Introduction
Implantable cardioverter defibrillator (ICD) implantation remains the mainstay of sudden cardiac death (SCD) prevention in high-risk patients. Among other complications of ICD therapy, inappropriate shocks negatively impact quality of life and may be associated with myocardial injury.1 In the context of the recent development of non-transvenous ICD therapies, current real-world data may help decision-making in this vulnerable patient cohort.
Methods
Study design
The Clinical and Device Functional Assessment of Real World ICD Patients (CARAT) trial is a prospective, multi-centre, international, observational, post-market study of all approved and commercialized CE and/or Food and Drug Administration ICD devices [VR/DR/cardiac resynchronisation therapy defibrillator (CRT-D)] from MicroPort CRM® (Clamart, France), registered at ClinicalTrials.gov (NCT02341768). The subjects provided informed consent during the index hospital stay of implantation in accordance with ISO 14155 and local regulations and were then followed-up over 2 years; a centralized and independent review of all ICD therapies were performed to adjudicate appropriateness.
Statistical analysis
Mean and standard deviation were calculated for continuous variables. For time-to-event variables, the Kaplan–Meier cumulative incidence with death as a competing risk method was used. The estimates were provided with time points every 2 months. Subjects who had prematurely discontinued without any event were censored at the date of discontinuation.
Results
Patient characteristics
From 1 July 2015, to 31 October 2017, a total of 2032 patients, either as a de novo implantation (n = 1463, 72.1%) or replacement (n = 565, 27.9%), were enrolled at 94 sites from 12 countries in Europe and North-America. As four patients were not implanted, 2028 patients (1601 males, 427 females) were analysed. Baseline characteristics are given in Table 1. Final follow-up visit after 24 months was available in 73.8% (n = 1496) study participants. Apart from death (9.4%, n = 169), reasons for premature study discontinuation were lost to follow-up in 5.0% (n = 102), withdrawal by the patient in 2.8% (n = 57), completion of the study without final visit in 2.2% (n = 45), non-fatal adverse events in 1.0% (n = 20), and other reasons in 6.7% (n = 136).
Table 1.
Baseline data
Population characteristics (n = 2028) | |
---|---|
Age | 67 years (IQR: 59–75) |
Females | 21.1% |
Primary prevention of SCD | 77.0% |
LVEF | 30% (IQR: 25–35) |
Ischaemic/idiopathic dilated/other CMP | 51%/29%/20% |
NYHA class I/II/III/IV | 15.9%/51.7%/31.0%/1.4% |
Advanced renal failure/diabetes | 15%/25.4% |
Chronic respiratory disease | 12.5% |
History of stroke | 7.5% |
HF hospitalization during last 6 months | 15.8% |
Paroxysmal/persistent/permanent AF | 11.4%/5%/9.2% |
Atrial flutter and/or atrial tachycardia | 3.5% |
AVB I/II/III | 7.2%/1.2%/2.2% |
SND | 3.5% |
LBBB/RBBB/IVCD | 23.5%/6.2%/2.0% |
Previous VT/SVT ablation | 3%/1.4% |
IQR, interquartile range; SCD, sudden cardiac death; VT, ventricular tachycardia; LBBB, left bundle branch abnormality; AF, atrial fibrillation; LVEF, left ventricular ejection fraction; CMP, cardiomyopathy; AVB, atrioventricular block; SND, sinus node disease; RBBB, right bundle branch block; IVCD, intraventricular conduction delay; SVT, supraventricular tachycardia; HF, heart failure; NYHA, New York Heart Association.
Tachyarrhythmia detection and therapy programming
Detection programming was very similar across CRT-D, dual chamber, and single chamber ICD models. A slow ventricular tachycardia (VT) zone was enabled in more than 50% of study participants from as low as 150 beats per minute (bpm), with the vast majority being programmed as a monitoring zone without therapy. A VT zone was enabled from 170 bpm on average, with AntiTachycardia Pacing 1 (ATP1) burst/AntiTachycardia Pacing 2 (ATP2) ramp and shocks in the vast majority of patients. A fast VT zone favouring antitachycardia pacing (ATP) delivery was activated in more than 80% of patients allowing rhythm detection until 240 bpm. There were only few and non-meaningful differences in tachyarrhythmia detection programming across countries or between patients receiving ICD therapy because of primary or secondary prevention of SCD. A so-called ‘shock box’ programming (without any ATP therapy, ventricular fibrillation (VF) zone between 198 and 206 bpm) was programmed in only 37 study participants (1.8%), mainly in patients with the ICD implanted for primary prevention of SCD.
Appropriate and inappropriate ICD therapies
Within 2 years, 13.4% (n = 250) and 3.8% (n = 69) received appropriate and inappropriate ICD therapies, respectively. Appropriate and inappropriate ICD shocks occurred in 5.0% (n = 95) and 1.6% (n = 29) of patients, respectively, as depicted in Figure 1.
Figure 1.
(A) Enrolment per country, ICD model, and type of SCD prevention. (B) Cumulative incidence of inappropriate shocks per type of ICD model. (C) Cumulative incidence of appropriate shocks per type of ICD model. ICD, implantable cardioverter defibrillator; SCD, sudden cardiac death.
Discussion
The large prospective international dataset of the CARAT study reports a remarkably low rate of inappropriate shocks in real-world contemporary ICD therapy (1.6% within 2 years). This rate slightly varies across ICD models; however, the confidence intervals overlap. Avoiding inappropriate shocks is a key component of quality of life and of ICD acceptance by patients.2
As the PARAD+® algorithm had been used as default tachyarrhythmia detection algorithm in the presence of a functional atrial lead, CARAT confirms results of previous studies such as OPTION3,4 or ISIS-ICD5 in terms of arrhythmia discrimination in dual-chamber devices. The present study also reports low inappropriate shocks in single-chamber devices, where arrhythmia discrimination is based on a combination of stability, acceleration, and ‘long cycle gap’ to identify atrial fibrillation.
In the PRAETORIAN trial with a median follow up of 49.1 months, rates of inappropriate ICD shocks were remarkably higher, both in patients implanted with a subcutaneous ICD (9.7%) and with a transvenous ICD (7.3%).6,7 The PIVOTAL trial testing the new extravascular ICD approach reported that even 9.7% of patients implanted with a substernal ICD lead received inappropriate shocks during a mean follow-up of 10.6 months.8 In transvenous ICD trials, inappropriate shocks were more common in the EU-CERT trial (7% within 2.7 years),9 the UMBRELLA study (5.0% at 2 years),10 and—to a much lesser degree—in the PainFree SST sub-study (2.8% in patients with a dual chamber ICD or CRT-D device and 3.7% in patients with a single chamber ICD within 2 years).11 The rate of inappropriate shocks in the VR patients in the present study (1.9% at 2 years) also compares favourably with the annual rate of 6.4% reported in a systematic review.12
Appropriate ICD therapy was delivered in 13.4% of the study cohort, whereas appropriate shocks were mandated in 5.0% of patients at 2 years. Consistently with previous studies,5 a lower rate of appropriate shocks is found in the CRT-D arm. The rate of appropriate ICD shocks in CARAT was lower than in recently published trials such as the EU-CERT trial (7.1% within 2.7 years)9 and the PRAETORIAN trial (19.5% in the subcutaneous ICD subgroup and 13.5% in the transvenous ICD subgroup, median follow-up 49.1 months)6,7 and comparable to the DANISH trial (11.5% within 5.5 years),13 potentially associated with a reduced proportion of ischaemic heart failure aetiology in the present study (51%).
The CARAT study further offers the opportunity to confirm the adoption of recommendations for ICD programming in clinical routine, according to the expert consensus on optimal ICD programming and testing.14,15 Prolonged detection settings for tachyarrhythmias with duration criteria of at least 6–12 s or 30 intervals were on average programmed shorter in CARAT patients, namely 20–24 cycles in the VT and 10 cycles in the VF zone, respectively; therefore, discrimination performance was obtained without compromise on the time to therapy delivery. Detection zones and therapy programming were globally well aligned with consensus, allowing to fully benefit from ICD tachyarrhythmia detection and therapy capabilities.
Contributor Information
Wolfgang Dichtl, University Hospital for Internal Medicine III (Cardiology and Angiology), Medical University Innsbruck, Anichstraße 35, A-6020 Innsbruck, Austria.
João De Sousa, Centro Hospitalar Lisboa Norte—Hospital de Santa Maria, Lisbon, Portugal.
Jose Manuel Rubin Lopez, H. Universitario Central de Asturias, Oviedo, Spain.
Enrique Garcia Campo, Complejo Hospitalario Universitario de Vigo, Vigo, Spain.
Klaus-Jürgen Gutleben, Herz- und Diabeteszentrum Bad Oeynhausen, Bad Oeynhausen, Germany.
Yann Poezevara, Clinical Affairs, MicroPort CRM, Paris, France.
Vincent Probst, CHU de Nantes—Hôpital Nord Laënnec, Nantes, France.
Funding
The CARAT trial was funded by Microport CRM® (Clamart, France).
Data availability
All data are available on a reasonable request to the sponsor of the trial.
References
- 1. Brewster J, Sexton T, Dhaliwal G, Charnigo R, Morales G, Parrott Ket al. Acute effects of implantable cardioverter-defibrillator shocks on biomarkers of myocardial injury, apoptosis, heart failure, and systemic inflammation. Pacing Clin Electrophysiol 2017;40:344–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Januszkiewicz Ł, Barra S, Providencia R, Conte G, De Asmundis C, Chun JKRet al. Long-term quality of life and acceptance of implantable cardioverter-defibrillator therapy: results of the European Heart Rhythm Association survey. Europace 2022;24:860–7. [DOI] [PubMed] [Google Scholar]
- 3. Kolb C, Sturmer M, Babuty D, Sick P, Davy JM, Molon Get al. Relation between detection rate and inappropriate shocks in single versus dual chamber cardioverter-defibrillator—an analysis from the OPTION trial. Sci Rep 2016;6:21748. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Kolb C, Sturmer M, Sick P, Reif S, Davy JM, Molon Get al. Reduced risk for inappropriate implantable cardioverter-defibrillator shocks with dual-chamber therapy compared with single-chamber therapy: results of the randomized OPTION study. JACC Heart Fail 2014;2:611–9. [DOI] [PubMed] [Google Scholar]
- 5. Ruiz-Granell R, Dovellini EV, Dompnier A, Khalighi K, García-Campo E, Olivier Aet al. Algorithm-based reduction of inappropriate defibrillator shock: results of the inappropriate shock reduction with PARAD+ rhythm discrimination-implantable cardioverter defibrillator study. Heart Rhythm 2019;16:1429–35. [DOI] [PubMed] [Google Scholar]
- 6. Knops RE, Olde Nordkamp LRA, Delnoy PHM, Boersma LVA, Kuschyk J, El-Chami MFet al. Subcutaneous or transvenous defibrillator therapy. N Engl J Med 2020;383:526–36. [DOI] [PubMed] [Google Scholar]
- 7. Knops RE, van der Stuijt W, Delnoy PPHM, Boersma LVA, Kuschyk J, El-Chami MFet al. Efficacy and safety of appropriate shocks and antitachycardia pacing in transvenous and subcutaneous implantable defibrillators: analysis of all appropriate therapy in the PRAETORIAN trial. Circulation 2022;145:321–9. [DOI] [PubMed] [Google Scholar]
- 8. Friedman P, Murgatroyd F, Boersma LVA, Manlucu J, O'Donnell D, Knight BPet al. ; Extravascular ICD Pivotal Study Investigators . . Efficacy and safety of an extravascular implantable cardioverter-defibrillator. N Engl J Med 2022;387:1292–302. [DOI] [PubMed] [Google Scholar]
- 9. Zabel M, Willems R, Lubinski A, Bauer A, Brugada J, Conen Det al. Clinical effectiveness of primary prevention implantable cardioverter-defibrillators: results of the EU-CERT-ICD controlled multicentre cohort study. Eur Heart J 2020;41:3437–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Fontenla A, Martínez-Ferrer JB, Alzueta J, Viñolas X, García-Alberola A, Brugada Jet al. Incidence of arrhythmias in a large cohort of patients with current implantable cardioverter-defibrillators in Spain: results from the UMBRELLA registry. Europace 2016;18:1726–34. [DOI] [PubMed] [Google Scholar]
- 11. Auricchio A, Schloss EJ, Kurita T, Meijer A, Gerritse B, Zweibel Set al. Low inappropriate shock rates in patients with single- and dual/triple-chamber implantable cardioverter-defibrillators using a novel suite of detection algorithms: PainFree SST trial primary results. Heart Rhythm 2015;12:926–36. [DOI] [PubMed] [Google Scholar]
- 12. Auricchio A, Hudnall JH, Schloss EJ, Sterns LD, Kurita T, Meijer Aet al. Inappropriate shocks in single-chamber and subcutaneous implantable cardioverter-defibrillators: a systematic review and meta-analysis. Europace 2017;19:1973–80. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Køber L, Thune JJ, Nielsen JC, Haarbo J, Videbæk L, Korup Eet al. Defibrillator implantation in patients with nonischemic systolic heart failure. N Engl J Med 2016;375:1221–30. [DOI] [PubMed] [Google Scholar]
- 14. Wilkoff BL, Fauchier L, Stiles MK, Morillo CA, Al-Khatib SM, Almendral Jet al. 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Europace 2016;18:159–83. [DOI] [PubMed] [Google Scholar]
- 15. Stiles MK, Fauchier L, Morillo CA, Wilkoff BL; ESC Scientific Document Group . 2019 HRS/EHRA/APHRS/LAHRS focused update to 2015 expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Europace 2019;21:1442–3. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data are available on a reasonable request to the sponsor of the trial.