Abstract
Proton therapy requires accurate dose calculation for treatment planning to ensure the conformal doses are precisely delivered to the targets. The conversion of CT numbers to material properties is a significant source of uncertainty for dose calculation. The aim of this study is to develop a physics-informed deep learning (PIDL) framework to derive accurate mass density and relative stopping power (RSP) maps from dual-energy computed tomography (DECT) images. The PIDL framework allows deep learning (DL) models to be trained with a physics loss function, which includes a physics model to constrain DL models. Five DL models were implemented including a fully connected neural network (FCNN), dual-FCNN (DFCNN), and three variants of residual networks (ResNet): ResNet-v1 (RN-v1), ResNet-v2 (RN-v2), and dual-ResNet-v2 (DRN-v2). An artificial neural network (ANN) and the five DL models trained with and without physics loss were explored to evaluate the PIDL framework. Two empirical DECT models were implemented to compare with the PIDL method. DL training data were from CIRS electron density phantom 062M (Computerized Imaging Reference Systems, Inc., Norfolk, VA). The performance of DL models was tested by CIRS adult male, adult female, and 5-year-old child anthropomorphic phantoms. For density map inference, the physics-informed RN-v2 was 3.3%, 2.9% and 1.9% more accurate than ANN for the adult male, adult female, and child phantoms. The physics-informed DRN-v2 was 0.7%, 0.6%, and 0.8% more accurate than DRN-v2 without physics training for the three phantoms, respectfully. The results indicated that physics-informed training could reduce uncertainty when ANN/DL models without physics training were insufficient to capture data structures or derived significant errors. DL models could also achieve better image noise control compared to the empirical DECT parametric mapping methods. The proposed PIDL framework can potentially improve proton range uncertainty by offering accurate material properties conversion from DECT.
1. Introduction
Proton therapy can deliver conformal doses to tumor targets and spare normal tissues with minimum irradiation to reduce patients’ short- or long-term side effects. Conventional proton treatment planning systems (TPS) utilize either analytical dose calculation (ADC) (Schuemann et al., 2015; Yepes et al., 2018; Liang et al., 2019) or Monte Carlo dose calculation (MCDC) algorithms (Paganetti et al., 2008; Saini et al., 2017; Chang et al., 2020) to design treatment plans. These dose calculation algorithms require information on the relative stopping power (RSP) or mass density (MD) of patients. This information is typically acquired from single-energy computed tomography (SECT) applying CT-number-to-RSP or CT-number-to-mass-density conversion curves, calibrated using the stoichiometric method (Schneider et al., 1996; Schneider et al., 2000). However, this method relies on linear fitting. The model performance can be compromised by the data quality (Dinh, 2013) such as data quantity and uncertainty of calibration materials. Besides, the linear model form limits the model performance even if extensive data are used (Ratner, 2011). Furthermore, materials with different compositions may have the same CT number from SECT, and the conventional calibration curves cannot settle this ambiguity problem. Hence, a margin of 3.5% is typically reserved for proton range uncertainty (Paganetti, 2012).
Dual-energy computed tomography (DECT) has been deployed to characterize human tissues for diagnosis (McCollough et al., 2015), and DECT's virtual monochromatic images (VMI) can improve beam-hardening artifacts and increase image quality (Yu et al., 2012). DECT can be also used to estimate effective atomic number () and relative electron density to that of water (), and these parametric maps can be used to determine proton RSP (Bär et al., 2017) for ADC (Wohlfahrt et al., 2017). In general, DECT can be utilized to infer more exact proton RSP than SECT (Yang et al., 2012). However, the DECT-based stoichiometric method (Bourque et al., 2014) used a fixed form of the fitting model, and it is not trivial to simultaneously include additional DECT’s images such as VMI with multiple energies, , and in an existed calibration model. To develop a new model to adapt additional DECT images, it generally requires extensive efforts to gain the insights and mechanistic understanding. As an alternative to mechanistic and semi-analytical models, it is viable to disclose the underlying correlations between DECT parametric maps and material mass density and RSP by modern machine learning (ML) methods.
Model inference of many traditional ML algorithms tend to follow Occam’s razor: the simplest model is preferable due to its interpretability and generalizability (Blumer et al., 1987; Domingos, 1999). However, the approach is limited by data complexity (Champion et al., 2019). On the contrary, deep learning (DL) (LeCun et al., 2015) can deploy hierarchical model layers to uncover complex structures from data. Hornik et al. (Hornik et al., 1989) proved that neural networks with multilayer perceptrons are universal approximators that can capture the properties of any measurable information. Su et al. (Su et al., 2018) used electron density phantoms to demonstrate that ML models achieved DECT parametric mapping for RSP with the uncertainty of ~9% and ~3% for lung and bone.
Physics-informed machine learning (Chang and Dinh, 2019; Karniadakis et al., 2021) has exhibited the advantage of integrating noisy data, insufficient data, and physics models during DL training. In the present work, we couple a physics model in the training of DL models and formulate a physics-informed DL framework to learn the human tissue surrogates from electron density phantoms and predict mass density and RSP maps for anthropomorphic phantoms with different body sizes and material compositions. The framework can adapt various DL models with different degrees of freedom. To evaluate the framework's performance for clinical applications, we investigate conditions by which DL models can benefit from the physics-informed training to work compatibly, stably, and effectively for DECT parametric mapping.
2. Materials and methods
2.1. Data acquisition and phantoms
The images of a CIRS electron density 062M phantom (Computerized Imaging Reference Systems, Inc., Norfolk, VA, USA) scanned with DECT was used to train DL models to establish correlations between DECT images and material properties. Table 1 gives the physical properties of CIRS 062M, which includes 17 inserts from 9 types of tissue surrogate. Each tissue surrogate has redundant inserts except for the bone 1250 mg/cc. A Siemens SOMATOM Definition Edge scanner was used to scan phantoms, and the CT acquisition and reconstruction parameters for all scans includes the tube voltage of 120 kVp, single collimation width of 0.6 mm, reconstruction method and kernel of sinogram-affirmed iterative reconstruction (SAFIRE) and Qr40 with iterative beam hardening correction using Siemens Syngo.CT VB20A. The phantom was scanned using TwinBeam dual-energy (TBDE) protocols with a CT dose index (CTDIvol, 32cm) of 11.57 mGy and effective milliampere-seconds () of 541.
Table 1.
Tissue surrogate | RSP | H | C | N | O | Mg | P | S | Cl | Ca | Ba | |||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CIRS062M | Lung (Inhale) | 0.203 | 0.202 | 67.8 | 8.59 | 65.92 | 3.52 | 19.27 | 1.69 | 1.01 | ||||
Lung (Exhale) | 0.494 | 0.492 | 67.3 | 8.87 | 66.00 | 2.38 | 20.41 | 0.61 | 1.73 | |||||
Adipose | 0.965 | 0.977 | 64.2 | 9.97 | 71.36 | 1.79 | 16.37 | 0.19 | 0.32 | |||||
BreastTissue | 0.996 | 1.003 | 65.2 | 9.60 | 70.33 | 1.92 | 17.02 | 0.19 | 0.94 | |||||
Muscle | 1.059 | 1.059 | 66.6 | 9.07 | 69.78 | 2.07 | 16.80 | 0.12 | 2.16 | |||||
Liver | 1.072 | 1.070 | 66.8 | 8.97 | 69.45 | 2.12 | 17.14 | 0.13 | 2.19 | |||||
Bone 200 mg/cc | 1.157 | 1.116 | 76.2 | 7.00 | 56.30 | 2.00 | 22.70 | 3.30 | 0.20 | 8.50 | ||||
Bone 800 mg/cc | 1.520 | 1.404 | 90.4 | 4.45 | 39.11 | 0.87 | 33.72 | 0.05 | 21.77 | 0.03 | ||||
Bone 1250 mg/cc | 1.830 | 1.647 | 101. 5 | 3.60 | 28.82 | 1.09 | 31.99 | 10.79 | 0.08 | 0.04 | 23.27 | 0.32 | ||
| ||||||||||||||
CIRS Atom | Lung | 0.205 | 0.204 | 67.8 | 8.59 | 65.89 | 3.52 | 19.29 | 1.69 | 1.01 | ||||
Breast | 0.991 | 0.982 | 70.4 | 9.60 | 70.26 | 1.93 | 17.00 | 0.20 | 9.40 | |||||
Soft Tissue | 1.055 | 1.041 | 70.4 | 8.47 | 57.44 | 1.65 | 24.59 | 7.62 | 0.19 | |||||
Spinal Cord | 1.065 | 1.035 | 73.4 | 7.36 | 54.27 | 2.17 | 26.59 | 9.37 | 0.22 | |||||
Brain | 1.069 | 1.049 | 72.2 | 8.16 | 53.60 | 1.53 | 26.49 | 9.98 | 0.19 | |||||
(M701/M702) | 1.586 | 1.410 | 88.3 | 4.83 | 37.03 | 0.97 | 35.66 | 6.19 | 0.05 | 15.24 | ||||
Bone (M705) | 1.518 | 1.422 | 85.5 | 5.23 | 41.63 | 1.11 | 33.31 | 3.54 | 0.05 | 15.09 |
The RSP of each tissue surrogate is calculated by Eq. (1) (Bichsel, 1969) where , , , , and are the relative electron density, electron mass, speed of light, proton velocity relative to the speed of light, and mean ionization potential of phantom media. The and in Eq. (1) are obtained using proton energy of 210 MeV and mean ionization potential of 78 eV (Seltzer S M, 2014).
(1) |
Two DECT scans were acquired with different insert arrangements, and each scan contained voxels. Figure 1 depicts the setup of DECT scans. During the scans, the electron density phantom was placed between solid water phantoms with a thickness of 10 cm to avoid CT artifacts.
The CIRS ATOM M701 (adult male), M702 (adult female), and M705 (5-year-old child) anthropomorphic phantoms (Computerized Imaging Reference Systems, Inc., Norfolk, VA) were used to validate DL results. Separate CT scans were made for the head-and-neck (HN), thorax, and pelvis regions of the phantoms, applying body-site specific TBDE CT protocols for each region. For a head-and-neck (HN) TBDE protocol, scans of M701, M702, and M705 contain , , and voxels. For M701 and M702, thorax TBDE scans include and voxels while pelvis scans involve and voxels, respectively. Table 2 shows the CTDIvol, 32cm and effective milliampere-seconds for different TBDE protocols at each site. For the 5-year-old child phantom, only a TBDE HN protocol was used. The material properties of the anthropomorphic phantoms are provided in Table 1. The tissue-mimicking materials are almost the same for each phantom except for bone. All DECT images were reconstructed using the same reconstruction setting with a diameter of 500 mm and slice thickness of 0.5 mm. Siemens Syngo.Via was used to generate DECT parametric maps such as relative electron density (), effective atomic number (), and VMI of 80, 135, and 190 keV. Table 3 and Table 4 give the reference and Syngo.Via values of and for the electron density and anthropomorphic phantoms.
Table 2.
CTDIvol, 32cm (mGy) / effective milliampere-seconds (mAseff) |
||||
---|---|---|---|---|
HN | Thorax | Pelvis | ||
M701 | 15.63 / 729 | 6.99 / 327 | 9.63 / 450 | |
M702 | 15.70 / 732 | 3.97 / 186 | 6.09 / 284 | |
M705 | 11.51 / 537 | - | - |
Table 3.
Reference | Syngo.Via | |||
---|---|---|---|---|
|
||||
Lung (Inhale) | 0.190 | 7.332 | 0.227 ± 0.021 | - |
Lung (Exhale) | 0.488 | 7.382 | 0.504 ± 0.021 | - |
Adipose | 0.949 | 6.370 | 0.965 ± 0.019 | 6.420 ± 0.972 |
BreastTissue | 0.976 | 6.766 | 0.992 ± 0.020 | 6.656 ± 0.815 |
Muscle | 1.042 | 7.383 | 1.035 ± 0.020 | 7.514 ± 0.626 |
Liver | 1.052 | 7.409 | 1.054 ± 0.021 | 7.459 ± 0.655 |
Bone 200 mg/cc | 1.116 | 10.055 | 1.107 ± 0.025 | 9.804 ± 0.442 |
Bone 800 mg/cc | 1.454 | 12.566 | 1.391 ± 0.029 | 12.802 ± 0.277 |
Bone 1250 mg/cc | 1.694 | 13.456 | 1.621 ± 0.036 | 13.652 ± 0.285 |
Table 4.
Reference | Syngo.Via | ||||||||
---|---|---|---|---|---|---|---|---|---|
|
|||||||||
|
|
||||||||
M701 | M702 | M705 | M701 | M702 | M705 | ||||
Lung | 0.200 | 7.332 | 0.207 ± 0.019 | 0.208 ± 0.018 | 0.216 ± 0.015 | - | - | - | |
Breast | 0.976 | 6.769 | - | 0.965 ± 0.022 | - | - | 7.282 ± 0.687 | - | |
Soft Tissue | 1.028 | 7.187 | 1.031 ± 0.029 | 1.031 ± 0.025 | 1.025 ± 0.014 | 7.110 ± 1.044 | 7.165 ± 0.859 | 7.342 ± 0.360 | |
Spinal Cord | 1.027 | 7.426 | 1.062 ± 0.031 | 1.060 ± 0.023 | 1.046 ± 0.013 | 6.691 ± 1.590 | 6.921 ± 1.139 | 7.537 ± 0.428 | |
Brain | 1.039 | 7.438 | 1.053 ± 0.011 | 1.047 ± 0.010 | 1.046 ± 0.008 | 7.408 ± 0.315 | 7.420 ± 0.303 | 7.433 ± 0.219 | |
Bone (M701/M702) | 1.495 | 11.634 | 1.437 ± 0.050 | 1.443 ± 0.042 | - | 11.741 ± 0.742 | 11.651 ± 0.654 | - | |
Bone (M705) | 1.436 | 11.481 | - | - | 1.387 ± 0.028 | - | - | 11.475 ± 0.284 |
2.1.1. RSP measurement for CIRS M701 adult male phantom using protons
We used a 208 MeV proton spot and Zebra (IBA Dosimetry, Germany) to measure the RSP of lung, soft tissue, brain, and bone surrogates for the CIRS M701 adult male phantom since the material compositions of these tissue surrogate are identical for the CIRS anthropomorphic phantoms. The proton spot was delivered by Varian ProBeam System (Varian Medical Systems, Palo Alto) with a cyclotron current of 20 nA and the spot sigma of 3.79 mm at the ISO center. Figure 2 depicts the experiment setup. The beam ISO center was set at the phantom surface. The thickness of each phantom slab is 25 mm. The 80% distal range (R80) differences from the measurement with and without phantom slabs were used to compute the water equivalent thickness (WET). The Zebra measurement includes range uncertainty of . The measured RSP can be obtained using WET divided by the thickness of phantom slabs. Then we derived the reference mass density using Eq. (1) by adjusting material mass densities to match the measured RSP. The measured RSP values for lung, soft tissue, brain, and bone surrogates are 0.201, 1.040, 1.050, and 1.410, and the reference mass densities are 0.202, 1.054, 1.070, and 1.517 g/cm3, respectively.
2.2. Physics-informed deep learning for DECT parametric mapping
Figure 3 portrays the physics-informed deep learning (PIDL) framework to achieve DECT parametric mapping and generate mass density and RSP maps. PyTorch (Paszke, 2019) was used for DL model implementation, and the model inputs include DECT parametric maps of , , VMI of 80, 135, and 190 keV, and DECT images from high/low energy spectra. At the training phase, central 79 CT slices from the two scans (given in Figure 1) of the electron density phantom were used, and we manually contoured circular regions of interest for each insert with a diameter of 24 pixels. Then each insert region from each DECT image was arranged to a 1-dimensional (1D) array such that the training input layer contains the size of 1242828 x 7 (total voxels x input images), and the output layer returns values of mass density and RSP simultaneously. Section 2.2.2. provides details about how the physics model tightly interacts with DL models utilizing physics quantities based on material compositions from a CIRS electron density phantom. This approach allows the model response to be constrained by physics knowledge. The performance of the PIDL framework is assessed by three CIRS anthropomorphic phantoms that include different body sizes and bone material compositions.
2.2.1. Deep learning models
The PIDL framework is not limited to DL models, and we implemented one traditional ML and five DL models in the framework to further explore relations of the predictive capability and model complexity. Figure 3(a)–(f) shows the six ML/DL models and the model complexity is increased from Figure3(a) to Figure 3(f) due to the increase of trainable parameters and model layers. The ML model in this work was artificial neural networks (ANN) and we implemented ANN with 30 hidden units according to the previous work conducted by Su et al. (Su et al., 2018). Figure 3(a) depicts the ANN structure applied in this study. The σ in Figure 3(a)–(c) denotes activation functions of ReLU (Nair and Hinton, 2010). Meanwhile, we implemented five DL models (Fukushima, 1980; He et al., 2016) to demonstrate the predictive capability of the proposed PIDL framework. Figure 3(b)–(f) show that the five DL models are a fully connected neural network (FCNN) (Bebis and Georgiopoulos, 1994), dual-FCNN (DFCNN) (Schmidhuber, 2015), and three variants of residual networks (ResNet) (Wang et al., 2018) including ResNet-v1 (RN-v1), ResNet-v2 (RN-v2), and dual-ResNet-v2 (DRN-v2). For the DL models with fully connected (FC) layers in Figure 3(b)–(c), there are 200 hidden units in each layer with layer normalization (LN) (Jimmy Lei Ba, 2016) to prevent gradient vanishing issues during training (Hochreiter et al., 2001). FCNN includes nine fully connected (FC) layers, while DFCNN has four FC and five dual-stream layers (Simonyan and Zisserman, 2014). The model complexity of DFCNN is increases compared to FCNN whereby DFCNN potentially can discover more structures from data (Naitzat et al., 2020). The detail structures of RN-v1 and RN-v2 are given in Appendix A. Figure 3(e) shows the structure of a ResNet (RN) block (marked with a green dashed line in Figure 3(e)). The utilization of RN blocks allows us to increase the model complexity compared to neural networks without residual implementation while preventing accuracy degradation issues (He et al., 2016). Figure 3(f) depicts the DRN-v2 architecture built based on two RN blocks. In Section 2.2.2, we described details about how to implement the physics model (given in Figure 3(g)) into a loss function for DL training.
2.2.2. Supervised loss function and physics loss
We use supervised DL models to establish correlations between DECT images and material mass densities and RSP. This inverse learning minimizes the loss function given by Eq. (2), which includes mean square error (MSE) loss and physics loss. The is equal to 1 for physics-informed training and the value is 0 for conventional training. The physics-informed deep learning (PIDL) and DL use identical ANN/DL model structures given in Figure 3(a)–(f) but with different loss functions during the model training. Therefore, ANN/DL models trained by physics-informed and conventional approaches will learn differently from data and result in distinct values of model parameters. For the conventional approach, DL models (given in Figure 3) are trained with only MSE loss, and we use abbreviations of ANN, FCNN, DFCNN, RN-v1, RN-v2, and DRN-v2 for the models trained with only MSE loss. For the physics-informed training, each model (given in Figure 3) is trained with both MSE and physics loss, and we use abbreviations of PANN, PFCNN, PDFCNN, PRN-v1, PRN-v2, and PDRN-v2 for the models trained with MSE and physics losses.
(2) |
Eq. (3) defines the MSE loss which is based on mass density and RSP with reference values for the electron density phantom (upper part of Table 1). The , , , , , and denote the mass density predicted by ANN/DL models, ground truth mass density, RSP predicted by ANN/DL models, ground truth RSP, total number of CT voxels, and ith voxel.
(3) |
Eq. (4) defines the physics loss where , , and denote total number of CT voxels, ith voxel, and measured quantities of interest that is CT numbers in this work. The represents physics insights that is empirical HU model in this work given by Eq. (5) (Rutherford et al., 1976; Jackson and Hawkes, 1981), assuming where is the linear attenuation coefficient.
(4) |
Eq. (5) is the physics model given in Figure 3(g), and the model is used to compute the physics loss given by Eq. (4). The , , and are energy-dependent coefficients corresponding to the photoelectric effect, coherent scattering, and incoherent scatting. These coefficients can be obtained by a least-square fit with VMI of 80keV from the scan of the CIRS 062M electron density phantom. The estimated values for , , and in Eq. (5) are , , and .
(5) |
By defining and equal to z3.62 and z1.86, the values can be derived by Eq. (6) (Mayneord, 1937; Spiers, 1946) where , , , and denote ith element, weight fraction, atomic number, and atomic mass.
(6) |
Eq. (7) defines the relative electron density () where is the electron density for water and it is a constant value of e−/cm3. The is the predicted mass density from the output of DL models.
(7) |
For the physics-informed training, the value of is queried during each training iteration, allowing the physics model given by Eq. (5) and in Figure 3(g) to be used to compute the physics loss defined by Eq. (4) to train a physics-informed DL model. The CT numbers in the physics loss (given by Eq. (4)) are acquired from the VMI with an energy of 80 keV from the scan of the CIRS 062M electron density phantom.
The 80-keV VMI is used because VMI with an energy around 80 keV can lead to the optimal image-noise ratio and beam-hardening ratio (Yu et al., 2012; Wohlfahrt et al., 2017; Wang et al., 2019) to reduce CT number variations between patients with different body sizes.
2.3. Empirical model for DECT parametric mapping
We implemented two empirical DECT models using MATLAB R2021a, and we compared the performance of DL models to the empirical models. Eq. (8) (Beaulieu et al., 2012) and Eq. (9) (Hünemohr et al., 2013; Kassaee et al., 2021) give empirical models to which can be used to estimate the material mass density and RSP maps from DECT where and are obtained from Siemens Syngo.Via. The linear mass density model involves considerable uncertainty for the inflated lung, and a constant value of 0.26 g/cm3 is assigned for it (Eq. (8)). The RSP model includes multiple correlations for different tissues, for example, soft tissues () and bony tissues (). Syngo.Via has a limitation that the is assigned to 0 for low-density tissues (), and, therefore, Eq. (9) includes a region of values < 0.5 for lung tissues.
(8) |
(9) |
2.4. Evaluation
To prepare the reference images of CIRS anthropomorphic phantoms, we first generated contours for each tissue surrogate from phantom images. We used RayStation 10B model-based segmentation (RayPhysics, 2021) to obtain contours for the brain, spinal cord, and lung. The threshold CT number was 250 HU for bone. The breast contour was manually contoured from phantom images. The soft tissue was first contoured by using RayStation 10 function to create external regions of interest (ROI). Then we subtracted the external ROI by each tissue contour to acquire the soft tissue contour. To avoid the partial volume and edge effects (Wohlfahrt et al., 2018; Polf et al., 2019), all contours were uniformly contracted by 3 mm, and the tissue materials within the contours were uniform for CIRS anthropomorphic phantoms. Then the mass density and RSP from Table 1 were allocated to each contour as the reference.
Eq. (10)–(11) give the evaluation metrics using absolute percentage error (APE) and mean APE (MAPE) where , , and represent the ith voxel, voxel quantity, and total voxels. For anthropomorphic phantoms, the APE is used to generate error maps for whole phantoms to visualize the error distribution, and the MAPE provides error estimation for each tissue contour. Both APE and MAPE were computed for each contour within CIRS anthropomorphic phantoms.
(10) |
(11) |
3. Results
We used MAPE to assess the performance of the PIDL framework regarding mass density and RSP map generation and compared the results to the conventional DL models (without the extra physics insight loss term, Eq. (4)) and the empirical models. In Section 3.1–3.3, the reference RSP and mass density for the CIRS anthropomorphic phantoms were from the vendor (lower part of Table 1). In Section 3.4, the reference RSP and mass density for the CIRS M701 adult male phantom were from the proton measurement as described in Section 2.1.1.
3.1. Site-specific analysis using an HN TBDE protocol for CIRS M701, M702, and M705 phantoms
Figure 4 depicts the quantitative evaluation results of the empirical and DL-based DECT parametric mapping for mass densities utilizing CIRS M701, M702, and M705 phantoms. For lung tissue surrogate, the empirical model results in significant uncertainty for low HU materials. Accordingly, the density of the inflated lung is set to be a constant value of 0.26 g/cm3 (Beaulieu et al., 2012), and the error is about 26.8%. In contrast, the DL models can achieve an error of ~1% for lung. Meanwhiles, DL models achieved lower errors than the empirical model for soft tissue, spinal cord, and brain. For M701 bone, the errors are 3.9% and 4.2% for the empirical model and RN-v2. However, the empirical model shows a significant standard deviation (SD) of 5.3%, while the SD of RN-v2 is 0.9%. Compared to the empirical model, DL models usually have a smaller SD. In case where the conventional DL models results in a high MAPE, the physics-informed DL models generally improve the accuracy of density map generations. For instance, the soft tissue errors are 2.2% and 2.4% for conventional ANN and DRN-v2, but the errors are reduced to 1.0% and 1.3% when training the DL models with physics constraints. The bone density generated by RN-v2 shows good agreement to the reference, with a small MAPE (0.2%), while the empirical model results in a MAPE of 4.1% for M705 bone density prediction. For the bone surrogate in M705, the conventional DRN-v2 yields a MAPE of 4.8% but the physics-informed training can reduce the error to 1.5%. In this case, DRN-v2 comprises complicated model architectures and can lead to large MAPE. However, these errors generally can be improved with constraints from the physics model, as shown in Figure 4(c5). Appendix B gives MAPE values of mass density and RSP comparisons for each model.
Figure 5 shows the MAPE of RSP by the empirical and DL models. The empirical model can generate RSP maps using Eq. 8, and the MAPE is 8.1% for lung while the error is 1.1% by RN-v1. Compared to the empirical model, the DL models consistently yield smaller MAPE and SD for soft tissues for M701 and M702. The empirical model can achieve smaller MAPE of brain and bone tissues compared to DL models,but the SD for the empirical model is about four times greater than for the DL models as demonstrated by M701 and M702. For M705 bone, RN-v2 leads to an error of but the empirical model results in an error of . ANN gives optimal MAPE of RSP for soft tissue, spinal cord, and brain at the HN site.
3.2. Site-specific analysis using a thorax TBDE protocol for CIRS M701 and M702 phantoms
Figure 6 shows MAPE values of mass densities and RSP for each tissue using the empirical models and ML/DL models with and without physics training including ANN, FCNN, DFCNN, RN-v1, RN-v2, and DRN-v2. The DL models show lower errors than the empirical models for most tissues. Meanwhiles, the physics-informed training can further reduce the errors for DL models without physics training. For instance, Figure 6(b1) and (b3) show that the errors of soft tissue and bone are 2.9% and 6.1% by DRN-v2 while the errors are 0.6% and 4.8% by physics-informed DRN-v2 (PDRN), respectfully. Appendix B gives MAPE values of mass density and RSP comparisons for each model.
3.3. Site-specific analysis using a pelvic TBDE protocol for CIRS M701 and M702 phantoms
Figure 7 shows the MAPE values of mass densities and RSP for M701 and M702 from the empirical model and ML/DL models with and without physics training. For M701 soft tissue and spinal cord given in Figure 7(a1) and (a2), DRN-v2 results in errors of 1.7% and 2.5%, and the physics-informed training can further lower these errors to 0.9% and 1.0%. Figure 7(c1)–(c2) and (d1)–(d2) depict that DL models can achieve smaller RSP MAPE than the empirical model for M701 soft tissue and spinal cord. Appendix B gives MAPE values of mass density and RSP comparisons for each model.
3.4. Analysis of CIRS M701 phantom using measured tissue-surrogate data
We compared the empirical and DL results to the measured data (given in Section 2.1.1) for CIRS M701 adult male phantom. The results in this section were averaged from the three treatment sites using the CIRS adult male phantom (M701) images scanned by HN, thorax, and pelvic TBDE protocols. Figure 8(a1)–(a4) depicts the MAPE values of the phantom mass densities for each tissue. DL models achieved lower errors than the empirical model. The physics-informed training can generally improve the model errors compared to the models without physics training. For instance, Figure 8(a2) and (a4) show that the errors from physics-informed ANN (PANN) are 1.3% and 2.5% for soft tissue and bone while the errors from ANN without physics training are 2.5% and 3.8%, respectfully. The physics-informed RN-v2 (PRN-v2) achieved the optimal errors of 0.8%, 1.1%, 0.2%, and 0.5% for lung, soft tissue, brain, and bone. Figure 8(b1)–(b4) show the MAPE values of RSP for each tissue from various models.
Figure 9 illustrates APE maps to evaluate error distributions of mass density for CIRS M701 adult male phantom. The VMI shows the anatomy of the phantom at three sites, including HN, thorax, and pelvis. The empirical and ANN model results in considerable uncertainty at the peripheral of the body and bone. When training an ANN with physics constraints, physics-informed ANN (PANN) error maps show improvement compared to the map by ANN. There is no significant difference between RN-v2 and PRN-v2. Both models can better control the errors across all tissues and the regions of the body peripheral compared to the results by the empirical model. Figure 10 shows the APE maps of RSP. The empirical and ANN models still involve significant uncertainty for the bone and regions of the body peripheral. However, the ANN model reveals optimal results for the soft tissue and brain. RN-v2 and PRN-v2 can still achieve the optimal APE for bone and lung compared to ANN and the empirical models.
4. Discussions
Su et al. (Su et al., 2018) employed machine learning methods to establish the correlations between DECT images and parametric maps such as RSP. In this work, we proposed a physics-informed deep learning (PIDL) framework, for the first time, to tightly incorporate a physics model to constrain DL training for DECT parametric mapping. The framework can predict the parametric maps of mass density and RSP to support proton MCDC and ADC for treatment planning. With physics constraints, machine learning models with simplistic structures such as ANN can also reduce mass density errors by 1.2%, 1.3%, and 1.3% for soft tissue, brain, and bone compared to ANN without physics training, as shown in Figure 8(a2)–(a4). Figure 8 also implies that the model errors do not tightly depend on different DL models tested in this work; instead, physics-informed training can improve the model accuracy when conventional DL models lead to large uncertainty. Meanwhiles, Figure 4, Figure 6, and Figure 8–9 show that the empirical DECT model given by Eq. (8) results in large uncertainty for lung surrogates from CIRS anthropomorphic phantoms. Since the model use a constant value for inflated lung, we can match the constant value to the reference mass density of lung surrogates to have zero uncertainty. However, this approach cannot be applied to patient study, and this result shows the limitation of Eq. (8).
The physics-based empirical models (Yang et al., 2010; Landry et al., 2013; Garcia et al., 2015; Zhu and Penfold, 2016) had demonstrated the capabilities of DECT parametric mapping. Nonetheless, the accuracy of those methods was impacted by CT noise and artifacts. Figure 9(c2) illustrates that the empirical model is affected by CT noise, and the phantom anatomy cannot be identified. In contrast, the DL models can suppress the noise, as shown in Figure 9(c2) and (c6) where the standard deviations of APE for whole bodies are 2.3% (empirical model) and 1.4% (PRN-v2), respectfully. Meanwhile, those methods generally involve different degrees of uncertainty due to model forms of regression methods such as linear fitting. The relations between CT numbers and quantities of interest may not always be described as a one-to-one function (Su et al., 2018). To prevent this ambiguity, DL with hierarchical structures has the potential to figure out the hidden correlations behind DECT data and achieve the accuracy and robustness of DL-based parametric mapping methods.
Hünemohr et al. (Hünemohr et al., 2014) concluded that the accuracy of mass density prediction dominates the uncertainty of RSP. Proton MCDC is accurate for particle transport in heterogeneous tissues (Schuemann et al., 2015; Huang et al., 2018; Lin et al., 2021) and can improve treatment planning quality. Accordingly, in the current work, we focus on using a physics model to constrain DL training based on the mass density, and Figure 8(a1)–(a4) depicts that PRN-v2 can achieve MAPE within 1.1%. Besides, when comparing DECT mass density mapping results to bone measurement data, both empirical and PRN-v2 reduce the errors by and . PRN-v2 performs adequately for both CIRS adult and child anthropomorphic phantoms. The model involves complicated convolutional layers to uncover structures of high-dimensional data and residual blocks to prevent accuracy degradation. We will incorporate PRN-v2 for future dosimetric study to further explore the impacts for treatment planning.
Figure 11 depicts density map prediction for two patients employing HN and pelvis TBDE scans with contrast. All density maps can qualitatively reflect the patients’ anatomy. Figure 11(a1)–(a4) display that the internal jugular vein (IJV) can impact the CT numbers that cause the variation of density curves by RN-v2 and PRN-v2 as shown in Figure 11(a5) and (b5) (pointed by red and blue arrows). Figure 11(a5) demonstrates that the density line profiles by DL models agree with the trend of the HU line profile by VMI. The empirical model is not affected by the variation of CT numbers induced by the contrast injection. It is because iodine is a high material, and it enhances the photo electric effect that impacts CT number. However, the contrast injected in IJV was OMNIPAQUE™ with the specific gravity of 1.406, and 100 ml of the iodine contrast was injected into the patient. The injection of contrast may cause local density variation. Meanwhile, DL models were not trained with iodine that increases the uncertainty from DL models. The patients with contrast become the limitation of the current PIDL framework due to lack of training data. Including more training datasets with iodine contrast may address this issue. Figure 11(b6) exhibits a line profile across soft tissue and bone, and both empirical and DL models agree with the trend of the VMI HU profile. Nevertheless, DL line profiles are smooth that implies the ability to suppress CT noise.
Although Figure 8 indicates that the uncertainty for RSP prediction is more significant than that of mass density prediction by DL models, Figure 8(b1)–(b4) still shows improvement achieved by DL models for lung, soft tissue, spinal cord, and bone compared to the empirical model. Meanwhile, the reference RSP values are calculated by Eq. (1), and the uncertainty of reference RSP comes from elemental compositions and values. For instance, Table 1 demonstrates that the density discrepancy between CIRS 062M muscle and CIRS M701 soft tissue is less than 0.5%, but the RSP and value differences are and due to different material compositions. Additional human tissue surrogates are required to broaden the coverage of training data and to accurate DL models. MC simulation (Hünemohr et al., 2013) can also be an alternative to derive RSP. Correlating DECT data to the parametric maps requires models to capture features from high dimensional data in the domain of extrapolation (Balestriero et al., 2021).
The DL framework serves as a high-order approximation for a well-defined physics problem that can be solved if the exact relative electron density and effective atomic number are known for patients. Hence, the accuracy of image-based relative electron density and effective atomic number dominate the uncertainty of physics-based models for material mass density and RSP inferences. Unlike mechanistic models (empirical models), the robustness of DL can be compromised if the training datasets are not sufficiently covered the physics problem that we attack. In this work, we aim to demonstrate the feasibility of using physics-informed training to enhance model accuracy for conventional ML and DL models. The evaluation of dosimetric impacts using the proposed PIDL framework with meat phantoms such as pig’s loin, belly, rib, and femur, is a crucial step for accessing the feasibility of clinical implementation.
Future work will incorporate additional physics models to regularize DL training to enhance accuracy and robustness (Chang and Dinh, 2019). The current model inputs include VMI of 135 keV and 190 keV because of their potential to mitigate artifacts from surgical implants of various sizes (Wellenberg et al., 2018a; Wellenberg et al., 2018b). Future investigation will likewise integrate the proposed framework to regulate DL-based metal artifact reduction methods (Gjesteby et al., 2019; Liao et al., 2020; Yu et al., 2021a; Yu et al., 2021b) to improve proton range uncertainty (Paganetti, 2012; Chang et al., 2020).
5. Conclusions
A physics-informed deep learning (PIDL) framework was developed to assimilate rich information from DECT and to generate mass density and RSP maps that can be used to inform proton analytical dose calculation and Monte Carlo dose calculation. Physics-informed deep learning models can enhance the accuracy of DECT parametric mapping for voxel-specific material mass densities, mainly when conventional DL models incorporate significant uncertainty. Machine learning models with basic structures can benefit the most from physics-informed training. DL models with complicated structures are not necessary to achieve the optimal DECT parametric mapping; alternatively, physics regularization can effectively improve the performance of DL models. The proposed framework also demonstrated the capability of suppressing CT noise, and the method has the potential to reduce proton range uncertainty to increase the quality of proton therapy.
Acknowledgments
This research is supported in part by the National Cancer Institute of the National Institutes of Health under Award Number R01CA215718.
Appendix A. Structure of ResNet-v1 (RN-v1) and ResNet-v2 (RN-v2)
Table A1 gives the structure of RN-v1 and RN-v2 as given in Figure 3(d)–(e). Each residual block includes two convolutional layers and one residual connection.
Table A1.
Network | Layer | Number of channels | Kernel size | Stride | Pad |
---|---|---|---|---|---|
ConvA | 64 | 7 | 2 | 3 | |
ConvB | 64 | 3 | 1 | 0 | |
| |||||
Residual Block A1 | Convolutional | 64 | 3 | 2 | 1 |
Convolutional | 64 | 3 | 1 | 1 | |
Residual | 64 | 1 | 2 | 0 | |
| |||||
Residual Block A2 | Convolutional | 128 | 3 | 2 | 1 |
Convolutional | 128 | 3 | 1 | 1 | |
Residual | 128 | 1 | 2 | 0 | |
| |||||
Residual Block A3 | Convolutional | 256 | 3 | 2 | 1 |
Convolutional | 256 | 3 | 1 | 1 | |
Residual | 256 | 1 | 2 | 0 | |
| |||||
Residual Block B1 | Convolutional | 64 | 2 | 2 | 1 |
Convolutional | 64 | 2 | 1 | 1 | |
Residual | 64 | 1 | 2 | 0 | |
| |||||
Residual Block B2 | Convolutional | 128 | 2 | 2 | 1 |
Convolutional | 128 | 2 | 1 | 1 | |
Residual | 128 | 1 | 2 | 0 | |
| |||||
Residual Block B3 | Convolutional | 256 | 2 | 2 | 1 |
Convolutional | 256 | 2 | 1 | 1 | |
Residual | 256 | 1 | 2 | 0 |
Appendix B. MAPE comparisons of mass densities and RSP for CIRS anthropomorphic phantoms
Table B1–B2 give the MAPE comparisons of mass densities and RSP using an HN TBDE protocol for CIRS adult male, adult female, and child phantoms. Table B3–B4 show the MAPR comparisons of mass densities and RSP using a thorax TBDE protocol for CIRS adult male and female phantoms. Table B5–B6 indicate the MAPE comparisons of mass densities and RSP using a pelvic TBDE protocol for CIRS adult male and female phantoms.
Table B1.
M701 | Lung | Soft Tissue | Spinal Cord | Brain | Bone | |
---|---|---|---|---|---|---|
Empirical Model | 26.8 ± 0.0 | 2.3 ± 2.2 | 1.4 ± 1.3 | 1.0 ± 0.8 | 3.9 ± 5.3 | |
| ||||||
Conventional | ANN | 9.0 ± 7.3 | 2.2 ± 1.3 | 1.6 ± 1.3 | 1.8 ± 0.6 | 8.3 ± 1.2 |
FCNN | 1.3 ± 3.3 | 1.2 ± 1.0 | 0.7 ± 1.0 | 0.2 ± 0.2 | 5.3 ± 0.8 | |
DFCNN | 2.0 ± 3.1 | 1.1 ± 0.8 | 0.8 ± 1.1 | 0.2 ± 0.2 | 5.6 ± 1.0 | |
RN-v1 | 0.9 ± 3.4 | 1.7 ± 1.8 | 0.8 ± 1.4 | 0.2 ± 0.2 | 4.1 ± 0.9 | |
RN-v2 | 1.1 ± 3.4 | 1.3 ± 1.2 | 0.7 ± 0.9 | 0.3 ± 0.2 | 4.2 ± 0.9 | |
DRN-v2 | 1.8 ± 4.2 | 2.4 ± 2.1 | 1.1 ± 1.2 | 0.4 ± 0.5 | 5.1 ± 1.6 | |
| ||||||
Physics-informed | ANN | 8.4 ± 7.2 | 1.0 ± 0.9 | 1.0 ± 0.9 | 0.5 ± 0.5 | 7.1 ± 1.1 |
FCNN | 1.7 ± 3.3 | 1.2 ± 0.7 | 0.7 ± 0.5 | 0.2 ± 0.2 | 4.9 ± 0.6 | |
DFCNN | 1.9 ± 3.3 | 1.1 ± 0.6 | 0.8 ± 0.5 | 0.3 ± 0.2 | 5.3 ± 0.7 | |
RN-v1 | 1.4 ± 3.7 | 1.3 ± 0.7 | 0.7 ± 0.5 | 0.2 ± 0.2 | 4.5 ± 0.7 | |
RN-v2 | 1.4 ± 3.6 | 1.2 ± 0.7 | 0.7 ± 0.5 | 0.2 ± 0.2 | 4.7 ± 0.9 | |
DRN-v2 | 1.3 ± 4.2 | 1.3 ± 0.7 | 0.7 ± 0.5 | 0.2 ± 0.2 | 5.0 ± 1.1 | |
| ||||||
M702 | Lung | Soft Tissue | Spinal Cord | Brain | Bone | |
| ||||||
Empirical Model | 26.8 ± 0.0 | 2.3 ± 2.2 | 1.2 ± 1.0 | 1.3 ± 0.9 | 3.5 ± 2.1 | |
| ||||||
Conventional | ANN | 10.0 ± 8.0 | 2.3 ± 1.4 | 1.7 ± 0.8 | 2.3 ± 0.6 | 8.4 ± 1.1 |
FCNN | 1.5 ± 0.9 | 1.2 ± 1.0 | 0.4 ± 0.5 | 0.2 ± 0.3 | 5.4 ± 0.7 | |
DFCNN | 2.0 ± 0.5 | 1.1 ± 0.9 | 0.4 ± 0.5 | 0.3 ± 0.3 | 5.6 ± 0.8 | |
RN-v1 | 0.9 ± 0.7 | 1.4 ± 1.5 | 0.4 ± 0.6 | 0.3 ± 0.3 | 4.1 ± 0.9 | |
RN-v2 | 1.2 ± 0.8 | 1.4 ± 1.2 | 0.4 ± 0.5 | 0.2 ± 0.3 | 4.2 ± 0.8 | |
DRN-v2 | 1.8 ± 0.9 | 2.2 ± 2.0 | 0.7 ± 0.7 | 0.5 ± 0.5 | 5.3 ± 1.6 | |
| ||||||
Physics-informed | ANN | 9.3 ± 8.4 | 1.0 ± 1.1 | 0.5 ± 0.6 | 0.6 ± 0.5 | 7.2 ± 1.2 |
FCNN | 1.8 ± 0.6 | 1.2 ± 0.9 | 0.3 ± 0.5 | 0.3 ± 0.3 | 4.8 ± 0.6 | |
DFCNN | 1.9 ± 0.6 | 1.1 ± 1.0 | 0.4 ± 0.5 | 0.3 ± 0.3 | 5.3 ± 0.7 | |
RN-v1 | 1.3 ± 0.8 | 1.3 ± 1.1 | 0.4 ± 0.4 | 0.2 ± 0.3 | 4.6 ± 0.8 | |
RN-v2 | 1.3 ± 0.8 | 1.3 ± 1.1 | 0.4 ± 0.5 | 0.2 ± 0.3 | 4.6 ± 0.8 | |
DRN-v2 | 1.1 ± 0.7 | 1.3 ± 1.1 | 0.4 ± 0.4 | 0.2 ± 0.3 | 4.8 ± 0.9 | |
| ||||||
M705 | Lung | Soft Tissue | Spinal Cord | Brain | Bone | |
| ||||||
Empirical Model | 26.8 ± 0.0 | 2.2 ± 1.5 | 1.2 ± 1.2 | 1.3 ± 0.8 | 4.1 ± 2.0 | |
| ||||||
Conventional | ANN | 7.4 ± 5.3 | 2.9 ± 0.8 | 1.9 ± 1.0 | 2.4 ± 0.5 | 8.5 ± 0.9 |
FCNN | 1.1 ± 1.7 | 0.7 ± 0.7 | 0.4 ± 0.7 | 0.3 ± 0.3 | 2.7 ± 0.9 | |
DFCNN | 1.8 ± 1.6 | 0.6 ± 0.6 | 0.3 ± 0.9 | 0.3 ± 0.2 | 3.2 ± 1.0 | |
RN-v1 | 0.6 ± 1.7 | 0.8 ± 0.6 | 0.4 ± 0.6 | 0.3 ± 0.3 | 0.1 ± 0.9 | |
RN-v2 | 0.9 ± 1.7 | 0.9 ± 0.6 | 0.4 ± 0.7 | 0.2 ± 0.2 | 0.2 ± 1.0 | |
DRN-v2 | 1.4 ± 1.7 | 1.3 ± 1.4 | 0.5 ± 0.7 | 0.6 ± 0.5 | 4.8 ± 1.3 | |
| ||||||
Physics-informed | ANN | 6.7 ± 5.2 | 1.1 ± 0.7 | 0.5 ± 0.9 | 0.7 ± 0.5 | 7.7 ± 0.9 |
FCNN | 1.6 ± 1.6 | 0.7 ± 0.6 | 0.3 ± 0.7 | 0.3 ± 0.2 | 1.5 ± 0.9 | |
DFCNN | 1.7 ± 1.7 | 0.6 ± 0.6 | 0.4 ± 0.8 | 0.3 ± 0.3 | 2.5 ± 1.1 | |
RN-v1 | 1.0 ± 1.8 | 0.8 ± 0.5 | 0.4 ± 0.6 | 0.3 ± 0.2 | 1.1 ± 1.0 | |
RN-v2 | 1.0 ± 1.7 | 0.8 ± 0.5 | 0.4 ± 0.7 | 0.2 ± 0.2 | 0.6 ± 0.9 | |
DRN-v2 | 0.7 ± 2.0 | 0.8 ± 0.5 | 0.4 ± 0.6 | 0.2 ± 0.2 | 1.5 ± 1.1 |
Table B2.
M701 | Lung | Soft Tissue | Spinal Cord | Brain | Bone | |
---|---|---|---|---|---|---|
Empirical Model | 8.1 ± 7.9 | 2.9 ± 2.6 | 2.7 ± 1.9 | 1.2 ± 0.9 | 4.2 ± 4.9 | |
| ||||||
Conventional | ANN | 9.6 ± 7.2 | 1.1 ± 1.0 | 1.7 ± 0.8 | 0.5 ± 0.4 | 8.0 ± 1.1 |
FCNN | 1.5 ± 3.3 | 2.1 ± 0.8 | 3.0 ± 0.6 | 1.8 ± 0.2 | 5.7 ± 0.7 | |
DFCNN | 2.4 ± 3.2 | 2.2 ± 0.6 | 3.1 ± 0.5 | 1.8 ± 0.2 | 5.8 ± 0.9 | |
RN-v1 | 1.1 ± 3.5 | 2.5 ± 1.1 | 3.0 ± 0.7 | 1.8 ± 0.3 | 4.6 ± 0.8 | |
RN-v2 | 1.3 ± 3.4 | 2.4 ± 0.8 | 3.1 ± 0.5 | 1.9 ± 0.2 | 4.8 ± 0.8 | |
DRN-v2 | 1.8 ± 3.8 | 3.0 ± 2.1 | 3.3 ± 1.3 | 1.8 ± 0.5 | 5.1 ± 1.3 | |
| ||||||
Physics-informed | ANN | 9.6 ± 7.2 | 1.4 ± 0.9 | 2.8 ± 0.9 | 1.5 ± 0.5 | 6.9 ± 1.2 |
FCNN | 2.0 ± 3.3 | 2.2 ± 0.7 | 3.0 ± 0.5 | 1.7 ± 0.2 | 5.3 ± 0.7 | |
DFCNN | 2.1 ± 3.3 | 2.2 ± 0.6 | 3.0 ± 0.5 | 1.7 ± 0.2 | 5.7 ± 0.8 | |
RN-v1 | 1.6 ± 4.0 | 2.4 ± 0.7 | 3.1 ± 0.5 | 1.9 ± 0.2 | 5.2 ± 0.8 | |
RN-v2 | 1.6 ± 3.6 | 2.3 ± 0.7 | 3.1 ± 0.5 | 1.9 ± 0.2 | 5.2 ± 1.0 | |
DRN-v2 | 1.3 ± 4.2 | 2.4 ± 0.7 | 3.1 ± 0.6 | 1.9 ± 0.2 | 5.1 ± 1.2 | |
| ||||||
M702 | Lung | Soft Tissue | Spinal Cord | Brain | Bone | |
| ||||||
Empirical Model | 8.9 ± 11.0 | 2.6 ± 2.4 | 1.6 ± 1.4 | 0.9 ± 0.8 | 3.9 ± 2.2 | |
| ||||||
Conventional | ANN | 10.5 ± 7.8 | 1.2 ± 1.2 | 1.3 ± 0.7 | 0.4 ± 0.4 | 8.0 ± 1.0 |
FCNN | 1.8 ± 0.9 | 2.1 ± 0.8 | 3.0 ± 0.5 | 1.6 ± 0.3 | 5.8 ± 0.6 | |
DFCNN | 2.3 ± 0.5 | 2.2 ± 0.6 | 3.0 ± 0.5 | 1.6 ± 0.2 | 5.8 ± 0.6 | |
RN-v1 | 1.1 ± 0.7 | 2.4 ± 0.9 | 2.9 ± 0.5 | 1.7 ± 0.3 | 4.6 ± 0.8 | |
RN-v2 | 1.5 ± 0.8 | 2.5 ± 0.7 | 3.0 ± 0.5 | 1.7 ± 0.3 | 4.8 ± 0.7 | |
DRN-v2 | 1.8 ± 0.9 | 2.7 ± 1.9 | 3.1 ± 0.9 | 1.6 ± 0.5 | 5.3 ± 1.3 | |
| ||||||
Physics-informed | ANN | 10.3 ± 8.0 | 1.4 ± 1.0 | 2.6 ± 0.7 | 1.1 ± 0.5 | 7.0 ± 1.1 |
FCNN | 2.1 ± 0.6 | 2.2 ± 0.7 | 3.0 ± 0.4 | 1.6 ± 0.2 | 5.3 ± 0.5 | |
DFCNN | 2.1 ± 0.6 | 2.2 ± 0.7 | 3.0 ± 0.4 | 1.6 ± 0.2 | 5.7 ± 0.6 | |
RN-v1 | 1.5 ± 0.8 | 2.4 ± 0.7 | 3.1 ± 0.4 | 1.7 ± 0.3 | 5.1 ± 0.6 | |
RN-v2 | 1.5 ± 0.8 | 2.4 ± 0.7 | 3.0 ± 0.5 | 1.7 ± 0.3 | 5.0 ± 0.7 | |
DRN-v2 | 1.1 ± 0.7 | 2.5 ± 0.7 | 3.1 ± 0.4 | 1.7 ± 0.3 | 4.9 ± 0.8 | |
| ||||||
M705 | Lung | Soft Tissue | Spinal Cord | Brain | Bone | |
| ||||||
Empirical Model | 6.4 ± 6.0 | 1.5 ± 1.4 | 1.5 ± 1.1 | 0.8 ± 0.7 | 4.6 ± 2.0 | |
| ||||||
Conventional | ANN | 8.4 ± 5.4 | 1.1 ± 0.8 | 1.2 ± 0.8 | 0.4 ± 0.4 | 8.4 ± 0.8 |
FCNN | 1.3 ± 1.7 | 1.6 ± 0.6 | 3.0 ± 0.6 | 1.6 ± 0.2 | 3.6 ± 0.7 | |
DFCNN | 2.3 ± 1.6 | 1.8 ± 0.4 | 3.0 ± 0.6 | 1.5 ± 0.2 | 4.0 ± 0.9 | |
RN-v1 | 0.8 ± 1.7 | 2.0 ± 0.4 | 3.0 ± 0.5 | 1.7 ± 0.3 | 1.3 ± 0.8 | |
RN-v2 | 1.2 ± 1.7 | 2.0 ± 0.4 | 3.0 ± 0.6 | 1.7 ± 0.3 | 1.4 ± 0.8 | |
DRN-v2 | 1.5 ± 2.0 | 1.6 ± 0.9 | 2.9 ± 0.6 | 1.6 ± 0.5 | 4.8 ± 1.1 | |
| ||||||
Physics-informed | ANN | 8.3 ± 5.4 | 0.7 ± 0.5 | 2.5 ± 0.7 | 1.0 ± 0.4 | 7.7 ± 0.8 |
FCNN | 1.9 ± 1.6 | 1.8 ± 0.5 | 2.9 ± 0.5 | 1.6 ± 0.2 | 2.6 ± 0.7 | |
DFCNN | 1.9 ± 1.6 | 1.8 ± 0.5 | 3.0 ± 0.5 | 1.6 ± 0.2 | 3.4 ± 0.9 | |
RN-v1 | 1.2 ± 1.8 | 1.9 ± 0.4 | 3.1 ± 0.4 | 1.7 ± 0.3 | 2.3 ± 0.8 | |
RN-v2 | 1.1 ± 1.7 | 2.0 ± 0.4 | 3.0 ± 0.5 | 1.7 ± 0.3 | 1.8 ± 0.8 | |
DRN-v2 | 0.7 ± 1.9 | 2.0 ± 0.4 | 3.0 ± 0.4 | 1.7 ± 0.2 | 2.5 ± 1.1 |
Table B3.
M701 | Lung | Breast | Soft Tissue | Spinal Cord | Bone | |
---|---|---|---|---|---|---|
Empirical Model | 26.8 ± 0.0 | - | 3.1 ± 2.5 | 3.4 ± 2.6 | 7.8 ± 3.9 | |
| ||||||
Conventional | ANN | 11.4 ± 7.2 | - | 2.6 ± 1.8 | 1.5 ± 1.3 | 8.5 ± 2.1 |
FCNN | 1.1 ± 1.6 | - | 1.5 ± 1.7 | 0.8 ± 0.9 | 5.6 ± 1.3 | |
DFCNN | 1.9 ± 1.5 | - | 1.4 ± 1.6 | 1.0 ± 1.0 | 5.6 ± 1.6 | |
RN-v1 | 0.7 ± 1.6 | - | 1.5 ± 1.9 | 1.7 ± 2.4 | 4.2 ± 1.2 | |
RN-v2 | 0.9 ± 1.5 | - | 1.4 ± 1.8 | 0.7 ± 0.8 | 4.3 ± 1.3 | |
DRN-v2 | 1.6 ± 1.8 | - | 2.5 ± 2.7 | 3.4 ± 2.4 | 6.3 ± 2.0 | |
| ||||||
Physics-informed | ANN | 10.6 ± 7.0 | - | 1.5 ± 1.5 | 1.4 ± 1.1 | 7.6 ± 2.1 |
FCNN | 1.6 ± 1.5 | - | 1.4 ± 1.6 | 0.9 ± 0.8 | 4.9 ± 1.1 | |
DFCNN | 1.8 ± 1.6 | - | 1.4 ± 1.6 | 1.0 ± 1.0 | 5.4 ± 1.4 | |
RN-v1 | 1.0 ± 1.7 | - | 1.4 ± 1.6 | 0.7 ± 0.8 | 4.7 ± 1.2 | |
RN-v2 | 1.0 ± 1.6 | - | 1.3 ± 1.7 | 0.7 ± 0.9 | 4.6 ± 1.2 | |
DRN-v2 | 0.8 ± 1.8 | - | 1.4 ± 1.6 | 0.7 ± 0.8 | 4.8 ± 1.3 | |
| ||||||
M702 | Lung | Breast | Soft Tissue | Spinal Cord | Bone | |
| ||||||
Empirical Model | 26.8 ± 0.0 | 3.4 ± 2.2 | 2.7 ± 2.2 | 2.5 ± 1.9 | 7.1 ± 6.9 | |
| ||||||
Conventional | ANN | 10.6 ± 6.5 | 2.5 ± 1.3 | 2.5 ± 1.4 | 1.2 ± 0.9 | 8.3 ± 1.8 |
FCNN | 1.2 ± 0.8 | 1.6 ± 0.8 | 1.2 ± 1.1 | 0.6 ± 0.6 | 5.6 ± 1.1 | |
DFCNN | 1.9 ± 0.5 | 1.4 ± 0.8 | 1.1 ± 1.0 | 0.8 ± 0.6 | 5.7 ± 1.3 | |
RN-v1 | 0.7 ± 0.7 | 1.9 ± 1.1 | 1.3 ± 1.4 | 1.4 ± 2.2 | 4.2 ± 1.1 | |
RN-v2 | 0.9 ± 0.7 | 1.8 ± 1.0 | 1.3 ± 1.2 | 0.6 ± 0.5 | 4.3 ± 1.1 | |
DRN-v2 | 1.5 ± 0.8 | 2.5 ± 1.6 | 2.1 ± 2.2 | 2.9 ± 1.9 | 6.1 ± 1.9 | |
| ||||||
Physics-informed | ANN | 9.7 ± 6.2 | 1.2 ± 1.0 | 1.2 ± 1.0 | 1.1 ± 0.9 | 7.4 ± 1.8 |
FCNN | 1.6 ± 0.5 | 1.5 ± 0.8 | 1.2 ± 1.0 | 0.7 ± 0.5 | 4.9 ± 0.9 | |
DFCNN | 1.7 ± 0.5 | 1.5 ± 0.8 | 1.2 ± 1.1 | 0.8 ± 0.7 | 5.4 ± 1.1 | |
RN-v1 | 1.0 ± 0.8 | 1.8 ± 1.0 | 1.2 ± 1.1 | 0.6 ± 0.4 | 4.8 ± 1.0 | |
RN-v2 | 1.0 ± 0.6 | 1.7 ± 0.9 | 1.2 ± 1.1 | 0.5 ± 0.5 | 4.6 ± 1.1 | |
DRN-v2 | 0.8 ± 0.7 | 1.6 ± 0.9 | 1.2 ± 1.1 | 0.6 ± 0.4 | 4.8 ± 1.1 |
Table B4.
M701 | Lung | Breast | Soft Tissue | Spinal Cord | Bone | |
---|---|---|---|---|---|---|
Empirical Model | 7.0 ± 5.8 | - | 3.2 ± 2.5 | 5.8 ± 3.8 | 8.7 ± 4.1 | |
| ||||||
Conventional | ANN | 12.3 ± 7.3 | - | 1.4 ± 1.4 | 2.7 ± 1.2 | 8.5 ± 1.8 |
FCNN | 1.4 ± 1.7 | - | 2.0 ± 1.2 | 3.0 ± 0.8 | 5.9 ± 1.1 | |
DFCNN | 2.3 ± 1.5 | - | 2.2 ± 1.1 | 3.2 ± 0.7 | 5.9 ± 1.4 | |
RN-v1 | 0.9 ± 1.6 | - | 2.3 ± 1.2 | 2.8 ± 1.3 | 4.8 ± 1.0 | |
RN-v2 | 1.2 ± 1.5 | - | 2.3 ± 1.2 | 3.0 ± 0.7 | 4.8 ± 1.1 | |
DRN-v2 | 1.6 ± 1.9 | - | 2.6 ± 2.0 | 6.0 ± 2.9 | 6.2 ± 1.8 | |
| ||||||
Physics-informed | ANN | 12.2 ± 7.3 | - | 1.4 ± 1.2 | 3.5 ± 1.3 | 7.7 ± 1.9 |
FCNN | 1.9 ± 1.5 | - | 2.1 ± 1.1 | 3.3 ± 0.6 | 5.3 ± 1.0 | |
DFCNN | 1.9 ± 1.6 | - | 2.1 ± 1.1 | 3.1 ± 0.7 | 5.7 ± 1.2 | |
RN-v1 | 1.2 ± 1.7 | - | 2.2 ± 1.1 | 3.1 ± 0.7 | 5.2 ± 1.0 | |
RN-v2 | 1.2 ± 1.7 | - | 2.2 ± 1.1 | 2.9 ± 0.8 | 5.0 ± 1.0 | |
DRN-v2 | 0.8 ± 1.7 | - | 2.3 ± 1.1 | 3.2 ± 0.8 | 4.9 ± 1.3 | |
| ||||||
M702 | Lung | Breast | Soft Tissue | Spinal Cord | Bone | |
| ||||||
Empirical Model | 6.7 ± 5.5 | 2.7 ± 1.9 | 2.7 ± 2.2 | 5.3 ± 3.1 | 8.0 ± 6.4 | |
| ||||||
Conventional | ANN | 11.4 ± 6.6 | 1.1 ± 0.9 | 1.2 ± 1.1 | 2.5 ± 1.0 | 8.2 ± 1.5 |
FCNN | 1.4 ± 0.8 | 0.8 ± 0.8 | 1.9 ± 0.9 | 3.2 ± 0.6 | 6.0 ± 0.9 | |
DFCNN | 2.3 ± 0.4 | 0.8 ± 0.7 | 2.1 ± 0.6 | 3.3 ± 0.5 | 6.0 ± 1.1 | |
RN-v1 | 0.9 ± 0.7 | 1.0 ± 0.9 | 2.3 ± 0.8 | 3.0 ± 1.1 | 4.7 ± 1.0 | |
RN-v2 | 1.2 ± 0.7 | 0.9 ± 0.8 | 2.3 ± 0.7 | 3.2 ± 0.5 | 4.8 ± 1.0 | |
DRN-v2 | 1.5 ± 0.9 | 1.3 ± 1.2 | 2.3 ± 1.7 | 5.8 ± 2.3 | 6.0 ± 1.6 | |
| ||||||
Physics-informed | ANN | 11.2 ± 6.5 | 1.4 ± 1.0 | 1.2 ± 0.9 | 3.5 ± 1.0 | 7.5 ± 1.6 |
FCNN | 1.9 ± 0.5 | 0.9 ± 0.8 | 2.1 ± 0.7 | 3.3 ± 0.4 | 5.4 ± 0.8 | |
DFCNN | 1.9 ± 0.5 | 0.8 ± 0.8 | 2.1 ± 0.7 | 3.2 ± 0.5 | 5.8 ± 0.9 | |
RN-v1 | 1.2 ± 0.8 | 0.9 ± 0.8 | 2.2 ± 0.7 | 3.3 ± 0.4 | 5.2 ± 0.9 | |
RN-v2 | 1.2 ± 0.7 | 0.9 ± 0.8 | 2.2 ± 0.7 | 3.1 ± 0.5 | 5.0 ± 0.9 | |
DRN-v2 | 0.8 ± 0.7 | 0.9 ± 0.8 | 2.2 ± 0.7 | 3.4 ± 0.5 | 5.0 ± 1.1 |
Table B5.
M701 | Soft Tissue | Spinal Cord | Bone | |
---|---|---|---|---|
Empirical Model | 2.5 ± 2.0 | 3.3 ± 2.7 | 3.3 ± 2.2 | |
| ||||
Conventional | ANN | 2.6 ± 1.3 | 2.3 ± 1.5 | 7.2 ± 1.3 |
FCNN | 1.0 ± 0.9 | 1.2 ± 1.4 | 4.7 ± 0.8 | |
DFCNN | 0.9 ± 0.9 | 1.0 ± 1.2 | 5.1 ± 1.0 | |
RN-v1 | 1.0 ± 1.2 | 1.1 ± 1.8 | 4.2 ± 0.7 | |
RN-v2 | 0.9 ± 0.9 | 1.1 ± 1.8 | 4.2 ± 0.6 | |
DRN-v2 | 1.7 ± 1.8 | 2.5 ± 2.5 | 3.9 ± 1.4 | |
| ||||
Physics-informed | ANN | 1.2 ± 1.0 | 1.4 ± 1.1 | 5.6 ± 1.4 |
FCNN | 1.0 ± 0.9 | 0.9 ± 1.1 | 4.7 ± 0.8 | |
DFCNN | 0.9 ± 0.9 | 1.1 ± 1.2 | 4.9 ± 0.8 | |
RN-v1 | 0.9 ± 0.8 | 0.9 ± 1.4 | 4.6 ± 0.8 | |
RN-v2 | 0.9 ± 0.9 | 1.2 ± 1.8 | 4.7 ± 1.0 | |
DRN-v2 | 0.9 ± 0.8 | 1.0 ± 1.6 | 4.9 ± 1.1 | |
| ||||
M702 | Soft Tissue | Spinal Cord | Bone | |
| ||||
Empirical Model | 2.3 ± 1.9 | 2.1 ± 1.8 | 3.2 ± 2.0 | |
| ||||
Conventional | ANN | 2.6 ± 1.2 | 1.6 ± 1.1 | 7.5 ± 1.2 |
FCNN | 0.9 ± 0.9 | 0.7 ± 0.8 | 4.9 ± 0.8 | |
DFCNN | 0.8 ± 0.8 | 0.6 ± 0.8 | 5.3 ± 1.1 | |
RN-v1 | 1.0 ± 1.1 | 0.7 ± 1.1 | 4.2 ± 0.8 | |
RN-v2 | 0.9 ± 0.9 | 0.6 ± 0.9 | 4.2 ± 0.8 | |
DRN-v2 | 1.6 ± 1.8 | 1.2 ± 1.4 | 4.2 ± 1.3 | |
| ||||
Physics-informed | ANN | 1.1 ± 0.9 | 1.0 ± 0.8 | 6.0 ± 1.3 |
FCNN | 0.9 ± 0.8 | 0.5 ± 0.7 | 4.9 ± 0.8 | |
DFCNN | 0.9 ± 0.9 | 0.7 ± 0.8 | 5.0 ± 0.9 | |
RN-v1 | 0.9 ± 0.8 | 0.6 ± 0.7 | 4.6 ± 0.9 | |
RN-v2 | 0.8 ± 0.8 | 0.6 ± 0.8 | 4.7 ± 1.0 | |
DRN-v2 | 0.9 ± 0.8 | 0.6 ± 0.8 | 5.0 ± 1.2 |
Table B6.
M701 | Soft Tissue | Spinal Cord | Bone | |
---|---|---|---|---|
Empirical Model | 2.6 ± 2.1 | 2.8 ± 2.2 | 3.5 ± 2.3 | |
| ||||
Conventional | ANN | 1.1 ± 0.9 | 1.3 ± 1.0 | 6.9 ± 1.1 |
FCNN | 1.7 ± 0.7 | 2.6 ± 1.1 | 5.2 ± 0.7 | |
DFCNN | 1.9 ± 0.6 | 2.7 ± 0.8 | 5.4 ± 0.9 | |
RN-v1 | 2.0 ± 0.6 | 2.4 ± 0.7 | 4.6 ± 0.6 | |
RN-v2 | 1.9 ± 0.6 | 2.5 ± 0.8 | 4.8 ± 0.5 | |
DRN-v2 | 2.1 ± 1.3 | 2.7 ± 1.7 | 4.1 ± 1.1 | |
| ||||
Physics-informed | ANN | 1.0 ± 0.8 | 2.0 ± 1.2 | 5.7 ± 1.2 |
FCNN | 1.8 ± 0.6 | 2.6 ± 0.8 | 5.2 ± 0.7 | |
DFCNN | 1.8 ± 0.6 | 2.8 ± 0.8 | 5.3 ± 0.7 | |
RN-v1 | 1.9 ± 0.6 | 2.4 ± 0.8 | 5.1 ± 0.7 | |
RN-v2 | 1.9 ± 0.6 | 2.5 ± 0.8 | 5.2 ± 0.9 | |
DRN-v2 | 1.9 ± 0.6 | 2.7 ± 0.8 | 5.0 ± 1.1 | |
| ||||
M702 | Soft Tissue | Spinal Cord | Bone | |
| ||||
Empirical Model | 2.4 ± 1.9 | 2.5 ± 1.9 | 3.4 ± 2.1 | |
| ||||
Conventional | ANN | 1.0 ± 0.9 | 1.5 ± 0.8 | 7.2 ± 1.1 |
FCNN | 1.7 ± 0.7 | 2.9 ± 0.7 | 5.4 ± 0.7 | |
DFCNN | 1.9 ± 0.5 | 2.9 ± 0.5 | 5.5 ± 0.9 | |
RN-v1 | 2.0 ± 0.6 | 2.7 ± 0.6 | 4.6 ± 0.7 | |
RN-v2 | 2.0 ± 0.6 | 2.8 ± 0.6 | 4.8 ± 0.7 | |
DRN-v2 | 2.1 ± 1.3 | 2.7 ± 1.2 | 4.3 ± 1.1 | |
| ||||
Physics-informed | ANN | 1.0 ± 0.7 | 2.6 ± 1.0 | 6.0 ± 1.2 |
FCNN | 1.8 ± 0.6 | 2.9 ± 0.5 | 5.3 ± 0.7 | |
DFCNN | 1.8 ± 0.6 | 3.0 ± 0.6 | 5.4 ± 0.7 | |
RN-v1 | 1.9 ± 0.6 | 2.8 ± 0.6 | 5.1 ± 0.8 | |
RN-v2 | 1.9 ± 0.6 | 2.8 ± 0.6 | 5.3 ± 0.9 | |
DRN-v2 | 2.0 ± 0.6 | 3.0 ± 0.6 | 5.1 ± 1.2 |
References
- Balestriero R, Pesenti J and LeCun Y 2021. Learning in High Dimension Always Amounts to Extrapolation arXiv preprint arXiv:2110.09485
- Bär E, Lalonde A, Royle G, Lu H-M and Bouchard H 2017. The potential of dual-energy CT to reduce proton beam range uncertainties Medical Physics 44 2332–44 [DOI] [PubMed] [Google Scholar]
- Beaulieu L, Carlsson Tedgren Å, Carrier J-F, Davis SD, Mourtada F, Rivard MJ, Thomson RM, Verhaegen F, Wareing TA and Williamson JF 2012. Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation Medical Physics 39 6208–36 [DOI] [PubMed] [Google Scholar]
- Bebis G and Georgiopoulos M 1994. Feed-forward neural networks IEEE Potentials 13 27–31 [Google Scholar]
- Bichsel H. Passage of charged particles through matter. California Univ., Berkeley (USA). Dept. of Physics) 1969.
- Blumer A, Ehrenfeucht A, Haussler D and Warmuth MK 1987. Occam's Razor Information Processing Letters 24 377–80 [Google Scholar]
- Bourque AE, Carrier J-F and Bouchard H 2014. A stoichiometric calibration method for dual energy computed tomography Physics in Medicine and Biology 59 2059–88 [DOI] [PubMed] [Google Scholar]
- Champion K, Lusch B, Kutz JN and Brunton SL 2019. Data-driven discovery of coordinates and governing equations Proceedings of the National Academy of Sciences 116 22445. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chang C-W and Dinh NT 2019. Classification of machine learning frameworks for data-driven thermal fluid models International Journal of Thermal Sciences 135 559–79 [Google Scholar]
- Chang C-W, Huang S, Harms J, Zhou J, Zhang R, Dhabaan A, Slopsema R, Kang M, Liu T, McDonald M, Langen K and Lin L 2020. A standardized commissioning framework of Monte Carlo dose calculation algorithms for proton pencil beam scanning treatment planning systems Medical Physics 47 1545–57 [DOI] [PubMed] [Google Scholar]
- Dinh N 2013. Validation data to support advanced code development NURETH-15, American Nuclear Society, Pisa, Italy
- Domingos P 1999. The Role of Occam's Razor in Knowledge Discovery Data Mining and Knowledge Discovery 3 409–25 [Google Scholar]
- Fukushima K 1980. Neocognitron: A self-organizing neural network model for a mechanism of pattern recognition unaffected by shift in position Biological Cybernetics 36 193–202 [DOI] [PubMed] [Google Scholar]
- Garcia LIR, Azorin JFP and Almansa JF 2015. A new method to measure electron density and effective atomic number using dual-energy CT images Physics in Medicine and Biology 61 265–79 [DOI] [PubMed] [Google Scholar]
- Gjesteby L, Shan H, Yang Q, Xi Y, Jin Y, Giantsoudi D, Paganetti H, De Man B and Wang G 2019. A dual-stream deep convolutional network for reducing metal streak artifacts in CT images Physics in Medicine & Biology 64 235003. [DOI] [PubMed] [Google Scholar]
- He K, Zhang X, Ren S and Sun J 2016. IEEE Conference on Computer Vision and Pattern Recognition (CVPR),27–30 June 2016 2016), vol. Series; ) pp 770–8 [Google Scholar]
- Hochreiter S, Bengio Y, Frasconi P and Schmidhuber J 2001. Gradient flow in recurrent nets: the difficulty of learning long-term dependencies A Field Guide to Dynamical Recurrent Neural Networks, IEEE Press [Google Scholar]
- Hornik K, Stinchcombe M and White H 1989. Multilayer feedforward networks are universal approximators Neural Networks 2 359–66 [Google Scholar]
- Huang S, Souris K, Li S, Kang M, Barragan Montero AM, Janssens G, Lin A, Garver E, Ainsley C, Taylor P, Xiao Y and Lin L 2018. Validation and application of a fast Monte Carlo algorithm for assessing the clinical impact of approximations in analytical dose calculations for pencil beam scanning proton therapy Medical Physics 45 5631–42 [DOI] [PubMed] [Google Scholar]
- Hünemohr N, Krauss B, Tremmel C, Ackermann B, Jäkel O and Greilich S 2013. Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogates Physics in Medicine and Biology 59 83–96 [DOI] [PubMed] [Google Scholar]
- Hünemohr N, Paganetti H, Greilich S, Jäkel O and Seco J 2014. Tissue decomposition from dual energy CT data for MC based dose calculation in particle therapy Medical Physics 41 061714. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jackson DF and Hawkes DJ 1981. X-ray attenuation coefficients of elements and mixtures Physics Reports 70 169–233 [Google Scholar]
- Jimmy Lei Ba JRK, Hinton Geoffrey E. 2016. Layer Normalization arXiv:1607.06450
- Karniadakis GE, Kevrekidis IG, Lu L, Perdikaris P, Wang S and Yang L 2021. Physics-informed machine learning Nature Reviews Physics 3 422–40 [Google Scholar]
- Kassaee A, Cheng C, Yin L, Zou W, Li T, Lin A, Swisher-McClure S, Lukens JN, Lustig RA, O'Reilly S, Dong L, Ms RH and Teo B-KK 2021. Dual-Energy Computed Tomography Proton-Dose Calculation with Scripting and Modified Hounsfield Units International Journal of Particle Therapy 8 62–72 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Landry G, Seco J, Gaudreault M and Verhaegen F 2013. Deriving effective atomic numbers from DECT based on a parameterization of the ratio of high and low linear attenuation coefficients Physics in Medicine and Biology 58 6851–66 [DOI] [PubMed] [Google Scholar]
- LeCun Y, Bengio Y and Hinton G 2015. Deep learning Nature 521 436–44 [DOI] [PubMed] [Google Scholar]
- Liang X, Li Z, Zheng D, Bradley JA, Rutenberg M and Mendenhall N 2019. A comprehensive dosimetric study of Monte Carlo and pencil-beam algorithms on intensity-modulated proton therapy for breast cancer Journal of Applied Clinical Medical Physics 20 128–36 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liao H, Lin WA, Zhou SK and Luo J 2020. ADN: Artifact Disentanglement Network for Unsupervised Metal Artifact Reduction IEEE Transactions on Medical Imaging 39 634–43 [DOI] [PubMed] [Google Scholar]
- Lin L, Taylor PA, Shen J, Saini J, Kang M, Simone CB II, Bradley JD, Li Z and Xiao Y 2021. NRG Oncology Survey of Monte Carlo Dose Calculation Use in US Proton Therapy Centers International Journal of Particle Therapy 8 73–81 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mayneord W. The significance of the roentgen Acta Int Union Against Cancer. 1937;2:271. [Google Scholar]
- McCollough CH, Leng S, Yu L and Fletcher JG 2015. Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications Radiology 276 637–53 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nair V and Hinton GE 2010. Rectified linear units improve restricted boltzmann machines. In: Proceedings of the 27th International Conference on International Conference on Machine Learning, (Haifa, Israel: Omnipress; ) pp 807–14 [Google Scholar]
- Naitzat G, Zhitnikov A and Lim L-H 2020. Topology of Deep Neural Networks Journal of Machine Learning Research 21 1–40 [Google Scholar]
- Paganetti H 2012. Range uncertainties in proton therapy and the role of Monte Carlo simulations Physics in Medicine and Biology 57 R99–R117 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Paganetti H, Jiang H, Parodi K, Slopsema R and Engelsman M 2008. Clinical implementation of full Monte Carlo dose calculation in proton beam therapy Physics in Medicine and Biology 53 4825–53 [DOI] [PubMed] [Google Scholar]
- Paszke A a G, Sam and Massa, Francisco and Lerer, Adam and Bradbury, James and Chanan, Gregory and Killeen, Trevor and Lin, Zeming and Gimelshein, Natalia and Antiga, Luca and Desmaison, Alban and Kopf, Andreas and Yang, Edward and DeVito, Zachary and Raison, Martin and Tejani, Alykhan and Chilamkurthy, Sasank and Steiner, Benoit and Fang, Lu and Bai, Junjie and Chintala, Soumith 2019. PyTorch: An Imperative Style, High-Performance Deep Learning Library: Curran Associates, Inc. [Google Scholar]
- Polf JC, Mille MM, Mossahebi S, Chen H, Maggi P and Chen-Mayer H 2019. Determination of proton stopping power ratio with dual-energy CT in 3D-printed tissue/air cavity surrogates Medical Physics 46 3245–53 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ratner B 2011. Statistical and Machine-Learning Data Mining: Techniques for Better Predictive Modeling and Analysis of Big Data: CRC Press, p. 388 [Google Scholar]
- RayPhysics 2021. RayStation 10B Reference Manual RaySearch Laboratories AB; ) [Google Scholar]
- Rutherford RA, Pullan BR and Isherwood I 1976. Measurement of effective atomic number and electron density using an EMI scanner Neuroradiology 11 15–21 [DOI] [PubMed] [Google Scholar]
- Saini J, Maes D, Egan A, Bowen SR, St James S, Janson M, Wong T and Bloch C 2017. Dosimetric evaluation of a commercial proton spot scanning Monte-Carlo dose algorithm: comparisons against measurements and simulations Physics in Medicine & Biology 62 7659–81 [DOI] [PubMed] [Google Scholar]
- Schmidhuber J 2015. Deep learning in neural networks: An overview Neural Networks 61 85–117 [DOI] [PubMed] [Google Scholar]
- Schneider U, Pedroni E and Lomax A 1996. The calibration of CT Hounsfield units for radiotherapy treatment planning Physics in Medicine and Biology 41 111–24 [DOI] [PubMed] [Google Scholar]
- Schneider W, Bortfeld T and Schlegel W 2000. Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions Physics in Medicine and Biology 45 459–78 [DOI] [PubMed] [Google Scholar]
- Schuemann J, Giantsoudi D, Grassberger C, Moteabbed M, Min CH and Paganetti H 2015. Assessing the Clinical Impact of Approximations in Analytical Dose Calculations for Proton Therapy Int J Radiat Oncol Biol Phys 92 1157–64 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Seltzer SMF-VJM, Andreo P, Bergstrom PM Jr Burns DT, Krajcar Bronic I, Ross CK, Salvat F 2014. key data for ionizing-radiation dosimetry: measurement standards and applications ICRU Publication 90 [Google Scholar]
- Simonyan K and Zisserman A 2014. Two-stream convolutional networks for action recognition in videos. In: Proceedings of the 27th International Conference on Neural Information Processing Systems - Volume 1, (Montreal, Canada: MIT Press; ) pp 568–76 [Google Scholar]
- Spiers FW 1946. Effective Atomic Number and Energy Absorption in Tissues The British Journal of Radiology 19 52–63 [DOI] [PubMed] [Google Scholar]
- Su K-H, Kuo J-W, Jordan DW, Van Hedent S, Klahr P, Wei Z, Al Helo R, Liang F, Qian P, Pereira GC, Rassouli N, Gilkeson RC, Traughber BJ, Cheng C-W and Muzic RF 2018. Machine learning-based dual-energy CT parametric mapping Physics in Medicine & Biology 63 125001. [DOI] [PubMed] [Google Scholar]
- Wang F, Han J, Zhang S, He X and Huang D 2018. CSI-Net: Unified human body characterization and pose recognition arXiv preprint arXiv:1810.03064
- Wang T, Ghavidel BB, Beitler JJ, Tang X, Lei Y, Curran WJ, Liu T and Yang X 2019. Optimal virtual monoenergetic image in “TwinBeam” dual-energy CT for organs-at-risk delineation based on contrast-noise-ratio in head-and-neck radiotherapy Journal of Applied Clinical Medical Physics 20 121–8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wellenberg RHH, Donders JCE, Kloen P, Beenen LFM, Kleipool RP, Maas M and Streekstra GJ 2018a. Exploring metal artifact reduction using dual-energy CT with pre-metal and post-metal implant cadaver comparison: are implant specific protocols needed? Skeletal Radiology 47 839–45 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wellenberg RHH, Hakvoort ET, Slump CH, Boomsma MF, Maas M and Streekstra GJ 2018b. Metal artifact reduction techniques in musculoskeletal CT-imaging European Journal of Radiology 107 60–9 [DOI] [PubMed] [Google Scholar]
- Wohlfahrt P, Möhler C, Hietschold V, Menkel S, Greilich S, Krause M, Baumann M, Enghardt W and Richter C 2017. Clinical Implementation of Dual-energy CT for Proton Treatment Planning on Pseudo-monoenergetic CT scans International Journal of Radiation Oncology*Biology*Physics 97 427–34 [DOI] [PubMed] [Google Scholar]
- Wohlfahrt P, Möhler C, Richter C and Greilich S 2018. Evaluation of Stopping-Power Prediction by Dual- and Single-Energy Computed Tomography in an Anthropomorphic Ground-Truth Phantom Int J Radiat Oncol Biol Phys 100 244–53 [DOI] [PubMed] [Google Scholar]
- Yang M, Virshup G, Clayton J, Zhu XR, Mohan R and Dong L 2010. Theoretical variance analysis of single- and dual-energy computed tomography methods for calculating proton stopping power ratios of biological tissues Physics in Medicine and Biology 55 1343–62 [DOI] [PubMed] [Google Scholar]
- Yang M, Zhu XR, Park PC, Titt U, Mohan R, Virshup G, Clayton JE and Dong L 2012. Comprehensive analysis of proton range uncertainties related to patient stopping-power-ratio estimation using the stoichiometric calibration Physics in Medicine and Biology 57 4095–115 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yepes P, Adair A, Grosshans D, Mirkovic D, Poenisch F, Titt U, Wang Q and Mohan R 2018. Comparison of Monte Carlo and analytical dose computations for intensity modulated proton therapy Physics in Medicine & Biology 63 045003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yu L, Leng S and McCollough CH 2012. Dual-Energy CT–Based Monochromatic Imaging American Journal of Roentgenology 199 S9–S15 [DOI] [PubMed] [Google Scholar]
- Yu L, zhang Z, Li X, Ren H, Zhao W and Xing L 2021a. Metal artifact reduction in 2D CT images with self-supervised cross-domain learning Physics in Medicine & Biology [DOI] [PubMed]
- Yu L, Zhang Z, Li X and Xing L 2021b. Deep Sinogram Completion With Image Prior for Metal Artifact Reduction in CT Images IEEE Transactions on Medical Imaging 40 228–38 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhu J and Penfold SN 2016. Dosimetric comparison of stopping power calibration with dual-energy CT and single-energy CT in proton therapy treatment planning Medical Physics 43 2845–54 [DOI] [PubMed] [Google Scholar]