Skip to main content
The British Journal of Ophthalmology logoLink to The British Journal of Ophthalmology
. 1989 Jul;73(7):570–573. doi: 10.1136/bjo.73.7.570

Validation of a photographic method of measuring corneal diameter.

J Robinson 1, K J Gilmore 1, A R Fielder 1
PMCID: PMC1041804  PMID: 2757999

Abstract

A photographic method for measuring corneal diameter using the Medical-Nikkor f200 mm lens is described. Measurements were compared with those obtained by calipers (46 eyes of 25 patients) and by placing a ruler either near the eye (123 eyes of 64 patients) or on the nose (98 eyes of 55 patients). Over all we found good correlation between photographic and caliper measurements (r = 0.94). No significant correlation was found between photographic measurements and estimates made with the ruler either near the eye or on the nose (r = 0.65 and 0.31 respectively). Modifications to our system are suggested which may provide an accurate, simple, non-invasive method of measuring corneal diameter in ophthalmic clinics.

Full text

PDF
570

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Aizawa F., Tagawa S., Ono K., Fukui H., Muto S. [Photographic estimation of corneal diameter]. Nihon Ganka Kiyo. 1971 Dec;22(12):1069–1072. [PubMed] [Google Scholar]
  2. Kiskis A. A., Markowitz S. N., Morin J. D. Corneal diameter and axial length in congenital glaucoma. Can J Ophthalmol. 1985 Apr;20(3):93–97. [PubMed] [Google Scholar]

Articles from The British Journal of Ophthalmology are provided here courtesy of BMJ Publishing Group