. 2023 Aug 1;18(4):989–996. doi: 10.26603/001c.83944

Validity, Reliability, and Efficiency of a Standard Goniometer, Medical Inclinometer, and Builder’s Inclinometer

June Hanks 1,, Betsy Myers 1
PMCID: PMC10399115  PMID: 37547826

Abstract

Background

Joint range of motion (ROM) is an important assessment to aid diagnostic and clinical decision-making for persons with a wide variety of neuromusculoskeletal conditions. The current clinical standard for assessing ROM is the standard goniometer (SG).

Purpose

The purpose of this study was to investigate the validity, reliability and time required to assess ROM using a standard goniometer (SG), medical inclinometer (MI), and builder’s digital inclinometer (BI).

Study Design

Cross-sectional study.

Methods

Fifty participants with no current shoulder, elbow, or forearm pain limiting movement were assessed by a single tester. The tester measured three repetitions of passive forearm and shoulder rotation with an SG, MI, and BI. Device order was randomized. Time to complete assessment with each device was measured.

Results

BI and MI were significantly faster than the SG (p < 0.001) for all motions. Inclinometer measurements were more reliable (average ICC = 0.933 for MI and 0.919 for BI) than SG measurements (average ICC = 0.822). There was good correlation between MI and BI and mean differences between devices was less than 2°. Correlations between the SG and the inclinometers ranged from poor to fair and mean differences between devices was 4°.

Conclusion

The BI and MI were reliable for measuring forearm and shoulder rotation. The poor correlation between the SG and inclinometers indicates that clinicians should utilize the same device for testing. Because time can be a barrier to clinician assessment, the greater efficiency and reliability of inclinometers warrants consideration as the new measurement standard. Standard patient and inclinometer positioning is recommended to enhance reliability.

Level of Evidence

2

Keywords: intrarater reliability, joint motion, goniometry, inclinometry, cost-effectiveness

INTRODUCTION

Assessment of joint range of motion (ROM) aids diagnostic and clinical decision-making for persons with a wide variety of neuromusculoskeletal conditions. The current clinical standard for assessing ROM is the standard goniometer (SG), that costs about $10. Visual estimation is often used in clinical practice as a quick and free alternative to SG. However, studies of visual estimation have shown lower intrarater and interrater reliability than assessment using a SG.1,2 Although methodological differences challenge direct comparisons among studies, interrater reliability using the SG varies among studies and is weak.1,3–8 Use of a SG requires lining up the goniometer arms with standard bony landmarks. While some landmarks are superficial and easily found, others, such as the greater trochanter, are more challenging, with as much as 15 mm difference in location recorded between raters,9 leading to significant goniometric measurement errors. Maintaining SG alignment can be challenging in certain populations, such as in children with cerebral palsy10 or in people with orthopedic conditions.3 Additionally, keeping alignment generally requires two hands to reduce the risk of error.11

The HALO® (HALO, HALO Medical Devices, Australia) is a hybrid between a goniometer and an inclinometer that utilizes a laser to intersect proximal and distal anatomical landmarks relative to the joint being measured. The HALO® device costs between $160 and $200.12 While the digitization of angular measurements removes one source of measurement error, the HALO® still requires proper alignment of the laser with proximal and distal landmarks. Intrarater and interrater reliability for the HALO® has been found to be similar or slightly less than the SG for shoulder rotation.13 Smart phone applications which utilize a magnetometer-based goniometer are also available for ROM assessment. Smart phone applications have greater intrarater and interrater reliability than the SG14 and they can be useful for patient self-assessment of ROM.15,16 However, smart phone applications have not been proven reliable across multiple joints or in all movements.17,18 They also require a dedicated clinic phone (or phones) to maintain proper infection control. The lack of consistent guidance regarding alignment may contribute to measurement error.13

Bubble inclinometers are handheld devices used to assess ROM in rehabilitation settings. Bubble inclinometers have a fluid-filled circular face. The fluid is a combination of a colored fluid and clear fluid. The interface of these two fluids moves with gravity, and this movement is used to measure motion against a rotating 360° dial. Bubble inclinometers are typically used to measure spinal motion. This measurement requires the placement of two devices at specific landmarks, zeroing each device, having the patient move through a ROM, and then reading the end measurements once the fluid has stopped shifting within the device. Bubble inclinometers, therefore, require increased time, and, in the case of spinal motion, require the use of two hands. Inaccurate assessment using the bubble inclinometer includes misplacement of the inclinometer on the body part, failure to maintain constant pressure of the inclinometer against the body part and tilting of the device during reading of the face of the device.19

The digital inclinometer is a handheld device placed against a body surface to measure angular position relative to the vertical or horizontal plane. The digital inclinometer does not require alignment with multiple reference points and requires only one hand, making it easier to use than the SG. Many digital inclinometers can store multiple measurements, improving efficiency by allowing multiple motions to be assessed without needing a break to record each motion. Unlike goniometry, there are no universally accepted procedures for using inclinometers to assess joint ROM. A digital inclinometer used in clinical and research settings, referred to as a medical inclinometer (MI), is expensive, with costs exceeding $400. A builder’s digital inclinometer (BI) commonly used by construction workers operates under the same principle as a MI and is much less expensive, ranging in cost from $10 to $50.

Purpose

Studies examining the reliability and time efficiency of the SG, MI and BI are not well represented in the literature. The purpose of this study was to investigate the validity, reliability and time required to assess ROM using a SG, MI, and BI. The research hypotheses were as follows: 1) measurements taken using the inclinometers would be faster than measurements taken with the goniometer; 2) inclinometer measurements would be more reliable than goniometric measurements; 3) measurements with the inclinometers would be highly correlated; 4) measurements taken with inclinometers would be correlated with goniometric measurements.

METHODS

Fifty healthy individuals (34 females and 16 males, 23 to 58 years of age) were recruited from a local university setting to participate in the study. To be eligible, participants were at least 18 years of age with no reported current shoulder, elbow, or forearm pain limiting movement. Participants wore short-sleeve shirts to allow identification of necessary bony landmarks for testing. The devices used for ROM assessment were SGs (360°, six-inch and 12-inch), MI (Acumar, Grayline Medical; Norwalk, CA), and BI (AccuRemote, San Clemente, CA) The study was approved by the Institutional Review Board of a public university. All participants provided written informed consent prior to participation.

All measurements were taken by the same tester with more than 30 years of clinical experience as a physical therapist. All measurements were performed according to the preferred position as described by Norkin and White.20 For this study, shoulder internal/external rotation was considered as glenohumeral internal/external rotation. A 12-inch SG was used for assessing shoulder internal/external rotation and 6-inch SG was used for assessing forearm pronation/supination. The inclinometers were placed on the dorsal surface of distal forearm for shoulder internal rotation and forearm pronation and on the ventral surface of the distal forearm for shoulder external and forearm supination (Figure 1).

Figure 1. Measurement of passive range of motion with inclinometers.

Figure 1.

A. Forearm pronation with BI. B. Forearm supination with MI. C. Shoulder external rotation with BI. D. Shoulder internal rotation with MI. BI = builder’s inclinometer, MI = medical inclinometer.

The order of joint testing was as follows: forearm pronation, forearm supination, shoulder external rotation, and shoulder internal rotation. The order of devices was randomly assigned for each participant. Immediately prior to assessment of each motion, the participant performed three repetitions of active ROM, followed by one repetition of passive ROM performed by the tester. With the participant in the standard test position, the participant performed a warm-up of three repetitions of active ROM, followed by one repetition of passive ROM performed by the tester to take the joint to end range of motion prior to passive ROM measurement. The warm-up was performed to minimize the increase in ROM with repeated measurements. The tester then positioned the participant in the neutral starting position for the given joint motion. For the first repetition with each device, the tester signaled the timer to begin timing the assessment and passively moved the participant to the end range of motion and aligned the measurement device. Once the device was properly aligned at the end of joint passive ROM, the tester signaled the timer to stop. The timer then read and recorded the amount of passive ROM as well as the time required to the nearest second.

The digital display on the MI ranged from 0-180° in 1° increments. The digital display of the BI ranged from 0.0° to 90.0° in 0.1° increments, with ROM beyond 90° displayed as –89.9° to –1.0°. For ROM greater than 90°, the recording researcher calculated the ROM (e.g. for a display reading of -80°, the recorded ROM was 100°). All information was recorded on a data recording form, transferred to a spreadsheet, and uploaded to the Statistical Package for the Social Sciences (SPSS) version 28.0 for analysis.

Statistical Analysis

The Shapiro-Wilk test indicated a non-normal data distribution; therefore, the Wilcoxon Signed Rank test with an alpha level of 0.05 was used to determine the difference in the time required for each motion with each device. Intrarater reliability for each device was calculated with the intraclass correlation coefficient (ICC) (3,k) two-way mixed model with absolute agreement using the three trials for each motion.21 Construct validity, determined by correlations between devices, was assessed using the ICC.22 Interpretation of correlation strength varies among authors.8,21–23 For this study, reliability was considered excellent for ICC values > 0.90; good for values 0.70-0.89; acceptable for values 0.60-0.69; fair for values 0.50-.59 and poor for values <0.50.

The absolute reliability of each device was quantified using the standard error of measurement (SEM) and minimal detectable change (MDC).22 The SEM is a measure of how much test scores vary about a “true” score. The SEM was calculated using the formula: SEM = standard deviation * √(1-r).22 The MDC is the smallest change that can be considered a true change as opposed to change due to measurement error. The MDC was calculated using the formula: MDC = SEM * 1.96 * √2 to determine the magnitude of change that would exceed the 90% confidence interval.22 Given the normal distribution of the difference scores between devices, the level of agreement between the BI and MI was visualized using Bland-Altman plots with 95% limits of agreement calculated using the formula: mean difference ± 1.96 * standard deviation.24

RESULTS

Table 1 provides the means and standard deviations of ROM measurements for each motion with each device. On average, the mean difference between the inclinometers was less than 2°. In contrast, the average mean differences between the SG and BI were 3° and 5° for the MI.

Table 1. Descriptive Measurement Data for Range of Motion.

Goniometer Builder Medical
Mean±SD SEM MDC Mean±SD SEM MDC Mean±SD SEM MDC
Pro 80±6.5 3.1 2.5 87±7.9 2.8 2.2 89±8.8 2.8 2.2
Sup 85±5.3 2.7 2.1 89±8.3 2.9 2.3 89±9.3 3.2 2.6
ER 93±6.4 2.3 1.8 94±9.4 2.4 1.9 94±7.7 2.1 1.7
IR 67±9.0 3.5 2.8 66±10.8 2.5 2.0 62±10.7 2.3 1.9

All values are in degrees. Builder = builder’s inclinometer; medical = medical inclinometer; SEM = standard error of measurement; MDC = minimal detectable change; Pro = pronation; Sup = supination; ER = shoulder external rotation; IR = shoulder internal rotation

Research Hypothesis 1: Time Required

Figure 2 demonstrates the mean time required to perform each measurement with each device. The Wilcoxon Signed Rank test indicated that the BI and MI inclinometers required significantly less time than the goniometer (p < 0.001) for all motions. Assessment using the MI required significantly less time than the BI for both pronation and supination, p = 0.004 and p = 0.005, respectively. Therefore, the first research hypothesis was supported.

Figure 2. Goni = goniometer; Builder = builder’s inclinometer; Medical = medical inclinometer; ER = external rotation; IR = internal rotation; * = goni significantly different than builder; † = goni significantly different than medical; ‡ = builder significantly different from medical.

Figure 2.

Research Hypothesis 2: Intrarater Reliability of Devices

The results of intrarater reliability with ICC and 95% confidence intervals for each device are listed in Table 2. For convenience, the SEM and MDC for each motion and each device have been included in Table 1. While all devices had good reliability (ICC > 0.70), SG measurements were less reliable than inclinometer measurements. Except for shoulder external rotation, the MI was the most reliable device for all motions, with all but one motion (supination) exceeding 0.90, equating to excellent reliability. While slightly better for the inclinometers, absolute reliability was good considering the low SEM across devices, ranging from 1.4° to 3.5°. The MDC and was low across devices, ranging from 1.7° to 2.8°. Therefore, the second research hypothesis was supported.

Table 2. Intrarater Reliability of Devices.

Motion Goni
ICC (95% CI)
Builder
ICC (95% CI)
Medical
ICC (95% CI)
Pronation 0.770 (0.663 - 0.852) 0.878 (0.802 - 0.927) 0.901 (0.848 - 0.939)
Supination 0.748 (0.611 - 0.844) 0.876 (0.812 - 0.923) 0.878 (0.814 - 0.924)
Shoulder ER 0.870 (0.802 – 0.919) 0.933 (0.890 - 0.960) 0.925 (0.883 – 0.954)
Shoulder IR 0.849 (0.772 – 0.905) 0.946 (0.916 - 0.967) 0.952 (0.925 – 0.971)

Goni = goniometer; Builder = builder’s inclinometer; Medical = medical inclinometer; ER = external rotation; IR = internal rotation

Research Hypothesis 3: Correlations Between Inclinometers

Inter-device correlations are listed in Table 3. Correlations were highest between the BI and the MI. The ICC values for pronation, supination, and shoulder internal rotation were all > 0.70, indicating good correlation between the BI and the MI. The correlation between inclinometers for shoulder external rotation was slightly less at 0.680. Therefore, the third research hypothesis was partially supported.

Table 3. Inter-Device Correlation.

Goni-Builder
ICC (95% CI)
Goni-Medical
ICC (95% CI)
Builder – Medical
ICC (95% CI)
Pronation 0.434 (-0.034 – 0.708) 0.377 (-0.080 – 0.676) 0.711* (0.539 – 0.826)
Supination 0.557 (0.212 – 0.755) 0.534 (0.278 – 0.714) 0.718* (0.551 – 0.829)
Shoulder ER 0.664 (0.477 – 0.794) 0.647 (0.453 – 0.783) 0.680 (0.497 – 0.805)
Shoulder IR 0.727* (0.565 – 0.835) 0.430 (0.145 – 0.641) 0.704* (0.427 – 0.843)

Goni = Goniometer; Builder = Builder’s Inclinometer; Medical = Medical Inclinometer; ER = external rotation; IR = internal rotation. * = ICC > 0.700

Research Hypothesis 4: Correlations Between the Goniometer and Inclinometers

As listed in Table 3, the correlations between the SG and the BI ranged from a low of 0.434 for forearm pronation ROM to a high of 0.727 for shoulder internal rotation. The correlations between the SG and the MI were similarly low, ranging from 0.377 for forearm pronation to 0.647 for shoulder external rotation. Therefore, the fourth research hypothesis was not supported.

Data Visualization

The Bland-Altman plots in Figure 3 and Figure 4 demonstrate the 95% limits of agreement between the BI and MI. More than 95% of measurements fall within these limits and are evenly distributed both above and below the mean throughout the ranges of motion. For pronation, supination, and shoulder external rotation, the mean differences between the BI and MI were less than 2°, with most measurements within ± 5°. For shoulder internal rotation, the mean differences were approximately 4°, with most measurements within ± 5°.

Figure 3. Bland Altman plot of the differences between the builder’s and medical inclinometer for A. Pronation and B. Supination.

Figure 3.

The solid black line is the mean difference between devices and the segmented lines demonstrate the 90% limits of agreement.

Figure 4. Bland Altman plot of the differences between the builder’s and medical inclinometer for A. Shoulder internal rotation and B. Shoulder external rotation.

Figure 4.

The solid black line is the mean difference between devices and the segmented lines demonstrate the 90% limits of agreement.

DISCUSSION

Accurate, reliable, and efficient measurements are important to detect motion deficits and changes over time. The SG is readily available in most clinics, has standardized methodology, and low cost. Use of the SG is taught in many medical professions relying on joint ROM assessment to determine baselines and the impact of interventions. However, the time efficiency and comparison of the SG to multiple types of inclinometers has not previously been fully described.

In this study, the inclinometers were more than twice as fast as the SG for measuring range of motion, equating to 3.2 to 6.0 seconds per measurement. Consider a clinician performing pre/post treatment ROM measurements for two motions, for example shoulder internal and external rotation, this could mean saving half a minute of time. When considering the course of a full-time work week, the time required equates to over 30 minutes that could be spent on patient management, clinical documentation, or limiting the need for overtime. The saving of time should entice both clinicians and clinic managers to incorporate inclinometers into clinical practice. While there was a statistically significant difference between the BI and MI for measuring pronation and supination, the nearly half second difference is not clinically relevant, such that, the adoption of either inclinometer device would improve efficiency.

Intrarater reliability for each device was good to excellent with the highest reliability demonstrated with the MI (average ICC value of 0.914), followed closely by the BI (average ICC value of 0.908), and the lowest ICC found with the SG (average ICC value of 0.809). In this study, the SG was the least reliable assessment tool; a result consistent with other studies that found inclinometry to be more reliable than goniometry.25,26 Nonetheless, the SG is the most used clinical tool to assess ROM. Cools et al.27 found high SG intrarater reliability for shoulder rotation (ICC ranging from 0.850 to 0.990). However, for that study, two individuals performed the measurements: one providing stabilization and one performing the measurement. In clinical practice, it is rare to have the assistance of another individual when performing ROM assessment.

Measurement error in the current study was low, with an MDC from 1.8° to 2.8 for the SG and from 1.7° to 2.0° for the inclinometers. Cools et al.27 reported a greater MDC for shoulder rotation, ranging from 4.4° to 8.0° for the SG and 4.0° to 6.4° for the inclinometer. Likewise, both the SEM and MDC for shoulder rotation with the SG and inclinometers in the current study were lower than those found by Correll et al.13 when using a SG and the Halo® device.

As expected, measurements taken with the BI and MI were more highly correlated with each other than with the SG. The average ICC value for BI and MI was 0.703 indicating good, but not excellent, consistency between devices. One reason for the small differences between the BI and MI could be due to the different device contact surfaces. The flat contact surface of the BI may allow for unintentional tilting of the device and thus align differently with body contours compared with the central rounded cutout of the MI. While the correlation between inclinometers was lower than expected, the mean differences between devices were quite small (0° to 4°). Clinically, these differences are unlikely to be relevant. However, given individual measurements varied by as much as 5°, it is recommended that clinicians use the same type of inclinometer when measuring a patient to ensure a change, or lack of change, has occurred. These suggestions are consistent with recommendations of other studies.3,28 Given the low cost and high reliability of the BI, clinicians and clinic managers may consider use of a BI rather than investing in the more expensive medical inclinometer.

Strengths

This study included participants of a broad range of ages without regard to prior injury or surgery, theoretically resulting in a wider range of scores than if only young, healthy individuals were tested. Having more variability in ROM is expected to prevent a misleadingly low reliability coefficient due to low score variability.22 This study utilized one tester who was blinded to measurement results, reducing the potential for variations in passive force application to achieve end ROM.

Limitations

This study included participants who were symptom free. Therefore, it is unclear if the results might apply to individuals who experience pain with testing. Since this study utilized only one tester, it is not possible to determine interrater reliability. Additionally, the tester was an experienced clinician and results may not generalize to an inexperienced clinician. The BI used in this study is typical of builder’s digital inclinometers, in that the device cannot record measurements greater than 90°. Angles greater than 90° required the recorder to calculate the true measurement which could have been a source of error. It is not known if other types of BI would perform similarly to the one used in this study.

CONCLUSION

The BI and MI were found to be reliable for measuring pronation, supination, and shoulder rotation. The low correlation between the SG and inclinometers indicates that clinicians should utilize the same device for testing. Because time can be a barrier to clinician measurement and remeasurement during an episode of care, the greater efficiency and reliability of inclinometers warrants consideration as the new measurement standard. In contrast to the SG, inclinometer assessment requires only one hand to manage the device. The MI cost may be prohibitive for routine clinical use. However, the lower BI cost makes the device clinically affordable. Future studies should examine the reliability of a BI capable of displaying angular measurements greater than 90°. Assessment of a greater number of joint motions using the BI and on various patient populations would be beneficial. Given the current lack of standardization, creating and utilizing a standard for patient and inclinometer positioning is recommended to enhance reliability.

Disclosures

The authors report no conflicts of interest.

References

  1. Reliability of goniometric measurements and visual estimates of knee range of motion obtained in a clinical setting. Watkins Michael A, Riddle Dan L, Lamb Robert L, Personius Walter J. Feb 1;1991 Physical Therapy. 71(2):90–96. doi: 10.1093/ptj/71.2.90. doi: 10.1093/ptj/71.2.90. [DOI] [PubMed] [Google Scholar]
  2. Inter-rater reliability for measurement of passive physiological range of motion of upper extremity joints is better if instruments are used: a systematic review. van de Pol Rachel J., van Trijffel Emiel, Lucas Cees. 2010Journal of Physiotherapy. 56(1):7–17. doi: 10.1016/s1836-9553(10)70049-7. doi: 10.1016/s1836-9553(10)70049-7. [DOI] [PubMed] [Google Scholar]
  3. Concurrent validity of digital inclinometer and universal goniometer in assessing passive hip mobility in healthy subjects. Roach S., San Juan J. G., Suprak D. N., Lyda M. 2013Int J Sports Phys Ther. 8(5):680–8. [PMC free article] [PubMed] [Google Scholar]
  4. Reliability of goniometric measurements in children with cerebral palsy: a comparative analysis of universal goniometer and electronic inclinometer. A pilot study. Herrero Pablo, Carrera Patricia, García Elena, Gómez-Trullén Eva M, Oliván-Blázquez Bárbara. Jul 10;2011 BMC Musculoskeletal Disorders. 12(1):155. doi: 10.1186/1471-2474-12-155. doi: 10.1186/1471-2474-12-155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Reliability of goniometric measurements. Boone Donna C., Azen Stanley P., Lin Chun-Mei, Spence Carol, Baron Carol, Lee Lynn. Nov 1;1978 Physical Therapy. 58(11):1355–1360. doi: 10.1093/ptj/58.11.1355. doi: 10.1093/ptj/58.11.1355. [DOI] [PubMed] [Google Scholar]
  6. Inter-rater reliability for measurement of passive physiological movements in lower extremity joints is generally low: a systematic review. van Trijffel Emiel, van de Pol Rachel J., Oostendorp Rob A.B., Lucas Cees. 2010Journal of Physiotherapy. 56(4):223–235. doi: 10.1016/s1836-9553(10)70005-9. doi: 10.1016/s1836-9553(10)70005-9. [DOI] [PubMed] [Google Scholar]
  7. The reliability and validity of wearable inertial sensors coupled with the Microsoft Kinect to measure shoulder range-of-motion. Beshara Peter, Chen Judy F., Read Andrew C., Lagadec Pierre, Wang Tian, Walsh William Robert. Dec 17;2020 Sensors. 20(24):7238. doi: 10.3390/s20247238. doi: 10.3390/s20247238. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Reliability of inclinometer and goniometric measurements of hip extension flexibility using the modified Thomas test. Clapis Phyllis A, Davis Susan Mercik, Davis Ross Otto. Jan;2008 Physiotherapy Theory and Practice. 24(2):135–141. doi: 10.1080/09593980701378256. doi: 10.1080/09593980701378256. [DOI] [PubMed] [Google Scholar]
  9. Reliability of intra- and inter-rater palpation discrepancy and estimation of its effects on joint angle measurements. Moriguchi Cristiane Shinohara, Carnaz Letícia, Silva Luciana Cristina Cunha Bueno, Salasar Luis Ernesto Bueno, Carregaro Rodrigo Luiz, Sato Tatiana de Oliveira, Coury Helenice Jane Cote Gil. Jun;2009 Manual Therapy. 14(3):299–305. doi: 10.1016/j.math.2008.04.002. doi: 10.1016/j.math.2008.04.002. [DOI] [PubMed] [Google Scholar]
  10. Reliability of inclinometer-derived passive range of motion measures in youth with cerebral palsy. Maltais Désirée B., Ferland Chantal, Perron Marc, Roy Jean-Sébastien. May 30;2019 Physical & Occupational Therapy In Pediatrics. 39(6):655–668. doi: 10.1080/01942638.2019.1597822. doi: 10.1080/01942638.2019.1597822. [DOI] [PubMed] [Google Scholar]
  11. Clinical measurement of range of motion. Review of goniometry emphasizing reliability and validity. Gajdosik Richard L., Bohannon Richard W. Dec 1;1987 Physical Therapy. 67(12):1867–1872. doi: 10.1093/ptj/67.12.1867. doi: 10.1093/ptj/67.12.1867. [DOI] [PubMed] [Google Scholar]
  12. Halo Medical Devices Halo 2 in 1 Digital Goniometer + Incliniometer. [2022-12-11]. https://a.co/d/btEB0rY
  13. Reliability and validity of the Halo digital goniometer for shoulder range of motion in healthy subjects. Correll Sarah, Field Jennifer, Hutchinson Heather, Mickevicius Gabby, Fitzsimmons Amber, Smoot Betty. Aug;2018 International Journal of Sports Physical Therapy. 13(4):707–714. doi: 10.26603/ijspt20180707. doi: 10.26603/ijspt20180707. [DOI] [PMC free article] [PubMed] [Google Scholar]
  14. Validation of a novel smartphone accelerometer-based knee goniometer. Ockendon Matthew, Gilbert Robin E. May 3;2012 Journal of Knee Surgery. 25(4):341–346. doi: 10.1055/s-0031-1299669. doi: 10.1055/s-0031-1299669. [DOI] [PubMed] [Google Scholar]
  15. Smartphone and universal goniometer for measurement of elbow joint motions: a comparative study. Behnoush Behnam, Tavakoli Nasim, Bazmi Elham, Nateghi Fard Fariborz, Pourgharib Shahi Mohammad Hossein, Okazi Arash, Mokhtari Tahmineh. Jun 11;2016 Asian journal of sports medicine. 7(2):e30668–e30668. doi: 10.5812/asjsm.30668. doi: 10.5812/asjsm.30668. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Within-day reliability of shoulder range of motion measurement with a smartphone. Shin Seung Han, Ro Du Hyun, Lee O.-Sung, Oh Joo Han, Kim Sae Hoon. Aug;2012 Manual Therapy. 17(4):298–304. doi: 10.1016/j.math.2012.02.010. doi: 10.1016/j.math.2012.02.010. [DOI] [PubMed] [Google Scholar]
  17. The smartphone inclinometer: A new tool to determine elbow range of motion? Vauclair Frédéric, Aljurayyan Abdulaziz, Abduljabbar Fahad H., Barimani Bardia, Goetti Patrick, Houghton Fiona, Harvey Edward J., Rouleau Dominique M. 2018European Journal of Orthopaedic Surgery & Traumatology. 28(3):415–421. doi: 10.1007/s00590-017-2058-x. doi: 10.1007/s00590-017-2058-x. [DOI] [PubMed] [Google Scholar]
  18. Reliability and validity varies among smartphone apps for range of motion measurements of the lower extremity: a systematic review. Hahn Sarah, Kröger Inga, Willwacher Steffen, Augat Peter. Nov 15;2021 Biomedical Engineering / Biomedizinische Technik. 66(6):537–555. doi: 10.1515/bmt-2021-0015. doi: 10.1515/bmt-2021-0015. [DOI] [PubMed] [Google Scholar]
  19. Validity of three lumbar sagittal motion measurement methods: surface inclinometers compared with radiographs. Samo Daniel G., Chen Shu-Pi C., Crampton Arthur R., Chen Edwin H., Conrad Karen M., Egan Linda, Mitton Joseph. Mar;1997 Journal of Occupational & Environmental Medicine. 39(3):209–216. doi: 10.1097/00043764-199703000-00010. doi: 10.1097/00043764-199703000-00010. [DOI] [PubMed] [Google Scholar]
  20. Norkin CC, White DJ. Measurement of Joint Motion: A Guide to Goniometry. FA Davis; Phildelphia, PA: [Google Scholar]
  21. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. Koo Terry K., Li Mae Y. Jun;2016 Journal of Chiropractic Medicine. 15(2):155–163. doi: 10.1016/j.jcm.2016.02.012. doi: 10.1016/j.jcm.2016.02.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Portney Leslie G, Watkins Mary P. In: Foundations of Clinical Research: Applications to Practice. Portney Leslie G, Watkins Mary P., editors. Pearson; Upper Sabble River, NJ: Statistical measures of reliability. [Google Scholar]
  23. Intraclass correlation – a discussion and demonstration of basic features. Liljequist David, Elfving Britt, Skavberg Roaldsen Kirsti. Jul 22;2019 PLoS One. 14(7):e0219854. doi: 10.1371/journal.pone.0219854. doi: 10.1371/journal.pone.0219854. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Understanding Bland Altman analysis. Giavarina Davide. 2015Biochemia Medica. 25(2):141–151. doi: 10.11613/bm.2015.015. doi: 10.11613/bm.2015.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. The reliability and concurrent validity of shoulder mobility measurements using a digital inclinometer and goniometer: A technical report. Kolber Morey J., Hanney William J. 2012International Journal of Sports Physical Therapy. 7(3):306–313. [PMC free article] [PubMed] [Google Scholar]
  26. Validity and reliability of a digital inclinometer to assess knee joint position sense in an open kinetic chain. Romero-Franco Natalia, Montaño-Munuera Juan Antonio, Fernández-Domínguez Juan Carlos, Jiménez-Reyes Pedro. May 1;2019 Journal of Sport Rehabilitation. 28(4):332–338. doi: 10.1123/jsr.2017-0221. doi: 10.1123/jsr.2017-0221. [DOI] [PubMed] [Google Scholar]
  27. Measuring shoulder external and internal rotation strength and range of motion: comprehensive intra-rater and inter-rater reliability study of several testing protocols. Cools Ann M., De Wilde Lieven, Van Tongel Alexander, Ceyssens Charlotte, Ryckewaert Robin, Cambier Dirk C. Oct;2014 Journal of Shoulder and Elbow Surgery. 23(10):1454–1461. doi: 10.1016/j.jse.2014.01.006. doi: 10.1016/j.jse.2014.01.006. [DOI] [PubMed] [Google Scholar]
  28. Reliability of shoulder range of motion comparing a goniometer to a digital level. Mullaney Michael J, McHugh Malachy P, Johnson Christopher P, Tyler Timothy F. Jun 17;2010 Physiotherapy Theory and Practice. 26(5):327–333. doi: 10.3109/09593980903094230. doi: 10.3109/09593980903094230. [DOI] [PubMed] [Google Scholar]

Articles from International Journal of Sports Physical Therapy are provided here courtesy of North American Sports Medicine Institute

RESOURCES