Abstract
INTRODUCTION--Over the past decade, several advances have been made in the management of tachycardias by pacing techniques, but limited data are available on the long-term outcome of patients treated with antitachycardia pacemakers. PATIENTS AND METHODS--An antitachycardia pacemaker, the Intermedics Intertach, was implanted in 22 (17 female) patients with supraventricular tachycardia over a five year period. All were selected after detailed evaluation and testing of a temporary antitachycardia pacemaker system showed that their arrhythmia could be stopped promptly, reliably, and under different physiological conditions. RESULTS--The 22 patients have been followed up for a mean period of 57.3 (range 19-76) months. All except one of the patients has had frequent episodes of tachycardia reliably ended by the pacemaker. Complications have occurred in seven patients, necessitating removal of the pacing system in four. Of the 18 patients who continue to have pacemakers, seven are being treated with beta blockers or verapamil; no other antiarrhythmic drugs are being taken. CONCLUSIONS--Antitachycardia pacing is an acceptable long-term option for carefully selected patients with supraventricular tachycardia, but even after extensive testing a substantial number of the patients may continue to require drug treatment. Furthermore, the widespread use of curative techniques for supraventricular arrhythmias (catheter ablation and surgery) has decreased the need for this palliative treatment.
Full text
PDF




Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Barold S. S., Ryan G. F., Goldstein S. The first implanted tachycardia-terminating pacemaker. Pacing Clin Electrophysiol. 1989 May;12(5):870–874. doi: 10.1111/j.1540-8159.1989.tb01912.x. [DOI] [PubMed] [Google Scholar]
- Charos G. S., Haffajee C. I., Gold R. L., Bishop R. L., Berkovits B. V., Alpert J. S. A theoretically and practically more effective method for interruption of ventricular tachycardia: self-adapting autodecremental overdrive pacing. Circulation. 1986 Feb;73(2):309–315. doi: 10.1161/01.cir.73.2.309. [DOI] [PubMed] [Google Scholar]
- De Belder M. A., Malik M., Ward D. E., Camm A. J. Pacing modalities for tachycardia termination. Pacing Clin Electrophysiol. 1990 Feb;13(2):231–248. doi: 10.1111/j.1540-8159.1990.tb05074.x. [DOI] [PubMed] [Google Scholar]
- Fromer M., Gloor H., Kus T., Shenasa M. Clinical experience with a new software-based antitachycardia pacemaker for recurrent supraventricular and ventricular tachycardias. Pacing Clin Electrophysiol. 1990 Jul;13(7):890–899. doi: 10.1111/j.1540-8159.1990.tb02126.x. [DOI] [PubMed] [Google Scholar]
- Griffin J. C., Sweeney M. The management of paroxysmal tachycardias using the Cybertach-60. Pacing Clin Electrophysiol. 1984 Nov;7(6 Pt 2):1291–1295. doi: 10.1111/j.1540-8159.1984.tb05697.x. [DOI] [PubMed] [Google Scholar]
- Holt P., Crick J. C., Sowton E. Antitachycardia pacing: a comparison of burst overdrive, self-searching and adaptive table scanning programs. Pacing Clin Electrophysiol. 1986 Jul;9(4):490–497. doi: 10.1111/j.1540-8159.1986.tb06604.x. [DOI] [PubMed] [Google Scholar]
- Jordaens L. J., Vandenbogaerde J. F., Van de Bruaene P., De Buyzere M. Transesophageal echocardiography for insertion of a physiological pacemaker in early pregnancy. Pacing Clin Electrophysiol. 1990 Aug;13(8):955–957. doi: 10.1111/j.1540-8159.1990.tb02138.x. [DOI] [PubMed] [Google Scholar]
- Li C. K., Shandling A. H., Nolasco M., Thomas L. A., Messenger J. C., Warren J. Atrial automatic tachycardia-reversion pacemakers: their economic viability and impact on quality-of-life. Pacing Clin Electrophysiol. 1990 May;13(5):639–645. doi: 10.1111/j.1540-8159.1990.tb02081.x. [DOI] [PubMed] [Google Scholar]
- McComb J. M., Jameson S., Bexton R. S. Atrial antitachycardia pacing in patients with supraventricular tachycardia: clinical experience with the Intertach pacemaker. Pacing Clin Electrophysiol. 1990 Dec;13(12 Pt 2):1948–1954. doi: 10.1111/j.1540-8159.1990.tb06922.x. [DOI] [PubMed] [Google Scholar]
- Nathan A., Hellestrand K., Ward D., Spurrell R., Camm J. Rate-related accelerating (autodecremental) atrial pacing for reversion of paroxysmal supraventricular tachycardia. J Electrocardiol. 1982 Jan;15(1):77–84. doi: 10.1016/s0022-0736(82)80048-4. [DOI] [PubMed] [Google Scholar]
- Spurrell R. A., Nathan A. W., Bexton R. S., Hellestrand K. J., Nappholz T., Camm A. J. Implantable automatic scanning pacemaker for termination of supraventricular tachycardia. Am J Cardiol. 1982 Mar;49(4):753–760. doi: 10.1016/0002-9149(82)91955-5. [DOI] [PubMed] [Google Scholar]
- den Dulk K., Kersschot I. E., Brugada P., Wellens H. J. Is there a universal antitachycardia pacing mode? Am J Cardiol. 1986 Apr 15;57(11):950–955. doi: 10.1016/0002-9149(86)90737-x. [DOI] [PubMed] [Google Scholar]