Abstract
The accuracy and reliability of measurement of coronary artery dimensions and detection of atherosclerotic lesions by high frequency epicardial echocardiography were compared with histopathological results. Ten pressure perfused human hearts were examined in vitro with a 10 MHz (Diasonics) transducer and a 7.5 MHz (Vingmed/Sonotron) transducer. There was close agreement between ultrasound and pathological measurements of coronary artery luminal diameter. Qualitative changes in wall structure such as diffuse wall thickening and calcification were readily identified; however, the resolution of the transducers was not high enough accurately to measure wall dimensions in normal coronary arteries. Coefficient of variation measurements for intra and inter observer variability (5.2% and 6.9% respectively) showed excellent reproducibility. The technique was accurate in identifying atherosclerotic lesions, imaging arteries distal to an occlusion, locating deeply sited arteries, and identifying complete obliteration of an artery. Intraoperative video playback and transducer miniaturisation may minimise problems caused by cardiac movement and restricted access. With these developments intraoperative assessment of coronary artery disease may become a real possibility.
Full text
PDF





Images in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Arnett E. N., Isner J. M., Redwood D. R., Kent K. M., Baker W. P., Ackerstein H., Roberts W. C. Coronary artery narrowing in coronary heart disease: comparison of cineangiographic and necropsy findings. Ann Intern Med. 1979 Sep;91(3):350–356. doi: 10.7326/0003-4819-91-3-350. [DOI] [PubMed] [Google Scholar]
- Gould K. L. Percent coronary stenosis: battered gold standard, pernicious relic or clinical practicality? J Am Coll Cardiol. 1988 Apr;11(4):886–888. doi: 10.1016/0735-1097(88)90227-6. [DOI] [PubMed] [Google Scholar]
- Marcus M. L., Skorton D. J., Johnson M. R., Collins S. M., Harrison D. G., Kerber R. E. Visual estimates of percent diameter coronary stenosis: "a battered gold standard". J Am Coll Cardiol. 1988 Apr;11(4):882–885. doi: 10.1016/0735-1097(88)90226-4. [DOI] [PubMed] [Google Scholar]
- McPherson D. D., Armstrong M., Rose E., Kieso R. A., Megan M., Hunt M., Hite P., Marcus M. L., Kerber R. E. High frequency epicardial echocardiography for coronary artery evaluation: in vitro and in vivo validation of arterial lumen and wall thickness measurements. J Am Coll Cardiol. 1986 Sep;8(3):600–606. doi: 10.1016/s0735-1097(86)80189-9. [DOI] [PubMed] [Google Scholar]
- Sahn D. J., Barratt-Boyes B. G., Graham K., Kerr A., Roche A., Hill D., Brandt P. W., Copeland J. G., Mammana R., Temkin L. P. Ultrasonic imaging of the coronary arteries in open-chest humans: evaluation of coronary atherosclerotic lesions during cardiac surgery. Circulation. 1982 Nov;66(5):1034–1044. doi: 10.1161/01.cir.66.5.1034. [DOI] [PubMed] [Google Scholar]
- Thomas A. C., Davies M. J. The demonstration of cardiac pathology using perfusion-fixation. Histopathology. 1985 Jan;9(1):5–19. doi: 10.1111/j.1365-2559.1985.tb02967.x. [DOI] [PubMed] [Google Scholar]
- Vlodaver Z., Frech R., Van Tassel R. A., Edwards J. E. Correlation of the antemortem coronary arteriogram and the postmortem specimen. Circulation. 1973 Jan;47(1):162–169. doi: 10.1161/01.cir.47.1.162. [DOI] [PubMed] [Google Scholar]
- White C. W., Wright C. B., Doty D. B., Hiratza L. F., Eastham C. L., Harrison D. G., Marcus M. L. Does visual interpretation of the coronary arteriogram predict the physiologic importance of a coronary stenosis? N Engl J Med. 1984 Mar 29;310(13):819–824. doi: 10.1056/NEJM198403293101304. [DOI] [PubMed] [Google Scholar]