Development and validation of an expanded panel of progestins using liquid chromatography-tandem triple quadrupole mass spectrometry (LC-MS/MS) to monitor protocol compliance in hormonal contraceptive pharmacokinetic/pharmacodynamic studies

Jeffrey T Jensen a, Bethany T Waites a,1, Emily R Boniface a, Sara McCrimmon a, Steven Blue b, David W Erikson b
PMCID: PMC10528587  NIHMSID: NIHMS1921728  PMID: 37499736

Abstract

We developed and validated use of ultra-high performance liquid chromatography-heated electrospray ionization-tandem triple quadrupole mass spectrometry (LC-MS/MS) to simultaneously analyze serum concentrations of ethinylestradiol (EE), dienogest (DNG), norelgestromin (NGMN), norethindrone (NET), gestodene (GSD), levonorgestrel (LNG), etonogestrel (ENG), segesterone acetate (NES), medroxyprogesterone acetate (MPA), and drospirenone (DRSP). The calibration range for all targets was 0.009 to 10 ng/ml, with lower limit of quantification of 0.009 ng/ml for all analytes except GSD (0.019 ng/ml). We used our assay to check compliance among participants in a clinical trial, confirmed use of DRSP in 11/13 study participants, and evidence of noncompliant progestins in 2 (LNG=1, NET =1). We conclude that this approach provides an accurate method to check protocol compliance in contraceptive clinical trials.

Keywords: progestins, assays, compliance, liquid chromatography

1. Introduction

Randomized trials control for unmeasured baseline confounding and provide the highest quality clinical evidence [1]. However, in small studies, randomization may not result in an even distribution of known confounders, let alone unmeasured behavioral factors. Several investigators have demonstrated nonadherence to treatment regimens as a factor influencing outcomes in contraceptive clinical trials [2, 3], and epidemiological studies evaluating effects of contraceptive methods on other health outcomes have also demonstrated participant non-compliance [4, 5]. Indeed, differences in compliance likely explain discrepancies in the Pearl Indexes between European and United States studies of the same oral contraceptive formulations.

Liquid chromatography-tandem triple quadrupole mass spectrometry (LC-MS/MS) provides a platform to simultaneously evaluate serum samples for a panel of natural and contraceptive steroids [6, 7]. Recently, investigators applying this approach documented that 17% of samples from women reporting injectable use, 62% of samples from self-reported OC users, and 8% of samples from self-reported implant users had no quantifiable hormones [5].

Here, we report the development and use of an expanded LC-MS/MS platform that includes all progestins marketed in the United States.

2. Materials and Methods

We previously reported details of methods used by the Endocrine Technologies Core (ETC) at the Oregon National Primate Research Center (ONPRC, Beaverton, OR) to develop an assay for simultaneous analysis of a panel of contraceptive and ovarian steroids [6]. We modified this approach to establish a new assay for simultaneous measurement of estradiol (E2); progesterone (P4), ethinylestradiol (EE), dienogest (DNG), norelgestromin (NGMN), norethindrone (NET), gestodene (GSD), levonorgestrel (LNG), etonogestrel (ENG), segesterone acetate (NES), medroxyprogesterone acetate (MPA), and drospirenone (DRSP) using ultra-high performance liquid chromatography-heated electrospray ionization-tandem triple quadrupole mass spectrometry (LC-MS/MS) on a Shimadzu Nexera-LCMS-8050 instrument (Kyoto, Japan). We used LabSolutions Software, V5.72 (Shimadzu) for all of the processing and analysis. Briefly, 200 μl of serum were mixed with 100 μl of 10:90 (v:v) LC-MS grade methanol (Honeywell Burdick & Jackson, St Louis, MO, USA): LC-MS grade water (Honeywell Burdick & Jackson, St Louis, MO, USA) containing a mix of synthetic [EE-d7, NGMN-d6, NET-d6, GSD-d6, LNG-d6, ENG-d7, NES-C3, MPA-d6, and DRSP-C3 (Toronto Research Chemicals, Toronto, CA)] and natural [estradiol (E2-d5, Cerilliant, Round Rock, Texas), progesterone (P4-C3, IsoSciences Ambler, PA, USA] isotopic standards and added to a 400 μl SLE+ extraction plate (Biotage, Uppsala, Sweden). Isotopic standard concentrations were determined so as to yield 1,000,000 – 2,000,000 peak area for positive mode targets (P4, DNG, LNG, NET, GSD, NGMN, ENG, NES, DRSP, and MPA) and 100,000 – 200,000 peak area for negative mode targets (E2, EE). Steroids were eluted with 3 x 600 μl dichloromethane (Sigma, St. Louis, MO) into 2 mL 96-well round bottom polypropylene plates (Analytical Sales & Services, Flanders, NJ, USA) containing 100 μL of 2-propanol (Sigma-Aldrich, St. Louis, MO, USA), dried with forced air, and reconstituted in 50 μl of 25% (v:v) methanol:water. For calibration curves, we spiked charcoal-stripped human serum (Golden West Biologicals) with unlabeled P4 (IsoSciences Ambler, PA, USA), E2 and EE (Cerilliant, Round Rock, TX), and DNG, NGMN, NET, GSD, LNG, ENG, NES, MPA, and DRSP standards (Toronto Research Chemicals, Toronto, CA) in methanol, diluted serially to final concentrations between 0.009 and 10 ng/ml in a 12-point curve including a 0 ng/ml blank, subjected 200 μl of standard to the SLE+ extraction procedure as described above, and then used the Shimadzu SIL-30CAMP autosampler to inject 15 μl samples onto a Raptor 2.7 μm Biphenyl 50 mm X 2.1 mm column with a matching 5 mm x 2.1 mm guard column (Restek, Bellefonte, PA). Mobile phase consisted of 0.15 mM ammonium fluoride (Sigma) in water (A), and methanol (B) with a flow rate of 0.5 ml/min. Using a Shimadzu Nexera LC-30AD system (LC), the gradient elution started at 60% B, increased (0.50-3.50 min) to 65% B (held 3.50-4.25 min), further increased (4.25-7.25 min) to 95% B and then (7.25-7.35 min) where it was held at 100% B (7.35-8.05 min) before being returned (8.05-8.15 min) to 60% B and held at 60% for column re-equilibration for a total of 9.55 minutes/sample. E2 and EE were detected in negative ion mode and all other targets were detected in positive ion mode, all with multiple reaction monitoring (MRM) using a Shimadzu LCMS-8050 tandem triple-quadrupole MS with heated electrospray ionization (ESI). More details regarding method validation and assay performance are included in the Supplemental Methods.

3. Results

We achieved a calibration range for all targets of 0.009 to 10 ng/ml, with lower limit of quantification (LLOQ) of 0.009 ng/ml for E2, P4, EE, MPA, LNG, ENG, NET, NGMN, NET, DNG, and DRSP and 0.019 ng/ml for GSD. Table 1 provides all data relating to method validation including intra-and inter-assay variability, detection limits, extraction efficiency, and matrix effects. Supplemental Table 1 provides the MRM transitions and retention times for each target. The working solution concentrations used for Internal Standards are provided in Supplemental Table 2. We provide a representative chromatogram as Supplemental Figure 1

Table 1.

Method validation parameters acquired during development and validation of an expanded panel of progestins using liquid chromatography-tandem triple quadrupole mass spectrometry (LC-MS/MS).

Hormone QC Target Concentration (ng/ml) Measured Concentration (ng/ml) Accuracy (%) Intra-Assay Precision (%CV) Inter-Assay Precision (%CV) Extraction Efficiency (Recovery) Matrix Effects LLOQ (ng/ml) ULOQ (ng/ml) MDL (ng/ml)
E2 Low 0.09 0.089 98.9 4.9 6.7 93.9% 100.3% 0.009 10 0.007
High 0.28 0.267 95.4 3.8 4.2
P4 Low 0.25 0.298 119.2 2.5 4.9 88.9% 83.5% 0.009 10 0.003
High 1.85 2.083 112.6 2.2 3.4
EE Low 0.07 0.071 101.4 4.2 5.2 99.7% 99.2% 0.009 10 0.003
High 0.23 0.238 103.5 2.6 2.9
NMGN Low 0.08 0.086 107.5 4.6 6.1 82.0% 87.8% 0.009 10 0.002
High 0.40 0.418 104.5 3.2 4.7
NET Low 0.06 0.062 103.3 3.4 5.8 91.7% 90.2% 0.009 10 0.003
High 0.44 0.399 90.7 2.5 3.6
GSD Low 0.05 0.049 98.0 6.4 10.6 97.8% 99.2% 0.019 10 0.013
High 0.30 0.313 104.3 2.2 8.3
LNG Low 1.60 1.658 103.6 2.3 4.4 98.9% 82.8% 0.009 10 0.002
High 3.80 3.950 103.9 1.0 4.4
ENG Low 0.08 0.093 116.3 3.6 8.3 96.4% 81.1% 0.009 10 0.008
High 0.53 0.605 114.2 1.4 5.3
NES Low 0.06 0.064 106.7 1.8 4.3 83.5% 93.2% 0.009 10 0.003
High 0.48 0.485 101.0 1.2 4.6
MPA Low 0.06 0.067 111.7 5.2 6.3 70.3% 89.6% 0.009 10 0.002
High 1.00 1.092 109.2 3.1 3.8
DRSP Low 0.09 0.097 107.8 2.4 5.1 86.0% 107.1% 0.009 10 0.002
High 0.44 0.460 104.5 0.5 4.4
DNG Low 0.09 0.083 92.2 6.1 7.3 94.9% 94.0% 0.009 10 0.002
High 0.44 0.350 80.0 1.8 5.0

QC = quality control; LLOQ = lower limit of quantification; ULOQ = upper limit of quantification; MDL = method detection limit; E2 = estradiol; P4 = progesterone; EE = ethinyl estradiol; NGMN = norelgestromin; DNG = dienogest; NET = norethindrone; GSD = gestodene; LNG = levonorgestrel; ENG = etonogestrel; NES = segesterone acetate; MPA = medroxyprogesterone acetate (MPA); DRSP = drospirenone.

We used the assay to evaluate samples obtained from a randomized controlled double-blind pilot study (Clinical Trials #NCT03418363) designed to evaluate the effects of an oral supplement on hepatic protein levels in current users of an EE/DRSP combined oral contraceptive (COC) for compliance using the study pill. The institutional review board at Oregon Health and Science University approved this study. Among the 13 participants who completed the study, 11 showed detectable concentrations of DRSP (Table 2). We did not identify any samples above the method detection limit, but below the lower limit of quantification. We found evidence of protocol non-compliance in two participants (NET and LNG one each).

Table 2.

Concentrations (ng/mL) of progestins detected using the LCMS/MS panel in serum from participants with reported use of an ethinyl estradiol/drospirenone combined pill.

Subject Drospirenone Levonorgestrel Norethindrone
1 25.27 ND ND
2 32.44 ND ND
3 17.57 ND ND
4 18.57 ND ND
5 28.38 ND ND
6 ND ND 3.30
7 13.01 ND ND
8 10.33 ND ND
9 23.40 ND ND
10 14.36 ND ND
11 36.21 ND ND
12 33.19 ND ND
13 ND 2.40 ND

ND = not detected

4. Discussion

We report the first validated serum assay for simultaneous evaluation of all synthetic contraceptive hormones in use in the United States. Of note, the assay includes DRSP and the recently approved DNG and NES. As the assay also includes GSD, a progestin commonly used in the United Kingdom and other countries, it provides a comprehensive platform for the evaluation of serum samples for contraceptive steroid use in clinical trials to assess protocol compliance or non-compliance. The assay has high sensitivity and reproducibility.

This assay allowed us to detect noncompliance with reported use of a DRSP OC. Two of the 13 samples from this clinical trial did not contain DRSP, and the assay detected LNG and NET (15% noncompliance).

A major strength of our assay is its high sensitivity to detect picogram quantities of all contraceptive progestins in clinical use, along with natural ovarian hormones, in a single test. However, as the minimum concentrations of a progestin required to exert a contraceptive effect (e.g. inhibition of ovulation) has not been rigorously determined for most ligands, determining whether the limits of our assay have clinical relevance will require additional pharmacokinetic and pharmacodynamic testing.

Noncompliance with study protocols decreases study validity, particularly in small pharmacokinetic and pharmacodynamic studies that rely on subjects’ use of a specific compound. Researchers and clinicians should take this into consideration when interpreting study results, and consider use of an objective measure to verify protocol compliance.

Supplementary Material

1
2
3

Implications:

The availability of a LC-MS/MS multi-progestin analysis panel for simultaneous evaluation of the most common contraceptive steroids approved worldwide could improve monitoring of compliance and protocol adherence in clinical trials.

Acknowledgements:

The authors would like to acknowledge OHSU’s Women’s Health Research Unit for the work and support given to this project. Thank you to Marci Messerle Forbes, Andrea O’Donnell, and Kise Bond for their help recruiting and enrolling patients.

Funding details:

Financial support: This project was funded by an Emerging Scholars in Family Planning grant (project number SFPRF12-ES22), awarded by the Society of Family Planning. The Endocrine Technologies Core at Oregon National Primate Research Center (ONPRC) is supported in part by NIH P51 OD011092 awarded to ONPRC.

Disclosure statement:

Dr. Jensen has received payments for consulting from Bayer Healthcare, Evofem, Hope Medicine, Foundation Consumer Healthcare, Mayne Pharma, Myovant, ViiV Healthcare, and TherapeuticsMD. OHSU has received research support from Abbvie, Bayer Healthcare, Daré, Estetra SPRL, Hope Medicine, Organon, Medicines360, Merck, Myovent, and Sebela. These companies and organizations may have a commercial or financial interest in the results of this research and technology. These potential conflicts of interest have been reviewed and managed by OHSU. None of these have direct conflict with this manuscript. No other authors have any potential conflicts of interest.

Footnotes

Clinical trial registration number: NCT03418363

References

  • [1].Koretz RL. Assessing the Evidence in Evidence-Based Medicine. Nutrition in clinical practice : official publication of the American Society for Parenteral and Enteral Nutrition. 2019;34:60–72. [DOI] [PubMed] [Google Scholar]
  • [2].Kaunitz AM, Portman D, Westhoff CL, Archer DF, Mishell DR Jr., Foegh M. Self-reported and verified compliance in a phase 3 clinical trial of a novel low-dose contraceptive patch and pill. Contraception. 2015;91:204–10. [DOI] [PubMed] [Google Scholar]
  • [3].Westhoff CL, Torgal AT, Mayeda ER, Shimoni N, Stanczyk FZ, Pike MC. Predictors of noncompliance in an oral contraceptive clinical trial. Contraception. 2012;85:465–9. [DOI] [PubMed] [Google Scholar]
  • [4].Achilles SL, Mhlanga FG, Musara P, Poloyac SM, Chirenje ZM, Hillier SL. Misreporting of contraceptive hormone use in clinical research participants. Contraception. 2018;97:346–53. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Pyra M, Lingappa JR, Heffron R, et al. Concordance of self-reported hormonal contraceptive use and presence of exogenous hormones in serum among African women. Contraception. 2018;97:357–62. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Blue SW, Winchell AJ, Kaucher AV, et al. Simultaneous quantitation of multiple contraceptive hormones in human serum by LC-MS/MS. Contraception. 2018;97:363–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Zhang J, Tang C, Oberly PJ, Minnigh MB, Achilles SL, Poloyac SM. A sensitive and robust UPLC-MS/MS method for quantitation of estrogens and progestogens in human serum. Contraception. 2019;99:244–50. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1
2
3

RESOURCES