Abstract
Introduction: Oxygen saturation is essential for medical care and is closely regulated within the body. Arterial blood gas (ABG) analysis is used to evaluate critically ill individuals' ventilation, oxygenation, acid-base status, and metabolic condition. Pulse oximetry is an easy and non-invasive way to measure the status of oxygen saturation non-invasively in clinical settings and provides a quick and precise assessment of oxygenation and reduces medical errors. SpO2 may not always be a reliable predictor of arterial oxygen saturation (SaO2), and hypoxemic, hemodynamically compromised, and critically ill patients may have lower SpO2 accuracy. A study is needed to assess and compare various oxygen saturation methods.
Aims and objectives: The study aimed to compare the oxygen saturation levels measured by pulse oximetry and ABG analysis in hypoxemic patients. The objectives were to compare the values between SaO2, PaO2, and SpO2 values obtained from the patients, and correlate the study parameters among both techniques.
Materials and methods: The study was conducted from February 2021 to June 2022 among the 102 hypoxemic patients who were admitted to the emergency and surgical intensive care unit (ICU) of Sree Balaji Medical College and Hospital in Chennai. Primary data on ABG analysis and pulse oximetry readings were collected from the study subjects. The patient and their past medical records, physical exam, chest x-ray findings, pulse oximetry, and ABG results were all reviewed. Each patient had their ABG, and pulse oximetry measured simultaneously. A comparison was made between SpO2 and partial pressure of oxygen (PaO2) and arterial oxygen saturation (SaO2) parameters using a paired t-test. The correlation was done against the SpO2 and ABG parameters and assessed for association using the correlation coefficient value; gender was also considered while correlating.
Results and discussion: An observational study was done among 102 study samples to comparatively analyze the oxygen saturation by two methods, namely pulse oximetry and ABG, in hypoxemic patients. While comparing the mean values of SaO2 and SpO2, they were 84.41 ± 4.24 and 80.58 ± 5.77, respectively, and this difference was statistically very significant (p < 0.001). While comparing the mean values of PaO2 and SaO2, they were 61.02 ± 5.01 and 84.41 ± 4.24, respectively, and this difference was statistically significant (p = 0.043). While comparing the mean values of PaO2 and SpO2, they were 61.02 ± 5.01 and 80.58 ± 5.77, respectively, and this difference was statistically significant (p = 0.054). Among the study population, with regard to the correlation factor, there is notably a very high and strong positive correlation between SaO2 and SpO2 and between SpO2 and PaO2. There was a negative correlation between SpO2 and finger abnormalities and between SpO2 and blood pressure.
Conclusion: The ABG method is considered the gold standard. When SpO2 levels fall below 90%, pulse oximetry may not be accurate enough to reliably assess oxygenation. In such cases, where alveolar hypoventilation is suspected, it is recommended to complement pulse oximetry with ABG studies. This is because ABG analysis provides a more comprehensive assessment of oxygenation and acid-base status, which can aid in the diagnosis and management of respiratory conditions.
Keywords: respiratory disease, arterial blood gas, pulse oximetry, oxygenation, acute hypoxemia
Introduction
Analysis of arterial blood gas (ABG) is frequently used to evaluate the ventilation, oxygenation, and acid-base status of critically unwell individuals. ABGs are obtained from arteries and provide a real-time indicator of the patient's oxygenation, ventilation, acid-base balance, and metabolic condition. Samples can be arterial, venous, or a combination of both. An arterial line blood sample or an arterial puncture can be used to measure an ABG. In critically ill patients, ABG analysis is crucial for making the correct diagnosis and selecting the best course of treatment. Respiratory failure is a good example of this function. ABG is the only study capable of identifying, describing, and quantifying respiratory failure. Primary oxygenation failure is indicated by the low or normal partial pressure of arterial carbon dioxide (PaCO2), low partial pressure of arterial oxygen (PaO2), and an alkaline pH, whereas ventilatory failure is indicated by the highest PaCO2, low PaO2, and pH. Pulse oximetry is used to measure the status of oxygen saturation (SpO2) non-invasively. It is a technique that is frequently utilized in clinical settings, including intensive care, surgery, and even outpatient clinics [1]. Pulse oximetry has been used in these contexts to lessen the ABG analysis and to improve the identification of conditions like hypoxic factors, which is indicated by an observed SpO2 value of less than 95% [2].
The rationale of the study was to analyze and compare finger oximeter and ABG readings in hypoxemic patients at a tertiary care hospital in Chennai. Technical issues with the measurement instrument or poor data interpretation can impact the precision and accuracy of SpO2 values. Pulse oximetry is an easy, non-invasive way to detect continuous blood oxygen saturation. Previous research has suggested that SpO2 may not always be a reliable predictor of SaO2. A study must be designed to assess and compare various oxygen saturation methods to determine the most reliable and effective method.
Materials and methods
The study was carried out among the 18-80 years old hypoxemic population who visited and got admitted to Sree Balaji Medical College and Hospital in Chennai for a period of 17 months (February 2021 to June 2022). The nature and purpose of the study were explained to the Institutional Review Board, Sree Balaji Medical College, and the Hospital following which ethical clearance was obtained to conduct the study (Institutional Human Ethical Committee of Sree Balaji Medical College and Hospital Ref No: 002/SBMC/IHEC/2021/1523). From February 2021 to June 2022, 102 patients at the emergency and surgical ICU at the Sree Balaji Medical College and Hospital in Chennai were the subject of the study. The purpose of the study was to compare and confirm the variables influencing the accuracy of the pulse oximeter in relation to ABG.
Primary data on ABG analysis and pulse oximetry readings were collected from the study subjects. The patient and their past medical records, physical exam, chest x-ray findings, pulse oximetry, and ABG results were all reviewed. After obtaining informed consent, patients who were admitted to the ICU throughout the study period, whether or not they had hypoxemia, were included in the study. Each patient admitted to the ICU had their ABG and pulse oximetry measured simultaneously. Before performing the Allen test, which evaluates the blood supply in the hands, the patient was asked to clench their fist for 30 seconds. After that, the pressure was put over the ulnar and radial arteries, and the hand was then opened. The Monitor - Nellcor Pulse Oximetry DS-100A Adult Finger Sensor was used to measure and analyze the pulse oximetry when the ABG was stopped in compliance with COBAS B 221 standards. A pulse oximeter was used to monitor SpO2 (Fingertip, China). In the arm opposite the one from which the arterial sample had been taken the finger probe device was placed on the index finger. SpO2 was measured twice at 0 and 2 minutes, and the average of the two measurements was calculated. In this ABG values served as a reference. ABG defines hypoxemia as having a partial oxygen pressure (PaO2) below 60 mm Hg or having a SpO2 or SaO2 below 90% [3].
The subjects included in the study were subjects consenting to the study, a population of 18-80 years of hypoxemic patients, pulse oximetry, and ABG analysis should be performed. simultaneously. The exclusion factors were individuals under the age of 18, patients with fingernails colored, blood samples out of a vein, and patients with methemoglobinemia and carbon monoxide poisoning were also excluded.
The sample size was determined from the G power statistical software tool. Values are recorded and noted in an Excel sheet decoded and coded against the standard values. Descriptive data and inferential statistics were calculated by Paired t-test and Pearson rank correlation coefficient test. Comparison made between SpO2 with PaO2, SaO2 parameters using paired t-test. Correlation was done against the SpO2 and ABG parameters and assessed for association using the correlation coefficient value, gender was also considered while correlating.
Statistical analysis version 21 of SPSS was used for the statistical analysis (IBM Corp., Armonk, NY, USA). Excel was used to first tabulate the data. The collected information was quantitative in nature; thus, the normality test was performed on it (Kolmogorov-Smirnov test). Since the data were determined to be regularly distributed, intra- and inter-group comparisons were made using parametric tests of significance. Mean and standard deviation was used to describe descriptive data that was quantitative in nature. SapO2 and SaO2 were compared using a paired t-test. Using Pearson's correlation, it was determined whether there was a correlation between vital signs and ABG parameters, vital signs and SapO2, hemoglobin concentration and ABG parameters, hemoglobin concentration, and SapO2, and ABG parameters and SapO2. In all tests, a p-value of <0.05 was considered the level of statistical significance.
Results
An observational study was done among 102 study samples to comparatively analyze the oxygen saturation by two methods namely: (1) Pulse oximetry and (2) ABG in hypoxemic patients in a tertiary care hospital in Chennai. The quantitative data were described using the mean and standard deviation. All continuous variables underwent the Kolmogorov-Smirnov test to evaluate their normality, and the results showed that the data were normally distributed. Therefore, both within-group and between-group comparisons were made using parametric tests of significance.
Table 1 shows the distribution of participants through each stage of the study. A total of 102 adults residing in Chennai, who complained of breathlessness were included in the study. The gender was distributed as 54 males and 48 females (Table 1). Study parameters like PaCO2, PaO2, and HCO3 for ABG and SpO2 for pulse oximeter are mentioned.
Table 1. Gender distribution of the study samples.
Gender | Frequency (n=102) | Percentage (%) |
Females | 48 | 47.05 |
Males | 54 | 52.94 |
Every patient underwent both kinds of oxygen saturation techniques simultaneously. Values like PaCO2, PaO2, and HCO3 for ABG and SpO2 for pulse oximeter were recorded.
Mean age of the study subjects
The mean age of the study population was 49.60 ± 17.09 years.
Mean values of the study parameters
The mean value of hemoglobin content was 9.70 ± 1.24. The mean value of blood pressure was 109.68 + 26.6. The mean value of pH was 7.35 ± 0.10. The mean value of PaO2 was 61.02 ± 5.01. The mean value of PaCO2 was 38.01 ± 7.54. The mean value of HCO3 was 18.51 ± 4.18. The mean value of SaO2 was 84.41 ± 4.24. The mean value of SpO2 was 80.58 ± 5.77 (Table 2).
Table 2. Descriptive statistics depicting the mean values of the study values.
Study Parameters | N | Mean | Std. Deviation |
Age (in years) | 102 | 49.6078 | 17.09968 |
Hemoglobin (g/dL) | 102 | 9.7049 | 1.24243 |
BP (mm Hg) | 102 | 109.6863 | 26.69037 |
pH | 102 | 7.3597 | 0.10102 |
PaO2 (mm Hg) | 102 | 61.0294 | 5.01178 |
PaCO2 (mm Hg) | 102 | 38.0196 | 7.54784 |
HCO3 (mmol/L) | 102 | 18.5196 | 4.18798 |
SaO2 (%) | 102 | 84.4118 | 4.24346 |
SpO2 (%) | 102 | 80.5882 | 5.77011 |
Finger abnormality | 102 | NA | NA |
The mean values of each study parameter are listed along with its standard deviation. Since it is a quantitative measurement, the mean and standard deviations are listed in the above-mentioned table. Since finger abnormalities are a qualitative measure, they cannot be quantified.
While comparing the mean values of SaO2 and SpO2 were 84.41 ± 4.24 and 80.58 ± 5.77, respectively, and this difference was statistically very highly significant (p< 0.001, Table 3). While comparing the mean values of PaO2 and SaO2 were 61.02 ± 5.01 and 84.41 ± 4.24, respectively, and this difference was statistically significant (p = 0.043, Table 3). While comparing the mean values of PaO2 and SpO2 were 61.02 ± 5.01 and 80.58 ± 5.77, respectively, and this difference was statistically significant (p = 0.054, Table 3). While comparing the mean values of BP and SpO2 were 109.68 ± 26.69 and 80.58 ± 5.77, respectively, and this difference was statistically highly significant (p= 0.007, Table 3). While comparing the mean values of BP and SaO2 were 109.68 ± 26.69 and 84.41 ± 4.24, respectively, and this difference was statistically very highly significant (p = 0.001, Table 3).
Table 3. Comparison of different variables.
Comparison of the Gold standard ABG (SaO2) and Pulse oximetry readings (SpO2) | ||||
Mean | Standard deviation | P-value | ||
Pair 1 | SaO2 (%) | 84.4118 | 4.24346 | <0.001*** |
SpO2 (%) | 80.5882 | 5.77011 | ||
Comparison of the Gold standard ABG - PaO2 and Pulse oximetry readings (SpO2) | ||||
Mean | Standard deviation | P-value | ||
Pair 2 | PaO2 (mm Hg) | 61.0294 | 5.01178 | 0.043* |
SaO2 (%) | 84.4118 | 4.24346 | ||
Comparison of the Gold standard ABG - PaO2 and Pulse oximetry readings (SpO2) | ||||
Mean | Standard deviation | P-value | ||
Pair 3 | PaO2 (mm Hg) | 61.0294 | 5.01178 | <0.001*** |
SpO2 (%) | 80.5882 | 5.77011 | ||
Comparison of the BP and Pulse oximetry readings (SpO2) | ||||
Mean | Standard deviation | P-value | ||
Pair 4 | BP (mm Hg) | 109.6863 | 26.69037 | <0.001*** |
SpO2 (%) | 80.5882 | 5.77011 | ||
Comparison of the Gold standard ABG - SaO2 and BP | ||||
Mean | Standard deviation | P-value | ||
Pair 5 | BP (mm Hg) | 109.6863 | 26.69037 | <0.001*** |
SaO2 (%) | 84.4118 | 4.24346 |
While comparing the mean values of SaO2 and SpO2, this difference was statistically very highly significant. While comparing the mean values of PaO2 and SaO2, this difference was statistically significant. While comparing the mean values of PaO2 and SpO2, this difference was statistically significant. While comparing the mean values of BP and SpO2, this difference was statistically highly significant. While comparing the mean values of BP and SaO2, this difference was statistically very highly significant.
Among the study population with regards to the correlation factor, there is notably a very high and strong positive correlation between SaO2 and SpO2 and between SpO2 and PaO2. There was a negative correlation between SpO2 and Finger abnormality and between SpO2 and Blood pressure (Table 4).
Table 4. Correlation of the values among both the procedures.
SpO2 | ||
Age (in years) | Pearson Correlation | 0.028 |
P-value | 0.777 | |
N | 102 | |
Hemoglobin (g/dL) | Pearson Correlation | 0.306 |
P-value | 0.002** | |
N | 102 | |
BP (mm Hg) | Pearson Correlation | -0.403 |
P-value | <0.001*** | |
N | 102 | |
pH | Pearson Correlation | 0.483 |
P-value | <0.001*** | |
N | 102 | |
PaO2 (mm Hg) | Pearson Correlation | 0.800 |
P-value | 0.043* | |
N | 102 | |
PaCO2 (mm Hg) | Pearson Correlation | 0.345 |
P-value | <0.001*** | |
N | 102 | |
HCO3 (mEq) | Pearson Correlation | -0.215 |
P-value | 0.030* | |
N | 102 | |
SaO2 (%) | Pearson Correlation | 0.842 |
P-value | <0.001*** | |
N | 102 | |
Finger abnormality | Pearson Correlation | -0.797 |
P-value | 0.008** | |
N | 102 |
Among the study population with regards to the correlation factor, there is notably a very high and strong positive correlation between SaO2 and SpO2 and between SpO2 and PaO2. There was a negative correlation between SpO2 and finger abnormality and between SpO2 and blood pressure.
Statistical significance vs clinical significance
Hence, the likelihood of obtaining a reliable result at times when there are hypoxic patients and blood pressure is significantly low or when there are confounding factors like finger abnormalities and lesions, with regards to ABG, the accuracy and sensitivity of obtaining true positive results are high compared to the finger pulse oximeter. Additionally, SpO2 can be used at times during emergencies but should not be relied upon as a standard procedure.
Discussion
The gold standard for learning about a critical patient's oxygenation, ventilation, and acid-base status is an ABG analysis. ABG may be used instead of a pulse oximeter in patients with acute illnesses who have a SpO2 below 90%; however, this may not always be the case when people are experiencing acute issues. The study contends that a pulse oximeter is an ineffective substitute for an ABG analyzer in patients with SpO2 below 90% since it cannot accurately estimate SaO2 or assess oxygenation. One of the most crucial components of patient management is the evaluation of the patient's oxygen saturation level, which is used to determine whether to start or increase oxygen therapy. There are some drawbacks to measuring ABG analysis, including the necessity to palpate the artery, the requirement for many punctures, an increased risk of infection, and the lack of a continuous monitoring procedure. A pulse oximeter, on the other hand, can be used to continually monitor the patient's saturation level while simultaneously making immediate attempts to rectify hypoxemia. It is also patient-friendly, non-invasive, and painless. The two basic assumptions of the pulse oximeter technology are that arterial blood flow is the only source of tissue pulsations in the body on which a pulse oximeter is used, and that hemoglobin exists in two states, either as hemoglobin or oxyhemoglobin.
A normal pulse oximetry saturation is regarded as being greater than 95%, albeit this definition may change depending on the patient's comorbidities. For patients with COVID-19, oxygen therapy is frequently titrated to maintain a SpO2 of at least >88%. Under usual circumstances, PaO2 can be predicted from SpO2 using a conventional oxyhemoglobin dissociation curve. The ABG measurement is more precise in these circumstances, such as carbon monoxide poisoning, where the pulse oximeter reading may not adequately reflect the oxygen-carrying capacity. Some COVID-19 patients have displayed the unsettling false assurance of having a 100% SpO2 when their PaO2 is less than 50 mm Hg. To prevent avoidable mistakes, it is crucial to consider when to rely on a saturation probe and which individuals will need an arterial blood analysis.
The accuracy of comparing the oxygen saturation measured by pulse oximetry with the ABG has been the subject of numerous studies. The data made available allow for a thorough evaluation of the effectiveness and performance of the proposed technique, and a comparison with the results obtained provides a clear image of which test should be used in the event of abnormalities. In contrast to our investigation, Nessler et al. [4] observed that the frontal pulse oximeter sensor was more reliable than the pulse oximeter on the fingers in detecting SpO2 (n=89%) in patients receiving vasopressors. Wilson et al. [5] advocated the use of ABG rather than SpO2 after observing a 2.7% discrepancy between SpO2 and SaO2 in emergency patients presenting with severe sepsis and septic shock. This was consistent with the findings of our investigation, in which even we found a respectable discrepancy between the SpO2 and SaO2 levels (the mean SaO2 value was 84.41± 4.24). The average SpO2 levels were 80.58 ± 5.77).
According to Jensen et al., recent meta-analysis of 74 trials (1976 to 1994) [6] on the value of SaO2 revealed by pulse oximetry, pulse oximeters were accurate to within 2% of SaO2 in the range of 70%-100% SaO2. Our study confines to almost similar results with the aforementioned vast study since even our study had almost at least 10%-30% similarity between SaO2 and SpO2 values as they almost change during extreme chronic conditions and in general - hypotensive conditions only. The Carter et al. [7] research discovered that pulse oximetry performance deteriorated below a SpO2 of 75%, which was consistent with the findings of our study. The readings tended to decrease below a saturation point of 75%, indicating that the pulse oximeter's performance was less than optimal compared to that under typical circumstances. The current study found an almost significant difference between the SpO2 and SaO2 values, which was consistent with Chiappini et al.'s investigations [8]. They found that the pulse oximeter's accuracy was only adequate for SpO2 readings under 82%, which was also true in the current study. A study on COPD patients by Razi and Akbari [9] included 152 participants. A 90.8% accuracy rate, a 93.3% sensitivity rate, and an 89.1% specificity rate were all achieved via pulse oximetry. The study asserted that pulse oximetry has high accuracy in estimating SaO2 in pulmonary disorders with SpO2 at 80%, in contrast to our investigation, which discovered that accuracy is estimated when SpO2 is larger than 90%. The results, however, are in line with those of our study of patients with SpO2 80%, which convinces us that utilizing a pulse oximeter as a substitute for an ABG analyzer to measure oxygenation is unsuitable.
In a study by Kanai et al. [10] that comprised 20,717 ABG samples from ICU patients, SpO2 likewise had a greater value when compared to SaO2. For the purpose of preventing hypoxemia, reducing mortality, and improving morbidity, the researchers advised keeping SpO2 above 92%. As the primary goal of the current study was to analyze the two methods separately, it was contrasted with the aforementioned study. In a similar way, several studies show that pulse oximeters are untrustworthy at low saturations because, as SaO2 declines, bias increases and precision (the standard deviation of the differences) decreases, increasing SpO2's SaO2 overestimation. There are many explanations offered as to why pulse oximeter performance is so subpar at low saturations. One of the reasons is the little variations in light-emitting diode output wavelength, which at low saturations result in proportionately larger errors. Another is that measurements of transmitted red light as opposed to infrared light at low saturations yield proportionally bigger inaccuracy due to the high extinction coefficient of reduced hemoglobin.
Elhossieny [11] conducted a study to identify the factors affecting the accuracy of pulse oximetry (SpO2) relative to arterial oxygen saturation SaO2. The study included 70 patients who were admitted to the surgical and emergency ICU. The results showed that there is a reduced gap between SpO2 and SaO2 in ABGs, indicating that SpO2 is more accurate than SaO2. Age, concurrent hypoxia, mechanical ventilation, vasopressor usage, oxygen mask use, and other factors all affected accuracy; sex and sepsis, however, had no statistically significant effects.
A study by Rauniyar et al. [12] examined ABG and pulse oximetry measurements of arterial oxygen saturation in ICU patients. Out of 101 individuals, 26 (25.7%) were hypoxemic based on SpO2 readings. The mean SaO2 and SpO2 values were 93.22% and 92.85%, respectively. Among those with SpO2 below 90%, mean SaO2 and SpO2 were 91.63% and 87.42%, respectively. For SpO2 below 80%, mean SaO2 and SpO2 were 63.40% and 71.80%, respectively. Non-hypoxic individuals had mean SpO2 and SaO2 of 95.773% and 95.654%, respectively. The agreement rate between SpO2 and SaO2 was 83.2%, and the sensitivity and specificity of pulse oximetry were 84.6% and 83%, respectively. The study suggests that pulse oximetry is highly accurate for measuring oxygen saturation (SpO2 > 90%) and can replace ABG measurements.
The purpose of the 2016 study conducted by Pentakota and Amalakanti [13] aimed to assess the effectiveness of portable pulse oximetry in identifying hypoxemia in emphysema and chronic bronchitis among 50 COPD patients in an Indian hospital. The researchers compared arterial oxygen saturation (SaO2) measured by a blood gas analyzer with oxygen saturation (SpO2) assessed by a pulse oximeter. Pulse oximetry showed promise, with high sensitivity (84.60%) in identifying hypoxemia in respiratory failure. However, in the comparison between chronic bronchitis and emphysema patients, pulse oximetry had lower sensitivity (82% vs 85%) and positive predictive value (69% vs 85%) in chronic bronchitis. The study concluded that invasive ABG analysis was more accurate than pulse oximetry. It also noted the higher variability of pulse oximeter readings in chronic bronchitis patients, suggesting limitations in using pulse oximetry as a diagnostic tool for this condition.
In a 2022 study by Rose et al. [14], the accuracy of pulse oximetry (SpO2) in determining oxygen saturation levels (SaO2) was investigated in COVID-19 patients admitted to ICUs and HDUs. The study found that SpO2 readings alone are not entirely reliable for determining SaO2 levels in COVID-19 patients. The researchers observed a bias between SpO2 and SaO2 measurements, with a mean difference of 0.86%. The study concluded that ABG analysis should be used alongside pulse oximetry to accurately assess oxygen saturation levels in COVID-19 patients in ICU/HDU settings.
In the current experiment, there was a significant connection between the two testing techniques when oxygenation saturation rose above 90%. ABG may be substituted with pulse oximetry in patients with acute diseases and SpO2 levels below 90% due to pulse oximetry’s high SaO2 estimation accuracy. According to the study, a pulse oximeter cannot effectively replace an ABG analyzer in patients with SpO2 < 90% since it is unable to precisely quantify SaO2 or gauge oxygenation. Numerous research has looked at the reliability of comparing the oxygen saturation measured by pulse oximetry with ABG [15,16].
Strengths of the study
Our study has several advantages. Each patient had a single measurement taken from a pair of them. There was essentially no time difference between the two measures because they were obtained concurrently, and the ABG analysis was completed right away. Fluctuations in oxygen levels that might have occurred over time were eliminated because the measurements were collected concurrently. This helped to enhance the validity and eliminate any bias from repeatedly measuring the same subject to acquire data. SpO2 can overstate SaO2, particularly in people with dark skin when saturation is poor.
Limitations
For instance, there may have been temperature control problems as a consequence of pinched or blocked tubing inside the analyzer or as a result of blockages or gaps in the sample stream. When compared to analyzers in a core or satellite laboratory, point-of-care testing equipment may be more susceptible to analytical errors and have a greater rate of error. Small sample size. All blood gas analyzers need to be maintained regularly by the manufacturer's instructions; the schedule should be modified to fit the requirements of the specific facility. This involves regular upkeep that is carried out daily, weekly, or monthly. If there are issues with quality control or worries about the analyzer's performance, corrective maintenance is necessary. The following list includes some of the recognized errors that might occur in our investigation. In accordance with the information given above, pulse oximetry is a painless, non-invasive, cost-effective continuous monitoring technique that can be used as a good substitute for an ABG, especially when SpO2 is below 90%, and has an accuracy of about 83.2%. However, it is not a good substitute for determining arterial oxygen saturation in situations with low oxygen saturation (SpO2-90%) and a critical state and is a poor predictor of hypoxemia.
Recommendations
More study is required to determine the accuracy and effectiveness of pulse oximetry in a large sample. Furthermore, although pulse oximetry evaluates arterial oxygenation, ventilation is not assessed by it. Therefore, an ABG assay is required to confirm any clinical suspicion of alveolar hypoventilation.
Conclusions
Thus, the study's findings show that, despite the portable, recently developed pulse oximeter's great accuracy in measuring oxygen saturation with SpO2 > 90%, the ABG method continues to be the Gold Standard process for reliably assessing clinical parameters under all circumstances. The study contends that, in contrast to the ABG analyzer, the precise determination of SaO2 and the assessment of hypoxemia by pulse oximeter is not a good option in patients with SpO2 < 90%. As a result, if alveolar hypoventilation is clinically suspected, pulse oximetry needs to be supported by ABG studies.
The authors have declared that no competing interests exist.
Human Ethics
Consent was obtained or waived by all participants in this study
Animal Ethics
Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.
References
- 1.Pulse oximetry. Jubran A. Crit Care. 1999;3:0–7. doi: 10.1186/cc341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Evaluation of hemodynamic and SpO2 variability during different stages of periodontal surgery. Amoian B, Rabíee M, Aghvami M, Milani S. J Indian Soc Periodontol. 2013;17:612–616. doi: 10.4103/0972-124X.119274. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.British Thoracic Society guidelines for home oxygen use in adults. Hardinge M, Annandale J, Bourne S, et al. Thorax. 2015;70 Suppl 1:0–43. doi: 10.1136/thoraxjnl-2015-206865. [DOI] [PubMed] [Google Scholar]
- 4.Comparison of pulse oximeter with arterial blood gas analysis in determination of SpO2 in COVID 19 at a tertiary care centre in central India. Priya. N, Ranjit. A, Ujwalla. G, et al. https://www.nveo.org/index.php/journal/article/view/564 Nat Volatiles Essent. 2021;8:1191–1197. [Google Scholar]
- 5.The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study. Wilson BJ, Cowan HJ, Lord JA, Zuege DJ, Zygun DA. BMC Emerg Med. 2010;10:9. doi: 10.1186/1471-227X-10-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults. Jensen LA, Onyskiw JE, Prasad NG, et al. Heart Lung. 1998;27:387–408. doi: 10.1016/s0147-9563(98)90086-3. [DOI] [PubMed] [Google Scholar]
- 7.Accuracy of two pulse oximeters at low arterial hemoglobin-oxygen saturation. Carter BG, Carlin JB, Tibballs J, Mead H, Hochmann M, Osborne A. Crit Care Med. 1998;26:1128–1133. doi: 10.1097/00003246-199806000-00040. [DOI] [PubMed] [Google Scholar]
- 8.Accuracy of a pulse oximeter in the measurement of the oxyhaemoglobin saturation. Chiappini F, Fuso L, Pistelli R. https://erj.ersjournals.com/content/11/3/716.short. Eur Respirat J. 1998;11:716–719. [PubMed] [Google Scholar]
- 9.Correlation of arterial blood gas measurements with venous blood gas values in mechanically ventilated patients. Razi E, Nasiri O, Akbari H, Razi A. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4153216/ Tanaffos. 2012;11:30–35. [PMC free article] [PubMed] [Google Scholar]
- 10.Reliability of oxygen saturation by pulse oximeter to detect hypoxemia of patients in the intensive care unit. Kanai R, Moriyama K, Kohyama T, et al. Anesthesia Analgesia. 2012;114:159. [Google Scholar]
- 11.Accuracy of pulse oximetry in comparison with arterial blood gas. Elhossieny KM. https://journals.lww.com/raic/Fulltext/2020/07010/Accuracy_of_pulse_oximetry_in_comparison_with.8.aspx Res Opin Anesth Intens Care. 2020;7:51–56. [Google Scholar]
- 12.Study of oxygen saturation by pulse oximetry and arterial blood gas in ICU patients: a descriptive cross-sectional study. Rauniyar NK, Pujari S, Shrestha P. JNMA J Nepal Med Assoc. 2020;58:789–793. doi: 10.31729/jnma.5536. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Pulse oximetry overestimates oxygen saturation in COPD. Amalakanti S, Pentakota MR. https://rc.rcjournal.com/content/61/4/423.short. Respirat Care. 2016;61:423–427. doi: 10.4187/respcare.04435. [DOI] [PubMed] [Google Scholar]
- 14.An observational study of simultaneous pulse oximetry and arterial oxygen saturation readings in intensive care unit/ high dependency unit in COVID-19 patients. Rose N, Sriram R, Sudheer R, et al. Asian J Med Sci. 2022;13:18–22. [Google Scholar]
- 15.Talwar D, Kumar S, Talwar D. IntechOpen; 2022. Non-Invasive Ventilation in Acute Hypoxemic Respiratory Failure. [Google Scholar]
- 16.Noninvasive ventilation in acute hypoxic respiratory failure in medical intensive care unit: a study in rural medical college. Bajaj A, Kumar S, Inamdar AH, Agrawal L. Int J Crit Illn Inj Sci. 2019;9:36–42. doi: 10.4103/IJCIIS.IJCIIS_40_18. [DOI] [PMC free article] [PubMed] [Google Scholar]