. 2023 May 30;13(6):591. doi: 10.3390/bios13060591

Discriminating Glioblastoma from Peritumoral Tissue by a Nanohole Array-Based Optical and Label-Free Biosensor

Víctor García-Milán 1,, Alfredo Franco 2,3,, Margarita Estreya Zvezdanova 3, Sara Marcos 4, Rubén Martin-Laez 1, Fernando Moreno 2,3, Carlos Velasquez 1,3,5, José L Fernandez-Luna 3,6,*
PMCID: PMC10296434  PMID: 37366956

Abstract

In glioblastoma (GBM) patients, maximal safe resection remains a challenge today due to its invasiveness and diffuse parenchymal infiltration. In this context, plasmonic biosensors could potentially help to discriminate tumor tissue from peritumoral parenchyma based on differences in their optical properties. A nanostructured gold biosensor was used ex vivo to identify tumor tissue in a prospective series of 35 GBM patients who underwent surgical treatment. For each patient, two paired samples, tumor and peritumoral tissue, were extracted. Then, the imprint left by each sample on the surface of the biosensor was individually analyzed, obtaining the difference between their refractive indices. The tumor and non-tumor origins of each tissue were assessed by histopathological analysis. The refractive index (RI) values obtained by analyzing the imprint of the tissue were significantly lower (p = 0.0047) in the peritumoral samples (1.341, Interquartile Range (IQR) 1.339–1.349) compared with the tumor samples (1.350, IQR 1.344–1.363). The ROC (receiver operating characteristic) curve showed the capacity of the biosensor to discriminate between both tissues (area under the curve, 0.8779, p < 0.0001). The Youden index provided an optimal RI cut-off point of 0.003. The sensitivity and specificity of the biosensor were 81% and 80%, respectively. Overall, the plasmonic-based nanostructured biosensor is a label-free system with the potential to be used for real-time intraoperative discrimination between tumor and peritumoral tissue in patients with GBM.

Keywords: glioblastoma, biosensor, extraordinary optical transmission, refractive index

1. Introduction

Glioblastoma (GBM) is the most common and lethal of primary brain tumors in adults, accounting for 49.1% of all malignant tumors with an incidence rate of 3.23 cases per 100,000 inhabitants in the United States [1]. Presently, GBM remains a major clinical challenge with a median survival of 14.6 months despite the current standard of care based on surgical resection followed by temozolomide, an alkylating agent, and radiotherapy [2].

The current paradigm of surgical treatment is to achieve maximal safe tumor resection, and the extent of resection is considered a prognostic factor for longer survival [3,4]. Nevertheless, due to GBM’s invasiveness and its diffuse parenchymal infiltration, complete tumor resection is often precluded [5]. Moreover, this infiltrative behavior results in a blurred tumor margin, which hampers the intraoperative identification of the tumor edge by the neurosurgeon [6].

In this context, several tools have been developed attempting to discriminate tumors from the surrounding parenchyma, such as neuronavigation, fluorescent marking, histopathological analysis, and confocal laser endomicroscopy (CLE) [7,8,9,10,11]. However, reliable identification of tumor margins in real time to attempt complete tumor removal remains a challenge today.

Biomedical applications of the extraordinary optical transmission (EOT) phenomenon [12] to discriminate different molecules and tissues have been widely explored in the last few decades [13,14,15]. The EOT phenomenon is usually observed in metal films with periodic nanohole arrays, and it is highly dependent on the changes in the optical properties of the medium in contact with the metal surface due to changes in plasmonic resonance conditions. Different plasmonic-based biosensor systems have been developed to detect very small amounts of biomolecules, taking advantage of the electric field enhancement in plasmonic materials [15,16,17]. Plasmonic biosensors have been proposed for the analysis of clinically relevant biomarkers, such as amyloid beta protein in Alzheimer’s disease [18] or prostate-specific antigen (PSA) in prostate cancer [19]. Moreover, EOT-based technology opens the possibility of using portable diagnostic devices made with miniaturized biosensors without the need for additional sophisticated equipment or specialized personnel to operate them [14,20].

Our group has previously reported the development of an optical system able to perform label-free, highly sensitive, and specific discrimination between live single tumor cells and normal cells in contact with a gold nanohole array [21,22,23]. The performance of the system was validated through the analysis of more than one thousand live cells [21]. We also described the use of the biosensor to study key features of cancer cells [24].

In the present study, we report the performance of the EOT-based biosensor for ex vivo discrimination between tumor and peritumoral tissue in patients with GBM, which provides proof of concept for its potential use as an intraoperative biosensing device.

2. Materials and Methods

2.1. Design and Subjects

A prospective study was performed, including a series of patients with GBM who were candidates for surgical treatment. Patients met all the following inclusion criteria: (1) adults ≥ 18 years of age; (2) newly diagnosed GBM; and (3) histopathology-confirmed GBM.

The collection of tissue specimens did not modify the standard of care for the patient. Patients’ data were treated confidentially, and research was conducted in accordance with the ethical standards of the Helsinki Declaration.

2.2. Clinical and Radiological Variables

Clinical and radiological data were collected and included age, sex, aim of surgery, tumor location, and presence of residual tumor following the surgical procedure.

Moreover, preoperative planification was determined in the preoperative magnetic resonance imaging (MRI) by a semi-automatic manual segmentation process using the Brainlab Elements software modules, Cranial 3.0 (Brainlab AG, Munich, Germany). The tumor area was defined as the contrast-enhanced tumor portion in the contrast-enhanced T1-weighted sequence (CE-T1W), and the peritumoral area as the peritumoral T2/FLAIR hyperintensity in the 3D FLAIR sequence.

2.3. Specimens Collection

Tumor and paired adjacent peritumoral tissue specimens were obtained from each case following the routine surgical sampling protocol for histopathological diagnosis. Specimens were collected by the neurosurgeon in the operating room during a standard craniotomy for tumor removal/biopsy. Surgery was performed in a standard manner, following our common institutional practice and the standard of care.

Both specimens were obtained by using MRI-based neuronavigation guidance and intraoperative fluorescence with 5-ALA or sodium fluorescein. The tumor specimen was collected from the tumor core, which was defined as the contrast enhancement T1 hyperintensity in the preoperative MRI. Areas of obvious necrosis and areas with high blood content were avoided. Comparably, the peritumoral sample was taken from the peritumoral normal-appearing parenchyma immediately adjacent to the tumor, avoiding areas of the cortex.

In all cases, the coordinates of the sample collection sites were saved in the navigation system, and screenshots were acquired to determine the precise site of the sampling (Figure 1a). The size of each sample was within a range of 5–8 mm.

Figure 1.

Figure 1

Overview of the workflow. (a) Axial (upper panels) and coronal (lower panels) MRI images showing the tumor area (red) of a patient with a left temporal lobe GBM. The tumor area is defined by the contrast-enhanced portion of the T1-weighted sequence and is preplanned in the Brainlab Elements software. Green lines pinpoint the specific sites where the tissue samples were taken. (b) Heatmap showing the shift of the plasmon resonance wavelength at each position on the biosensor due to the RI of the imprint. A total of 144 measurements covering the entire biosensor were obtained for each tissue sample. A representative histogram showing the RI values of the imprints left by peritumoral (green) and tumor (red) tissues is shown. RIU, refractive index units. (c) Schematic representation of tumor and peritumoral areas of a GBM showing the tumor border and the necrotic tumor core. The right image shows a representative hematoxylin-eosin-stained section revealing the tumor-peritumoral tissue border.

2.4. Biosensor Optimization and Specimen Processing

The features of the biosensor have already been described in detail [21]. In brief, it consists of a gold film with a square array of holes periodically distributed along an area of 500 m × 500 m made by electron beam lithography (EBL). The hole’s diameter is 220 nm, the array periodicity is 550 nm, and the film thickness is 60 nm. All biosensors were characterized and cleaned before use. Characterization was carried out to determine their optical sensitivity in nm/RIU units and to calibrate the response of each biosensor along the entire surface, as described elsewhere [23]. In order to remove organic residues, biosensors were exposed to piranha solution and plasma cleaning procedures, as reported [21].

Each tumor and peritumoral tissue sample was taken with tweezers and placed on the corresponding biosensor to cover the entire nanohole array. This process has been integrated as part of the routine histopathological examination protocol following standards applied to GBM surgery. In order to get an imprint of the tissue, samples remained in contact with the biosensor surface for 1 min. At the end of this time, the tissue was removed, placed in a properly identified formalin-containing tube, and the histology was assessed by an experienced pathologist (S.M.). The excess liquid on the biosensors was evaporated at room temperature for 5–10 min, and the biosensors were stored at 4 °C until optical analysis.

2.5. Optical System and Setup

The imprint left by the tissue on the biosensor was analyzed by means of the optical transmission spectral shift due to the refractive index (RI) of the imprint. The spectral measurements were carried out in regions restricted to a diameter of 40 m, in such a way that each imprint was analyzed 144 times for statistical significance (Figure 1b).

The optical system used for the measurements has already been described in detail [21,23]. Briefly, it is composed of a motorized Nikon Eclipse microscope with a 20× objective, an Andor Shamrock 500i spectrograph, and a cooled Idus CCD camera coupled to the spectrograph. The transmission spectra were processed and analyzed using in-house software. A scheme of the optical system is represented in Figure 2.

Figure 2.

Figure 2

Main parts of the biosensor device. (a) Halogen lamp as source of light. (b) Adapted upright microscope. (c) Sample holder with the biosensor (in yellow with vertical lines) at the bottom of a well that is filled with saline buffer during the measurements. (d) Scanning electron micrograph of the nanostructured gold film. (e) Optical fiber. (f) Spectrophotometer. (g) Representative spectral shift due to optical differences between GBM tumors and peritumoral tissues. As a background reference, the spectrum of phosphate buffered saline (PBS) is also included.

2.6. Pathological Anatomy

After the imprint process, tissue samples were formalin-fixed and paraffin-embedded for histopathological analysis. In brief, 3 μm sections were obtained, and hematoxylin-eosin staining was performed according to standard protocols. Sections were examined under a conventional optical microscope (Figure 1c). Histological diagnosis was completed according to the 2021 WHO classification [25].

2.7. Statistical Analysis

A descriptive analysis of each clinical and radiological variable was performed. The Kolmogorov-Smirnov test was used to study the distribution of each variable. Medians and interquartile ranges (IQR) were calculated for variables with a non-normal distribution, while those with a normal distribution were recorded as the mean and standard deviation (SD).

A Wilcoxon-Mann-Whitney test (WMW) was used to compare non-normal distributed quantitative variables between tumor and peritumoral samples. Statistical significance was defined as a two-tailed p-value less than 5% (p < 0.05). The statistical analysis was performed using GraphPad Prism version 9.0.0 for Mac OS X (GraphPad Software, San Diego, CA, USA).

The receiver operating characteristic (ROC) curve was used to analyze the critical values to differentiate tumor from peritumoral tissue, and sensitivity and specificity values were obtained. Finally, thresholds for RI were set following the Youden index optimization criteria [26].

3. Results

3.1. Demographics

A total of 35 GBM patients who underwent surgery at our institution from May 2019 to March 2023 were included in the study. Four cases were excluded due to technical problems with the biosensor and/or the optical measurements. The main clinical, radiological, and demographic features are summarized in Table 1.

Table 1.

Clinical, sociodemographic, and radiological features.

Case Age Sex Side Location Aim of Surgery Residual Tumor
1 71 Female Right Frontal Resection No
2 59 Female Right Temporal Biopsy NA
3 72 Male Right Frontal Biopsy NA
4 49 Male Right Frontal Resection No
5 75 Female Right Frontal Resection Yes
6 69 Male Right Temporal Biopsy NA
7 50 Male Left Frontal Resection Yes
8 71 Male Right Frontal Resection No
9 72 Female Right Frontal Biopsy Yes
10 71 Female Right Parietal Resection NA
11 48 Male Bilateral Frontal Biopsy NA
12 59 Male Right Temporal Resection Yes
13 58 Female Right Temporal Resection Yes
14 42 Male Right Frontal Resection Yes
15 69 Male Right Occipital Resection No
16 70 Male Right Temporal Resection No
17 76 Female Right Frontal Biopsy NA
18 73 Male Left Frontal Resection No
19 58 Male Left Frontal Biopsy NA
20 77 Female Left Temporal Resection Yes
21 66 Male Left Occipital Resection No
22 68 Male Left Frontal Resection Yes
23 71 Female Left Parietal Resection No
24 65 Male Right Frontal Resection Yes
25 51 Male Left Frontal Resection No
26 65 Female Right Occipital Resection Yes
27 68 Male Left Temporal Biopsy NA
28 49 Male Right Frontal Resection Yes
29 78 Female Right Frontal Resection No
30 68 Male Left Occipital Resection Yes
31 61 Male Right Temporal Resection Yes

NA, data not available.

A total of 31 paired tumor-peritumoral samples were analyzed by an expert pathologist, and all tumor samples were histopathologically confirmed GBM (grade IV, WHO 2021) [25].

3.2. Biosensor Performance

A schematic representation of the entire process is shown in Figure 3, including the proposed adaptation of the ex vivo biosensor system to a setup in the operating room.

Figure 3.

Figure 3

Representation of the experimental system and the proposed setup in the operating room. The upper panel shows the different steps, starting with tissue collection, which is deposited onto the biosensor. Then, tissue is removed, and optical measurements of the imprints left by both GBM and peritumoral tissues are obtained. The lower panel shows a representation of the proposed setup that would be used during a surgical procedure. Light passes through the fiber towards the biosensor, located at the tip of a holder, and is reflected back to a microspectrometer for real-time identification of the tissue. Arrows indicate the direction of the light. Illustration adapted from templates by BioRender.com.

Tissue samples were obtained using neuronavigation guidance and immediately deposited on the biosensor. The RI of tissue imprints was then analyzed for both tumor and peritumoral samples (Table 2). In some cases, postoperative histopathologic examination revealed that the peritumoral tissue was GBM. The RI difference is considered positive when tumor RI is higher than peritumoral RI and negative when peritumoral RI is higher than or very similar to tumor RI. Mean RI values were 1.341 (IQR 1.339–1.349) for peritumoral imprint and 1.350 (IQR 1.344–1.363) for tumor imprint (p = 0.0047).

Table 2.

RI values of tumor and peritumoral imprints on the biosensor.

Case Peritumoral
RI
Tumor RI RI
Difference
Peritumoral Histology * Tumor
Histology *
1 1.339 1.362 POS Peritumoral GBM
2 1.345 1.351 POS Peritumoral GBM
3 1.339 1.350 POS Peritumoral GBM
4 1.352 1.356 POS Peritumoral GBM
5 1.337 1.349 POS Peritumoral GBM
6 1.346 1.365 POS Peritumoral GBM
7 1.352 1.348 NEG GBM GBM
8 1.342 1.344 NEG GBM GBM
9 1.335 1.348 POS Peritumoral GBM
10 1.333 1.344 POS Peritumoral GBM
11 1.343 1.372 POS Peritumoral GBM
12 1.354 1.337 NEG Peritumoral GBM
13 1.339 1.350 POS Peritumoral GBM
14 1.335 1.344 POS Peritumoral GBM
15 1.336 1.334 NEG Peritumoral GBM
16 1.353 1.366 POS Peritumoral GBM
17 1.340 1.367 POS Peritumoral GBM
18 1.356 1.361 POS Peritumoral GBM
19 1.346 1.349 POS GBM GBM
20 1.341 1.344 POS Peritumoral GBM
21 1.349 1.368 POS Peritumoral GBM
22 1.345 1.343 NEG Peritumoral GBM
23 1.349 1.373 POS Peritumoral GBM
24 1.330 1.343 POS Peritumoral GBM
25 1.341 1.342 NEG Peritumoral GBM
26 1.481 1.399 NEG Peritumoral GBM
27 1.341 1.359 POS Peritumoral GBM
28 1.344 1.359 POS Peritumoral GBM
29 1.342 1.337 NEG GBM GBM
30 1.317 1.340 POS Peritumoral GBM
31 1.389 1.363 NEG GBM GBM

POS, positive; NEG, negative; * Postoperative histopathologic examination.

The RI difference was positive in 22 of the 31 cases (71%), confirming the capacity of the biosensor to distinguish between the tumor and the surrounding peritumoral tissue (Figure 4). The histopathological analysis revealed that in case 19, the peritumoral sample was GBM, indicating a false-positive result of the biosensor.

Figure 4.

Figure 4

Optical measurements of tissue imprints. Histograms show the RI values of the imprints left on the biosensor by peritumoral (green) and tumor (red) tissue. A magnified image of a representative histogram is shown in Figure 1B.

In cases 7, 8, 29, and 31, the RI difference was negative despite the fact that both tissues were GBM based on the histological assignment. In the other 5 cases, 12, 15, 22, 25, and 26, a negative RI difference was also obtained, but this time the histology confirmed the tumor and peritumoral origin of the tissues. A likely explanation for these unexpected results could be contamination with red blood cells, as they have a high RI (1.370–1.420) [27]. Although areas with high blood content were avoided, the presence of limited numbers of red cells might interfere with optical measurements.

The receiver operating characteristic (ROC) curve was used to assess the sensitivity and specificity for each of the possible cut-off points for the discrimination of tumor and peritumoral tissue (Figure 5). The highest Youden index, labeled J in the Receiver Operating Characteristic (ROC) curve, was used to provide an optimal RI cut-off point of 0.003. Under these conditions, the sensitivity and specificity were 81% and 80%, respectively. The area under the curve was 0.8779 (0.7571–0.9988, 95% confidence interval, p < 0.0001), indicating that the biosensor is a suitable system to distinguish GBM from its surrounding tissue.

Figure 5.

Figure 5

Receiver operating characteristic plot showing the biosensor’s capacity to discriminate between tumor and peritumoral tissues.

Moreover, the variation of spectral measurements from a single sample was around 0.151 nm, which gives an uncertainty of less than 0.001 RIU.

4. Discussion

4.1. Discrimination Capacity of the Biosensor

In this study, we prospectively evaluated the accuracy of an EOT-based tool in providing ex-vivo intraoperative discrimination between the GBM tumor and the peritumoral tissue. Other technologies are working on near-real-time assessment for tumor discrimination in GBM [11]. However, this is the first study on the potential use of EOT-based systems as intraoperative biosensing devices for real-time detection of GBM tumor margins during the surgical act.

Interestingly, our optical system was able to discriminate between GBM and peritumoral tissue without any need for tissue preparation or labeling, raising the possibility of implementing such technology during tumor removal in the operating room.

The ROC curve analysis observed in Figure 5 shows the sensitivity and specificity of each of the possible cut-off points of the EOT-based biosensor. In order to determine the optimal cut-off point to discriminate tumors from peritumoral tissue, the highest Youden index (0.61) was used. The biosensor showed a sensitivity of 81% and a specificity of 80%. A high sensitivity of the system is interpreted as the probability of obtaining a positive result for those tissues that are truly GBM, which allows discrimination from the surrounding parenchymal tissue. On the other hand, high specificity consists of the probability that in a non-tumor tissue, the biosensor will detect RI values corresponding to the peritumoral area.

The data presented here support the application of the nanoplasmonic biosensor to differentiate tumor tissue during tumor resection and may entail several potential advantages over the systems currently being used for the same purpose.

Histopathological analysis is the current gold standard for the identification of tumor tissue, but its definitive result is only provided days after surgery [28]. Intraoperative pathological assessment of tumor margins may guide neurosurgeon intervention, but it does not provide real-time results and requires interpretation by an experienced neuropathologist [29,30]. On the contrary, the nanoplasmonic biosensor can efficiently differentiate between tumor tissue and the surrounding parenchyma and could be used as a real-time discrimination system in GBM surgeries. The system would not require interpretation or special training, as the biosensor could provide a different signal depending on the tissue analyzed.

Other intraoperative techniques used to discriminate tumor tissue, such as fluorescence-guided surgery and confocal laser endomicroscopy (CLE), rely on the pre- or intraoperative administration of a fluorophore to enhance the tumor tissue [9,10,11]. Intraoperative fluorophores commonly used in GBM surgery include 5-aminolevulinic acid (5-ALA) [9,31] and sodium fluorescein [32,33], which are administered to the patient before or during surgery. Although it has been demonstrated that tumor fluorescence derived from 5-ALA enables more complete resections, achieving a progression-free survival improvement [9], fluorescence fades away from the tumor core, and its interpretation becomes challenging at the infiltrative margins [5]. Some authors described key factors that limit the sensitivity of this approach, like low spatial resolution due to the “averaging effect” or poor optical detection for disseminated and sparse cells in tumor margins [31]. Additionally, although 5-ALA and sodium fluorescence have been described as well-tolerated compounds with a low rate of side effects in high-grade glioma resection surgery [34], some studies have described mild side effects, such as intraoperative hypotension in a small proportion of patients [35,36], photosensitivity, and allergic reactions.

Similarly, CLE is a novel labeling technique that allows in vivo digital biopsies for the visualization of cellular structures by using previously injected sodium fluorescein [11]. Recent studies have shown that CLE is a useful tool for the accurate diagnosis of brain tumor lesions by obtaining high-quality intraoperative images of different targets (tumor and peritumoral areas or the interface between both) [37,38]. However, CLE images require interpretation by experts in the field. In addition, fluorescence-based techniques may have side effects, and pregnant patients or patients with hypersensitivity to the fluorophore or renal failure are excluded [38].

On the other hand, the use of the plasmonic biosensor in this context is a label-free method that can accurately discriminate GBM from the peritumoral tissue based on the EOT, which relies on the biophysical features of the tissue. Contrary to fluorescence-guided surgery and the CLE, the plasmonic biosensor would not require any procedure to prepare the patient for surgery, avoiding possible alterations of the tissue as well as potential side effects. In addition, the biosensor does not require interpretation of results because it gives an immediate yes or no response based on a cut-off point for refractive index values.

There are other image-based systems, such as neuronavigation, intraoperative ultrasound (ioUS), and intraoperative MRI (ioMRI). The first one relies on the placement of sensors during surgery that correlate the location with a preoperative neuroimaging study (CT or MRI). However, during surgery, the brain parenchyma undergoes distortion due to cerebrospinal fluid loss, edema, and tumor resection. These changes, known as “brain shift”, make the preoperative neuronavigator image less reliable during surgery [7,8]. ioUS provides real-time feedback, but it has a high user dependency and requires a long learning curve and a complex interpretation of results [39,40]. Lastly, ioMRI can improve the extent of resection by identifying residual tumors during glioma surgery. However, the process of introducing and maintaining iMRI is very costly and time-consuming without real-time feedback [41,42,43], and its use may also increase complications due to prolonged operating time [44,45].

Overall, the nanoplasmonic biosensor capacity for discrimination between tumor and peritumoral tissue, alone or in combination with other methods, may have a place in the workflow for the surgical treatment of GBM.

4.2. Limitations and Future Perspectives

Our study included a limited number of patients, and the results will need to be replicated in larger cohorts based on multicenter studies.

In addition, further optimization of the nanostructure of the plasmonic biosensor could improve performance in the identification of tumor tissue, increasing sensitivity. We realize that other biological components, including red blood cells, may interfere with the RI measurements, but both tissues, GBM and PT, are so close together that we would not expect a contamination bias in a specific tissue. Moreover, tissues were gently flushed with saline buffer once the tissue was deposited on the biosensor, similar to the flushing with saline to remove blood from the surgical field. The sensitivity and specificity of the biosensor are now about 80%. In order to improve specificity, we first need to gain an understanding of the biological basis of the optical differences between tissues, which will reveal the potential contaminants that might interfere with optical measurements.

We are now working on technical issues for a future adaptation of the biosensor to the surgical workflow. Finally, studies are under way to analyze the components in the tumor and peritumoral tissues responsible for the discrimination capacity of the biosensor.

5. Conclusions

The plasmonic-based nanostructured biosensor has been shown to efficiently discriminate between tumor and peritumoral tissue ex vivo in patients with GBM. The experimental workflow starts with small pieces of GBM and peritumoral tissue that are deposited on the surface of the biosensor. The imprints left by the tissues are then analyzed with an adapted upright microscope connected to a spectrometer to obtain optical measurements. The sensitivity and specificity of the biosensing system were determined to be about 80%, and the area under the curve was 0.8779 (0.7571–0.9988, 95% confidence interval, p < 0.0001), indicating that the biosensor provides a suitable procedure to distinguish GBM from its peritumoral tissue. This is the first step in the road map to transfer the experimental setup to an in vivo system able to discriminate between both tissues in real time, which would assist decision-making during surgery.

To achieve this goal, further optimization of the optical system to improve specificity, an understanding of the biological bases for these optical differences, and studies in larger cohorts of patients undergoing surgery for GBM will need to be done.

Acknowledgments

The authors are grateful to Olga Gutierrez and Paula Calzada for their technical help.

Author Contributions

Conceptualization, F.M., C.V. and J.L.F.-L.; experimental work, V.G.-M., A.F., M.E.Z. and S.M.; project administration, R.M.-L.; resources, F.M. and J.L.F.-L.; validation, A.F., R.M.-L. and J.L.F.-L.; writing—original draft, V.G.-M., M.E.Z. and A.F.; writing—review and editing C.V., R.M.-L., S.M. and J.L.F.-L. All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

This study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Ethics Committee of Marques de Valdecilla University Hospital.

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

All data are included in the article.

Conflicts of Interest

The authors declare no conflict of interest.

Funding Statement

This work was supported by the Ministerio de Ciencia e Innovación grant PID2021-128220NB-100, the Instituto de Salud Carlos III through grant DTS18/00141, co-funded by the European Regional Development Fund/European Social Fund, “A way to make Europe/Investing in your future”, and the Instituto de Investigación Valdecilla (IDIVAL) (APG/03).

Footnotes

Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

References

  • 1.Ostrom Q.T., Cioffi G., Waite K., Kruchko C., Barnholtz-Sloan J.S. CBTRUS Statistical Report: Primary Brain and Other Central Nervous System Tumors Diagnosed in the United States in 2014–2018. Neuro Oncol. 2021;23:III1–III105. doi: 10.1093/neuonc/noab200. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Stupp R., Mason W.P., van den Bent M.J., Weller M., Fisher B., Taphoorn M.J.B., Belanger K., Brandes A.A., Marosi C., Bogdahn U., et al. Radiotherapy plus Concomitant and Adjuvant Temozolomide for Glioblastoma. N. Engl. J. Med. 2005;352:987–996. doi: 10.1056/NEJMoa043330. [DOI] [PubMed] [Google Scholar]
  • 3.Graus F., Bruna J., Pardo J., Escudero D., Vilas D., Barceló I., Brell M., Pascual C., Crespo J.A., Erro E., et al. Patterns of Care and Outcome for Patients with Glioblastoma Diagnosed during 2008-2010 in Spain. Neuro Oncol. 2013;15:797–805. doi: 10.1093/neuonc/not013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Tejada Solís S., Plans Ahicart G., Iglesias Lozano I., de Quintana Schmidt C., Fernández Coello A., Hostalot Panisello C., Ley Urzaiz L., García Romero J.C., Díez Valle R., González Sánchez J., et al. Glioblastoma Treatment Guidelines: Consensus by the Spanish Society of Neurosurgery Tumor Section. Neurocirugia. 2020;31:289–298. doi: 10.1016/j.neucir.2020.06.001. [DOI] [PubMed] [Google Scholar]
  • 5.Velásquez C., Mansouri S., Mora C., Nassiri F., Suppiah S., Martino J., Zadeh G., Fernández-Luna J.L. Molecular and Clinical Insights into the Invasive Capacity of Glioblastoma Cells. J. Oncol. 2019;2019:1740763. doi: 10.1155/2019/1740763. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Bastola S., Pavlyukov M.S., Yamashita D., Ghosh S., Cho H., Kagaya N., Zhang Z., Minata M., Lee Y., Sadahiro H., et al. Glioma-Initiating Cells at Tumor Edge Gain Signals from Tumor Core Cells to Promote Their Malignancy. Nat. Commun. 2020;11:4660. doi: 10.1038/s41467-020-18189-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Kubben P.L., ter Meulen K.J., Schijns O.E.M.G., ter Laak-Poort M.P., van Overbeeke J.J., van Santbrink H. Intraoperative MRI-Guided Resection of Glioblastoma Multiforme: A Systematic Review. Lancet Oncol. 2011;12:1062–1070. doi: 10.1016/S1470-2045(11)70130-9. [DOI] [PubMed] [Google Scholar]
  • 8.Wirtz C.R., Albert F.K., Schwaderer M., Heuer C., Staubert A., Tronnier V.M., Knauth M., Kunze S. The Benefit of Neuronavigation for Neurosurgery Analyzed by Its Impact on Glioblastoma Surgery. Neurol. Res. 2000;22:354–360. doi: 10.1080/01616412.2000.11740684. [DOI] [PubMed] [Google Scholar]
  • 9.Stummer W., Pichlmeier U., Meinel T., Wiestler O.D., Zanella F., Reulen H.J., ALA-Glioma Study Group Fluorescence-Guided Surgery with 5-Aminolevulinic Acid for Resection of Malignant Glioma: A Randomised Controlled Multicentre Phase III Trial. Lancet Oncol. 2006;7:392–401. doi: 10.1016/S1470-2045(06)70665-9. [DOI] [PubMed] [Google Scholar]
  • 10.Stummer W., Koch R., Valle R.D., Roberts D.W., Sanai N., Kalkanis S., Hadjipanayis C.G., Suero Molina E. Intraoperative Fluorescence Diagnosis in the Brain: A Systematic Review and Suggestions for Future Standards on Reporting Diagnostic Accuracy and Clinical Utility. Acta Neurochir. 2019;161:2083–2098. doi: 10.1007/s00701-019-04007-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Acerbi F., Pollo B., De Laurentis C., Restelli F., Falco J., Vetrano I.G., Broggi M., Schiariti M., Tramacere I., Ferroli P., et al. Ex Vivo Fluorescein-Assisted Confocal Laser Endomicroscopy (CONVIVO® System) in Patients With Glioblastoma: Results From a Prospective Study. Front. Oncol. 2020;10:606574. doi: 10.3389/fonc.2020.606574. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Ebbesen T.W., Lezec H.J., Ghaemi H.F., Thio T., Wolff P.A. Extraordinary Optical Transmission through Sub-Wavelength Hole Arrays. Nature. 1998;391:667–669. doi: 10.1038/35570. [DOI] [Google Scholar]
  • 13.Liang W., Wang S., Festa F., Wiktor P., Wang W., Magee M., Labaer J., Tao N. Measurement of Small Molecule Binding Kinetics on a Protein Microarray by Plasmonic-Based Electrochemical Impedance Imaging. Anal. Chem. 2014;86:9860–9865. doi: 10.1021/ac5024556. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Soler M., Estévez M.-C., Cardenosa-Rubio M., Astua A., Lechuga L.M. How Nanophotonic Label-Free Biosensors Can Contribute to Rapid and Massive Diagnostics of Respiratory Virus Infections: COVID-19 Case. ACS Sens. 2020;5:2663–2678. doi: 10.1021/acssensors.0c01180. [DOI] [PubMed] [Google Scholar]
  • 15.Zhou X.L., Yang Y., Wang S., Liu X.W. Surface Plasmon Resonance Microscopy: From Single-Molecule Sensing to Single-Cell Imaging. Angew. Chem. Int. Ed. 2020;59:1776–1785. doi: 10.1002/anie.201908806. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Toma K., Kano H., Offenhäusser A. Label-Free Measurement of Cell-Electrode Cleft Gap Distance with High Spatial Resolution Surface Plasmon Microscopy. ACS Nano. 2014;8:12612–12619. doi: 10.1021/nn505521e. [DOI] [PubMed] [Google Scholar]
  • 17.Escobedo C. On-Chip Nanohole Array Based Sensing: A Review. Lab A Chip. 2013;13:2445–2463. doi: 10.1039/c3lc50107h. [DOI] [PubMed] [Google Scholar]
  • 18.Lee Y.K., Lee K.S., Kim W.M., Sohn Y.S. Detection of Amyloid-Β42 Using a Waveguide-Coupled Bimetallic Surface Plasmon Resonance Sensor Chip in the Intensity Measurement Mode. PLoS ONE. 2014;9:e98992. doi: 10.1371/journal.pone.0098992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Ertürk G., Özen H., Tümer M.A., Mattiasson B., Denizli A. Microcontact Imprinting Based Surface Plasmon Resonance (SPR) Biosensor for Real-Time and Ultrasensitive Detection of Prostate Specific Antigen (PSA) from Clinical Samples. Sens. Actuators B Chem. 2016;224:823–832. doi: 10.1016/j.snb.2015.10.093. [DOI] [Google Scholar]
  • 20.Mauriz E., Dey P., Lechuga L.M. Advances in Nanoplasmonic Biosensors for Clinical Applications. Analyst. 2019;144:7105–7129. doi: 10.1039/C9AN00701F. [DOI] [PubMed] [Google Scholar]
  • 21.Franco A., Vidal V., Gómez M., Gutiérrez O., Martino M., González F., Moreno F., Fernández-Luna J.L. A Label-Free Optical System with a Nanohole Array Biosensor for Discriminating Live Single Cancer Cells from Normal Cells. Nanophotonics. 2021;11:315–328. doi: 10.1515/nanoph-2021-0499. [DOI] [Google Scholar]
  • 22.Barreda Á.I., Otaduy D., Martín-Rodríguez R., Merino S., Fernández-Luna J.L., González F., Moreno F. Electromagnetic Behavior of Dielectric Objects on Metallic Periodically Nanostructured Substrates. Opt. Express. 2018;26:11222–11237. doi: 10.1364/OE.26.011222. [DOI] [PubMed] [Google Scholar]
  • 23.Franco A., Otaduy D., Barreda A.I., Fernández-Luna J.L., Merino S., González F., Moreno F. Optical Inspection of Manufactured Nanohole Arrays to Bridge the Lab-Industry Gap. Opt. Laser Technol. 2019;116:48–57. doi: 10.1016/j.optlastec.2019.03.010. [DOI] [Google Scholar]
  • 24.Carcelen M., Vidal V., Franco A., Gomez M., Moreno F., Fernandez-Luna J.L. Plasmonic Biosensing for Label-Free Detection of Two Hallmarks of Cancer Cells: Cell-Matrix Interaction and Cell Division. Biosensors. 2022;12:674. doi: 10.3390/bios12090674. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Louis D.N., Perry A., Wesseling P., Brat D.J., Cree I.A., Figarella-Branger D., Hawkins C., Ng H.K., Pfister S.M., Reifenberger G., et al. The 2021 WHO Classification of Tumors of the Central Nervous System: A Summary. Neuro Oncol. 2021;23:1231–1251. doi: 10.1093/neuonc/noab106. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Martínez-Camblor P., Pardo-Fernández J.C. The Youden Index in the Generalized Receiver Operating Characteristic Curve Context. Int. J. Biostat. 2019;15:20180060. doi: 10.1515/ijb-2018-0060. [DOI] [PubMed] [Google Scholar]
  • 27.Park Y.K., Diez-Silva M., Popescu G., Lykotrafitis G., Choi W., Feld M.S., Suresh S. Refractive Index Maps and Membrane Dynamics of Human Red Blood Cells Parasitized by Plasmodium Falciparum. Proc. Natl. Acad. Sci. USA. 2008;105:13730–13735. doi: 10.1073/pnas.0806100105. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Louis D.N., Perry A., Reifenberger G., von Deimling A., Figarella-Branger D., Cavenee W.K., Ohgaki H., Wiestler O.D., Kleihues P., Ellison D.W. The 2016 World Health Organization Classification of Tumors of the Central Nervous System: A Summary. Acta Neuropathol. 2016;131:803–820. doi: 10.1007/s00401-016-1545-1. [DOI] [PubMed] [Google Scholar]
  • 29.Chatterjee S. Artefacts in Histopathology. J. Oral Maxillofac. Pathol. 2014;18:111–116. doi: 10.4103/0973-029X.141346. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Plesec T.P., Prayson R.A. Frozen Section Discrepancy in the Evaluation of Central Nervous System Tumors. Arch. Pathol. Lab. Med. 2007;131:1532–1540. doi: 10.5858/2007-131-1532-FSDITE. [DOI] [PubMed] [Google Scholar]
  • 31.Wei L., Roberts D.W., Sanai N., Liu J.T.C. Visualization Technologies for 5-ALA-Based Fluorescence-Guided Surgeries. J. Neurooncol. 2019;141:495–505. doi: 10.1007/s11060-018-03077-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Acerbi F., Broggi M., Schebesch K.M., Höhne J., Cavallo C., De Laurentis C., Eoli M., Anghileri E., Servida M., Boffano C., et al. Fluorescein-Guided Surgery for Resection of High-Grade Gliomas: A Multicentric Prospective Phase II Study (FLUOGLIO) Clin. Cancer Res. 2018;24:52–61. doi: 10.1158/1078-0432.CCR-17-1184. [DOI] [PubMed] [Google Scholar]
  • 33.Restelli F., Bonomo G., Monti E., Broggi G., Acerbi F., Broggi M. Safeness of Sodium Fluorescein Administration in Neurosurgery: Case-Report of an Erroneous Very High-Dose Administration and Review of the Literature. Brain Spine. 2022;2:101703. doi: 10.1016/j.bas.2022.101703. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Kiesel B., Wadiura L.I., Mischkulnig M., Makolli J., Sperl V., Borkovec M., Freund J., Lang A., Millesi M., Berghoff A.S., et al. Efficacy, Outcome, and Safety of Elderly Patients with Glioblastoma in the 5-Ala Era: Single Center Experience of More than 10 Years. Cancers. 2021;13:6119. doi: 10.3390/cancers13236119. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Lau D., Hervey-Jumper S.L., Chang S., Molinaro A.M., McDermott M.W., Phillips J.J., Berger M.S. A Prospective Phase II Clinical Trial of 5-Aminolevulinic Acid to Assess the Correlation of Intraoperative Fluorescence Intensity and Degree of Histologic Cellularity during Resection of High-Grade Gliomas. J. Neurosurg. 2016;124:1300–1309. doi: 10.3171/2015.5.JNS1577. [DOI] [PubMed] [Google Scholar]
  • 36.Chung I.W.H., Eljamel S. Risk Factors for Developing Oral 5-Aminolevulenic Acid-Induced Side Effects in Patients Undergoing Fluorescence Guided Resection. Photodiagnosis Photodyn. Ther. 2013;10:362–367. doi: 10.1016/j.pdpdt.2013.03.007. [DOI] [PubMed] [Google Scholar]
  • 37.Belykh E., Zhao X., Ngo B., Farhadi D.S., Byvaltsev V.A., Eschbacher J.M., Nakaji P., Preul M.C. Intraoperative Confocal Laser Endomicroscopy Ex Vivo Examination of Tissue Microstructure During Fluorescence-Guided Brain Tumor Surgery. Front. Oncol. 2020;10:599250. doi: 10.3389/fonc.2020.599250. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Xu Y., Abramov I., Belykh E., Mignucci-Jiménez G., Park M.T., Eschbacher J.M., Preul M.C. Characterization of Ex Vivo and in Vivo Intraoperative Neurosurgical Confocal Laser Endomicroscopy Imaging. Front. Oncol. 2022;12:979748. doi: 10.3389/fonc.2022.979748. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Dixon L., Lim A., Grech-Sollars M., Nandi D., Camp S. Intraoperative Ultrasound in Brain Tumor Surgery: A Review and Implementation Guide. Neurosurg. Rev. 2022;45:2503–2515. doi: 10.1007/s10143-022-01778-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Mahboob S., McPhillips R., Qiu Z., Jiang Y., Meggs C., Schiavone G., Button T., Desmulliez M., Demore C., Cochran S., et al. Intraoperative Ultrasound-Guided Resection of Gliomas: A Meta-Analysis and Review of the Literature. World Neurosurg. 2016;92:255–263. doi: 10.1016/j.wneu.2016.05.007. [DOI] [PubMed] [Google Scholar]
  • 41.Blystad I., Marcel Warntjes J.B., Smedby O., Lundberg P., Larsson E.M., Tisell A. Quantitative MRI for Analysis of Peritumoral Edema in Malignant Gliomas. PLoS ONE. 2017;12:e0177135. doi: 10.1371/journal.pone.0177135. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Rogers C.M., Jones P.S., Weinberg J.S. Intraoperative MRI for Brain Tumors. J. Neuro-Oncol. 2021;151:479–490. doi: 10.1007/s11060-020-03667-6. [DOI] [PubMed] [Google Scholar]
  • 43.Giordano M., Samii A., Lawson McLean A.C., Bertalanffy H., Fahlbusch R., Samii M., Di Rocco C. Intraoperative Magnetic Resonance Imaging in Pediatric Neurosurgery: Safety and Utility. J. Neurosurg. Pediatr. 2017;19:77–84. doi: 10.3171/2016.8.PEDS15708. [DOI] [PubMed] [Google Scholar]
  • 44.Rao G. Intraoperative MRI and Maximizing Extent of Resection. Neurosurg. Clin. N. Am. 2017;28:477–485. doi: 10.1016/j.nec.2017.05.003. [DOI] [PubMed] [Google Scholar]
  • 45.Fujii Y., Ogiwara T., Watanabe G., Hanaoka Y., Goto T., Hongo K., Horiuchi T. Intraoperative Low-Field Magnetic Resonance Imaging-Guided Tumor Resection in Glioma Surgery: Pros and Cons. J. Nippon. Med. Sch. 2022;89:269–276. doi: 10.1272/jnms.JNMS.2022_89-301. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

All data are included in the article.