. 2023 Mar 2;33(3):591–610. doi: 10.1007/s00062-023-01271-5

Computed Tomography of the Head

A Systematic Review on Acquisition and Reconstruction Techniques to Reduce Radiation Dose

Michael Dieckmeyer 1,2,, Nico Sollmann 2,3,4, Karina Kupfer 2, Maximilian T Löffler 2,5, Karolin J Paprottka 2, Jan S Kirschke 2,3, Thomas Baum 2
PMCID: PMC10449676  PMID: 36862232

Abstract

In 1971, the first computed tomography (CT) scan was performed on a patient’s brain. Clinical CT systems were introduced in 1974 and dedicated to head imaging only. New technological developments, broader availability, and the clinical success of CT led to a steady growth in examination numbers. Most frequent indications for non-contrast CT (NCCT) of the head include the assessment of ischemia and stroke, intracranial hemorrhage and trauma, while CT angiography (CTA) has become the standard for first-line cerebrovascular evaluation; however, resulting improvements in patient management and clinical outcomes come at the cost of radiation exposure, increasing the risk for secondary morbidity. Therefore, radiation dose optimization should always be part of technical advancements in CT imaging but how can the dose be optimized? What dose reduction can be achieved without compromising diagnostic value, and what is the potential of the upcoming technologies artificial intelligence and photon counting CT? In this article, we look for answers to these questions by reviewing dose reduction techniques with respect to the major clinical indications of NCCT and CTA of the head, including a brief perspective on what to expect from current and future developments in CT technology with respect to radiation dose optimization.

Keywords: Iterative reconstruction, Dose reduction, Low dose, Radiation exposure, Head CT, CT angiography

Key Points

  1. Advanced iterative reconstruction enables significant dose reduction in head CT.

  2. For CTA of the head, the combination with lower CM volume can be particularly effective for improved patient care.

  3. AI and PCCT can be expected to enable additional dose reductions of at least 40%.

  4. For novel techniques, careful evaluation of image quality and diagnostic performance is essential to ensure diagnostic benefits head CT.

Introduction

Over the last decades, we have seen a steady growth in the number of computed tomography (CT) examinations [13]. New technological developments, broader availability of the required hardware and software, as well as physician and patient demands are the main reasons for increased clinical application of CT [1, 2]. Although magnetic resonance imaging (MRI) has developed tremendously over the last decades, in particular for head imaging, and has become more available, CT still is the workhorse of head imaging for multiple indications, including non-contrast CT (NCCT) in cerebral ischemia and stroke, assessment of intracranial hemorrhage (ICH), headaches, or acute neurologic deficits, as well as for first-line diagnostics in loss of consciousness and trauma evaluation in the emergency setting [4]. Detailed assessment of cerebral blood supply has become the clinical standard through the use of CT angiography (CTA) and CT perfusion (CTP) [5, 6]. Therefore, CT of the head is essential for fast and accurate diagnosis, optimized patient management and treatment. Besides MRI, which is generally used in less acute clinical settings, CT makes up a large proportion of the daily neuroradiological workload in head imaging.

However, CT comes with the inherent downside of ionizing radiation, which may cause radiation-induced malignancies [7, 8]. In the USA, approximately 2% of future cancer cases are assumed to be attributable to the current application of medical imaging [3, 9], and CT exposure was estimated to be responsible for 1% of total cancer mortality [10]. Therefore, the principle to keep radiation exposure as low as reasonably achievable (ALARA) is fundamental [11, 12].

Unfortunately, CT-related radiation exposure in daily clinical routine still shows strong inter-institutional and intra-institutional variations, as well-defined and universal reference standards are usually not available [13, 14]. General recommendations are complicated to determine considering different scanner models and technologies, which considerably influence radiation exposure. Using the technique only when the clinical value outweighs risks and costs, and restricting the scan volume to the clinical task are undoubtedly the most effective ways to limit CT-related radiation exposure in order to protect patients. Besides a careful risk-benefit assessment, developments on both the acquisition and the reconstruction side have led to an optimized trade-off between image quality (IQ) and radiation exposure [15, 16]. A widely applied dose reduction approach is the optimization of acquisition parameters, including tube voltage, tube current, and contrast medium (CM) volume. Tube current is expressed either directly (as mA) or indirectly in terms of tube current-time product (as mAs). Modern clinical CT systems use automatic exposure control (AEC) by means of automatic tube current modulation (ATCM), which is based on patient habitus, z‑axis modulation, and rotational modulation [1719].

Increased image noise and artifacts are the drawbacks of most dose reduction methods but could at least partially be compensated by adequate image reconstruction techniques. Developments in iterative reconstruction (IR) techniques have led to major improvements in recent years, which have therefore become an essential component of CT dose reduction [16]. Before the emergence of IR, filtered back projection (FBP) was the standard image reconstruction for clinical CT. This fast and robust method is based upon the exact mathematical relation between measured projection and reconstructed image data [16, 20]. However, particularly for dose-reduced acquisitions, FBP-reconstructed images can suffer from low quality, as noise-free data are assumed and image noise is amplified in the filtering process. In contrast, IR can reduce image noise through iterative filtering or modeling of data acquisition physics [16]. Currently, hybrid IR (HIR), such as iDose (Philips Healthcare, Best, The Netherlands), ASIR (GE Healthcare, Milwaukee, WI, USA), or SAFIRE (Siemens Healthineers, Erlangen, Germany), is the most frequently used IR technique, providing increased reduction of image noise and reconstruction time through iterative filtering of both projection and image data. Benefits in terms of achievable IQ, in particular at low doses, are only exceeded by model-based IR (MBIR) algorithms, which use advanced models in an iterative process of backward and forward projections. However, the exceptional level of noise reduction and achievable dose reduction comes at the cost of high computational efforts, limiting accessibility and usability in daily clinical practice. Yet, artificial intelligence (AI) could help to overcome this limitation by using a convolutional neural network (CNN) trained with (simulated) low-dose (LD) data to reconstruct standard-dose (SD) high-quality CT images [2124].

Although the use of dose reduction techniques for head CT is increasing, they have not yet been systematically reviewed. Therefore, the purpose of this article was to review clinical applications of dose reduction and LD techniques for NCCT and CTA of the head. The objective was to analyze achievable dose reductions, effects on IQ, and diagnostic performance with respect to the most relevant pathologies that are frequently assessed by head CT applications. Availability and quality of the included literature has a high degree of heterogeneity with respect to dose reduction methods, reported dose parameters, and clinical applications. Therefore, we concentrated on five major clinical indications: (i) NCCT of cerebral ischemia and stroke, (ii) NCCT of intracranial hemorrhage, (iii) NCCT of the head without specified indications, (iv) CTA of intracranial aneurysms, and (v) CTA of other cerebrovascular diseases and carotid artery disease.

Material and Methods

Search Strategy

A search of PubMed (http://www.ncbi.nlm.nih.gov/pubmed) was performed to identify studies evaluating methods to reduce radiation dose for NCCT and CTA of the head with respect to the following clinical indications: (i) NCCT of cerebral ischemia and stroke, (ii) NCCT of ICH, (iii) CTA of intracranial aneurysms, and (iv) CTA of other cerebrovascular diseases and carotid artery disease. Additionally, NCCT studies without specified indications were identified by screening of references of included studies. The search was conducted by two persons (radiologists with 7 and 4 years of experience, respectively) without a beginning search date (search end date 19 July 2022). Uncertainties about inclusion of a respective article, if present, were resolved by consensus through discussion with a third person (board-certified consultant in radiology, 11 years of experience).

The literature search was performed according to the Preferred Reporting Items for Systematic reviews and Meta-Analyses (PRISMA) guidelines (Fig. 1; [25, 26]). The used search terms for PubMed are available in the appendix.

Fig. 1.

Fig. 1

PubMed search flow diagram according to the Preferred Reporting Items for Systematic reviews and Meta-Analyses (PRISMA) guidelines [25, 26]

Inclusion Criteria

Studies were included if they met the following inclusion criteria: (1) study population: studies performed in humans including adult or pediatric patients; (2) study design: retrospective or prospective; (3) indications and scanning type: diagnostic NCCT for present or suspected intracranial pathology, or CTA for evaluation of the intracranial vasculature and/or carotid arteries and (4) purpose: comparison of LD to SD protocols using CT data acquired at different dose levels, CT data acquired at a single dose level and additionally simulated at different dose levels, or CT data including a dose comparison between patient subgroups.

Exclusion Criteria

Studies were not considered if they met the following exclusion criteria: (1) article type: case reports, case series, conference abstracts, letters, editorials, reviews, meta-analyses, or surveys; (2) language of publication other than English; (3) studies in cadavers, phantoms, or animals and (4) studies with other purposes (e.g., exclusive comparison of CM volumes, comparison of shielding techniques, medical staff radiation exposure reports).

Extraction of Data

The following information was extracted from the selected articles: (1) author(s); (2) year of publication; (3) number of subjects (n) of the entire study and relevant patient subgroups (e.g., SD group, LD group); (4) details on group comparisons (if applicable); (5) details on the used CT system, including number of detector rows, vendor, and model name; (6) image acquisition parameters; (7) image reconstruction algorithms and parameters; (8) dose reduction (in %) and (9) reported dose values: CT dose index (CTDIvol), dose length product (DLP), and/or effective dose (E).

Results

Study Selection

The search via PubMed resulted in 1469 publications after removal of duplicates. During screening of titles and abstracts, 1333 records were discarded. The assessment of full-text articles led to the removal of 107 records, and the reference screening of included studies led to the addition of 8 articles, resulting in 37 publications that were included in the qualitative synthesis for this systematic review (Fig. 1).

Study Characteristics

The 37 selected studies covered NCCT of cerebral ischemia and stroke (n = 6), NCCT of ICH (n = 4), NCCT without specified indications (n = 8), CTA of intracranial aneurysms (n = 10), and CTA of other cerebrovascular diseases (n = 10).

Patients

The total number of subjects (n) as well as the number of subjects in the SD group(s) and LD group(s) were extracted. Furthermore, the number of included CT examinations or subject numbers for relevant subgroups (e.g., used CT system, administered CM dose) were extracted when provided. Total numbers ranged from n = 20 [27] to n = 305 patients [28].

CT System, Acquisition and Reconstruction Parameters

All except one of the studies used multi-detector CT (MDCT). One study published in 2005 used single-detector CT [29]. Tube voltages from 70 to 150 kV were used. Dose reduction in LD protocols was achieved by a reduction of tube voltage, tube current, or by a combination of both. Reduced tube currents were determined using different approaches: (i) fixed mA values or ranges, or (ii) reference mA values or ranges in the case of ATCM. Reporting of mA was heterogeneous, including reference values, mean or median values, and ranges. Thus, statistics on the reported numbers would not be meaningful to present. As an alternative or in addition to tube current, some studies reported tube current-time products, which take into account the exposure time. As tube current-time product is proportional to dose, the reported mAs values can to a certain extent be considered a measure of radiation exposure.

Image reconstruction by FBP was reportedly used in 20 studies. The use of IR gradually increased with the year of the publication, including IR in image space (IRIS, n = 2), HIR (n = 23), and MBIR (n = 6). IR was used to create LD protocols and compared to SD protocols with FBP in 13 studies [27, 3041]. The reconstruction technique was not reported in 10 studies. These studies were published in 2016 or earlier, making it reasonable to assume that FBP was used.

Dose Reporting and Dose Reduction Calculation

Studies reported doses as CTDIvol (n = 33), DLP (n = 29), and E (n = 26). Specifically, E is derived by multiplying the DLP with a conversion factor for a specific CT examination and is commonly regarded as the most appropriate indicator of stochastic radiation risk. Different DLP to E conversion factors were used from previous publications, which depended on patient age, scanned anatomical region, acquisition parameters, and time of publication. Comparable conversion factors for the head between 0.0021 and 0.0024 mSv/(mGy * cm) were used in the majority of studies [4246], while two studies used conversion factors for the head and neck of 0.0031 and 0.0048 mSv/(mGy * cm) [38, 41], and one study used a dedicated dose calculation tool for DLP to E conversion [27].

Dose reductions were explicitly reported or retrospectively calculated from reported dose values. Achieved dose reductions ranged from 10–87%, not taking into account simulated LD studies. Two studies used simulated LD data, either by virtually lowered tube currents [47] or artificial noise insertion into CT projection data [48]. Dose values were reported as mean (with or without standard deviation) or median (with or without minimum, maximum, and interquartile ranges). The mean was extracted in favor of median when both values were provided. In Tables 1234 and 5 dose values are provided for the SD group(s), LD group(s), and other subgroups (CM volume, used CT system) where reasonably applicable.

Table 1.

Dose reduction in NCCT of cerebral ischemia and stroke

Author Year Subjects
(n)
Comparison CT system Acquisition parameters Reconstruction name (level) Dose reduction (%) CTDIvol (mGy) DLP
(mGy * cm)
E
(mSv)
Paprottka [49] 2021

131

131SD

131LD

SD vs. LD 128-MDCT (Philips Ingenuity Core, Philips Healthcare, Best, The Netherlands)

120 kV, 343 mA, 300 mAsSD

120 kV, 229 mA, 200 mAsLD

HIR (iDose 4)

MBIR (IMR 3)

33CTDI

34DLP

46.6SD

31.2LD

673.6SD

441.9LD

NA
Kapur [50] 2021

423

391SD

32LD

SD vs. LD 256-MDCT (Philips iCT, Philips Healthcare, Best, The Netherlands)

120 kV, 330 mAsSD

120 kV, 280 mAsLD

HIR (iDose 4)

10CTDI

8DLP

8E

52.4SD

47.4LD

1061.9SD

972.9LD

2.6SD

2.4LD

Bricout [51] 2015

26

13SD

13LD

SD vs. LD 64-MDCT (Siemens SOMATOM Definition AS, Siemens Healthineers, Erlangen, Germany)

120 kV, 350 mAsam, SD

100 kV, 400 mAsam, LD

HIR (SAFIRE 1)

23CTDI

21DLP

42.6SD

33.0LD

662.0SD

524.0LD

1.4SD

1.1LD

Bodelle [52] 2015

51

30SD

21LD

SDFBP vs. LDFBP, HIR 256/128-MDCT (Siemens SOMATOM Definition Flash/AS, Siemens Healthineers, Erlangen, Germany)

120 kV, 340 mAsSD

120 kV, 260 mAsLD

FBP

HIR (SAFIRE 1–5)

22E NA NA

2.3SD

1.8LD

Ben-David [53] 2014

30

30LD

30SD

SD vs. LD 64-MDCT (Philips Brilliance, Philips Healthcare, Best, The Netherlands)

120 kV, 445 mAsSD

80 kV, 1200 mAsLD

NA 22CTDI

59.0SD

46.0LD

NA NA
Zacharia [54] 2011

200

100SD

100LD

SD vs. LD 16-MDCT (Siemens Sensation, Siemens Healthineers, Erlangen, Germany)

120 kV, 300 mAsSD

120 kV, 300 mAsam, LD

NA

36CTDI

35DLP

NA NA NA

am automatic tube current modulation, CTDI based on volumetric CT dose index, E based on effective dose, FBP filtered back projection, HIR hybrid iterative reconstruction, LD low-dose, MBIR model-based iterative reconstruction, SD standard-dose

Table 2.

Dose reduction in NCCT of intracranial hemorrhage

Author Year Subjects (n) Comparison CT System Acquisition parameters Reconstruction name (level) Dose reduction (%) CTDIvol (mGy) DLP
(mGy * cm)
E
(mSv)
Fletcher [48] 2019 83 SD vs. LDS2 128-MDCT (Siemens SOMATOM Definition FLASH/AS+, Siemens Healthineers, Erlangen, Germany)

120 kV, 250 eff. mAs SD

120 kV, 25–200 eff. mAsLD

FBP

HIR (SAFIRE 2)

20–90

38.1SD

3.8–30.5LD

NA NA
Scholtz [55] 2017

123

36SD, SECT

30SD, DECT

32LD, SECT

25LD, DECT

SDSECT vs

SDDECT vs. LDSECT vs. LDDECT

192-MDCT (Siemens, Siemens Healthineers, Erlangen, Germany)

120 kV, 270 mAs SECT

80/150 kV, 410/273 mAs DECT

120 kV, 270 mAsam, SECT

80/150 kV, 410/273 mAsam, DECT

MBIR (ADMIRE)

18CTDI, SECT

27CTDI, DECT

25DLP, SECT

24DLP, DECT

39.5SD, SECT

41.0SD, DECT

32.2LD, SECT

30.0LD, DECT

771.5SD, SECT

770.6SD, DECT

575.0LD, SECT

587.0LD, DECT

NA
Kaul [37] 2016

78

18SD

22LD1

20LD2

18LD3

SDFBP vs. LD1FBP vs. LD2ASIR20 vs. LD3ASIR30/40 64-MDCT (GE Lightspeed VCT, GE Healthcare, Milwaukee, WI, USA)

120 kV, 100–300 mAam, SD

100 kV, 100–300 mAam, LD1–3

FBP

HIR (ASIR 20–40%)

20/43/66CTDI

23/34/64DLP

31.9SD

25.4LD1

18.3LD2

10.8LD3

396.0SD

304.0LD1

260.0LD2

141.0LD3

NA
Bodelle [56] 2014

94

54SD

40LD

SD vs. LD 128-MDCT (Siemens SOMATOM Definition Flash/AS, Siemens Healthineers, Erlangen, Germany)

120 kV, 340 mAsSD

120 kV, 260 mAsLD

FBP

HIR (SAFIRE 1–5)

29DLP NA

1045.0SD

744.0LD

2.4SD

1.7LD

am automatic tube current modulation, ASIR20/30/40 adaptive statistical iterative reconstruction at level 20/30/40% (available options: 0–50% in 10% increments), CTDI based on CTDIvol, DECT dual energy CT, LD low-dose, LD1 low-dose protocol 1, LD2 low-dose protocol 2, LD3 low-dose protocol 3, S2 10/20/50/80% of SD corresponding to 25/50/100/200 eff. mAs using simulated noise insertion into CT projection data, SD standard-dose, SECT single energy CT

Table 3.

Dose reduction in other NCCT of the head

Author Year Subjects (n) Comparison CT System Acquisition parameters Reconstruction name (level) Dose reduction (%) CTDIvol
(mGy)
DLP
(mGy * cm)
E
(mSv)
Kaul [36] 2016

177

71SD

86LD1

20LD2

SDFBP vs. LD1ASIR20 vs. LD2ASIR30 64-MDCT (GE Lightspeed VCT, GE Healthcare, Milwaukee, WI, USA)

120 kV, 100–300 mAam, SD

120 kV, 100–300 mAam, LD

FBP

HIR (ASIR 20%, 30%)

41LD1

73LD2

51.6SD

30.2LD1

13.9LD2

768.0SD

455.0LD1

204.0LD2

1.6SD

1.1LD1

0.4LD2

Ozdoba [34] 2014

75

50SD

25LD

SDFBP vs. LDHIR

16/64-MDCT (Siemens SOMATOM Sensation, Siemens Healthineers, Erlangen, Germany)SD

128-MDCT (Siemens Somatom Definition Edge, Siemens Healthineers, Erlangen, Germany)LD

120 kV, 220/380 mAsSD

100 kV, 230 mAsLD

FBP

HIR (SAFIRE)

40

57.2SD

34.9LD

964.0SD

574.1LD

NA
Korn [33] 2013

60

30SD

30LD

SDFBP vs. LDFBP, HIR 128-MDCT (Siemens Somatom Definition Flash, Siemens Healthineers, Erlangen, Germany)

120 kV, 320 mAsSD

120 kV, 255 mAsLD

FBP

HIR (SAFIRE 3)

20

59.7SD

47.8LD

1042.0SD

829.0LD

2.2SD

1.7LD

Kilic [28] 2013

305

152SD

153LD

SDFBP vs. LDASIR 16-MDCT (GE Lightspeed VCT)

100/120 kV, 200–420 mASD

100/120 kV, 100–290 mALD

FBP

HIR (ASIR 30%)

30

38.8pf/29.0cer SD

27.7pf/20.2cer LD

466.5SD

329.2LD

2.2SD

1.5LD

Korn [57] 2012

90

30SD

30LD1

30LD2

SD vs. LD1 vs. LD2 128-MDCT (Siemens SOMATOM Definition Flash, Siemens Healthineers, Erlangen, Germany)

120 kV, 320 mAsSD

120 kV, 275 mAs LD1

120 kV, 225 mAs LD2

FBP

IRIS

15SD-LD1

30SD-LD2

60.1SD

51.8LD1

42.3LD2

1043.0SD

890.0LD1

733.0LD2

2.2SD

1.8LD1

1.5LD2

Rapalino [32] 2012

150

50SD

100LD

SDFBP vs. LDASIR 64-MDCT (GE Discovery CT750HD, GE Healthcare, Milwaukee, WI, USA)

120 kV, 250 mA, 175 mAsSD

120 kV, 200 mA, 140 mAsLD

FBP

HIR (ASIR 20–100%)

26

66.5SD

49.7LD

1270.3SD

932.3LD

2.7SD

2.0LD

Becker [31] 2012

150

50SD

50LD1

50LD2

SDFBP vs. LD1FBP vs. LD2IRIS MDCT (Siemens Somatom Definition Flash)

120 kV, 320 mAsSD

120 kV, 390 mAsam, LD1, LD2

FBP

IRIS

24CTDI

20DLP

60.0SD

46.0LD1

45.0LD2

887.0SD

722.0LD1

708.0LD2

NA
Kilic [30] 2011

149

51SD

98LD

SDFBP vs. LDASIR 16-MDCT (GE Brightspeed, GE Healthcare, Milwaukee, WI, USA)

140pf/120cer kV, 170pf/270cer mA SD

140pf/120cer kV, 125pf/190cer mA SD

FBP

HIR (ASIR 30%)

26pf, CTDI

35cer, CTDI

31DLP

93.5pf, SD

59.4cer, SD

69.1pf, LD

38.6cer, LD

1081.3SD

748.6LD

2.3SD

1.6LD

am automatic tube current modulation, ASIR adaptive statistical iterative reconstruction, ASIR20/30 adaptive statistical iterative reconstruction at level 20/30/40% (available options: 0–50% in 10% increments), cer cerebrum, CTDI based on CTDIvol, FBP filtered back projection, HIR hybrid iterative reconstruction, IRIS iterative reconstruction in image space, LD low-dose, LD1 low-dose protocol 1, LD2 low-dose protocol 2, pf posterior fossa, SAFIRE sinogram-affirmed iterative reconstruction, SAFIRE1–5 SAFIRE at level 1–5, SD standard-dose, SD-LD1 dose reduction between standard-dose protocol and low-dose protocol 1, SD-LD2 dose reduction between standard-dose protocol dose and low-dose protocol

Table 4.

Dose reduction in CTA of intracranial aneurysm

Author Year Subjects (n) Comparison CT System Acquisition parameters Reconstruction name (level) Dose reduction (%) CTDIvol (mGy) DLP
(mGy * cm)
E
(mSv)
Chen [40] 2017

100

50SD

50LD

SDFBP, SC vs. LDHIR, LC 128-MDCT (Siemens SOMATOM Definition Flash, Siemens Healthineers, Erlangen, Germany)

120 kVam, SD

70 kVam, LD

FBP

HIR (SAFIRE 4)

81CTDI

82DLP

33.4SD

6.4LD

629.1SD

116.0LD

1.3SD

0.2LD

Nagayama [39] 2017

75

37SD

38LD

SDFBP, SC vs. LDHIR, LC 128-MDCT (Siemens SOMATOM Definition AS+, Siemens Healthineers, Erlangen, Germany)

120 kV, 350 mAsam, SD

80 kVATVS, 431 mAsam, LD

FBP

HIR (SAFIRE 3)

65CTDI

62E

41.8SD

14.8LD

NA

1.6SD

0.6LD

Ni [58] 2016

204

102SD

102LD

SDSC vs. LDLC 128-MDCT (Siemens SOMATOM Definition, Siemens Healthineers, Erlangen, Germany)

120 kV, 230 mAam, SD

80 kV, 230 mAam, LD

NA

73CTDI

73DLP

25.9SD

7.0LD

507.0SD

136.7LD

1.1SD

0.3LD

Yang [59] 2016

80

40SD

40LD

SD vs. LD 128-DS-MDCT (Siemens SOMATOM Definition Flash)

80/sn140 kV, 300/150 mAsSD

80/sn140 kV, 200/100 mAsLD

FBP

HIR (SAFIRE 3)

34CTDI

31DLP

20.6SD

13.6LD

378.3SD

259.4LD

0.8SD

0.5LD

Tang [60] 2015

294

148SD

146LD

SD vs. LD 64-MDCT (GE Lightspeed, GE Healthcare, Milwaukee, WI, USA)

120 kV, 641 mASD

100 kV, 380 mALD

NA

36CTDI

36DLP

55.7SD

35.9LD

932.6SD

594.8LD

2.0SD

1.3LD

Chen [35] 2015

100

50SD

50LD

SDFBP vs. LDHIR 128-MDCT (Siemens SOMATOM Definition Flash)

120 kVam, SD

70 kVam, LD

FBP

HIR (SAFIRE 1–5)

80CTDI

81DLP

33.7SD

6.6LD

609.9SD

118.0LD

1.3SD

0.2LD

Luo [61] 2014

120

40SD

40LD1

40LD2

SD vs. LD1 vs. LD2 128-MDCT (Siemens SOMATOM Definition)

120 kV, 230 mAsam, 70 ml CM SD

100 kV, 230 mAsam, 30 ml CM LD1

80 kV, 230 mAsam, 30 ml CM LD2

NA

45SD-LD1, CTDI

73SD-LD2, CTDI

51LD1-LD2, CTDI

44SD-LD1, DLP

74SD-LD2, DLP

52LD1-LD2, DLP

26.1SD

14.4LD1

7.0LD1

515.0SD

286.0LD

136.0LD2

1.1SD

0.6LD

0.3LD2

Kidoh [62] 2013

98

32 SD, SC

33 SD, LC

33 LD, LC

SD vs. LD

SC vs. LC

256-MDCT (Brilliance iCT, Philips Healthcare, Best, The Netherlands)

100 kV, 641 mA, 800 eff. mAs SD

80 kV, 923 mA, 1365 eff. mAsLD

370 mg/kg CM SC

296 mg/kg CM LC

NA 16CTDI

59.2SD

49.7LD

NA NA
Sun [63] 2012

48

24SD

24LD

SD vs. LD 320-MDCT (Aquilion One, Toshiba Medical Systems, Canon Medical Systems Corporation, Ōtawara, Tochigi, Japan)

120 kV, 200 mA, 150 mAsSD

80 kV, 200 mA, 150 mAsLD

NA

70CTDI

70DLP

54.6SD

16.6LD

876.6SD

265.6LD

1.8SD

0.6LD

Waaijer [64] 2007

40

20SD

20LD

SD vs. LD 16 MDCT (MX 8000 IDT, Philips Healthcare, Best, The Netherlands)

120 kV, 200 mAsSD

90 kV, 330 mAsLD

NA 24

27.2SD

20.6LD

NA NA

am automatic tube current modulation, ATVS Automated tube voltage selection with 120 kV reference, CM contrast medium, CTDI based on CTDIvol, E based on effective dose, FBP filtered back projection, HIR hybrid iterative reconstruction, LC low contrast medium concentration, LD low-dose, LD1 low-dose protocol 1 using 120 kV and 150 mAs, LD2 low-dose protocol 2 using 100 kV and 230 mAs, RD reference dose using 120 kV, RD-SD dose reduction between reference-dose protocol and standard-dose protocol, RD-LD dose reduction between reference-dose protocol and low-dose protocol, SAFIRE sinogram-affirmed iterative reconstruction, SAFIRE1–5 SAFIRE at level 1–5, SC standard contrast medium concentration, SD standard-dose, SD-LD dose reduction between standard-dose protocol and low-dose protocol, SD-LD1 dose reduction between standard-dose protocol and low-dose protocol 1, SD-LD2 dose reduction between standard-dose protocol dose and low-dose protocol 2, sn140 use of a selective proton shield (SPS) with a 140 kV X-ray tube with additional tin filtration, LD1-LD2 dose reduction between low-dose protocol 1 and low-dose protocol 2

Table 5.

Dose reduction in CTA of other cerebrovascular diseases

Author Year Subjects (n) Comparison CT System Acquisition parameters Reconstruction name (level) Dose reduction (%) CTDIvol (mGy) DLP
(mGy * cm)
E
(mSv)
Sollmann [47] 2019 30 SD vs. LDS1 128-MDCT (Philips Ingenuity Core; Philips Healthcare, Best, The Netherlands)

120 kV, 263 mA, 130 mAsam, SD

13–65 mAsLD

HIR (SIR) 50–90

8.5SD

0.9–4.2LD

NA NA
Annoni [65] 2019

205

100SD

105LD

SD vs. LD

64-MDCT (GE Discovery 750 HD, GE Healthcare, Milwaukee, WI, USA)SD

256-MDCT (GE Revolution, GE Healthcare, Milwaukee, WI, USA)LD

100 kV, 213–600 mAam, SD

80 kV, 120–500 mAam, LD

HIR (ASIR 50%) SD

HIR (ASIR‑V 50%)LD

87CTDI

86DLP

20.5SD

2.7LD

735.3SD

100.7LD

1.6SD

0.2LD

Wang [66] 2018

101

53SD

48LD

SD vs. LD 256-MDCT (Philips Brilliance iCT; Philips Healthcare, Best, The Netherlands)

100 kV, 220 mAsSD

80 kV, 220 mAsLD

FBP

HIR (iDose 4)

MBIR (IMR 1)

51SD-LD

39RD-SD

70RD-LD

16.9SD

8.3LD

27.6RD

540.0SD

267.2LD

884.4RD

1.1SD

0.6LD

1.9RD

Chen [41] 2018

121

39SD

41LD1

41LD2

SDFBP vs. LD1ASIR‑V vs. LD2 ASIR‑V 256-MDCT (GE Revolution)

120 kV, 10–350 mAam, SD

120 kV, 10–350 mAam, LD1

100 kV, 10–400 mAam, LD2

FBP

HIR (ASIR‑V 50%)

24SD-LD1, CTDI

37SD-LD2, CTDI

25SD-LD1, DLP

40SD-LD2, DLP

9.1SD

6.9LD1

5.7LD2

393.0SD

293.0LD1

236.0LD2

9.1SD

6.9LD1

5.7LD2

Leithner [67] 2018 43 SD vs. LD 192-MDCT (Siemens SOMATOM Force, Siemens Healthineers, Erlangen, Germany)

120 kV, 95/59 mAsam, SD

90 kV, 95 mAsam, LD

MBIR (ADMIRE 3) 40DLP

2.8SD

1.7LD

185.4SD

110.6LD

0.4SD

0.2LD

Chen [38] 2017

50

25SD

25LD

SDSC, FBP vs. LDLC, MBIR 192-MDCT (Siemens SOMATOM Force)

100 kV, 700 mAsam, 40 ml/kg CM SD

70 kV, 700 mAsam, 25 ml CMLD

FBP

MBIR (ADMIRE 3)

54CTDI

56DLP

4.6SD

2.1LD

190.8SD

84.2LD

0.9SD

0.4LD

Kayan [68] 2016

101

50SD

51LD

SDSC vs. LDLC 192-MDCT (Siemens Definition AS, Siemens Healthineers, Erlangen, Germany)

100 kV, 150 mAam, 1 ml/kg CM SD

80 kV, 150 mAam, 0.5 ml/kg CMLD

NA 48DLP NA

225.7SD

116.6LD

NA
Moloney [27] 2016 20 SDFBP, ASIR vs. LDASIR vs. LDMBIR 64-MDCT (GE Lightspeed VCT, GE Healthcare, Milwaukee, WI, USA)

100 kV, 60–230 mAam, SD

100 kV, 30–150 mAam, LD

FBP

HIR (ASIR)

MBIR (Veo)

50 NA

688.0SD

341.3LD

1.8SD

3.7LD

Bricout [51] 2015

26

13SD

13LD

SDSAFIRE1 vs. LDSAFIRE3 64-MDCT (Siemens SOMATOM Definition AS, Siemens Healthineers, Erlangen, Germany)

100 kV, 250 mAsam, SD

100 kV, 220 mAsam, LD

HIR (SAFIRE 1, 3)

31CTDI

26DLP

24.5SD

17.0LD

369.0SD

273.0LD

0.8SD

0.6LD

Bahner [29] 2005 29 SD vs. LD SDCT (Siemens SOMATOM Plus 4, Siemens Healthineers, Erlangen, Germany)

120 kV, 200 mA, 150 mAsSD

80 kV, 255 mA, 255 mASLD

NA

38CTDI

43E

21.9SD

13.5LD

NA

0.7SD

0.4LD

ADMIRE advanced modeled iterative reconstruction, am automatic tube current modulation, ASIR first generation version of ASIR, ASIR‑V third generation version of ASIR, CM contrast medium, CTDI based on CTDIvol, E based on effective dose, FBP filtered back projection, HIR hybrid iterative reconstruction, LC low contrast medium concentration, LD low-dose, LD1 low-dose protocol 1, LD2 low-dose protocol 2, MBIR model-based iterative reconstruction, RD reference dose protocol, S1 10/25/50% of SD using simulated lower tube currents, SC standard contrast medium concentration, SD standard-dose, SD-LD dose reduction between standard-dose protocol and low-dose protocol, SD-LD1 dose reduction between standard-dose protocol and low-dose protocol 1, SD-LD2 dose reduction between standard-protocol dose and low-dose protocol 2, SIR statistical iterative reconstruction

Outcome Measures

Quantitative Measures

Quantitative outcome measures included physical metrics of objective image noise and contrast as well as other quantitative parameters. In total, 34 studies reported on quantitative image noise as the standard deviation of Hounsfield units (n = 27) and/or signal-to-noise ratio (SNR, n = 29), measured in one or multiple standardized regions of interest (ROIs). For NCCT, most frequent ROI locations were cortical and nuclear gray matter, white matter, and cerebrospinal fluid (CSF). For CTA, most frequent ROI locations were the lumen of large intracranial arteries (anterior cerebral artery, ACA; middle cerebral artery, MCA; posterior cerebral artery, PCA; basilar artery, BA), internal carotid arteries (ICA), and common carotid arteries (CCA), as well as adjacent muscles and brain parenchyma. Contrast-to-noise ratio (CNR) was reported in 29 studies and was usually determined between gray and white matter for NCCT, and between arterial lumen and adjacent brain parenchyma or muscles for CTA. Other quantitative parameters were reported in 8 CTA studies, including arterial attenuation measured in HU (n = 4, [29, 64, 66, 68]) and aneurysm diameters (n = 4, [5860, 63]).

Qualitative Measures

Purely quantitative outcome measures are important to enable a comparable IQ assessment [69]. However, more subjective outcome measures are needed to assess the utility of NCCT and CTA at different doses for the clinical application or diagnostic question. The most frequently reported qualitative measures comparable across all included studies were subjective IQ (n = 26), containing common subcategories for some studies (e.g., overall IQ, artifacts, image contrast and sharpness), followed by subjective image noise (n = 16) and diagnostic confidence (n = 15, alternatively termed diagnostic utility, ability, reliability, or acceptability). These measures usually used 3–5 point Likert scales. Furthermore, other application-specific variables were evaluated as qualitative outcome measures and are described in the corresponding sections. In 24 studies, qualitative items were rated by 2 or more readers, and inter-observer agreement (IOA) was reported using the intraclass correlation coefficient (ICC), Cohen’s kappa, or Fleiss’ kappa [70, 71]. In the majority of studies IOA was at least substantial (> 0.6).

For NCCT, two studies assessed diagnostic performance for findings corresponding to acute neurologic deficits [48] or for the detection of ICH [55], reporting classification metrics (accuracy, sensitivity, and specificity) as well as reader agreement rules and jackknife alternative free-response receiver operating characteristic figure of merit (JAFROC-FOM) [48, 72].

For CTA, diagnostic performance for the detection and size measurements of intracranial aneurysms was assessed in 7 studies, using digital subtraction angiography (DSA) as the reference standard. Reported metrics included classification metrics (n = 4; [58, 60, 61, 63]), aneurysm detection numbers or rates (n = 5; [35, 40, 59, 60, 63]), and correlation coefficients of aneurysm diameters (n = 3; [59, 60, 63]), 5 studies [27, 38, 65, 67, 68] investigated the grade of stenosis of the ICA or extracranial and intracranial arteries using the European Carotid Surgery Trial (ECST) [73] or North American Symptomatic Carotid Endarterectomy Trial (NASCET) [74] criteria. Stenosis gradings were used for subgroup analyses and assessment of diagnostic performance.

Dose Reduction in NCCT

Dose reduction in NCCT was considered in 18 articles, including 6 studies focusing on ischemia and stroke and 4 studies focusing on ICH. The remaining 8 NCCT studies had unspecified or mixed indications, e.g., evaluation of trauma, skull fracture, amnesia, loss of consciousness, seizure, headache, vomiting, focal neurological deficit, coagulopathy, treatment with anticoagulants, or tumor staging. For two studies comparing SD and LD protocols of multi-modal CT (NCCT, CTA, and CTP), only the NCCT results were extracted [50, 51]. The results are summarized in Tables 12 and 3.

Cerebral Ischemia and Stroke

Dose reductions ranged from 10 to 36% without impairment of IQ (based on 4 studies, [5052, 54]). Reported doses in terms of CTDIvol and DLP ranged from 31.2 to 59.0 mGy and 441.9–1061.9 mGy * cm, respectively. The highest dose reduction of 36% was reported in 200 patients by Zacharia et al., who compared SD and LD protocols by using ATCM in z‑direction for the LD protocol [54].

In the context of stroke imaging, the detection of ischemic lesions is an important indication of NCCT. In comparison to SD-CT, the conspicuousness of ischemic demarcation could be shown to be preserved by using HIR instead of FBP [52], or it could even be improved by using MBIR instead of HIR for LD-CT (Fig. 2; [49]). Furthermore, excellent IOA (range: 0.80–0.93) regarding overall IQ, anatomic detail, gray-white matter differentiation, and conspicuousness of ischemic demarcation was demonstrated at reduced dose [49, 50, 52]. Furthermore, gray to white matter contrast per dose was found to be markedly increased in the context of acute stroke scans using a reduced dose protocol [53].

Fig. 2.

Fig. 2

Initial SD NCCT (upper row) and follow-up LD NCCT (lower row) of a 41-year-old man with left-sided MCA infarction. Axial images were reconstructed using HIR and MBIR. The demarcated area in the left hemisphere shows better conspicuousness due to higher CNR with MBIR compared to HIR. CNR contrast-to-noise ratio, HIR hybrid iterative reconstruction, LD low-dose, MBIR model-based iterative reconstruction, MCA middle cerebral artery, NCCT non-contrast computed tomography, SD standard-dose

Intracranial Hemorrhage

Four NCCT studies focused on or included patients for ICH evaluation, using heterogeneous study designs and methodologies. Not taking into account simulated LD protocols, dose reductions ranged from 18 to 66% while maintaining sufficient quantitative and qualitative IQ [37, 48, 55, 56].

Reported doses in terms of CTDIvol and DLP ranged from 10.8–41.0 mGy and 141.0–1045.0 mGy * cm, respectively. The highest dose reduction of 66% was reported by Kaul et al., who compared an SD and multiple LD protocols by using reduced tube voltage and ATCM combined with increasing levels of HIR in pediatric patients. A dose reduction of 34% still provided adequate image quality and diagnosis-related confidence. The corresponding LD protocol was therefore recommended for everyday clinical practice [37].

In a study of 94 consecutive ICH patients by Bodelle et al., HIR was recommended over FBP for the evaluation of brain structures and ICH detection when using an LD protocol, enabling a considerable dose reduction of 29% [56].

Scholtz et al. compared SD to LD protocols using a dual-source MDCT in single-energy (SECT) and dual-energy (DECT) mode. With respect to ICH detection, both SECT and DECT achieved excellent sensitivity and specificity at a significant dose reduction of about 25%. The authors recommended the routine use of ATCM and MBIR to reduce dose in ICH evaluation for both SECT and DECT [55].

By inserting artificial noise into MDCT projection data acquired at 250 effective mAs, Fletcher et al. generated LD images corresponding to 25–200 effective mAs using FBP and HIR. Diagnostic performance assessed by JAFROC-FOM was shown to be non-inferior for the 100 effective mAs LD data reconstructed with HIR. As nowadays IR can be considered the standard reconstruction method for most centers, a dose reduction potential of 60% can be derived from this study [48].

NCCT Without Specified Indications

The remaining NCCT studies included patients with mixed or unspecified indications. All 8 studies reported maintained or even improved subjective IQ that was diagnostically acceptable while achieving dose reductions ranging mainly from 20 to 40% [28, 3034, 57]. In all studies, this was achieved with (mainly hybrid) IR. The commonly used objective measures of IQ were SNR and CNR, primarily assessed within or between gray and white matter, and additionally within the CSF space in a subset of studies [28, 30, 31, 33, 57]. Comparing LD and SD protocols, SNR was slightly decreased to unchanged, while CNR showed mostly no relevant differences. Reported doses in terms of CTDIvol and DLP ranged from 13.9–93.5 mGy and 204.0–1270.3 mGy * cm, respectively. The highest dose reduction was 73% using a strong level of HIR (30% with ASIR; GE Healthcare, Milwaukee, WI, USA) in the LD protocol compared to FBP in the SD protocol [36]. Furthermore, reported IOA ranged from moderate (0.54–0.60) regarding noise, sharpness, and diagnostic acceptability [30], to substantial (0.68–0.71) regarding overall IQ [33, 57], artifacts [30], and diagnosis-related confidence [36], to excellent (0.85–0.6) regarding noise, contrast and overall diagnosability [36]. Non-standard IQ evaluation parameters used without dedicated reported IOA included distinctness of posterior fossa contents [57] and visibility of small structures [32].

Dose Reduction in CTA of the Head and Neck

Dose reduction in CTA of the head and neck was considered in 20 articles, 10 studies investigated intracranial aneurysms, including 5 studies that used a lower CM volume in the LD group [39, 40, 58, 61, 62]. Another 10 studies focused on other cerebrovascular diseases, including 2 studies that used a lower CM volume in the LD group [38, 68]. For one study comparing SD and LD protocols of multi-modal CT (NCCT, CTA, and CTP), only the CTA results were extracted [51]. Results are summarized in Tables 4 and 5.

Intracranial Aneurysms

Dose reductions ranged from 16–82%, and none of the studies showed a relevant impairment of subjective IQ for the LD compared to the SD protocol. Specifically, studies using lower CM volume in the LD group consistently demonstrated improved IQ [39, 40, 58, 61, 62]. Reported doses in terms of CTDIvol and DLP ranged from 6.4–59.2 mGy and 116.0–932.6 mGy * cm, respectively. The highest dose reduction of 82% was reported by Chen et al., who used HIR for the LD protocol acquired with reduced tube voltage and CM volume [40].

In addition to the detection of aneurysms of the brain-supplying arteries, aneurysm size measurements provide relevant information derived from CTA. Diameters were shown to be unaffected in LD-CTA compared to SD-CTA [59], and not significantly different from DSA measurements for both LD-CTA and SD-CTA [63]. DSA is the reference standard for the evaluation of aneurysms and was used in four studies to assess diagnostic performance. High diagnostic accuracy (90–99% for LD-CTA vs. 94–100% for SD-CTA), sensitivity (80–95% vs. 91–100%), and specificity (93–100% vs. 93–100%) without statistically significant differences between each CTA protocol and DSA, or between the two CTA protocols were demonstrated [58, 60, 61, 63]. In a different approach to assess diagnostic capabilities, Yang et al. used surgical verification as reference, and 29/29 (100%) of the aneurysms in the LD group were detected compared to 25/26 (96%) in the SD group [59].

Furthermore, IOA regarding overall IQ and other qualitative outcome measures was substantial to excellent (0.60–1.00) in the majority of studies. Other subjective outcome measures included vascular sharpness [35, 39, 40], noise [40], arterial contrast [39], calcifications [35] as well as visibility of blood vessels (in particular small arteries, arteries near the skull base and surgical clips, and peripheral veins) [39, 62].

Other Cerebrovascular Diseases

Not taking into account simulated LD protocols, dose reductions ranged from 25 to 87% [27, 38, 41, 51, 6568], with the majority of studies reporting maximum values of 40–70% [27, 38, 41, 6668] while showing maintained or even improved subjective IQ that was diagnostically acceptable. Reported doses in terms of CTDIvol and DLP were 1.7–24.5 mGy and 84.2–735.3 mGy * cm, respectively. The highest dose reduction of 87% was reported by Annoni et al., who used a newer version of HIR for the LD protocol acquired at reduced tube voltage [65].

Diagnostic performance with respect to ICA stenosis or extracranial and intracranial arterial stenosis was analyzed using different approaches. Two studies used DSA as the reference and reported high diagnostic accuracy (94–99%), sensitivity (91–100%), and specificity (94–99%) for LD-CTA [38, 65]. Annoni et al. additionally demonstrated an excellent correlation regarding the stenosis degree between LD-CTA and DSA (r = 0.98) [65]. Two studies showed excellent IOA (0.81–1.00) for detection and grading of ICA stenosis without a statistically significant difference between SD-CTA and LD-CTA [27, 67].

In a different study design, Sollmann et al. applied virtually lowered tube current in 30 patients who underwent CTA. Even at 25% of the original dose, good vascular contrast and clearly detectable arteries were demonstrated (Fig. 3). As a result, with the adequate type of HIR, all large vessel occlusions or dissections (15/15) could be detected with medium to high diagnostic confidence and excellent IOA (1.00) [47]. Furthermore, IOA regarding overall IQ and diagnostic confidence was substantial to excellent (0.75–1.00) when reported in other studies on that matter [41, 65, 66].

Fig. 3.

Fig. 3

SD (100% tube current, upper row) and simulated LD CTA using virtual tube current reduction (25% of original tube current, bottom row). a Left column: 81-year-old woman with right-sided dissection of the extracranial ICA (red circle). Right column: 84-year-old man with left-sided thrombotic MCA occlusion of the M1 segment (red circle). b Magnified views of the relevant pathologies. The shown axial images were reconstructed with statistical iterative reconstruction, a type of HIR, with low regularization to generate images close to clinical appearance. CTA computed tomography angiography, HIR hybrid iterative reconstruction, ICA internal carotid artery, LD low-dose, MCA middle cerebral artery, SD standard-dose

Discussion

In this article, dose reduction techniques for NCCT and CTA of the head were systematically reviewed and 37 studies representing the most common clinical indications were included. LD-CT and SD-CT were most frequently compared between a different combination of tube settings and image reconstruction techniques.

For NCCT of cerebral ischemia and stroke, achieved dose reductions ranged from 10 to 36%. The majority of studies reduced the dose by at least 22% and maintained objective and subjective IQ, while the use of more advanced IR improved visualization of ischemic demarcation. For NCCT of intracranial hemorrhage, slightly higher dose reductions of 18–66% with sufficient IQ were achieved, while simulated LD-CT suggested a dose reduction potential of 60%. Achieved dose reductions for NCCT with unspecified indication were in a comparable range of 20–40% (Fig. 4).

Fig. 4.

Fig. 4

Development of dose reduction over the last two decades in computed tomography of the head. For studies reporting multiple dose reduction values or ranges, the highest value with acceptable image quality and/or diagnostic performance is displayed

In the context of NCCT, it has to be mentioned that the required dose for differentiation of normal from pathological tissue depends on the contrast difference. For the detection of cerebral ischemia, the contrast differences are very small and therefore a higher dose is required compared to the assessment of ICH, where the contrast difference is significantly higher.

Given the only slightly decreased or even equivalent IQ and diagnostic acceptability of LD protocols and ongoing improvements in image reconstruction, there could be more potential for future dose reduction beyond the reported 20–40% in NCCT of the head. However, increased dose reductions must be used with care. In the abdominal region, the diagnostic performance of low contrast lesions has been reported to be impaired when dose reduction exceeded 25% [75, 76]. Such low contrast lesions are key diagnostic questions in NCCT, particularly in cerebral ischemia and stroke. Diagnostic performance must therefore be thoroughly evaluated before higher dose reductions, facilitated by advanced IR or other novel reconstruction techniques, are implemented.

For CTA of intracranial aneurysms, dose reduction ranged from 16 to 82%, while 6 of 10 studies even achieved maximum values of 65% or higher (Fig. 4). Despite the substantial dose reduction, subjective IQ and diagnostic performance for aneurysm detection were non-inferior. Hence, CTA could be a less invasive and dose-intensive alternative to DSA for accurate evaluation of intracranial aneurysms. Achieved dose reduction in CTA of other cerebrovascular diseases and carotid artery disease was at a comparable level. Similarly, subjective IQ and diagnostic performance with respect to arterial stenosis were not negatively affected. Overall, it can be concluded that tube voltage reduction, advanced IR, and CM volume optimization may represent the most effective combination for dose reduction in head CTA. Beyond radiation exposure, patient care can be improved via lower contrast media-related morbidity, as long as long as CM volume is reduced with care, in order to not impair the visibility of intracranial vessels and associated diagnostic performance.

When comparing dose reductions of NCCT and CTA, one important difference regarding the required contrast has to be acknowledged: in comparison to NCCT of cerebral ischemia or ICH, blood vessels in CTA are high-contrast objects which therefore require a significantly lower dose per se.

Until now, dose reduction in head CT has mainly been achieved through modified tube settings while advanced reconstruction techniques ensured acceptable IQ and high diagnostic performance. Most of these advances have been enabled by software innovations, but current and future developments in CT hardware will also very likely increase dose reduction. The clinical introduction of photon-counting CT (PCCT) represents a revolution in CT imaging, also with respect to radiation exposure [77, 78]. At the same radiation dose as conventional CT, better gray to white matter differentiation due to higher CNR and less image noise has been demonstrated. The ensuing high potential of PCCT for IQ improvement in NCCT of the head could translate into a radiation dose reduction of approximately 40% [79]. At a comparable dose as conventional single-energy CT, PCCT has been reported to reduce beam-hardening artifacts and improve IQ in arterial segments close to surrounding bone [80]. In combination with its ability of spectral material decomposition, it thus bears great potential to further reduce radiation dose for intracranial and carotid CTA, particularly in the presence of calcifications and plaques. Furthermore, the increasing clinical use of PCCT implies the routine acquisition of spectral data. With this evolution, virtual non-contrast (VNC) scans could make NCCT scans redundant in stroke or other head CT protocols requiring non-contrast and contrast-enhanced images, and thereby save a considerable amount of radiation dose. Clinical translation of AI in CT imaging is steadily increasing, which can be expected to unveil additional dose reduction opportunities for head CT, primarily on the image reconstruction side [2529]. Beyond the application of CNNs to accelerate MBIR algorithms, deep-learning techniques can enable the implementation of more complex functions in IR models [16]. Given that existing study results can be validated and reliably translated into the clinical routine, the use of AI will further reduce the required dose in head CT.

The present review article is not without limitations. Methodology and design of the included studies are quite heterogeneous, mostly due to the development of CT hardware and software over time, different CT manufacturers and models, and the fact that different departments have established different head CT protocols. As a result, different dose parameters (CTDIvol, DLP, E) with different ranges of values have been used to determine the amount of dose reduction. Furthermore, absolute definitions of SD and LD protocol cannot be reasonably established in the context of the included studies. Consequently, the start parameters for dose reduction quantification could not be normalized which potentially compromises objectivity and comparability of dose reduction.

In conclusion, considerable dose reduction in NCCT and CTA of the head can be realized by global approaches, such as decreased tube voltage and ATCM, while advances in IR algorithms ensure diagnostic image quality. For CTA examinations, the combination with lower CM volume can be particularly effective for improved patient care. In the upcoming years, the ongoing clinical transition of novel CT acquisition and reconstruction techniques can be expected to enable additional dose reductions of 40% or more, and potentially further CM volume reduction. However, it will be essential that the image quality and diagnostic performance is thoroughly evaluated to guarantee that the benefits of head CT are not compromised.

Acknowledgments

Funding

MD and NS have received funding by the German Society of Musculoskeletal Radiology (Deutsche Gesellschaft für Muskuloskelettale Radiologie [DGMSR]). JSK and TB have received funding by the German Research Foundation (Deutsche Forschungsgemeinschaft [DFG]; project 432290010). JSK has received funding by European Research Council (ERC) under the European Union’s Horizon 2020 research and innovation programme (grant agreement No 963904—Bonescreen—ERC-2020-POC-LS).

Abbreviations

ACA

Anterior cerebral artery

ADMIRE

Advanced modeled iterative reconstruction

AEC

Automatic exposure control

AI

Artificial intelligence

ALARA

As low as reasonably achievable

ASIR

Adaptive statistical iterative reconstruction

ATCM

Automatic tube current modulation

BA

Basilar artery

CCA

Common carotid artery

CM

Contrast medium

CNN

Convolutional neural network

CNR

Contrast-to-noise ratio

CT

Computed tomography

CTA

Computed tomography angiography

CTDIvol

Volumetric CT dose index

CTP

Computed tomography perfusion

DECT

Dual-energy CT

DLP

Dose length product

DSA

Digital subtraction angiography

E

Effective dose

ECST

European Carotid Surgery Trial

FBP

Filtered back projection

HIR

Hybrid iterative reconstruction

ICA

Internal carotid artery

ICC

Intraclass correlation coefficient

ICH

Intracranial hemorrhage

IMR

Iterative model reconstruction

IOA

Interobserver agreement

IQ

Image quality

IR

Iterative reconstruction

JAFROC-FOM

Jackknife alternative free-response receiver operating characteristic figure of merit

LD

Low-dose

LD-CT

Low-dose computed tomography

LD-CTA

Low-dose computed tomography angiography

MBIR

Model-based iterative reconstruction

MCA

Middle cerebral artery

MDCT

Multi-detector computed tomography

NA

Not available

NASCET

North American Symptomatic Carotid Endarterectomy Trial

NCCT

Non-contrast computed tomography

PCA

Posterior cerebral artery

PCCT

Photon counting CT

PRISMA

Preferred Reporting Items for Systematic reviews and Meta-Analyses

ROI

Region of interest

SAFIRE

Sinogram-affirmed iterative reconstruction

SD

Standard-dose

SD-CT

Standard-dose CT

SECT

Single-energy CT

SIR

Statistical iterative reconstruction

SNR

Signal-to-noise ratio

VNC

Virtual non-contrast

Appendix

PubMed Search Terms

Dose Reduction in NCCT of Cerebral Ischemia and Stroke:

((computed tomography) OR (CT)) AND ((low-dose) OR (low dose) OR (dose reduction) OR (low-kilovolt) OR (low kilovolt) OR (low-kV) OR (low kV) OR (iterative reconstruction)) AND ((stroke) OR (cerebral ischemia) OR (cerebral infarction) OR (cerebral infarct) OR (brain ischemia) OR (brain infarction) OR (brain infarct) OR (demarcation)).

Dose Reduction in NCCT of Intracranial Hemorrhage (ICH):

((computed tomography) OR (CT)) AND ((low-dose) OR (low dose) OR (dose reduction) OR (low-kilovolt) OR (low kilovolt) OR (low-kV) OR (low kV) OR (iterative reconstruction)) AND ((cerebral bleeding) OR (cerebral hemorrhage) OR (cerebral haemorrhage) OR (intracerebral hemorrhage) OR (intracerebral haemorrhage) OR (intracranial hemorrhage) OR (intracranial haemorrhage) OR (ICH) OR (brain bleeding) OR (brain hemorrhage) OR (brain haemorrhage) OR (subarachnoid bleeding) OR (subdural hematoma) OR (subdural haematoma) OR (subdural hemorrhage) OR (subdural haemorrhage) OR (SDH) OR (subdural bleeding) OR (epidural hematoma) OR (epidural haematoma) OR (epidural hemorrhage) OR (epidural haemorrhage) OR (extradural hematoma) OR (extradural haematoma) OR (epidural bleeding)).

Dose Reduction in CTA of Intracranial Aneurysms:

((computed tomography angiography) OR (CT angiography) OR (CTA) OR (computed tomography angiogram) OR (CT angiogram)) AND ((low-dose) OR (low dose) OR (dose reduction) OR (low-kilovolt) OR (low kilovolt) OR (low-kV) OR (low kV) OR (iterative reconstruction)) AND ((cerebral aneurysm) OR (brain aneurysm)).

Dose Reduction in CTA of Other Cerebrovascular Diseases and Carotid Artery Disease:

((computed tomography angiography) OR (CT angiography) OR (CTA) OR (computed tomography angiogram) OR (CT angiogram)) AND ((low-dose) OR (low dose) OR (dose reduction) OR (low-kilovolt) OR (low kilovolt) OR (low-kV) OR (low kV) OR (iterative reconstruction)) AND ((stroke) OR (cerebral ischemia) OR (cerebral infarction) OR (cerebral infarct) OR (brain ischemia) OR (brain infarction) OR (brain infarct) OR (vessel occlusion) OR (LVO) OR (thrombus) OR (vessel stenosis) OR (vessel dissection) OR (arterial stenosis) OR (arterial dissection) OR (carotid stenosis) OR (carotid dissection) OR (cerebral vascular malformation) OR (brain vascular malformation) OR (cerebral arteriovenous malformation) OR (brain arteriovenous malformation) OR (cerebral AVM) OR (brain AVM)).

Funding

Open Access funding enabled and organized by Projekt DEAL.

Conflict of interest

M. Dieckmeyer, N. Sollmann, K. Kupfer, M.T. Löffler, K.J. Paprottka, J.S. Kirschke and T. Baum declare that they have no competing interests.

Contributor Information

Michael Dieckmeyer, Email: michael.dieckmeyer@tum.de.

Nico Sollmann, Email: nico.sollmann@tum.de.

Karina Kupfer, Email: karina.kupfer@tum.de.

Maximilian T. Löffler, Email: m.loeffler@tum.de

Karolin J. Paprottka, Email: karolin.paprottka@tum.de

Jan S. Kirschke, Email: jan.kirschke@tum.de

Thomas Baum, Email: thomas.baum@tum.de.

References

  • 1.Smith-Bindman R, Kwan ML, Marlow EC, Theis MK, Bolch W, Cheng SY, Bowles EJA, Duncan JR, Greenlee RT, Kushi LH, Pole JD, Rahm AK, Stout NK, Weinmann S, Miglioretti DL. Trends in use of medical imaging in US health care systems and in Ontario, Canada, 2000–2016. JAMA. 2019;322:843–856. doi: 10.1001/jama.2019.11456. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Smith-Bindman R, Miglioretti DL, Johnson E, Lee C, Feigelson HS, Flynn M, Greenlee RT, Kruger RL, Hornbrook MC, Roblin D, Solberg LI, Vanneman N, Weinmann S, Williams AE. Use of diagnostic imaging studies and associated radiation exposure for patients enrolled in large integrated health care systems, 1996–2010. JAMA. 2012;307:2400–2409. doi: 10.1001/jama.2012.5960. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Brenner DJ, Hall EJ. Computed tomography—an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–2284. doi: 10.1056/NEJMra072149. [DOI] [PubMed] [Google Scholar]
  • 4.Radiology ACo. ACR-ASNR-SPR practice parameter for the performance of computed tomography (CT) of the head. 2020. https://www.acr.org/-/media/ACR/Files/Practice-Parameters/CT-Brain.pdf?la=en. Accessed 12 Aug 2022.
  • 5.Radiology ACo. ACR-ASNR-SPR practice parameter for the performance and interpretation of cervicocerebral computed tomography angiography (CTA). 2020. https://www.acr.org/-/media/ACR/Files/Practice-Parameters/CervicoCerebralCTA.pdf?la=en. Accessed 12 Aug 2022.
  • 6.Radiology ACo. ACR-ASNR-SPR practice parameter for the performance of computed tomography (CT) perfusion in neuroradiologic imaging. 2022. https://www.acr.org/-/media/ACR/Files/Practice-Parameters/CT-Perfusion.pdf?la=en. Accessed 12 Aug 2022.
  • 7.Smith-Bindman R, Lipson J, Marcus R, Kim KP, Mahesh M, Gould R, Berrington de Gonzalez A, Miglioretti DL. Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med. 2009;169:2078–2086. doi: 10.1001/archinternmed.2009.427. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Richards PJ, George J. Diagnostic CT radiation and cancer induction. Skelet Radiol. 2010;39:421–424. doi: 10.1007/s00256-009-0819-2. [DOI] [PubMed] [Google Scholar]
  • 9.Berrington de Gonzalez A, Mahesh M, Kim KP, Bhargavan M, Lewis R, Mettler F, Land C. Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med. 2009;169:2071–2077. doi: 10.1001/archinternmed.2009.440. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Sodickson A, Baeyens PF, Andriole KP, Prevedello LM, Nawfel RD, Hanson R, Khorasani R. Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology. 2009;251:175–184. doi: 10.1148/radiol.2511081296. [DOI] [PubMed] [Google Scholar]
  • 11.Bevelacqua JJ. Practical and effective ALARA. Health Phys. 2010;98(Suppl 2):S39–S47. doi: 10.1097/HP.0b013e3181d18d63. [DOI] [PubMed] [Google Scholar]
  • 12.Prasad KN, Cole WC, Haase GM. Radiation protection in humans: extending the concept of as low as reasonably achievable (ALARA) from dose to biological damage. Br J Radiol. 2004;77:97–99. doi: 10.1259/bjr/88081058. [DOI] [PubMed] [Google Scholar]
  • 13.Demb J, Chu P, Nelson T, Hall D, Seibert A, Lamba R, Boone J, Krishnam M, Cagnon C, Bostani M, Gould R, Miglioretti D, Smith-Bindman R. Optimizing radiation doses for computed tomography across institutions: dose auditing and best practices. JAMA Intern Med. 2017;177:810–817. doi: 10.1001/jamainternmed.2017.0445. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Smith-Bindman R, Moghadassi M, Wilson N, Nelson TR, Boone JM, Cagnon CH, Gould R, Hall DJ, Krishnam M, Lamba R, McNitt-Gray M, Seibert A, Miglioretti DL. Radiation doses in consecutive CT examinations from five University of California medical centers. Radiology. 2015;277:134–141. doi: 10.1148/radiol.2015142728. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Lell MM, Kachelriess M. Recent and upcoming technological developments in computed tomography: High speed, low dose, deep learning, Multienergy. Invest Radiol. 2020;55:8–19. doi: 10.1097/RLI.0000000000000601. [DOI] [PubMed] [Google Scholar]
  • 16.Willemink MJ, Noel PB. The evolution of image reconstruction for CT-from filtered back projection to artificial intelligence. Eur Radiol. 2019;29:2185–2195. doi: 10.1007/s00330-018-5810-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Kalender WA, Wolf H, Suess C. Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements. Med Phys. 1999;26:2248–2253. doi: 10.1118/1.598738. [DOI] [PubMed] [Google Scholar]
  • 18.Mulkens TH, Bellinck P, Baeyaert M, Ghysen D, Van Dijck X, Mussen E, Venstermans C, Termote JL. Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology. 2005;237:213–223. doi: 10.1148/radiol.2363041220. [DOI] [PubMed] [Google Scholar]
  • 19.Lee CH, Goo JM, Ye HJ, Ye SJ, Park CM, Chun EJ, Im JG. Radiation dose modulation techniques in the multidetector CT era: from basics to practice. Radiographics. 2008;28:1451–1459. doi: 10.1148/rg.285075075. [DOI] [PubMed] [Google Scholar]
  • 20.Fleischmann D, Boas FE. Computed tomography—old ideas and new technology. Eur Radiol. 2011;21:510–517. doi: 10.1007/s00330-011-2056-z. [DOI] [PubMed] [Google Scholar]
  • 21.Wolterink JM, Leiner T, Viergever MA, Isgum I. Generative adversarial networks for noise reduction in low-dose CT. IEEE Trans Med Imaging. 2017;36:2536–2545. doi: 10.1109/TMI.2017.2708987. [DOI] [PubMed] [Google Scholar]
  • 22.Chen H, Zhang Y, Kalra MK, Lin F, Chen Y, Liao P, Zhou J, Wang G. Low-dose CT with a residual encoder-decoder convolutional neural network. IEEE Trans Med Imaging. 2017;36:2524–2535. doi: 10.1109/TMI.2017.2715284. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Chen H, Zhang Y, Zhang W, Liao P, Li K, Zhou J, Wang G. Low-dose CT via convolutional neural network. Biomed Opt Express. 2017;8:679–694. doi: 10.1364/BOE.8.000679. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Missert AD, Yu L, Leng S, Fletcher JG, McCollough CH. Synthesizing images from multiple kernels using a deep convolutional neural network. Med Phys. 2020;47:422–430. doi: 10.1002/mp.13918. [DOI] [PubMed] [Google Scholar]
  • 25.Moher D, Liberati A, Tetzlaff J, Altman DG, Group P. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA Statement. Open Med. 2009;3:e123–e130. [PMC free article] [PubMed] [Google Scholar]
  • 26.Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gotzsche PC, Ioannidis JP, Clarke M, Devereaux PJ, Kleijnen J, Moher D. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. PLoS Med. 2009;6:e1000100. doi: 10.1371/journal.pmed.1000100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Moloney F, Murphy KP, Twomey M, Crush L, Canniffe EM, McLaughlin PD, Moore N, O’Keeffe M, O’Neill S, Manning BM, Wyse G, Fanning N, O’Connor OJ, Maher MM. Low-dose carotid computed tomography angiography using pure iterative reconstruction. J Comput Assist Tomogr. 2016;40:833–839. doi: 10.1097/RCT.0000000000000436. [DOI] [PubMed] [Google Scholar]
  • 28.Kilic K, Erbas G, Guryildirim M, Konus OL, Arac M, Ilgit E, Isik S. Quantitative and qualitative comparison of standard-dose and low-dose pediatric head computed tomography: a retrospective study assessing the effect of adaptive statistical iterative reconstruction. J Comput Assist Tomogr. 2013;37:377–381. doi: 10.1097/RCT.0b013e31828426de. [DOI] [PubMed] [Google Scholar]
  • 29.Bahner ML, Bengel A, Brix G, Zuna I, Kauczor HU, Delorme S. Improved vascular opacification in cerebral computed tomography angiography with 80 kVp. Invest Radiol. 2005;40:229–234. doi: 10.1097/01.rli.0000155281.32319.52. [DOI] [PubMed] [Google Scholar]
  • 30.Kilic K, Erbas G, Guryildirim M, Arac M, Ilgit E, Coskun B. Lowering the dose in head CT using adaptive statistical iterative reconstruction. AJNR Am J Neuroradiol. 2011;32:1578–1582. doi: 10.3174/ajnr.A2585. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Becker HC, Augart D, Karpitschka M, Ulzheimer S, Bamberg F, Morhard D, Neumaier K, Graser A, Johnson T, Reiser M. Radiation exposure and image quality of normal computed tomography brain images acquired with automated and organ-based tube current modulation multiband filtering and iterative reconstruction. Invest Radiol. 2012;47:202–207. doi: 10.1097/RLI.0b013e31823a86d5. [DOI] [PubMed] [Google Scholar]
  • 32.Rapalino O, Kamalian S, Kamalian S, Payabvash S, Souza LC, Zhang D, Mukta J, Sahani DV, Lev MH, Pomerantz SR. Cranial CT with adaptive statistical iterative reconstruction: improved image quality with concomitant radiation dose reduction. AJNR Am J Neuroradiol. 2012;33:609–615. doi: 10.3174/ajnr.A2826. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Korn A, Bender B, Fenchel M, Spira D, Schabel C, Thomas C, Flohr T, Claussen CD, Bhadelia R, Ernemann U, Brodoefel H. Sinogram affirmed iterative reconstruction in head CT: improvement of objective and subjective image quality with concomitant radiation dose reduction. Eur J Radiol. 2013;82:1431–1435. doi: 10.1016/j.ejrad.2013.03.011. [DOI] [PubMed] [Google Scholar]
  • 34.Ozdoba C, Slotboom J, Schroth G, Ulzheimer S, Kottke R, Watzal H, Weisstanner C. Dose reduction in standard head CT: first results from a new scanner using iterative reconstruction and a new detector type in comparison with two previous generations of multi-slice CT. Clin Neuroradiol. 2014;24:23–28. doi: 10.1007/s00062-013-0263-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Chen GZ, Zhang LJ, Schoepf UJ, Wichmann JL, Milliken CM, Zhou CS, Qi L, Luo S, Lu GM. Radiation dose and image quality of 70 kVp cerebral CT angiography with optimized sinogram-affirmed iterative reconstruction: comparison with 120 kVp cerebral CT angiography. Eur Radiol. 2015;25:1453–1463. doi: 10.1007/s00330-014-3533-y. [DOI] [PubMed] [Google Scholar]
  • 36.Kaul D, Kahn J, Huizing L, Wiener E, Grupp U, Boning G, Ghadjar P, Renz DM, Streitparth F. Reducing radiation dose in adult head CT using iterative reconstruction—A clinical study in 177 patients. Rofo. 2016;188:155–162. doi: 10.1055/s-0041-107200. [DOI] [PubMed] [Google Scholar]
  • 37.Kaul D, Kahn J, Huizing L, Wiener E, Boning G, Renz DM, Streitparth F. Dose reduction in paediatric cranial CT via iterative reconstruction: a clinical study in 78 patients. Clin Radiol. 2016;71:1168–1177. doi: 10.1016/j.crad.2016.06.115. [DOI] [PubMed] [Google Scholar]
  • 38.Chen Y, Zhang X, Xue H, Zhu Y, Wang Y, Li Y, Zhang Z, Jin Z. Head and neck angiography at 70 kVp with a third-generation dual-source CT system in patients: comparison with 100 kVp. Neuroradiology. 2017;59:1071–1081. doi: 10.1007/s00234-017-1901-4. [DOI] [PubMed] [Google Scholar]
  • 39.Nagayama Y, Nakaura T, Tsuji A, Urata J, Furusawa M, Yuki H, Hirarta K, Oda S, Kidoh M, Utsunomiya D, Yamashita Y. Cerebral bone subtraction CT angiography using 80 kVp and sinogram-affirmed iterative reconstruction: contrast medium and radiation dose reduction with improvement of image quality. Neuroradiology. 2017;59:127–134. doi: 10.1007/s00234-016-1776-9. [DOI] [PubMed] [Google Scholar]
  • 40.Chen GZ, Fang XK, Zhou CS, Zhang LJ, Lu GM. Cerebral CT angiography with iterative reconstruction at 70kVp and 30 mL iodinated contrast agent: Initial experience. Eur J Radiol. 2017;88:102–108. doi: 10.1016/j.ejrad.2016.12.037. [DOI] [PubMed] [Google Scholar]
  • 41.Chen CW, Chen PA, Chou CC, Fu JH, Wang PC, Hsu SH, Lai PH. Combination of adaptive statistical iterative reconstruction-V and lower tube voltage during craniocervical computed tomographic angiography yields better image quality with a reduced radiation dose. Acad Radiol. 2019;26:e233–e240. doi: 10.1016/j.acra.2018.07.019. [DOI] [PubMed] [Google Scholar]
  • 42.Bongartz G, Golding SJ, Jurik AG, Leonardi M, van Persijn van Meerten E, Rodríguez R, Schneider K, Calzado A, Geleijns J, Jessen KA, Panzer W, Shrimpton PC, Tosi G. European guidelines for multislice computed tomography. 2004. [Google Scholar]
  • 43.Committee DICC . AAPM Report No. 096—The measurement, reporting, and management of radiation dose in CT. College Park: Committee DICC; 2008. pp. 20740–23846. [Google Scholar]
  • 44.Huda W, Ogden KM, Khorasani MR. Converting dose-length product to effective dose at CT. Radiology. 2008;248:995–1003. doi: 10.1148/radiol.2483071964. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology. 2010;257:158–166. doi: 10.1148/radiol.10100047. [DOI] [PubMed] [Google Scholar]
  • 46.Shrimpton PC, Jansen JT, Harrison JD. Updated estimates of typical effective doses for common CT examinations in the UK following the 2011 national review. Br J Radiol. 2016;89:20150346. doi: 10.1259/bjr.20150346. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Sollmann N, Mei K, Riederer I, Schon S, Kopp FK, Loffler MT, Hedderich DM, Maegerlein C, Friedrich B, Kreiser K, Boeckh-Behrens T, Kirschke JS, Rummeny EJ, Zimmer C, Noel PB, Baum T. Tube current reduction in CT angiography: how low Can we go in imaging of patients with suspected acute stroke? AJR Am J Roentgenol. 2019;213:410–416. doi: 10.2214/AJR.18.20954. [DOI] [PubMed] [Google Scholar]
  • 48.Fletcher JG, DeLone DR, Kotsenas AL, Campeau NG, Lehman VT, Yu L, Leng S, Holmes DR, 3rd, Edwards PK, Johnson MP, Michalak GJ, Carter RE, McCollough CH. Evaluation of lower-dose spiral head CT for detection of Intracranial findings causing neurologic deficits. AJNR Am J Neuroradiol. 2019;40:1855–1863. doi: 10.3174/ajnr.A6251. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Paprottka KJ, Kupfer K, Riederer I, Zimmer C, Beer M, Noel PB, Baum T, Kirschke JS, Sollmann N. Impact of dose reduction and iterative model reconstruction on multi-detector CT imaging of the brain in patients with suspected ischemic stroke. Sci Rep. 2021;11:22271. doi: 10.1038/s41598-021-01162-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Kapur SC, Kapur J, Sharma VK. Radiation exposure during computerized tomography-based neuroimaging for acute ischemic stroke: a case-control study. J Integr Neurosci. 2021;20:605–611. doi: 10.31083/j.jin2003064. [DOI] [PubMed] [Google Scholar]
  • 51.Bricout N, Estrade L, Boustia F, Kalsoum E, Pruvo JP, Leclerc X. Reduced-dose CT protocol for the assessment of cerebral vasospasm. Neuroradiology. 2015;57:1211–1218. doi: 10.1007/s00234-015-1585-6. [DOI] [PubMed] [Google Scholar]
  • 52.Bodelle B, Wichmann JL, Scholtz JE, Lehnert T, Vogl TJ, Luboldt W, Schulz B. Iterative reconstruction leads to increased subjective and objective image quality in cranial CT in patients with stroke. AJR Am J Roentgenol. 2015;205:618–622. doi: 10.2214/AJR.15.14389. [DOI] [PubMed] [Google Scholar]
  • 53.Ben-David E, Cohen JE, Nahum Goldberg S, Sosna J, Levinson R, Leichter IS, Gomori JM. Significance of enhanced cerebral gray-white matter contrast at 80 kVp compared to conventional 120 kVp CT scan in the evaluation of acute stroke. J Clin Neurosci. 2014;21:1591–1594. doi: 10.1016/j.jocn.2014.03.008. [DOI] [PubMed] [Google Scholar]
  • 54.Zacharia TT, Kanekar SG, Nguyen DT, Moser K. Optimization of patient dose and image quality with z-axis dose modulation for computed tomography (CT) head in acute head trauma and stroke. Emerg Radiol. 2011;18:103–107. doi: 10.1007/s10140-010-0908-5. [DOI] [PubMed] [Google Scholar]
  • 55.Scholtz JE, Wichmann JL, Bennett DW, Leithner D, Bauer RW, Vogl TJ, Bodelle B. Detecting intracranial hemorrhage using automatic tube current modulation with advanced modeled iterative reconstruction in Unenhanced head single- and dual-energy dual-source CT. AJR Am J Roentgenol. 2017;208:1089–1096. doi: 10.2214/AJR.16.17171. [DOI] [PubMed] [Google Scholar]
  • 56.Bodelle B, Klein E, Naguib NN, Bauer RW, Kerl JM, Al-Butmeh F, Wichmann JL, Ackermann H, Lehnert T, Vogl TJ, Schulz B. Acute intracranial hemorrhage in CT: benefits of sinogram-affirmed iterative reconstruction techniques. AJNR Am J Neuroradiol. 2014;35:445–449. doi: 10.3174/ajnr.A3801. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Korn A, Fenchel M, Bender B, Danz S, Hauser TK, Ketelsen D, Flohr T, Claussen CD, Heuschmid M, Ernemann U, Brodoefel H. Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol. 2012;33:218–224. doi: 10.3174/ajnr.A2749. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Ni QQ, Chen GZ, Schoepf UJ, Klitsie MAJ, De Cecco CN, Zhou CS, Luo S, Lu GM, Zhang LJ. Cerebral CTA with low tube voltage and low contrast material volume for detection of intracranial aneurysms. AJNR Am J Neuroradiol. 2016;37:1774–1780. doi: 10.3174/ajnr.A4803. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Yang B, Gao Y, Yang YY, Zhao W. Influence of selective photon shield combined with sinogram-affirmed iterative reconstruction on image quality and radiation dose during dual-energy CT angiography-mediated diagnosis of intracranial aneurysms. Int J Neurosci. 2016;126:1084–1091. doi: 10.3109/00207454.2015.1121389. [DOI] [PubMed] [Google Scholar]
  • 60.Tang K, Li R, Lin J, Zheng X, Wang L, Yin W. The value of cerebral CT angiography with low tube voltage in detection of intracranial aneurysms. Biomed Res Int. 2015;2015:876796. doi: 10.1155/2015/876796. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Luo S, Zhang LJ, Meinel FG, Zhou CS, Qi L, McQuiston AD, Schoepf UJ, Lu GM. Low tube voltage and low contrast material volume cerebral CT angiography. Eur Radiol. 2014;24:1677–1685. doi: 10.1007/s00330-014-3184-z. [DOI] [PubMed] [Google Scholar]
  • 62.Kidoh M, Nakaura T, Ogata T, Takashima H, Yoshikawa M, Uemura S, Harada K, Yamashita Y. Subtracted 3D CT angiography for the evaluation of intracranial aneurysms in 256-slice multidetector CT: usefulness of the 80-kVp plus compact contrast medium bolus protocol. Eur Radiol. 2013;23:3012–3019. doi: 10.1007/s00330-013-2928-5. [DOI] [PubMed] [Google Scholar]
  • 63.Sun G, Ding J, Lu Y, Li M, Li L, Li GY, Zhang XP. Comparison of standard- and low-tube voltage 320-detector row volume CT angiography in detection of intracranial aneurysms with digital subtraction angiography as gold standard. Acad Radiol. 2012;19:281–288. doi: 10.1016/j.acra.2011.11.004. [DOI] [PubMed] [Google Scholar]
  • 64.Waaijer A, Prokop M, Velthuis BK, Bakker CJ, de Kort GA, van Leeuwen MS. Circle of Willis at CT angiography: dose reduction and image quality—reducing tube voltage and increasing tube current settings. Radiology. 2007;242:832–839. doi: 10.1148/radiol.2423051191. [DOI] [PubMed] [Google Scholar]
  • 65.Annoni AD, Montorsi P, Andreini D, Pontone G, Mancini ME, Muscogiuri G, Formenti A, Mushtaq S, Trabattoni P, Dainese L, Pepi M. Submillisievert CT angiography for carotid arteries using wide array CT scanner and latest iterative reconstruction algorithm in comparison with previous generations technologies: Feasibility and diagnostic accuracy. J Cardiovasc Comput Tomogr. 2019;13:41–47. doi: 10.1016/j.jcct.2019.01.009. [DOI] [PubMed] [Google Scholar]
  • 66.Wang X, Zhu C, Li J, Degnan AJ, Jiang T, Lu J. Knowledge-based iterative model reconstruction: Comparative image quality with low tube voltage cerebral CT angiography. Medicine (Baltimore) 2018;97:e11514. doi: 10.1097/MD.0000000000011514. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Leithner D, Wichmann JL, Mahmoudi S, Martin SS, Albrecht MH, Vogl TJ, Scholtz JE. Diagnostic yield of 90-kVp low-tube-voltage carotid and intracerebral CT-angiography: effects on radiation dose, image quality and diagnostic performance for the detection of carotid stenosis. Br J Radiol. 2018;91:20170927. doi: 10.1259/bjr.20170927. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Kayan M, Demirtas H, Turker Y, Kayan F, Cetinkaya G, Kara M, Orhan Celik A, Umul A, Yilmaz O, Recep Aktas A. Carotid and cerebral CT angiography using low volume of iodinated contrast material and low tube voltage. Diagn Interv Imaging. 2016;97:1173–1179. doi: 10.1016/j.diii.2016.06.005. [DOI] [PubMed] [Google Scholar]
  • 69.Månsson LG. Methods for the evaluation of image quality: a review. Radiat Prot Dosimetry. 2000;90:89–99. doi: 10.1093/oxfordjournals.rpd.a033149. [DOI] [Google Scholar]
  • 70.Cohen J. A coefficient of agreement for nominal scales. Educ Psychol Meas. 1960;20:37–46. doi: 10.1177/001316446002000104. [DOI] [Google Scholar]
  • 71.Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–174. doi: 10.2307/2529310. [DOI] [PubMed] [Google Scholar]
  • 72.Chakraborty DP. Recent advances in observer performance methodology: jackknife free-response ROC (JAFROC) Radiat Prot Dosimetry. 2005;114:26–31. doi: 10.1093/rpd/nch512. [DOI] [PubMed] [Google Scholar]
  • 73.European Carotid Surgery Trialists’ Collaborative Group Randomised trial of endarterectomy for recently symptomatic carotid stenosis: final results of the MRC European Carotid Surgery Trial (ECST) Lancet. 1998;351:1379–1387. doi: 10.1016/S0140-6736(97)09292-1. [DOI] [PubMed] [Google Scholar]
  • 74.North American Symptomatic Carotid Endarterectomy Trial Collaborators. Barnett HJM, Taylor DW, Haynes RB, Sackett DL, Peerless SJ, Ferguson GG, Fox AJ, Rankin RN, Hachinski VC, Wiebers DO, Eliasziw M. Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. N Engl J Med. 1991;325:445–453. doi: 10.1056/NEJM199108153250701. [DOI] [PubMed] [Google Scholar]
  • 75.McCollough CH, Yu L, Kofler JM, Leng S, Zhang Y, Li Z, Carter RE. Degradation of CT low-contrast spatial resolution due to the use of iterative reconstruction and reduced dose levels. Radiology. 2015;276:499–506. doi: 10.1148/radiol.15142047. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Mileto A, Guimaraes LS, McCollough CH, Fletcher JG, Yu L. State of the art in abdominal CT: the limits of iterative reconstruction algorithms. Radiology. 2019;293:491–503. doi: 10.1148/radiol.2019191422. [DOI] [PubMed] [Google Scholar]
  • 77.Willemink MJ, Persson M, Pourmorteza A, Pelc NJ, Fleischmann D. Photon-counting CT: technical principles and clinical prospects. Radiology. 2018;289:293–312. doi: 10.1148/radiol.2018172656. [DOI] [PubMed] [Google Scholar]
  • 78.Flohr T, Petersilka M, Henning A, Ulzheimer S, Ferda J, Schmidt B. Photon-counting CT review. Phys Med. 2020;79:126–136. doi: 10.1016/j.ejmp.2020.10.030. [DOI] [PubMed] [Google Scholar]
  • 79.Pourmorteza A, Symons R, Reich DS, Bagheri M, Cork TE, Kappler S, Ulzheimer S, Bluemke DA. Photon-counting CT of the brain: in vivo human results and image-quality assessment. AJNR Am J Neuroradiol. 2017;38:2257–2263. doi: 10.3174/ajnr.A5402. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 80.Symons R, Reich DS, Bagheri M, Cork TE, Krauss B, Ulzheimer S, Kappler S, Bluemke DA, Pourmorteza A. Photon-counting computed tomography for vascular imaging of the head and neck: first in vivo human result. Invest Radiol. 2018;53:135–142. doi: 10.1097/RLI.0000000000000418. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Clinical Neuroradiology are provided here courtesy of Springer

RESOURCES