleral Lenses for Keratoconus and Irregular Astigmatism: ClinicalTrials.gov Identifier: NCT04698590. Available at: https://clinicaltrials.gov/ct2/show/NCT04698590. Accessed June 27, 2022.
  • 12.OVITZ. Custom Wavefront Guided Lenses Available at: https://www.ovitz.us. Accessed June 27, 2022.
  • 13.LaserFit®. Personalized Wavefront-Guided Scleral Contact Lenses. Available at: https://laserfitlens.com/personalized-wavefront-guided-lenses. Accessed June 27, 2022.
  • 14.Sawides L, de Gracia P, Dorronsoro C, et al. Vision is Adapted to the Natural Level of Blur Present in the Retinal Image. PLoS One 2011;6:e27031. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.de Gracia P, Dorronsoro C, Marin G, et al. Visual Acuity under Combined Astigmatism and Coma: Optical and Neural Adaptation Effects. J Vis 2011;11:5. [DOI] [PubMed] [Google Scholar]
  • 16.Applegate RA, Marsack JD, Ramos R, Sarver EJ. Interaction between Aberrations to Improve or Reduce Visual Performance. J Cataract Refract Surg 2003;29:1487–95. [DOI] [PubMed] [Google Scholar]
  • 17.Hess RF, Thompson B, Baker DH. Binocular Vision in Amblyopia: Structure, Suppression and Plasticity. Ophthalmic Physiol Opt 2014;34:146–62. [DOI] [PubMed] [Google Scholar]
  • 18.Jiménez JR, Castro JJ, Jiménez R, Hita E. Interocular Differences in Higher-Order Aberrations on Binocular Visual Performance. Optom Vis Sci 2008;85:174–9. [DOI] [PubMed] [Google Scholar]
  • 19.Wood IC. Stereopsis with Spatially-Degraded Images. Ophthalmic Physiol Opt 1983;3:337–40. [PubMed] [Google Scholar]
  • 20.Ukwade MT, Bedell HE, Harwerth RS. Stereopsis is Perturbed by Vergence Error. Vision Res 2003;43:181–93. [DOI] [PubMed] [Google Scholar]

RESOURCES