A review of the effect of skin pigmentation on pulse oximeter accuracy

Raghda Al-Halawani 1,, Peter H Charlton 1,2, Meha Qassem 1,, Panayiotis A Kyriacou 1
PMCID: PMC10391744  PMID: 37172609

Abstract

Objective. Pulse oximetry is a non-invasive optical technique used to measure arterial oxygen saturation (SpO2) in a variety of clinical settings and scenarios. Despite being one the most significant technological advances in health monitoring over the last few decades, there have been reports on its various limitations. Recently due to the Covid-19 pandemic, questions about pulse oximeter technology and its accuracy when used in people with different skin pigmentation have resurfaced, and are to be addressed. Approach. This review presents an introduction to the technique of pulse oximetry including its basic principle of operation, technology, and limitations, with a more in depth focus on skin pigmentation. Relevant literature relating to the performance and accuracy of pulse oximeters in populations with different skin pigmentation are evaluated. Main Results. The majority of the evidence suggests that the accuracy of pulse oximetry differs in subjects of different skin pigmentations to a level that requires particular attention, with decreased accuracy in patients with dark skin. Significance. Some recommendations, both from the literature and contributions from the authors, suggest how future work could address these inaccuracies to potentially improve clinical outcomes. These include the objective quantification of skin pigmentation to replace currently used qualitative methods, and computational modelling for predicting calibration algorithms based on skin colour.

Keywords: pulse oximetry, photoplethysmography, COVID-19, skin pigmentation, accuracy

1. Introduction

A pulse oximeter is a non-invasive and low cost, optical device, based on a two wavelength photoplethysmography (PPG) system, used for the continuous monitoring of arterial oxygenation. Pulse oximetry has been found to work on the finger, the ear, the bridge of the nose, the nasal septum, the temple over the temporal artery, and on the foot or palm in infants. Arterial oxygen saturation measurement plays a central role in the diagnosis of sleep-related respiratory disorders such as obstructive sleep apnoea (OSA), monitoring and treatment of respiratory diseases by detecting hypoxaemia, assisting the titration of supplemental oxygen treatment in preterm neonates, and many more (Kyriacou and Allen 2021). Pulse oximetry is now routinely used in a wide range of clinical settings, ranging from hospital care to primary care to home monitoring (Kelleher 1989, Morris et al 1989, Tremper and Barker 1989, Wilkins et al 1989, Falconer and Robinson 1990, Eichhorn 1993, Lawless 1994, Wahr et al 1995, Dumas et al 1996, Shah et al 2012). Pulse oximeters have been shown to be an effective tool for identifying the need for hospitalisation in initially non-severe and possibly high-risk COVID-19 patients, or for discharging known or suspected Covid patients (Levitan 2020, Luks and Swenson 2020, Philip et al 2020, Quaresima and Ferrari 2020, Rodriguez 2020, Teo 2020, Tobin et al 2020, Shah 2020, Brouqui 2021, Cajanding 2021, Cysewska-Sobusiak 2021, Gootenberg et al 2021, Greenhalgh et al 2021, Lancet et al 2021, Michard et al 2021, Ngiam et al 2021, Wilson-Baig et al 2021). Hence, pulse oximetry is now widely used in clinical and consumer settings (Teo 2020), underlining the need to ensure it is as accurate and reliable as possible. Figure 1 provides a succinct summary of the key events in the history of pulse oximeter technology.

Figure 1.

Figure 1.

Key events in the history of pulse oximeter technology.

There are rising concerns of the differing performance of pulse oximetry in patients with different skin pigmentations (Wang and Poh 1985, Cecil et al 1988, Gabrielczyk and Buist 1988, Ries et al 1989, Cahan et al 1990, Jubran and Tobin 1990, Ralston et al 1991, Volgyesi and Spahr-Schopfer 1991, 1992, Lee et al 1993, Bothma et al 1996, Adler et al 1998, Bickler et al 2005, Reuss 2005, Feiner et al 2007, Witting and Scharf 2008, Foglia et al 2017, Ebmeier et al 2018, Murphy and Omar 2018, Guber et al 2019, Smith and Hofmeyr 2019, Badgujar et al 2020, Dyer 2020, Michael et al 2020, Philip et al 2020, Tobin et al 2020, Baker et al 2021, Hunasikatti 2021, Philip et al 2021, Sjoding et al 2021, Todd 2021, Valbuena et al 2021, Vesoulis et al 2021, Whitehead 2021, Wiles et al 2021, Wong et al 2021, Baker and Wilson 2022, Bangash et al 2022, Burnett et al 2022, Crooks et al 2022, Fawzy et al 2022, Ferrari et al 2022, Henry et al 2022, Holder and Wong 2022, Knight et al 2022, Kyriacou et al 2022, Okunlola et al 2022, Shi et al 2022, Tobin and Jubran 2022, Wiles et al 2022). The concept that the accuracy of pulse oximetry may differ between people of different skin pigmentations is not new. Rather, its importance has increased in the Covid-19 pandemic (Greenhalgh et al 2021), since the deterioration of blood oxygenation is one of the primary symptoms of severe Covid-19. The limitations of pulse oximetry, including the impact of ambient light, difficulties in obtaining measurements during low perfusion and motion, etc, are known (Jubran 2015) and guidance on how to minimise their effect is available (GOV.UK 2021). However, the increased mortality rate amongst ethnic minority patients since 2020 from the Covid-19 outbreak has raised the question of whether differential accuracy of pulse oximetry due to skin pigmentation may be a contributing factor to this health inequality. Consequently, this review aims to summarise the literature on the effect of skin pigmentation on pulse oximeter accuracy, and based on this, recommend solutions for minimising these effects.

2. Principles of pulse oximetry

2.1. The gold standard: blood gas analysis

Blood gas analysis is currently the gold standard for blood oxygenation measurement. It was the only method available before the introduction of non-invasive blood oxygenation measurements. Blood gas analysers consist of three electrodes that measure pH, partial pressure of carbon dioxide (PaCO2) and partial pressure of oxygen (PaO2) at 37 °C using extracted blood samples (Severinghaus 2006).

Two main indicators are used to reveal the amount of oxygen present in blood, namely functional arterial oxygen saturation (SaO2) and fractional arterial oxygen saturation (FrO2). Both indices represent the percentage content of oxygen-bound haemoglobin (HbO2) against the total haemoglobin present. This is typically performed by measuring the concentration of the different haemoglobin species of interest. SaO2, expressed as a percentage, is then determined by the ratio of HbO2 concentration over the sum concentration of HbO2 and deoxyhaemoglobin (HHb) in blood

SaO2=[HbO2]HbO2+[HHb].

Similarly, the fractional saturation, FrO2, is determined by the proportion of oxygenated haemoglobin over the total concentration of all haemoglobin species

FrO2=[HbO2]HbO2+HHb+COHb+[MetHb],

where [COHb], and [MetHb] are respectively the concentrations of carboxyhaemoglobin and methaemoglobin. FrO2 is a more accurate method for determining oxygenation of blood, especially when the presence of other dyshaemoglobins (haemoglobins that do not bind to oxygen) is suspected due to particular physiological conditions.

2.2. Peripheral oxygen saturation

Pulse oximetry is non-invasive, low-cost, safe and is currently performed routinely on all surgical patients during admission, intraoperatively and postoperatively. It is more readily available than arterial blood gas analysis for the measurement of oxygen saturation and provides continuous monitoring. It has largely replaced the blood gas method in many clinical settings, unless carbon dioxide or acid-base status is specifically required (Smith and Hofmeyr 2019).

Pulse oximeters (figure 2) estimate functional arterial oxygen saturation by measuring changes in light absorbance in the arterioles over time, as they contain a higher concentration of HbO2 relative to HHb. The photoplethysmogram (PPG) is used to detect blood volume changes and to differentiate between absorbance of arterial blood and other absorbers (skin, bone, venous blood). A good choice of wavelength is where there are large differences in the extinction coefficients (ε) of HbO2 and HHb (Webster ) (figure 3). Another criterion for the wavelength selection is the relative flatness of the absorption spectra around the chosen wavelength (Moyle 1994, Mannheimer et al 1997b). The two conventional wavelengths used in pulse oximetry are the 660 nm (red) and 940 nm (near infra-red). Measurement of arterial oxygen saturation by pulse oximetry is denoted by SpO2, while the term SaO2 is generally reserved for arterial oxygen saturation measured by blood gas analysis in extracted blood (Nitzan et al 2020).

Figure 2.

Figure 2.

A pulse oximeter device for the non-invasive measurement of arterial oxygen saturation, available from pharmacies for home monitoring.

Figure 3.

Figure 3.

Absorption spectra of oxygenated (HbO2) and deoxygenated (HHb) haemoglobin between the visible and near-infra-red region. Molar extinction coefficients of both haemoglobin species are shown with respect to the wavelengths of interest in pulse (Kyriacou and Allen 2021). Reprinted from Kyriacou et al 2022, Copyright (2022), with permission from Elsevier.

The Beer–Lambert law forms the basis of light absorption measurements where the effects of scattering are negligible (Mannheimer et al 1997a). It relates the attenuation of light to the properties of the material through which the light is travelling, which is assumed to be an absorbing-only medium (Nitzan et al 2020):

It=I0eεCd,

where It is the intensity of transmitted (i.e. received) light, I0 is the intensity of the incident light, d is the optical pathlength of the tissue under investigation (the source-detector distance in reflectance mode), ε is the molar extinction coefficient of tissue layers and chromophores such as haemoglobin, melanin, water, etc, and C refers to their

2.2.1. Concentrations

The Beer–Lambert law does not account for the scattering of light as it passes through tissue. To address this, the Beer–Lambert law can be modified to account for scattering by: (i) introducing a term, G, representing ‘light-loss due to scattering’, and (ii) replacing the pathlength, d, with the optical pathlength, l, where the optical pathlength represents the actual distance travelled by light, which is longer than the pathlength due to scattering (Nitzan et al 2020). The modified Beer–Lambert law is (Nitzan et al 2020):

It=I0e(GεCl).

The modified Beer–Lambert law can be used to estimate SpO2 as described in (Nitzan et al 2020). To do so, measurements of the intensity of transmitted light, I t , are extracted at points of minimum and maximum absorptions during the cardiac cycle (see figure 4). This produces a pair of equations which are combined to eliminate I0. It is assumed that G is approximately constant throughout a cardiac cycle, thereby eliminating G from the resulting equation. I is also assumed to be constant throughout a cardiac cycle. This produces the following equation relating the PPG measurements (I D and I S at maximum and minimum absorptions respectively) to the maximal change in haemoglobin concentration during a cardiac cycle, ΔC (Nitzan et al 2020):

IDIS/IS=εΔCl.

This equation is applied twice, once to I D and I S extracted from a PPG signal obtained using a red light, and then again to I D and I S extracted from a PPG signal obtained using infra-red light. Assuming that ΔC is approximately equal at the two wavelengths, the two resulting equations are combined to produce the ‘ratio of ratios’, R (Nitzan et al 2020):

R(IDIS)/IS660(IDIS)/IS940=ε660l660ε940l940

where the subscripts ‘660’ and ‘940’ indicate the two wavelengths. The extinction coefficients can then be expressed in terms of blood oxygen saturation, SaO2, using (Nitzan et al 2020):

ε=εOSaO2+εD1SaO2,

where ε O and ε D are extinction coefficients for oxygenated and deoxygenated blood respectively. Substituting equation (7) into equation (6), and rearranging, an expression is obtained relating SpO2 to known quantities (ε) and the measured R:

SpO2=εD660Rl940l660εD940Rl940l660εO940εD940+εD660εO660.

Note that all these values are known constants expect R and the optical pathlengths at the two wavelengths. Consequently, most modern pulse oximeters convert measurements of R to SpO2 using a ‘look-up table’ (Webster 1997b), based on the assumption that l 940 /l 660 remains constant between individuals, using the following equation:

R=AC660DC660AC940DC940,

where [ac660] and [dc660] represent the alternating current and direct current from tissue layers at 660 nm (a red wavelength) respectively, and likewise for [ac940] and [dc940] at 940 nm (an infra-red wavelength). Manufacturers calibrate pulse oximeters empirically by correlating the measured ratio (R) of AC/DC signals (figure 5). Most modern pulse oximeters convert these ratios to SpO2 using a ‘look-up table’ (Webster 1997b).

Figure 4.

Figure 4.

Intensity of red and infra-red PPG signals during diastolic (Id) and systolic (Is) absorption.

Figure 5.

Figure 5.

PPG parameters used for the calculation of the ratio of ratios (R). The AC amplitude at a generic wavelength λ is obtained from the difference between minimum and maximum absorption (squares) during the cardiac cycle. The DC component is the average light intensity. These two parameters are extracted from red and infra-red wavelengths for calculation of the ratio of ratios (R) in equation (5) (Kyriacou and Allen 2021). Reprinted from Kyriacou et al 2022, Copyright (2022), with permission from Elsevier.

This approach assumes that the relationship between R and SaO2 is constant across different individuals, which would be a reasonable assumption if the inter and intra variability of skin components in individuals are disregarded. For instance, the magnitude of absorbance and the optical pathlength is influenced by the presence of melanin in tissue. Since there is a directly proportional relationship between melanin concentration and μ a , the value of μ a_epidermis in individuals with dark skin is greater than in individuals with fair skin, which results in increased light absorption (Bashkatov et al 2000). Additionally, the type of light scattering, i.e. Rayleigh and Mie scattering, is determined by the size of the scatterer relative to the selected wavelength, which has been found to be larger for individuals with darker skin (Zonios et al 2001). As a result, photons would have a tendency to travel in a more forward direction (Mie scattering) in individuals with dark skin. Hence, current pulse oximeter calibrations may not be captured for wide populations, which may be contributing to the differing performance of pulse oximeters in patients with different skin pigmentations.

2.3. Modalities of pulse oximetry

There are two geometrical configurations of pulse oximetry, namely transmittance and reflectance mode. In transmittance mode PPG, the source and detector are placed opposite to one another across the surface of the region of interest (ROI), most commonly the finger. Therefore, the optical path length is affected by the amount of light scattering within the tissue, as well as the tissue thickness, which alters the distance between the fixed source and detector. However, in reflectance mode PPG, the source and detector are placed adjacent to each other and the light detected is back scattered and reflected within the tissue. Reflectance PPG offers greater flexibility than transmittance PPG in regards to the position of the sensors in different anatomical locations, such as the forehead. The source-detector separation can be altered by changing the distance between the source and detector, resulting in differences in the behaviour of light. For instance, if the desired outcome is to achieve maximum photon penetration depth, the source-detector separation can be increased in order to see the interaction of light with deeper tissue. In both modalities, the light-tissue interactions that take place vary depending on the wavelength of the light source, which alters the optical properties of the tissue components.

2.4. Limitations of pulse oximetry

It is important to acknowledge the limitations and assumptions of pulse oximetry before delving further into the impact and/or possible causes of skin pigmentation on the accuracy of pulse oximetry measurement. This is because inaccurate measurements may result from a combination of certain conditions, than in the presence of either condition alone.

As previously stated, pulse oximeters measure functional oxygen saturation, which as a result, assumes that the only haemoglobin species bounding with oxygen and present in blood are deoxygenated and oxygenated haemoglobin. However, other dyshaemoglobins such as carboxyhaemoglobin (COHb) and methaemoglobin (MetHb), may be present in the blood due to higher concentrations or partial pressures of carbon monoxide and/or other gases and chemical compounds). The most common case is carboxyhaemoglobin, which is a molecule of haemoglobin that has bound reversibly with carbon monoxide (CO). In the visible range of the light spectrum, COHb presents a very similar absorption profile to oxyhaemoglobin, thus making it very difficult to distinguish between the two haemoglobins. This does not cause any particular concern in healthy subjects, but, in particular cases such as carbon monoxide poisoning or smokers, COHb may be present in higher concentrations in blood. This could potentially result in erroneous SpO2 estimations, although they have been seen to minimally effect measurements of SpO2 (Feiner et al 2013).

Furthermore, accurate estimation of blood oxygen saturation may be compromised under conditions of poor blood perfusion. To estimate SpO2 accurately, pulse oximeters rely on the optical measurements of arterial pulsations (i.e. PPG signals) within the tissue bed. However, these pulsations only make up 2%−5% of the total optical absorption profile recorded and are significantly dependent on an adequate blood flow/supply to the measurement area. When blood flow to certain anatomical areas is compromised due to, for instance, hypovolemia, hypotension or hypothermia, the arterial pulsations recorded by the pulse oximeter may be significantly reduced or disappear, causing inaccurate SpO2 measurements. A reasonable solution is to move the sensor from a peripheral site, such as the finger to a more central location with adequate perfusion. However, this may be not ideal in cases of severe critical illness such as patients experiencing a high degree of hypoperfusion. Another cause of falsely low SpO2 is changes in venous volume due to venous pulsations, which can result from a number of scenarios including tightly placed finger probes, or again, in critically ill-patients suffering from heart complications.

Pulse oximeters are well known to provide inaccurate readings when the light absorption profile of red and/or infra-red light is corrupted. Nail polish and artificial fingernails have been reported to affect pulse oximeter readings measured at the fingertips. More specifically, dark colours of nail polish such as black or blue can cause false readings and lead to inaccuracies (Coté et al 1988, Çiçek et al 2011, Yönt et al 2014). As this is a well-known limitation of pulse oximetry, removing nail polish or changing the measurement site can eliminate this particular issue. Contrarily to the popular understanding that nail polish causes false SpO2 readings, some studies have determined a limited or insignificant impact on SpO2 readings (Rodden et al 2007, Yamamoto et al 2008, Balaraman et al 2020). However, the reported differences in the literature may be due to methodological differences and more standardised studies are required to assess the existence of such interference across different pulse oximeter manufactures, nail polish colours, and larger populations.

Another limitation is the effect of light pollution (ambient light) (Mathes et al 2008, Schulte et al 2012, Saito et al 2017) and movement artefacts. Any external light, aside from the discrete wavelengths employed by the instrument, can interfere with measurement if it reaches the photodetector. Usually, ambient light pollution may cause offsets in the sensor or cause additional issues due to the flickering of lamps at 50 Hz (i.e. this frequency may vary across different countries), thus introducing unwanted noise in the PPG signals, which can possibly lead to erroneous SpO2 values. For this reason, pulse oximetry sensors should be optically insulated from any external light, and they should be positioned carefully to avoid ambient light reaching the photodetector. Using filtered photodetectors may alleviate the problem, but any light component in the red to the near-infra-red range of the spectrum (i.e. the effective range used in pulse oximetry) can still enter the sensor if not properly screened. Also, if pulse oximetry sensors are not properly positioned on the measurement tissue, mismatches in the red and infra-red absorption (penumbra effect) may cause false SpO2 readings. The penumbra effect re-emphasises the importance of correct probe positioning. Inappropriate probe placement leads to error readings, especially on the small fingers of neonates and infants. This effect occurs when the probe is not symmetrically placed, such that the pathlength between the two LEDs and the photodetector is unequal, causing one wavelength to be ‘overloaded’. Also, the emitted light can be projected tangentially to the detector (in reflectance probes), sometimes without crossing an arterial bed, leading to optical shunting. Repositioning of the probe often leads to improvement in saturation readings.

Furthermore, one of the ‘nemesis’ of pulse oximetry is sensor/tissue movement artefact and this should be limited during measurements. Good probe design and attachment as per application needs can help mitigate against this. Random extra-arterial movements can be easily recognised, but periodical movements, synchronous with the cardiac cycle, can be erroneously interpreted as ‘signal’ by the instrument and cause inaccurate SpO2 measurements. Advancements in signal processing techniques and technologies have helped to minimise the issue of movement artefacts in pulse oximetry (Chacon et al 2018, Louie et al 2018, Banik et al 2020). Some proprietary algorithms such as the Masimo Signal Extraction (SET®) (Goldman et al 2000) are already incorporated in modern pulse oximeters and are showing a robust estimation of SpO2 during motion (Hoskin and Granger 2001, Shah et al 2012).

Lastly, pulse oximeter readings can be impacted by intravenous and intradermal dyes (methylene blue, indocyanine green, patent blue, nitrobenzene and indigo carmine), which are used to identify anatomical structures (e.g. cancerous tissue, lymph nodes, etc) in diagnostic imaging or surgical operations. Usage of pulse oximeters during or shortly after the induction of these dyes has shown to have resulted in falsely low SpO2 values without actual desaturation. These errors occur as the pulse oximeter cannot distinguish between the absorption of light by intravenous dye and the absorption by haemoglobin species. For example, methylene blue has its peak spectral absorption at 668 nm, hence it absorbs most of the light emitted by the red emitter. This absorption is interpreted by the pulse oximeter as the presence of reduced haemoglobin, leading to inaccurate estimation of SpO2 (Sriganesh et al 2013, Ishiyama et al 2015). Therefore, caution must be taken when performing dye injections with concomitant pulse oximetry measurements.

In conventional practice, the effects of light scattering are accounted for via empirical calibration of the oximeter device, which appears to work well, but only up to a certain point (Webster 1997b). This is an unavoidable limitation of pulse oximeters as they can only be as accurate as their empirical calibration curves. These calibration curves are obtained from healthy subjects by measuring the red: infra-red ratio (5) while varying the inspired fraction of oxygen and, simultaneously, measuring SaO2 by blood-gas analysis and collecting dual-wavelength PPG signals. Understandably, researchers and pulse oximetry manufacturers are limited in the degree of hypoxaemia inducible in these volunteers, to an SaO2 of approximately 70%–75%. Therefore, the shape of the calibration curve below these levels must be extrapolated, with potential implications for the accuracy of pulse oximetry at low saturation levels. One of the limitations of this traditional calibration method is the limited range of oxygen saturation that can be acquired. Ethical issues prevent intentional desaturation of healthy subjects below a certain point due to risk of hypoxic brain damage. Also, another potential limitation of such calibration studies is perhaps the lack of balance between participating volunteers from all categories of skin pigmentation. Notably, the assumptions inherently made during an empirical calibration are valid only for a limited range of saturations and become invalid under extreme conditions.

3. Skin pigmentation as a limitation of pulse oximetry

Pulse oximeters use the optical properties of HbO2 and HHB in the visible and near infra-red regions to estimate blood oxygen saturation. Differences between arterial oxygen saturation measured by blood gas analysis and pulse oximetry was documented soon after their invention in the 1980s (Wang and Poh 1985, Cecil et al 1988, Gabrielczyk and Buist 1988, Ries et al 1989), yet the problem remains neither fully corrected nor fully understood (Sjoding et al 2021). A summary of the studies which compared the accuracy of pulse oximeters in groups of different race or skin pigmentation is provided in table 1. These were identified via a number of databases such as PubMed, Web of Science, and Google Scholar, using the following search terms/phrases: ‘pulse oximeter skin pigmentation’, ‘racial bias in pulse oximetry’, ‘pulse oximetry covid-19’, and ‘silent hypoxaemia’. Sources that were not identified using the search terms were retrieved from the references of identified publications. To be included, studies had to compare SpO2 against reference SaO2 measurements obtained via blood gas analysis on subjects with different skin pigmentation.

Table 1.

Pertinent studies that compared the accuracy of pulse oximeters in groups of different race or skin pigmentation (available to access).

Study No. subjects Setting Method for stratifying skin pigmentation Statistical information Findings
Wang et al (1985) (Wang and Poh 1985) 31 ‘pigmented patients’ Not available Not available Not available Biox III oximeter gave readings that closely approximated oxygen saturation measurements from arterial blood, confirming the usefulness of non-invasive oximetry even in pigmented patients.
Gabrielczyk and Buist (1988) 21 (4 ‘racially pigmented’) Post-cardiac surgery Unclear Bias ± SD (Overall): +0.6 ± 1.6% a Statistically insignificant difference in accuracy of SpO2 between skin pigmentations.
Cecil et al (1988) 152 (136 White, 1 Asian, 15 Black) Not stated Scale of 1 to 3—Light, medium, or dark pigment Bias (Overall): Nellcor, +0.59%. Ohmeda, −0.897% a Both oximeters displayed a statistically significant, but clinically insignificant bias when compared with arterial blood oxyhaemoglobin. Greater accuracy in SpO2 measurements taken from Black patients was observed by the Ohmeda oximeter.
Ries et al (1989) 187 Distribution of racial groups not stated Laboratory Munsell colour system Bias (Overall): Ohmeda, +1.4%. Hewlett Packard, −0.6% a Readings were slightly less accurate in patient groups with darker skin, suggesting that dark skin colour may. affect the performance and accuracy of ear oximeters.
Jubran and Tobin (1990) 54 (29 Black, 25 White) Intensive Care Visual Bias ± SD (White): 2.2 ± 1.8% a SpO2 was less accurate and less precise in Black patients.
Bias ± SD (Black): 3.3 ± 2.7% a
Cahan et al (1990) 28 (22 White, 6 Black) Hypoxia Laboratory Unclear Bias ± SD (White): 1.9 ± 5.1% a Pulse oximetry values can be higher in Black subjects than in White subjects, especially at saturations below 80%.
Bias ± SD (Black): 5.1 ± 4.6% a
Lee et al (1993) 33 (22 Chinese, 6 Malaysian, 5 Indian) Intensive Care Race Bias ± SD (Overall): +0.82 ± 2.6% a SpO2 was most accurate in Chinese patients, and less accurate (greater overestimation) in Malaysian and Indian patients. It was also less accurate at low saturations and in patients with elevated bilirubin levels.
Range: −4.9% − 10.5%
Bothma et al (1996) 100 darkly pigmented adults Intensive Care Reflectance spectrophotometer Bias (Overall): −1.0% − 1.2% a The accuracy of pulse oximetry is not adversely affected by skin pigmentation, and it remains a useful oxygenation-monitoring device in darkly pigmented patients.
LOA: −6.6% − 6.6% across all pulse oximeters
SD: 1.9%–2.4% across all pulse oximeters
Adler et al (1998) 278 (34 Dark, 101 Intermediate, 143 Light) Emergency Department Munsell colour system Bias ± SD (Light): +2.5 ± 4.6% a The accuracy and precision of SpO2 was not affected by skin pigmentation. Signal quality was poorer in a great proportion of patients with dark skin pigmentation than intermediate or light.
Bias ± SD (Intermediate): +2.8 ± 5.2% a
Bias ± SD (Dark): +2.2 ± 3.7% a
Bickler et al (2005) 21 (11 Dark, 10 Light) Hypoxia Laboratory Race Bias ± SD (Light): +0.37 ± 3.20% a SpO2 was less accurate (greater overestimation) in Darkly pigmented subjects. This bias was greater at lower saturations, and differed between pulse oximeters.
Bias ± SD (Dark): +3.56 ± 2.45% a
Feiner et al (2007) 36 (17 Dark, 7 Intermediate, 12 Light) Hypoxia Laboratory Race Bias range for intermediate and dark skin: +4.5%–+4.9% (SaO2 = 60% − 70%) a SpO2 was less accurate (overestimation) in subjects with Dark and Intermediate pigmentation at lower saturations for five out of six combinations of pulse oximeters and probe types (the exception being the Masimo Radical with adhesive probe).
+2.4%–+3.6% (SaO2 = 70% − 80%)
Witting and Scharf (2008) 837 (577 African American, 260 White) Emergency department Self-report Not available African American group without hypoxaemia was associated with a 0.8-unit increase in SpO2 values, and without a change in precision. African American females and White females had an average SpO2 higher than African American men and White men, respectively. Clinicians should regard these findings with particular significance.
Foglia et al (2017) (Foglia et al 2017) 35 (14 Dark, 21 Light) Infants with congenital heart disease Munsell colour system Bias ± SD (Overall): Nellcor, +3.9 ± 2.0% Masimo, +0.8 ± 2.4% a No significant difference in SpO2 accuracy between patients with Dark and Light pigmentation.
Ebmeier et al (2018) 394 Distribution of racial groups not stated Intensive Care Fitzpatrick scale Bias (Light versus Medium): −0.9% a The accuracy of SpO2 was influenced by skin pigmentation and pulse oximeter model (Masimo underestimating, Philips overestimating).
Bias (Light versus Dark): −2.4% a
LOA: −4.4%–4.4%
Murphy and Omar (2018) 146 (6 Light, 111 Medium, 29 Dark) Intensive Care Massey New I mmigrant Survey skin colour scale Bias ± SD (Overall): +1.64 ± 0.15 g dL−1a The degree of skin pigmentation does not appear to influence the magnitude of bias, rather the increasing severity of illness and decreasing lower mean arterial blood pressure.
LOA (Overall): −1.05 g dL−1–4.33 g dL−1
Smith and Hofmeyr (2019) 220 12 Type I, 28 Type II, 69 Type III, 45 Type IV, 28 Type V, and 38 Type VI Perioperative areas e.g. pre-assessment clinics, recovery rooms, operating theatres and intensive care units Fitzpatrick scale Bias (Overall): −0.55% a Darker skin pigmentation showed no trend to an effect on the accuracy of oxygen saturation measured using portable fingertip pulse oximeter and a conventional bedside pulse oximeter.
LOA: −3.25%–2.16%
Sjoding et al (2021) 10,001 (1,326 Black, 8,675 White) Intensive Care, and Inpatients receiving oxygen Race Not available SpO2 was less accurate (overestimation) in Black than White patients. Hidden hypoxemia (arterial oxygen saturation of <88% and SpO2 of 92%–96%) was nearly three times as common in Black than White patients.
Valbuena et al (2021) 372 (65 Asian, 51 Black, 70 Hispanic, 186 White) Intensive Care Race Incidence of occult hypoxaemia: 10.2% (White), 21.5% (Black), 8.6% (Hispanic), and 9.2% (Asian) Hidden hypoxemia (blood gas arterial oxygen saturation of <88% and SpO2 of 92%–96%) was more common in Black patients than Asian, Hispanic or White patients.
Vesoulis et al (2021) 294 (124 Black, 170 White) Neonatal Intensive Care Race Bias (White): +0.72% a SpO2 was less accurate (overestimation) in Black than White patients. Hidden hypoxemia (arterial oxygen saturation of <85% and SpO2 ≥90%) was more common (although not significantly so) in Black than White patients.
Bias (Black): +1.73% a
Wiles et al (2021) 194 (34 Asian, 19 Black, 6 Other, 135 White) Critical Care Race Bias (White): + 0.28% b SpO2 was less accurate (overestimation) in Black, than Asian, than White patients (although no statistical test was used). Correlation between SpO2 and arterial oxygen saturation was lower in Black patients than Asian or White patients.
LOA (White): −1.79%–2.35%
Bias (Asian): −0.33% b
LOA (Asian): −2.47%–1.80%
Bias (Black): −0.75% b
LOA (Black): −3.47–1.97
Okunlola et al (2022) 491 (108 Dark, 383 Light) Hypoxia Laboratory Unclear Not available SpO2 was less accurate (overestimation) in Dark than Light to Medium skin pigmentations.
Wong et al (2021) 87,971 (1,919 Asian, 26,032 Black, 2,397 Hispanic, 57,632 White) Intensive Care and other Hospital Wards Race Proportion of patients with hidden hypoxaemia: The incidence of hidden hypoxemia (arterial oxygen saturation of <88% and SpO2 ≥88%) was greatest in Black, then Hispanic, Asian and finally White patients. It was associated with greater organ dysfunction 24 h later, and higher in-hospital mortality.
White: 4.9%
Asian: 4.9%
Hispanic: 6.0%
Black: 6.9%
Bangash et al (2022) 16,818 (81.2% White, 11.7% Asian, 4.0% Black, 3.2% Other ethnicities) Hospital Ethnicity Relative to White patients: Pulse oximetry tends to overestimate O2 saturation, and this is more pronounced in patients of Black ethnicity. These differences resulted in 6.1% versus 8.7% of White versus Black patients classified as normoxic on SpO2 who were hypoxic on the gold standard SaO2 reading.
Bias (Asian): 0.5pp greater a
Bias (Black): 0.8pp greater a
Bias (Other): 0.3pp greater a
Shi et al (2022) 6,505 (4,897 adults and 1,608 children) 27 out of 32 in hospital, none at home 15 studies measured skin pigmentation and 22 referred only to ethnicity Bias ± SD (Light): −0.35 ± 1.49% a Pulse oximetry may overestimate oxygen saturation in people with dark skin and people whose ethnicity is reported as Black/African American, compared with SaO2, although the overestimation may be quite small in hospital settings. The clinical importance of any overestimation will depend on the particular clinical circumstance.
Distribution of racial groups not stated
LOA (Light): −1.87–4.09
Bias ± SD (Intermediate): −0.58 ± 1.47% a
LOA (Intermediate): −3.46–2.30
Bias ± SD (Dark): +1.11 ± 1.52% a
LOA (Dark): −3.27–2.58
Baker and Wilson (2022) 75 (39 Black, 36 White) Hypoxia Laboratory Unclear Bias ± SD (White): −0.05 ± 1.35 a There was no clinically significant difference in the accuracy or bias between Black and White subjects monitored with Masimo SET pulse oximetry.
Arms (White): 1.35
Bias ± SD (Black): −0.2 ± 1.40 a
Arms (Black): 1.42
Crooks et al (2022) 2997 Distribution of ethnic groups not stated Hospital Ethnicity Bias (White): +3.2% a Pulse oximetry overestimated arterial oxygen saturations compared to blood gas measurement across all ethnicity groups when SpO2 measurements were below 90%, and underestimated these when SpO2 measurements were above 95%. However, individuals with Black, Asian or mixed ethnicity had a higher reading for oxygen saturation as measured by pulse oximetry than blood gas compared to individuals with a White ethnicity.
Bias (Asian): +5.1% a
Bias (Black): +5.4% a
Bias (Mixed): +6.9% a
Wiles et al (2022) 178 (126 White, 30 South Asian, 13 Black, and 9 other ethnic origin) Intensive Care Ethnicity Bias (White): −0.25% b Bias was greater in patients of non-White ethnic origin. The study also found that pulse oximetry is less accurate in patients diagnosed with COVID-19 and receiving mechanical ventilation.
LOA (White): −4.75%–4.23%
Bias (South Asian): −0.96% b
LOA (South Asian):
−5.62%–3.71%
Bias (Black): −1.72% b
LOA (Black): −6.8%–3.36%
Bias (Other): −1.21% b
LOA (Other): −5.48%–3.05%
Henry et al (2022) 26,603 (24,493 White, 1,263 Black, 574 Asian, 273 American Indian) Intensive Care Self-identification Incidence of occult hypoxaemia: White (3.6%), Black (6.2%), Asian and American Indian (6.6%) Occult hypoxemia is more common in Black patients compared with White patients and is associated with increased mortality, suggesting potentially important outcome implications for undetected hypoxemia. It is imperative to validate pulse oximetry with expanded racial inclusion.
Burnett et al (2022) 151,070 (16,011 Black, 21,223 Hispanic, 70,722 White, 8,121 Asian, 34,993 other) Unclear Self-identification Bias ± SD (White): −0.20 ± 6.3% a Self-reported Black and Hispanic race/ethnicity are associated with a greater prevalence of intraoperative occult hypoxemia in the SpO2 range of 92% to 100% when compared with self-reported White race/ethnicity.
Bias ± SD (Hispanic): +0.5 ± 7.9% a
Bias ± SD (Asian): +0.2 ± 6.5% a
Bias ± SD (Other): +0.1 ± 5.9% a
Bias ± SD (Black): + 0.6 ± 9.1% a
Fawzy et al (2022) 1216 (63 Asian, 478 Black, 215 Hispanic, and 460 White) Referral centres and community hospitals Self-identification Bias (Relative to White Patients) b Pulse oximetry overestimated arterial oxygen saturation among Asian, Black, and Hispanic patients (ethnic minority groups) compared with White patients with COVID-19. This contributes to unrecognised or delayed recognition of eligibility to receive COVID-19 therapies.
Asian: −1.73%
Hispanic: −1.13%
Black: −1.23%

LOA: Limits of agreement, SD: Standard deviation, pp: Percentage points

a

Bias= SpO2 - SaO2

b

Bias= SpO2 - SaO2.

3.1. Overestimation of arterial oxygen saturation

The mean difference between blood oxygen saturation measurements obtained by the two approaches is the ‘bias’, hence, ‘mean bias’ was referred to the average bias recorded within a racial subgroup. Out of the 28 studies identified in table 1, 22 studies found that SpO2 was overestimated in those with darker skin relative to reference SaO2 measurements obtained by blood gas analysis (Ries et al 1989, Cahan et al 1990, Jubran and Tobin 1990, Lee et al 1993, Adler et al 1998, Bickler et al 2005, Feiner et al 2007, Witting and Scharf 2008, Foglia et al 2017, Ebmeier et al 2018, Hunasikatti 2021, Sjoding et al 2021, Valbuena et al 2021, Vesoulis et al 2021, Wiles et al 2021, Wong et al 2021, Baker and Wilson 2022, Bangash et al 2022, Crooks et al 2022, Okunlola et al 2022, Shi et al 2022, Wiles et al 2022). Additionally, over half (57%) of the studies documented an increase in bias for subjects from all racial sub groups as they became less saturated (Cahan et al 1990, Jubran and Tobin 1990, Volgyesi and Spahr-Schopfer 1991, Lee et al 1993, Bothma et al 1996, Bickler et al 2005, Feiner et al 2007, Foglia et al 2017, Ebmeier et al 2018, Smith and Hofmeyr 2019, Vesoulis et al 2021, Wiles et al 2021, Wong et al 2021, Bangash et al 2022, Crooks et al 2022, Henry et al 2022, Shi et al 2022). For White subjects, bias values were observed between −0.35% and 3.2% and between 0.6% and 5.1% for Black subjects. Precision, which is given by the standard deviation (SD), was found to range between 1.8% and 6.3% for White subjects and between 2.7% and 9.1% for Black subjects. Evidently, bias and precision are seen to have a wider range in Black subjects relative to White subjects, which indicates that some pulse oximetry measurements are more accurate in individuals with fair skin.

In infants, a 1.5-fold overestimation of SpO2 was observed in Black infants relative to White infants for a SaO2 range between 85% and 100% (Vesoulis et al 2021). However, some biases in infants may be caused by cyanosis and not skin pigmentation (Mahle et al 2009, Ross et al 2014) although cyanosis is often observed differently in people with different skin tones (blue/purple in fair skin, grey/green in intermediate skin, and grey/white in dark skin). A mixture of fetal haemoglobin (HbF) and adult haemoglobin (HbA) could result in an underestimation of SpO2 of 3%–4%, or rotation of sensors every 12 h (to avoid skin injury) can cause inaccurate readings, as well as increase in melanin production following treatment in Black and Asian patients after phototherapy for jaundice patients (Vesoulis et al 2021). Such considerations are as important for rectifying bias caused in White subjects, specifically hypoxic patients. Hence, more work must be done on investigating bias at low arterial oxygen saturations (<90%), as this may provide some explanation about the overestimation of SpO2, and the occurrence of suboptimal function, which was observed two times more frequently in Black patients relative to White patients (Adler et al 1998).

Moreover, pulse oximeters were found to have a root-mean squared error (Arms) of more than 3% in infants overall, with a greater discrepancy in Black infants (9.5% for Black infants and 8.9% for White infants) (Vesoulis et al 2021). Other studies recorded root-mean square values less than 3% (1.35% for White patients and 1.42% for Black patients using the Masimo SET) (Baker and Wilson 2022), (1.08% for White patients, 1.13% for Asian patients, and 1.56% for Black patients) (Wiles et al 2021). There is a greater urgency to address the huge inaccuracies observed in infants relative to adults, and to perform thorough testing of consumer marketed pulse oximeters, including new generation models (Rosychuk et al 2012), in order to prevent misleading interpretation of pulse oximeters readings, especially by home users (Okunlola et al 2022). The OpenOximetry.org Project is actively working to better understand the impact of skin pigmentation on oximeter accuracy and to develop new strategies to eliminate this source of error (Open Critical Care, n.d.). This includes laboratory testing on human subjects to determine the performance of existing oximeters used in a laboratory setting, and whether they perform differently in some patients in a clinical setting.

3.2. Incidence of hypoxaemia

As expected, the overestimation of SpO2 in Black subjects leads to an increased incidence of occult hypoxaemia (SaO2 < 88% yet SpO2 = 92%–96%) (Jubran and Tobin 1990, Michael et al 2020, Todd 2021, Valbuena et al 2021, Vesoulis et al 2021, Wong et al 2021, Burnett et al 2022, Henry et al 2022, Wiles et al 2022) ((Baker et al 2021) is a notable exception). Occult hypoxaemia was found to occur three times more frequently in Black subjects than in White subjects (Michael et al 2020, Valbuena et al 2021), two times more frequently in Black subjects than in White subjects (Burnett et al 2022), and in 9.2% of Black infants and 7.7% of White infants (Vesoulis et al 2021). Infants spend much of their hospitalisation period with oxygen saturations above 85%, so any small change in bias is critical, and can have adverse effects on mortality post-surgery (Nafiu et al 2020). Relying on pulse oximeters to triage patients and adjust supplemental O2 may place Black patients with increased hypoxaemia at risk (Todd 2021).

To correct for these levels of hypoxaemia amongst different racial groups, SpO2 targets of 92% and 95% were found to be reliable in predicting a satisfactory level of oxygenation in White and Black patients respectively (Jubran and Tobin 1990). These thresholds were determined for patients who recorded a PaO2 of 60 mmHg or above, however, it was more difficult to correct hypoxaemia while avoiding O2 toxicity for Black patients who recorded an average PaO2 level of 83 ± 31 mmHg (Jubran and Tobin 1990, Ralston et al 1991). In another study, these results were also consistent after the exclusion of patients with high COHb (seen to overestimate SpO2 (Adler et al 1998)), diabetes, and adjusting for age and sex and cardiovascular score, which indicates that the inaccuracies appear to be related to skin pigmentation. It is important to note that not all Black patients had occult hypoxaemia, but it is clear that there is a variation in risk due to race (Michael et al 2020), extending to greater organ dysfunction and higher in hospital mortality, even when age, sex, and sequential organ failure assessment score were adjusted (Wong et al 2021). Similarly in a recent retrospective study, Black patients diagnosed with Covid-19 were always at risk of unrecognised oxygen treatment eligibility based on pulse oximetry measurement (Fawzy et al 2022).

3.3. Methodologies

Conventionally, skin pigmentation is described by self-report or ethnic classification, without consideration of environmental factors affecting the quantity and distribution of melanin in the epidermis, such as exposure to ultraviolet B radiation. We identified some studies whose methods and/or outcomes did not comply with the United States Food and Drug Administration (FDA). According to regulations, 15% of the participant pool should be darkly pigmented. This requirement was met in (Wong et al 2021, Baker and Wilson 2022) (White pool = 65.5%, Black pool 29.6%, White pool = 55.44%, Black pool = 44.56%), but not in other studies (Guber et al 2019, Michael et al 2020, Valbuena et al 2021, Wiles et al 2021, Bangash et al 2022, Wiles et al 2022). Despite this, the inaccuracies seen in Black or darkly skinned subjects mentioned above did not change, although it was difficult to make conclusive statements in studies with a small number of volunteers (Gabrielczyk and Buist 1988).

The magnitude of the biases may have occurred due to a number of reasons. Firstly, some studies did not obtain simultaneous SpO2–SaO2 pairs as conducted in (Ebmeier et al 2018, Smith and Hofmeyr 2019, Baker et al 2021), since SaO2 measurements are prone to fluctuations over short periods of time. Instead, some measurements were taken a minimum of 4 min apart (Wiles et al 2021), 7.5 min apart (Bangash et al 2022), 10 min apart (Michael et al 2020), and up to 30 min apart (Crooks et al 2022). It is also important to consider that breathing variations, body temperature, anaemia, excessive motion, high levels of COHb and MetHb (Adler et al 1998), incorrectly applied probes/probe type (Feiner et al 2007), blood gas contamination by air in syringe (Vesoulis et al 2021), and perfusion status (Okunlola et al 2022), may have resulted in a small bias difference, especially in critically-ill patients (Wiles et al 2021).

Furthermore, only two studies (Ebmeier et al 2018, Guber et al 2019) used the Fitzpatrick scale (as recommended by the FDA) to categorise skin tones. Others chose to use the Munsell chart (Adler et al 1998, Feiner et al 2007, Foglia et al 2017) or mostly report race by self-identification (Witting and Scharf 2008, Vesoulis et al 2021, Bangash et al 2022, Wiles et al 2022). Only one study quantified pigment using reflectance spectrophotometry (Bothma et al 1996), which is the most objective method for classifying different skin colours.

3.4. Statistical significance

The majority of the data and information presented in this section share similar trends, however, not all these studies showed statistically significant differences. From table 1, 18 out of 30 studies (60%) concluded that their results showed a statistically significant relationship between skin pigmentation and overestimated SpO2 in subjects with darker skin (Cecil et al 1988, Gabrielczyk and Buist 1988, Ries et al 1989, Jubran and Tobin 1990, Lee et al 1993, Adler et al 1998, Bickler et al 2005, Feiner et al 2007, Sjoding et al 2021, Valbuena et al 2021, Vesoulis et al 2021, Wong et al 2021, Bangash et al 2022, Burnett et al 2022, Crooks et al 2022, Fawzy et al 2022, Okunlola et al 2022, Shi et al 2022, ). Some studies were either not able to conclude statistical significance due to limiting factors (e.g. sample size), or simply did not mention the statistical significance of the data (Wang and Poh 1985, Cahan et al 1990, Witting and Scharf 2008, Smith and Hofmeyr 2019, Wiles et al 2022). Additionally, there appears to be no clear guidelines for determining the level of clinically significant inaccuracy (or difference in accuracies). For instance, the differences in accuracy between groups were deemed to be clinically significant in Wiles et al (2021) and Volgyesi and Spahr-Schopfer (1991), and clinically insignificant in Bothma et al (1996), Adler et al (1998), and Baker and Wilson (2022), despite both sharing overlapping limits of agreements. Whilst it is beyond the scope of this narrative review to synthesise the results of different studies, we believe that the evidence indicates that at least some pulse oximeters are less accurate in people with darker skin pigmentation. Several recent large-scale studies have found lower accuracy and/or greater levels of occult hypoxaemia in people with darker skin pigmentation (Wong et al 2021, Bangash et al 2022, Burnett et al 2022, Crooks et al 2022, Fawzy et al 2022, Henry et al 2022, Shi et al 2022). We note that accuracy may differ between different pulse oximeter models, and future meta-analyses could consider this.

4. Natural and environmental factors of skin in relation to photoplethysmography (PPG)

4.1. Determinants of skin colour and appearance

Skin is a highly complex organ, and its colour and appearance are influenced by several factors. These include the presence of melanin, keratin, carotene, and haemoglobin, and other characteristics that differ with race such as hydration, texture, and homogeneity. The human skin is made up of multiple layers, each with their own distinctive optical properties that govern the absorption and scattering mechanisms of light.

Melanin is the primary determinant of skin colour in people with darker skin (Naik and Farrukh 2022). Carotene, keratin, oxygenated haemoglobin and water are other chromophores which lead to the differing absorption and reflection of light and hence varying shades of skin. The pink hue found in skin types I–II (from the Fitzpatrick scale) is produced by the combination of haemoglobin and oxyhaemoglobin, and the yellow-orange hue in skin types V–VI results from the combination of melanin and carotenes (Rawlings 2006). Natural melanin levels are determined by genetics, but can be influenced by quality of life, hormones, extrinsic and intrinsic aging, and skin pigment disorders, all of which alter the absorption coefficient and scattering coefficient of skin over time.

Melanin is produced in melanocyte cells, which are typically localised in the basal layer of the epidermis. The number of melanocytes does not account for the differences in skin colour, but rather, it is the variation in size, quantity, location, and distribution of melanosomes within melanocytes (where melanin pigment is synthesised) that contributes to the formation of skin colour (Szabó et al 1969, Toda et al 1972, Johnson et al 1998). In fair skin, melanosomes are smaller, found in clusters, degrade more quickly relative to more pigmented skin (Jothishankar and Stein 2019), and are synthesised in very acidic conditions (Ancans et al 2001, Fuller et al 2001, Watabe et al 2004). Contrastingly to darkly pigmented skin, melanosomes are larger, more dispersed along the basal layer, and synthesise at a higher pH of 6.8 (Jothishankar and Stein 2019).

Depending on the patient, it is important to select an appropriate measurement site to obtain reliable bio-optical data (Mantri and Jokerst 2022). This may differ from one patient to another, especially for those with dark skin pigmentation, as they are more prone to post-inflammatory hypopigmentation or hyperpigmentation relative to individuals with fair skin (Fitzpatrick et al 2018). Additionally, the colour of the nail bed in a Black person is much lighter than their skin, and so the behaviour of light through a finger (transmittance mode pulse oximetry) can differ relative to the behaviour of light through a palm (reflectance mode pulse oximetry).

4.2. The influence of skin pigmentation on the optical properties of skin

Dermatologists often use the Fitzpatrick scale, which describes skin colour based on a response to Sun exposure (Fitzpatrick et al 2018) which has been used to stratify skin pigmentation in experimental studies (Cahan et al 1990, Murphy and Omar 2018). The volume fraction of melanosomes in the epidermis layer of human skin ranges between 1% and 43% for very fair to very dark pigmentation, which in turn, represents a large variation in the average epidermal absorption coefficient that is also wavelength-dependent. Based on these skin types and their corresponding melanosome volume fractions as reported by Kanellis (2019) for fair, medium, and dark skin, the average absorption coefficient of the epidermis may be calculated using the following equation (Jacques 1998):

μa_epidermiscm1=fmelanin6.6×1011λ3.33+1fmelanin0.244+85.3eλ15466.2,

where f melanin is the volume fraction of the epidermis occupied by melanosomes, and λ is the wavelength of light in nanometres (nm).

This yields to absorption coefficients of 0.72 mm−1, 4.20 mm−1, and 8.24 mm−1 for fair (f melanin = 2.55%), medium (f melanin = 15.5%), and dark skin (f melanin = 30.5%) respectively using red light, and suggests that darkly skinned subjects may absorb nearly two times more red light than medium skin and over eleven times more than fair skin. Contrastingly, the absorptions of infra-red light by different skin types are more similar in magnitude, with absorption coefficients of 0.24 mm−1, 1.31 mm−1, and 2.59 mm−1 for fair, medium, and dark skin respectively. It is important to note that these values would vary between subjects in the same skin pigmentation group, and that also, these values be may not be entirely representative as the equations may be oversimplified. However, from a quantitative standpoint, these values may be indicative of the magnitude of change in the optical properties of the epidermis as melanin concentration vary.

Furthermore, the amount of light reflected from within or on the tissue surface contributes to the magnitude of light intensity output. Diffuse reflection is a type of reflection that arises from subsurface scattering of light in rough surfaces such as human tissue. The amount of diffuse reflection that takes place is influenced by the optical and geometric properties of human skin i.e., refractive index and angle of incidence (So-Ling and Li 2001). Fair skin types possess a low absorption coefficient and hence, more light can escape via diffuse reflectance. However, dark skin types have a high absorption coefficient, reducing the amount of diffuse reflectance, and tend to have a low scattering coefficient due to reduced blood supply to the skin. As a result, surface reflection relies on the combined effects of absorption and scattering based on skin pigmentation, hydration, sebum production, collagen network, and skin homogeneity.

Overall, it is important to distinguish between the characteristics of skin and their optical properties in order to understand the underpinning cause of overestimation of oxygen saturation in darkly pigmented subjects. It is possible that the distribution of melanin, colour, or racial differences, together or individually, are resulting in the inaccuracies observed in pulse oximetry measurement. The effect of each of these contributors on the optical properties of skin must be studied.

5. Clinical implications

We have established the effects of skin pigmentation on the accuracy of pulse oximetry measurement, and so it is vital to understand the impact of such outcomes to improve patient and consumer care as well as to inform clinicians of cautionary measures.

Pulse oximeters are widely used to identify and monitor signs of disease, and to help clinicians make informed diagnostic decisions. With the remote assistance of healthcare professionals, pulse oximeters have undoubtedly reduced the unnecessary admission of patients suffering from acute Covid-19 by allowing users to track their oxygen saturation levels at home. However, with the plethora of evidence suggesting their greater level of inaccuracy in patients with darker skin tones, we question the efficacy of this device in the recent global pandemic and in other clinical scenarios. Other clinical scenarios include the assessment and/ or diagnosis of sleep apnoea, newborn screening, hypoxaemia, and the administration of supplemental oxygen therapy.

Firstly, if pulse oximetry used in polysomnography is inaccurate in Black or Asian patients, related studies may need to be revisited (Philip et al 2021). For instance, several studies have documented Black males younger than 39 years and between 50 and 59 years to have a higher apnea-hypopnea index compared to White men of the same ages (where pulse oximetry measurements are used to assess the apnea-hypopnea index). Being a Black male younger than 40 years of age increased the apnea-hypopnea index by 3.21 breathing pauses per hour of sleep compared to a White man in the same age range with the same body mass index.

Secondly, pulse oximetry is often used to screen newborn babies to identify low blood oxygen saturations associated with critical congenital heart defects (Brown et al 2020). A baby can experience change in skin pigmentation from birth up until 12 weeks of age due to jaundice as a natural physiological response, or from breast milk, breast feeding failure, haemolysis, or inadequate liver function. Therefore, it may be difficult to obtain accurate measurements of SpO2, or even monitor changes in SpO2 if jaundice levels, and consequently skin colour, are inconsistent. It may be that the build-up of bilirubin in the blood has a greater impact on bias than melanin in infants, since the effect on bias is smaller than in adults (Vesoulis et al 2021).

Thirdly, fair skin types may inflict marginal influence on the pulse oximeter, with more significant errors in dark pigmented skin. As a result, its impact on clinical decision making could be significant at threshold values for diagnosis of hypoxaemia, with overestimated SpO2 leading to clinically important hypoxaemia remaining undetected and untreated. In addition, underestimated SpO2 readings have the potential to be harmful too, resulting in unnecessary treatment with oxygen (and the risk of hyperoxaemia) and wider impacts such as delayed hospital discharge (Shi et al 2022). In the context of profound hypoxaemia, clinicians should not rely on oximetry alone or on isolated readings (Knight et al 2022).

In clinical situations where oxygen is given despite normal SpO2, including carbon monoxide poisoning and states of increased cellular demand such as shock, sepsis, or major trauma, the need for supplemental oxygen is based on clinical judgement and severity of underlying condition (Cajanding 2021). Oxygen treatment is not without its risks, as we have already seen the higher prevalence of O2 toxicity in Black patients. Complications including nasal irritation, dry nose and throat, hyperoxia-related vasoconstriction, bacterial contamination of delivery systems, and lung injury, are common and can be potentially harmful (Cajanding 2021).

6. Recommendations

To address these issues in the short-term, inaccurate pulse oximeter models that are used in clinical settings or are publicly available for at home use must be identified, especially if their accuracy varies with skin pigmentation. Interestingly, some pulse oximeter models are cheaper yet perform similarly to more expensive and internationally standardised models when measuring hypoxaemia in healthy subjects (Lipnick et al 2016). Potentially, another temporary solution is to assign different SpO2 thresholds for the identification of hypoxaemia according to the patient’s skin pigmentation (Wong et al 2021) when using pulse oximeters that comply with FDA standards (i.e. Arms< 3%), provided that skin pigmentation is objectively quantified. Lastly, it may not be ideal to use pulse oximeters to monitor changes in SpO2 for oxygen administration based on saturation thresholds (Michard et al 2021), as there is research to indicate otherwise (Perkins et al 2003). Meanwhile, more studies should be conducted in real life settings to investigate the utility of pulse oximetry under a variety of conditions and circumstances.

It appears that the presence of melanin in the human epidermis may influence SpO2 estimation independently of blood oxygen saturation (Okunlola et al 2022). Melanin has the highest ability to absorb light in the visible and infra-red regions (characterised by μa) relative to all other skin chromophores and layers. Hence, it is a key contributor to the intensity of transmitted light in both transmittance and reflectance pulse oximetry (4). The majority of studies presented in this review have investigated the accuracy of transmittance finger pulse oximetry in medical settings. However, it would also be useful to explore the accuracy of oxygen saturation in subjects using smartphone applications, which rely on measurement of reflected light (Knight et al 2022). There has been some evidence to indicate that transmittance-based pulse oximeters possess a higher accuracy than reflectance-based pulse oximeters (Bangash et al 2022). Therefore, their use in medical settings may not be appropriate until algorithms to adjust for the reduced signal to noise ratio and other characteristics of phone cameras have been addressed (Luks and Swenson 2020).

A computational analysis of the effect of skin pigmentation on the accuracy of pulse oximetry in both PPG modes would be useful in determining the behaviour of light in a certain ROI, and the comparison of different outcomes (Huong and Ngu 2014). Across the years, Monte Carlo modelling has been a widely used tool for representing the stochastic nature of biological tissue, as it is a mathematical technique which involves modelling the probability of different outcomes with random variables. Although previous MC models (Wiles et al 2022) have briefly studied melanin between concentrations of 5% and 20% in their tissue model, there seemed to be no effect on calibration at a SpO2 of 40% (fetal pulse oximetry). Such models can be adapted and extended to look at the full range of melanin concentrations and observe the bio-optical changes for a broader range of arterial oxygen saturation levels. It is important to note here that the current equation (equation (10)) used to quantify the absorption coefficients of the epidermis, melanin, etc, can be oversimplified, and not sufficiently replicated in in-silico models. An alternative approach for modelling melanin at a more structural and molecular level can be achieved using the time-dependent density functional theory (Meng and Kaxiras 2008). Absorbance spectra can be simulated for all the constituent eumelanin monomers, which would be useful in understanding the influences of changes in melanin, particularly when interpreting results from in vitro experiments. However, this is a very timely procedure and therefore impractical from an engineering perspective and limited in the ability to optimise the sensor geometry, if required. Therefore, utilising a combination of modelling techniques must be considered when simulating melanin for a more accurate representation of skin pigmentation.

Furthermore, it may be useful to re-consider the wavelengths at which conventional pulse oximeters operate. There have been discussions on the influence of skin pigmentation on other applications such as PPG-based HR monitoring. They concluded that green light resulted in increased errors in individuals with darker skin tones due to the increased absorption of light by melanin at shorter wavelengths (Fine et al 2021), and that there were difficulties in choosing a wavelength which was insensitive to movement artefacts, poor skin perfusion and darker skin tones (Lemay et al 2021). As described earlier in figure 4, the estimation of SpO2 involves calculating the ratio of the normalised amplitudes of two pulse waves measured using red and infra-red light. Each normalised amplitude is calculated by dividing the pulse wave amplitude by the amount of light transmitted through the tissue. The higher concentration of melanin in patients with darker skin might affect the amount of light transmitted at both wavelengths, particularly red. Due to high absorption by melanin, less red light has been seen to travel through the finger, requiring a higher AC gain for better visualisation of the signal (Cahan et al 1990).

Recently, Hayet al (2018) found that use of a novel two infra-red pulse oximeter does not require empirical calibration. They expect that the accuracy of SpO2 measurement is to be greater than that of the red and infra-red pulse oximeter because of the small inter-person variability of the optical path lengths difference between the two wavelengths, resulting from the lesser impact of melanin in the near infra-red region. Another possible engineering solution is the use of multi-wavelength pulse oximeters, which consider the optical properties of multiple constituents in the region of interest. There is currently a prototype pulse oximeter which uses four short wavelengths (orange, red, blue, and green), and has shown to measure carboxyhaemoglobin saturation (SpCO) and methaemoglobin saturation (SpMet) more accurately than the Masimo Radical 7 (Chong et al 2017). However, further work is needed to test the prototype on a larger number of volunteers, and to ensure that they are racially diverse.

Race adjustments could potentially be built into future pulse oximeters, however, a manual adjustment based on the patient’s skin type may be the most feasible approach in the immediate short term until a concrete solution is found (Philip et al 2021). There is a need to collate PPG data from healthy dark-skinned subjects for saturations below and above 70% (Jubran and Tobin 1990, Calvin 1992, Lee et al 1993, Todd 2021, Holder and Wong 2022, Tobin and Jubran 2022) to observe the changes between PPG AC and DC with skin pigmentation. The correlation between SpO2 and R, must also be better established, as it may be unreasonable to assume a single relationship that relates R to SpO2 for all skin pigmentations. Perhaps pulse oximeters could use different relationships for different skin pigmentations, requiring either automatic recognition of skin pigmentation, or selection by the operator (Bickler et al 2005), although the latter could potentially lead to a considerable source of human error. Furthermore, calibration algorithms and calibration data employed in commercial pulse oximeters should be made available in order to improve awareness within the public and medical community of the possible bias introduced by calibration algorithm and data (Ferrari et al 2022).

Lastly, a key observation from the literature is the need to objectively quantify skin pigmentation. This can be achieved using RGB colour imaging (which indicates the level of red, green, and blue) (Verdugo-Naranjo et al 2022) and reflectance colourimetry, including the Mexameter® (MX 18), Colorimeter® (CL 400), and SkinColorCatch® (previously DermaCatch). The latter measure a combination of melanin and haemoglobin index, L*a*b (L* = lightness, a* = red/green coordinate, and b* = yellow/blue coordinate), Individual Typology Angle (ITA), or L*C*h (L* = lightness, C* = chroma, h = hue angle) (Ly et al 2020). Interpreting these quantitative results is done by correlating the Fitzpatrick, ITA, and melanin index (Baquié and Kasraee 2014, Wilkes et al 2015, Agache et al 2017, Visscher 2017). Although there are a variety of methods to objectively measure skin pigment, Dermacatch® has shown a significantly higher specificity and reproducibility than Mexameter® in the measurement of skin pigmentation and erythema (Agache et al 2017). Overall, the connotations associated with the ‘racial bias’ in pulse oximetry can be re-defined, by establishing that this diagnostic inaccuracy may be due to differences in skin colour, texture, hydration, etc, rather than solely the quantity and quality of melanin in skin.

7. Conclusions

We have analysed the literature on the effect of skin pigmentation on pulse oximeter accuracy, and which reports that SpO2 is frequently overestimated in Black adults and infants, and in subjects with darker skin. As a result, these patients are more likely to experience occult hypoxaemia than White subjects, which may lead to delayed medical attention. This phenomenon has been revisited over 30 years after the introduction of pulse oximeters, since the reliance on such devices has increased during the Covid-19 pandemic, particularly for identifying hypoxaemia in the home. Although there are many questions regarding the accuracy of pulse oximeters possibly due to influence of skin pigmentation, they are still utilised in a variety of clinical applications and settings. Therefore, there is an urgent need to address this issue. We propose potential areas to investigate in the near future, such as the immediate identification of inaccurate pulse oximeters, the investigation of multi-wavelength pulse oximeters in subjects with different skin tones, obtaining more data from darkly skinned subjects to implement in-built calibration options, to objectively quantify skin pigmentation, and the development of computational models to predict differing bio-optical outcomes. Future work must include the development of pulse oximeter design and technology to eliminate bias associated with skin pigmentation, as well as all other known limitations.

Acknowledgments

PHC acknowledges funding from the British Heart Foundation under Grant FS/20/20/34626.

Data availability statement

All data that support the findings of this study are included within the article (and any supplementary information files).

References

  1. Abay T Y, Kyriacou P A. Photoplethysmography in oxygenation and blood volume measurements. In: Kyriacou P, Allen J, editors. Photoplethysmography: Technology, Signal Analysis, and Applications. 1st edn Amsterdam: Elsevier; 2021. [Google Scholar]
  2. Adler J N, Hughes L A, Vivilecchia R, Camargo C A. Effect of skin pigmentation on pulse oximetry accuracy in the emergency department. Acad. Emerg. Med. 1998;5:965–70. doi: 10.1111/j.1553-2712.1998.tb02772.x. [DOI] [PubMed] [Google Scholar]
  3. Agache F, Fanian H, Maibach I, Humbert P. Agache’s Measuring the Skin. 2nd edn. Cham, Switzerland: Springer; 2017. pp. pp 549–62. [Google Scholar]
  4. Ancans J, et al. Melanosomal pH controls rate of melanogenesis, eumelanin/phaeomelanin ratio and melanosome maturation in melanocytes and melanoma cells. Exp. Cell. Res. 2001;268:26–35. doi: 10.1006/excr.2001.5251. [DOI] [PubMed] [Google Scholar]
  5. Anliker U, et al. AMON: a wearable multiparameter medical monitoring and alert system. IEEE Trans. Inf. Technol. Biomed. 2004;8:415–27. doi: 10.1109/TITB.2004.837888. [DOI] [PubMed] [Google Scholar]
  6. Aoyagi T, Kishi M, Yamaguchi K, Watanabe S. Improvement of earpiece oximeter. Abstracts of the 13th annual meeting of the Japan Society of Medical Electronics and Biomedical Engineering ; Osaka. 1974. pp. pp 90–1. [Google Scholar]
  7. Badgujar K C, Badgujar A B, Dhangar D V, Badgujar V C. Importance and use of pulse oximeter in COVID-19 pandemic: general factors affecting the sensitivity of pulse oximeter. Indian Chem. Eng. 2020;62:374–84. doi: 10.1080/00194506.2020.1845988. [DOI] [Google Scholar]
  8. Baker K, et al. Performance of five pulse oximeters to detect hypoxaemia as an indicator of severe illness in children under five by frontline health workers in low resource settings - a prospective, multicentre, single-blinded, trial in Cambodia, Ethiopia, South Sudan, and Uganda. E. Clin. Med. 2021;38 doi: 10.1016/j.eclinm.2021.101040. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Baker S J, Wilson C W. Racial effects on Masimo pulse oximetry: a laboratory study. J Clin Monit Comput. 2022;37:567–74. doi: 10.1007/s10877-022-00927-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Balaraman K, Rausch D, Bellomy R, Chong K, Kameoka A, Simmons Z, Jenkins C, Sumibcay T, Liu E, Wong K. 1167: effect of nail polish on oxygen saturation readings. Crit. Care Med. 2020;48:561. doi: 10.1097/01.ccm.0000643592.11619.5c. [DOI] [Google Scholar]
  11. Bangash M N, et al. Determining the impact of ethnicity on the accuracy of measurements of oxygen saturations: A retrospective cohort study. eClinicalMedicine. 2022;48 doi: 10.1101/2021.11.21.21266662. [DOI] [PMC free article] [PubMed] [Google Scholar]
  12. Banik P P, Hossain S, Kwon T-H, Kim H, Kim K-D. Development of a wearable reflection-type pulse oximeter system to acquire clean PPG signals and measure pulse rate and SpO2 with and without finger motion. Electronics. 2020;9:1905. doi: 10.3390/electronics9111905. [DOI] [Google Scholar]
  13. Baquié M, Kasraee B. Discrimination between cutaneous pigmentation and erythema: comparison of the skin colorimeters dermacatch and mexameter. Skin Res. Technol. 2014;20:218–27. doi: 10.1111/srt.12109. [DOI] [PubMed] [Google Scholar]
  14. Barker S J, Curry J, Redford D, Morgan S. Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study. Anesthesiology. 2006;105:892–7. doi: 10.1097/00000542-200611000-00008. [DOI] [PubMed] [Google Scholar]
  15. Bashkatov A N, Genina E A, et al. Optical properties of melanin in the skin and skinlike phantoms. Proceedings of the SPIE; 2000. [DOI] [Google Scholar]
  16. Bickler P E, Feiner J R, Severinghaus J W. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005;102:715–9. doi: 10.1097/00000542-200504000-00004. [DOI] [PubMed] [Google Scholar]
  17. Bothma P A, Joynt G M, Lipman J, Hon H, Mathala B, Scribante J, Kromberg J. Accuracy of pulse oximetry in pigmented patients. S. Afr. Med. J. 1996;86:594–6. [PubMed] [Google Scholar]
  18. Brouqui P, Amrane S, Million M, Cortaredona S, Parola P, Lagier J C, Raoult D. Asymptomatic hypoxia in COVID-19 is associated with poor outcome. Int. J. Infect. Dis. 2021;102:233–8. doi: 10.1016/j.ijid.2020.10.067. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Brown S, Liyanage S, Mikrou P, Singh A, Ewer A K. Lancet. vol 396. Amsterdam: Elsevier; 2020. Newborn pulse oximetry screening in the UK: a 2020 survey; p. p 881. [DOI] [PubMed] [Google Scholar]
  20. Burnett G W, Stannard B, Wax D B, Lin H M, Pyram-Vincent C, DeMaria S, Levin M A. Self-reported race/ethnicity and intraoperative occult hypoxemia: a retrospective cohort study. Anesthesiology. 2022;136:688–96. doi: 10.1097/ALN.0000000000004153. [DOI] [PubMed] [Google Scholar]
  21. Cahan C, Decker M J, Hoekje P L, Strohl K P. Agreement between noninvasive oximetric values for oxygen saturation. Chest. 1990;97:814–9. doi: 10.1378/chest.97.4.814. [DOI] [PubMed] [Google Scholar]
  22. Cajanding R. Oxygen use and saturation targets in patients with COVID-19: are we giving too much or aiming too low? Nurs. Crit. Care. 2021;27:282–5. doi: 10.1111/nicc.12709. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Calvin J E, et al. Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health. CMAJ. 1992;146:703–12. [PMC free article] [PubMed] [Google Scholar]
  24. Cecil W T, Thorpe K J, Fibuch E E, Tuohy G F. A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters. J. Clin. Monit. 1988;4:31–6. doi: 10.1007/BF01618105. [DOI] [PubMed] [Google Scholar]
  25. Chacon P J, Pu L, da Costa T H, Shin Y-H, Ghomian T, Shamkhalichenar H, Wu H-C, Irving B A, Choi J-W. A wearable pulse oximeter with wireless communication and motion artifact tailoring for continuous use. IEEE Trans. Biomed. Eng. 2018;66:1505–13. doi: 10.1109/TBME.2018.2874885. [DOI] [PubMed] [Google Scholar]
  26. Chong A V, Terosiet M, Histace A, Romain O. Prototype of a multispectral and multiwavelength pulse oximeter based on a BQJ sensor. 2017 15th IEEE Int. New Circuits and Systems Conf. (NEWCAS); 2017. pp. pp 373–6. [Google Scholar]
  27. Çiçek H S, Gümü S , S, Deniz Ö, Yildiz S, Açikel C H, Çakir E, Tozkoparan E, Uçar E, Bilgiç H. Effect of nail polish and henna on oxygen saturation determined by pulse oximetry in healthy young adult females. Emerg. Med. J. 2011;28:783–5. doi: 10.1136/emj.2010.096073. [DOI] [PubMed] [Google Scholar]
  28. Coté C J, Goldstein E A, Fuchsman W H, Hoaglin D C. The effect of nail polish on pulse oximetry. Anesth. Analg. 1988;67:683–6. [PubMed] [Google Scholar]
  29. Crooks C J, et al. Pulse oximeters’ measurements vary across ethnic groups: an observational study in patients with Covid-19 infection. Eur. Respiratory J. 2022;59:1–4. doi: 10.1183/13993003.03246-2021.. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Cysewska-Sobusiak A. Problems with evaluating readings of pulse oximeters used at home during the ongoing COVID-19 pandemic. Photonics Lett. Poland. 2021;13:19–21. doi: 10.4302/plp.v13i1.1093. [DOI] [Google Scholar]
  31. Dumas C, Wahr J A, Tremper K K. Clinical evaluation of a prototype motion artifact resistant pulse oximeter in the recovery room. Anesthesia Analgesia. 1996;83:269–72. doi: 10.1213/00000539-199608000-00012. [DOI] [PubMed] [Google Scholar]
  32. Dyer O. Pulse oximetry may underestimate hypoxaemia in black patients, study finds. Brit. Med. J. 2020;371:m4926. doi: 10.1136/bmj.m4926. [DOI] [Google Scholar]
  33. Ebmeier S J, et al. A two centre observational study of simultaneous pulse oximetry and arterial oxygen saturation recordings in intensive care unit patients. Anaesthesia Intensive Care. 2018;46:297–303. doi: 10.1177/0310057X1804600307. [DOI] [PubMed] [Google Scholar]
  34. Eichhorn J H. Pulse oximetry as a standard of practice in anesthesia. Anesthesiology. 1993;78:423–6. [PubMed] [Google Scholar]
  35. Falconer R J, Robinson B J. Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers. Br. J. Anaesth. 1990;65:552–7. doi: 10.1093/bja/65.4.552. [DOI] [PubMed] [Google Scholar]
  36. Fawzy A, et al. Racial and ethnic discrepancy in pulse oximetry and delayed identification of treatment eligibility among patients with COVID-19. JAMA Intern. Med. 2022;21224:1–9. doi: 10.1001/jamainternmed.2022.1906. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. FDA Inaccuracy of pulse oximeters based on skin pigmentation requires additional FDA regulation. Chest 2022 [Google Scholar]
  38. Feiner J R, Rollins M D, Sall J W, Eilers H, Au P, Bickler P E. Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia. Anesth. Analg. 2013;117:847–58. doi: 10.1213/ANE.0b013e31828610a0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  39. Feiner J R, Severinghaus J W, Bickler P E. Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth. Analg. 2007;105:6. doi: 10.1213/01.ane.0000285988.35174.d9. [DOI] [PubMed] [Google Scholar]
  40. Ferrari M, Quaresima V, Scholkmann F. Pulse oximetry, racial bias and statistical bias: further improvements of pulse oximetry are necessary. Ann. Intensive Care. 2022;12:12. doi: 10.1186/s13613-022-00992-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Fine J, Branan K L, Rodriguez A J, Boonya-ananta T A, Ramella-Roman J C, McShane M J, Coté G L. Sources of Inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors. 2021;11:126. doi: 10.3390/bios11040126. [DOI] [PMC free article] [PubMed] [Google Scholar]
  42. Fitzpatrick J E, High W A, Kyle W L. Introduction to Clinical Dermatology. In: Fitzpatrick J E, et al., editors. Urgent care dermatology Symptom-Based Diagnosis. s.I.: Elsevier; 2017. pp. pp 1–6. [Google Scholar]
  43. Foglia E E, Whyte R K, Chaudhary A, Mott A, Chen J, Propert K J, Schmidt B. The effect of skin pigmentation on the accuracy of pulse oximetry in infants with hypoxemia. J. Pediatr. 2017;182:375–7. doi: 10.1016/j.jpeds.2016.11.043. [DOI] [PMC free article] [PubMed] [Google Scholar]
  44. Fuller B B, Spaulding D T, Smith D. Regulation of the catalytic activity of preexisting tyrosinase in black and caucasian human melanocyte cell cultures. Exp. Cell. Res. 2001;262:197–208. doi: 10.1006/excr.2000.5092. [DOI] [PubMed] [Google Scholar]
  45. Gabrielczyk M R, Buist R J. Pulse oximetry and postoperative hypothermia. An evaluation of the Nellcor N-100 in a cardiac surgical intensive care unit. Anaesthesia. 1988;43:402–4. doi: 10.1111/j.1365-2044.1988.tb09025.x.. [DOI] [PubMed] [Google Scholar]
  46. Goldman J M, Petterson M T, Kopotic R J, Barker S J. Masimo signal extraction pulse oximetry. J. Clin. Monit. Comput. 2000;16:475–83. doi: 10.1023/A:1011493521730. [DOI] [PubMed] [Google Scholar]
  47. Gootenberg D B, et al. Developing a pulse oximetry home monitoring protocol for patients suspected with COVID-19 after emergency department discharge. BMJ Health Care Inform. 2021;28:e100330. doi: 10.1136/bmjhci-2021-100330. [DOI] [PMC free article] [PubMed] [Google Scholar]
  48. GOV.UK The Use and Regulation of Pulse Oximeters (Information for Healthcare Professionals). 2021. Available https://gov.uk/guidance/the-use-and-regulation-of-pulse-oximeters-information-for-healthcare-professionals .
  49. Guber A, Epstein S G, Kohn S, Shitrit D. Wrist-sensor pulse oximeter enables prolonged patient monitoring in chronic lung diseases. J. Med. Syst. 2019;43:230. doi: 10.1007/s10916-019-1317-2. [DOI] [PubMed] [Google Scholar]
  50. Greenhalgh T, Knight M, Inda-Kim M, Fulop N J, Leach J, Vindrola-Padros C. Remote management of covid-19 using home pulse oximetry and virtual ward support. Brit. Med. J. 2021;372:n677. doi: 10.1136/bmj.n677.. [DOI] [PubMed] [Google Scholar]
  51. Hay 0 Y, Cohen M, Nitzan I, Kasirer Y, Shahroor-karni S, Yitzhaky Y, Engelberg S, Nitzan M. Pulse Oximetry with two infrared wavelengths without calibration in extracted arterial blood. Sensors. 2018;18:3457. doi: 10.3390/s18103457. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. Haynes A B, et al. Safe surgery saves lives study group. A surgical safety checklist to reduce morbidity and mortality in a global population. N. Engl. J. Med. 2009;360:491–9. doi: 10.1056/NEJMsa0810119. [DOI] [PubMed] [Google Scholar]
  53. Henry N R, Hanson A C, Schulte P J, Warner N S, Manento M N, Weister T J, Warner M A. Disparities in hypoxemia detection by pulse oximetry across self-identified racial groups and associations with clinical outcomes. Crit. Care Med. 2022;50:204–11. doi: 10.1097/CCM.0000000000005394. [DOI] [PMC free article] [PubMed] [Google Scholar]
  54. Holder A L, Wong A I. The big consequences of small discrepancies: why racial differences in pulse oximetry errors matter. Crit. Care Med. 2022;50:335–7. doi: 10.1097/CCM.0000000000005447. [DOI] [PMC free article] [PubMed] [Google Scholar]
  55. Hoppe-Seyler F. Ueber die chemischen und optischen Eigenschaften des Blutfarbstoffs. Archiv. f. pathol. Anat. 1864;29:233–5. doi: 10.1007/BF01937823. [DOI] [Google Scholar]
  56. Hoskin R W, Granger R. Intraoperative decrease in pulse oximeter readings following injection of isosul-fan blue. Can. J. Anaesth. 2001;48:38–40. doi: 10.1007/BF03019812. [DOI] [PubMed] [Google Scholar]
  57. Hüfner G. Ueber das oxyhiimoglobin des pferdes. Z. physiol. Chern. 1884;8:338–65. [Google Scholar]
  58. Hunasikatti M. Racial bias in accuracy of pulse oximetry and its impact on assessments of hypopnea and T90 in clinical studies. J. Clin. Sleep Med. 2021;17:1145. doi: 10.5664/jcsm.9178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  59. Huong, Ngu X. The application of extended modified lambert beer model for measurement of blood carboxyhemoglobin and oxyhe-moglobin saturation. J. Innov. Opt. Health Sci. 2014;07:1450026. doi: 10.1142/S1793545814500266. [DOI] [Google Scholar]
  60. Ishiyama T, Kotoda M, Asano N, Ikemoto K, Mitsui K, Sato H, Matsukawa T, Sessler D I. The effects of patent blue dye on peripheral and cerebral oxyhaemoglobin saturations. Anaesthesia. 2015;70:429–33. doi: 10.1111/anae.12932. [DOI] [PubMed] [Google Scholar]
  61. Jacques S L. Skin Optics summary. Oregon Medical Laser Center News. 1998. [Online]. Available: https://omlc.org/news/jan98/skinoptics.html .
  62. Johnson Jr B L, Moy R L, White G M. Ethnic Skin: Medical and Surgical. St. Louis, Missouri: Mosby; 1998. pp. pp 3–5. [Google Scholar]
  63. Jothishankar B, Stein S L. Impact of skin color and ethnicity. Clin. Dermatol. 2019;37:418–29. doi: 10.1016/j.clindermatol.2019.07.009. [DOI] [PubMed] [Google Scholar]
  64. Jubran A. Pulse oximetry. Crit. Care. 2015;19:272. doi: 10.1186/s13054-015-0984-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  65. Jubran A, Tobin M J. Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients. Chest. 1990;97:1420–5. doi: 10.1378/chest.97.6.1420. [DOI] [PubMed] [Google Scholar]
  66. Kanellis V G. A review of melanin sensor devices. Biophys. Rev. 2019;11(6):843–9. doi: 10.1007/s12551-019-00581-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  67. Kelleher J F. Pulse oximetry. J. Clin. Monit. 1989;5:37–62. doi: 10.1007/BF01618369. [DOI] [PubMed] [Google Scholar]
  68. Knight M J, Subbe C P, Inada-Kim M. Racial discrepancies in oximetry: where do we stand? Anaesthesia. 2022;77:129–31. doi: 10.1111/anae.15635. [DOI] [PubMed] [Google Scholar]
  69. Kramer K. Bestimmung des sauerstoffgehaltes und der himoglobin konzentration in himoglobinlslungen und himolysierten blut auf lichtelektrischen wege. Z. Biol. 1934;95:126–34. [Google Scholar]
  70. Kyriacou P A, Charlton P H, Al-Halawani R, Shelley K H. Inaccuracy of pulse oximetry with dark skin pigmentation: clinical implications and need for improvement. Br J. Anaesth. 2022;130:e33. doi: 10.1016/j.bja.2022.03.011. [DOI] [PubMed] [Google Scholar]
  71. Lancet E A, et al. Prehospital hypoxemia, measured by pulse oximetry, predicts hospital outcomes during the New York City COVID-19 pandemic. J. Am. Coll. Emerg. Physicians Open. 2021;2:e12407. doi: 10.1002/emp2.12407. [DOI] [PMC free article] [PubMed] [Google Scholar]
  72. Lawless S T. Crying wolf: false alarms in a pediatric intensive care unit. Crit. Care Med. 1994;22:981–5. [PubMed] [Google Scholar]
  73. Lee K H, Hui K P, Tan W C, Lim T K. Factors influencing pulse oximetry as compared to functional arterial saturation in multi-ethnic Singapore. Singap. Med. J. 1993;34:385–7. [PubMed] [Google Scholar]
  74. Lemay M, et al. Chapter 18 -Applications of optical cardiovascular monitoring. In: E Sazonov., editor. Edward Sazonov, Wearable Sensors. New York: Academic; 2021. pp. pp 487–517. [Google Scholar]
  75. Levitan R M. Pulse oximetry as a biomarker for early identification and hospitalization of COVID-19 pneumonia. Acad Emerg Med. 2020;27:785–6. doi: 10.1111/acem.14052. [DOI] [PMC free article] [PubMed] [Google Scholar]
  76. Lipnick M S, Feiner J R, Au P, Bernstein M, Bickler P E. The accuracy of 6 Inexpensive pulse oximeters not cleared by the food and drug administration: the possible global public health implications. Anesth. Analg. 2016;123:338–45. doi: 10.1213/ANE.0000000000001300. [DOI] [PubMed] [Google Scholar]
  77. Louie A, Feiner J R, Bickler P E, Rhodes L, Bernstein M, Lucero J. Four types of pulse oximeters accurately detect hypoxia during low perfusion and motion. Anesthesiol. J. Am. Soc. Anesthesiol. 2018;128:520–30. doi: 10.1097/ALN.0000000000002002. [DOI] [PubMed] [Google Scholar]
  78. Luks A M, Swenson E R. Pulse oximetry for monitoring patients with COVID-19 at home. potential pitfalls and practical guidance. Ann. Am. Thorac. Soc. 2020;17:1040–6. doi: 10.1513/AnnalsATS.202005-418FR. [DOI] [PMC free article] [PubMed] [Google Scholar]
  79. Ly B C K, et al. Research techniques made simple: cutaneous colorimetry: a reliable technique for objective skin color measurement. J. Investigative Dermatology. 2020;140:3–12. doi: 10.1016/j.jid.2019.11.003. [DOI] [PubMed] [Google Scholar]
  80. Mahle W T, et al. American heart association congenital heart defects committee of the council on cardiovascular disease in the young, council on cardiovascular nursing, and interdisciplinary council on quality of care and outcomes research; american academy of pediatrics section on cardiology and cardiac surgery, and committee on fetus and newborn. Role of pulse oximetry in examining newborns for congenital heart disease: a scientific statement from the american heart association and american academy of pediatrics. Circulation. 2009;120:447–58. doi: 10.1161/CIRCULATIONAHA.109.192576. [DOI] [PubMed] [Google Scholar]
  81. Mannheimer P D, Casciani J R, Fein M E, Nierlich S L. Wavelength selection for low-saturation pulse oximetry. IEEE Trans. Biomed. Eng. 1997a;44:148–58. doi: 10.1109/10.554761. [DOI] [PubMed] [Google Scholar]
  82. Mannheimer P D, Fein M E, Casciani J R. Physio-optical considerations in the design of fetal pulse oximetry sensors. Eur. J. Obstet. Gynecol. Reprod. Biol. 1997b;72:9–19. doi: 10.1016/s0301-2115(97)02713-9. [DOI] [PubMed] [Google Scholar]
  83. Mantri Y, Jokerst J V. Impact of skin tone on photoacoustic oximetry and tools to minimize bias. Biomed. Opt. Express. 2022;13:875–87. doi: 10.1364/BOE.450224. [DOI] [PMC free article] [PubMed] [Google Scholar]
  84. Mathes A M, Kreuer S, Schneider S O, Ziegeler S, Grundmann U. The performance of six pulse oximeters in the environment of neuronavigation. Anesth. Analg. 2008;107:541–4. doi: 10.1213/ane.0b013e31817e6778. [DOI] [PubMed] [Google Scholar]
  85. Mayerhöfer T G, Pahlow S, Popp J. The bouguer-beer–lambert law: shining light on the obscure. Chem. Phys. Chem. 2020;21:2029–46. doi: 10.1002/cphc.202000464. [DOI] [PMC free article] [PubMed] [Google Scholar]
  86. Meng S, Kaxiras E. Theoretical models of eumelanin protomolecules and their optical properties. Biophys. J. 2008;94:2095–105. doi: 10.1529/biophysj.107.121087. [DOI] [PMC free article] [PubMed] [Google Scholar]
  87. Michael W S, et al. Racial bias in pulse oximetry measurement. New Engl. J. Med. 2020;383:2477–8. doi: 10.1056/NEJMc2029240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  88. Michard F, Shelley K, L’Her E. COVID-19: pulse oximeters in the spotlight. J. Clin. Monit. Comput. 2021;35:11–4. doi: 10.1007/s10877-020-00550-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  89. Millikan G A. The oximeter, an instrument for measuring continuously the oxygen saturation of arterial blood in man. Rev. Sci. Instrum. 1942;13:434–44. doi: 10.1063/1.1769941. [DOI] [Google Scholar]
  90. Morris R W, Nairn M, Torda T A. A comparison of fifteen pulse oximeters: I. A clinical comparison; II. A test of performance under conditions of poor perfusion. Anaesth. Intensive Care. 1989;17:62–73. doi: 10.1177/0310057X8901700113. [DOI] [PubMed] [Google Scholar]
  91. Moyle J T B. Principles and Practice Series: Pulse Oximetry. London: BMJ Publishing Group; 1994. [Google Scholar]
  92. Murphy S M, Omar S. The clinical utility of noninvasive pulse co-oximetry hemoglobin measurements in dark-skinned critically ill patients. Anesth. Analg. 2018;126:1519–26. doi: 10.1213/ANE.0000000000002721. [DOI] [PubMed] [Google Scholar]
  93. Nafiu O O, Mpody C, Kim S S, Uffman J C, Tobias J D. Race, postoperative complications, and death in apparently healthy children. Pediatrics. 2020;146:e20194113. doi: 10.1542/peds.2019-4113. [DOI] [PubMed] [Google Scholar]
  94. Naik P P, Farrukh S N. Influence of ethnicities and skin color variations in different populations: a review. Skin Pharmacol Physiol. 2022;35:65–76. doi: 10.1159/000518826. [DOI] [PubMed] [Google Scholar]
  95. Ngiam J N, Chew N W, Sia C H, Kong W K, Poh K K. Silent hypoxia: pulse oximetry and its relation to COVID-19 in Singapore. Singap. Med. J. 2021;64:222–5. doi: 10.11622/smedj.2021134. [DOI] [PMC free article] [PubMed] [Google Scholar]
  96. Nicolai L. Uber Sichtbarmachung, Verlauf und chemische Kinetik der Oxyhemoglobinreduktion im ebenden Gewebe, besonders in der menschlichen. Arch. Ges. Physiol. 1932;229:372–89. doi: 10.1007/BF01754476. [DOI] [Google Scholar]
  97. Nitzan M, Nitzan I, Arieli Y. The various oximetric techniques used for the evaluation of blood oxygenation. Sensors (Basel). 2020;20:4844. doi: 10.3390/s20174844. [DOI] [PMC free article] [PubMed] [Google Scholar]
  98. Okunlola O E, Lipnick M S, Batchelder P B, Bernstein M, Feiner J R, Bickler P E. Pulse oximeter performance, racial inequity, and the work ahead. Respir. Care. 2022;67:252–7. doi: 10.4187/respcare.09795. [DOI] [PubMed] [Google Scholar]
  99. Open Critical Care Oxygen FAQ: Skin pigmentation. [Online] https://opencriticalcare.org/faq-category/skin/
  100. Perkins G D, McAuley D F, Giles S, Routledge H, Gao F. Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation? Crit. Care. 2003;7:R67. doi: 10.1186/cc2339. [DOI] [PMC free article] [PubMed] [Google Scholar]
  101. Philip K E J, et al. Working accuracy of pulse oximetry in COVID-19 patients stepping down from intensive care: a clinical evaluation. BMJ Open Respir. Res. 2020;7:e000778. doi: 10.1136/bmjresp-2020-000778. [DOI] [PMC free article] [PubMed] [Google Scholar]
  102. Philip K E J, Tidswell R, McFadyen C. Racial bias in pulse oximetry: more statistical detail may help tackle the problem. Brit. Med. J. 2021;372:n298. doi: 10.1136/bmjresp-2020-000778. [DOI] [PubMed] [Google Scholar]
  103. Quaresima V, Ferrari M. COVID-19: efficacy of prehospital pulse oximetry for early detection of silent hypoxemia. Crit. Care. 2020;24:501. doi: 10.1186/s13054-020-03185-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  104. Ralston A C, Webb R K, Runciman W B. Potential errors in pulse oximetry: III. Effects of interferences, dyes, dyshaemoglobins and other pigments. Anaesthesia. 1991;46:291–5. doi: 10.1111/j.1365-2044.1991.tb11501.x. [DOI] [PubMed] [Google Scholar]
  105. Rawlings A V. Ethnic skin types: are there differences in skin structure and function? Int. J. Cosmet. Sci. 2006;28:79–93. doi: 10.1111/j.1467-2494.2006.00302.x. [DOI] [PubMed] [Google Scholar]
  106. Reuss J L. Multilayer modeling of reflectance pulse oximetry. IEEE Trans. Biomed. Eng. 2005;52:153–9. doi: 10.1109/TBME.2004.840188. [DOI] [PubMed] [Google Scholar]
  107. Ries A L, Prewitt L M, Johnson J J. Skin color and ear oximetry. Chest. 1989;96:287–90. doi: 10.1378/chest.96.2.287. [DOI] [PubMed] [Google Scholar]
  108. Rodden A M, Spicer L, Diaz V A, Steyer T E. Does fingernail polish affect pulse oximeter readings? Intensive Crit. Care Nurs. 2007;23:51–5. doi: 10.1016/j.iccn.2006.08.006. [DOI] [PubMed] [Google Scholar]
  109. Rodriguez C. Using pulse oximetry to monitor high-risk patients with COVID-19 at home. Nursing. 2020;50:15–6. doi: 10.1097/01.NURSE.0000718376.94916.eb. [DOI] [PubMed] [Google Scholar]
  110. Ross P A, Newth C J, Khemani R G. Accuracy of pulse oximetry in children. Pediatrics. 2014;133:22–9. doi: 10.1542/peds.2013-1760. [DOI] [PubMed] [Google Scholar]
  111. Rosychuk R J, Hudson-Mason A, Eklund D, Lacaze-Masmonteil T. Discrepancies between arterial oxygen saturation and functional oxygen saturation measured with pulse oximetry in very preterm infants. Neonatology. 2012;101:14–9. doi: 10.1159/000326797. [DOI] [PubMed] [Google Scholar]
  112. Saito J, Kitayama M, Kato R, Hirota K. Interference with pulse oximetry by the stealth station image guidance system. J. Clin. Rep. 2017;3:6. doi: 10.1186/s40981-017-0076-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  113. Schulte T E, Ohnoutka J R, Agrawal A. BrainLAB interference with pulse oximetry during stereotactic brain biopsy. J. Clin. Anesth. 2012;24:675. doi: 10.1016/j.jclinane.2012.03.010. [DOI] [PubMed] [Google Scholar]
  114. Severinghaus J W. Arterial blood gases. In: Laurent G J, Shapiro S D, editors. Encyclopedia of Respiratory Medicine. New York: Academic; 2006. pp. pp 144–50. [Google Scholar]
  115. Shah N, Ragaswamy H B, Govindugari K, Estanol L. Performance of three new-generation pulse oximeters during motion and low perfusion in volunteers. J. Clin. Anesth. 2012;24:385–91. doi: 10.1016/j.jclinane.2011.10.012. [DOI] [PubMed] [Google Scholar]
  116. Shah S, Majmudar K, Stein A, Gupta N, Suppes S, Karamanis M, Capannari J, Sethi S, Patte C. Novel use of home pulse oximetry monitoring in COVID-19 patients discharged from the emergency department identifies need for hospitalization. Acad. Emerg. Med. 2020;27:681–92. doi: 10.1111/acem.14053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  117. Shi C, et al. The effects of skin pigmentation on the accuracy of pulse oximetry in measuring oxygen saturation: a systematic review and meta-analysis. BMC Medicine. 2022;20:267. doi: 10.1186/s12916-022-02452-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  118. Sjoding M, Iwashyna T J, Valley T S. More on racial bias in pulse oximetry measurement. Reply. N. Engl. J. Med. 2021;384:1278. doi: 10.1056/NEJMc2101321. [DOI] [PubMed] [Google Scholar]
  119. Smith R N, Hofmeyr R. Perioperative comparison of the agreement between a portable fingertip pulse oximeter v. a conventional bedside pulse oximeter in adult patients (COMFORT trial) S. Afr. Med. J. 2019;109:154–8. doi: 10.7196/SAMJ.2019.v109i3.13633. [DOI] [PubMed] [Google Scholar]
  120. So-Ling C, Li L. A multi-layered reflection model of natural human skin. Computer Graphics International; 2001. pp. pp 249–56. [DOI] [Google Scholar]
  121. Squire J. Instrument for measuring quantity of blood and its degree of oxygenation in web of the hand. Clin. Sci. 1940;4:331–9. [Google Scholar]
  122. Sriganesh K, Vinay B, Bhadrinarayan V. Indocyanine green dye administration can cause oxygen desaturation. J. Clin. Monit. Comput. 2013;27:371. doi: 10.1007/s10877-012-9423-4. [DOI] [PubMed] [Google Scholar]
  123. Stokes G G. On the Reduction and Oxidation of the Colouring Matter of the Blood. Proc R. Soc. 1864;13:355–64. [Google Scholar]
  124. Szabó G, et al. Racial differences in the fate of melanosomes in human epidermis. Nature. 1969;222:1081–2. doi: 10.1038/2221081a0. [DOI] [PubMed] [Google Scholar]
  125. Teo J. Early detection of silent hypoxia in Covid-19 pneumonia using smartphone pulse oximetry. J. Med. Syst. 2020;44:134. doi: 10.1007/s10916-020-01587-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  126. Tobin M J, Jubran A. Pulse oximetry, racial bias and statistical bias. Ann. Intensive Care. 2022;12:2. doi: 10.1186/s13613-021-00974-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  127. Tobin M J, Laghi F, Jubran A. Why COVID-19 silent hypoxemia is baffling to physicians. Am. J. Respir. Crit. Care. Med. 2020;202:356–60. doi: 10.1164/rccm.202006-2157CP. [DOI] [PMC free article] [PubMed] [Google Scholar]
  128. Toda K, et al. Alteration of racial differences in melanosome distribution in human epidermis after exposure to ultraviolet light. Nat. New Biol. 1972;236:143–5. doi: 10.1038/newbio236143a0. [DOI] [PubMed] [Google Scholar]
  129. Todd B. Pulse oximetry may be inaccurate in patients with darker skin. AJN Am. J. Nurs. 2021;121:16. doi: 10.1097/01.NAJ.0000742448.35686.f9. [DOI] [PubMed] [Google Scholar]
  130. Tremper K K, Barker S J. Pulse oximetry. Anesthesiology. 1989;70:98–108. doi: 10.1097/00000542-198901000-00019. [DOI] [PubMed] [Google Scholar]
  131. Tytler J A, Seeley H F. The nellcor N-101 pulse oximeter. Anaesthesia. 1986;41:302–5. doi: 10.1111/j.1365-2044.1986.tb12793.x. [DOI] [PubMed] [Google Scholar]
  132. Valbuena V S M, et al. Racial bias in pulse oximetry measurement among patients about to undergo extracorporeal membrane oxygenation in 2019-2020: a retrospective cohort study. Chest. 2021;161:971–8. doi: 10.1016/j.chest.2021.09.025. [DOI] [PMC free article] [PubMed] [Google Scholar]
  133. Verdugo-Naranjo I, et al. Radiodermatitis grade estimation by RGB color imaging. Artif. Life Robot. 2022;27:58–63. doi: 10.1007/s10015-022-00739-w. [DOI] [Google Scholar]
  134. Vesoulis Z, et al. Racial discrepancy in pulse oximeter accuracy in preterm infants. J. Perinatol. 2021;42:79–85. doi: 10.1038/s41372-021-01230-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  135. Visscher M O. Skin color and pigmentation in ethnic skin. Facial Plast. Surg. Clin. North Am. 2017;25:119–25. doi: 10.1016/j.fsc.2016.08.011. [DOI] [PubMed] [Google Scholar]
  136. Volgyesi G A, Spahr-Schopfer I. Does skin pigmentation affect the accuracy of pulse oximetry? an in vitro study. Anesthesiology. 1991;75:A406. doi: 10.1097/00000542-199109001-00406. [DOI] [Google Scholar]
  137. Wahr J A, Tremper K K, Diab M. Pulse oximetry. Respir. Care Clin. N Am. 1995;1:77–105. [PubMed] [Google Scholar]
  138. Wang Y T, Poh S C. Noninvasive oximetry in pigmented patients. Ann. Acad. Med. 1985;14:427–9. [PubMed] [Google Scholar]
  139. Watabe H, et al. Regulation of tyrosinase processing and trafficking by organellar pH and by proteasome activity. J. Biol. Chem. 2004;279:7971–81. doi: 10.1074/jbc.M309714200. [DOI] [PubMed] [Google Scholar]
  140. Webster J G, editor. Design of Pulse Oximeters. 1st edn. Boca Raton, FL: CRC Press; 1997. [DOI] [Google Scholar]
  141. Westhorpe R N, Ball C. The pulse oximeter. Anaesthesia Intensive Care. 2008;36:767. doi: 10.1177/0310057X0803600602. [DOI] [PubMed] [Google Scholar]
  142. Whitehead C T. More on racial bias in pulse oximetry measurement. N. Engl. J. Med. 2021;384:1278. doi: 10.1056/NEJMc2101321. [DOI] [PubMed] [Google Scholar]
  143. Wiles M D, et al. The effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis: a single-centre, retrospective analysis. Anaesthesia. 2021;77:143–52. doi: 10.1111/anae.15581. [DOI] [PMC free article] [PubMed] [Google Scholar]
  144. Wiles M D, et al. Effect of patient ethnicity on the accuracy of peripheral pulse oximetry in patients with COVID-19 pneumonitis requiring mechanical ventilation. Anaesthesia. 2022;77:489–91. doi: 10.1111/anae.15656. [DOI] [PubMed] [Google Scholar]
  145. Wilkes M, Wright C Y, du Plessis J L, Reeder A. Fitzpatrick skin type, individual typology angle, and melanin index in an african population: steps toward universally applicable skin photosensitivity assessments. JAMA Dermatol. 2015;151:902–3. doi: 10.1001/jamadermatol.2015.0351. [DOI] [PubMed] [Google Scholar]
  146. Wilkins C J, Moores M, Hanning C D. Comparison of pulse oximeters: effects of vasoconstriction and venous engorgement. Br. J. Anaesth. 1989;62:439–44. doi: 10.1093/bja/62.4.439. [DOI] [PubMed] [Google Scholar]
  147. Wilson-Baig N, McDonnell T, Bentley A. Discrepancy between SpO2 and SaO2 in patients with COVID-19. Anaesthesia. 2021;76:6–7. doi: 10.1111/anae.15228. [DOI] [PMC free article] [PubMed] [Google Scholar]
  148. Witting M D, Scharf S M. Diagnostic room-air pulse oximetry: effects of smoking, race, and sex. Am. J. Emerg. Med. 2008;26:131–6. doi: 10.1016/j.ajem.2007.04.002. [DOI] [PubMed] [Google Scholar]
  149. Wong A I, et al. Analysis of discrepancies between pulse oximetry and arterial oxygen saturation measurements by race and ethnicity and association with organ dysfunction and mortality. JAMA Netw. Open. 2021;4:e2131674. doi: 10.1001/jamanetworkopen.2021.31674. [DOI] [PMC free article] [PubMed] [Google Scholar]
  150. Wood E H. A single scale absolute reading ear oximeter. Proc Staff MeetMayo Clin. 1950;25:384–91. [PubMed] [Google Scholar]
  151. Yamamoto L G, Yamamoto J A, Yamamoto J B, Yamamoto B E, Yamamoto P P. Nail polish does not significantly affect pulse oximetry measurements in mildly hypoxic subjects. Respir. Care. 2008;53:1470–4. [PubMed] [Google Scholar]
  152. Yönt G H, Korhan E A, Dizer B. The effect of nail polish on pulse oximetry readings. Intensive Crit. Care Nurs. 2014;30:111–5. doi: 10.1016/j.iccn.2013.08.003. [DOI] [PubMed] [Google Scholar]
  153. Zonios G, Bykowski J, Kollias N. Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy. J. Invest. Dermatol. 2001;117:1452–7. doi: 10.1046/j.0022-202x.2001.01577.x. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

All data that support the findings of this study are included within the article (and any supplementary information files).