Infection responsive coatings to reduce biofilm formation and encrustation of urinary catheters

Anthony J Slate 1,, Ocean E Clarke 2, Mina Kerio 3, Jonathan Nzakizwanayo 4, Bhavik Anil Patel 5, Brian V Jones 6,
PMCID: PMC10294568  PMID: 37296327

Abstract

Aims

The care of patients undergoing long-term urethral catheterization is frequently complicated by Proteus mirabilis infection. This organism forms dense, crystalline biofilms, which block catheters leading to serious clinical conditions. However, there are currently no truly effective approaches to control this problem. Here, we describe the development of a novel theranostic catheter coating, to simultaneously provide early warning of blockage, and actively delay crystalline biofilm formation.

Methods and Results

The coating comprises of a pH sensitive upper polymer layer (poly(methyl methacrylate-co-methacrylic acid); Eudragit S 100®) and a hydrogel base layer of poly(vinyl alcohol), which is loaded with therapeutic agents (acetohydroxamic acid or ciprofloxacin hydrochloride) and a fluorescent dye, 5(6)-carboxyfluorescein (CF). The elevation of urinary pH due to P. mirabilis urease activity results in the dissolution of the upper layer and release of cargo agents contained in the base layer. Experiments using in vitro models, which were representative of P. mirabilis catheter-associated urinary tract infections, demonstrated that these coatings significantly delay time taken for catheters to block. Coatings containing both CF dye and ciprofloxacin HCl were able to provide an average of ca. 79 h advanced warning of blockage and extend catheter lifespan ca. 3.40-fold.

Conclusions

This study has demonstrated the potential for theranostic, infection-responsive coatings to form a promising approach to combat catheter encrustation and actively delay blockage.

Keywords: catheter-associated urinary tract infection, crystalline biofilm, Proteus mirabilis, theranostic, infection responsive coating


Significance and Impact of Study.

This study builds on previous research conducted in our group and further optimizes the proposed, infection responsive, urinary catheter coating. For the first time, therapeutic agents have been incorporated into this theranostic coating design. This resulted in an advanced warning of blockage (e.g., dye release) and actively delays catheter blockage times (e.g., therapeutic agent release into the bladder). Such coatings represent a major advance in the treatment of catheter-associated urinary tract infections (CAUTIs).

Introduction

Indwelling urinary catheters are deployed in a wide variety of circumstances related to the management of bladder dysfunction, as well as the management and monitoring of urinary output (Stickler, 2008). The duration of catheterization can range from short-term use in clinical settings (up to 7 days), to long-term use (> 28 days) in patients cared for in the community (Stickler, 2008; Loveday et al., 2014; Stickler, 2014; Shackley et al., 2017). In keeping with the prolific use of these devices, catheter-associated urinary tract infections (CAUTIs) continue to be among the most prevalent healthcare-associated infections in both hospital and community care settings; with up to 80 % of all healthcare-associated urinary tract infections (UTIs) related to the use of indwelling catheters (Parker et al., 2017; Clark and Wright, 2019; Yakusheva et al., 2019). Even with the most scrupulous nursing care and rigorous application of sterile closed drainage systems, bacteriuria is almost inevitable in patients undergoing long-term catheterization (> 28 days), and this group of patients are particularly vulnerable to resulting complications (Nicolle, 2014).

In particular, the encrustation and blockage of catheters is a common and serious complication, which may be experienced by up to 50 % of individuals undergoing long-term catheterization (Nicolle, 2014). If undetected, blockage can lead to the reflux of infected urine to the upper urinary tract, which may subsequently result in pyelonephritis, septicaemia, and endotoxic shock (Stickler and Zimakoff, 1994; Jordan et al., 2015). Proteus mirabilis is a common pathogen of the catheterized urinary tract, associated with up to 44 % of CAUTIs, and is the primary cause of catheter blockage (O'Hara et al., 2000; Jacobsen et al., 2008; Nicolle, 2014). This organism forms dense crystalline biofilms, which encrust catheters and occlude urine flow (Griffith et al., 1976; Cox and Hukins, 1989; Clapham et al., 1990; Dumanski et al., 1994; Stickler and Zimakoff, 1994; Morris et al., 1999, Jones et al., 2005). This process is driven by the production of a potent urease enzyme by P. mirabilis, which generates ammonia through the hydrolysis of urea present in urine, elevating urinary pH (Griffith et al., 1976). Under these alkaline conditions, the precipitation of calcium phosphate (apatite) and magnesium ammonium phosphate (struvite) occurs, producing microcrystalline aggregates that become incorporated within developing P. mirabilis biofilms (Hedelin et al., 1984; Cox and Hukins, 1989). Integration of crystal aggregates into biofilms further stabilizes and enhances their growth leading to the mineralization of biofilms, the formation of crystalline biofilm structures, and the eventual blockage of catheters (Griffith et al., 1976; Cox and Hukins, 1989; Clapham et al., 1990; Dumanski et al., 1994; Stickler and Zimakoff, 1994; Morris et al., 1999; Jones et al., 2005).

Despite the potentially severe consequences for patients, catheter blockage is typically not detected until more serious conditions arise. The initial bacterial colonization of the catheterized urinary tract is usually asymptomatic, and blockage can occur rapidly with little warning (Kohler‐Ockmore and Feneley, 1996; Tambyah and Maki, 2000; Stickler et al., 2006; Long et al., 2014; Stickler, 2014). This is particularly problematic for individuals cared for in the community where continual clinical surveillance is not feasible. Although a range of catheters with antimicrobial coatings are available, their ability to control infection during even short-term catheterization is poor, and currently all available catheter types are vulnerable to encrustation (Stickler, 2014; Nzakizwanayo et al., 2016; Cortese et al., 2018). Therefore, novel approaches, which provide advanced warning, or prevent infection and catheter blockage, are clearly needed and would considerably improve patient wellbeing.

To address this, we have recently developed a prototype infection-responsive catheter coating that can provide early warning of catheter blockage (Milo et al., 2016; Milo et al., 2017). This coating comprised a pH sensitive outer layer (Eudragit S 100 ®), which sealed a lower hydrogel basement layer (poly(vinyl alcohol); PVA) containing the self-quenching fluorescent dye 5(6)-carboxyfluorescein. The elevation of urinary pH by P. mirabilis results in degradation of the upper Eudragit S 100 ® layer and release of the fluorescent dye to give advanced warning of the potential for catheter blockage ( Figure. 1). Here, we expand on these initial studies, describing the further development of this coating technology to include delivery of therapeutic agents to actively delay catheter blockage. Evaluation of these novel coatings in representative models of CAUTI demonstrated the potential for these to provide a theranostic approach for the control of catheter blockage, by simultaneously treating infection (therapeutic) while providing advanced warning of catheter blockage (diagnostic).

Figure 1.

Figure 1.

Infection responsive coating design and function. (a) Schematic showing urease-mediated hydrolysis of urea to generate ammonia, which elevates urinary pH during P. mirabilis infection. (b) Overview of coating design and response to P. mirabilis infection. Elevation of urinary pH driven by P. mirabilis urease activity leads to dissolution of the upper Eudragit S 100® ‘trigger’ layer and release of cargo molecules contained in the lower hydrogel layer.

Materials and methods

Bacterial strains, media, and routine culture

Proteus mirabilis clinical strain RS1 (Royal Sussex County Hospital) was used throughout this study. All chemicals, reagents, and growth media were obtained from Fisher Scientific (UK), Merck (UK), and Oxoid (UK), unless otherwise stated. Bacteria were routinely cultured in Lysogeny Broth [LB (Miller): 5 g L−1 yeast extract, 10 g L−1 tryptone, 10 g L−1 sodium chloride] at 37 °C with shaking for 18 h; for the isolation of single colonies of P. mirabilis and the suppression of swarming motility, strains were grown on no salt-LB (NSLB) agar (5 g L−1 yeast extract, 10 g L−1 tryptone, 20 g L−1 agar).

Catheter coating protocol

Catheter coating production was adapted from protocols originally described by Milo et al., (2016), and coatings were applied to the external surface of the catheter from the catheter tip to the start of the retention balloon.

Salinization

The catheter surface was modified to promote the attachment of the hydrogel/polymer coating. This salinization process covered the surface with organofunctional alkoxysilane molecules, which increased the hydrophobicity of the catheter surface (forming covalent -Si-O-Si- bonds). The tips of the catheters were submerged in a solution of ammonia hydroxide (33 % v/v) and hydrogen peroxide (30 % v/v) (at a 1:1 ratio) for 30 minutes with sonication. The catheters were rinsed in sterile deionized water (thrice) and dried under a constant stream of nitrogen. The catheter tips were placed in 1 % v/v (3-aminopropyl)triethoxysilane (APTES) in dimethylformamide overnight (at room temperature). The catheters were washed three times in sterile deionized water and dried under a constant stream of nitrogen.

PVA Preparation

High molecular weight poly(vinyl alcohol) (PVA; 146,000 – 186,000g mol−1) was produced at 30 % w/v in sterile deionized water. This was heated to 97 °C with constant aggregation for 10 minutes. Once dissolved and cooled, the therapeutic payload was added at a 1:1 ratio. The final PVA concentration was 15 % w/v. If 5(6)-carboxyfluorescein (CF) was utilized this was added at a final concentration of 100 mM, whilst acetohydroxamic acid (AHA) and ciprofloxacin hydrochloride were incorporated at 5 mg mL−1.

Preparation of Eudragit S 100®

The pH sensitive, Eudragit S 100® polymer was prepared by combining, 109.1 mL acetone, 163.5 mL isopropanol, and 10.7 mL deionized water, this solution was then aliquoted into two equal volumes (141.65 mL). Eudragit polymer (15.6 g) was added to one of the diluent mixtures and homogenized at room temperature. To the other half of the diluent mixture, hydrated magnesium silicate (7.81 g, talc - an anti-tacking agent) and triethyl citrate (1.4 mL) were added. This was agitated for 10 minutes using a high shear mixer ( 1,000rpm). The solutions were then combined (slowly with constant agitation) and this was filtered (Grade 1, 85 mm; Whatman, UK) using a Büchner filter under vacuum to remove the hydrated magnesium silicate. This step was repeated three times. The polymer solution was stored at 4 °C until required.

Catheter coating

Foley urinary catheters (BARDIA® AQUAFIL® All Silicone Foley Catheter; Bard Limited, UK) were coated between the balloon and the tip (whilst deflated) with 1 mL of the prepared 15 % w/v PVA containing the therapeutic payload. The catheter eyelet was obstructed, during the coating process, to inhibit polymer blockage. This was then stored at − 20 °C overnight to promote cryogenic crosslinking. The catheters were thawed at room temperature for 4 h before being dip-coated with the secondary pH responsive polymer Eudragit S 100®. The dip coating procedure was carried out manually, with the catheters being vertically submerged into the polymer for 3 seconds followed by a 5-minute solvent evaporation time, to ensure full Eudragit S 100® coverage, this process was repeated 20 times. The catheters were dried at 4 °C overnight. After 18 h, the barrier blocking the eyelet was replaced and the catheters were coated using the aforementioned process a further 10 times. Once coated, the catheters were stored at 4 °C, overnight. This process was repeated once more. Following this, the catheter eyelet was unblocked and Eudragit S 100® was added a further five times to seal the PVA layer around the eyelet (total Eudragit S 100® coatings: 45). This coating ensured the therapeutic payload below was adequately sealed prior to experimentation. The catheters were then stored at 4 °C until required.

Growth curves

In a 96-well plate, wells containing 100 µL of tryptone soya broth (TSB) or TSB supplemented with 100 mM 5(6)-carboxyfluorescein were inoculated with 105 CFU mL−1 of P. mirabilis strain RS1 from a mid-log phase culture. Uninoculated wells served as negative controls. The plate was then incubated in a plate reader (Multiskan™ FC Microplate Reader) at 37 °C while shaking, and optical density (OD600 nm) was measured at 1 hour intervals over 24 hours. Data from three biological replicates (n = 3) were pooled to create a mean overall growth curve.

Artificial urine

The artificial urine (AU) medium used throughout this study has previously been described (Nzakizwanayo et al., 2019). Briefly, the AU was initially prepared as a 5 × concentrated stock solution containing; sodium disulfate (11.5 g L−1), magnesium chloride (hexahydrate) (3.25 g L−1), sodium chloride (23 g L−1), trisodium citrate (3.25 g L−1), sodium oxalate (0.1 g L−1), potassium dihydrogen orthophosphate (14 g L−1), potassium chloride (8 g L−1), ammonium chloride (5 g L−1), gelatine (25 g L−1), and tryptone soya broth (5 g L−1). Stock solutions of urea (125 g) and calcium chloride dihydrate (3.25 g) were sterilized separately in 400 mL sterile deionized water using a Nalgene Vacuum Filter System (0.44 μm nitrocellulose membrane; Sartorius, UK), while other components were sterilized by autoclaving. For use in bladder models, all components were combined and diluted to 1 × strength using sterile deionized water (3.6 L), the final pH was adjusted to 6.1 prior to experimentation.

In vitro models of the catheterized urinary tract

Bladder models were conducted as previously described (Nzakizwanayo et al., 2019). The bladder models consisted of a double-walled glass chamber (50 mm internal chamber diameter) maintained at 37 °C by a water jacket supplied from a circulating water bath. All silicone foley catheters (BARDIA® AQUAFIL®; Bard Limited, UK) which were either uncoated or coated with the aforementioned polymer design were inserted into the 'bladder' and retention balloons inflated with 10 mL sterile saline. The catheter was then attached to a drainage bag to form a sterile, closed drainage system. Artificial urine was supplied to the 'bladder' at a constant flow rate of ca. 0.75 mL minute−1 (calculated individually for each model). The in vitro bladder model systems were inoculated with 10 mL of a P. mirabilis culture containing ∼109 CFU mL−1 and bacterial cells were allowed to establish within the model for 1 h before flow was activated. The number of viable cells in the residual urine and the pH of the medium were measured at the start of the experiment and following catheter blockage. The pH was also measured at 4 h to determine the effect the coating had on the AU.

Scanning electron microscopy (SEM)

To visualize catheter encrustation and coating structure/dissolution SEM was conducted with minimal sample preparation as per Holling et al., (2014). Coated Foley catheters were removed from the in vitro bladder models after 10 h incubation. Samples were visualized after a 10 h incubation period as this allowed all coatings (and controls) to be visualized prior to blockage, but after a significant time of elevated pH, in order to determine the extent of catheter coating degradation. Catheter sections were recovered for visualization by making initial incisions through the coating directly below the eyelet, and a second distal incision to remove a 1 cm section. Samples were mounted on aluminium stubs and placed in a desiccator containing silica gel beads (2 - 5mm), overnight. SEM was conducted using a Hitachi SU3900 large chamber, variable pressure SEM (Hitachi High-Tech Corporation, Japan).

Statistical analysis

Statistical analysis was conducted by performing one-way or two-way ANOVAs as appropriate, coupled with Dunnett’s correction for multiple comparisons, using GraphPad Prism (version 8.4.3; GraphPad Software, USA).

Results

Impact of optimized coatings on catheter blockage

The initial optimization of our coating production involved increasing the proportion of PVA in the hydrogel basement layer and improvements to techniques for application of the upper pH responsive Eudragit S 100 ® layer. To determine the baseline performance of these optimized coatings, we compared the blockage times of catheters with ‘empty’ coatings devoid of cargo molecules, and those containing CF dye only, to standard uncoated All Silicone Foley catheters ( Figure. 2).

Figure 2.

Figure 2.

Effect of optimized coatings on P. mirabilis -mediated catheter blockage. (a) Results of in vitro bladder model experiments simulating established infection (109 CFU mL−1) showing impact of optimized coatings on catheter blockage, pH elevation, and survival of P. mirabilis strain RS1. Measurement of pH and enumeration of remaining viable cells in residual urine was conducted at catheter blockage. No Coating = standard All Silicone Foley catheters without coating; Empty = All Silicone Foley catheters with coating lacking any cargo molecules in the hydrogel layer; 100 mM CF Dye = All Silicone Foley catheters with coatings containing 100 mM CF dye in the hydrogel layer. Data represents the mean of five biological replicates ( = 5) and error bars show standard error of the mean, * P ≥ .05, ***P ≥ .0001 compared to uncoated catheters. (b) Impact of 100 mM 5(6)-carboxyfluorescein on growth of P. mirabilis RS1 in TSB. Data represents the mean of three biological replicates and error bars show standard error of the mean. * - ****P ≥ .05 - .00001. (c) Scanning electron micrographs of catheter coatings after 10 h incubation in sterile uninoculated bladder models (No P. mirabilis ), or models inoculated with P. mirabilis (+ P. mirabilis ).

In representative models of P. mirabilis CAUTI, simulating an established high-level infection (109 CFU mL−1), empty coatings alone significantly increased time to blockage by an average of 7.8 h compared with uncoated catheters ( Figure. 2a). The inclusion of 100 mM of 5(6)-carboxyfluorescein in the hydrogel layer of the coating provided further increases in catheter blockage time, on average 16.9 h longer than for uncoated catheters ( Figure. 2a). Coating application did not alter the flow rate of urine through the catheters, the average flow rate of the catheters throughout this study was 0.88 (± 0.012) mL minute−1.

The pH of residual urine at the time of catheter blockage was not significantly different in any models with coated catheters compared to uncoated controls ( Figure. 2a). However, a significant reduction in viable cell numbers was observed in models with coatings containing the CF dye ( Figure. 2a), and further investigation revealed a significant impact of the CF dye on P. mirabilis growth at concentrations used in the coatings ( Figure. 2b).

Consistent with our previous work (Milo et al., 2016), coatings containing CF dye also elicited a visible urine colour change, which was most pronounced ∼ 4 - 5h after activation of models inoculated with P. mirabilis , providing, ∼ 29 h advance warning of catheter blockage on average. SEM analysis of catheter sections after 10 h of incubation in the bladder models confirmed stability of coatings in uninoculated control models, and the deterioration of coatings in models infected with P. mirabilis ( Figure. 2c).

Incorporation of therapeutic agents into catheter coatings

To evaluate the ability of our infection responsive coatings to deliver therapeutic agents, we subsequently incorporated either the urease inhibitor acetohydroxamic acid (AHA, marketed commercially as Lithostat®), or the antibiotic ciprofloxacin hydrocholoride at 5 mg mL−1 into the PVA layer. Release of this amount of each drug into the residual bladder model urine was estimated to generate concentrations well above the indicated therapeutic range for Lithostat® [∼ 30µg mL−1; (Mission Pharmacal Company 2023)], and the P. mirabilis strain RS1 MIC for ciprofloxacin HCl (0.125 µg mL−1). Coating variants containing CF dye in combination with therapeutic agents, as well as those containing therapeutic agents alone, were produced and their impact on P. mirabilis crystalline biofilm formation was again evaluated in our infection models.

The inclusion of AHA in coatings resulted in no significant increase in time to blockage compared to the empty coating, but incorporation of AHA in combination with CF dye resulted in a nominal but statistically significant increase in blockage time ( Figure. 3a). In contrast, incorporation of ciprofloxacin HCl in coatings alone or in combination with CF dye, produced notable increases in catheter blockage time and increased catheter lifespan by 2.82-fold and 3.40-fold, respectively, compared to the empty coatings ( Figure. 3a). In coatings containing CF dye in conjunction with therapeutic agents, a visual indication of coating activation was also evident after model activation, which provided an average of ∼ 36 h advance warning of blockage for coatings containing AHA and ∼ 79 h for coatings containing ciprofloxacin HCl. No significant differences in urinary pH or viable bacterial counts were observed at time of blockage for any coating evaluated (compared to empty coatings), but viable cells counts in models with ciprofloxacin HCl coatings were lower than for models with ciprofloxacin HCl and CF dye containing coatings ( Figure. 3b and c).

Figure 3.

Figure 3.

Impact of theranostic coatings on catheter blockage. Theranostic coatings containing CF dye, and either the urease inhibitor acetohydroxamic acid or the antibiotic ciprofloxacin HCl, were evaluated using in vitro bladder models. (a) Time taken for catheters to block; (b) pH of residual bladder model urine at time of catheter blockage; and (c) viable cell counts in residual urine at time of catheter blockage . Empty = All Silicone Foley catheters with coating lacking any cargo molecules in the hydrogel layer; CF Dye ONLY = All Silicone Foley catheters with coatings containing 100 mM 5(6)-carboxyfluorescein dye (only) in the hydrogel layer; AHA = acetohydroxamic acid (Lithostat®) at 5 mg mL−1 in the coating; CIP = ciprofloxacin at 5 mg mL−1 in the coating; and CF dye = 5(6)-carboxyfluorescein dye at 100 mM in coating. Data represents the mean of five biological replicates ( n = 5) and error bars show standard error of the mean. * P ≥ .05, ****P ≥ .00001, compared to empty coatings.

Discussion

Although a wide range of approaches to control biofilm formation on urethral catheters have been described, and often appear to provide adequate protection against many common species of uropathogenic bacteria, development of effective strategies to control encrustation by P. mirabilis poses a particular challenge (Stickler, 2008, 2014; Milo et al., 2019; Pelling et al., 2019). In this study, our prototype theranostic coatings were able to provide both advanced warning of catheter blockage as well as actively increasing the time taken for P. mirabilis to block catheters. While both therapeutic agents tested in our coatings may be administered orally for the treatment of UTIs, localized delivery options have the potential to avoid many of the systemic side effects associated with these drugs and provide a more effective way to combat CAUTIs (Benderev, 1988; Zacchè et al., 2015; Wan et al., 2018; Marei et al., 2021; Moussa et al., 2021).

Due to the key role of the P. mirabilis urease enzyme in the formation of crystalline biofilms and blockage of urinary catheters, inhibition of this enzyme is an attractive approach to control catheter encrustation by this organism. Urease inhibitors have previously been explored as potential therapeutic treatments to reduce catheter encrustation, and have shown some promise for control of crystalline biofilm formation (Morris and Stickler, 1998; Follmer, 2010). In particular, the structural analogue of urea, AHA is currently marketed under the name Lithostat®, and indicated for use in the treatment of UTIs involving urease-producing species and stone formation. However, whilst AHA has demonstrated effective inhibition of urolithiasis in some clinical trials, it has severe side effects for many patients when taken orally. These include headaches, gastrointestinal upset, thrombophlebitis, and dermatitis, which limit its clinical application; it is currently only recommended as a treatment of last resort (Williams et al., 1984; Griffith et al., 1988; Georgi, 2004; York et al., 2015; Zisman, 2017; Abou Chakra et al., 2020).

However, in our experiments, delivery of AHA via our coating system did not provide significant protection against catheter blockage. The failure of Lithostat® to delay catheter blockage in these experiments is most likely due to the burst release nature of drug delivery afforded by this coating, coupled with the continual flow of artificial urine in the bladder models system (Cho et al., 2003; Milo et al., 2017; Nzakizwanayo et al., 2019; Lin et al., 2021). Under these conditions, the concentration of Lithostat® achieved in residual bladder urine on coating activation will be continually reduced by urine flow through, and it appears that this does not remain at a sufficient level for long enough to inhibit crystal formation significantly. Although the current coating design containing 5 mg mL−1 AHA is expected to result in the generation of concentrations in residual bladder model urine in the range of ∼ 150 – 250µg mL−1, which is well above the expected therapeutic range for this drug (indicated as 8 - 30µg mL−1 in the prescribing information; (Mission Pharmacal Company, 2023) further optimization of the coating to increase the concentration of AHA delivered may improve performance. The timing of release of AHA may also be a factor, in that pH elevation and crystal formation will already be initiated once coatings have been activated under the conditions tested in our infection models. This could impede the ability of Lithostat® to effectively inhibit further crystal formation when delivered locally via our coating. In addition, the recommended use of Lithostat® is as an adjunct to antibiotic therapy (Mission Pharmacal Company, 2023), and combining this with antibacterial agents in coatings should be explored in subsequent studies.

In contrast, the incorporation of ciprofloxacin in our coatings provided significant increases in time to blockage and tripled catheter life span. Fluroquinolones, such as ciprofloxacin, are broad spectrum antibiotics used routinely in UTI therapies (Brown, 2006; Cao et al., 2021). According to the National Institute of Health and Care Excellence (NICE) guidelines, ciprofloxacin is recommended as a first choice oral and intravenous antibiotic for patients suffering upper UTI symptoms (National Institute of Health and Care Excellence, 2018). However, recent concerns in regards to the safety profile of ciprofloxacin have been raised, and include reports of disabling and potentially long-lasting side effects (such as tendonitis, arthralgia, and neuropathies associated with paraesthesia, fatigue, memory impairment, depression, and sleep disorders), which have led to new restrictions and precautions in the use of ciprofloxacin (Zacchè et al., 2015). As such, the potential for more targeted localized delivery of these agents in response to infection could help prevent problematic side effects, and is also an important feature of strategies aimed at reducing the spread of antibiotic resistance.

Our results are congruent with those of Zhou and colleagues, who also described the localized delivery of ciprofloxacin via an infection-responsive catheter coating (Zhou et al., 2018). As in our experiments, the coating-mediated delivery of ciprofloxacin was also able to provide significant increases in time to P. mirabilis mediated catheter blockage (Zhou et al., 2018). However, despite reporting comparable blockage time for uncoated catheters, the increased time to blockage afforded by coatings tested in this study (up to 52.5 h for coatings with ciprofloxacin HCl alone) was notably greater than the increases attained by Zhou et al., (2018) (up to 20 h). Incorporation of CF dye in our coating along with ciprofloxacin HCl was able to provide ∼ 79 h advanced warning of catheter blockage, also considerably more than ∼ 28 h (based on initial colour change indicative of pH elevation) reported for analogous coatings by Zhou et al., (2018).

Although similar concentrations of ciprofloxacin HCl were incorporated into our coatings and those of Zhou et al., (5 and 4.5 mg mL−1, respectively), the differences in coating performance are likely attributable to variation in coating designs and the resulting release profiles afforded. The encapsulation of ciprofloxacin HCl into vesicles is reported to provide a slower sustained release in coatings developed by (Zhou et al., 2018). While there are potentially clear advantages to a slow sustained release, and such innovations may allow drugs such as Lithostat® to work more effectively when delivered locally via infection-responsive coatings, the delivery of antimicrobials to inhibit biofilm formation may work less well in this approach. In contrast, the more rapid burst release expected from our coating design would be expected to expose P. mirabilis cells to high initial concentrations of ciprofloxacin HCl at early stages of biofilm formation. These differences in release profile could have contributed to the increased catheter lifespan observed in our coating system.

This hypothesis is congruent with previous studies of ciprofloxacin susceptibility in P. mirabilis crystalline biofilms, which demonstrated key roles of biomineralization in protecting cells from this antibiotic and preventing penetration into the biofilm (Li et al., 2016). As such a greater exposure to ciprofloxacin HCl at earlier stages of biofilm formation, when susceptibility is higher, would be expected to have a more pronounced impact. Alternatively, the acidic nature of ciprofloxacin could in itself be a factor if its release also serves to temporarily buffer urinary pH and further delay crystal formation, and this mechanism would also be expected to be more pronounced in a burst release system. In addition, anti-urease activity has been reported by ciprofloxacin (and other fluroquinolones) due to their carboxylic group interacting with the nickel atoms present at the active site of the urease enzyme (Kafarski and Talma, 2018), which may also play a role in the increase in blockage times observed.

However, it should be noted that in both our study and that of Zhou et al., the elimination of P. mirabilis from the bladder model system was not achieved, and cell population had recovered by the time catheters blocked (Zhou et al., 2018). Although not statistically significant, the recovery of bacterial populations was potentially more pronounced for coatings containing ciprofloxacin and CF dye in our experiments compared to ciprofloxacin alone (Figure. 3). This most likely reflects the longer time taken for catheters to block in models with ciprofloxacin HCl and CF dye-coated catheters (average of ∼ 14 h longer compared to models with ciprofloxacin only coatings). This increased time to blockage is most likely due to the synergistic effects of ciprofloxacin and CF dye, but will also result in greater dilution of released drugs prior to blockage due to continual urine flow through the system. This dilution effect may result in greater recovery of the bacterial population before the catheters block.

A further notable result of this study is that ‘empty’ coatings devoid of any cargo molecules, as well as those containing the CF dye alone, were able to provide significant increases in time to catheter blockage compared to uncoated controls. This is in contrast to our previous studies, where empty coatings did not impact time to blockage compared to uncoated catheters (Milo et al., 2016). In addition, we did not previously observe any inhibitory impact of the CF dye alone on blockage times, although its effect on P. mirabilis growth was not evaluated (Milo et al., 2016). Furthermore, the current coating design has been modified and optimized and these changes are likely responsible for the impact of empty coatings on blockage time. This likely stems from a mechanical disruption of biofilm formation when coatings are activated. The dissolution of the upper Eudragit S 100® layer and the subsequent detachment and degradation of coating (as observed via SEM), should serve to also remove biofilms that have begun to form on these surfaces and around the catheter eyelet. The eyelet in particular is a primary site of blockage and disruption of biofilms that begin to form in this region would be expected to have a notable impact on overall time to blockage (Stickler et al., 1999; Stickler, 2008; Stickler and Feneley, 2010; Holling et al., 2014).

These observations may point to additional attributes of the infection responsive coatings that could be optimized and exploited to control biofilm formation on urinary catheters. This aspect of our coating technology may be particularly relevant in the context of antibiotic resistance. For example, although ciprofloxacin release would not be expected to provide protection against resistant strains of P. mirabilis, the physical effects of coating degradation on biofilm formation should be pertinent regardless of the antibiotic resistance profile of infecting P. mirabilis strains. In this context, the incorporation of other non-antibiotic payloads into coatings capable of inhibiting bacterial growth or biofilm formation could also be explored to counter potential issues with antibiotic resistance. For example, the drugs thioridazine and fluoxetine have previously been reported to reduce P. mirabilis biofilm formation through efflux inhibition (Nzakizwanayo et al., 2017), whilst, the use of novel antimicrobial peptides to control bacterial infection and encrustation on ureteral stents has recently been described (Yao et al., 2022).

Overall, this study has demonstrated the potential for theranostic, infection-responsive coatings to form a promising approach to combat catheter encrustation and blockage. Further studies are required to improve coating performance and production, this includes optimization of the agents incorporated, and understanding parameters such as shelf-life, stability, and biocompatibility. Ultimately, the performance of final coating designs will need to be evaluated in clinical trials to clearly establish efficacy, safety, and benefits to patients. However, this proof-of-concept study shows that the continued development of this coating technology has the potential to address a clear and unmet clinical need, which could provide considerable benefits to many patients.

Acknowledgement

The authors would like to thank Dr. Matthew Case of the Material and Chemical Characterisation Facility (MC²) at the University of Bath (doi.org/10.15125/mx6j-3r54) for their technical support and guidance in this work.

Contributor Information

Anthony J Slate, Department of Life Sciences, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.

Ocean E Clarke, Department of Life Sciences, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.

Mina Kerio, School of Applied Sciences, University of Brighton, Brighton BN2 4GJ, United Kingdom.

Jonathan Nzakizwanayo, Department of Life Sciences, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.

Bhavik Anil Patel, School of Applied Sciences, University of Brighton, Brighton BN2 4GJ, United Kingdom.

Brian V Jones, Department of Life Sciences, University of Bath, Claverton Down, Bath BA2 7AY, United Kingdom.

Conflict of interest

The authors declare no conflict of interest.

Funding

This work was primarily supported by funding from the Dunhill Medical Trust (RPGF1906\171). O.E.C. is supported by a University of Bath PhD studentship. J.N. is also supported by funding from the Wellcome Trust (206 854/Z/17/Z).

Author contributions

Anthony J. Slate (Conceptualization, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing), Ocean E. Clarke (Investigation, Methodology, Writing – review & editing), Mina Kerio (Methodology), Jonathan Nzakizawayano (Investigation, Writing – review & editing), Bhavik Anil Patel (Conceptualization) and Brian V. Jones (Conceptualization, Funding acquisition, Project administration, Supervision, Writing – review & editing).

Data availability

The data underlying this article will be shared on reasonable request to the corresponding authors.

References

  1. Abou Chakra M, Dellis AE, Papatsoris AGet al. Established and recent developments in the pharmacological management of urolithiasis: an overview of the current treatment armamentarium. Expert Opin Pharmacother. 2020;21:85–96.. 10.1080/14656566.2019.1685979. [DOI] [PubMed] [Google Scholar]
  2. Benderev TV. Acetylcysteine for urinary tract mucolysis. J Urol. 1988;139:353–4.. 10.1016/S0022-5347(17)42412-8. [DOI] [PubMed] [Google Scholar]
  3. Brown PD. Ciprofloxacin for the management of urinary tract infection. Womens Health (Lond Engl). 2006;2:509–16.. 10.2217/17455057.2.4.509. [DOI] [PubMed] [Google Scholar]
  4. Cao D, Shen Y, Huang Yet al. Levofloxacin versus ciprofloxacin in the treatment of urinary tract infections: evidence-based analysis. Front Pharmacol. 2021;12:658095. 10.3389/fphar.2021.658095. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Cho YW, Park JH, Kim SHet al. Gentamicin-releasing urethral catheter for short-term catheterization. J Biomater Sci Polym Ed. 2003;14:963–72.. 10.1163/156856203322381447. [DOI] [PubMed] [Google Scholar]
  6. Clapham L, McLean R, Nickel Jet al. The influence of bacteria on struvite crystal habit and its importance in urinary stone formation. J Cryst Growth. 1990;104:475–84.. 10.1016/0022-0248(90)90150-J. [DOI] [Google Scholar]
  7. Clark M, Wright M-D. Antisepsis for Urinary Catheter Insertion: A Review of Clinical Effectiveness and Guidelines. Ottawa: Canadian Agency for Drugs and Technologies in Health, 2019. [PubMed] [Google Scholar]
  8. Cortese YJ, Wagner VE, Tierney Met al. 2018; Review of catheter-associated urinary tract infections and in vitro urinary tract models. J Healthc Eng. 2018;2018: 2986742. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Cox AJ, Hukins DW. Morphology of mineral deposits on encrusted urinary catheters investigated by scanning electron microscopy. J Urol. 1989;142:1347–50.. 10.1016/S0022-5347(17)39095-X. [DOI] [PubMed] [Google Scholar]
  10. Dumanski AJ, Hedelin H, Edin-Liljegren Aet al. Unique ability of the Proteus mirabilis capsule to enhance mineral growth in infectious urinary calculi. Infect Immun. 1994;62:2998–3003.. 10.1128/iai.62.7.2998-3003.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Follmer C. Ureases as a target for the treatment of gastric and urinary infections. J Clin Pathol. 2010;63:424–30.. 10.1136/jcp.2009.072595. [DOI] [PubMed] [Google Scholar]
  12. Georgi K. American Urological Association. Patient Care. 2004;38:10–11. [Google Scholar]
  13. Griffith DP, Khonsari F, Skurnick JHet al. A randomized trial of acetohydroxamic acid for the treatment and prevention of infection-induced urinary stones in spinal cord injury patients. J Urol. 1988;140:318–24.. 10.1016/S0022-5347(17)41592-8. [DOI] [PubMed] [Google Scholar]
  14. Griffith DP, Musher D, Itin C. Urease. The primary cause of infection-induced urinary stones. Invest Urol. 1976;13:346350. [PubMed] [Google Scholar]
  15. Hedelin H, Eddeland A, Larsson Let al. The composition of catheter encrustations, including the effects of allopurinol treatment. Br J Urol. 1984;56:250–4.. 10.1111/j.1464-410X.1984.tb05382.x. [DOI] [PubMed] [Google Scholar]
  16. Holling N, Lednor D, Tsang Set al. Elucidating the genetic basis of crystalline biofilm formation in Proteus mirabilis. Infect Immun. 2014;82:1616–26.. 10.1128/IAI.01652-13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Jacobsen SM, Stickler DJ, Mobley HLet al. Complicated catheter-associated urinary tract infections due to Escherichia coli and Proteus mirabilis. Clin Microbiol Rev. 2008;21:26–59.. 10.1128/CMR.00019-07. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Jones BV, Mahenthiralingam E, Sabbuba NAet al. Role of swarming in the formation of crystalline Proteus mirabilis biofilms on urinary catheters. J Med Microbiol. 2005;54:807–13.. 10.1099/jmm.0.46123-0. [DOI] [PubMed] [Google Scholar]
  19. Jordan RPC, Malic S, Waters MGet al. Development of an antimicrobial urinary catheter to inhibit urinary catheter encrustation. Microbiol Discov. 2015;3.1. 10.7243/2052-6180-3-1. [DOI] [Google Scholar]
  20. Kafarski P, Talma M. Recent advances in design of new urease inhibitors: a review. J Adv Res. 2018;13:101–12.. 10.1016/j.jare.2018.01.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Kohler-Ockmore J, Feneley R. Long-term catheterization of the bladder. Br J Urol. 1996;77:347–51.. 10.1046/j.1464-410X.1996.09074.x. [DOI] [PubMed] [Google Scholar]
  22. Li X, Lu N, Brady HRet al. Ureolytic biomineralization reduces Proteus mirabilis biofilm susceptibility to ciprofloxacin. Antimicrob Agents Chemother. 2016;60:2993–3000.. 10.1128/AAC.00203-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Lin EMJ, Lay CL, Subramanian GSet al. Control release coating for urinary catheters with enhanced released profile for sustained antimicrobial protection. ACS Appl Mater Interfaces. 2021;13:59263–74.. 10.1021/acsami.1c17697. [DOI] [PubMed] [Google Scholar]
  24. Long A, Edwards J, Thompson Ret al. A clinical evaluation of a sensor to detect blockage due to crystalline biofilm formation on indwelling urinary catheters. BJU Int. 2014;114:278–85.. 10.1111/bju.12577. [DOI] [PubMed] [Google Scholar]
  25. Loveday HP, Wilson JA, Pratt RJet al. Epic3: national evidence-based guidelines for preventing healthcare-associated infections in NHS hospitals in England. J Hosp Infect. 2014;86:S1–70.. 10.1016/S0195-6701(13)60012-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Marei MM, Jackson R, Keene DJB. Intravesical gentamicin instillation for the treatment and prevention of urinary tract infections in complex paediatric urology patients: evidence for safety and efficacy. J Pediatr Urol. 2021;17:65.e1–11.. 10.1016/j.jpurol.2020.08.007. [DOI] [PubMed] [Google Scholar]
  27. Milo S, Hathaway H, Nzakizwanayo Jet al. Prevention of encrustation and blockage of urinary catheters by Proteus mirabilis via pH-triggered release of bacteriophage. J Mater Chem B. 2017;5:5403–11.. 10.1039/C7TB01302G. [DOI] [PubMed] [Google Scholar]
  28. Milo S, Nzakizwanayo J, Hathaway HJet al. Emerging medical and engineering strategies for the prevention of long-term indwelling catheter blockage. Proc Inst Mech Eng H. 2019;233:68–83.. 10.1177/0954411918776691. [DOI] [PubMed] [Google Scholar]
  29. Milo S, Thet NT, Liu Det al. An in-situ infection detection sensor coating for urinary catheters. Biosens Bioelectron. 2016;81:166–72.. 10.1016/j.bios.2016.02.059. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Mission Pharmacal Company . 2023; Lithostat (Acetohydroxamic Acid) Tablets—full prescribing information. https://missionpharmacal.com/lithostat-acetohydroxamic-acid/ (5 June 2023, date last accessed).
  31. Morris NS, Stickler D, McLean RJ. The development of bacterial biofilms on indwelling urethral catheters. World J Urol. 1999;17:345–50.. 10.1007/s003450050159. [DOI] [PubMed] [Google Scholar]
  32. Morris NS, Stickler DJ. The effect of urease inhibitors on the encrustation of urethral catheters. Urol Res. 1998;26:275–9.. 10.1007/s002400050057. [DOI] [PubMed] [Google Scholar]
  33. Moussa M, Chakra MA, Papatsoris AGet al. Bladder irrigation with povidone-iodine prevent recurrent urinary tract infections in neurogenic bladder patients on clean intermittent catheterization. Neurourol Urodyn. 2021;40:672–9.. 10.1002/nau.24607. [DOI] [PubMed] [Google Scholar]
  34. National Institute of Health and Care Excellence . 2018; Urinary tract infection (catheter-associated): antimicrobial prescribing. National Institute of Health and Care Excellence (NICE guideline [NG113]: London, UK. https://www.nice.org.uk/guidance/ng113. [Google Scholar]
  35. Nicolle LE. Catheter associated urinary tract infections. Antimicrob Resist Infect Control. 2014;3:2014. 10.1186/2047-2994-3-23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. Nzakizwanayo J, Hanin A, Alves DRet al. Bacteriophage can prevent encrustation and blockage of urinary catheters by Proteus mirabilis. Antimicrob Agents Chemother. 2016;60:1530–6.. 10.1128/AAC.02685-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. Nzakizwanayo J, Pelling H, Milo Set al. (2019) An in vit ro bladder model for studying catheter-associated urinary tract infection and associated analysis of biofilms. Methods Mol Biol. 2021;139–58. [DOI] [PubMed] [Google Scholar]
  38. Nzakizwanayo J, Scavone P, Jamshidi Set al. Fluoxetine and thioridazine inhibit efflux and attenuate crystalline biofilm formation by Proteus mirabilis. Sci Rep. 2017;7:12222. 10.1038/s41598-017-12445-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  39. O'Hara CM, Brenner FW, Miller JM. Classification, identification, and clinical significance of Proteus, Providencia, and Morganella. Clin Microbiol Rev. 2000;13:534–46.. 10.1128/CMR.13.4.534. [DOI] [PMC free article] [PubMed] [Google Scholar]
  40. Parker V, Giles M, Graham Let al. Avoiding inappropriate urinary catheter use and catheter-associated urinary tract infection (CAUTI): a pre-post control intervention study. BMC Health Serv Res. 2017;17:1–9.. 10.1186/s12913-017-2268-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Pelling H, Nzakizwanayo J, Milo Set al. Bacterial biofilm formation on indwelling urethral catheters. Lett Appl Microbiol. 2019;68:277–93.. 10.1111/lam.13144. [DOI] [PubMed] [Google Scholar]
  42. Shackley DC, Whytock C, Parry Get al. Variation in the prevalence of urinary catheters: a profile of National Health Service patients in England. BMJ Open. 2017;7:e013842. [DOI] [PMC free article] [PubMed] [Google Scholar]
  43. Stickler D, Feneley R. The encrustation and blockage of long-term indwelling bladder catheters: a way forward in prevention and control. Spinal Cord. 2010;48:784–90.. 10.1038/sc.2010.32. [DOI] [PubMed] [Google Scholar]
  44. Stickler D, Zimakoff J. Complications of urinary tract infections associated with devices used for long-term bladder management. J Hosp Infect. 1994;28:177–94.. 10.1016/0195-6701(94)90101-5. [DOI] [PubMed] [Google Scholar]
  45. Stickler DJ. Bacterial biofilms in patients with indwelling urinary catheters. Nat Rev Urol. 2008;5:598–608.. 10.1038/ncpuro1231. [DOI] [PubMed] [Google Scholar]
  46. Stickler DJ. Clinical complications of urinary catheters caused by crystalline biofilms: something needs to be done. J Intern Med. 2014;276:120–9.. 10.1111/joim.12220. [DOI] [PubMed] [Google Scholar]
  47. Stickler DJ, Jones SM, Adusei GOet al. A clinical assessment of the performance of a sensor to detect crystalline biofilm formation on indwelling bladder catheters. BJU Int. 2006;98:1244–9.. 10.1111/j.1464-410X.2006.06562.x. [DOI] [PubMed] [Google Scholar]
  48. Stickler DJ, Morris NS, Winters C. Simple physical model to study formation and physiology of biofilms on urethral catheters. Methods Enzymol. 1999;3310:494–501.. 10.1016/S0076-6879(99)10037-5. [DOI] [PubMed] [Google Scholar]
  49. Tambyah PA, Maki DG. Catheter-associated urinary tract infection is rarely symptomatic: a prospective study of 1,497 catheterized patients. Arch Intern Med. 2000;160:678–82. [DOI] [PubMed] [Google Scholar]
  50. Wan Q, Xiong G, Liu Get al. Urothelium with barrier function differentiated from human urine-derived stem cells for potential use in urinary tract reconstruction. Stem Cell Res Ther. 2018;9:304. 10.1186/s13287-018-1035-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. Williams JJ, Rodman JS, Peterson CM. A randomized double-blind study of acetohydroxamic acid in struvite nephrolithiasis. N Engl J Med. 1984;311:760–4.. 10.1056/NEJM198409203111203. [DOI] [PubMed] [Google Scholar]
  52. Yakusheva O, Costa DK, Bobay KLet al. Variability in catheter-associated asymptomatic bacteriuria rates among individual nurses in intensive care units: an observational cross-sectional study. PLoS One. 2019;14:e0218755. 10.1371/journal.pone.0218755. [DOI] [PMC free article] [PubMed] [Google Scholar]
  53. Yao Q, Zhang J, Chen B. Mussel-inspired clickable antibacterial peptide coating on ureteral stents for encrustation prevention. ACS Appl Mater Interfaces. 2022;14:36473–86.. 10.1021/acsami.2c09448. [DOI] [PubMed] [Google Scholar]
  54. York NE, Borofsky MS, Lingeman JE. Risks associated with drug treatments for kidney stones. Expert Opin Drug Saf. 2015;14:1865–77.. 10.1517/14740338.2015.1100604. [DOI] [PubMed] [Google Scholar]
  55. Zacchè MM, Srikrishna S, Cardozo L. Novel targeted bladder drug-delivery systems: a review. Res Rep Urol. 2015;7:169–78. [DOI] [PMC free article] [PubMed] [Google Scholar]
  56. Zhou J, Hou S, Li Let al. Theranostic infection-responsive coating to in situ detect and prevent urinary catheter blockage. Adv Mater Interfaces. 2018;5:1801242. 10.1002/admi.201801242. [DOI] [Google Scholar]
  57. Zisman AL. Effectiveness of treatment modalities on kidney stone recurrence. CJASN. 2017;12:1699–708.. 10.2215/CJN.11201016. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data underlying this article will be shared on reasonable request to the corresponding authors.