Auscultatory Blood Pressure Measurement—Effect of Pressure on the Head of the Stethoscope

Sol Londe, Thomas S Klitzner
PMCID: PMC1021735  PMID: 6495724

Abstract

Excessive pressure on the stethoscope head in auscultatory blood pressure measurement does not affect systolic blood pressure value but it does erroneously lower diastolic readings and frequently causes the sounds to persist to zero. Consequently, the lightest possible pressure should be placed on the stethoscope head.

Full text

PDF
193

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. BORDLEY J., 3rd, CONNOR C. A. R., HAMILTON W. F., KERR W. J., WIGGERS C. J. Recommendations for human blood pressure determinations by sphygmomanometers. Circulation. 1951 Oct;4(4):503–509. doi: 10.1161/01.cir.4.4.503. [DOI] [PubMed] [Google Scholar]
  2. Kirkendall W. M., Feinleib M., Freis E. D., Mark A. L. Recommendations for human blood pressure determination by sphygmomanometers. Subcommittee of the AHA Postgraduate Education Committee. Circulation. 1980 Nov;62(5):1146A–1155A. [PubMed] [Google Scholar]
  3. LANGE R. L., HECHT H. H. Genesis of pistol-shot and Korotkoff sounds. Circulation. 1958 Nov;18(5):975–978. doi: 10.1161/01.cir.18.5.975. [DOI] [PubMed] [Google Scholar]
  4. O'Brien E. T., O'Malley K. ABC of blood pressure measurement: technique. Br Med J. 1979 Oct 20;2(6196):982–984. doi: 10.1136/bmj.2.6196.982. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Western Journal of Medicine are provided here courtesy of BMJ Publishing Group

RESOURCES