. 2023 Aug 23;35(8):082120. doi: 10.1063/5.0157801

Enhanced capillary pumping using open-channel capillary trees with integrated paper pads

Jodie C Tokihiro 1, Wan-chen Tu 1, Jean Berthier 1, Jing J Lee 1, Ashley M Dostie 1, Jian Wei Khor 1, Madeleine Eakman 1, Ashleigh B Theberge 1,2,1,2,a), Erwin Berthier 1,a)
PMCID: PMC10479884  PMID: 37675268

Abstract

The search for efficient capillary pumping has led to two main directions for investigation: first, assembly of capillary channels to provide high capillary pressures, and second, imbibition in absorbing fibers or paper pads. In the case of open microfluidics (i.e., channels where the top boundary of the fluid is in contact with air instead of a solid wall), the coupling between capillary channels and paper pads unites the two approaches and provides enhanced capillary pumping. In this work, we investigate the coupling of capillary trees—networks of channels mimicking the branches of a tree—with paper pads placed at the extremities of the channels, mimicking the small capillary networks of leaves. It is shown that high velocities and flow rates (7 mm/s or 13.1 μl/s) for more than 30 s using 50% (v/v) isopropyl alcohol, which has a 3-fold increase in viscosity in comparison to water; 6.5 mm/s or 12.1 μl/s for more than 55 s with pentanol, which has a 3.75-fold increase in viscosity in comparison to water; and >3.5 mm/s or 6.5 μl/s for more than 150 s with nonanol, which has a 11-fold increase in viscosity in comparison to water, can be reached in the root channel, enabling higher sustained flow rates than that of capillary trees alone.

NOMENCLATURE

Latin, lower case

p

Perimeter (mm)

p0

Perimeter of root channel (mm)

pn

Perimeter of branch n (mm)

t

Time (s)

tn

Time at extremity of branch n (s)

t 0

Time at exit of the root channel (s)

z

Travel distance (mm)

z 0

Travel distance (root channel) (mm)

zn

Travel distance in branch n (mm)

zp

Travel distance in the paper pad (mm)

Latin, upper case

An

Characteristic treelength (n branches) (mm)

C

Coefficient of the LWR law (mm2/s)

K

Permeability (paper pad) (mm2)

L

Length of a branch (mm)

L 0

Length of root channel (mm)

P

Pressure (Pa)

Pj

Pressure at node j (Pa)

P cap

Capillary pressure (Pa)

S

Cross-sectional area (mm2)

S 0

Cross-sectional area (root channel) (mm2)

Sp 0

Cross-sectional area (paper pad entrance) (mm2)

V

Velocity (mm/s)

Vn

Velocity (branch n) (mm/s)

V 0

Velocity (root channel) (mm/s)

Greek, lower case

α

Homothetic factor (non-dimensional)

β

Geometrical angle of the circular pad (rd)

γ

Liquid–air surface tension (mN/m)

λ

Friction length (mm)

μ

Liquid viscosity (mPa s)

τ

Time (for paper pads) (s)

θ

Young angle (rd)

θ*

Generalized Cassie angle (rd)

ϕ

porosity (paper pad) (non-dimensional)

Greek, upper case

Gradient operator (1/mm)

Σn

Characteristic number for the tree (non-dimensional)

I. INTRODUCTION

Simple and autonomous microfluidic systems can be designed by use of capillary forces. In such systems, bulky active pumps1,2 are not needed. However, a major drawback in a capillarity-based system is the decrease in the flow velocity—and the flow rate—with time. According to the Lucas–Washburn–Rideal (LWR) law, the decrease in the velocity is proportional to the inverse of the square root of time.3–5 To overcome this drawback, different capillary pump designs have been developed.

Many passive designs have been developed for capillary-driven flows in closed channels;6–9 however, few have been proposed in the case of open channels. The electrowetting-based pumping device proposed by Satoh et al. is one of the first pumping designs in open geometries but requires the addition of electric actuation.10 The most current open-pumping systems rely on evaporation. Evaporation from a rectangular open channel has been documented by Kolliopoulos et al.11 and Lynn and Dandy,12 and pumping has been set from a reservoir or fibrous pads by Zimmermann et al.13 However, these methodologies are restricted to low boiling point liquids and entail long experiment durations. The search for efficient open-channel pumping based solely on geometrical features is currently progressing. Srinivasan has proposed a geometrical diffuser for zero gravity pumping in space vanes,14 and Guo et al. have developed an interesting system combining capillarity in closed channels and additional pumping from paper pads has been documented.15

In this work, the pumping mechanism involves the use of networks of small open channels where the capillary pressure is high, or there are matrices of fibers (often paper pads) with a high wicking power. These capillary pumps are placed behind the “region of interest” where the biological or chemical processes are performed and can be used in multiple applications including separation methods or sample processing.

Capillary trees for capillary pumping in closed (confined) channels have been developed in the past.16 Examples include triple tree line capillary pumps used for performing immunoassays,17 microstructures for simple and advanced capillary pumping,18 and multilayers of microfluidic paper to generate the capillary flow.19,20 On the other hand, it was shown that microporous and fibrous structures, such as paper pads or porous membranes, provide efficient pumping properties due to their high wicking power.21–23

In the capillary-driven microfluidics field, open systems are of special interest.24–26 These systems remove at least one “wall” of the microfluidic channel (often the top wall), providing easy access to the flowing liquid in the channels. We have previously found that a capillary tree can be used to maintain a high value of velocity in the root channel—the channel of interest for a given application—in open microfluidic devices.27

In this work, we show that these open capillary tree channels can easily be connected to paper pads (mimicking the small capillary networks of leaves) to further extend capillary pumping in the root channel, which is an innovation from our previous work.28 Extending periods of high flow velocity is an important problem in open capillary microfluidics. Here, we have chosen to use an open channel instead of a closed capillary tree due to fabrication limitations such as difficulty in adjusting a cover plate over the trees with minimal to no leakage of the flowing solvent. Along with our previous work, this work expands the toolbox for open-channel microfluidics, which is implemented in a broad range of fields of studies, such as diagnostics, biomedical science, pharmaceutical science, space science, and integrated analysis systems. The removal of the top wall of the channel enables direct access to the flowing fluid for the addition, removal, or manipulation of the channel contents. Thus, together with this work and prior work,28 we demonstrate a method for sustaining periods of high fluid velocity through the coupling of paper pads (current work) and bifurcating capillary trees (prior work).

In itself, an “infinite” capillary tree with decreasing channels has been shown to be favorable for pumping, but practically, it is impossible to design and fabricate such a device. In addition, paper alone is limited by the high friction of the liquid in the microporous media. Hence, the additive structure constituted by the paper pads replaces the downscaling of the trees and enhances the global pumping effect, first coming from the capillary tree and then from the fibrous paper “leaves.” The paper pads are placed in milled receptacles at the end of the branched channels. We used a geometric design of capillary trees where the root channel is successively divided in a cascade of daughter channels, and the cross sections of the daughter channels are progressively decreased at a ratio of 0.85. With this design, a high flow rate in the root channel can be achieved and sustained with such designs using bifurcations and paper pads (Fig. 1). The multiplication of the tree branches avoids the velocity decrease in the root channel predicted by the LWR law in the case of a constant cross-section channel. The flow rate increases at the bifurcation due to the increase in the capillary pressure. Then, decreases due to wall friction, but the flow rate in the root channel is twice that of the first branches, four times that in the second branches, and so on. On the other hand, the multiplication of the pads in parallel palliates the friction effect in each pad on the flow rate in the root channel.

FIG. 1.

FIG. 1.

Diagram of an open-channel microfluidic device utilizing a capillary tree channel and paper pads.

A closed form model for the flow dynamics is derived here, coupling the formulation of the capillary tree flow with the paper pads. In this manuscript, it will be shown that high capillary velocities are obtained even in the case of highly viscous fluids, pentanol, and nonanol. While there are publications where paper pads are used to drive flow,15,23 to the best of our knowledge, this is the first study to couple paper pads to bifurcating trees, specifically in open-channel systems. We show here that by combining homothetic capillary tree channels with paper pads in an open microfluidic device, such designs maintain a high liquid velocity (7 mm/s or 13.1 μl/s) for more than 30 s using 50% (v/v) isopropyl alcohol, which has a 3-fold increase in viscosity in comparison to water; 6.5 mm/s or 12.1 μl/s for more than 55 s with pentanol, which has a 3.75-fold increase in viscosity in comparison to water; and >3.5 mm/s or 6.5 μl/s for more than 150 s with fluid, which has a viscosity that is 11 times higher than water in the root channel for more than 1 min.

II. MATERIALS AND METHODS

A. Fabrication of capillary tree channels

The device consists of winding serpentine-shaped channels (the root channel), a large inlet in which the liquid is introduced using a pipette, three levels of branches, and semicircle paper pads at the extremity of the last set of channel branches (Fig. 1). The dimensions of the channels are listed in Table SII; an engineering drawing is included in Fig. SI; and the computer aided design files are included in the supplementary material. The widths and depths of the channels are homothetically reduced by a factor of 0.85 after each bifurcation. The turns in the winding channels do not affect the capillary flow in the absence of capillary filaments,29 as the rounded bottom avoids the formation of filaments observed in channels of rectangular cross section.30 The average wall friction length of the root channel is estimated to be λ 259 μm from our preceding work.27,28 It was shown that the average friction length produces the value of the average wall friction, τ, by the formula, τ=μV/λ,21 where μ is the liquid viscosity.

The device was designed using a computer aided design (CAD) software (Solidworks 2017, Waltham, MA), and the design files were converted to G-code using a computer aided manufacturing (CAM) software (Fusion 360). Channels were milled in poly(methyl methacrylate) (PMMA) sheets (3.175 mm thick, #8560K239; McMaster-Carr, Santa Fe Springs, CA). To create round bottom channels, endmills with a cutter diameter of 1/32 in. (TR-2-0312-BN) and 1/64 in. (TR-2-0150-BN) were used (Performance Micro Tool, Janesville, WI). The devices were fabricated via micromilling on a Datron Neo computer numerical control (CNC) mill (Datron, Germany). The channel bottom is estimated to have a few micrometers of roughness, which is one magnitude below the roughness values that were observed by Lade et al.31 that would produce fluctuations in flow velocity.

B. Paper pads

Whatman #1 paper (Whatman Grade 1 Qualitative Filter Paper, #28450-160, VWR Scientific, San Francisco, CA) was cut into half circle shapes using a plotter cutter (Graphtec Corporation, Yokohama, Japan). A tight contact between the paper pads and the outlets of the last set of tree channels is essential to maintaining the capillary flow of the fluid. The main characteristics of the paper pads are listed in the supplementary material, Table SIII. The capillary pressure depends on the liquid that is used and the values are in alignment with the experimental results.

C. Solvents

The physical properties of the solvents are indicated in the supplementary material, Table SI. To mitigate evaporation of the solvents, pentanol and nonanol, which are low volatile solvents (boiling points are 139 and 213 °C, respectively), were used. Both pentanol and nonanol have been colored with either Solvent Yellow 7 or with Solvent Green 3 (Sigma-Aldrich) at concentrations of 0.50 and 1.43 mg/ml, respectively. Aqueous isopropyl alcohol (IPA) (VWR Scientific) was used at a concentration 50% (v/v) and colored with 0.60% yellow or 1.2% blue food coloring (McCormick).

D. Capillary trees with integrated paper pads flow experiments

To obtain fluid travel distance and velocity data, 2 ml of the yellow-dyed fluid was pipetted into the reservoir of the device. Once the fluid front reached the end of level 1, a 200 μl refill of the yellow fluid was added. After the yellow fluid wetted the paper pad, 500 μl of the blue-dyed fluid was pipetted into the fluid reservoir with a 200 μl refill after the blue fluid front reached the end of level 1. Data were reported up to the point with when the paper pad was saturated at 90 s for 50% (v/v) isopropyl alcohol, 57 s for pentanol, and 154 s for nonanol for trials 1–3, respectively (Fig. 3 and Figs. SII and SIII).

FIG. 3.

FIG. 3.

Travel distance and velocity over time for 50% (v/v) IPA solution, pentanol, and nonanol solutions in an open channel capillary tree with paper pads. (a) Experimentally determined travel distances in the bifurcating capillary tree vs time for the first phase of the flow (orange dots) and the second phase of the flow (blue dots) after the flow has reached the pads, and blue colored solution has been added to the inlet. Travel distance was determined by measuring the distance of the fluid front in the device in the recorded videos. The black lines are the theoretical results. (b) Velocities in the root channel vs time when the tip of the flow is in the capillary tree (orange dots) and in the paper pads (blue dots). Velocities were determined using Eq. (5).

E. Imaging and analysis

Videos of the progression of the solvent flow in the device were recorded using a Nikon-D5300 ultra-high resolution single lens reflective (SLR) camera. The images of the videos were extracted automatically by the code. The location of the tip of the 50% IPA flow was pinpointed using MATLAB software, while the travel distance of the pentanol and nonanol flows was measured manually using the Fiji (ImageJ) software. For the manual analysis, the scale was set for an individual trial with the “Set Scale” tool. The fluid front was pinpointed with the “Segmented Line” tool, and the “Measure” function was used to calculate the total distance traveled along the capillary tree. Each data were from every 10 frames (yellow) or 30 frames (blue).

III. RESULTS AND DISCUSSION

A. Theory

In the first phase, the flow advances in the capillary tree, dividing itself at each bifurcation. The device is designed so that the tree is symmetrical, and, thus, all fluidic channels have the same length. This motion has already been analyzed in our prior work.27 We describe here a theoretical approach for open capillary channel bifurcating trees coupled to paper pads. The notations and associated definitions are given in the Nomenclature section.

Let us recall that the marching distance in the open root channel (the channel before the beginning of the bifurcations) is given by

z0=2λγcosθ0*μt, (1)

where λ is the average wall friction length,29 γ is the surface tension, μ is the viscosity, t is the time, θ* is the generalized Cassie angle,32 and index, 0, refers to the root channel, so that θ0* is the generalized Cassie angle in the root channel. Using the pressure at each node (bifurcation) plus the homothetic relation for the channel perimeters, and cross sections and the mass conservation equation, one finds the expression of zn for the marching distance in the channels after the nth bifurcations:

zn=An1+1+αnCAn2(ttn1), (2)

where C=2λγcosθ*μ, α is the homothetic ratio, tn−1 is the time at which the liquid enters the nth channel, and An is a geometrical factor which depends on the channel lengths L0 to Ln1 and α. The algebra leading to this expression is lengthy and fully developed in the supplementary material, Sec. 1. Note that (2) differs from the LWR expression where zt1/2.

Figure 2 (Multimedia view) shows the flow of fluid in a device that combines a capillary tree and paper pads. When the flow reaches the paper pads, the wicking of the pads is governed by Darcy's law,33,34

Vp=KμϕP=KμϕPcapPjzp, (3)

where Pcap is the capillary pressure of the paper, Pj is the pressure at the channel-paper pad junction, K is the permeability of the pad, and φ is its porosity. The index, p, refers to the paper, and the triplet (Pcap, K, φ) characterizes the paper strip.35,36 The capillary pressure of the paper pad must be larger—in absolute value—than the pressure, Pj, at the last nodes so that the liquid continues to flow in the pads (Vp > 0). This condition is easily satisfied even if the capillary pressure in the tree increases after each bifurcation: Pcap,n=pnγcosθn*/Sn=αnp0γcosθ0*/S0=αnPcap,0. If we remark that αnPcap,050 Pa, and that the capillary pressure in most paper pads is of the order of 3000 Pa at least (Table SIII), more than 25 bifurcations will still produce a capillary pressure inferior to that of the pad—considering α = 0.85. The derivation of the flow motion in the pads (coupled to the tree) is detailed in the supplementary material, Sec. 2. Note that three assumptions are used in the model. First, the paper pads are homogeneous (i.e., there is no region of higher or lower porosity). Hence, saturation is neglected, and the sharp front assumption is used.37–39 This observation is confirmed by experiments.23,37

FIG. 2.

FIG. 2.

Still images of an open-channel device filled with yellow and blue nonanol solutions. (a) Yellow liquid is pipetted in the inlet of the open channel. (b) Progression of the yellow liquid in the root channel and capillary tree. (c) Blue liquid is pipetted in the inlet after the yellow liquid reaches the paper pad. (d) Progression of blue liquid in the open channel. Scale bar is 1 cm. Multimedia available online. Download video file (975.9KB, mp4) DOI: 10.1063/5.0157801.1

Second, it is assumed that the dilatation of the paper fibers from the liquid is negligible; therefore, the porosity, φ, is constant everywhere in the pad. Third, the cellulose fibers do not absorb the wicking liquid, and, thus, the mass conservation of the flowing liquid is independent of time.

Equation (3) can be solved using the expression of the pressure, Pj, found in the first phase (flow in the tree) and assuming a circular or flat contact line in the conical (angle β) or rectangular pads (β = 0). The travel distance in the paper is then

zp=an1+anδSp,0S01+1+S0Sp,02b1+anδτan2, (4)

where an=2nΣnKp0L0λS0, δ=S02βhp, and b=2KμϕPcap. Here, p0 corresponds to the total perimeter of the cross section of the root channel, S0 is the cross-section surface area, hp is the thickness of the paper pad, β is the paper pad cone angle, τ is the time counted from the moment when the liquid reaches the pads, and Σn=1+L12α4L0++Ln2α4nL0. Note that if β is zero, then δ is infinite, resulting in anδ to cancel out in Eq. (4). The two first parameters, an and δ, have the dimension of length, while the unit for b is mm2/s, and Σn is dimensionless. The S0Sp,0 is the ratio between the cross-sectional area of the root channel and that of the paper pad (at the junction with the tree). Using the mass conservation equation, the velocity in the root channel when the liquid wicks the paper is given by

Vroot=2nVpϕSpS0=ϕSp,0S0zpδS0Sp,0b2aΣn1+S0Sp,02b1+anδτan2 (5)

where zp is given by (4).

B. Comparison with experiments

The travel distance produced by Eq. (S12) (supplementary material Sec. 2) has been checked against the experiments using colored solutions of 50% IPA (v/v), pentanol, and nonanol (Fig. 3) flowing in the homothetic tree of ratio α = 0.85. A representative trial for each fluid is shown in Fig. 3. Raw data for the travel distance for each fluid are presented in Fig. SII. The travel distances in the tree, as measured by the progression of the fluid front, are well matched by the theory. The theory assumes a perfect flow with a circular fluid front in the semi-circular paper pad and perfect synchronization between the pads; however, in our experiments, the fluid front deviates from the theory due to a not-perfectly circular liquid front in the paper pad, imperfect synchronization between the flows in the different branches, and occasional leaking in the space below the pads. Nonetheless, this does not preclude the data. Note that a velocity jump is predicted by the model at the tree/pad junction due to the sudden change of capillary pressure. However, this velocity jump is not high due to the friction along the whole capillary tree. For this reason and because of the connection between channel and pads, this jump is hardly seen in the experiments.

In the case of 50% IPA, the coefficient, γμ2λcosθ*, was found to be approximately 40 mm/s1/2 ( Table SI). In the paper pads, a good fit was found for a capillary pressure of 3000 Pa. Root channel velocities on the order of 7 mm/s were obtained for 60 s (Fig. 3) for 50% IPA. The contact of the liquid with the pads occurs at 35 s for both 50% IPA and pentanol, and 95 s for nonanol (Fig. 3). Raw velocity data plots are presented in Fig. SIII.

For the case of pentanol, which is a fluid with an “intermediary” viscosity of 3.75  mPa s, the pentanol also has a good fit for the travel distances in the capillary tree, which is obtained for the value γμ2λcosθ* = 40 mm/s1/2. Root channel velocities on the order of 6.5 mm/s were obtained for at least 55 s (Fig. 3).

The case of nonanol is of great interest due to its high viscosity of 0.011 Pa s. A very good fit for the travel distances in the capillary tree is obtained for the value γμ2λcosθ* = 23.7 mm/s1/2, and a capillary pressure Pcap = 5500 Pa in the paper pad. Root channel velocities on the order of 3.5 mm/s were obtained for 150 s (Fig. 3).

The fluctuations that were observed in the velocity measurements (Fig. 3) are due to the discretization, dz/dt. The velocity is the derivative of the travel distance, and derivation amplifies the fluctuations. The only physical variation is at the junction between the channel and the paper due to abrupt capillary pressure changes.

C. Discussion

Efficient capillary pumping has been the subject of many investigations. In this problem, three parameters must be considered: (1) the maximum velocity of the flow, (2) the maximum flow rate, and (3) the duration of the pumping. Prior research has focused on obtaining the highest possible velocities, but typically only for a short time and a moderate volumetric flow rate. For example, Reches et al.40 have obtained interesting velocities of 2 cm/s for water in treated threads of wool, but along a length of 2 cm, corresponding to a duration time of 1 s. In Table I, a review of the literature for various methods of capillary pumping is summarized. The table indicates that it is difficult to obtain high velocities (larger than 1 mm/s) for a long duration (larger than 30 s) in capillary-based systems.

TABLE I.

Literature review of capillary pumping.

Reference Velocity Flow rate Duration Channel
Liquid (mm/s) (μl/s) (s) Characteristics
Watera n/a 0.13 15 Large array of PCB pillars spaced 50–300 μm
Waterb 15 n/a 20 Multilayer paper pads (Whatmann #1), 100–200 μm
Waterc n/a 0.07 n/a Large array of pillars spaced 50 μm–SI chips
Waterd 0.3 n/a 1000 Paper pads width 1–3 cm (rectangular or circular sectors), Millipore nitrocellulose
Watere 10 n/a 10 Multilayer paper pads paper, 175 μm, gap 36 μm (Whatmann #1)
Glycerol solutionsf 0.4 n/a <10 3D printed channels, Width 300 μm, acrylonitrile-butadiene-styrene
Parylene/Waterg 0.4 n/a <10 Channels, 10 μm, parylene
Water (Wool, Threads)h 20 n/a 1 Fiber threads length 1 cm, cotton, wool, polyester
This study
Nonanol >3.5 6.5 >150 Milled capillary trees
50% IPA 7 13.1 >60 Width 1 mm to 300 μm
Pentanol 6.5 12.1 >55 PMMA
a

Reference 41.

b

Reference 19.

c

Reference 42.

d

Reference 21.

e

Reference 43.

f

Reference 31.

g

Reference 44.

h

Reference 40.

In our study, it is assumed that the capillary tree is “symmetrical,” i.e., all branches at the same level of ramification are identical. The channels are relatively larger in size due to CNC milling constrictions; however, this theory is applicable to smaller dimensions up to 300 μm in width when employing other types of fabrication methods. High velocities are obtained with our device where the dimensions are at the upper side of the microscale limits. If we remark that the root channel velocity is approximately proportional to the square root of the friction length and that the friction length decreases proportionally with the channel dimension, a homothetical reduction of the channel dimension of a factor, n, will result in the reduction of the velocity of a factor n. For example, if the channel cross section is decreased by 4 (200 μm width), the velocity will be reduced by a factor of 2. Still, high velocities are obtained with the device for microscale channels. The progression of the liquid is the same for each path and is obtained by using the extended LWR law that states that the dynamics of the flow results from the balance between the capillary force on the advancing meniscus and the wall friction along the path.3–5,24,25 Intrinsically, capillary-driven flow velocities decrease as capillary length decreases. Here, our device enables longer capillary lengths in the root channel due to the multiplication of daughter branches while maintaining high flow velocities, allowing for applications in diagnostics and bioanalytical chemistry.24 Note that the approach proposed here is also applicable to closed systems by using the friction length of the closed channels instead of open capillary channels.24,25 In the microporous paper pads, the wicking liquid velocity is given by Darcy's law33,34 and determined by three parameters: permeability, porosity, and capillary pressure.35,36

To validate the closed form model derived in this work, experiments were performed using nonanol, isopropyl alcohol, and pentanol solutions in open microfluidic channels. We chose these liquids because they wet native PMMA without surface treatment (θ = 13°–47°, Table SI). Furthermore, evaporation of these liquids is slow compared to the timescale of the capillary flow. Hence, evaporation is not taken into account for this study.3,45–47 High capillary velocities are obtained even in the case of highly viscous nonanol, which has a viscosity eleven times that of water (Table SI). In the case of the less viscous IPA aqueous solution, velocities higher than 1 cm/s were obtained.

The device presented here is of interest for its ability to combine velocity, flow rate, and duration. This innovation builds on our previous work28 to further enable extended periods of high velocity flow in open microfluidic channels, which is desired for capillary microfluidics research and device development for fields such as chemical, biomedical, biological, space science, and materials research. In the future, the device can be optimized to achieve higher velocity. A longer length of the last branch of the capillary tree can enable a higher velocity in the root channel. The choice of the paper matrix is of great importance, especially the two parameters K/φ (Leverett parameter) and capillary pressure (Pcap). Longer pads would allow longer duration of the high velocity flow. Additionally, the device can be micromilled using different materials, such as polystyrene, which enables wider application.48 Devices can be oxygen plasma treated to allow flow of aqueous solutions (cell culture media, biological fluids, etc.).26,48

IV. CONCLUSION

In this work, the dynamics of the capillary flow circulating in a capillary tree with paper pads placed at the extremities of the capillary tree branches has been investigated. A model for the dynamics of the flow in the capillary tree has been coupled to a model for the flow in the paper pads. This coupling has been validated against experiments performed with milled open PMMA channels and paper pads. It is first shown that capillary trees with homothetically decreasing cross-sectional areas (in a ratio of 0.85) maintain the flow velocity in the root channel. Moreover, the presence of paper pads at the extremities of the branches prolongs the duration of the high flow rate pumping. The present analysis demonstrates the possibility of obtaining high velocities and flow rates (7 mm/s or 13.1 μl/s) for more than 30 s for 50% (%v/v) IPA, 6.5 mm/s or 12.1 μl/s for more than 55 s with pentanol, and flow rates of >3.5 mm/s or 6.5 μl/s for more than 150 s for nonanol. These flow rates are nearly constant (save for periodic jumps due to experimental fluctuations) if conical-shaped paper pads are used as suggested in the literature.49,50 For volatile fluids, the employment of evaporation could extend the duration of these high flow rates. Furthermore, we envision many areas of future application including using our method to push the limits of viscous fluid flow in open channels as we can enable the sustained passive flow of complex biological fluids such as saliva or blood. This device also has the potential to be applied to biological experiments such as in vitro cell culture or analytical methods involving biological fluids. Furthermore, the method presented here can be adapted for diverse applications of open microfluidics, including chemical synthesis, chemical analysis, thermics, materials science, or space science, allowing for diverse applications at the frontiers of science.

SUPPLEMENTARY MATERIAL

See the supplementary material for the open capillary padding design file, and additional information on physical properties of solvents and paper pads, channel dimensions, theoretical derivations, and engineering drawing of microfluidic device.

ACKNOWLEDGMENTS

This work was funded by NIH Grant No. R35GM128648. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. We also acknowledge the M.J. Murdock Foundry for Translational Research. We would also like to thank David N. Phan for writing the code for video extraction.

Contributor Information

Ashleigh B. Theberge, Email: mailto:abt1@uw.edu.

Erwin Berthier, Email: mailto:erwin.berthier@gmail.com.

AUTHOR DECLARATIONS

Conflict of Interest

The authors acknowledge the following potential conflicts of interest in companies pursuing open microfluidic/analytic technologies: E.B.: Tasso, Inc., Salus Discovery, LLC, and Stacks to the Future, LLC; A.B.T.: Stacks to the Future, LLC. The work in this manuscript is not related to these companies.

Author Contributions

Jodie C. Tokihiro, Wan-chen Tu, and Jean Berthier contributed equally to this work.

Jodie C. Tokihiro: Data curation (equal); Formal analysis (equal); Investigation (equal); Methodology (equal); Project administration (equal); Validation (equal); Visualization (equal); Writing – original draft (equal); Writing – review & editing (equal). Wan-chen Tu: Data curation (equal); Formal analysis (equal); Investigation (equal); Methodology (equal); Project administration (equal); Validation (equal); Visualization (equal); Writing – original draft (equal); Writing – review & editing (equal). Jean Charles Berthier: Conceptualization (equal); Data curation (equal); Formal analysis (equal); Methodology (equal); Supervision (equal); Validation (equal); Writing – original draft (equal); Writing – review & editing (equal). Jing J. Lee: Data curation (equal); Formal analysis (equal); Investigation (equal); Methodology (equal); Software (equal); Writing – original draft (equal). Ashley M. Dostie: Data curation (equal); Investigation (equal); Writing – original draft (equal). Jian Wei Khor: Conceptualization (equal); Formal analysis (equal); Writing – review & editing (equal). Madeleine Eakman: Investigation (equal); Writing – review & editing (equal). Ashleigh B. Theberge: Conceptualization (equal); Formal analysis (equal); Funding acquisition (equal); Project administration (equal); Supervision (equal); Writing – review & editing (equal). Erwin Berthier: Conceptualization (equal); Formal analysis (equal); Funding acquisition (equal); Supervision (equal); Writing – review & editing (equal).

DATA AVAILABILITY

The data that support the findings of this study are available within the article and its supplementary material.

References

  • 1. Gervais L., Hitzbleck M., and Delamarche E., “ Capillary-driven multiparametric microfluidic chips for one-step immunoassays,” Biosens. Bioelectron. 27(1), 64 (2011). 10.1016/j.bios.2011.06.016 [DOI] [PubMed] [Google Scholar]
  • 2. Safavieh R. and Juncker D., “ Capillarics: Pre-programmed, self-powered microfluidic circuits built from capillary elements,” Lab Chip 13(21), 4180 (2013). 10.1039/c3lc50691f [DOI] [PubMed] [Google Scholar]
  • 3. Lucas R., “ Ueber das Zeitgesetz des kapillaren Aufstiegs von Flüssigkeiten,” Kolloid-Z. 23, 15 (1918). 10.1007/BF01461107 [DOI] [Google Scholar]
  • 4. Washburn E. W., “ The dynamics of capillary flow,” Phys. Rev. 17, 273 (1921). 10.1103/PhysRev.17.273 [DOI] [Google Scholar]
  • 5. Rideal E. K., “ CVIII. On the flow of liquids under capillary pressure,” London, Edinburgh Dublin Philos. Mag. J. Sci. 44(264), 1152 (1922). 10.1080/14786441008634082 [DOI] [Google Scholar]
  • 6. Juncker D., Schmid H., Drechsler U., Wolf H., Wolf M., Michel B., de Rooij N., and Delamarche E., “ Autonomous microfluidic capillary system,” Anal. Chem. 74(24), 6139 (2002). 10.1021/ac0261449 [DOI] [PubMed] [Google Scholar]
  • 7. Cesaro-Tadic S., Dernick G., Juncker D., Buurman G., Kropshofer H., Michel B., Fattinger C., and Delamarche E., “ High-sensitivity miniaturized immunoassays for tumor necrosis factor? Using microfluidic systems,” Lab Chip 4(1), 563 (2004). 10.1039/b408964b [DOI] [PubMed] [Google Scholar]
  • 8. Zimmermann M., Schmid H., Hunziker P., and Delamarche E., “ Capillary pumps for autonomous capillary systems,” Lab Chip 7(1), 119 (2007). 10.1039/B609813D [DOI] [PubMed] [Google Scholar]
  • 9. Vestad T., Marr D. W. M., and Oakey J., “ Flow control for capillary-pumped microfluidic systems,” J. Micromech. Microeng. 14, 1503 (2004). 10.1088/0960-1317/14/11/010 [DOI] [Google Scholar]
  • 10. Satoh W., Hosono H., and Suzuki H., “ On-chip microfluidic transport and mixing using electrowetting and incorporation of sensing functions,” Anal. Chem. 77(21), 6857 (2005). 10.1021/ac050821s [DOI] [PubMed] [Google Scholar]
  • 11. Kolliopoulos P., Jochem K. S., Francis L. F., and Kumar S., “ Capillary flow of evaporating liquid solutions in open rectangular microchannels,” J. Fluid Mech. 938, A22 (2022). 10.1017/jfm.2022.140 [DOI] [Google Scholar]
  • 12. Lynn N. S. and Dandy D. S., “ Passive microfluidic pumping using coupled capillary/evaporation effects,” Lab Chip 9, 3422 (2009). 10.1039/b912213c [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Zimmermann M., Bentley S., Schmid H., Hunziker P., and Delamarche E., “ Continuous flow in open microfluidics using controlled evaporation,” Lab Chip 5(12), 1355 (2005). 10.1039/b510044e [DOI] [PubMed] [Google Scholar]
  • 14. Srinivasan R., “ Estimating zero-g flow rates in open channels having capillary pumping vanes,” Int. J. Numer. Methods Fluids 41(4), 389 (2003). 10.1002/fld.446 [DOI] [Google Scholar]
  • 15. Guo W., Hansson J., and van der Wijngaart W., “ Capillary pumping independent of the liquid surface energy and viscosity,” Microsyst. Nanoeng. 4, 2 (2018). 10.1038/s41378-018-0002-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Olanrewaju A., Beaugrand M., Yafia M., and Juncker D., “ Capillary microfluidics in microchannels: From microfluidic networks to capillaric circuits,” Lab Chip 18(16), 2323 (2018). 10.1039/C8LC00458G [DOI] [PubMed] [Google Scholar]
  • 17. Pla-Roca M. and Juncker D., “ PDMS microfluidic capillary systems for patterning proteins on surfaces and performing miniaturized immunoassays,” Methods Mol. Biol. 671, 177 (2011). 10.1007/978-1-59745-551-0 [DOI] [PubMed] [Google Scholar]
  • 18. Delamarche E., Juncker D., and Schmid H., “ Microfluidics for processing surfaces and miniaturizing biological assays,” Adv. Mater. 17, 2911–2933 (2005). 10.1002/adma.200501129 [DOI] [Google Scholar]
  • 19. Channon R. B., Nguyen M. P., Scorzelli A. G., Henry E. M., Volckens J., Dandy D. S., and Henry C. S., “ Rapid flow in multilayer microfluidic paper-based analytical devices,” Lab Chip 18(5), 793 (2018). 10.1039/C7LC01300K [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Channon R. B., Nguyen M. P., Charles S. H., and Dandy D. S., “ Multilayered microfluidic paper-based devices: Characterization, modeling, and perspectives,” Anal. Chem. 91(14), 8966 (2019). 10.1021/acs.analchem.9b01112 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Mendez S., Fenton E. M., Gallegos G. R., Petsev D. N., Sibbett S. S., Stone H. A., Zhang Y., and López G. P., “ Imbibition in porous membranes of complex shape: Quasi-stationary flow in thin rectangular segments,” Langmuir 26(2), 1380 (2010). 10.1021/la902470b [DOI] [PubMed] [Google Scholar]
  • 22. Wang X., Zhou J., and Papautsky I., “ Vortex-aided inertial microfluidic device for continuous particle separation with high size-selectivity, efficiency, and purity,” Biomicrofluidics 7(4) 044119 (2013). 10.1063/1.4818906 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Kokalj T., Park Y., Vencelj M., Jenko M., and Lee L. P., “ Self-powered imbibing microfluidic pump by liquid encapsulation: SIMPLE,” Lab Chip 14(22), 4329 (2014). 10.1039/C4LC00920G [DOI] [PubMed] [Google Scholar]
  • 24. Berthier J., Brakke K. A., and Berthier E., Open Microfluidics (Wiley Publishers, 2016).
  • 25. Berthier J., Berthier E., and Theberge A. B., Open-Channel Microfluidics: Fundamentals and Applications ( Morgan and Claypool Publishers, 2019). [Google Scholar]
  • 26. Berthier E., Dostie A. M., Lee U. N., Berthier J., and Theberge A. B., “ Open microfluidic capillary systems,” Anal. Chem. 91(14), 8739 (2019). 10.1021/acs.analchem.9b01429 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Lee J. J., Berthier J., Kearney K. E., Berthier E., and Theberge A. B., “ Open-channel capillary trees and capillary pumping,” Langmuir 36(43), 12795 (2020). 10.1021/acs.langmuir.0c01360 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Lee J. J., Berthier J., Theberge A. B., and Berthier E., “ Capillary flow in open microgrooves: Bifurcations and networks,” Langmuir 35(32), 10667 (2019). 10.1021/acs.langmuir.9b01456 [DOI] [PubMed] [Google Scholar]
  • 29. Berthier J., Brakke K. A., Gosselin D., Navarro F., Belgacem N., and Chaussy D., “ Spontaneous capillary flow in curved, open microchannels,” Microfluid. Nanofluidics 20, 100 (2016). 10.1007/s10404-016-1766-6 [DOI] [Google Scholar]
  • 30. Berthier J., Brakke K. A., Gosselin D., Huet M., and Berthier E., “ Metastable capillary filaments in rectangular cross-section open microchannels,” AIMS Biophys. 1(1), 31 (2014). 10.3934/biophy.2014.1.31 [DOI] [Google Scholar]
  • 31. Lade R. K., Hippchen E. J., Macosko C. W., and Francis L. F., “ Dynamics of capillary-driven flow in 3D printed open microchannels,” Langmuir 33(12), 2949 (2017). 10.1021/acs.langmuir.6b04506 [DOI] [PubMed] [Google Scholar]
  • 32. Berthier J., Gosselin D., and Berthier E., “ A generalization of the Lucas–Washburn–Rideal law to composite microchannels of arbitrary cross section,” Microfluid. Nanofluid. 19, 497 (2015). 10.1007/s10404-014-1519-3 [DOI] [Google Scholar]
  • 33. Darcy H., Les Fontaines Publiques de La Ville de Dijon ( Paris, 1856). [Google Scholar]
  • 34. Whitaker S., “ Flow in porous media I: A theoretical derivation of Darcy's law,” Transp. Porous Media 1, 3 (1986). 10.1007/BF01036523 [DOI] [Google Scholar]
  • 35. Amico S. C. and Lekakou C., “ Axial impregnation of a fiber bundle. Part 1: Capillary experiments,” Polym. Compos. 23(2), 249 (2002). 10.1002/pc.10429 [DOI] [Google Scholar]
  • 36. Ashari A. and Vahedi Tafreshi H., “ General capillary pressure and relative permeability expressions for through-plane fluid transport in thin fibrous sheets,” Colloids Surf., A 346(1–3), 114 (2009). 10.1016/j.colsurfa.2009.06.001 [DOI] [Google Scholar]
  • 37. Dane J. H., Hofstee C., and Corey A. T., “ Simultaneous measurement of capillary pressure, saturation, and effective permeability of immiscible liquids in porous media,” Water Resour. Res. 34(12), 3687, 10.1029/1998WR900026 (1998). [DOI] [Google Scholar]
  • 38. Babchin A. J., Bentsen R., Faybishenko B., and Geilikman M. B., “ On the capillary pressure function in porous media based on relative permeabilities of two immiscible fluids: Application of capillary bundle models and validation using experimental data,” Adv. Colloid Interface Sci 233, 176 (2016). 10.1016/j.cis.2015.07.001 [DOI] [PubMed] [Google Scholar]
  • 39. Zarandi M. A. F., Pillai K. M., and Kimmel A. S., “ Spontaneous imbibition of liquids in glass-fiber wicks. Part I: Usefulness of a sharp-front approach,” AIChE J. 64(1), 294 (2018). 10.1002/aic.15965 [DOI] [Google Scholar]
  • 40. Reches M., Mirica K. A., Dasgupta R., Dickey M. D., Butte M. J., and Whitesides G. M., “ Thread as a matrix for biomedical assays,” ACS Appl. Mater. Interfaces 2(6), 1722 (2010). 10.1021/am1002266 [DOI] [PubMed] [Google Scholar]
  • 41. Vasilakis N., Papadimitriou K. I., Morgan H., and Prodromakis T., “ High-performance PCB-based capillary pumps for affordable point-of-care diagnostics,” Microfluid. Nanofluid. 21(6), 103 (2017). 10.1007/s10404-017-1935-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Safavieh R., Tamayol A., and Juncker D., “ Serpentine and leading-edge capillary pumps for microfluidic capillary systems,” Microfluid. Nanofluid. 18, 357 (2015). 10.1007/s10404-014-1454-3 [DOI] [Google Scholar]
  • 43. Schaumburg F. and Berli C. L. A., “ Assessing the rapid flow in multilayer paper-based microfluidic devices,” Microfluid. Nanofluid. 23(98), 98 (2019). 10.1007/s10404-019-2265-3 [DOI] [Google Scholar]
  • 44. Yang L.-J., Yao T.-J., and Tai Y.-C., “ The marching velocity of the capillary meniscus in a microchannel,” J. Micromech. Microeng. 14, 220 (2004). 10.1088/0960-1317/14/2/008 [DOI] [Google Scholar]
  • 45. Goedecke N., Eijkel J., and Manz A., “ Evaporation driven pumping for chromatography application,” Lab Chip 2(4), 219 (2002). 10.1039/b208031c [DOI] [PubMed] [Google Scholar]
  • 46. Nie C., Frijns A. J. H., Mandamparambil R., and den Toonder J. M. J., “ A microfluidic device based on an evaporation-driven micropump,” Biomed. Microdevices 17, 47 (2015). 10.1007/s10544-015-9948-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Liu H., Zhang X., Hong Z., Pu Z., Yao Q., Shi J., Yang G., Mi B., Yang B., Liu X., Jiang H., and Hu X., “ A bioinspired capillary-driven pump for solar vapor generation,” Nano Energy 42, 115 (2017). 10.1016/j.nanoen.2017.10.039 [DOI] [Google Scholar]
  • 48. Berry S. B., Lee J. J., Berthier J., Berthier E., and Theberge A. B., “ Droplet incubation and splitting in open microfluidic channels,” Anal. Methods 11(35), 4528 (2019). 10.1039/C9AY00758J [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49. Dosso F. D., Tripodi L., Spasic D., Kokalj T., and Lammertyn J., “ Innovative hydrophobic valve allows complex liquid manipulations in a self-powered channel-based microfluidic device,” ACS Sensors 4(3), 694 (2019). 10.1021/acssensors.8b01555 [DOI] [PubMed] [Google Scholar]
  • 50. Elizalde E., Urteaga R., and Berli C. L. A., “ Rational design of capillary-driven flows for paper-based microfluidics,” Lab Chip 15(10), 2173 (2015). 10.1039/C4LC01487A [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

See the supplementary material for the open capillary padding design file, and additional information on physical properties of solvents and paper pads, channel dimensions, theoretical derivations, and engineering drawing of microfluidic device.

Data Availability Statement

The data that support the findings of this study are available within the article and its supplementary material.


Articles from Physics of Fluids are provided here courtesy of American Institute of Physics

RESOURCES