Abstract
Traditional hand-held ultrasound probe has some limitations in prostate biopsy. Improving the localization and accuracy of ultrasound probe will increase the detection rate of prostate cancer while biopsy techniques remain unchanged. This paper designs a manipulator for transrectal ultrasound probe, which assists doctors in performing prostate biopsy and improves the efficiency and accuracy of biopsy procedure. The ultrasound probe manipulator includes a position adjustment module that can lock four joints at the same time. It reduces operating time and improves the stability of the mechanism. We use the attitude adjustment module designed by double parallelogram RCM mechanism, the ultrasound probe can realize centering and prevent its radial motion. The self-weight balance design helps doctors operate ultrasound probe without weight. Using MATLAB to analyze the manipulator, the results show that the workspace of the mechanism can meet the biopsy requirements. And simulate the centering effect of the ultrasound probe when the attitude is adjusted at different feeding distances, the results show that the ultrasound probe is centering stability. Finally, the centering and joint interlocking tests of the physical prototype are completed. In this paper, a 7-DOF manipulator for transrectal ultrasound probe is designed. The mechanism is analyzed for kinematics, workspace analysis, simulation of centering effects, development of a physical prototype and related experimental research. The results show that the surgical demand workspace is located inside the reachable workspace of the mechanism and the joint locking of the manipulator is reliable.
Keywords: Prostate biopsy, ultrasound probe, RCM, manipulator, location, joint interlocking
Introduction
Transrectal prostate biopsy is the gold standard for the diagnosis of prostate cancer. 1 Currently, the commonly method used in prostate biopsy is to perform a biopsy of the prostate through the rectum under the guidance of ultrasound images,2,3 the ultrasound probe is equipped with a needle guide for transrectal access to the prostate (see Figure 1). However, there are still some problems in the biopsy procedure. During the biopsy targeting process, even experienced urologist can’t accurately dissect the target tissues in the expected biopsy procedure. 4 It is also difficult for doctors to perform repeated sampling on the same target. The ultrasound probe is inserted into the patient’s rectum during the biopsy, there is radial motion causing damage to the rectum when adjusting the probe attitude. Related study demonstrated that the use of robotic-assisted ultrasound probe for biopsy can significantly reduce repeated biopsy errors, improve precision and accuracy and increase the average significant prostate cancer detection rate. 5 Robotic prostate biopsy is an emerging technology, Miah et al. reported on of the largest series of men who underwent robotically assisted prostate biopsy. The results have demonstrated a high detection of clinically significant prostate cancer when combined with only limited near-field sampling, shorten the learning curve and added technological security. 6 Schneider et al. designed a 3-DOF motorized reobotic probe, which can provide accurate needle placement and greatly reducing contact with the muscles and nerves. The device is mounted to the ultrasound device, resulting in accurate registration of the needle end-effector with respect to the ultrasound image. 7 Phee et al. designed a percutaneous biopsy of the prostate robot system. Combined with the three-dimensional image reconstruction, the doctor determines the prostate puncture point and the perineal entry point based on experience. The robot system can automatically calculate the optimal needle path and perform three-dimensional simulation before the needle is inserted. The doctor manually operates the biopsy device to obtain the lesion tissue. 8 Yu et al. developed a robotic system with 16 degrees of freedom (DOF) (9 DOF for the positioning module and 7 DOF for the surgery module) for prostate brachytherapy. The system can obtain ultrasound images of the lesion area by rotating operation in an automatic or manual manner.9,10 The experimental results show that the system has achieved good therapeutic effects. Fichtinger et al. developed a robotic system that includes ultrasound image navigation and treatment planning system (TPS). The most important feature of the system is that a small 4 degrees of freedom (DOF) parallel robot is installed directly above the ultrasound probe, so that the coordinate system of the parallel robot and the ultrasound probe adjustment mechanism is the same, ensuring that the system’s workflow and calibration standards have not changed. 11 Stoianovici Dan et al. designed a robotic device for holding and manipulating the ultrasound probe for intraoperative ultrasound image navigation. The TRUS robot has 4 DOFs and is controlled remotely by the surgeon. Robotic-assisted prostate biopsy can reduce prostate motion and deformation errors.12–14 Virtrami et al. designed a manipulator for assisting endorectal prostate biopsy, possessed 6 DOFs. The manipulator has a free mode and a locked mode. Besides, the urologist can switch them at will. Clinical trials were performed with or without the assistance of manipulator. The main purpose was to compare the accuracy of biopsy. The results show that the use of manipulator has a statistically significant increase in biopsy accuracy.15,16 Although active prostate interventional robots have been extensively studied, due to their complex systems and high cost, it takes longer to obtain medical device license certification and longer training time for use. It is more difficult to promote and apply while passive platforms with auxiliary and support functions are also the same. It can meet the requirements of biopsy procedure, so it is necessary to study the passive ultrasound probe manipulator of transrectal puncture.
Figure 1.

Transrectal biopsy of the prostate.
During the biopsy procedure, the ultrasound probe should always be operated with the center of the anus as the rotation center to avoid damage to the rectum and alleviate the pain of the patient. The key technology of the passive prostate assisted interventional device is the centering motion of the transrectal ultrasound probe, and the use of a remote center of motion (RCM) mechanism can achieve the centering motion of the ultrasound probe, thereby liberating the doctor’s hands. Various RCM mechanisms are widely used in robotic-assisted minimally invasive surgery (MIS). There are eight types of RCM mechanisms used for MIS robots, such as isocenters, cicular tracking arcs, parallelograms, synchronous belt transmission, spherical linkages, parallel manipulators, compliant mechanisms and passive RCMs. 17 And the double parallelogram mechanism is most widely used RCM mechanism in MIS robots. 18 Taylor et al. first proposed the concept of “remote-center-of-motion robot” and designed a laparoscopic surgical robot, which consists of a 3-axis linear xyz stage and a 2-axis four-link parallel mechanism providing two rotations. The yaw rotation center and the pitch rotation center of the mechanism intersect at the remote motion center. The posture of the puncture needle can be adjusted arbitrarily in these 2-axis, the position of remote motion center can be adjusted by the expansion of the connecting rod. 19 Planar RCM mechanisms like the double parallelogram mechanism only provide 1 DOF, Li et al. proposed an RCM mechanism type synthesis method based on the intersecting motion planes. By constructing several planar mechanisms with movable planes whose common intersecting line passes through a fix point.20,21 Nisar et al. presents a new RCM mechanism which provides 2 DOFs, including roll and translation DOF, the novelty of the design is that it provides a significantly smaller footprint than the existing 2-DOF plan.22,23 Although the current RCM mechanisms designed by research scholars can meet the requirements of surgery, they all have the problems of complex structure, poor stability, and large space occupation.
In order to solve these problems, this paper analyzes the actual working space and functional requirements of transrectal prostate biopsy. Based on the integration of the current prostate intervention robot system and the remote end adjustment mechanism, a novel 2-DOF RCM mechanism is designed based on the double parallelogram mechanism for ultrasound probe yaw DOF and pitch DOF, so that operate the ultrasound probe to move around a fixed point. In order to ensure the safety of the biopsy and improve the positioning accuracy, reliable locking is required when the ultrasound probe is at any pose in space. Based on this, this paper designs a 7-DOF manipulator for transrectal ultrasound probe, which can realize continuous motion around the centering point, provide reliable joint interlocking, adopt gravity self-balancing design and assist the doctor to operate the ultrasound probe without weight to complete the biopsy procedure.
7-DOF Manipulator structure design
Analysis of the operable space of mechanism
During the transrectal prostate biopsy procedure, in order to reduce the patient’s pain, the left lying position is adopted, as shown in Figure 2. The back of the body is 30–45° from the edge of the operating table, the thigh is 45–60° from the edge of the bed, the distance of the prostate from the bed surface is 150–250 mm. The left lying position can well expose the anus, which provides convenience for the ultrasound probe into the rectum.
Figure 2.

Operable space of mechanism.
Transmission scheme
In view of the particularity of the passive transrectal ultrasound probe manipulator, it is necessary to design a suitable configuration so that it can realize the position and attitude adjustment, rotation and feed adjustment of the ultrasound probe. The first is to analyze the configuration of the position adjustment module. After summary, there are currently six types of robot arm configurations: cartesian, cylindrical, polar, articulated, SCARA, parallel robot, 24 This paper compares various structure types of the manipulator from several indicators such as workspace, occupied space, positioning accuracy, intuitiveness of movement, and the effect of joint motion on the end-effector posture. The specific analysis and comparison are as show in Tables 1 and 2. 25
Table 1.
Comparison of typical robot arm structure.
| Types | Cartesian | Cylindrical | Polar | SCARA | Articulated | Parallel |
|---|---|---|---|---|---|---|
| Structure diagram |
|
|
|
|
|
|
| Joint configuration | PPP | RPP | RRP | RRP | RRR | RPR |
| Workspace | Small | Middle | Big | Middle | Big | Small |
| Occupied space | Big | Middle | Small | Middle | Small | Big |
| Positioning accuracy | High | Medium | Low | Medium | Low | High |
| Intuitive movement | Strong | Medium | Poor | Medium | Poor | Strong |
| Impact on posture | No | Small | Middle | Small | Big | Small |
Table 2.
The advantages and disadvantages of various types of robot arm.
| Types | Advantage | Disadvantages |
|---|---|---|
| Cartesian | 1. The structural rigidity of the overall system is large; 2. Independent joint movement and convenient driving; 3. The solution of kinematics is simple without singular state; 4. It is easy to design and process. |
1. The system structure takes up a lot of space; 2. The operating space is small; 3. Cannot go around the target object for operation; 4. Poor operation flexibility. |
| Cylindrical | 1. Has a simple kinematics model; 2. Simple design and processing, easy to realize; 3. The operation is flexible, convenient and powerful. |
1. Limited working space; 2. There is overlap in the work area; 3. Mobile joints are not easy to protect. |
| Polar | 1. Large working space; 2. The system takes up little space; 3. Flexible and convenient operation. |
1. Has a complex kinematics model; 2. The design and processing are complicated and difficult to achieve; 3. Mobile joints are not easy to protect. |
| SCARA | 1. Lightweight structure, small size and light weight; 2. Simple dynamic model; 3. Quick response. |
1. The design and processing are more complicated; 2. Poor structural rigidity; 3. The motion relationship is more complicated and the control calculation is large. |
| Articulated | 1. Has high flexibility; 2. Large operation space, compact structure and small space; 3. All adopt rotary joints, which can reach any part of the workpiece with little interference; 4. Suitable for motor drive. |
1. It has a complex kinematics model, complex motion relationships, and a large amount of control calculation; 2. The design and processing are complex and difficult to achieve; 3. It is not intuitive to determine the position and posture of the end piece; 4. Poor structural rigidity. |
| Parallel | 1. Large stiffness to weight ratio; 2. Fast response speed; 3. It is convenient for modular design and low manufacturing cost; 4. Convenient installation, suitable for frequent relocation and easy maintenance. |
1. Small working space; 2. The legs, platform, and base interfere with each other in space; 3. It has a singular position. After the mechanism reaches this position, the platform cannot be pushed by the motor. |
From the above analysis, it can be seen that the SCARA robot has a light structure, small size, light weight, large working space, high positioning accuracy, strong intuitive movement and meets the requirements of surgical operation instruments. At present, medical robots are also used SCARA type structure, the work is more reliable. The articulated robot has compact structure, small space occupation, and high flexibility. Therefore, the configuration of the position adjustment module designed based on the advantages of the SCARA robot and the articulated robot is shown in Figure 3. The module has 3 DOFs, and four rotary joints. The joint link of the joint 2 and the joint 3 adopts a parallelogram mechanism composed of two links, so as to ensure that the end coordinate system only moves along the X axis, Y axis, and Z axis.
Figure 3.
Structure diagram of 7-DOF manipulator.
For the attitude adjustment mechanism, choose a 2-DOF structure. In order to enable the ultrasound probe to achieve centering motion around the center point of the anus during the biopsy procedure, the configuration of the attitude adjustment adopts an RCM mechanism. At present, there are two main ways to achieve telecentric positioning: one is to achieve telecentric motion through multi-joint coupling, but this method is mainly realized by an algorithm, which requires high stability and reliability of the algorithm. Moreover, the safety is poor, the other is through mechanism constraints. 26 Because of the high safety of this method, it is widely used in the research of minimally invasive medical robots. 27 There are four main ways to restrain the attitude mechanism: spherical mechanism, double-arc slide rail mechanism, single-arc slide rail mechanism and telecentric parallelogram mechanism. The comparative analysis is shown in Table 3. Because the attitude adjustment module is located at the end of the position adjustment module, the RCM mechanism must be as lightweight as possible to reduce the load. After comprehensive consideration, the dual parallelogram mechanism was selected for the attitude adjustment module.
Table 3.
Comparison of remote center of motion.
| Mechanism type | Structure diagram | Structural features |
|---|---|---|
| Spherical |
|
Simple structure, poor stability, complicated analysis, and convenient driving. |
| Double-arc slide rail |
|
The structure is complex, bulky, difficult to process, and difficult to drive. |
| Single-arc slide rail |
|
The structure is simple, the space is large, the processing accuracy is high, and the driving is difficult. |
| Telecentric parallelogram |
|
The structure is simple, the weight is light, the configuration is diverse, and the driving is convenient. |
The mechanical structure of the 7-DOF manipulator mainly includes a 3-DOF position adjustment module, a 2-DOF attitude adjustment module, a 2-DOF rotation and feed module of ultrasound probe (see Figure 3). The design scheme can complete the basic functions of operating the ultrasound probe and make the mechanism flexible enough. The motion relationship between the three modules is completely decoupled and does not interfere with each other, which reduces the complexity and design difficulty of the mechanism.
Attitude adjustment module design
The attitude adjustment module adopts an RCM mechanism based on a double parallelogram. In order to facilitate the description of the characteristics of the mechanism, the schematic diagram is shown in Figure 4, the double parallelogram is parallelogram ADCF and parallelogram BCIJ. In order to conveniently describe the characteristics of the mechanism, the local coordinate system is established in Figure 4. When the link DF swings α angle with point D as the center, the double parallelogram mechanism has the following characteristics:
Figure 4.

Schematic diagram of attitude adjustment module.
Parallel characteristics, AD // BI // CJ, AC // DF // IJ;
Centering characteristics, the position of the intersection point H of the AD extension line and the JI extension line is always unchanged.
When the doctor performs the biopsy procedure, the interserction point H is at the center point of anus (see Figure 5). In order to prevent the attitude adjustment module from interfering with the patient’s body, according to the size of the human body, , , the length of the longitudinal link For , considering the rotation of the ultrasound probe and the size occupied by the feed module, take , , the length of the transverse link is . According to the physiological characteristics of the human body, the general space occupied by the prostate is a rectangular parallelepiped of . 28 Figure 5 shows the spatial state of the prostate in the left lateral position of the patient and the spatial analysis of the pendulum angle of the ultrasound probe scanning the prostate. β is the swing angle of the ultrasound probe in the transverse section, and the pitch angle γ is the swing angle of the ultrasound probe in the longitudinal section. Because the distance from the prostate to the center of the anus in human body is 60 mm, 29 . considering that the volume of the prostate will increase by after filling with water, the space volume of the prostate is when calculating the swing angle of the ultrasound probe. Point O is the central point of the space volume of the prostate, M, K are the midpoint of the longitudinal section, P is the endpoint, then:
Figure 5.

Analysis of swing angle of ultrasound probe.
| (1) |
| (2) |
| (3) |
In order for the ultrasound probe to cover the entire area of the prostate, β and γ should satisfy the following equation:
| (4) |
| (5) |
then , can be obtained:
Rotation and feed module design
The mounting block is connected to the attitude adjustment module, and two parallelly placed polished rod are slid in the mounting block for achieving the feed freedom of the ultrasound probe along the center of the anus. The fixed ring is mounted on the polished rod, and the rotating cylinder is nested in the fixed ring. The rotating ring is mounted on the other end of the rotating cylinder to restrict the axial movement of the rotating cylinder in the fixed ring (see Figure 6). The side of the fixed ring defines a circular groove with a central angle of 90°, and a limit pin is mounted on the rotating ring. And sliding in the circular groove can limit the rotation angle of the rotating cylinder in the fixed cylinder, which realizes the rotation DOF of the ultrasound probe (see Figure 7). In this design, six hand-screws are used to mount the two ends of the rotating cylinder. A rubber clamping block is mounted on each of the hand-screws. The ultrasound probe can be clamped or loosened by manually adjusting the position of each of the hand-screws.
Figure 6.

Rotation and feed module of ultrasound probe structure diagram.
Figure 7.

Rotation and feed module of ultrasound probe structure diagram.
Joint interlocking mechanism design
The joint 5 and the joint 6 of the attitude adjustment module are rotating joints for setting the scanning posture of the ultrasound probe relative to the prostate. Once the ultrasound probe reaches the required posture, the joint is locked to meet the operational precision of the terminal device. Because the attitude adjustment module needs to be frequently locked or moved during the biopsy procedure, it is necessary to ensure the operational flexibility and the convenience of locking. The electromagnetic power off brake can be selected as the locking part of the joint, the above requirements are well met (see Figure 8). When the electromagnetic brake is powered off and then the brake is engaged, the ultrasound probe can stably maintain a given posture. However, when the electromagnetic brake is disengaged, the ultrasound probe and the attitude adjustment module will naturally fall due to their own weight, so doctors need more force to operate the ultrasound probe, which is easy to cause operational fatigue. In this way, the gravity self-balancing design is adopted. Add a spring to the linkage to balance the heavy moment caused by its own weight (see Figure 9). When the transverse link is in the horizontal position and the feed distance of the ultrasound probe is 0 mm, the gravity moment generated by the attitude adjustment module on the pitch axis is the largest. For the convenience of calculation, the total gravity moment is decomposed into three component moments for solution, as shown in Figure 10. Using the mass properties in the SolidWorks software, the mass of the two transverse links is kg, the mass of the two longitudinal links is kg, the mass of the ultrasound probe rotation and feed module is kg and find the centroids of the parts get mm, mm, mm, .
Figure 8.

Structure of yaw and pitch rotation joint.
Figure 9.

Improved mechanism of system.
Figure 10.

Total gravity moment of pitch axis.
The total gravity moment of the attitude adjustment module is:
The position adjustment module has a total of four joints. If each joint is locked separately, it will not only increase the weight of the module, but also occupy the doctor’s operating mechanism time. The designed joint interlocking mechanism solves this problem by separating the structures of the joint 1 and the joint 3 into upper and lower parts, as shown in Figure 11. The joint 1 connected to the upper arm in the upper half continues to separate into two parts, as shown in Figure 12. When a pair of forces F of equal size and opposite directions are applied to the A-side and B-side of the joint 1, the joint 1 can clamp both the post and the connecting rod due to the compression, thereby limiting the θ1 and θ2. When a parallelogram link is used to transmit the force F to the A-side and B-side of the joint 4, the joint 4 can clamp the connecting rod and the rotating shaft mounted on the arm, thereby limiting the angle of θ3 and θ4. The applied force F is generated by the hand-tightening bolts installed from the A-side of the joint 1 and the nuts fixed to the B-side of the joint 1. The interlock mechanism not only realizes the simultaneous locking of the four joints in the position adjustment module, simplifies the doctor’s operation, but also has a simple structure, low cost, and light weight.
Figure 11.

Interlocking mechanism diagram.
Figure 12.

Improved structure of joint 1.
Workspace analysis and prototype experimental study
Workspace analysis
The manipulator needs to have a large working space, high precision, and flexible pose during the biopsy procedure. Therefore, it is necessary to analyze the motion analysis and attitude adjustment of the manipulator. It is also an intuitive factor that reflects the performance of the manipulator. The doctor operates an ultrasound probe to reach the remote motion center of the mechanism to the center point of the patient’s anus. Because the attitude adjustment module dose not affect the position of the remote motion center, the set of point that the remote motion center can reach is the motion space of the mechanism. The connection between the end point of the scanning portion of the ultrasound probe and the end point of the hand-held part can express the motion posture of the ultrasound probe.
According to the motion range of each joint of the position adjustment module to set the simulation parameters as shown in Table 4. As can be seen from the Figure 13, the red area represents the range of the position of the central point of the anal periphery when patients of different sizes lie on the operating bed. And the blue area represents the range of movement of the remote motion center when the ultrasound probe position adjustment mechanism adjusts the position. The range of motion of the telecentric point completely covers the position range of the central point, so the mechanism can well meet the needs of patients of different sizes for position adjustment.
Table 4.
Position adjustment simulation parameters.
| θ 1 | θ 2 | θ 4 | θ 5 | θ 6 | a 6 | α7 |
|---|---|---|---|---|---|---|
| −90~90 | 0~38 | −150~150 | 0 | 0 | 0 | 0 |
Figure 13.
Workspace of remote center point: (a) three-dimensional workspace maps, (b) XOZ plane projection, (c) XOY plane projection, and (d) YOZ plane projection.
The attitude adjustment module enables the ultrasound probe to perform a centering function of swinging around a certain point and simulates the centering effect of the ultrasound probe. The motion parameters are shown in Table 5. When the feed distance , the simulation results are shown in Figure 14. The attitude of the ultrasound probe at different joint angles, the position coordinates of the remote motion center of the mechanism are . When the yaw and pitch angles are adjusted according to a given angle, the scanning end of the ultrasound probe is always at the remote motion center of the mechanism and a cone-shaped space with the point as a vertex is formed.
Table 5.
Attitude simulation parameters of ultrasound probe.
| θ1/deg | θ2/deg | θ4/deg | θ5/deg | θ6/deg | α7/deg |
|---|---|---|---|---|---|
| −30 | 20 | 80 | −20~20 | −25~25 | 0 |
Figure 14.
Attitude of ultrasound probe at 0 mm feeding, these lines to represent the center line of the probe with randomly given colors: (a) three-dimensional plots ( ), and (b) XOY plane projection ( ).
In order to observe the changes of the attitude of the ultrasound probe with different feeds, keep the joint parameters of the position adjustment module and the attitude adjustment module unchanged and only change the feed distance of the ultrasound probe to . The simulation results are shown in Figure 15. The coordinates of the remote motion center of the mechanism are still , therefore, the ultrasound probe still achieves attitude adjustment with the remote motion center as the center point after feeding 30 mm. Change the feed distance of the ultrasound probe to and keep the parameters of the other joints unchanged. Simulate the attitude of the ultrasound probe, the result is shown in Figure 16. As can be seen from the figure, the coordinates of the intersections of all the line segments intersecting in space are still . which meets the design.
Figure 15.
Attitude of ultrasound probe at 30 mm feeding: (a) three-dimensional plots ( ) and (b) XOY plane projection ( ).
Figure 16.
Attitude of ultrasound probe at 60 mm feeding: (a) three-dimensional plots ( ) and (b) XOY plane projection ( ).
Manipulator for prostate biopsy experiment
Before the biopsy, the position adjustment module is dragged by the doctor to achieve the positioning of the mechanism in the vertical and horizontal directions. When the end of the ultrasound probe reaches the center point of the patient’s anus, the position adjustment module is locked to ensure the positioning accuracy (see Figure 17). According to the clinical requirements, the rotation angle ranges from to , the ranges from to , the and the ranges from to . The experimental instrument is an electronic digital angle measuring instrument with an accuracy of 0.1°. Before the measurement, we set the initial position of each joint and then measure separately. The experimental results are shown in Table 6.
Figure 17.

Physical prototype of manipulator.
Table 6.
Angle measurement of each joint in position adjustment module.
| Angle | ||||
|---|---|---|---|---|
| Target range | ||||
| Initial position description | The rod 1 is parallel to the edge of the operating bed ( ) | Rod 1 and rod 2 are collinear ( ) | Rod 2 and rod 3 are collinear ( ) | Rod 3 and rod 4 are collinear ( ) |
| Initial position diagram |
|
|
|
|
| Measurement result |
The experimental results of joint angle measurement show that each joint of the position adjustment module can continuously rotate and reach the required angle range, which satisfies the design requirements.
The distribution of the prostate position relative to the base coordinate system in different somatotype patients from 240 mm to 400 mm in X direction, from 0 mm to 120 mm in Y direction, from 150 mm to 250 mm in Z direction. This area is the surgical demand working space that the remote center point of the mechanism must reach. The shape is a rectangular parallelepiped with eight vertices. Adjust the end of the ultrasound probe to the remote center point of the mechanism, then lock the attitude adjustment module. Adjust the position adjustment module to test the ability of the mechanism to meet the requirements of the surgical workspace, as shown in Figure 18.
Figure 18.
Workspace accessibility measurement: (a) x = 240, y = 0, z = 150, (b) x = 400, y = 0, z = 150, (c) x = 400, y = 0, z = 250, (d) x = 240, y = 0, z = 250, (e) x = 240, y = 120, z = 150, (f) x = 400, y = 120, z = 150, (g) x = 240, y = 120, z = 250, (h) x = 400, y = 120, z = 250, and (i) x = 320, y = 80, z = 200.
After testing, the mechanism can successfully reach the eight vertices at the boundary of the surgical workspace and any positions in the interior to meet the surgical requirements.
Mechanism error analysis
The motion characteristics of the RCM mechanism is the range of the remote motion center in the space when the ultrasound probe performs the attitude adjustment. Under the existing experimental conditions, the motion performance of the RCM mechanism can be evaluated by measuring the error between the spatial fixed point and the remote motion center. Because the positional change of the remote motion center in space is three-dimensional, and a fixed point in space can only measure the error of the RCM mechanism in the two-dimensional plane, two sets of experiments are needed. When the yaw or pitch angle is changed separately, the resulting errors are measured separately, as shown in Figures 19 and 20.
Figure 19.

Measurement of yaw remote center point error.
Figure 20.

Measurement of pitch remote center point error.
Experimental variables: adjust the yaw angle to , , , , , and adjust the pitch angle to , , , , under different feed rates (0 mm, 30 mm, 60 mm).
The experimental data obtained by multiple measurements is shown in Figure 21. The marked points on the coordinate paper are processed and measured, the distance from the remote motion center to the fixed point of the space is within 4 mm. The anal canal of human is 2.5–5 cm in length and 3 cm in diameter when distended, 30 and the diameter of the ultrasound probe (Endocavity 3D 8838, BK Ultrasound) for biopsy is about 16 mm. Therefore, the error of the remote motion center is within an acceptable range, and will not cause harm and discomfort to the patient. According to the results of the two sets of experiments, when the ultrasound probe adjusts the yaw and pitch angles under different feed rates, the positional change of the remote motion center in the space is within the deformation range of the anus, which has high precision that can meet the surgical requirements.
Figure 21.
Remote center point error under different feed rates: (a) remote center point error under the motion of yaw angle and (b) remote center point error under the motion of pitch angle.
Conclusion
In this paper, a manipulator for transrectal ultrasound probe is designed, which includes a position adjustment module that can simultaneously lock four joints. Furthermore, it can improve he stability of the surgeon’s gesture. The attitude adjustment module designed by the double parallelogram RCM mechanism enables the ultrasound probe to achieve the centering function in order to prevent its radial motion. The rotation and feed module of ultrasound probe is designed for easy installation of puncture probe and can clamp the irregular shape ultrasound probe. Weight self-balancing design is used to help doctors to operate ultrasound probe without weight.
Through the workspace analysis of the mechanism, the surgical demand workspace is located inside the reachable workspace of the mechanism, which can meet the surgical requirements. When the attitude is adjusted under different feed distances, the centering effect of the ultrasound probe is simulated. The results show that the centering effect of the ultrasound probe is stable, thus verifying the position and attitude adjustment module of the ultrasound probe meet the design requirements.
The physical prototype of the manipulator is developed and tested. It is proved that the mechanism can flexibly reach any position in working space of the prostate biopsy and locking is reliable. The centering effect of the remote motion center when adjusting the attitude of the ultrasound probe is measured, and the displacement range of remote motion center in the space is within 4 mm. The results show that the centering accuracy of the ultrasound probe when adjusting the attitude under different feed rate meet the surgical requirements.
Author biographies
Yongde Zhang received his B. Sc., M. Sc. and Ph. D. degrees all from Harbin Institute of Technology in 1988, 1993 and 1999, respectively. Now he is a professor and Ph.D. supervisor in Harbin University of Science and Technology. His main research interests include medical robot, education robot and biomimetic robot. The email id of author is zhangyd@hrbust.edu.cn
Dexian Liang received his B. Sc. degrees from Harbin University of Science and Technology in 2017. Now he is a M. Sc. candidate in Harbin University of Science and Technology. His main research interests include medical robot, motion planning. The email id of author is liangdx@hrbust.edu.cn.
Liyi Sun received her B. Sc. degrees from Harbin University of Science and Technology in 2016. Now she is a M. Sc. candidate in Harbin University of Science and Technology. Her main research interests include medical robot, biomimetic robot. The email id of author is sunliyi0906@gmail.com
Xiaowei Guo received his B. Sc. and M. Sc. degrees all from Harbin University of Science and Technology in 2015 and 2018. His main research interests include medical robot, mechanical Design. The email id of author is 904767305@qq.com
Jingang Jiang received his B. Sc., M. Sc. and Ph. D. degrees all from Harbin University of Science and Technology in 2005, 2008 and 2013, respectively. Now he is a professor and Ph.D. supervisor in Harbin University of Science and Technology. His main research interests include medical robot, biomimetic robot. The email id of author is jiangjingang@hrbust.edu.cn
Sihao Zuo received his B. Sc. and M. Sc. degrees all from Harbin University of Science and Technology in 2014 and 2018. Now he is a Ph. D. candidate in Harbin University of Science and Technology. His main research interests include medical robot, control algorithm. The e-mail id of author is bk_robot@163.com
Yanhua Zhang received her B. Sc., M. Sc. degrees all from Harbin Medical University in1985, 2000. Now she is a chief physician in ultrasound imaging department, the affiliated tumour hospital of Harbin Medical University. Her main research interests in ultrasound diagnosis. The email id of author is zhangyanhua09@126.com
Footnotes
The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Funding: The author(s) disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This work was supported by the National Natural Science Foundation of China (Grant No. 51675142) and by the Natural Science Foundation of Heilongjiang Province of China (Grant No. ZD2018013).
ORCID iDs: Yongde Zhang
https://orcid.org/0000-0003-4602-1563
Jingang Jiang
https://orcid.org/0000-0003-0491-9236
References
- 1.Candefjord S, Ramser K, Lindahl OA.Technologies for localization and diagnosis of prostate cancer. J Med Eng Technol 2009; 33(8): 585–603. [DOI] [PubMed] [Google Scholar]
- 2.Nazir B.Pain during transrectal ultrasound-guided prostate biopsy and the role of periprostatic nerve block: what radiologists should know. Korean J Radiol 2014; 15(5): 543–553. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Cicione A, Cantiello F, Damiano R.Transrectal ultrasound guided prostate biopsy. In: Rane A, Turna B, Autorino R, et al. (eds.) Practical tips in urology. London: Springer-Verlag, 2017, pp.269–279. [Google Scholar]
- 4.Han M, Chang D, Kim C, et al. Geometric evaluation of systematic transrectal ultrasound guided prostate biopsy. J Urol 2012; 188(6): 2404–2209. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Kaye DR, Stoianovici D, Han M.Robotic ultrasound and needle guidance for prostate cancer management: review of the contemporary literature. Curr Opin Urol 2014; 24(1): 75–80. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Miah S, Servian P, Patel A, et al. A prospective analysis of robotic targeted MRI-US fusion prostate biopsy using the centroid targeting approach. J Robot Surg 2020; 14(1): 69–74. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Schneider CM, Okamura AM, Fichtinger G. A robotic system for transrectal needle insertion into the prostate with integrated ultrasound. In: Proceedings of the 2004 IEEE international conference on robotics and automation, New Orleans, 26 April–1 May 2004, pp.365–370. New York: IEEE. [Google Scholar]
- 8.Phee L, Xiao D, Yuen J, et al. Ultrasound guided robotic system for transperineal biopsy of the prostate. In: Proceedings of the 2005 IEEE international conference on robotics and automation, Barcelona, Spain, 18–22 April 2005, pp.1315–1320. New York: IEEE. [Google Scholar]
- 9.Yu Y, Podder TK, Zhang YD, et al. Robotic system for prostate brachytherapy. Comput Aided Surg 2007; 12(6): 366–370. [DOI] [PubMed] [Google Scholar]
- 10.Yu Y, Podder T, Zhang Y, et al. Robot-assisted prostate brachytherapy. In: Proceedings of the 2006 international conference on medical image computing and computer-assisted intervention, Copenhagen, Denmark, 1–6 October 2006, pp.41–49. Berlin: Springer. [DOI] [PubMed] [Google Scholar]
- 11.Fichtinger G, Fiene JP, Kennedy CW, et al. Robotic assistance for ultrasound-guided prostate brachytherapy. Med Image Anal 2008; 12(5): 535–545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Han M, Kim C, Mozer P, et al. Tandem-robot assisted laparoscopic radical prostatectomy to improve the neurovascular bundle visualization: a feasibility study. Urology 2011; 77(2): 502–506. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Stoianovici D, Kim C, Schäfer F, et al. Endocavity ultrasound probe manipulators. IEEE ASME Trans Mechatron 2013; 18(3): 914–921. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Kim C, Schäfer F, Chang D, et al. Robot for ultrasound-guided prostate imaging and intervention. In: Proceedings of the 2011 IEEE/RSJ international conference on intelligent robots and systems, San Francisco, 25–30 September 2011, pp.943–948. New York: IEEE. [Google Scholar]
- 15.Vitrani MA, Baumann M, Reversat D, et al. Prostate biopsies assisted by comanipulated probe-holder: first in man. Int J Comput Assist Radiol Surg 2016; 11(6): 1153–1161. [DOI] [PubMed] [Google Scholar]
- 16.Chalard R, Reversat D, Morel G, et al. Precisely positioning the tip of an instrument inserted through an orifice with a free wrist robot: application to prostate biopsies. Int J Comput Assist Radiol Surg 2018; 13(5): 611–618. [DOI] [PubMed] [Google Scholar]
- 17.Kuo CH, Dai JS, Dasgupta P.Kinematic design considerations for minimally invasive surgical robots: an overview. Int J Med Robotics Comput Assist Surg 2012; 8(2): 127–145. [DOI] [PubMed] [Google Scholar]
- 18.Kong K, Li J, Zhang H, et al. Kinematic design of a generalized double parallelogram based remote center-of-motion mechanism for minimally invasive surgical robot. J Med Devices 2016; 10(4): 041006. [Google Scholar]
- 19.Taylor RH, Funda J, Eldridge B, et al. A telerobotic assistant for laparoscopic surgery. IEEE Eng Med Biol Mag 1995; 14(3): 279–288. [Google Scholar]
- 20.Li J, Zhang G, Müller A, et al. A family of remote center of motion mechanisms based on intersecting motion planes. J Mech Des 2013; 135(9): 091009. [Google Scholar]
- 21.Li J, Zhang G, Xing Y, et al. A class of 2-degree-of-freedom planar remote center-of-motion mechanisms based on virtual parallelograms. J Mech Rob 2014; 6(3): 031014. [Google Scholar]
- 22.Nisar S, Endo T, Matsuno F.Design and kinematic optimization of a two degrees-of-freedom planar remote center of motion mechanism for minimally invasive surgery manipulators. J Mech Rob 2017; 9(3): 031013. [Google Scholar]
- 23.Nisar S, Endo T, Matsuno F.Design and optimization of a 2-degree-of-freedom planar remote center of motion mechanism for surgical manipulators with smaller footprint. Mech Mach Theory 2018; 129: 148–161. [Google Scholar]
- 24.Liu D, Wang TM, Zhang YR, et al. Structure synthesis of surgical robot orienting to minimally invasive surgery. Robot 2003; 25(3): 132–135. [Google Scholar]
- 25.Li QZ.Research and development of robotic-assisted microsurgery system. Tianjin: Tianjin University, 2004. [Google Scholar]
- 26.Fu YL, Li K, Pan B.Mechanical structure design and kinematics simulation of endoscope operating robot. Mech Eng 2007; 24(6): 62–65. [Google Scholar]
- 27.Aksungur S.Remote Center of Motion (RCM) mechanisms for surgical operations. Int J Appl Math Electron Comput 2015; 3(2): 119–126. [Google Scholar]
- 28.Kim SB, Cho IC, Min SK.Prostate volume measurement by transrectal ultrasonography: comparison of height obtained by use of transaxial and midsagittal scanning. Korean J Urol 2014; 55(7): 470–474. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Holmes DR, Davis BJ, Goulet CC, et al. Shape characterization of the prostate and urethra applicable to the design of a trans-urethral ultrasound imaging probe for prostate brachytherapy guidance. Int J Radiat Oncol Bio Phys 2005; 63: S548–S549. [Google Scholar]
- 30.Duthie HL.Progress report. Anal continence. Gut 1971; 12(10): 844–852. [DOI] [PMC free article] [PubMed] [Google Scholar]