Abstract
Visual electrophysiology affords direct, quantitative, objective assessment of visual pathway function at different levels, and thus yields information complementary to, and not necessarily obtainable from, imaging or psychophysical testing. The tests available, and their indications, have evolved, with many advances, both in technology and in our understanding of the neural basis of the waveforms, now facilitating more precise evaluation of physiology and pathophysiology. After summarising the visual pathway and current standard clinical testing methods, this review discusses, non-exhaustively, several developments, focusing particularly on human electroretinogram recordings. These include new devices (portable, non-mydiatric, multimodal), novel testing protocols (including those aiming to separate rod-driven and cone-driven responses, and to monitor retinal adaptation), and developments in methods of analysis, including use of modelling and machine learning. It is likely that several tests will become more accessible and useful in both clinical and research settings. In future, these methods will further aid our understanding of common and rare eye disease, will help in assessing novel therapies, and will potentially yield information relevant to neurological and neuro-psychiatric conditions.
Subject terms: Retinal diseases, Retina
摘要 摘要
视觉电生理学能直接、定量、客观地评估不同水平的视觉通路功能, 因此可获得从影像学或心理物理测试中不一定能获得的互补信息。随着技术进步和对波形神经基础的理解的加深, 现有的测试及其指南的发展有助于更精确地评估视觉同路生理学和病理生理学的改变。本文总结了视觉通路和当前的标准临床测试方法, 不完全地讨论了几个设备, 尤其关注人类视网膜电图记录。其中包括新的设备(便携式、免散瞳、多模式)、新的测试指南(包括旨在分离视杆驱动和视锥驱动反应, 和监测视网膜适应的指南), 以及建模和机器学习的使用等分析方法的发展。在临床和研究中, 有几种测试可能便于获取并且具有实用性。在未来, 这些方法将进一步帮助我们了解眼部的常见和罕见病, 有助于评估新的治疗方法, 并可能提供与神经和神经精神疾病相关的信息。
Introduction
Transformational advances have occurred in recent decades in multimodal imaging of the retina and visual pathway. Such techniques, however, largely convey information on anatomical structure. Electrophysiology permits direct, objective and quantitative assessment of function, both at the level of the visual cortex (visual evoked potentials, VEPs) and at the level of the retina (electroretinogram, ERG). This review will briefly summarise the visual pathway, list current standard electrophysiological testing protocols, and then describe some recent advances and anticipated future developments, with particular focus on the ERG.
The neuronal pathway of visual signals
After refraction and transmission by the optical media, light hits the retina, isomerising photopigment and initiating the phototransduction cascade in the outer segments of the rod and cone photoreceptors. This results in the shut-off of a cationic current (that was flowing into the outer segments in the dark), leading to hyperpolarisation of the photoreceptors (the membrane potential becoming more negative). This in turn leads to reduction in glutamate release at the synapse between photoreceptors and bipolar cells. ON bipolar cells depolarise (membrane potential becomes more positive) in response to this, whilst OFF bipolar cells hyperpolarise. Bipolar cells synapse with ganglion cells, whose axons form the optic nerve. Other neuronal cell types in the retina, horizontal cells and amacrine cells, modify transmission at these synapses, subserving lateral interactions between photoreceptors and bipolar cells respectively, such that the information transmitted in the ganglion cell axons represent highly processed features of the visual environment.
The optic nerves (formed from the axons of the retinal ganglion cells) from each eye meet at the optic chiasm: the fibres originating from the nasal retina of each eye decussate to join the contralateral optic tract, whilst those originating in the temporal retina continue in the ipsilateral optic tract. The ganglion cell axons synapse with neurons in the lateral geniculate nucleus, with the latter neurons then projecting in the optic radiations to the primary visual cortex, located in the occipital lobes. Thus, the right visual cortex receives input from the temporal retina of the right eye and the nasal retina of the left eye (both responding to stimuli in the left visual field), whilst the left visual cortex receives input from the temporal retina of the left eye and the nasal retina of the right eye (thus responding to stimuli in right visual space).
Visual electrophysiology tests
Full-field electroretinogram
The electroretinogram (ERG) represents the summed electrical response of the retina to light stimuli. The potential difference is measured between a corneal electrode and a reference skin electrode (the latter usually placed on the temple), with a ground electrode commonly placed on the forehead. The full-field ERG [1] is elicited by flashes that stimulate the whole of the retina. This typically consists of a negative deflection, the a-wave, which largely reflects the hyperpolarisation of the photoreceptors, followed by a positive deflection, the b-wave, that largely reflects the depolarisation of the ON bipolar cells. The amplitudes and peak times of these components are conventionally measured, as shown in the upper panels of Fig. 1. (The lower panels of Fig. 1 highlight additional features of the waveforms, which will be discussed in a later section.)
Fig. 1. ERG responses to white flashes delivered in the dark-adapted and light-adapted state.
Traces show ERG responses (averaged from several flash presentation) recorded from a healthy individual to strong flashes (10 photopic cd s m−2, corresponding to the DA10 of the ISCEV standard) delivered in the dark following dark adaptation (lefthand panels) and to standard flashes (3 photopic cd s m−2, corresponding to the LA3 of the ISCEV standard) delivered on a standard white background (30 photopic cd m−2) in the light-adapted state (righthand panels). Labels in the upper panels highlight the quantitative parameters usually reported (a-wave and b-wave amplitudes and peak times); the oscillatory potentials (OPs) on the rising limb of the b-wave may be qualitatively evaluated or quantified with additional filtering. The lower panels highlight several components in the waveforms and the underlying retinal processes or neuronal origins that could potentially be interrogated quantitatively with more sophisticated analyses (including mathematical modelling or machine learning techniques).
Selective attenuation of the b-wave suggests a locus of dysfunction that occurs after phototransduction (for example owing to impairment of bipolar cell function or reduced synaptic transmission between photoreceptors and bipolar cells) [2, 3], whilst attenuation of the a-wave indicates impairment of phototransduction. Stimuli are conventionally delivered in the dark-adapted state, to assess rod system function, and in the light-adapted state, to assess cone system function. As well as flash stimuli in the light-adapted state, a flicker stimulus (conventionally at 30 Hz, which cannot be resolved temporally by the rod system) is also used to evaluate cone system function [1]. With conventional full-field ERG stimuli, therefore, retina-wide dysfunction can be detected, and impairments can be localised to the rod system or cone system, and to the outer retina (photoreceptor outer segments) or inner retina (impairments that occur after phototransduction, affecting the photoreceptor bipolar cell synapse or bipolar cell signals).
Pattern electroretinogram
The pattern ERG [4] is elicited by a checkerboard stimulus (with alternating black and white squares reversing a few times a second) that stimulates the central 15 degrees of retina, thus testing macular function. A positive component (peaking at around 50 ms, hence termed the P50) is seen followed by a negative component (the N95, with a trough around 95 ms). Abnormalities of the pattern ERG can arise from dysfunction of the macula, including the macular cone photoreceptors, and also the macular retinal ganglion cells (in the latter case, the N95 is more severely affected than the P50).
Multifocal electroretinogram
The multifocal ERG [5] also gives information relating to more localised areas of retina than the full-field ERG. Here, an area of retina of around 40 or 50 degrees is stimulated by an array of light and dark hexagons, each of which reverses in luminance according to a pseudorandom sequence. The resulting ERG recording is then mathematically correlated with each hexagon’s sequence of illumination to yield waveforms (typically a negative deflection, followed by a positive, followed by a second negative, deflection, labelled N1, P1, and N2 respectively) corresponding to each specific area of retina (the area stimulated by the relevant hexagon).
Electro-oculogram
The electro-oculogram (EOG) [6] measures the standing potential of the eye, relating to a potential across the retinal pigment epithelium (RPE). Skin electrodes are placed at the inner and outer canthus of each eye, and patients make horizontal saccades between two targets. As the eyes move, the potential across the RPE reverses relative to the electrodes, resulting in deflections seen in the recording. In the dark, the amplitudes of these deflections fall to a minimum (the “dark trough”), and in the light, they rise to a peak (the “light peak”) over the course of several minutes. The light peak:dark trough ratio is selectively reduced in certain conditions (whilst the ERG is normal), typically in Best disease.
Visual evoked potentials
Visual evoked potentials (VEPs) [7] are recorded from electrodes placed on the scalp and overlying the visual cortex. These can be evoked by flash stimuli or patterned stimuli (the latter are in a checkerboard configuration of light and dark squares, similar to the stimulus for the pattern ERG). Abnormalities can arise from impairments anywhere in the visual pathway, and so VEPs are usually best interpreted in the context of ERG findings, for example in distinguishing between a retinopathy/maculopathy or optic neuropathy. Pattern reversal VEPs (in which the black and white squares reverse) show least variability between individuals. Pattern onset VEPs can give objective estimates of visual acuity (by evaluating the smallest check sizes that appear to be resolved at the level of the visual cortex) [8, 9], helpful in cases of suspected non-organic visual loss. Flash VEPs can be particularly useful in cases of poorer cooperation and fixation, or when optical factors might degrade a patterned image.
VEPs can aid in evaluating optic nerve disease and multi-channel VEPs can help detect intracranial misrouting associated with albinism. In albinism, the majority of retinal ganglion cell axons cross at the chiasm rather than solely those from the nasal retina. Thus, whilst in healthy individuals a monocular stimulus will produce similar VEP responses over both hemispheres, in albinism, a monocular stimulus produces a larger response over the contralateral hemisphere.
ISCEV standard protocols
Visual electrophysiology waveforms are dependent upon stimulus and recording parameters, as well as physiological or pathological changes in the visual pathway. The International Society for Clinical Electrophysiology of Vision (ISCEV) defines standard protocols, which enables some uniformity across laboratories and in publications. The standards are revised periodically. Currently there are standards for the full-field ERG, the pattern ERG, multifocal ERG, EOG and VEP [1, 4–7]. Table 1 lists the standard protocols along with the year of most recent update, and summarises the relevant stimuli. ISCEV also publishes extended protocols [9–16], which will be discussed later.
Table 1.
International Society of Clinical Electrophysiology of Vision (ISCEV) standard protocols with reference to latest updated versions at the current time (Jan 2023).
Test | Latest update | Authors | Summary Comments |
---|---|---|---|
Full-field ERG | 2022 | Robson et al. | White full-field stimuli delivered in the dark-adapted (DA) state (following 20 min dark adaptation) and in the light-adapted (LA) state (in the presence of a 30 cd m−2 white background, following 10 min adaptation to this background if dark-adapted responses have been recorded first). DA responses are to 0.01, 3 and 10 cd s m−2 flashes. LA responses are to 30 Hz flicker and to flashes (both 3 cd s m−2). Pupils should be dilated. If non-mydriatic responses are recorded, stimulus strength should be adjusted to give equivalent retinal illuminance. The document also includes a non-standard abbreviated protocol that can be helpful for example with very young patients. |
Multi-focal ERG | 2021 | Hoffmann et al. | Stimulus display consists of 61 or 103 scaled hexagons (recommended total recording time of 4 or 8 min respectively), with frame rate between 60 and 75 Hz. Luminance of elements in the light state should be 100 cd m−2 or more. Pupils are dilated. Fixation is important. |
Pattern ERG | 2012 | Bach et al. | Stimulus is contrast-reversing checkerboard pattern (luminance of white areas should be 80 cd m−2 or more), with reversal rate of 4.0 reversals per second. Pupils are not dilated; fixation and optimal refraction are important. Stimulus field should be 15 degrees in width and height. Check size is 0.8 degrees. |
Electro-oculogram | 2017 | Constable et al. | Stimuli presented within a uniform field (integrating sphere). Pupils are dilated. Electrodes placed at canthi. Patient makes saccades between two red fixation lights (15 degrees left and right of centre) that alternate every second for 10 s in every minute, in the dark for 15 min, and then in a white light background (100 cd m−2) for 15 min. The light peak to dark trough ratio is calculated. |
Visual evoked potentials | 2016 | Odom et al. | VEPs are recorded to pattern (checkerboard) stimuli (subtending 15 degrees; mean luminance 50 cd m−2) and to flash stimuli, presented monocularly. Pattern-reversal VEPs (2 reversals per second), and pattern onset/offset VEPs are recorded to checkerboard stimuli with large (1 degree) and small (0.25 degree) checks. Flash VEPs are elicited by a flash (3 cd m−2 s) subtending 20 degrees or more of visual field. A single midline occipital active electrode is used (multi-channel recording is not part of the basic standard, but needed to assess chiasmal and post-chiasmal pathway dysfunction and chiasmal misrouting). Pupils not dilated. |
Stimulus strengths are in photopic units. The reader is referred to the relevant documents for full details.
Some current and future developments
There are numerous recent and anticipated developments in the field of visual electrophysiology. In 2021, a special section in this journal featured several relevant articles, each focusing on a particular aspect, summarising current knowledge, providing new insights and future directions [3, 17–24]. For the remainder of the current article, some advances, particularly relating to human ERGs, will be highlighted, categorised broadly into new devices, new stimulus protocols, and novel analyses. These are also summarised in Table 2.
Table 2.
Examples of developments in visual electrophysiology (list not exhaustive), some of which are described in more detail in the text.
Developments | Examples |
Advances in devices |
•Portable/handheld equipment •Real-time pupil monitoring and adjusting of stimulus strength to deliver desired retinal illuminance, allowing non-mydriatic recordings •Multimodal platforms combining psychophysical and electrophysiological testing, or combining high-resolution fundus imaging (including OCT) with electrophysiology |
Stimulus protocols |
•ISCEV extended protocols •Protocols to separate dark-adapted rod and cone responses or to selectively modulate specific photoreceptor classes (silent substitution) •Dynamic tracking of retinal light and dark adaptation using electrophysiology |
Novel analyses |
•Waveform transformations •Fitting of mechanistic models to derive parameters of phototransduction or outer retinal current flows in the living human eye •Artificial intelligence/machine learning algorithms •Genetic analyses (association of electrophysiology parameters with common and rare genetic variants) |
Advances in devices
For decades, xenon flash guns and incandescent lamps were used to provide flash and background stimuli. Developments in LED technology allowed more energy efficient devices and more precise control and stability of stimuli. Over 20 years ago, one hurdle in the investigation of kinetics of recovery of human cone photoreceptor circulating current at sub-second time scales following very intense steady bleaching exposures was the time taken to extinguish (or for the subject to move away from) the steady bleaching background [25]. With the availability of ultrabright LEDs, bright bleaching backgrounds could be extinguished rapidly, and bright xenon flashes could be delivered within tens of milliseconds following background extinction, helping establish that cone circulating current recovers extremely rapidly: following a bleaching exposure estimated to bleach around 90% of the cone photopigment, cone circulating current, as estimated by the bright-flash ERG appeared to recover fully within 100 ms (and substantially within 20 ms) [26]. For rods, following a similar large bleach, recovery of circulating current takes at least 20 min; [27] thus, recovery of current in cones was shown to occur over 10,000 times faster [26].
Portable nonmydriatic devices
Whilst handheld stimulators have been available for many years, more sophisticated portable devices have been developed more recently. The RETeval device (LKC Technologies) can deliver near full-field stimuli equivalent to the ISCEV standard, and can be used with a variety of electrodes, including “Sensor Strip” skin electrodes, which are placed on the skin below the lower lid margin and which incorporate a recording, reference and ground electrode within the same strip. The device includes a camera, so the subject’s eye can be viewed (and eye closure noted) by the operator. The device also measures pupil diameter and can adjust stimulus strength accordingly (to match the retinal illuminance delivered by standard stimuli through a dilated pupil), which allows use with natural pupils, removing the need for pharmacological mydriasis. Figure 2 illustrates the device and example waveforms. Many publications have emerged reporting its use in diverse conditions including diabetic retinopathy, a range of inherited retinal diseases, birdshot chorioretinopathy, glaucoma, idiopathic intracranial hypertension, and others [28–47].
Fig. 2. A portable nonmydriatic stimulus and recording system (RETeval, LKC Technologies).
A Demonstration of positioning of device over the eye from which recordings are being taken. The electrodes being used are the “Sensor Strip” skin electrodes (LKC Technologies), but the device can also be used in conjunction with other electrode types. B Example ERG responses recorded with this technique in response to a stimulus equivalent to the standard light-adapted flash (each trace is the average of approximately 30 flash presentations; the green and orange traces are from two successive series of flashes). C View of device screen prior to initiation of nonmydriatic recordings. The device has an inbuilt video camera, so the subject’s eye is visible. The device also detects the pupil (highlighted by blue circle) and will adjust stimulus strength to deliver the retinal illuminance equivalent to that delivered by standard stimuli through a dilated pupil. D Responses to a stimulus equivalent to the standard light-adapted 30 Hz flicker.
Multimodal devices
Another important development in device technology is multimodality. Both historically and currently, separate pieces of equipment are used for psychophysical testing (including perimetry), retinal imaging and electrophysiology. More recently, some manufacturers have combined psychophysical testing capability and electrophysiology within the same device, for example dark adaptometry, full-field stimulus testing (FST), combined with electrophysiology (Diagnosys LLC) [48, 49]. Fundus cameras have also been used in conjunction with ERG to reliably localise the area of stimulation [50, 51]. Very recently, a system has been developed incorporating both optical coherence tomography (OCT) and ERG recording, aiming to stimulate very small areas of retina with OCT guidance, and record responses (Nanoscope Technologies LLC) [52].
Newer stimulus protocols
ISCEV extended protocols
In addition to the well-established ISCEV standard protocols, several extended protocols [9–16], endorsed by ISCEV, have been developed to probe, in more detail, aspects of visual function. These are summarised in Table 3. Some have been incorporated in portable devices, with findings reported in several studies [3, 31, 33, 35, 36, 39]. An ISCEV guide has been published reviewing the range of standard and extended protocols [53].
Table 3.
ISCEV extended protocols with summary notes and year of latest update. All stimulus strengths are in photopic units unless otherwise specified.
Protocol | Latest update | Authors | Summary Comments |
---|---|---|---|
VEP estimation of visual acuity | 2021 | Hamilton et al. | Guidelines for recording pattern VEPs to range of check sizes to establish spatial frequency limit and relating this to visual acuity. Helpful to estimate visual acuity when conventional psychophysical testing not possible or reliable. |
Derivation and analysis of strong flash rod-isolated ERG a-wave | 2020 | Brigell et al. | ERGs recorded to a strong white flash (30 cd s m−2) under dark adaptation and additionally following light adaptation in standard background (30 cd m−2). Estimated rod-driven ERG a-wave derived by subtracting light-adapted (cone-driven) response from the dark-adapted (rod and cone-driven ERG). Potential value in characterisation of diseases selectively affecting rod phototransduction or numbers of rod photoreceptors. |
S-cone ERG | 2020 | Perlman et al. | ERGs recorded in response to a blue flash (450–470 nm) delivered on a longer wavelength (yellow, amber or orange) background (background peak wavelength between 570–620 nm, luminance 300 cd m−2). Helpful in distinguishing between achromatopsia and blue cone monochromacy, and in confirming diagnosis of enhanced S-cone syndrome. |
Stimulus response series, light-adapted ERG | 2019 | McCulloch et al. | Light-adapted ERGs recorded to 9 flash stimuli (0.03 to 300 cd s m−2) in presence of standard background (30 cd m−2) with guidance on fitting stimulus-response relation. Shorter protocol also provided. Enables more comprehensive characterisation of cone system function than standard light-adapted stimuli, including characterisation of “photopic hill” phenomenon (b-wave amplitude increasing with stimulus strength to peak and then falling to non-zero plateau). |
Stimulus response series, dark-adapted b-wave | 2019 | Johnson et al. | Dark-adapted ERGs recorded in response to large range of flash strengths (typically from −3.5 to 0.5 log cd s m−2) with guidance on fitting of stimulus-response relation. Provides more comprehensive characterisation of rod system function than standard dark-adapted stimuli. |
Photopic negative response (PhNR) | 2018 | Frishman et al. | The PhNR is a negative waveform component following the b-wave and arises from retinal ganglion cells (RGCs). The recommended stimulus is a red (630–660 nm) LED flash (1.0–2.5 cd s m−2) on a blue (450–485 nm) LED background (100 scotopic cd m−2). Potential use in evaluating RGC function (in glaucoma and other optic neuropathies). |
Dark-adapted red flash | 2018 | Thompson et al. | Dark-adapted ERG to dim red flash (0.3 cd s m−2), shows an early (30–50 ms) positive cone-driven peak followed by a later (~100 ms) larger rod-driven peak. Allows assessment of dark-adapted cone system, helpful in evaluating cone system dysfunction (particularly retinopathy associated with variants in RGS9 or R9AP), and also in determining origin (rod or cone) of residual standard dark-adapted ERGs. |
On-Off ERG | 2018 | Sustar et al. | ERGs recorded to long duration (150 or 200 ms) stimuli. Stimuli are white (150–350 cd m−2) on a white background (30 cd m−2), or orange (620 nm) stimuli on a green (560 nm) background. The long stimulus allows separate evaluation of response to stimulus onset and offset. Helpful in distinguishing conditions selectively impairing ON bipolar signals (e.g., complete congenital stationary night blindness and melanoma-associated-retinopathy) from those impairing both ON and OFF signals (incomplete congenital stationary night blindness and other disorders). |
The On-Off ERG entails the delivery of long duration (>100 ms) stimuli so that retinal responses to onset and offset of stimulation can be separately evaluated. This can be helpful in evaluating conditions in which ON bipolar cell signals are selectively attenuated (including complete congenital stationary night blindness and melanoma-associated retinopathy) [16]. Delivery of dim red flashes in the dark-adapted state can yield assessment of the dark-adapted cone system [15]. The photopic negative response is thought to be best elicited by red flashes delivered on a blue background [14]. It derives from retinal ganglion cells, and has been shown to be attenuated in optic neuropathies, including glaucoma. Stimulus response series have been published for the dark-adapted and light-adapted ERGs [12, 13], where ERGs are recorded in response to a wider range of flash strengths than in the standard for full-field ERGs. Selective S-cone ERGs (elicited by blue flashes delivered on a longer wavelength background) can help confirm the diagnosis of enhanced S-cone syndrome (typically related to bi-allelic variants in NR2E3) and demonstrate S-cone preservation in blue cone monochromacy (helping distinguish it electrophysiologically from achromatopsia) [11]. A published protocol also exists to derive the dark-adapted rod-driven a-wave response to strong flashes (by subtraction of the cone-driven component yielded by the response to the same flash delivered on the standard ISCEV light-adapting background) [10]. Recently, a protocol has been published for estimation of visual acuity from the pattern VEP, using different check sizes to establish the spatial frequency limit [9].
Parsing dark-adapted rod and cone responses
Separating rod and cone contributions to responses to strong flashes delivered in the dark can pose a particular challenge, since both rods and cones are stimulated. The extended protocol mentioned above [10] specifies delivery of the same flash in the presence of the standard ISCEV white background to estimate the cone-driven contribution in the dark. However, the standard white background is likely to significantly light adapt the cone system, so the response elicited by a flash delivered on this background is likely to differ somewhat from the cone-driven response to a flash delivered in the dark. Alternative methods take advantage of the different spectral sensitivities of rods and cones: rods are maximally sensitive to shorter (bluer) wavelengths than the majority of cones (which are L- and M-cones, whilst S-cones are relatively sparse). Thus, a blue background can be applied which saturates the rods, but minimally light-adapts the majority of the cones. Responses recorded to flashes delivered on this background will be cone-driven, with no rod-driven component. These responses will more closely match the dark-adapted cone system response than those recorded on the standard white background. If these responses (recorded on the dim blue rod-saturating background) are subtracted from responses to the same flashes delivered in the dark, the result provides an estimate of the rod-driven response (Fig. 3, upper panels). Several studies have used this method [25, 26, 54–58], including a recent investigation into ERG associations of a common myopia-associated variant [58]. In that study, the technique revealed an association specifically with cone-driven, but not rod-driven, responses.
Fig. 3. Examples of non-standard stimulus protocols to interrogate specific aspects of retinal function.
A, B Upper panels show a method to separate rod and cone-driven contributions to dark-adapted flash responses. A Blue traces show ERG responses to blue flashes of a range of strengths delivered in the dark-adapted state (each trace averaged from multiple flash presentations). These responses contain contributions from rod and cone systems. The red traces show responses to the same flashes, but delivered on a blue background. The background strength was 2.9 photopic cd m−2 and 34 scotopic cd m−2. Delivered through a dilated pupil, the background illuminance (approximately 1500 to 1700 scotopic trolands) is likely to be sufficient to saturate the rods (which are thought to be largely saturated by backgrounds in excess of 1000 scotopic trolands), but minimally adapt the cones. Thus, the flash responses recorded are cone-driven with no rod-driven components. B Traces plotted are the result of mathematical subtraction of the red traces from the blue traces in panel A. These traces represent the estimated isolated rod-driven component (with the a-waves largely reflecting current flows in the rod photoreceptors). C, D Lower panels show an ERG method of tracking recovery of rod system sensitivity in the dark following a bright light exposure. C ERG responses to dim flashes of fixed strength (0.02 scotopic cd m−2) delivered at different times in the dark after steady state exposure to a standard white background (30 photopic cd m−2). The smaller amplitude responses were recorded at earlier times following extinction of the background, whilst the larger amplitude responses were recorded at later times. D Amplitudes of responses in C plotted as a function of post-bleach time. Amplitudes have been normalised to the estimated final dark-adapted level (denoted by the horizontal dashed line).
Other methods of separating rod and cone contributions exist, including a double flash protocol whereby flashes are delivered at early times (around 1 second) following a very strong flash [54, 59]. At this time, rods will still be in saturation, and hence the response will be cone-driven. It is also possible to separate photoreceptor classes at particular levels of adaptation (rods and cones, and even L and M cones, which have considerable overlap in wavelength sensitivity) using “silent substitution” methods whereby the spectral composition of stimuli is manipulated such that certain photoreceptor classes (for example L or M cones) can be selectively modulated whilst maintaining the same stimulation of the other photoreceptors. This has been used psychophysically and also in electrophysiology [60–62].
Tracking retinal adaptation with the ERG
Conventionally, ERGs are recorded in a steady (either dark-adapted or light-adapted) state. It is possible, however, to use the ERG to dynamically track adaptation to different backgrounds. Whilst dark adaptation has mainly been measured psychophysically, an ERG-based method allows objective and quantitative tracking of recovery of rod and cone sensitivity at the level of the retina. Typically, following a bleaching exposure (steady exposure to bright illumination or to a series of strong flashes), stimuli of a fixed strength are delivered repeatedly in the dark. The ERG elicited by such stimuli is initially of low amplitude, but the amplitude recovers with time in the dark. These studies have given insights into kinetics of rod and cone photopigment regeneration and recovery of photoreceptor outer segment current [27, 56, 57, 63–67]. Figure 3 (lower panels) shows an ERG method of using dim flashes to track rod system recovery in the dark, variations of which have been used in several investigations [63, 65, 66, 68].
Such methods also helped demonstrate the physiological basis of the phenomenon of transient smartphone “blindness” [68]. In these cases, patients reported recurrent monocular nyctalopia, initially eluding medical explanation. Following careful history-taking, the reduced visual sensitivity occurred following smartphone viewing whilst lying in bed on one side (such that one eye was inadvertently covered by the pillow; this eye became dark adapted, whilst the uncovered eye light adapted). By recording ERGs to dim flashes delivered to both eyes at different times following monocular smartphone viewing, it was shown that the phenomenon was simply one of retinal adaptation and demonstrable at the level of rod-driven bipolar cell responses [68]. This ERG method of tracking recovery of rod or cone sensitivity may have utility in conditions in which dark adaptation is selectively affected (including vitamin A deficiency, Sorsby fundus dystrophy, and age-related macular degeneration). Use of the ERG in tracking rod and cone dark adaptation has recently been extensively reviewed, along with discussion of potential applications [69].
Novel analyses
Figure 1 shows standard dark-adapted and light-adapted full-field flash ERG responses. The upper panels, as mentioned earlier, show the parameters that are conventionally measured and reported. The lower panels highlight multiple components in the waveform and potential inferences that could be made, with more advanced quantitative analyses, regarding physiological and pathophysiological processes in specific retinal neuronal groups. Several papers have used more sophisticated analyses, such as discrete wavelet transforms [70–72], applied for example to separately evaluate ON and OFF components in responses [70]. Frequently, novel analyses are used in conjunction with specifically designed protocols, hence many of these developments have occurred simultaneously.
Mathematical models of phototransduction and outer retinal current flows
A mechanistic model, taking into account the likely kinetics of each of the activation stages of phototransduction, was formulated over 30 years ago by Lamb and Pugh [73]. This was shown to provide a remarkably close fit, using a single set of parameters, to changes in the photoreceptor outer segment current in response to a range of flash strengths. This model was also shown to provide a reasonable fit to the leading edge of the ERG a-wave, and was fitted to both rod and cone-driven a-waves in numerous studies [25, 54, 57, 66, 74–81].
However, limitations to this approach have been identified [79, 82]. The model did not provide a close fit to ERG responses to very strong flashes, and so a capacitive time constant was incorporated to improve the fit [77]. Also, the cone-driven a-wave has been shown to contain significant post-receptoral (OFF bipolar cell) signals [58, 83, 84], and so will not solely reflect the cone photoreceptor current. Importantly, also, the a-wave trough and initial recovery (termed by some the a-wave “nose”) seen in the dark-adapted response to strong flashes has been shown to be likely to arise from currents flowing in other parts of the photoreceptor [82].
Robson and Frishman have described a more comprehensive mathematical model (taking into account both outer segment as well as photoreceptor axonal currents) that provides a good fit to the rod-driven a-wave elicited by strong flashes, including the a-wave trough and the initial subsequent recovery towards baseline (the nose) prior to the main part of the b-wave [82]. Figure 4 illustrates application of the model to rod-driven responses to a strong flash. Such models could have applicability in photoreceptor diseases, helping elucidate the altered current flows in the outer retina of affected patients.
Fig. 4. Illustration of mathematical models fitted to rod-driven a-wave response to strong flash.
A Derived dark-adapted rod-driven response to strong blue flash from a healthy individual; this is the largest amplitude trace from Fig. 3B. This trace is also reproduced in panels B and C. B The red dashed curve shows a fit based on the model of Lamb & Pugh, developed for the outer segment photocurrent. The curve was fitted with the assumption that the photoreceptor response is truncated as a result of the encroaching of post-receptoral signals giving rise to the b-wave (not accounted for in the model), and hence the curve continues downward past the a-wave trough. C Red dashed curve shows a fit based on the later model of Robson & Frishman. This model explicitly takes into account current flows in other parts of the photoreceptor layer, showing that the a-wave trough region in the bright-flash ERG is consistent with arising from current flows proximal to the photoreceptor outer segments.
Artificial intelligence algorithms
Artificial intelligence (AI) is being applied to many areas of healthcare showing high levels of accuracy, equivalent to experts. There have been investigations applying AI or machine-learning techniques to electrophysiology data [34, 85–90]. These include studies of ERG data that may have applicability in conditions including glaucoma [88], hydroxychloroquine retinopathy [87], and even autism spectrum disorder [86] and depression [89], as well as studies of VEP data to improve estimation of acuity [90]. A recent study demonstrated that machine learning techniques could be successfully applied to develop automated ERG phenotypic classification of patients with Stargardt disease (arising from variants in the ABCA4 gene) [85]. This study also used the ERG data to provide a severity score for each ABCA4 variant (over 1000 different disease-causing variants have been described), which could help give prognostic information for patients based on their genotype. As well as potentially assisting with rapid diagnosis (which could have widespread application given the relative scarcity of ERG expertise), AI techniques could in future yield clinically relevant classifications using subtle features in the waveform data that are undetectable by human experts. AI techniques could also be used to integrate, quantitatively, electrophysiology recordings with imaging and other phenotypic (or genotypic) data to generate predictions of diagnoses, together with levels of certainty.
Investigating effects of common genetic variants
Combing electrophysiological analyses with genetic data can be powerful, not just in the field of rare diseases such as ABCA4-retinopathy, but also in interrogating effects of common genetic variants on retinal function. ERGs in healthy twins show high heritability [91], indicating that genetic factors are likely to contribute significantly to variance in response parameters in the general population. Large genome-wide association studies have identified many variants associated with common eye diseases. The common variant that is consistently most strongly associated with myopia is near the gene GJD2 which encodes retinal gap junctions. In the study described earlier [58], a mixed linear model was used, with ERG parameters as outcomes, and allelic dosage at this locus as a predictor, to test the hypothesis that aspects of retinal electrophysiology might associate with this locus. An association was found with some cone-driven ERG parameters, consistent with a possible role for altered cone-driven signals in the pathogenesis of myopia. Studies of this type, particularly if many loci are to be investigated, require large numbers of genotyped individuals with ERG data, and it is anticipated that future investigations in larger cohorts will be informative.
Concluding remarks
Electrophysiology has provided insights into the physiology and pathophysiology of the human retina and visual pathway for over a century. International standardisation of testing protocols, with regular updates, has been a major advance in bringing about consistency in methods and allowing data generated in laboratories across the world to be meaningfully interpreted. As retinal imaging has advanced tremendously in recent decades, with high resolution cross-sectional images now acquired as a standard in many clinical settings, and single photoreceptors visualised using adaptive optics techniques in research settings, the indications for electrophysiology have evolved, but the tests are far from obsolete. Uniquely, electrophysiology permits direct, quantitative in vivo evaluation of visual pathway function at the level of the retina and visual cortex, and thus provides information not obtainable from imaging. Additionally, such functional testing could potentially detect neuronal dysfunction, for example in diabetic eye disease, prior to changes observable on imaging.
This review is not exhaustive, and many advances have not been covered. With development of newer devices (including portable and multimodal technology), and more refined, including more rapid, testing protocols, combined with novel, AI-assisted analyses, it is likely that tests will become more accessible and continue to yield valuable clinical and scientific information. This will have relevance to common and rare diseases of the eye and visual pathway, and also, given similarities between retinal and brain circuitry, potentially to wider neurological and neuropsychiatric disease [71, 85, 88, 92–94].
Acknowledgements
The author is grateful to Professor Trevor Lamb FRS and Professor John Robson FRS for prior discussions relating to fitting of a-wave models, and for providing model simulations.
Funding
OAM is funded by the Wellcome Trust (Grant 206619/Z/17/Z), and has received research funding from Fight for Sight (UK), the Thomas Pocklington Trust, the Birdshot Uveitis Society, Moorfields Eye Charity and the NIHR Biomedical Research Centre at Moorfields Eye Hospital and the UCL Institute of Ophthalmology. The views expressed are those of the author and not the funding organisations.
Author contributions
OAM conceived and wrote the manuscript.
Data availability
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
Competing interests
The author declares no competing interests.
Footnotes
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Robson AG, Frishman LJ, Grigg J, Hamilton R, Jeffrey BG, Kondo M, et al. ISCEV Standard for full-field clinical electroretinography (2022 update) Doc Ophthalmol. 2022;144:165–77. doi: 10.1007/s10633-022-09872-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Audo I, Robson AG, Holder GE, Moore AT. The negative ERG: clinical phenotypes and disease mechanisms of inner retinal dysfunction. Surv Ophthalmol. 2008;53:16–40. doi: 10.1016/j.survophthal.2007.10.010. [DOI] [PubMed] [Google Scholar]
- 3.Jiang X, Mahroo OA. Negative electroretinograms: genetic and acquired causes, diagnostic approaches and physiological insights. Eye. 2021;35:2419–37. doi: 10.1038/s41433-021-01604-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Bach M, Brigell MG, Hawlina M, Holder GE, Johnson MA, McCulloch DL, et al. ISCEV standard for clinical pattern electroretinography (PERG): 2012 update. Doc Ophthalmol. 2013;126:1–7. doi: 10.1007/s10633-012-9353-y. [DOI] [PubMed] [Google Scholar]
- 5.Hoffmann MB, Bach M, Kondo M, Li S, Walker S, Holopigian K, et al. ISCEV standard for clinical multifocal electroretinography (mfERG) (2021 update) Doc Ophthalmol. 2021;142:5–16. doi: 10.1007/s10633-020-09812-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Constable PA, Bach M, Frishman LJ, Jeffrey BG, Robson AG, International Society for Clinical Electrophysiology of Vision. ISCEV Standard for clinical electro-oculography (2017 update) Doc Ophthalmol. 2017;134:1–9. doi: 10.1007/s10633-017-9573-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Odom JV, Bach M, Brigell M, Holder GE, McCulloch DL, Mizota A, et al. ISCEV standard for clinical visual evoked potentials: (2016 update) Doc Ophthalmol. 2016;133:1–9. doi: 10.1007/s10633-016-9553-y. [DOI] [PubMed] [Google Scholar]
- 8.Hamilton R, Bach M, Heinrich SP, Hoffmann MB, Odom JV, McCulloch DL, et al. VEP estimation of visual acuity: a systematic review. Doc Ophthalmol. 2021;142:25–74. doi: 10.1007/s10633-020-09770-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Hamilton R, Bach M, Heinrich SP, Hoffmann MB, Odom JV, McCulloch DL, et al. ISCEV extended protocol for VEP methods of estimation of visual acuity. Doc Ophthalmol. 2021;142:17–24. doi: 10.1007/s10633-020-09780-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Brigell M, Jeffrey BG, Mahroo OA, Tzekov R. ISCEV extended protocol for derivation and analysis of the strong flash rod-isolated ERG a-wave. Doc Ophthalmol. 2020;140:5–12. doi: 10.1007/s10633-019-09740-4. [DOI] [PubMed] [Google Scholar]
- 11.Perlman I, Kondo M, Chelva E, Robson AG, Holder GE. ISCEV extended protocol for the S-cone ERG. Doc Ophthalmol. 2020;140:95–101. doi: 10.1007/s10633-019-09730-6. [DOI] [PubMed] [Google Scholar]
- 12.McCulloch DL, Kondo M, Hamilton R, Lachapelle P, Messias AMV, Robson AG, et al. ISCEV extended protocol for the stimulus-response series for light-adapted full-field ERG. Doc Ophthalmol. 2019;138:205–15. doi: 10.1007/s10633-019-09685-8. [DOI] [PubMed] [Google Scholar]
- 13.Johnson MA, Jeffrey BG, Messias AMV, Robson AG. ISCEV extended protocol for the stimulus-response series for the dark-adapted full-field ERG b-wave. Doc Ophthalmol. 2019;138:217–27. doi: 10.1007/s10633-019-09687-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Frishman L, Sustar M, Kremers J, McAnany JJ, Sarossy M, Tzekov R, et al. ISCEV extended protocol for the photopic negative response (PhNR) of the full-field electroretinogram. Doc Ophthalmol. 2018;136:207–11. doi: 10.1007/s10633-018-9638-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Thompson DA, Fujinami K, Perlman I, Hamilton R, Robson AG. ISCEV extended protocol for the dark-adapted red flash ERG. Doc Ophthalmol. 2018;136:191–7. doi: 10.1007/s10633-018-9644-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Sustar M, Holder GE, Kremers J, Barnes CS, Lei B, Khan NW, et al. ISCEV extended protocol for the photopic On-Off ERG. Doc Ophthalmol. 2018;136:199–206. doi: 10.1007/s10633-018-9645-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Thompson D. Introduction to the special section on Visual electrophysiology 2021: a coming of age. Eye. 2021;35:2339–40. doi: 10.1038/s41433-021-01678-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hamilton R. Clinical electrophysiology of vision-commentary on current status and future prospects. Eye. 2021;35:2341–3. doi: 10.1038/s41433-021-01592-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.O’Neill EK, Smith R. Visual electrophysiology in the assessment of toxicity and deficiency states affecting the visual system. Eye. 2021;35:2344–53. doi: 10.1038/s41433-021-01663-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Handley SE, Šuštar M, Tekavčič, Pompe M. What can visual electrophysiology tell about possible visual-field defects in paediatric patients. Eye. 2021;35:2354–73. doi: 10.1038/s41433-021-01680-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Al-Nosairy KO, Hoffmann MB, Bach M. Non-invasive electrophysiology in glaucoma, structure and function-a review. Eye. 2021;35:2374–85. doi: 10.1038/s41433-021-01603-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Marmoy OR, Viswanathan S. Clinical electrophysiology of the optic nerve and retinal ganglion cells. Eye. 2021;35:2386–405. doi: 10.1038/s41433-021-01614-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Cornish EE, Vaze A, Jamieson RV, Grigg JR. The electroretinogram in the genomics era: outer retinal disorders. Eye. 2021;35:2406–18. doi: 10.1038/s41433-021-01659-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Thompson DA, Handley SE, Henderson RH, Marmoy OR, Gissen P. An ERG and OCT study of neuronal ceroid lipofuscinosis CLN2 Battens retinopathy. Eye. 2021;35:2438–48. doi: 10.1038/s41433-021-01594-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Paupoo AA, Mahroo OA, Friedburg C, Lamb TD. Human cone photoreceptor responses measured by the electroretinogram [correction of electoretinogram] a-wave during and after exposure to intense illumination. J Physiol. 2000;529:469–82. doi: 10.1111/j.1469-7793.2000.00469.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Kenkre JS, Moran NA, Lamb TD, Mahroo OA. Extremely rapid recovery of human cone circulating current at the extinction of bleaching exposures. J Physiol. 2005;567:95–112. doi: 10.1113/jphysiol.2005.088468. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Thomas MM, Lamb TD. Light adaptation and dark adaptation of human rod photoreceptors measured from the a-wave of the electroretinogram. J Physiol. 1999;518:479–96. doi: 10.1111/j.1469-7793.1999.0479p.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Kato K, Nagashima R, Matsubara H, Ikesugi K, Tsukitome H, Matsui Y, et al. Transient Increase of Flicker Electroretinography Amplitudes after Cataract Surgery: Association with Postoperative Inflammation. Ophthalmol Sci. 2022;3:100243. doi: 10.1016/j.xops.2022.100243. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.You JY, Dorfman AL, Gauvin M, Vatcher D, Polomeno RC, Little JM, et al. Comparing the RETeval® portable ERG device with more traditional tabletop ERG systems in normal subjects and selected retinopathies. Doc Ophthalmol. 2022. 10.1007/s10633-022-09903-w. [DOI] [PMC free article] [PubMed]
- 30.Friedel EBN, Schäfer M, Endres D, Maier S, Runge K, Bach M, et al. Electroretinography in adults with high-functioning autism spectrum disorder. Autism Res. 2022;15:2026–37. doi: 10.1002/aur.2823. [DOI] [PubMed] [Google Scholar]
- 31.Hanson JVM, Weber C, Pfäffli OA, Bassler D, McCulloch DL, Gerth-Kahlert C. Flicker electroretinogram in newborn infants. Doc Ophthalmol. 2022;145:175–84. doi: 10.1007/s10633-022-09889-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Yamashita T, Kato K, Kondo M, Miki A, Araki S, Goto K, et al. Photopic negative response recorded with RETeval system in eyes with optic nerve disorders. Sci Rep. 2022;12:9091. doi: 10.1038/s41598-022-12971-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Waldie AM, Hobby AE, Chow I, Cornish EE, Indusegaran M, Pekacka A, et al. Electrophysiological assessment in Birdshot Chorioretinopathy: Flicker Electroretinograms recorded with a handheld device. Transl Vis Sci Technol. 2022;11:23. doi: 10.1167/tvst.11.5.23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Diao T, Kushzad F, Patel MD, Bindiganavale MP, Wasi M, Kochenderfer MJ, et al. Comparison of machine learning approaches to improve diagnosis of optic neuropathy using photopic negative response measured using a handheld device. Front Med. 2021;8:771713. doi: 10.3389/fmed.2021.771713. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Han KJ, Moon H, Woo JM, Min JK. Using reteval system flicker electroretinography for evaluation of dense vitreous hemorrhage. Retina. 2022;42:73–79. doi: 10.1097/IAE.0000000000003263. [DOI] [PubMed] [Google Scholar]
- 36.Raharja A, Leo SM, Chow I, Indusegaran M, Hammond CJ, Mahroo OA, et al. Exploratory study of the association between the severity of idiopathic intracranial hypertension and electroretinogram photopic negative response amplitude obtained using a handheld device. Life. 2021;11:437. doi: 10.3390/life11050437. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Haseoka T, Inagaki R, Kurata K, Arai S, Takagi Y, Suzuki H, et al. Usefulness of handheld electroretinogram system for diagnosing blue-cone monochromatism in children. Jpn J Ophthalmol. 2021;65:23–29. doi: 10.1007/s10384-020-00782-9. [DOI] [PubMed] [Google Scholar]
- 38.Carter P, Gordon-Reid A, Shawkat F, Self JE. Comparison of the handheld RETeval ERG system with a routine ERG system in healthy adults and in paediatric patients. Eye (Lond) 2021;35:2180–9. doi: 10.1038/s41433-020-01221-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Brigell MG, Chiang B, Maa AY, Davis CQ. Enhancing. Transl Vis Sci Technol. 2020;9:40. doi: 10.1167/tvst.9.9.40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Kato K, Sugawara A, Nagashima R, Ikesugi K, Sugimoto M, Kondo M. Factors affecting photopic negative response recorded with RETeval system: Study of young healthy subjects. Transl Vis Sci Technol. 2020;9:19. doi: 10.1167/tvst.9.9.19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Kita Y, Holló G, Saito T, Momota Y, Kita R, Tsunoda K, et al. RETeval portable electroretinogram parameters in different severity stages of Glaucoma. J Glaucoma. 2020;29:572–80. doi: 10.1097/IJG.0000000000001509. [DOI] [PubMed] [Google Scholar]
- 42.Ji X, McFarlane M, Liu H, Dupuis A, Westall CA. Hand-held, dilation-free, electroretinography in children under 3 years of age treated with vigabatrin. Doc Ophthalmol. 2019;138:195–203. doi: 10.1007/s10633-019-09684-9. [DOI] [PubMed] [Google Scholar]
- 43.Liu H, Ji X, Dhaliwal S, Rahman SN, McFarlane M, Tumber A, et al. Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data. Doc Ophthalmol. 2018;137:169–81. doi: 10.1007/s10633-018-9660-z. [DOI] [PubMed] [Google Scholar]
- 44.Hobby AE, Kozareva D, Yonova-Doing E, Hossain IT, Katta M, Huntjens B, et al. Effect of varying skin surface electrode position on electroretinogram responses recorded using a handheld stimulating and recording system. Doc Ophthalmol. 2018;137:79–86. doi: 10.1007/s10633-018-9652-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Grace SF, Lam BL, Feuer WJ, Osigian CJ, Cavuoto KM, Capo H. Nonsedated handheld electroretinogram as a screening test of retinal dysfunction in pediatric patients with nystagmus. J Aapos. 2017;21:384–8. doi: 10.1016/j.jaapos.2017.06.022. [DOI] [PubMed] [Google Scholar]
- 46.Fukuo M, Kondo M, Hirose A, Fukushima H, Ikesugi K, Sugimoto M, et al. Screening for diabetic retinopathy using new mydriasis-free, full-field flicker ERG recording device. Sci Rep. 2016;6:36591. doi: 10.1038/srep36591. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Maa AY, Feuer WJ, Davis CQ, Pillow EK, Brown TD, Caywood RM, et al. A novel device for accurate and efficient testing for vision-threatening diabetic retinopathy. J Diabetes Complicat. 2016;30:524–32. doi: 10.1016/j.jdiacomp.2015.12.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Klein M, Birch DG. Psychophysical assessment of low visual function in patients with retinal degenerative diseases (RDDs) with the Diagnosys full-field stimulus threshold (D-FST) Doc Ophthalmol. 2009;119:217–24. doi: 10.1007/s10633-009-9204-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Messias K, Jägle H, Saran R, Ruppert AD, Siqueira R, Jorge R, et al. Psychophysically determined full-field stimulus thresholds (FST) in retinitis pigmentosa: relationships with electroretinography and visual field outcomes. Doc Ophthalmol. 2013;127:123–9. doi: 10.1007/s10633-013-9393-y. [DOI] [PubMed] [Google Scholar]
- 50.Miyake Y, Yanagida K, Kondo T, Yagasaki K. Simultaneous recording of local ERG and VER under direct fundus observation with infrared television fundus camera (author’s transl) Nippon Ganka Gakkai Zasshi. 1981;85:1521–33. [PubMed] [Google Scholar]
- 51.Sugita T, Kondo M, Piao CH, Ito Y, Terasaki H. Correlation between macular volume and focal macular electroretinogram in patients with retinitis pigmentosa. Invest Ophthalmol Vis Sci. 2008;49:3551–8. doi: 10.1167/iovs.08-1954. [DOI] [PubMed] [Google Scholar]
- 52.Carlson M, Kim S, Aparicio-Domingo S, Li KV, Puig B, Batabyal S, et al. OCT guided micro-focal ERG system with multiple stimulation wavelengths for characterization of ocular health. Sci Rep. 2022;12:4009. doi: 10.1038/s41598-022-07622-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Robson AG, Nilsson J, Li S, Jalali S, Fulton AB, Tormene AP, et al. ISCEV guide to visual electrodiagnostic procedures. Doc Ophthalmol. 2018;136:1–26. doi: 10.1007/s10633-017-9621-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Friedburg C, Thomas MM, Lamb TD. Time course of the flash response of dark- and light-adapted human rod photoreceptors derived from the electroretinogram. J Physiol. 2001;534:217–42. doi: 10.1111/j.1469-7793.2001.t01-1-00217.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Mahroo OA, Lamb TD. Recovery of the human photopic electroretinogram after bleaching exposures: estimation of pigment regeneration kinetics. J Physiol. 2004;554:417–37. doi: 10.1113/jphysiol.2003.051250. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Mahroo OA, Lamb TD. Slowed recovery of human photopic ERG a-wave amplitude following intense bleaches: a slowing of cone pigment regeneration? Doc Ophthalmol. 2012;125:137–47. doi: 10.1007/s10633-012-9344-z. [DOI] [PubMed] [Google Scholar]
- 57.Mahroo OA, Ban VS, Bussmann BM, Copley HC, Hammond CJ, Lamb TD. Modelling the initial phase of the human rod photoreceptor response to the onset of steady illumination. Doc Ophthalmol. 2012;124:125–31. doi: 10.1007/s10633-012-9316-3. [DOI] [PubMed] [Google Scholar]
- 58.Jiang X, Xu Z, Soorma T, Tariq A, Bhatti T, Baneke AJ, et al. Electrical responses from human retinal cone pathways associate with a common genetic polymorphism implicated in myopia. Proc Natl Acad Sci USA. 2022;119:e2119675119. doi: 10.1073/pnas.2119675119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Verdon WA, Schneck ME, Haegerstrom-Portnoy G. A comparison of three techniques to estimate the human dark-adapted cone electroretinogram. Vis Res. 2003;43:2089–99. doi: 10.1016/s0042-6989(03)00330-4. [DOI] [PubMed] [Google Scholar]
- 60.Kremers J. The assessment of L- and M-cone specific electroretinographical signals in the normal and abnormal human retina. Prog Retin Eye Res. 2003;22:579–605. doi: 10.1016/s1350-9462(03)00049-1. [DOI] [PubMed] [Google Scholar]
- 61.Jacob MM, Pangeni G, Gomes BD, Souza GS, da Silva Filho M, Silveira LC, et al. The spatial properties of L- and M-cone inputs to electroretinograms that reflect different types of post-receptoral processing. PLoS One. 2015;10:e0121218. doi: 10.1371/journal.pone.0121218. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Huchzermeyer C, Kremers J. Selective stimulation of the different photoreceptor classes by silent substitution in psychophysical and electroretinographic measurements. Klin Monbl Augenheilkd. 2022;239:1433–9. doi: 10.1055/a-1937-9901. [DOI] [PubMed] [Google Scholar]
- 63.Cameron AM, Mahroo OA, Lamb TD. Dark adaptation of human rod bipolar cells measured from the b-wave of the scotopic electroretinogram. J Physiol. 2006;575:507–26. doi: 10.1113/jphysiol.2006.108027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Binns AM, Margrain TH. Evaluating retinal function in age-related maculopathy with the ERG photostress test. Invest Ophthalmol Vis Sci. 2007;48:2806–13. doi: 10.1167/iovs.06-0392. [DOI] [PubMed] [Google Scholar]
- 65.Ruseckaite R, Lamb TD, Pianta MJ, Cameron AM. Human scotopic dark adaptation: Comparison of recoveries of psychophysical threshold and ERG b-wave sensitivity. J Vis. 2011;11:2. doi: 10.1167/11.8.2. [DOI] [PubMed] [Google Scholar]
- 66.Dimopoulos IS, Tennant M, Johnson A, Fisher S, Freund PR, Sauvé Y. Subjects with unilateral neovascular AMD have bilateral delays in rod-mediated phototransduction activation kinetics and in dark adaptation recovery. Invest Ophthalmol Vis Sci. 2013;54:5186–95. doi: 10.1167/iovs.13-12194. [DOI] [PubMed] [Google Scholar]
- 67.Lamb TD, Corless RM, Pananos AD. The kinetics of regeneration of rhodopsin under enzyme-limited availability of 11-cis retinoid. Vis Res. 2015;110:23–33. doi: 10.1016/j.visres.2015.02.014. [DOI] [PubMed] [Google Scholar]
- 68.Alim-Marvasti A, Bi W, Mahroo OA, Barbur JL, Plant GT. Transient smartphone “Blindness”. N. Engl J Med. 2016;374:2502–4. doi: 10.1056/NEJMc1514294. [DOI] [PubMed] [Google Scholar]
- 69.Jiang X, Mahroo OA. Human retinal dark adaptation tracked in vivo with the electroretinogram: insights into processes underlying recovery of cone- and rod-mediated vision. J Physiol. 2022;600:4603–21. doi: 10.1113/JP283105. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Gauvin M, Sustar M, Little JM, Brecelj J, Lina JM, Lachapelle P. Quantifying the ON and OFF contributions to the Flash ERG with the discrete wavelet transform. Transl Vis Sci Technol. 2017;6:3. doi: 10.1167/tvst.6.1.3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Constable PA, Marmolejo-Ramos F, Gauthier M, Lee IO, Skuse DH, Thompson DA. Discrete wavelet transform analysis of the electroretinogram in autism spectrum disorder and attention deficit hyperactivity disorder. Front Neurosci. 2022;16:890461. doi: 10.3389/fnins.2022.890461. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Sarossy M, Crowston J, Kumar D, Weymouth A, Wu Z. Time-frequency analysis of ERG with discrete wavelet transform and matching pursuits for glaucoma. Transl Vis Sci Technol. 2022;11:19. doi: 10.1167/tvst.11.10.19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Lamb TD, Pugh EN., Jr A quantitative account of the activation steps involved in phototransduction in amphibian photoreceptors. J Physiol. 1992;449:719–58. doi: 10.1113/jphysiol.1992.sp019111. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Hood DC, Birch DG. Light adaptation of human rod receptors: the leading edge of the human a-wave and models of rod receptor activity. Vis Res. 1993;33:1605–18. doi: 10.1016/0042-6989(93)90027-t. [DOI] [PubMed] [Google Scholar]
- 75.Breton ME, Schueller AW, Lamb TD, Pugh EN., Jr Analysis of ERG a-wave amplification and kinetics in terms of the G-protein cascade of phototransduction. Invest Ophthalmol Vis Sci. 1994;35:295–309. [PubMed] [Google Scholar]
- 76.Hood DC, Birch DG. Rod phototransduction in retinitis pigmentosa: estimation and interpretation of parameters derived from the rod a-wave. Invest Ophthalmol Vis Sci. 1994;35:2948–61. [PubMed] [Google Scholar]
- 77.Smith NP, Lamb TD. The a-wave of the human electroretinogram recorded with a minimally invasive technique. Vis Res. 1997;37:2943–52. doi: 10.1016/s0042-6989(97)00094-1. [DOI] [PubMed] [Google Scholar]
- 78.Jacobson SG, Kemp CM, Cideciyan AV, Macke JP, Sung CH, Nathans J. Phenotypes of stop codon and splice site rhodopsin mutations causing retinitis pigmentosa. Invest Ophthalmol Vis Sci. 1994;35:2521–34. [PubMed] [Google Scholar]
- 79.Cideciyan AV, Jacobson SG. An alternative phototransduction model for human rod and cone ERG a-waves: normal parameters and variation with age. Vis Res. 1996;36:2609–21. doi: 10.1016/0042-6989(95)00327-4. [DOI] [PubMed] [Google Scholar]
- 80.Cideciyan AV, Zhao X, Nielsen L, Khani SC, Jacobson SG, Palczewski K. Null mutation in the rhodopsin kinase gene slows recovery kinetics of rod and cone phototransduction in man. Proc Natl Acad Sci USA. 1998;95:328–33. doi: 10.1073/pnas.95.1.328. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Cideciyan AV, Jacobson SG, Gupta N, Osawa S, Locke KG, Weiss ER, et al. Cone deactivation kinetics and GRK1/GRK7 expression in enhanced S cone syndrome caused by mutations in NR2E3. Invest Ophthalmol Vis Sci. 2003;44:1268–74. doi: 10.1167/iovs.02-0494. [DOI] [PubMed] [Google Scholar]
- 82.Robson JG, Frishman LJ. The rod-driven a-wave of the dark-adapted mammalian electroretinogram. Prog Retin Eye Res. 2014;39:1–22. doi: 10.1016/j.preteyeres.2013.12.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Bush RA, Sieving PA. A proximal retinal component in the primate photopic ERG a-wave. Invest Ophthalmol Vis Sci. 1994;35:635–45. [PubMed] [Google Scholar]
- 84.Robson JG, Saszik SM, Ahmed J, Frishman LJ. Rod and cone contributions to the a-wave of the electroretinogram of the macaque. J Physiol. 2003;547:509–30. doi: 10.1113/jphysiol.2002.030304. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Glinton SL, Calcagni A, Lilaonitkul W, Pontikos N, Vermeirsch S, Zhang G, et al. Phenotyping of ABCA4 Retinopathy by Machine Learning Analysis of Full-Field Electroretinography. Transl Vis Sci Technol. 2022;11:34. doi: 10.1167/tvst.11.9.34. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Manjur SM, Hossain MB, Constable PA, Thompson DA, Marmolejo-Ramos F, Lee IO, et al. Detecting autism spectrum disorder using spectral analysis of electroretinogram and machine learning: Preliminary results. Annu Int Conf IEEE Eng Med Biol Soc. 2022;2022:3435–8. doi: 10.1109/EMBC48229.2022.9871173. [DOI] [PubMed] [Google Scholar]
- 87.Habib F, Huang H, Gupta A, Wright T. MERCI: a machine learning approach to identifying hydroxychloroquine retinopathy using mfERG. Doc Ophthalmol. 2022;145:53–63. doi: 10.1007/s10633-022-09879-7. [DOI] [PubMed] [Google Scholar]
- 88.Gajendran MK, Rohowetz LJ, Koulen P, Mehdizadeh A. Novel machine-learning based framework using electroretinography data for the detection of early-stage glaucoma. Front Neurosci. 2022;16:869137. doi: 10.3389/fnins.2022.869137. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Schwitzer T, Le Cam S, Cosker E, Vinsard H, Leguay A, Angioi-Duprez K, et al. Retinal electroretinogram features can detect depression state and treatment response in adults: A machine learning approach. J Affect Disord. 2022;306:208–14. doi: 10.1016/j.jad.2022.03.025. [DOI] [PubMed] [Google Scholar]
- 90.Bach M, Heinrich SP. Acuity VEP: improved with machine learning. Doc Ophthalmol. 2019;139:113–22. doi: 10.1007/s10633-019-09701-x. [DOI] [PubMed] [Google Scholar]
- 91.Bhatti T, Tariq A, Shen T, Williams KM, Hammond CJ, Mahroo OA. Relative genetic and environmental contributions to variations in human retinal electrical responses quantified in a twin study. Ophthalmology. 2017;124:1175–85. doi: 10.1016/j.ophtha.2017.03.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92.Bubl E, Dörr M, Riedel A, Ebert D, Philipsen A, Bach M, et al. Elevated background noise in adult attention deficit hyperactivity disorder is associated with inattention. PLoS One. 2015;10:e0118271. doi: 10.1371/journal.pone.0118271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Lee IO, Skuse DH, Constable PA, Marmolejo-Ramos F, Olsen LR, Thompson DA. The electroretinogram b-wave amplitude: a differential physiological measure for Attention Deficit Hyperactivity Disorder and Autism Spectrum Disorder. J Neurodev Disord. 2022;14:30. doi: 10.1186/s11689-022-09440-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Schwitzer T, Leboyer M, Laprévote V, Louis Dorr V, Schwan R. Using retinal electrophysiology toward precision psychiatry. Eur Psychiatry. 2022;65:e9. doi: 10.1192/j.eurpsy.2022.3. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.