Use of jet mixing devices with an oxygen concentrator.

M B Dobson 1
PMCID: PMC1021101  PMID: 1494771

Abstract

Jet mixing devices ("Venturi" devices) can be used in conjunction with domiciliary oxygen concentrators and provide delivered oxygen concentrations similar to those obtained with medical oxygen, though with the devices delivering higher concentrations (above 30% oxygen) the total flow is substantially reduced. A jet mixing device driven by a domiciliary concentrator would be valuable in various circumstances, especially in developing countries and also for infants and for patients with upper respiratory tract infections who are breathing through the mouth.

Full text

PDF
1060

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Canet J., Sanchis J. Performance of a low flow O2 Venturi mask: diluting effects of the breathing pattern. Eur J Respir Dis. 1984 Jan;65(1):68–73. [PubMed] [Google Scholar]
  2. Johns D. P., Streeton J. A., Rochford P. D. An air-entrainment device for preparing precision gas mixtures. J Med Eng Technol. 1983 May-Jun;7(3):140–143. doi: 10.3109/03091908309032578. [DOI] [PubMed] [Google Scholar]
  3. Jones H. A., Turner S. L., Hughes J. M. Performance of the large-reservoir oxygen mask (Ventimask). Lancet. 1984 Jun 30;1(8392):1427–1431. doi: 10.1016/s0140-6736(84)91930-5. [DOI] [PubMed] [Google Scholar]
  4. Kittredge P. Neither Venturi nor Bernoulli. Lancet. 1983 Jan 22;1(8317):182–182. doi: 10.1016/s0140-6736(83)92779-4. [DOI] [PubMed] [Google Scholar]