Abstract
The heart rate response of 59 patients aged 17-79 years implanted with seven different types of rate responsive pacemakers was evaluated during graded exercise treadmill testing and during standardised daily activities. The heart rate response in patients with pacemakers was compared with the chronotropic response in 20 healthy controls of similar age and sex distribution who performed identical protocols. All pacemaker types adequately simulated the control heart rate response during the graded exercise treadmill test except during the early stages of exercise. However, during everyday activities, the response of ventricular rate responsive (VVIR) pacemakers was varied. Activity sensing systems rapidly overresponded to staircase descent, to changes in walking speed, and to suitcase lifting with the pacemaker arm, and these systems did not respond to mental stress. "Physiological" sensors (QT and minute ventilation units) responded slowly to rapid changes in physiological demand. The QT pacemaker patients did respond to mental stress but showed a paradoxical increase in rate during the recovery phases of burst exercise protocols such as staircase ascent/descent and walking deceleration. Dual chamber pacemakers in VDD, DDD, and DDDR modes most closely simulated the normal chronotropic response during everyday activities. Graded exercise treadmill testing, in isolation, may not be the best way to asses or program the heart rate response in patients with the heart rate adaptive pacemakers because changes in heart rate during everyday activities may deviate considerably from the normal sinus response despite satisfactory simulation of the normal chronotropic response during treadmill testing.
Full text
PDF






Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Alt E., Heinz M., Hirgstetter C., Emslander H. P., Daum S., Blömer H. Control of pacemaker rate by impedance-based respiratory minute ventilation. Chest. 1987 Aug;92(2):247–252. doi: 10.1378/chest.92.2.247. [DOI] [PubMed] [Google Scholar]
- Alt E., Hirgstetter C., Heinz M., Blömer H. Rate control of physiologic pacemakers by central venous blood temperature. Circulation. 1986 Jun;73(6):1206–1212. doi: 10.1161/01.cir.73.6.1206. [DOI] [PubMed] [Google Scholar]
- Alt E., Völker R., Wirtzfeld A., Ulm K. Survival and follow-up after pacemaker implantation: a comparison of patients with sick sinus syndrome, complete heart block, and atrial fibrillation. Pacing Clin Electrophysiol. 1985 Nov;8(6):849–855. doi: 10.1111/j.1540-8159.1985.tb05904.x. [DOI] [PubMed] [Google Scholar]
- Ausubel K., Furman S. The pacemaker syndrome. Ann Intern Med. 1985 Sep;103(3):420–429. doi: 10.7326/0003-4819-103-3-420. [DOI] [PubMed] [Google Scholar]
- Beyersdorf F., Kreuzer J., Happ T. A., Zegelman M., Satter P. Increase in cardiac output with rate-responsive pacemaker. Ann Thorac Surg. 1986 Aug;42(2):201–205. doi: 10.1016/s0003-4975(10)60519-x. [DOI] [PubMed] [Google Scholar]
- Cammilli L., Alcidi L., Papeschi G., Wiechmann V., Padeletti L., Grassi G. Preliminary experience with the pH-triggered pacemaker. Pacing Clin Electrophysiol. 1978 Oct;1(4):448–457. doi: 10.1111/j.1540-8159.1978.tb03506.x. [DOI] [PubMed] [Google Scholar]
- Faerestrand S., Breivik K., Ohm O. J. Assessment of the work capacity and relationship between rate response and exercise tolerance associated with activity-sensing rate-responsive ventricular pacing. Pacing Clin Electrophysiol. 1987 Nov;10(6):1277–1290. doi: 10.1111/j.1540-8159.1987.tb04964.x. [DOI] [PubMed] [Google Scholar]
- Humen D. P., Kostuk W. J., Klein G. J. Activity-sensing, rate-responsive pacing: improvement in myocardial performance with exercise. Pacing Clin Electrophysiol. 1985 Jan;8(1):52–59. doi: 10.1111/j.1540-8159.1985.tb05723.x. [DOI] [PubMed] [Google Scholar]
- Karlöf I. Haemodynamic effect of atrail triggered versus fixed rate pacing at rest and during exercise in complete heart block. Acta Med Scand. 1975 Mar;197(3):195–206. [PubMed] [Google Scholar]
- Lau C. P., Camm A. J. Role of left ventricular function and Doppler-derived variables in predicting hemodynamic benefits of rate-responsive pacing. Am J Cardiol. 1988 Nov 1;62(13):906–911. doi: 10.1016/0002-9149(88)90891-0. [DOI] [PubMed] [Google Scholar]
- Lau C. P., Mehta D., Toff W. D., Stott R. J., Ward D. E., Camm A. J. Limitations of rate response of an activity-sensing rate-responsive pacemaker to different forms of activity. Pacing Clin Electrophysiol. 1988 Feb;11(2):141–150. doi: 10.1111/j.1540-8159.1988.tb04535.x. [DOI] [PubMed] [Google Scholar]
- Lau C. P., Wong C. K., Leung W. H., Cheng C. H., Lo C. W. A comparative evaluation of a minute ventilation sensing and activity sensing adaptive-rate pacemakers during daily activities. Pacing Clin Electrophysiol. 1989 Sep;12(9):1514–1521. doi: 10.1111/j.1540-8159.1989.tb06156.x. [DOI] [PubMed] [Google Scholar]
- Lipkin D. P., Buller N., Frenneaux M., Ludgate L., Lowe T., Webb S. C., Krikler D. M. Randomised crossover trial of rate responsive Activitrax and conventional fixed rate ventricular pacing. Br Heart J. 1987 Dec;58(6):613–616. doi: 10.1136/hrt.58.6.613. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Maisch B., Langenfeld H. Rate adaptive pacing--clinical experience with three different pacing systems. Pacing Clin Electrophysiol. 1986 Nov;9(6 Pt 2):997–1004. doi: 10.1111/j.1540-8159.1986.tb06660.x. [DOI] [PubMed] [Google Scholar]
- Nordlander R., Pehrsson S. K., Aström H., Karlsson J. Myocardial demands of atrial-triggered versus fixed-rate ventricular pacing in patients with complete heart block. Pacing Clin Electrophysiol. 1987 Sep;10(5):1154–1159. doi: 10.1111/j.1540-8159.1987.tb06135.x. [DOI] [PubMed] [Google Scholar]
- Rickards A. F., Donaldson R. M., Thalen H. J. The use of QT interval to determine pacing rate: early clinical experience. Pacing Clin Electrophysiol. 1983 Mar;6(2 Pt 2):346–356. doi: 10.1111/j.1540-8159.1983.tb04371.x. [DOI] [PubMed] [Google Scholar]
- Rossi P., Plicchi G., Canducci G., Rognoni G., Aina F. Respiratory rate as a determinant of optimal pacing rate. Pacing Clin Electrophysiol. 1983 Mar;6(2 Pt 2):502–510. doi: 10.1111/j.1540-8159.1983.tb04399.x. [DOI] [PubMed] [Google Scholar]
- Rossi P. Rate-responsive pacing: biosensor reliability and physiological sensitivity. Pacing Clin Electrophysiol. 1987 May;10(3 Pt 1):454–466. doi: 10.1111/j.1540-8159.1987.tb04507.x. [DOI] [PubMed] [Google Scholar]
- Rossi P., Rognoni G., Occhetta E., Aina F., Prando M. D., Plicchi G., Minella M. Respiration-dependent ventricular pacing compared with fixed ventricular and atrial-ventricular synchronous pacing: aerobic and hemodynamic variables. J Am Coll Cardiol. 1985 Sep;6(3):646–652. doi: 10.1016/s0735-1097(85)80126-1. [DOI] [PubMed] [Google Scholar]
- Stangl K., Wirtzfeld A., Heinze R., Laule M., Seitz K., Göbl G. A new multisensor pacing system using stroke volume, respiratory rate, mixed venous oxygen saturation, and temperature, right atrial pressure, right ventricular pressure, and dP/dt. Pacing Clin Electrophysiol. 1988 Jun;11(6 Pt 1):712–724. doi: 10.1111/j.1540-8159.1988.tb06021.x. [DOI] [PubMed] [Google Scholar]
- Wirtzfeld A., Goedel-Meinen L., Bock T., Heinze R., Liss H. D., Munteanu J. Central venous oxygen saturation for the control of automatic rate-responsive pacing. Pacing Clin Electrophysiol. 1982 Nov;5(6):829–835. doi: 10.1111/j.1540-8159.1982.tb06564.x. [DOI] [PubMed] [Google Scholar]
- Yasue H., Takizawa A., Nagao M., Nishida S., Horie M., Kubota J., Omote S., Takaoka K., Okumura K. Long-term prognosis for patients with variant angina and influential factors. Circulation. 1988 Jul;78(1):1–9. doi: 10.1161/01.cir.78.1.1. [DOI] [PubMed] [Google Scholar]