Iatrogenic Hypoxemia and Atrial Septal Defect Due to Electrical Storm Ablation After Left Ventricular Assist Device: A Case Report

Ryan Wang 1, Darcy J Mainville 2, Alexandra Vacaru 1,, Ioana Pasca 3,2
Editors: Alexander Muacevic, John R Adler
PMCID: PMC10287845  PMID: 37362482

Abstract

A 59-year-old male with an implantable cardiac defibrillator, left ventricular assist device, and refractory ventricular tachycardia presented with hypoxemia due to a post-ablation iatrogenic atrial septal defect. Left ventricular assist devices generate pressure gradients that may exacerbate intracardiac shunts and can precipitate significant hypoxemia.

Keywords: atrial septal defect, implantable cardiac defibrillator, left ventricular assist device, transesophageal echocardiogram, ventricular tachycardia, arterial blood gas, hypoxemia, catheter ablation

Introduction

Electrical storm is hallmarked by recurrent episodes of ventricular tachycardia (VT), ventricular fibrillation, or multiple appropriate shocks from an implantable cardiac defibrillator (ICD) over a short period, typically 24 hours. Management of electrical storm is challenging, and primary intervention involves the identification and correction of precipitating factors, such as myocardial ischemia as well as arrhythmia suppression with antiarrhythmics and beta-blockers. Refractory arrhythmias may warrant catheter ablation [1,2]. Although mechanical circulatory support may be used in the management of electrical storm, ventricular arrhythmias may occur after left ventricular assist device (LVAD) implantation. One study showed that 23% of patients developed at least one ventricular arrhythmia after LVAD placement. Approximately 4% of LVAD patients experience five or more episodes of arrhythmia events in the immediate post-operative period [3]. Catheter ablation is not only feasible with LVAD but is safe and effective. Even in patients with incessant VT, catheter ablation reduces VT recurrence [3,4]. We present the management of a patient with LVAD and electrical storm and describe the development of refractory hypoxemia after catheter ablation. Health Insurance Portability and Accountability consent was obtained from the patient for publication of this study.

Case presentation

A 59-year-old male with acute left anterior descending ST-elevation myocardial infarction, failed percutaneous coronary intervention, left ventricular ejection fraction of 20%, and recurrent VT was transferred to our institution for a higher level of care. After a failed percutaneous coronary intervention, he developed persistent monomorphic VT and recurrent ICD shocks refractory to lidocaine and amiodarone.

He was evaluated for LVAD implantation for ischemic cardiomyopathy and refractory VT. Concurrent epicardial cryoablation was planned to address recurrent ventricular arrhythmias. The HeartMate 3 LVAD (Chicago, IL: Abbott) was implanted, and cryoablation was performed adjacent to the LVAD core. Post-operatively, he was started on procainamide up to a maximum of 4 mg/min, but he continued to experience recurrent episodes of VT. Electrophysiology recommended endocardial catheter ablation.

Procedure

Left ventricular access was obtained via a standard transseptal approach. After initial transseptal access was obtained, extensive ablation of the left and right ventricular aspects of the interventricular septum was performed. After ablation, the clinical VT was no longer inducible.

Post-procedure, he remained intubated on mechanical ventilation with a large alveolar-arterial gradient and a P/F ratio of 93 on 100% FiO2. His arterial blood gas (ABG) on 100% FiO2 was pH 7.40, PaCO2 33 mmHg, PaO2 94 mmHg, and HCO3 19.9 mEq (Table 1). A bedside transesophageal echocardiogram (TEE) demonstrated a 3 mm transseptal puncture atrial septal defect (ASD) with moderate right to left shunt across the defect (Figure 1). His hypoxemia was attributed to the right-left shunt and closure of the defect was advised.

Table 1. Arterial blood gas pre- and post-ASD closure.

ASD: atrial septal defect; BE: base excess

Variables pH PCO2 (mmHg) PO2 (mmHg) HCO3 (mEq) BE (mmol/L) Lactate (mmol/L) FiO2
Pre-ASD closure 7.40 33 94 19.9 3.7 1.0 100%
Post-ASD closure 7.34 39 198 20.6 4.1 1.0 100%

Figure 1. TEE visualization of ASD.

Figure 1

TEE: transesophageal echocardiogram; ASD: atrial septal defect

He returned to the cardiac catheterization laboratory for ASD closure 48 hours post-ablation. A 30-mm Cardioform device (Newark, DE: W.L. Gore and Associates) was deployed under fluoroscopy and TEE guidance, and the ASD was closed. Prior to device deployment, he was on 100% FiO2 with a PaO2 of 100 mmHg. After ASD closure, his PaO2 increased to 200 mmHg on 100% FiO2. His ABG with the same ventilator settings on 100% FiO2 after device deployment and closure of his ASD was pH 7.34, PaCO2 39 mmHg, PaO2 198 mmHg, and HCO3 20.6 mEq (Table 1). His right atrial pressure was 13 mmHg pre-device placement and 14 mmHg post-device placement. Left atrial pressure was 8 mmHg. His ICD generator was also changed as the battery was depleted from delivery of recurrent shocks.

After ASD closure, hypoxemia resolved and arterial saturation improved. He was extubated to a high-flow nasal cannula 48 hours later. His hemodynamic status continued to improve, and he was discharged to acute rehabilitation.

Discussion

Predominant strategies for the management of electrical storm include antiarrhythmic medical therapy, beta-blockers, and catheter ablation [1,5]. Some studies have demonstrated refractory electrical storm managed with mechanical circulatory support, but VT may also be exacerbated by LVAD placement. LVAD implantation may be associated with electrolyte changes, repolarization abnormalities, myocardial scar, and ventricular chamber collapse, which may precipitate arrhythmia [6].

When electrical storm is refractory to medical therapy, endocardial ablation may be indicated. Although catheter ablation has been described in patients with LVAD, the complications uniquely associated with catheter ablation in the LVAD population have not [3,4]. In a limited series, Sacher et al. described the outcomes of catheter ablation in 34 patients with LVAD. Among these, one patient developed cardiogenic shock with acidosis, one experienced a transient ischemic attack, one experienced a stroke eight days after ablation, and one required a blood transfusion with two units of packed red blood cells for a groin hematoma [7].

The patient presented suffered from refractory electrical storm after LVAD implantation and epicardial ablation. An endocardial approach was required to address the residual arrhythmogenic foci. Transseptal puncture is routinely performed for left ventricular access, and the small residual ASD usually self-seals. Persistent iatrogenic ASD has been described after pulmonary vein isolation for atrial fibrillation [8,9]. However, the majority of iatrogenic ASDs are associated with inconsequential left-to-right shunt and usually close within 12 months [10]. A few studies have described the closure of ASD after LVAD [11,12]. However, our case is one of the first to describe refractory hypoxemia from an iatrogenic ASD in a patient with LVAD after catheter ablation. 

Persistence of a right-to-left shunt requires a significant pressure gradient between the right and left atrium. The LVAD offloads the left ventricle and decreases left atrial pressure. When unloading of the left side of the heart is coupled with elevated right atrial pressures, the pressure gradient may be sufficient to support a persistent right-to-left shunt. Our patient likely had elevated right atrial pressure from moderate tricuspid regurgitation. Furthermore, implantation of an LVAD significantly increases the workload of the right ventricle. In patients with cardiomyopathy and reduced cardiac output, the right ventricle has acclimated to a lower cardiac output state. When an LVAD is implanted, the right ventricle is forced to adapt to much higher cardiac output and may precipitate right ventricular dysfunction and additional increases in right atrial pressure [13]. Transseptal puncture requires careful consideration in the LVAD population, persistent hypoxemia should raise concern for the presence of a right-to-left shunt [14]. We advocate for continuous oximetry and careful attention to oxygen requirements post-procedure. Consider TEE evaluation in patients with unexplained hypoxemia.

Conclusions

Endocardial catheter ablation is a safe and effective treatment for electrical storm. Although mechanical circulatory support has been used to reduce the burden of arrhythmia, LVAD may also be associated with persistent ventricular arrhythmias. When a transseptal puncture is performed in a patient with an LVAD, conditions are optimal to support persistence of a right-to-left shunt, and we recommend post-ablation TEE to evaluate for intracardiac shunts. Right-to-left shunts may cause significant hypoxemia and warrant emergent closure. LVAD implantation dramatically alters intracardiac pressures and can exacerbate or even reverse intracardiac shunts.

The authors have declared that no competing interests exist.

Human Ethics

Consent was obtained or waived by all participants in this study

References

  • 1.The evaluation and management of electrical storm. Eifling M, Razavi M, Massumi A. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3066819/ Tex Heart Inst J. 2011;38:111–121. [PMC free article] [PubMed] [Google Scholar]
  • 2.Electrical storm in patients with an implanted defibrillator: a matter of definition. Israel CW, Barold SS. Ann Noninvasive Electrocardiol. 2007;12:375–382. doi: 10.1111/j.1542-474X.2007.00187.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Catheter ablation for ventricular tachyarrhythmias in patients supported by continuous-flow left ventricular assist devices. Garan AR, Iyer V, Whang W, et al. ASAIO J. 2014;60:311–316. doi: 10.1097/MAT.0000000000000061. [DOI] [PubMed] [Google Scholar]
  • 4.Endocardial catheter ablation of ventricular tachycardia in patients with ventricular assist devices. Dandamudi G, Ghumman WS, Das MK, Miller JM. Heart Rhythm. 2007;4:1165–1169. doi: 10.1016/j.hrthm.2007.05.029. [DOI] [PubMed] [Google Scholar]
  • 5.Venturing into ventricular arrhythmia storm: a systematic review and meta-analysis. Nayyar S, Ganesan AN, Brooks AG, Sullivan T, Roberts-Thomson KC, Sanders P. Eur Heart J. 2013;34:560–571. doi: 10.1093/eurheartj/ehs453. [DOI] [PubMed] [Google Scholar]
  • 6.The burden of ventricular arrhythmias following left ventricular assist device implantation. Griffin JM, Katz JN. Arrhythm Electrophysiol Rev. 2014;3:145–148. doi: 10.15420/aer.2014.3.3.145. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Characteristics of ventricular tachycardia ablation in patients with continuous flow left ventricular assist devices. Sacher F, Reichlin T, Zado ES, et al. Circ Arrhythm Electrophysiol. 2015;8:592–597. doi: 10.1161/CIRCEP.114.002394. [DOI] [PubMed] [Google Scholar]
  • 8.Persistent iatrogenic atrial septal defects after pulmonary vein isolation: long-term follow-up with contrast transesophageal echocardiography. Davies A, Gunaruwan P, Collins N, Barlow M, Jackson N, Leitch J. J Interv Card Electrophysiol. 2017;48:99–103. doi: 10.1007/s10840-016-0193-7. [DOI] [PubMed] [Google Scholar]
  • 9.Persistence of iatrogenic atrial septal defect after pulmonary vein isolation - an underestimated risk? Hammerstingl C, Lickfett L, Jeong KM, et al. Am Heart J. 2006;152:362–365. doi: 10.1016/j.ahj.2006.04.034. [DOI] [PubMed] [Google Scholar]
  • 10.The incidence and long-term clinical outcome of iatrogenic atrial septal defects secondary to transseptal catheterization with a 12F transseptal sheath. Singh SM, Douglas PS, Reddy VY. Circ Arrhythm Electrophysiol. 2011;4:166–171. doi: 10.1161/CIRCEP.110.959015. [DOI] [PubMed] [Google Scholar]
  • 11.Percutaneous closure of patent foramen ovale for refractory hypoxemia after HeartMate II left ventricular assist device placement. Kapur NK, Conte JV, Resar JR. https://pubmed.ncbi.nlm.nih.gov/17827518/ J Invasive Cardiol. 2007;19:268–270. [PubMed] [Google Scholar]
  • 12.Transcatheter closure of patent foramen ovale after left ventricular assist device implantation: intraoperative decision making. Kavarana MN, Rahman FA, Recto MR, Dowling RD. J Heart Lung Transplant. 2005;24 doi: 10.1016/j.healun.2005.02.007. [DOI] [PubMed] [Google Scholar]
  • 13.The ABCs of left ventricular assist device echocardiography: a systematic approach. Ammar KA, Umland MM, Kramer C, et al. Eur Heart J Cardiovasc Imaging. 2012;13:885–899. doi: 10.1093/ehjci/jes090. [DOI] [PubMed] [Google Scholar]
  • 14.Electrical storm after correction of an uncomplicated congenital atrial septal defect in an adult: a case report. Liang Y, Hei F, Guan Y. BMC Cardiovasc Disord. 2021;21 doi: 10.1186/s12872-021-02164-6. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Cureus are provided here courtesy of Cureus Inc.

RESOURCES