Abstract
Objective—To assess the accuracy of measurement of area of the lumen, and sensitivity, and specificity of detection of atheroma in coronary arteries in vitro with a commercially available 20 MHz intravascular ultrasound system.
Setting—A teaching hospital department of cardiology with the support of the department of cardiovascular pathology.
Procedure—10 segments of coronary artery were removed from cadaver hearts. Intravascular ultrasound imaging was performed at fixed levels and the vessels were then sectioned and photographed before histological preparation. An independent blinded observer measured luminal area and assessed the presence of atheroma on the intravascular ultrasound images of 76 vessel sections (304 quadrants). The sensitivity and specificity of detection of atheroma was assessed in comparison with the histologically prepared sections. Luminal areas from intravascular ultrasound, photographs of cross sections of the vessels and histological sections were compared with the technique of limits of agreement.
Results—Overall 36% of the 304 quadrants studied histologically had identifiable atheroma. Intravascular ultrasound sensitivity for atheroma was 0·593 and the specificity was 0·839. The positive predictive value was 0·674, and the relative risk 3·139. Values for area of the vessel lumen were on average 9·4 mm2 (confidence interval (CI) 8·6–10·2 mm2) larger than those measured from photographs and 10·7 (CI 9·8–11·6 mm2) larger than those measured from the histological sections.
Conclusions—The intravascular ultrasound system assessed in this study significantly overestimated coronary vessel luminal area and had low sensitivity and specificity for detection of atheroma. Improvements in image resolution are required before this system can provide useful information on coronary artery size and morphology.
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Bland J. M., Altman D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986 Feb 8;1(8476):307–310. [PubMed] [Google Scholar]
- Bom N., ten Hoff H., Lancée C. T., Gussenhoven W. J., Bosch J. G. Early and recent intraluminal ultrasound devices. Int J Card Imaging. 1989;4(2-4):79–88. doi: 10.1007/BF01745137. [DOI] [PubMed] [Google Scholar]
- Crowley R. J., von Behren P. L., Couvillon L. A., Jr, Mai D. E., Abele J. E. Optimized ultrasound imaging catheters for use in the vascular system. Int J Card Imaging. 1989;4(2-4):145–151. doi: 10.1007/BF01745144. [DOI] [PubMed] [Google Scholar]
- DeRouen T. A., Murray J. A., Owen W. Variability in the analysis of coronary arteriograms. Circulation. 1977 Feb;55(2):324–328. doi: 10.1161/01.cir.55.2.324. [DOI] [PubMed] [Google Scholar]
- Grondin C. M., Dyrda I., Pasternac A., Campeau L., Bourassa M. G., Lespérance J. Discrepancies between cineangiographic and postmortem findings in patients with coronary artery disease and recent myocardial revascularization. Circulation. 1974 Apr;49(4):703–708. doi: 10.1161/01.cir.49.4.703. [DOI] [PubMed] [Google Scholar]
- Gussenhoven E. J., Essed C. E., Lancée C. T., Mastik F., Frietman P., van Egmond F. C., Reiber J., Bosch H., van Urk H., Roelandt J. Arterial wall characteristics determined by intravascular ultrasound imaging: an in vitro study. J Am Coll Cardiol. 1989 Oct;14(4):947–952. doi: 10.1016/0735-1097(89)90471-3. [DOI] [PubMed] [Google Scholar]
- Gussenhoven E. J., Essed C. E., Lancée C. T., Mastik F., Frietman P., van Egmond F. C., Reiber J., Bosch H., van Urk H., Roelandt J. Arterial wall characteristics determined by intravascular ultrasound imaging: an in vitro study. J Am Coll Cardiol. 1989 Oct;14(4):947–952. doi: 10.1016/0735-1097(89)90471-3. [DOI] [PubMed] [Google Scholar]
- Gussenhoven W. J., Essed C. E., Frietman P., Mastik F., Lancée C., Slager C., Serruys P., Gerritsen P., Pieterman H., Bom N. Intravascular echographic assessment of vessel wall characteristics: a correlation with histology. Int J Card Imaging. 1989;4(2-4):105–116. doi: 10.1007/BF01745140. [DOI] [PubMed] [Google Scholar]
- Hodgson J. M., Graham S. P., Savakus A. D., Dame S. G., Stephens D. N., Dhillon P. S., Brands D., Sheehan H., Eberle M. J. Clinical percutaneous imaging of coronary anatomy using an over-the-wire ultrasound catheter system. Int J Card Imaging. 1989;4(2-4):187–193. doi: 10.1007/BF01745149. [DOI] [PubMed] [Google Scholar]
- Kalbfleisch S. J., McGillem M. J., Simon S. B., DeBoe S. F., Pinto I. M., Mancini G. B. Automated quantitation of indexes of coronary lesion complexity. Comparison between patients with stable and unstable angina. Circulation. 1990 Aug;82(2):439–447. doi: 10.1161/01.cir.82.2.439. [DOI] [PubMed] [Google Scholar]
- Liu M. W., Roubin G. S., King S. B., 3rd Restenosis after coronary angioplasty. Potential biologic determinants and role of intimal hyperplasia. Circulation. 1989 Jun;79(6):1374–1387. doi: 10.1161/01.cir.79.6.1374. [DOI] [PubMed] [Google Scholar]
- Mallery J. A., Tobis J. M., Griffith J., Gessert J., McRae M., Moussabeck O., Bessen M., Moriuchi M., Henry W. L. Assessment of normal and atherosclerotic arterial wall thickness with an intravascular ultrasound imaging catheter. Am Heart J. 1990 Jun;119(6):1392–1400. doi: 10.1016/s0002-8703(05)80190-5. [DOI] [PubMed] [Google Scholar]
- Moriuchi M., Tobis J. M., Mahon D., Gessert J., Griffith J., McRae M., Moussabeck O., Henry W. L. The reproducibility of intravascular ultrasound imaging in vitro. J Am Soc Echocardiogr. 1990 Nov-Dec;3(6):444–450. doi: 10.1016/s0894-7317(14)80360-x. [DOI] [PubMed] [Google Scholar]
- Nissen S. E., Grines C. L., Gurley J. C., Sublett K., Haynie D., Diaz C., Booth D. C., DeMaria A. N. Application of a new phased-array ultrasound imaging catheter in the assessment of vascular dimensions. In vivo comparison to cineangiography. Circulation. 1990 Feb;81(2):660–666. doi: 10.1161/01.cir.81.2.660. [DOI] [PubMed] [Google Scholar]
- Pandian N. G., Kreis A., O'Donnell T. Intravascular ultrasound estimation of arterial stenosis. J Am Soc Echocardiogr. 1989 Nov-Dec;2(6):390–397. doi: 10.1016/s0894-7317(89)80040-9. [DOI] [PubMed] [Google Scholar]
- Pandian N. G., Kreis A., Weintraub A., Motarjeme A., Desnoyers M., Isner J. M., Konstam M., Salem D. N., Millen V. Real-time intravascular ultrasound imaging in humans. Am J Cardiol. 1990 Jun 1;65(20):1392–1396. doi: 10.1016/0002-9149(90)91334-3. [DOI] [PubMed] [Google Scholar]
- Potkin B. N., Bartorelli A. L., Gessert J. M., Neville R. F., Almagor Y., Roberts W. C., Leon M. B. Coronary artery imaging with intravascular high-frequency ultrasound. Circulation. 1990 May;81(5):1575–1585. doi: 10.1161/01.cir.81.5.1575. [DOI] [PubMed] [Google Scholar]
- Pratt C. M., Hallstrom A., Theroux P., Romhilt D., Coromilas J., Myles J. Avoiding interpretive pitfalls when assessing arrhythmia suppression after myocardial infarction: insights from the long-term observations of the placebo-treated patients in the Cardiac Arrhythmia Pilot Study (CAPS) J Am Coll Cardiol. 1991 Jan;17(1):1–8. doi: 10.1016/0735-1097(91)90697-8. [DOI] [PubMed] [Google Scholar]
- Tobis J. M., Mallery J. A., Gessert J., Griffith J., Mahon D., Bessen M., Moriuchi M., McLeay L., McRae M., Henry W. L. Intravascular ultrasound cross-sectional arterial imaging before and after balloon angioplasty in vitro. Circulation. 1989 Oct;80(4):873–882. doi: 10.1161/01.cir.80.4.873. [DOI] [PubMed] [Google Scholar]
- Tobis J. M., Mallery J., Mahon D., Lehmann K., Zalesky P., Griffith J., Gessert J., Moriuchi M., McRae M., Dwyer M. L. Intravascular ultrasound imaging of human coronary arteries in vivo. Analysis of tissue characterizations with comparison to in vitro histological specimens. Circulation. 1991 Mar;83(3):913–926. doi: 10.1161/01.cir.83.3.913. [DOI] [PubMed] [Google Scholar]
- ten Hoff H., Korbijn A., Smith T. H., Klinkhamer J. F., Bom N. Imaging artifacts in mechanically driven ultrasound catheters. Int J Card Imaging. 1989;4(2-4):195–199. doi: 10.1007/BF01745150. [DOI] [PubMed] [Google Scholar]