The Role of Novel Digital Clinical Tools in the Screening or Diagnosis of Obstructive Sleep Apnea: Systematic Review

Miguel Duarte 1,✉,#, Pedro Pereira-Rodrigues 1,2,3,#, Daniela Ferreira-Santos 1,2,3,#
Editor: Taiane de Azevedo Cardoso
Reviewed by: Zilu Liang, Manuel Casal-Guisande
PMCID: PMC10413091  PMID: 37494079

Abstract

Background

Digital clinical tools are a new technology that can be used in the screening or diagnosis of obstructive sleep apnea (OSA), notwithstanding the crucial role of polysomnography, the gold standard.

Objective

This study aimed to identify, gather, and analyze the most accurate digital tools and smartphone-based health platforms used for OSA screening or diagnosis in the adult population.

Methods

We performed a comprehensive literature search of PubMed, Scopus, and Web of Science databases for studies evaluating the validity of digital tools in OSA screening or diagnosis until November 2022. The risk of bias was assessed using the Joanna Briggs Institute critical appraisal tool for diagnostic test accuracy studies. The sensitivity, specificity, and area under the curve (AUC) were used as discrimination measures.

Results

We retrieved 1714 articles, 41 (2.39%) of which were included in the study. From these 41 articles, we found 7 (17%) smartphone-based tools, 10 (24%) wearables, 11 (27%) bed or mattress sensors, 5 (12%) nasal airflow devices, and 8 (20%) other sensors that did not fit the previous categories. Only 8 (20%) of the 41 studies performed external validation of the developed tool. Of these, the highest reported values for AUC, sensitivity, and specificity were 0.99, 96%, and 92%, respectively, for a clinical cutoff of apnea-hypopnea index (AHI)≥30. These values correspond to a noncontact audio recorder that records sleep sounds, which are then analyzed by a deep learning technique that automatically detects sleep apnea events, calculates the AHI, and identifies OSA. Looking at the studies that only internally validated their models, the work that reported the highest accuracy measures showed AUC, sensitivity, and specificity values of 1.00, 100%, and 96%, respectively, for a clinical cutoff AHI≥30. It uses the Sonomat—a foam mattress that, aside from recording breath sounds, has pressure sensors that generate voltage when deformed, thus detecting respiratory movements, and uses it to classify OSA events.

Conclusions

These clinical tools presented promising results with high discrimination measures (best results reached AUC>0.99). However, there is still a need for quality studies comparing the developed tools with the gold standard and validating them in external populations and other environments before they can be used in clinical settings.

Trial Registration

PROSPERO CRD42023387748; https://www.crd.york.ac.uk/prospero/display_record.php?RecordID=387748

Keywords: obstructive sleep apnea, diagnosis, digital tools, smartphone, wearables, sensor, polysomnography, systematic review, mobile phone

Introduction

Background

Obstructive sleep apnea (OSA) is a sleep-related breathing disorder characterized by transitory periods of breathing cessation due to partial (hypopnea) or complete (apnea) obstruction of the respiratory tract that affects ventilation during sleep. Repeated episodes of upper airway obstruction during sleep may, understandably, result in sleep fragmentation, nonrestorative sleep, and excessive daytime somnolence [1]. Furthermore, it has a great impact on multiple organ systems and is associated with hypertension, cardiovascular morbidities (eg, arrhythmias, coronary artery, and cerebrovascular diseases), and decrements in cognitive function [2].

The prevalence of this disease varies greatly depending on the population being studied and on how OSA is defined (eg, testing methodology, scoring criteria used, and apnea-hypopnea index [AHI] threshold) [3]. Despite the heterogeneity in population prevalence studies, this number is high, as nearly one-seventh or one billion of the world’s adult population is thought to have some degree of OSA [4-6]. These numbers have been on an upward trajectory, partially because of the increasing number of individuals with excess weight and obesity in high-income countries, as this is one of the causal factors of this pathology [6,7]. Nevertheless, despite being a highly prevalent disease, many cases remain undiagnosed and untreated, resulting in a decrease in quality of life, along with an increase in the incidence of adverse events and overall mortality [8].

The current gold-standard method for the diagnosis of OSA is overnight polysomnography, which takes place in a sleep laboratory with the attendance of a sleep technician [3], and where oxygen saturation, oronasal airflow, respiratory movement, electroencephalogram, body position, electromyogram, electrooculogram, and electrocardiogram are recorded [3,9].

Traditionally, sleep studies have been categorized into type I (or polysomnography), type II, type III, and type IV. Contrary to type I, types II to IV are unattended. Type II studies use the same number of monitoring sensors as the gold standard but are performed outside of the sleep laboratory, normally at the patient’s home. Unfortunately, they lack technical quality because problems such as sensor displacement or malfunction cannot be addressed by sleep technicians. Type III studies, also known as home sleep apnea tests (HSATs), have already been validated and use devices that measure otherwise limited cardiopulmonary parameters: 2 respiratory variables (eg, airflow and breathing effort), oxygen saturation, and a cardiac variable (eg, heart rate [HR] or electrocardiogram). Finally, type IV sleep studies are the most limited type of sleep study, using devices that measure only 1 or 2 parameters, typically HR or oxygen saturation [3].

Although the polysomnography provides detailed and highly accurate results, it is a time-consuming, labor-intensive, and expensive test [6], as it requires the patient to stay overnight in the sleep laboratory, a sleep technician to attend the study, and manual scoring of the data to produce the results, just to list some disadvantages [3]. This causes sleep laboratories to be unable to keep up with demand, often with long waiting lists and inaccessibility to a large part of the population [10].

The use of digital tools and innovative devices is a rapidly expanding area of research and has the potential to revolutionize the way health services are delivered, increasing access to health care in an easier way and at lower costs [11]. They can be an invaluable addition for health care professionals, as they provide many different functions, ranging from clinical decision support systems to data collection [12].

Wearable devices provide a level of unobtrusiveness that is not achievable with standard techniques, conceivably allowing faster OSA screening, along with improved long-term characterization and follow-up because of the possibility of day-to-day use. Subsequently, research on the use of these instruments in the diagnosis of OSA has been growing rapidly in recent years, with numerous vital signs and sleep parameters being monitored and strategies being used [13]. For example, some rely on movement analysis during sleep using accelerometers (actigraphy), snoring audio processing using tracheal and ambient microphones, and oxygen saturation measurement [14].

In contrast, smartphone-based health care platforms are emerging as an innovative solution owing to their ability to integrate, in the same device or in combination with other wireless wearable devices, several of the essential sensors to obtain the desired physiological variables for sleep-related disease diagnosis [14]. In addition, because of their ability to monitor sleep over long periods in the home setting, wrist-worn sleep devices, such as smartwatches and fitness trackers, are gaining attention from the sleep medicine community, using photoplethysmography, microphones, accelerometers, HR, or oximetry data [15]. Sheet-shaped or under-the-mattress sensors are also upcoming technologies that use pressure sensors for the detection of respiratory efforts based on accompanying thoracic movements [16], which spares the patient from being restrained by attached sensors and consequently allows a more natural and comfortable sleep experience.

Therefore, the integration of these contemporary and latest devices and platforms has the potential to improve patient care and grant better access to screening or diagnostic tests, allowing for quicker diagnosis, monitoring, and treatment of patients with OSA [13].

Objective

Given the shortcomings of the current gold standard and the promising features of the new innovative digital clinical tools, this systematic review aimed to identify, gather, and analyze the most accurate digital tools used for OSA screening or diagnosis in the adult population. We intend to identify individuals with a higher risk of developing this disease, which would benefit the most from a full in-laboratory polysomnography to confirm the diagnosis, relieving some of the pressure on this field, following a rule-in approach.

Methods

This systematic review was carried out according to the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) guidelines [17], and the protocol was registered in the PROSPERO under the reference CRD42023387748.

Search Strategy and Selection Criteria

A comprehensive literature search, without any restrictions, was conducted using the PubMed, Scopus, and Web of Science databases for articles published until November 2022. Specific queries were used for each platform, which can be found in Multimedia Appendix 1. Subsequently, a manual search was performed using relevant references from the included studies and relevant reviews on the matter. If there was no access to the full-text article on the web, the respective authors were contacted to obtain it.

Articles were independently selected by 2 reviewers (blinded to each other’s assessment, MD and DFS), applying predefined criteria to each article’s title and abstract, and in the second phase, to the integral texts of the selected articles. Divergent opinions were resolved by consensus. These processes were conducted using Rayyan (Qatar Computing Research Institute), a web and mobile app that helps expedite the initial screening of articles for systematic reviews [18].

Included in this review were studies that reported on adult patients with suspected OSA or OSA diagnosis (population) and assessed the accuracy of digital clinical tools for the screening or diagnosis of OSA (exposure and comparator) while having polysomnography as a gold standard (outcome). Studies that evaluated the accuracy of digital tools in pregnant women or the pediatric population that used HSAT or other types of sleep studies as the reference test, as well as interventions using only one portion of the data obtained by polysomnography as the index test (eg, pulse oximetry, electroencephalography, and electrocardiogram), were excluded.

Data Extraction

Once the articles were selected, data were extracted to a prespecified Microsoft Excel spreadsheet by 2 reviewers (MD and DFS) blinded to each other’s assessment, which included the following: (1) article information: title, authors, publication date, country, and journal and (2) methods: participant selection, sample size, execution or nonexecution of in-laboratory polysomnography, prevalence of OSA, type of digital tool analyzed in comparison with polysomnography, inclusion and exclusion criteria, and potential bias.

To enhance the comprehension and analysis of the data from all included studies, we grouped the digital clinical tools into five categories: (1) smartphone-based tools, (2) wearable tools, (3) bed or mattress sensors, (4) nasal airflow devices, and (5) other digital tools. The last category was created to avoid further subdivisions, as some tools did not fit the previous ones.

For each type of tool, specific data were extracted, including population (n), clinical cutoff values for the diagnosis of OSA and severity classification (AHI), area under the curve (AUC), sensitivity, and specificity values.

Moreover, an additional division was made regarding the validation of the developed tool. If the studies merely developed a digital tool or developed and tested it on the same collected population, the results were presented for that group and represented by “D”—derivation group—in the tables. If the study was developed and validated on a different population, those results were presented and represented by “V”—validation group. In addition, a further subdivision could be performed regarding subject- or event-wise validation because the results obtained from each method should not be directly compared. Even so, given the few included studies that performed event-wise validation, this subdivision was not considered. However, these studies are mentioned with a footnote in the respective tables.

Studies were presented by the year of publication within each category. Any missing information is reported in the tables of the Results section by “—” (meaning “not available”), and the AHI cutoff for which the best metrics were obtained is marked in italics. A special note to one article presents the results for several sensor positions, where we only show the best results. In addition, each manuscript was checked for the definition of hypopnea, namely, the percentage of desaturation, but as this definition was not clear in most of the studies, we did not consider it further.

Finally, as we intended to identify and select patients with a high probability of having OSA suitable to perform polysomnography, tools with high specificity values were considered the best, following a rule-in approach.

Risk of Bias

At 2 points in time, 2 reviewers (MD and DFS) assessed the risk of bias in all 41 included studies. It was performed by analyzing and answering a total of 10 questions from the Joanna Briggs Institute critical appraisal tool for diagnostic test accuracy studies [19]. All answers can be found in the tables and are represented by symbols according to their risk of bias. A green minus sign is presented in the table if a low risk of bias was found for a question. A red plus sign is presented if a substantial risk of bias is found. A yellow question mark is presented if the risk of bias was unclear. If the question did not apply to our specifically analyzed studies, they were indicated as not applicable.

Regarding the questions, each was replaced with the letter Q in the tables, followed by the number of the question: Q1, “Was a consecutive or random sample of patients enrolled?” Q2, “Was a case-control design avoided?” Q3, “Did the study avoid inappropriate exclusions?” Q4, “Were the index test results interpreted without knowledge of the results of the reference standard?” Q5, “If a threshold was used, was it prespecified?” Q6, “Is the reference standard likely to correctly classify the target condition?” Q7, “Were the reference standard results interpreted without knowledge of the results of the index test?” Q8, “Was there an appropriate interval between the index test and reference standard?” Q9, “Did all patients receive the same reference standard?” and Q10, “Were all patients included in the analysis?”

Given that the index test (digital tool) and the gold standard (polysomnography) were performed simultaneously in all studies included, Q8 was not applicable to any of the designs.

Results

Overview

We retrieved 1714 articles, of which 477 were duplicates. From the 1237 articles, after in-depth scrutiny, we retained 41 papers that met the inclusion criteria, as shown in Figure 1.

Figure 1.

Figure 1

Flow diagram showcasing the article selection process. PSG: polysomnography.

Disagreements were observed between the reviewers during both phases of the analysis. The overall rate of concordance in the title and abstract screening was 87%, whereas that in the integral version was 92%.

The gold-standard diagnostic test, polysomnography, was performed in all the studies included in our review. Some studies were unclear about the overall context in which the polysomnography was performed and did not report any details about the polysomnography data collection (eg, setting, equipment used, number of channels, and overall conditions); therefore, we excluded them from the analysis.

The oldest digital clinical tool was developed in 2002 and consists of a small, lightweight device worn underneath the nose and above the upper lip, which identifies and counts nasal airflow cessations through the night and uses it to predict the probability of OSA [20]. In contrast, in 2022, a variety of tools were developed and tested, namely a smartphone-based method [21], a wearable adhesive patch [22], 2 radar devices [23,24], and an audio recorder [25]. The frequency distribution of the tools included in our systematic review is shown in Table 1.

Table 1.

Frequency distribution of the tools included in our systematic review (n=41).

Digital tools Number of articles, n (%)
Smartwatch 4 (10)
Smart pillow 1 (2)
Smartphone 7 (17)
Ring 1 (2)
Respiration belt 1 (2)
Radar 5 (12)
Neck device 1 (2)
Nasal air flow 5 (12)
Microphones 2 (5)
Mattress sensor 10 (24)
Garment 1 (2)
Bone conducted vibration 1 (2)
Adhesive patch 2 (5)

Overall, the largest sample size was 620 patients [26], whereas the smallest consisted of only 5 patients [27], with a median sample size of 53 patients with suspected OSA. The overall prevalence of OSA was reported in 21 of the 41 included studies, with values ranging from 51% [28] to 100% [29].

AHI was used to define OSA presence or absence, as well as to stratify patients by severity level, with the chosen cutoff in each paper stated in the respective table. A considerable number of studies did not report the cutoff value (12/41, 29%), whereas among those who chose only 1 definition cutoff, the most frequent was AHI≥15 (8/41, 19%), followed by AHI≥5 (3/41, 7%). One study [30] used a different cutoff from the rest, AHI≥20, whereas another study [31] performed a comparison between the different severity-level subgroups using AHI=5, AHI=15, and AHI=30 instead of comparing the presence or absence of the disease for a specific AHI cutoff, thus impairing the comparison with the rest. The remaining studies (17/41, 41%) presented several severity cutoffs and corresponding results for each, often AHI≥5, AHI≥15, and AHI≥30.

Smartphone-Based Tools

Of the 41 included articles, 7 (17%) analyzed the screening or diagnostic ability of smartphone-based tools (Table 2) by using 1 or more of the sensors incorporated in the smartphones. A total of 3 studies used a microphone to capture audio signals to detect the patient’s respiratory effort, which were then processed and analyzed to predict the diagnosis [21,32,33], with 1 study [33] concomitantly using the smartphone’s accelerometer. In contrast, 2 studies created apps as a form of screening using models that can predict the risk of OSA from patient variables [26] or process signals to detect patterns of sleep-disordered breathing [28]. The remaining 2 studies transformed the smartphone into a sonar device that emits sound and then captures its reflection, using it to classify the respiratory movements of the patient [34,35].

Table 2.

Study characteristics of the smartphone-based tools categorya.

Study, year Digital tool OSAb definition Sample size, n (group type: D=derivation group, V=validation group) OSA prevalence, n (%) Area under the curve, (95% CI) Sensitivity, % (95% CI) Specificity, % (95% CI)
Al-Mardini et al [33], 2014 Smartphone (audio; accelerometer) c 15 (D) 8 (53) 100 (—) 86 (—)
Nandakumar et al [34], 2014 Smartphone (sonar; audio) 37 (D) 26 (70)
Narayan et al [32], 2018 Smartphone (audio) AHId≥15 32 (D); 59 (V) 48 (53) 0.87 (—) 94 (—) 63 (—)
Lyon et al [28], 2019 Smartphone (app; sonar) AHI≥15 162 (D) 83 (51) 0.95 (—) 94 (—) 97 (—)
Haberfeld et al [26], 2020 Smartphone (app) 620 (D) 357 (58) Male: 0.61 (—);
female: 0.62 (—)
Male: 86 (—);
female: 56 (—)
Tiron et al [35], 2020 Smartphone (sonar; audio) AHI≥15 128 (D); 120 (V) 0.92 (0.85-0.95) 88 (77-95) 80 (68-89)
Castillo-Escario et al [21], 2022e Smartphone (audio) 16 (D) 0.88 (—) 72 (—) 89 (—)

aWhen external validation was performed, the results are only presented for the validation group.

bOSA: obstructive sleep apnea.

cNot available.

dAHI: apnea-hypopnea index.

eUsed event-wise validation.

OSA definition was stated in only 43% (3/7) of studies, all of which used an AHI≥15 cutoff. The largest sample size recorded was 620 patients, whereas the smallest sample comprised only 15 patients. The prevalence of OSA was reported in 71% (5/7) of articles, ranging from 51% to 70%. Regarding the studies that only performed internal validation, the best AUC, sensitivity, and specificity values were 0.95, 94%, and 97%, respectively, for an AHI≥15 cutoff [28]. As for the studies that performed external validation, 2 studies were found, namely, Narayan et al [32] that reported AUC, sensitivity, and specificity values of 0.87, 94%, and 63%, respectively, and Tiron et al [35] that reported values of 0.92 (95% CI 0.85-0.95), 88% (95% CI 67%-95%), and 80% (95% CI 68%-89%), respectively.

Table 3 shows the risk of bias for studies within this category. In the patient selection domain, 1 study [33] had a high risk of bias, 3 studies had a low risk of bias [21,28,34], and 3 others had an unclear risk of bias [26,32,35]. For the index test domain, 4 studies presented a high risk [21,26,33,34], namely in questions Q4 and Q5, with 2 studies having a low risk [28,32], and 1 study being unclear [35].

Table 3.

Risk of bias for smartphone-based tools category using the Joanna Briggs Institute critical appraisal checklist for diagnostic test accuracy studies.

Study, year Risk of bias

Patient selection Index test

Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 Q9 Q10
Al-Mardini et al [33], 2014 Inline graphic a Inline graphic b Inline graphic c graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/Ad graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Nandakumar et al [34], 2014 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Narayan et al [32], 2018 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Lyon et al [28], 2019 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Haberfeld et al [26], 2020 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Tiron et al [35], 2020 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Castillo-Escario et al [21], 2022 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg

aIndicates unclear risk of bias.

bIndicates high risk of bias.

cIndicates low risk of bias.

dN/A: not applicable.

Wearable Tools

Table 4 shows the 10 (24%) of the 41 wearable tools included in our systematic review. The device used in 4 of them was a smartwatch, making use of their built-in sensors, such as photoplethysmography signals or an accelerometer [36-39]. For the other wearables, there was a garment or body-worn sensor that recorded electrocardiogram, thoracic movements, and positional signals [40]; a respiration belt that registered thoracic movement during respiration [41]; and a neck device that recorded snoring and position using a microphone and accelerometer, respectively [42]. In addition, 2 adhesive patches were included: one that recorded electrocardiogram and actigraphy signals that were patched to the chest [43], and one that recorded blood saturation on the hand using a photoelectric reflex sensor [22]. Finally, a ring device capturing oxygen saturation, photoplethysmography signals, and accelerometer data are also shown in this table [44].

Table 4.

Study characteristics of the wearable tools categorya.

Study, year Digital tool OSAb definition Sample size, n (group type: D=derivation group, V=validation group) OSA prevalence, n (%) Area under the curve (95% CI) Sensitivity, % (95% CI) Specificity, % (95% CI)
Levendowski et al [42], 2014 Neck device AHIc>5d; AHI≥10; AHI≥15 20 (D); 24 (V) e 78 (—)d;
81 (—);
87 (—)
92 (—)d;
87 (—);
88 (—)
Selvaraj and Narasimhan [43], 2014 Adhesive patch sensors (chest) AHI≥15 53 (D) 32 (60) 94 (83-98) 79 (65-88)
Ben Azouz et al [40], 2018 Garment or body-worn sensor 32 (D)
Hayano et al [39], 2020 Smartwatch AHI>15 41 (D) 22 (54) 82 (—) 89 (—)
Chen et al [37], 2021 Smartwatch 20 (D) 96 (—)
Ganglberger et al [41], 2021 Respiration belt 404 (D) 248 (61) 0.97 (—)
Chen et al [38], 2021 Smartwatch AHI≥5; AHI≥15d; AHI≥30 20 (D) 17 (85) 0.81 (—);
0.93 (—)d;
0.80 (—)
77 (50-93);
86 (57-98)d;
80 (44-98)
100 (29-100);
100 (54-100)d;
80 (44-98)
Yeh et al [44], 2021 Ring AHI≥5; AHI≥15; AHI≥30d 78 (D) 43 (55) 0.93 (0.88-0.98);
0.96 (0.90-1.00);
0.96 (0.92-1.00)d
100 (92-100);
93 (77-99);
71 (42-92)d
3 (0-15);
74 (59-85);
95 (87-99)d
Fedorin and Slyusarenko [36], 2021 Smartwatch AHI>15 107 (D); 69 (V)
Xu et al [22], 2022 Adhesive patch sensor (palm) AHI≥5; AHI≥15d 196 (D) 0.95 (0.91-0.98);
0.95 (0.92-0.98)d
93 (—);
92 ()d
77 (—);
89 ()d

aIf the study used different clinical cutoff values for obstructive sleep apnea diagnosis, the results are only presented for the best-achieved cutoff and marked in italics. When external validation was performed, the results are only presented for the validation group.

bOSA: obstructive sleep apnea.

cAHI: apnea-hypopnea index.

dBest-achieved cutoff.

eNot available.

OSA definition was stated in 70% (7/10) of studies. The AHI≥15 cutoff was used in 3 of them, whereas the rest presented 2 or 3 severity cutoffs (AHI≥5, AHI≥15, and AHI≥30). The largest sample size was 404 patients, whereas the smallest had only 20 patients. The prevalence of OSA was reported in 50% (5/10) of articles and varied between 54% and 85%. Considering studies that only performed internal validation, the best AUC, sensitivity, and specificity values were 0.93, 86% (95% CI 57%-98%), and 100% (95% CI 54%-100%), respectively, for an AHI≥15 cutoff [38]. Once again, 2 studies [36,42] also performed external validation in a distinct group. Levendowski et al [42] reported sensitivity and specificity values of 78% and 92%, respectively, for AHI≥5 cutoff. In the study by Fedorin and Slyusarenko [36], these measures have not been reported. As previously stated, in 1 study, the authors presented results for 3 positions of the sensor, but we chose to include only the best position [43].

In Table 5, we can see that, for the patient selection domain, only 1 study showed a high risk of bias [43], with 6 studies having a low risk [22,36,37,40,41,44], and 3 studies being unclear [38,39,42]. As for the index test domain, 6 studies presented a high risk [36-38,40-42] among questions Q5 and Q10, and the other 4 studies showed a low risk of bias [22,39,43,44].

Table 5.

Risk of bias for wearable tools category using Joanne Briggs Institute critical appraisal checklist for diagnostic test accuracy studies.

Study, year Risk of bias

Patient selection Index test

Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 Q9 Q10
Levendowski et al [42], 2014 Inline graphic a graphic file with name jmir_v25i1e47735_fig4.jpg Inline graphic b graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/Ac graphic file with name jmir_v25i1e47735_fig4.jpg Inline graphic d
Selvaraj and Narasimhan [43], 2014 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Ben Azouz et al [40], 2018 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Hayano et al [39], 2020 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Chen et al [37], 2021 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Ganglberger et al [41], 2021 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Chen et al [38], 2021 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg
Yeh et al [44], 2021 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Fedorin and Slyusarenko [36], 2021 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg
Xu et al [22], 2022 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg

aIndicates low risk of bias.

bIndicates unclear risk of bias.

cN/A: not applicable.

dIndicates high risk of bias.

Bed or Mattress Sensors

Bed or mattress-based sensors, for which the description can be found in Table 6, are also a big part of our pool of articles, with 11 (27%) of the 41 studies analyzing the performance and screening capability of these pressure-based sensing devices in the context of OSA [16,30,45-52]. In addition to under-the-mattress sensors, there is also a smart pillow paired with an oximeter, which offers diagnostic capability by detecting apnea events and interventional intent by being able to inflate and change its conformation and deobstruct the patient’s airway [29].

Table 6.

Study characteristics of the bed or mattress sensors categorya.

Study, year Digital tool OSAb definition Sample size, n (group type: D=derivation group, V=validation group) OSA prevalence, n (%) Area under the curve, (95% CI) Sensitivity, % (95% CI) Specificity, % (95% CI)
Agatsuma et al [16], 2009 Under-the-mattress sensor AHIc≥5; AHI≥15; AHI≥30d 201 (D); 165 (V) e 100 (—);
100 (—);
92 ()d
41 (—);
90 (—);
98 (—)d
Tenhunen et al [52], 2013 Under-the-mattress sensor AHI>5; AHI>15d; AHI>30 157 (D) 77 (—);
95 ()d;
94 (—)
81 (—);
92 (—)d;
82 (—)
Zhang et al [29], 2013f Smart pillow + oximeter 40 (D) 40 (100)
Tsukahara et al [30], 2014 Under-the-mattress sensor AHI≥20 101 (D) 90 (—) 90 (—)
Hwang et al [46], 2014 Under-the-mattress sensor AHI >5; AHI>15d; AHI>20 32 (D) 26 (81) 0.98 (—);
0.99 (—)d;
0.98 (—)
100 (—);
100 (—)d;
92 (—)
75 (—);
92 (—)d;
92 (—)
Norman et al [50], 2014 Under-the-mattress sensor AHI≥5; AHI≥15; AHI≥30d 43 (D) 35 (81) 0.94 (—);
0.97 (—);
1.00 (—)d
94 (—);
88 (—);
100 (—)d
77 (—);
91 (—);
96 (—)d
Mora et al [48], 2015 Under-the-mattress sensor AHI≥5 24 (D) 13 (54)
Meng et al [47], 2016 Under-the-mattress sensor AHI≥5d; AHI≥15; AHI≥30 131 (D) 0.98 (—)d;
0.98 (—);
0.98 (—)
95 (—)c;
90 (—);
90 (—)
100 (—)d;
97 (—);
95 (—)
Davidovich et al [45], 2016 Under-the-mattress sensor AHI≥15 96 (D) 64 (67) 88 (—) 89 (—)
Mosquera-Lopez et al [49], 2019 Under-the-mattress sensor AHI>5 14 (D) 8 (57) 89 (—) 77 (—)
Sadek et al [51], 2020f Under-the-mattress sensor 10 (D) 57 (—) 45 (—)

aIf the study used different clinical cutoff values for the diagnosis of obstructive sleep apnea, the results are only presented for the best-achieved cutoff and marked in italics. When external validation was performed, the results are only presented for the validation group. Studies that used event-wise validation are mentioned in a footnote.

bOSA: obstructive sleep apnea.

cAHI: apnea-hypopnea index.

dBest-achieved cutoff.

eNot available.

fUsed event-wise validation.

OSA definition was stated in 82% (9/11) of the studies. The majority presented 3 severity cutoffs (AHI≥5, AHI≥15, and AHI≥30), with 1 study using values of AHI>5, AHI>15, and AHI>20. For the rest, 2 studies used a cutoff of AHI≥5, 1 study used AHI≥15, and the other used AHI≥20. The largest sample size recorded was 366 patients, whereas the smallest sample size consisted of only 10. The prevalence of OSA was reported in 55% (6/11) of articles, ranging from 54% to 100%. With respect to studies that only performed internal validation, the best AUC, sensitivity, and specificity values were 1.00, 100%, and 96%, respectively, for an AHI≥30 cutoff [50]. This study used the Sonomat, a foam mattress that, aside from recording breath sounds, has pressure sensors that generate voltage when deformed, thus detecting respiratory movements, and using it to classify OSA events. In this category, only 1 study externally validated the device; Agatsuma et al [16] reported sensitivity and specificity values of 92% and 98%, respectively, for an AHI≥30 cutoff, with no mention of the AUC value.

Table 7 shows, in the patient selection domain, 3 studies with a high risk of bias [29,47,50], 6 with a low risk [16,30,45,48,49,51], and 2 with unclear risk [46,52]. In turn, in the index test domain, 4 studies had a high risk [29,47,51,52], namely in questions Q5 and Q10, and the other 7 studies presented a low risk of bias [16,30,45,46,48-50].

Table 7.

Risk of bias for bed or mattress sensors category using the Joanna Briggs Institute critical appraisal checklist for diagnostic test accuracy studies.

Study, year Risk of bias

Patient selection Index test

Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 Q9 Q10
Agatsuma et al [16], 2009 Inline graphic a graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/Ab graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Tenhunen et al [52], 2013 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg Inline graphic c graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg Inline graphic d
Zhang et al [29], 2013 graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg
Tsukahara et al [30], 2014 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Hwang et al [46], 2014 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg
Norman et al [50], 2014 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Mora et al [48], 2015 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Meng et al [47], 2016 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg
Davidovich et al [45], 2016 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Mosquera-Lopez et al [49], 2019 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Sadek et al [51], 2020 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg

aIndicates low risk of bias.

bN/A: not applicable.

cIndicates unclear risk of bias.

dIndicates high risk of bias.

Nasal Airflow Devices

In Table 8, we list the 5 (12%) nasal airflow devices out of the 41 tools. Of all the studies, 3 tested the accuracy of an under-the-nose pressure sensor named SleepStrip for detecting sleep events and diagnosing OSA [20,53,54]. Another study analyzed an under-the-nose pressure sensor that has not yet been marketed [27]. Finally, one piece of equipment was used to measure the nasal airflow using a nasal cannula [55].

Table 8.

Study characteristics of nasal airflow devices categorya.

Study, year Digital tool OSAb definition Sample size, n (group type: D=derivation group, V=validation group) OSA prevalence, n (%) Area under the curve, (95% CI) Sensitivity, % (95% CI) Specificity, % (95% CI)
Shochat et al [20], 2002 Under-the-nose pressure sensor SleepStrip AHIc>10; AHI>20; AHI>40d 288 (D) e 86 (—);
80 (—);
80 (—)e
57 (—);
70 (—);
86 (—)d
Wong et al [55], 2008 Nasal cannula 34 (D)
Ozmen et al [54], 2011 Under-the-nose pressure sensor SleepStrip AHI>10; AHI>20; AHI>40d 64 (D)
0.80 (0.68-0.89);
0.84 (0.72-0.92);
0.94 (0.85-0.98)d
83 (—);
80 (—);
94 (—)d
77 (—);
87 (—);
94 (—)d
Dinç et al [53], 2014 Under-the-nose pressure sensor SleepStrip AHI≥5; AHI≥15; AHI>30d 41 (D) 33 (81) 0.77 (0.61-0.94);
0.82 (0.73-1.00);
0.91 (0.79-1.00)d
54 (—);
44 (—);
45 (—)d
100 (—);
100 (—);
100 (—)d
Jin and Sánchez-Sinencio [27], 2015 Under-the-nose pressure sensor 5 (D)

aIf the study used different clinical cutoff values for the diagnosis of obstructive sleep apnea, the results are only presented for the best-achieved cutoff and marked in italics. When external validation was performed, the results are only presented for the validation group.

bOSA: obstructive sleep apnea.

cAHI: apnea-hypopnea index.

dBest-achieved cutoff.

eNot available.

OSA definition was only stated in the 3 studies using SleepStrip, with 1 of the studies using the usual cutoffs of AHI≥5, AHI≥15, and AHI≥30, and the other 2 studies using different cutoffs (AHI>10, AHI>20, and AHI>40). The largest sample size recorded was 288 patients, whereas the smallest sample comprised only 5. Only 1 (20%) out of the 5 studies reported values for the prevalence of OSA, registering 81% [53]. Contrary to the previous categories, none of the included studies performed external validations. The best reported AUC, sensitivity, and specificity values were 0.94 (95% CI 0.85-0.98), 94%, and 94%, respectively, for a clinical cutoff of AHI>40 [54].

This category showed the highest risk of bias (Table 9). All studies showed a high [54] or unclear [20,27,53,55] risk of bias in the patient selection domain. In the index test domain, there was only 1 study [53] with a low risk, whereas 4 studies [20,27,54,55] had a high risk of bias spanning questions Q5 and Q10.

Table 9.

Risk of bias for nasal airflow devices category using the Joanna Briggs Institute critical appraisal checklist for diagnostic test accuracy studies.

Study, year Risk of bias

Patient selection Index test

Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 Q9 Q10
Shochat et al [20], 2002 Inline graphic a graphic file with name jmir_v25i1e47735_fig4.jpg Inline graphic b graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/Ac graphic file with name jmir_v25i1e47735_fig4.jpg Inline graphic d
Wong et al [55], 2008 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg
Ozmen et al [54], 2011 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg
Dinç et al [53], 2014 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Jin and Sánchez-Sinencio [27], 2015 graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig2.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg

aIndicates low risk of bias.

bIndicates unclear risk of bias.

cN/A: not applicable.

dIndicates high risk of bias.

Other Digital Tools

A total of 8 (20%) out of the 41 studies included tools that did not fit the previous categories, which are shown in Table 10. Among these studies, there are 5 that used radar technology to screen for OSA, which are novel devices in the shape of noncontact bedside sensors that use radio waves to detect and measure thoracic movement and respiration [23,24,56-58]. Audio recording using noncontact microphones was also featured in 2 studies, with posterior sleep sound analysis using algorithms and deep learning methods [25,59]. In addition, 1 study detected snoring using an unconventional method by capturing its vibration using a bone-conducted transducer [31].

Table 10.

Study characteristics of other digital tools categorya.

Study, year Digital tool OSAb definition Sample size, n (group type: D=derivation group, V=validation group, T=test group, V=validation group) OSA prevalence, n (%) Area under the curve, (95% CI) Sensitivity, % (95% CI) Specificity, % (95% CI)
Alshaer et al [59], 2013 Face frame with a microphone attached AHIc≥10; AHI≥15d 32 (D) e 100 (—);
89 (—)d
85 (—);
96 (—)d
Zaffaroni et al [57], 2013 Radar AHI≥5; AHI≥15d; AHI≥30 74 (D) 60 (81) 0.90 (—);
0.97 (—)d;
0.96 (—)
98 (—);
90 (—)d;
84 (—)
47 (—);
92 (—)d;
89 (—)
Crinion et al [56], 2020 Radar AHI≥15 67 (D); 55 (V) 53 (79) 0.85 (—) 88 (—) 67 (—)
Xin et al [31], 2021 Bone-conducted transducer AHI=5; AHI=15d; AHI=30 28 (D) 23 (82) 0.91 (—);
1.00 (—)d;
1.00 (—)
91 (—);
100 (—)d;
92 (—)
100 (—);
100 (—)d;
100 (—)
Zhao et al [58], 2021 Radar—OrbSense AHI≥5; AHI≥15; AHI≥30d 359 (D) 0.90 (—);
0.94 (—);
0.97 (—)d
96 (93-98);
90 (84-93);
89 (81-93)d
56 (44-68);
81 (74-87);
94 (90-97)d
Wei et al [23], 2022 Radar; ring AHI≥5 67 (D) 57 (85) 0.85 (—) 100 (—) 70 (—)
Zhuang et al [24], 2022f Radar 10 (D); 2 (V)
Wang et al [25], 2022 Audio recorder AHI≥5; AHI≥15; AHI≥30d 116 (D); 19 (T); 59 (V) 0.94 (—);
0.98 (—);
0.99 (—)d
94 (—);
89 (—);
96 (—)d
83 (—);
96 (—);
92 (—)d

aIf the study used different clinical cutoff values for the diagnosis of obstructive sleep apnea, the results are only presented for the best-achieved cutoff and marked in italics. When external validation was performed, the results are only presented for the validation group. Studies that used event-wise validation are mentioned in a footnote.

bOSA: obstructive sleep apnea.

cAHI: apnea-hypopnea index.

dBest-achieved cutoff.

eNot available.

fUsed event-wise validation.

OSA definition was stated in 88% (7/8) of the studies. Some of them used severity cutoffs to define OSA, with 4 studies using the usual 3 cutoff points (AHI≥5, AHI≥15, and AHI≥30), and 1 study using only 2 cutoffs (AHI≥10 and AHI≥15). A total of 1 study used AHI≥5 as the OSA definition and another study used AHI≥15. The largest sample size recorded was 359 patients, whereas the smallest had only 12 patients. The prevalence of OSA was reported in 50% (4/8) of the articles and varied between 79% and 85%. With regard to the studies that only performed internal validation, the best AUC, sensitivity, and specificity values were 0.97, 89% (95% CI 81%-93%), and 94% (95% CI 90%-97%), respectively, for a cutoff of AHI≥30 [58]. External validation was performed in 3 studies within this category. One of them applied the tool to a cohort of male participants aged between 18 and 70 years, with a clinical diagnosis of hypertension and receiving antihypertensive medication [56]. The other 2 studies [24,25] validated the tools in a different group of 59 and 2 patients with similar characteristics as the derivation group. Crinion et al [56] reported AUC, sensitivity, and specificity values of 0.85, 88%, and 67%, respectively, for an AHI≥15 cutoff, whereas the study by Wang et al [25] presented values of 0.99, 96%, and 92%, respectively, for an AHI≥30. By contrast, the study by Zhuang et al [24] did not report any discrimination measures.

The studies included in this category had a lower risk of bias in the entire assembly, as shown in Table 11. In the patient selection domain, the risk of bias was negligible, whereas in the index test domain, only 2 studies faltered in questions Q5 [24] and Q10 [59].

Table 11.

Risk of bias for other digital tools category using the Joanna Briggs Institute critical appraisal checklist for diagnostic test accuracy studies.

Study, year Risk of bias

Patient selection Index test

Q1 Q2 Q3 Q4 Q5 Q6 Q7 Q8 Q9 Q10
Alshaer et al [59], 2013 Inline graphic a graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/Ab graphic file with name jmir_v25i1e47735_fig4.jpg Inline graphic c
Zaffaroni et al [57], 2013 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Crinion et al [56], 2020 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Xin et al [31], 2021 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Zhao et al [58], 2021 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Wei et al [23], 2022 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Zhuang et al [24], 2022 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig3.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg
Wang et al [25], 2022 graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg N/A graphic file with name jmir_v25i1e47735_fig4.jpg graphic file with name jmir_v25i1e47735_fig4.jpg

aIndicates low risk of bias.

bN/A: not applicable.

cIndicates high risk of bias.

Discussion

Principal Findings

As previously stated, this review aimed to gather the available evidence on upcoming digital tools in the screening or diagnosis of OSA, with a total of 41 tools that presented promising results, showing high discrimination measures (best results reaching AUC values higher than 0.99). This was done by grouping digital tools based on the technologies used, making discrimination measures comparable. Furthermore, we did not intend to replace the current gold standard, polysomnography, as the American Academy of Sleep Medicine guideline recommendations explicitly state that clinical tools...should not be used to diagnose OSA in adults in the absence of polysomnography or home sleep apnea testing [3]. However, digital devices and other mobile health tools can play an important adjuvant role in this process, which is also recognized by the American Academy of Sleep Medicine guidelines. It states that in non-sleep clinic settings, these tools may be more helpful to identify patients who are at increased risk for OSA [3] proving our rule-in approach. Recent reviews have assessed the potential use of digital tools in sleep-breathing disorders. The study by Behar et al [60] reviewed existing smartphone apps being used, particularly in OSA screening. They focused essentially on studies that applied questionnaires via an app or that used built-in smartphone sensors and characteristics, such as the accelerometer and the ability to record sleep sounds. However, it lacks a comparison of these proposed smartphone-based tools with the gold standard and the respective discrimination measures. The study by Kim et al [61] concentrated on the reliability of smartphones in the screening of moderate to severe OSA. In addition to the fact that our review covers a more versatile set of digital clinical tools, we also considered all diagnostic cutoffs, thus evaluating the use of these tools in the screening or diagnosis of all levels of OSA. We also found 2 other systematic reviews with similar aims to ours, but they only featured articles published until 2017. The studies by Mendonça et al [9] and Rosa et al [11] included an extensive array of new digital tools, some still in the research project phase, and others already commercially available. Nevertheless, both reviews included studies without polysomnography as the gold standard, allowing the use of HSAT as a comparison and reference test.

Of the 41 included studies, 7 were smartphone-based tools; 10 were wearables; 11 used bed or mattress sensors; 5 measured nasal airflow; and 8 used other technologies such as radar devices, adhesive patches, or microphones. Out of all of them, only 8 performed external validation of the developed digital tool, whereas 27, the majority, merely performed internal validation. In addition, it is worth mentioning that 8 studies did not present discrimination measures. Regarding internal validation studies, most included bootstrapping or cross-validation techniques.

For the group of studies that only performed internal validation, the one with the highest reported accuracy was that of Norman et al [50]. They used a foam mattress (Sonomat) that, aside from recording breath sounds, has pressure sensors that generate voltage when deformed, thus detecting respiratory movements and classifying OSA events. The highest AUC, sensitivity, and specificity values were 1.00, 100%, and 96%, respectively, for a clinical cutoff AHI≥30. When looking at the studies that externally validated the proposed tools, the study that arose as the best, by Wang et al [25], showcased AUC, sensitivity, and specificity values of 0.99, 96%, and 92%, respectively, for an AHI≥30. The proposed tool consists of a noncontact audio recorder that records sleep sounds, which are then analyzed using a deep learning technique that automatically detects sleep apnea events, calculates AHI, and identifies OSA. An overall note should be made for the studies that used mattress sensors, as they revealed some of the best sensitivity, specificity, and AUC values among those that only performed internal validation. In turn, nasal airflow devices showed high specificity, but lacked sensitivity.

On the basis of the currently available published data, contactless devices, such as audio recorders jointly with machine learning techniques, were also shown to have the most significant potential for screening, diagnosis, and possibly monitoring OSA, being a promising path forward. Future work can follow this strategy to further validate these tools because they are still in the development and testing phases.

Clinical questionnaires, such as the STOP-Bang, Berlin, and NoSAS (Neck, Obesity, Snoring, Age, Sex) scores, can help identify patients at increased risk of OSA [3]. Although they are easy to perform and validate in different populations, they do not offer any advantages over digital clinical tools. Given the possibility of day-to-day use and signal recording, the latter can improve the long-term characterization and follow-up of individuals with sleep-breathing disorders. On the basis of the sensitivity and specificity values, STOP-Bang reached 84% and 54%, respectively [62]. For primary care patients, the Berlin questionnaire achieved values of 86% and 77%, respectively [63]. Finally, in the general population, the NoSAS score had sensitivity and specificity values ranging from 79% to 85% and 69% to 77%, respectively [64]. When comparing this with our best results (internally or externally validated), we can see that most digital clinical tools achieved higher sensitivity and specificity values (eg, the best externally validated tool had 96% sensitivity and 92% specificity).

It is important to consider the limitations and strengths of our methodology as well as those of the included studies. Although we cannot be certain that we retrieved all the published literature on the topic, we are confident that our methodology is adequate. The fact that the search was performed in 3 different search engines (one related to health sciences and 2 with a broader spectrum) minimized this risk. Furthermore, it is worth noting that a great part of the available work has substantial gaps in terms of the study design. Although all studies performed an appropriate statistical analysis, many lacked a satisfactory number of participants in both the test and validation groups. At the time of patient enrollment, the reasons for performing polysomnography were also not clear in all manuscripts. The prevalence of OSA varies from 51% to 100%, with some studies not describing this proportion. In addition, most studies evaluated symptomatic patients referred to a sleep clinic and did not reflect the prevalence of OSA in the overall population. Crucial measures to assess diagnostic capabilities, such as sensitivity, specificity, and AUC, have often not been reported. Nearly all digital tools were tested in a controlled laboratory setting, and given the potential use of these devices as an accessible and less expensive method to screen for OSA, it is paramount to invest in further research to test their performance at the home level, where multiple factors might reduce the accuracy of such technologies. Moreover, most studies have developed and tested these devices, but external validation is still lacking. Given the paucity of studies with a comparison with formal in-laboratory polysomnography, additional studies should also be performed in an attempt to validate such digital clinical tools.

Nevertheless, after analyzing the Joanna Briggs Institute checklist results, we believe that we face an overall considerably low risk of bias in both domains. The most common reason for a high risk of bias was the lack of an OSA definition or cutoff for which the discrimination measures were calculated (Q5). However, most studies that did not present the definition did not report the discrimination measures of interest. In addition, a considerable number of studies selected different cutoff values from those described in the guidelines, which makes it difficult to compare the results with similar studies. Another reason was that several designs were faltered in Q3: “Did the study avoid inappropriate exclusions? This is mostly because of an imprecise description of the reasons that led to those exclusions, as in one study where patients with back pain were excluded because this would make the use of the mattress sensor more difficult, which can overestimate the applicability of this type of tool [47]. Moreover, several studies did not include the results for all selected patients (P10), stating that it was attributable to a lack of space [29] or unreturned devices [54]. Even so, it is important to mention that we used a low threshold to consider the answer to the questions as “no” or “unclear.”

Conclusions

Sleep medicine is a prime field for the use of digital tools and novel unobtrusive technologies. Although they hold great promise, they are still in an early stage of development. This systematic review sheds light on the potential of such devices for the screening or diagnosis of OSA, as they are probably the future of research and development in this field. Although they cannot replace the gold standard of polysomnography, they can greatly assist in large-scale screening and increase the accessibility of the general population to sleep studies. Despite the promising results, this study also highlights the need for future high-quality studies, more robust clinical data, and strategies for care implementation, with the validation of the developed tools in external populations and home environments before they can be used and recommended in a clinical setting.

Acknowledgments

This study was supported by National Funds through Fundação para a Ciência e a Tecnologia, I.P., within the CINTESIS, R&D Unit (reference UIDP/4255/2020).

Abbreviations

AHI

apnea-hypopnea index

AUC

area under the curve

HR

heart rate

HSAT

home sleep apnea test

NoSAS

Neck, Obesity, Snoring, Age, Sex

OSA

obstructive sleep apnea

PRISMA

Preferred Reporting Items for Systematic Reviews and Meta-Analyses

Multimedia Appendix 1

Search queries used for each database—PubMed, Scopus, and Web of Science.

Footnotes

Conflicts of Interest: None declared.

References

  • 1.Strollo Jr PJ, Rogers RM. Obstructive sleep apnea. N Engl J Med. 1996 Jan 11;334(2):99–104. doi: 10.1056/NEJM199601113340207. [DOI] [PubMed] [Google Scholar]
  • 2.Peppard PE, Young T, Barnet JH, Palta M, Hagen EW, Hla KM. Increased prevalence of sleep-disordered breathing in adults. Am J Epidemiol. 2013 May 01;177(9):1006–14. doi: 10.1093/aje/kws342. https://europepmc.org/abstract/MED/23589584 .kws342 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Kapur VK, Auckley DH, Chowdhuri S, Kuhlmann DC, Mehra R, Ramar K, Harrod CG. Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: an American academy of sleep medicine clinical practice guideline. J Clin Sleep Med. 2017 Mar 15;13(3):479–504. doi: 10.5664/jcsm.6506. https://europepmc.org/abstract/MED/28162150 .jc-17-00035 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Benjafield AV, Ayas NT, Eastwood PR, Heinzer R, Ip MS, Morrell MJ, Nunez CM, Patel SR, Penzel T, Pépin JL, Peppard PE, Sinha S, Tufik S, Valentine K, Malhotra A. Estimation of the global prevalence and burden of obstructive sleep apnoea: a literature-based analysis. Lancet Respir Med. 2019 Aug;7(8):687–98. doi: 10.1016/S2213-2600(19)30198-5. https://europepmc.org/abstract/MED/31300334 .S2213-2600(19)30198-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Senaratna CV, Perret JL, Lodge CJ, Lowe AJ, Campbell BE, Matheson MC, Hamilton GS, Dharmage SC. Prevalence of obstructive sleep apnea in the general population: a systematic review. Sleep Med Rev. 2017 Aug;34:70–81. doi: 10.1016/j.smrv.2016.07.002.S1087-0792(16)30064-8 [DOI] [PubMed] [Google Scholar]
  • 6.O'Mahony AM, Garvey JF, McNicholas WT. Technologic advances in the assessment and management of obstructive sleep apnoea beyond the apnoea-hypopnoea index: a narrative review. J Thorac Dis. 2020 Sep;12(9):5020–38. doi: 10.21037/jtd-sleep-2020-003. https://europepmc.org/abstract/MED/33145074 .jtd-12-09-5020 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Franklin KA, Lindberg E. Obstructive sleep apnea is a common disorder in the population-a review on the epidemiology of sleep apnea. J Thorac Dis. 2015 Aug;7(8):1311–22. doi: 10.3978/j.issn.2072-1439.2015.06.11. https://europepmc.org/abstract/MED/26380759 .jtd-07-08-1311 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Kendzerska T, Mollayeva T, Gershon AS, Leung RS, Hawker G, Tomlinson G. Untreated obstructive sleep apnea and the risk for serious long-term adverse outcomes: a systematic review. Sleep Med Rev. 2014 Feb;18(1):49–59. doi: 10.1016/j.smrv.2013.01.003.S1087-0792(13)00007-5 [DOI] [PubMed] [Google Scholar]
  • 9.Mendonça F, Mostafa SS, Ravelo-García AG, Morgado-Dias F, Penzel T. Devices for home detection of obstructive sleep apnea: a review. Sleep Medicine Reviews. 2018 Oct;41:149–60. doi: 10.1016/j.smrv.2018.02.004. [DOI] [PubMed] [Google Scholar]
  • 10.Stewart SA, Skomro R, Reid J, Penz E, Fenton M, Gjevre J, Cotton D. Improvement in obstructive sleep apnea diagnosis and management wait times: a retrospective analysis of home management pathway for obstructive sleep apnea. Can Respir J. 2015 May;22(3):167–70. doi: 10.1155/2015/516580. doi: 10.1155/2015/516580.16633 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Rosa T, Bellardi K, Viana Jr A, Ma Y, Capasso R. Digital health and sleep-disordered breathing: a systematic review and meta-analysis. J Clin Sleep Med. 2018 Sep 15;14(9):1605–20. doi: 10.5664/jcsm.7346. https://europepmc.org/abstract/MED/30176971 .jc-18-00060 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Free C, Phillips G, Felix L, Galli L, Patel V, Edwards P. The effectiveness of M-health technologies for improving health and health services: a systematic review protocol. BMC Res Notes. 2010 Oct 06;3:250. doi: 10.1186/1756-0500-3-250. https://bmcresnotes.biomedcentral.com/articles/10.1186/1756-0500-3-250 .1756-0500-3-250 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Manoni A, Loreti F, Radicioni V, Pellegrino D, Della Torre L, Gumiero A, Halicki D, Palange P, Irrera F. A new wearable system for home sleep Apnea testing, screening, and classification. Sensors (Basel) 2020 Dec 08;20(24):7014. doi: 10.3390/s20247014. https://www.mdpi.com/resolver?pii=s20247014 .s20247014 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Camara MA, Castillo Y, Blanco-Almazan D, Estrada L, Jane R. mHealth tools for monitoring obstructive sleep Apnea patients at home: proof-of-concept. Annu Int Conf IEEE Eng Med Biol Soc. 2017 Jul;2017:1555–8. doi: 10.1109/EMBC.2017.8037133. [DOI] [PubMed] [Google Scholar]
  • 15.de Zambotti M, Cellini N, Goldstone A, Colrain IM, Baker FC. Wearable sleep technology in clinical and research settings. Med Sci Sports Exerc. 2019 Jul;51(7):1538–57. doi: 10.1249/MSS.0000000000001947. https://europepmc.org/abstract/MED/30789439 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Agatsuma T, Fujimoto K, Komatsu Y, Urushihata K, Honda T, Tsukahara T, Nomiyama T. A novel device (SD-101) with high accuracy for screening sleep apnoea-hypopnoea syndrome. Respirology. 2009 Nov;14(8):1143–50. doi: 10.1111/j.1440-1843.2009.01627.x.RES1627 [DOI] [PubMed] [Google Scholar]
  • 17.Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, Shamseer L, Tetzlaff JM, Akl EA, Brennan SE, Chou R, Glanville J, Grimshaw JM, Hróbjartsson A, Lalu MM, Li T, Loder EW, Mayo-Wilson E, McDonald S, McGuinness LA, Stewart LA, Thomas J, Tricco AC, Welch VA, Whiting P, Moher D. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021 Mar 29;372:n71. doi: 10.1136/bmj.n71. http://www.bmj.com/lookup/pmidlookup?view=long&pmid=33782057 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Ouzzani M, Hammady H, Fedorowicz Z, Elmagarmid A. Rayyan—a web and mobile app for systematic reviews. Syst Rev. 2016 Dec 5;5(1):210. doi: 10.1186/s13643-016-0384-4. https://systematicreviewsjournal.biomedcentral.com/articles/10.1186/s13643-016-0384-4 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Campbell JM, Klugar M, Ding S, Carmody DP, Hakonsen SJ, Jadotte YT, White S, Munn Z. Diagnostic test accuracy systematic reviews. In: Aromataris E, Munn Z, editors. JBI Manual for Evidence Synthesis. New York, NY: JBI Publication; 2020. [Google Scholar]
  • 20.Shochat T, Hadas N, Kerkhofs M, Herchuelz A, Penzel T, Peter JH, Lavie P. The SleepStrip: an apnoea screener for the early detection of sleep apnoea syndrome. Eur Respir J. 2002 Jan;19(1):121–6. doi: 10.1183/09031936.02.00227302. http://erj.ersjournals.com/cgi/pmidlookup?view=long&pmid=11843310 . [DOI] [PubMed] [Google Scholar]
  • 21.Castillo-Escario Y, Werthen-Brabants L, Groenendaal W, Deschrijver D, Jane R. Convolutional neural networks for Apnea detection from smartphone audio signals: effect of window size. Annu Int Conf IEEE Eng Med Biol Soc. 2022 Jul;2022:666–9. doi: 10.1109/EMBC48229.2022.9871396. [DOI] [PubMed] [Google Scholar]
  • 22.Xu Y, Ou Q, Cheng Y, Lao M, Pei G. Comparative study of a wearable intelligent sleep monitor and polysomnography monitor for the diagnosis of obstructive sleep apnea. Sleep Breath. 2023 Mar;27(1):205–12. doi: 10.1007/s11325-022-02599-x. https://europepmc.org/abstract/MED/35347656 .10.1007/s11325-022-02599-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Wei Z, Xu J, Li W, Wang X, Qin Z, Zhou J, Wang W. Evaluation of a non-contact ultra-wideband bio-radar sleep monitoring device for screening of sleep breathing disease. Sleep Breath. 2022 Jun;26(2):689–96. doi: 10.1007/s11325-021-02424-x.10.1007/s11325-021-02424-x [DOI] [PubMed] [Google Scholar]
  • 24.Zhuang Z, Wang F, Yang X, Zhang L, Fu CH, Xu J, Li C, Hong H. Accurate contactless sleep apnea detection framework with signal processing and machine learning methods. Methods. 2022 Sep;205:167–78. doi: 10.1016/j.ymeth.2022.06.013.S1046-2023(22)00153-0 [DOI] [PubMed] [Google Scholar]
  • 25.Wang B, Tang X, Ai H, Li Y, Xu W, Wang X, Han D. Obstructive sleep apnea detection based on sleep sounds via deep learning. Nat Sci Sleep. 2022 Nov 08;14:2033–45. doi: 10.2147/NSS.S373367. https://europepmc.org/abstract/MED/36394068 .373367 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Haberfeld C, Sheta A, Hossain MS, Turabieh H, Surani S. SAS mobile application for diagnosis of obstructive sleep apnea utilizing machine learning models. Proceedings of the 11th IEEE Annual Ubiquitous Computing, Electronics & Mobile Communication Conference; UEMCON '20; October 28-31, 2020; New York, NY, USA. 2020. pp. 522–9. https://ieeexplore.ieee.org/document/9298041 . [DOI] [Google Scholar]
  • 27.Jin J, Sánchez-Sinencio E. A home sleep apnea screening device with time-domain signal processing and autonomous scoring capability. IEEE Trans Biomed Circuits Syst. 2015 Feb;9(1):96–104. doi: 10.1109/TBCAS.2014.2314301. [DOI] [PubMed] [Google Scholar]
  • 28.Lyon G, Tiron R, Zaffaroni A, Osman A, Kilroy H, Lederer K, Fietze I, Penzel T. Detection of sleep apnea using sonar smartphone technology. Annu Int Conf IEEE Eng Med Biol Soc. 2019 Jul;2019:7193–6. doi: 10.1109/EMBC.2019.8857836. [DOI] [PubMed] [Google Scholar]
  • 29.Zhang J, Zhang Q, Wang Y, Qiu C. A real-time auto-adjustable smart pillow system for sleep apnea detection and treatment. Proceedings of the 12th international conference on Information processing in sensor networks; IPSN '13; April 8-11, 2013; Philadelphia, PA, USA. 2013. pp. 179–90. https://dl.acm.org/doi/10.1145/2461381.2461405 . [DOI] [Google Scholar]
  • 30.Tsukahara M, Sakao S, Jujo T, Sakurai T, Terada J, Kunii R, Tanabe N, Tatsumi K. The accuracy and uncertainty of a sheet-type portable monitor as a screening device to identify obstructive sleep apnea-hypopnea syndrome. Intern Med. 2014;53(12):1307–13. doi: 10.2169/internalmedicine.53.2208. https://joi.jlc.jst.go.jp/DN/JST.JSTAGE/internalmedicine/53.2208?from=PubMed .DN/JST.JSTAGE/internalmedicine/53.2208 [DOI] [PubMed] [Google Scholar]
  • 31.Xin Y, Liu T, Liu H, Liu L, Li Y, Hou T, Liu C, Zhu J, Lin T, Jin S. A bone-conduction transducer-based detection system for sleep apnea screening in the family units. IEEE Sens J. 2021 Mar 15;21(6):8411–20. doi: 10.1109/jsen.2020.3048310. https://ieeexplore.ieee.org/document/9311653/ [DOI] [Google Scholar]
  • 32.Narayan S, Shivdare P, Niranjan T, Williams K, Freudman J, Sehra R. Noncontact identification of sleep-disturbed breathing from smartphone-recorded sounds validated by polysomnography. Sleep Breath. 2019 Mar;23(1):269–79. doi: 10.1007/s11325-018-1695-6. https://europepmc.org/abstract/MED/30022325 .10.1007/s11325-018-1695-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Al-Mardini M, Aloul F, Sagahyroon A, Al-Husseini L. Classifying obstructive sleep apnea using smartphones. J Biomed Inform. 2014 Dec;52:251–9. doi: 10.1016/j.jbi.2014.07.004. https://linkinghub.elsevier.com/retrieve/pii/S1532-0464(14)00151-8 .S1532-0464(14)00151-8 [DOI] [PubMed] [Google Scholar]
  • 34.Nandakumar R, Gollakota S, Watson NF. Contactless sleep apnea detection on smartphones. Proceedings of the 13th Annual International Conference on Mobile Systems, Applications, and Services; MobiSys '15; May 18-22, 2015; Florence, Italy. 2015. pp. 45–57. https://dl.acm.org/doi/10.1145/2742647.2742674 . [DOI] [Google Scholar]
  • 35.Tiron R, Lyon G, Kilroy H, Osman A, Kelly N, O'Mahony N, Lopes C, Coffey S, McMahon S, Wren M, Conway K, Fox N, Costello J, Shouldice R, Lederer K, Fietze I, Penzel T. Screening for obstructive sleep apnea with novel hybrid acoustic smartphone app technology. J Thorac Dis. 2020 Aug;12(8):4476–95. doi: 10.21037/jtd-20-804. https://europepmc.org/abstract/MED/32944361 .jtd-12-08-4476 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Fedorin I, Slyusarenko K. Consumer smartwatches as a portable PSG: LSTM based neural networks for a sleep-related physiological parameters estimation. Annu Int Conf IEEE Eng Med Biol Soc. 2021 Nov;2021:849–452. doi: 10.1109/EMBC46164.2021.9629597. [DOI] [PubMed] [Google Scholar]
  • 37.Chen X, Xiao Y, Tang Y, Fernandez-Mendoza J, Cao G. ApneaDetector: detecting sleep apnea with smartwatches. Proc ACM Interact Mob Wearable Ubiquitous Technol. 2021 Jun 24;5(2):1–22. doi: 10.1145/3463514. https://dl.acm.org/doi/abs/10.1145/3463514 . [DOI] [Google Scholar]
  • 38.Chen Y, Wang W, Guo Y, Zhang H, Chen Y, Xie L. A single-center validation of the accuracy of a photoplethysmography-based smartwatch for screening obstructive sleep apnea. Nat Sci Sleep. 2021 Sep 16;13:1533–44. doi: 10.2147/NSS.S323286. https://europepmc.org/abstract/MED/34557047 .323286 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Hayano J, Yamamoto H, Nonaka I, Komazawa M, Itao K, Ueda N, Tanaka H, Yuda E. Quantitative detection of sleep apnea with wearable watch device. PLoS One. 2020 Nov 09;15(11):e0237279. doi: 10.1371/journal.pone.0237279.PONE-D-20-22289 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Ben Azouz A, Issa A, Knief P, Kane T, Coyle S, Costello R, McGuigan KG, Diamond D. Evaluation of use of wearable sensor garment in home screening for sleep apnea events. Proceedings of the 2018 International Conference on Promising Electronic Technologies; ICPET '18; October 3-4, 2018; Deir El-Balah, Palestine. 2018. pp. 104–7. https://ieeexplore.ieee.org/document/8531238 . [DOI] [Google Scholar]
  • 41.Ganglberger W, Bucklin AA, Tesh RA, Da Silva Cardoso M, Sun H, Leone MJ, Paixao L, Panneerselvam E, Ye EM, Thompson BT, Akeju O, Kuller D, Thomas RJ, Westover MB. Sleep apnea and respiratory anomaly detection from a wearable band and oxygen saturation. Sleep Breath. 2022 Sep;26(3):1033–44. doi: 10.1007/s11325-021-02465-2. https://europepmc.org/abstract/MED/34409545 .10.1007/s11325-021-02465-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Levendowski DJ, Veljkovic B, Seagraves S, Westbrook PR. Capability of a neck worn device to measure sleep/wake, airway position, and differentiate benign snoring from obstructive sleep apnea. J Clin Monit Comput. 2015 Feb;29(1):53–64. doi: 10.1007/s10877-014-9569-3. https://europepmc.org/abstract/MED/24599632 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Selvaraj N, Narasimhan R. Automated prediction of the apnea-hypopnea index using a wireless patch sensor. Annu Int Conf IEEE Eng Med Biol Soc. 2014;2014:1897–900. doi: 10.1109/EMBC.2014.6943981. [DOI] [PubMed] [Google Scholar]
  • 44.Yeh E, Wong E, Tsai CW, Gu W, Chen PL, Leung L, Wu IC, Strohl KP, Folz RJ, Yar W, Chiang AA. Detection of obstructive sleep apnea using Belun Sleep Platform wearable with neural network-based algorithm and its combined use with STOP-Bang questionnaire. PLoS One. 2021 Oct 11;16(10):e0258040. doi: 10.1371/journal.pone.0258040. https://dx.plos.org/10.1371/journal.pone.0258040 .PONE-D-21-13761 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Davidovich ML, Karasik R, Tal A, Shinar Z. Sleep apnea screening with a contact-free under-the-mattress sensor. Comput Cardiol. 2016;46:849–52. doi: 10.22489/cinc.2016.246-183. https://www.cinc.org/archives/2016/pdf/246-183.pdf . [DOI] [Google Scholar]
  • 46.Hwang SH, Lee HJ, Yoon HN, Jung DW, Lee YJ, Lee YJ, Jeong DU, Park KS. Unconstrained sleep apnea monitoring using polyvinylidene fluoride film-based sensor. IEEE Trans Biomed Eng. 2014 Jul;61(7):2125–34. doi: 10.1109/TBME.2014.2314452. [DOI] [PubMed] [Google Scholar]
  • 47.Meng L, Xu H, Guan J, Yi H, Wu H, Yin S. Validation of a novel sleep-monitoring system for diagnosing obstructive sleep apnea: a comparison with polysomnography. Exp Ther Med. 2016 Nov;12(5):2937–41. doi: 10.3892/etm.2016.3721. https://europepmc.org/abstract/MED/27882098 .ETM-0-0-3721 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Mora GG, Kortelainen JM, Hernandez ER, Tenhunen M, Bianchi AM, Mendez MO. Evaluation of pressure bed sensor for automatic SAHS screening. IEEE Trans Instrum Meas. 2015 Jul;64(7):1935–43. doi: 10.1109/tim.2014.2366976. https://ieeexplore.ieee.org/document/6980123 . [DOI] [Google Scholar]
  • 49.Mosquera-Lopez C, Leitschuh J, Condon J, Hagen CC, Rajhbeharrysingh U, Hanks C, Jacobs PG. Design and evaluation of a non-Contact Bed-mounted sensing device for AutomatedIn-home detection of obstructive sleep apnea:a pilot study. Biosensors (Basel) 2019 Jul 22;9(3):90. doi: 10.3390/bios9030090. https://www.mdpi.com/resolver?pii=bios9030090 .bios9030090 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Norman MB, Middleton S, Erskine O, Middleton PG, Wheatley JR, Sullivan CE. Validation of the Sonomat: a contactless monitoring system used for the diagnosis of sleep disordered breathing. Sleep. 2014 Sep 01;37(9):1477–87. doi: 10.5665/sleep.3996. https://europepmc.org/abstract/MED/25142565 .sp-00541-13 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Sadek I, Heng TT, Seet E, Abdulrazak B. A new approach for detecting sleep apnea using a contactless bed sensor: comparison study. J Med Internet Res. 2020 Sep 18;22(9):e18297. doi: 10.2196/18297. https://www.jmir.org/2020/9/e18297/ v22i9e18297 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Tenhunen M, Elomaa E, Sistonen H, Rauhala E, Himanen SL. Emfit movement sensor in evaluating nocturnal breathing. Respir Physiol Neurobiol. 2013 Jun 15;187(2):183–9. doi: 10.1016/j.resp.2013.03.009.S1569-9048(13)00088-8 [DOI] [PubMed] [Google Scholar]
  • 53.Dinç AE, Yılmaz M, Tutar H, Aydil U, Kızıl Y, Damar M, Kemaloğlu YK. Reliability of SleepStrip as a screening test in obstructive sleep apnea patients. Eur Arch Otorhinolaryngol. 2014 Oct;271(10):2813–8. doi: 10.1007/s00405-014-3087-2. [DOI] [PubMed] [Google Scholar]
  • 54.Ozmen OA, Tüzemen G, Kasapoğlu F, Ozmen S, Coşkun H, Ursavaş A, Karadağ M, Onart S. The reliability of SleepStrip as a screening test in obstructive sleep apnea syndrome. Kulak Burun Bogaz Ihtis Derg. 2011 Jan;21(1):15–9. [PubMed] [Google Scholar]
  • 55.Wong KK, Jankelson D, Reid A, Unger G, Dungan G, Hedner JA, Grunstein RR. Diagnostic test evaluation of a nasal flow monitor for obstructive sleep apnea detection in sleep apnea research. Behav Res Methods. 2008 Feb;40(1):360–6. doi: 10.3758/brm.40.1.360. [DOI] [PubMed] [Google Scholar]
  • 56.Crinion SJ, Tiron R, Lyon G, Zaffaroni A, Kilroy H, Doheny E, O'Hare E, Boyle P, Russell A, Traynor M, Kent BD, Ryan S, McNicholas WT. Ambulatory detection of sleep apnea using a non-contact biomotion sensor. J Sleep Res. 2020 Feb;29(1):e12889. doi: 10.1111/jsr.12889. [DOI] [PubMed] [Google Scholar]
  • 57.Zaffaroni A, Kent B, O'Hare E, Heneghan C, Boyle P, O'Connell G, Pallin M, de Chazal P, McNicholas WT. Assessment of sleep-disordered breathing using a non-contact bio-motion sensor. J Sleep Res. 2013 Apr;22(2):231–6. doi: 10.1111/j.1365-2869.2012.01056.x. [DOI] [PubMed] [Google Scholar]
  • 58.Zhao R, Xue J, Dong XS, Zhi H, Chen J, Zhao L, Zhang X, Li J, Penzel T, Han F. Screening for obstructive sleep apnea using a contact-free system compared with polysomnography. J Clin Sleep Med. 2021 May 01;17(5):1075–82. doi: 10.5664/jcsm.9138. https://europepmc.org/abstract/MED/33576734 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Alshaer H, Levchenko A, Bradley TD, Pong S, Tseng WH, Fernie GR. A system for portable sleep apnea diagnosis using an embedded data capturing module. J Clin Monit Comput. 2013 Jun;27(3):303–11. doi: 10.1007/s10877-013-9435-8. [DOI] [PubMed] [Google Scholar]
  • 60.Behar J, Roebuck A, Domingos JS, Gederi E, Clifford GD. A review of current sleep screening applications for smartphones. Physiol Meas. 2013 Jul;34(7):R29–46. doi: 10.1088/0967-3334/34/7/R29. [DOI] [PubMed] [Google Scholar]
  • 61.Kim DH, Kim SW, Hwang SH. Diagnostic value of smartphone in obstructive sleep apnea syndrome: a systematic review and meta-analysis. PLoS One. 2022 May 19;17(5):e0268585. doi: 10.1371/journal.pone.0268585. https://dx.plos.org/10.1371/journal.pone.0268585 .PONE-D-21-33551 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Chung F, Yegneswaran B, Liao P, Chung SA, Vairavanathan S, Islam S, Khajehdehi A, Shapiro CM. STOP questionnaire: a tool to screen patients for obstructive sleep apnea. Anesthesiology. 2008 May;108(5):812–21. doi: 10.1097/ALN.0b013e31816d83e4. https://pubs.asahq.org/anesthesiology/article-lookup/doi/10.1097/ALN.0b013e31816d83e4 .00000542-200805000-00008 [DOI] [PubMed] [Google Scholar]
  • 63.Netzer NC, Stoohs RA, Netzer CM, Clark K, Strohl KP. Using the Berlin questionnaire to identify patients at risk for the sleep apnea syndrome. Ann Intern Med. 1999 Oct 05;131(7):485–91. doi: 10.7326/0003-4819-131-7-199910050-00002.199910050-00002 [DOI] [PubMed] [Google Scholar]
  • 64.Marti-Soler H, Hirotsu C, Marques-Vidal P, Vollenweider P, Waeber G, Preisig M, Tafti M, Tufik SB, Bittencourt L, Tufik S, Haba-Rubio J, Heinzer R. The NoSAS score for screening of sleep-disordered breathing: a derivation and validation study. The Lancet Respiratory Medicine. 2016 Sep;4(9):742–8. doi: 10.1016/s2213-2600(16)30075-3. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Multimedia Appendix 1

Search queries used for each database—PubMed, Scopus, and Web of Science.


Articles from Journal of Medical Internet Research are provided here courtesy of JMIR Publications Inc.